Many assistive products have no product-specific ISO or national standard. When such a product is being registered, placed on the EU market, or questioned by a buyer who wants a test report, the first question is which standard to test to. ISO 21856:2022, Assistive products — General requirements and test methods, was written for this situation. This guide covers what the standard applies to, how it relates to product standards, what its clauses deal with, how to choose the clauses that apply, and what the SUNGO Hefei laboratory's accreditation covers and excludes.
Key points
- ISO 21856:2022 was prepared by ISO/TC 173. It is the first edition, published in July 2022, and replaces ISO 16201:2006. The European adoption, EN ISO 21856:2022, supersedes EN ISO 16201:2006 and EN 12182:2012.
- It applies to assistive products that are considered medical devices and are intended to alleviate or compensate for a disability, but not to products that work by administering pharmaceutical substances to the user.
- It does not replace product-specific standards. The Introduction advises users to check first for a more relevant standard; ISO 21856 can then supplement a product standard that does not cover one of its requirements.
- At SUNGO, ISO 21856 and EN ISO 21856 are listed in the Hefei laboratory's IAS accreditation scope (TL-1119), with clauses 4.3, 5.2, 5.3, 8.3, 11 and 17.4 excluded, and reports are issued under that IAS accreditation. The IAS scope states no edition year; please confirm the edition with the laboratory before ordering.
- Biological tests under clause 5.3, Biocompatibility and toxicity (cytotoxicity, sensitization, irritation, implantation and similar), are subcontracted by SUNGO to a laboratory with the relevant competence. In line with accreditation rules, the client's agreement is obtained before subcontracting, and the report identifies the subcontracted items and the subcontractor. The other excluded clauses must be arranged separately.
The standard and its scope
Edition and origin
As of 2026-10-22, the ISO website lists ISO 21856:2022 as Edition 1, published 2022-07, status Published, ISO/TC 173 (Assistive products), ICS 11.180.01. The Foreword says this first edition cancels and replaces ISO 16201:2006, with the scope changed to assistive products in general. The Introduction says the document is based on EN 12182:2012.
EN ISO 21856:2022, dated July 2022, was prepared by ISO/TC 173 with CEN/TC 293, Assistive products and accessibility. Its European foreword says it supersedes EN ISO 16201:2006 and EN 12182:2012 and was to be given national-standard status by January 2023 at the latest. The endorsement notice says CEN approved the text of ISO 21856:2022 as EN ISO 21856:2022 without modification. Whether EN ISO 21856 has been cited as a harmonised standard under the EU MDR should be checked against the current list in the Official Journal of the EU; this guide takes no position on it.
Scope
Clause 1 covers general requirements and test methods for assistive products considered to be medical devices and intended to alleviate or compensate for a disability, excluding products that achieve their purpose by administering pharmaceutical substances. The notes add that assistive products are medical devices in some jurisdictions but not in others, and that other International Standards cover particular product types.
On ISO 9999, one sentence is enough: Note 3 to the scope says not every item in ISO 9999 is a medical device, so using the standard or some of its clauses for non-medical assistive products is left to the parties concerned, and clause 3.3 notes that its definition of "assistive product" differs from ISO 9999 because it is restricted to medical devices.
How it relates to product-specific standards
The Introduction says the standard provides safety requirements and recommendations for assistive products not covered by another International Standard. Users should first check for a more relevant standard; where a product standard does not cover a requirement in ISO 21856, ISO 21856 can be used as a supplement, and it can also serve as reference material for drafting product standards. It also notes that the same product can face different requirements in a hospital, in home care or in an institution.
Examples: wheelchairs have the ISO 7176 series; walking aids manipulated by both arms have the ISO 11199 series; walking aids manipulated by one arm, such as elbow crutches and walking sticks with three or more legs, have the ISO 11334 series; assistive products for personal hygiene that support users have ISO 17966; hoists for the transfer of persons have ISO 10535:2021. Test these products to their product standard first, and use ISO 21856 only for real risks the product standard leaves uncovered. Whether SUNGO's accreditation covers these product standards, which parts it covers and which clauses it excludes depend on the certificate scopes; ISO 10535 is not within SUNGO's accreditation scope.
Is there a Chinese counterpart?
China has a recommended national standard with a similar title: GB/T 41697-2022, Assistive products for persons with disability — General requirements and test methods, published 2022-10-12 and in force from 2023-02-01, under the Ministry of Civil Affairs. The national standards information platform does not list an international standard that it adopts, so whether its clauses match ISO 21856 one-to-one should be checked in the standard text. GB/T 41697 is not within SUNGO's accreditation scope.
What the clauses cover
ISO 21856:2022 has 28 clauses and three informative annexes: A, general recommendations; B, environmental and consumer-related guidance; C, guidelines for accessible information. The table groups the clauses by topic, following the ISO 21856:2022 table of contents. For requirements, test methods and pass/fail criteria, refer to the standard text.
| Clause | Topic | When to look closely |
|---|---|---|
| 4 (4.1–4.11) | General requirements: risk analysis and management, intended performance and technical documentation, clinical evaluation and investigation, products that can be dismantled, fasteners, load limits, immobilising means, usability, design for persons with sensory and cognitive impairments, accessibility, feedback | Every product |
| 5 | Materials: flammability, biocompatibility and toxicity, contaminants and residues, infection and microbiological contamination (cleaning and disinfection, machine-washable products, animal tissue), resistance to corrosion | Depends on materials and use; 5.5.3 and 5.5.4 for washable products or animal-derived materials |
| 6 | Emitted sound and vibration, including audible warning devices | Motors, moving mechanisms, audible warnings |
| 7, 8 | Electromagnetic compatibility; electrical safety (charge level indicators, electrically heated blankets and pads, ingress of liquids or particulate matter, pendant controls) | Products with electrical functions |
| 9, 10 | Overflow, spillage, leakage and ingress of liquids; surface temperature | Products that hold or contact liquids, or have heated parts |
| 11 | Sterility: requirements, sterilization processes, maintenance of sterility in transit | Products supplied sterile |
| 12–16 | Safety of moving parts (squeezing, mechanical wear, emergency stopping); means to prevent falling out (including side rails); prevention of traps (holes and clearances, V-shaped openings); folding and locking mechanisms; carrying handles | Moving or folding parts, side rails, handles |
| 17 | Products that support or suspend users: static forces, dynamic forces, tips | Products that carry or suspend the user |
| 18–21 | Portable and mobile products; surfaces, corners, edges and protruding parts; hand-held products; products for children | Product form and user group |
| 22–24 | Stability; forces in soft tissues; ergonomic principles | Most products |
| 25–28 | Information supplied by the manufacturer (pre-sale, user and service information, labelling); packaging; test report; accessible information | Every product |
Clause 4: risk analysis decides which clauses apply
Clause 4.1 requires hazards to be identified and risks estimated using the procedures in ISO 14971 and, where relevant, ISO 12100. No single product uses every clause, so the risk management file should record, clause by clause, whether each applies and why; the test plan and report then take their clause list from that record.
Clause 4.6, Load limits, should be read with the definitions in 3.16. Maximum user mass is the greatest permissible mass of the person using the product; minimum user mass is the lowest user mass needed to activate or use it; maximum load, or safe working load, is the maximum external mechanical load permitted on the equipment or a part of it in normal use. As the note explains, a product that carries a user, an assistant and possibly accessories has a maximum load above the maximum user mass, while for a product the user only leans on, the maximum user mass is the higher figure. These declared values feed the relevant tests and should be fixed before testing is ordered.
Clauses 5 and 11: materials and sterility
Clauses 5.4 (contaminants and residues), 5.5 (infection and microbiological contamination) and 5.6 (resistance to corrosion) are within the Hefei IAS scope. For the requirements of 5.5.2, Cleaning and disinfection, refer to the standard text; please send the manufacturer's cleaning and disinfection instructions with the samples. Clauses 5.2 (flammability) and 5.3 (biocompatibility and toxicity) are excluded. According to the table of contents, clause 5.2 covers not only upholstered parts, mattresses, bed bases, bedding and textiles (5.2.2) but also polymeric parts, electrical components and wiring (5.2.3–5.2.5). The ignitability standards for textiles, mattresses and upholstered furniture in the normative references (ISO 12952-1/-2, EN 597-1/-2, EN 1021-2), and the cable and fire-test standards (IEC 60332-1-2, IEC 60695-11-10, UL 1581), are not within SUNGO's accreditation scope.
Clause 11 only matters for products supplied sterile. Normative references related to sterilization and sterile packaging include EN 556-1, ISO 11135, ISO 11137-1/-2, ISO 25424 and ISO 11607-1; how each is used should be checked in the standard text. Because clause 11 is excluded, these tests must be ordered separately under their own standards. Whether the SUNGO Shanghai laboratory can take them on depends on its certificate scope; see the sterilization validation and sterile packaging validation pages.
Clauses 7 and 8: products with electrical functions
The normative references include IEC 60601-1:2005+AMD1:2012+AMD2:2020, IEC 60601-1-2, IEC 60601-1-11 and IEC 60529; how far each clause relies on them should be checked in the standard text. Clause 7 (electromagnetic compatibility) and clause 8 (electrical safety, except 8.3) are covered by the Hefei laboratory's IAS accreditation (TL-1119), and IEC 60601-1, IEC 60601-1-2 and IEC 60601-1-11 are also listed as separate standards in that IAS accreditation (TL-1119). Clause 8.3, Electrically heated blankets, pads and similar flexible heating appliances, is excluded, and the related IEC 60601-2-35 is not within SUNGO's accreditation scope. See the medical electrical safety and EMC pages.
Clause 17: products that support or suspend users
Clause 17 splits static forces into products that support users (17.2.1) and products that suspend them (17.2.2), covers dynamic forces in 17.3, and gives requirements and test methods for tips, covering friction and durability, in 17.4. Clause 17.4 is excluded. ISO 24415-1 (friction of tips for walking aids) and ISO 24415-2 (durability of tips for crutches), which deal with the same subject, are not within SUNGO's accreditation scope.
Deciding which clauses apply to your product
- Search for a product-specific standard first (ISO, EN, Chinese national and industry standards) and record what you searched and concluded.
- If one exists, use it as the main basis, then go through ISO 21856 and add only the real risks the product standard leaves uncovered.
- If none exists, use ISO 21856 as the main basis. Clauses that do not depend on product form, such as clauses 4 and 25, should normally be checked; decide the rest by features such as moving or folding parts, supporting or suspending the user, hand-held use, use by children and electrical functions.
- Define the use environment (hospital, home care or institution), since requirements can differ between them.
- Record in the risk management file whether each clause applies, with reasons where it does not, and list items under excluded clauses separately.
SUNGO accreditation scope (as written on the certificate)
These rows are taken from the IAS accreditation scope of Hefei SUNGO Medical Technology Co., Ltd. (TL-1119, effective 2026-06-04). The standard numbers and titles in the Item column, and the limitation text after "Certificate text:", are copied from the certificate; the Clauses covered column restates that limitation, and the other notes are this guide's own.
| Item | Clauses covered | Notes / limitations |
|---|---|---|
| ISO 21856 Assistive products - General requirements and test methods | Whole standard except 4.3, 5.2, 5.3, 8.3, 11, 17.4 | Certificate text: except for 4.3 Clinical evaluation, 5.2 Flammability, 5.3 Biocompatibility and toxicity, 8.3 Electrically heated blankets, pads and similar flexible heating appliances, 11 Sterility, 17.4 Requirements and test method for tips. No edition year stated. |
| EN ISO 21856 Assistive products - General requirements and test methods (ISO 21856:2022) | Whole standard except 4.3, 5.2, 5.3, 8.3, 11, 17.4 | Same exclusions. No year in the standard number; the title refers to ISO 21856:2022. |
| IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance | Whole standard | No edition year stated. |
| IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | Whole standard | No edition year stated. |
| IEC 60601-1-11 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment | Whole standard | No edition year stated. |
How the excluded clauses are handled:
| Excluded clause | Subject | Handling |
|---|---|---|
| 4.3 | Clinical evaluation (Clinical evaluation and investigation in the table of contents) | Requirements as stated in the standard text; outside this accreditation scope |
| 5.2 | Flammability | Outside this accreditation scope; order separately |
| 5.3 | Biocompatibility and toxicity | The biological tests involved (cytotoxicity, sensitization, irritation, implantation and similar) are subcontracted by SUNGO to a laboratory with the relevant competence; in line with accreditation rules, the client's agreement is obtained first, and the report identifies the subcontracted items and the subcontractor. Other requirements of 5.3 are as stated in the standard text and are outside this accreditation scope |
| 8.3 | Electrically heated blankets, pads and similar flexible heating appliances | Outside this accreditation scope; order separately |
| 11 | Sterility | Outside this accreditation scope; order sterilization and sterile-packaging tests separately under their own standards |
| 17.4 | Requirements and test method for tips | Outside this accreditation scope; order separately |
Where the testing is done and under which accreditation
Testing to ISO 21856 and EN ISO 21856 is done at the Hefei laboratory. Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity, with its laboratory in Zipeng Town, Feixi County, Hefei. Shanghai SUNGO Medical Technology Co., Ltd. (laboratory at 958 Kangqiao East Road, Pudong, Shanghai) is also a separate legal entity, and its certificate scopes do not include these standards.
- Accreditation on the report: reports to these two standards are issued under the Hefei laboratory's IAS accreditation (TL-1119).
- Edition: the IAS scope states no edition year; confirm with the laboratory which edition the IAS report will cite before ordering.
- Electrical safety and EMC: covered by the Hefei laboratory's IAS accreditation (TL-1119).
- Biocompatibility: handled through the subcontracting arrangement in the table above.
- Acceptance: whether a regulator, notified body or buyer accepts the report depends on their requirements, and SUNGO makes no promise about it. Please check with them before you submit.
For the full split of accreditations between the two sites, see SUNGO Medical Device Testing Scope: CNAS, CMA and IAS in Shanghai and Hefei. To check a certificate scope and the marks on a report yourself, see How to Check an Accreditation Scope and Read the Mark on a Report. For wheelchair and assistive product testing in general, see the wheelchair and assistive product service page.
Preparing a submission
- Product description: intended use, intended users, use environment, target markets, and whether the product is regulated as a medical device there.
- Standards search: which product-specific standards you checked, whether any apply, and which risks need ISO 21856 as a supplement.
- The risk management file, or at least an ISO 14971 hazard analysis summary with clause-by-clause applicability decisions. See Turning an ISO 14971 Risk Analysis into a Test List.
- Declared values: maximum user mass and maximum load (safe working load), plus minimum user mass where relevant.
- Configuration list: models, configurations, accessories and detachable parts, with the adjustment range and locking method of adjustable or folding mechanisms.
- Draft instructions for use and labels; clause 25 covers pre-sale, user and service information and labelling.
- Cleaning and disinfection instructions; whether the product is machine-washable or contains animal-derived materials.
- For electrical products: power supply, battery and charging arrangement, pendant controls, and the manufacturer's declared enclosure protection rating.
- Whether the product is supplied sterile, and a materials list for scoping the subcontracted biocompatibility work.
- Samples in their market packaging; the number of samples is agreed with the laboratory as part of the test plan.
Frequently asked questions
My product already has a product-specific standard. Do I still need ISO 21856?
The product standard comes first. ISO 21856 can supplement it where it does not cover a requirement, so whether you add any clauses depends on whether your risk analysis shows risks the product standard leaves uncovered.
What is the difference between ISO 21856 and EN ISO 21856?
The EN ISO 21856:2022 endorsement notice says CEN approved the ISO 21856:2022 text without modification. The European version adds a European foreword and supersedes EN ISO 16201:2006 and EN 12182:2012. Both are listed separately in the Hefei IAS scope with the same exclusions.
Can I order biocompatibility and flammability testing from SUNGO together?
SUNGO subcontracts the biological tests under clause 5.3 (cytotoxicity, sensitization, irritation, implantation and similar) to a laboratory with the relevant competence; the client's agreement is obtained before subcontracting, and the report identifies the subcontracted items and the subcontractor. Clause 5.2, Flammability, is outside SUNGO's accreditation scope for this standard and must be ordered separately.
Where are electrical safety and EMC tested for powered assistive products?
At the Hefei laboratory, covered by its IAS accreditation (TL-1119). Clause 8.3, Electrically heated blankets, pads and similar flexible heating appliances, is excluded.
When you ask for a quotation, please give the product name and model, intended use and users, use environment, target markets, any product-specific standards found, maximum user mass and maximum load, whether the product has electrical functions or is supplied sterile, and the edition the report should cite. The laboratory will use this to confirm the applicable clauses and anything that needs separate arrangements. Please send your request through the contact page.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.