Two different tests often arrive under the same label, "sterility test". One is the test of sterility used in sterilization process validation under ISO 11737-2:2019, or its Chinese identical adoption GB/T 19973.2-2025, where product has first received a reduced exposure to the sterilizing agent. The other is the sterility test in General Chapter 1101 of the Pharmacopoeia of the People's Republic of China (ChP), Volume IV, which is performed on product that has already been through an aseptic process or a sterilization process. They differ in purpose, samples and the meaning of a positive, and citing the wrong one can leave you with an unusable report. Below we compare the two from the standards' own text, explain what changed in GB/T 19973.2-2025, and set out SUNGO Shanghai's accredited scope and submission requirements.
Key takeaways
- Different purposes: ISO 11737-2 covers tests of sterility performed when defining, validating or maintaining a sterilization process, on product exposed to a treatment reduced relative to routine sterilization. ChP 1101 is the pharmacopoeial sterility test, performed on product after an aseptic process or a sterilization process.
- Not interchangeable: ISO 11737-2 states that it does not apply to sterility testing for routine product release, to the pharmacopoeial test for sterility, or to sterility testing used to demonstrate shelf life, stability or package integrity.
- Positives mean different things: in validation, the number of positive tests is test data (e.g. for radiation dose setting or a dose audit) and is judged against what the sterilization standard specifies for the chosen method. In the pharmacopoeial test, results are judged according to the result interpretation rules in the pharmacopoeia text.
- The 2025 Chinese edition: GB/T 19973.2-2025 was published on 2025-01-24, took effect on 2026-02-01, replaces GB/T 19973.2-2018 in full and is an identical adoption of ISO 11737-2:2019.
- Which report SUNGO can issue: all of these tests are done in the Shanghai laboratory. ISO 11737-2:2019 is accredited by CNAS. ISO 11737-2 is also accredited by IAS (the IAS scope gives no year and marks it Third edition; confirm the edition with the laboratory before an IAS report is issued). GB/T 19973.2-2025 is covered by CMA only, so no CNAS or IAS report is available under that number. ChP 1101 is covered by CNAS, CMA and IAS (IAS limited to medical device testing).
The standards and their scope
ISO 11737-2:2019
ISO 11737-2:2019 is the third edition, published in December 2019 by ISO/TC 198, Sterilization of health care products, replacing the 2009 second edition. According to the ISO website, this edition was reviewed and confirmed in 2025 and remains current. Clause 1 says it specifies general criteria for tests of sterility on medical devices that have been exposed to a sterilizing treatment reduced relative to what routine processing would use. These tests are meant for defining, validating or maintaining a sterilization process.
The same clause lists four exclusions: sterility testing for routine release of sterilized product; performing a pharmacopoeial test for sterility; tests used to demonstrate product shelf life, stability and/or package integrity; and culturing of biological indicators or inoculated products. A note adds that the first two are not requirements of sterilization process standards such as ISO 11137-1 and ISO 11137-2.
The introduction explains the logic: a reduced exposure, often called a fractional exposure or verification dose, reveals how resistant the product's naturally occurring contamination is to the sterilizing agent, and the exposed items are then tested individually. Examples include establishing a radiation sterilization dose and showing that an established dose remains valid.
GB/T 19973.2-2025
According to China's national standards information platform, GB/T 19973.2-2025, Sterilization of health care products, Microbiological methods, Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process, is a recommended national standard. It was published on 2025-01-24, took effect on 2026-02-01, replaces GB/T 19973.2-2018 in full, is an identical adoption of ISO 11737-2:2019, and sits with the national technical committee for disinfection technology and equipment. The 2018 edition, which adopted ISO 11737-2:2009, was withdrawn on 2026-02-01. The English title also changed, from "Sterilization of medical devices" to "Sterilization of health care products".
Because the two Chinese editions adopt the 2019 and 2009 ISO editions, the technical changes between them track the main changes listed in the ISO 11737-2:2019 foreword (check the GB/T 19973.2-2025 foreword itself for its exact wording):
- New requirement: the time between manufacture of the test samples and exposure to the sterilizing agent should be as short as possible.
- New requirement: samples stay immersed in the culture medium during incubation, with a rationale where that is not possible.
- Added guidance: packaging is not usually tested for sterility unless it is an integral part of the product.
- Added guidance on what a "controlled environment" for tests of sterility means.
- Added guidance: if method suitability still fails after several attempts to eliminate inhibitory substances, a reduction in them may be accepted with a rationale and risk assessment.
- Added guidance that microorganisms from positive growth should generally be identified.
- Added guidance on periodically demonstrating ongoing method suitability, so that small changes do not build up over time.
- A new table showing the typical split of responsibilities between manufacturer and laboratory (Annex B).
ChP General Chapter 1101
ChP Volume IV General Chapter 1101 is the sterility test method. ISO 11737-2 calls this kind of test a "test for sterility": a technical operation specified in a pharmacopoeia and performed on product following an aseptic process or exposure to a sterilization process. SUNGO's Shanghai scope lists both the 2020 and 2025 editions of Chapter 1101, and we test to the edition cited in the client's product standard or technical requirements. For the number of items, the quantity tested, the culture media, the incubation conditions and the interpretation of results, refer to the text of the edition cited; this article does not restate those values.
"Test of sterility" vs "test for sterility"
Clause 3 of ISO 11737-2:2019 keeps these terms apart. A test of sterility is performed in development, validation or requalification to determine whether viable microorganisms are present on product or portions of it, after a treatment reduced from the full process. A test for sterility is the pharmacopoeial test. Read protocols and reports with care.
Side-by-side comparison
| Aspect | ISO 11737-2:2019 / GB/T 19973.2-2025 | ChP General Chapter 1101 |
|---|---|---|
| Purpose | Defining, validating and maintaining a sterilization process, e.g. radiation dose setting and sterilization dose audits | Pharmacopoeial sterility test of product |
| State of the product | Exposed to a treatment reduced relative to routine sterilization, e.g. a verification dose | After an aseptic process or a sterilization process |
| Number of items and batches | Rationale must be documented; the number may be set by the applicable sterilization standard | As specified in the pharmacopoeia text |
| What is tested | The entire product whenever practicable; a sample item portion (SIP) where the sterilization standard permits, with its adequacy demonstrated | Quantity as specified in the pharmacopoeia text |
| Methods | Three general approaches: direct immersion, removal and transfer of microorganisms, filtration of liquid products | As specified in the pharmacopoeia text |
| Media and incubation | Term 3.4 defines "culture condition" as the combination of growth media and manner of incubation (which can include temperature and time); see the standard for the requirements | As specified in the pharmacopoeia text |
| Positive results | Validation data, judged against what the sterilization standard specifies for the method used; organisms should generally be identified | Handled under the pharmacopoeia's result interpretation rules |
Product selection, batches and product families
Under ISO 11737-2, selected items must represent routine production, including packaging materials and processes. Where products are grouped into a family, the rationale for including each product is recorded and the test product must represent the whole family. The rationale for the number of items and batches is documented too; a note says the relevant sterilization standard may describe this. For radiation, ISO 11137-2 defines product families mainly by bioburden, the numbers and types of microorganisms on the product.
Sample item portion (SIP)
Whenever practicable the whole product is tested; a sample item portion (SIP) may be used only where the applicable sterilization standard allows. SIP selection depends on bioburden distribution. If bioburden is known to be even, the SIP can come from any part of the item. If known to be uneven, the SIP either represents each material proportionally or is the portion presenting the most severe challenge to the process, in which case the relationship between that portion's bioburden and the whole item's should be established. If the distribution is unknown, the SIP represents each material proportionally. The SIP can be calculated from length, mass, volume or surface area, and its adequacy must be demonstrated; ISO 11137-2, for example, sets criteria for this. Bioburden itself is determined under ISO 11737-1; see How to Determine Bioburden: Key Points of ISO 11737-1.
Packaging of test items
Product or SIPs should be packaged as in routine production. Any difference is documented, and the packaging must still allow the intended treatment, maintain microbiological status and give the sterilizing agent equivalent access.
Three general methods
Clause 6 gives three approaches: direct immersion of product in culture medium (or adding medium to it) and incubating, keeping it immersed throughout where possible and justifying any exception such as buoyant materials; removing microorganisms from the product and transferring them to medium; and filtering liquid products, then immersing the filter in medium. Factors that influence method design, such as the parts of the product labelled sterile and the product's physical and chemical nature, must be considered and recorded. Clauses 7 and 8 cover assessment and maintenance of the test method.
Reading a positive result
Validation items were never fully sterilized, so a positive does not mean the product has failed. ISO 11137-2 defines a positive test of sterility as detectable microbial growth from product or portions of it, and a false positive as growth from extraneous contamination or turbidity from product-medium interaction. Per its introduction, dose setting methods use results from items given less than the sterilization dose to predict the dose needed for the target sterility assurance level, Method VDmax uses them to substantiate a preselected dose, and dose audits confirm an established dose remains effective. Whether a given number of positives is acceptable depends on what ISO 11137-2 specifies for the method chosen.
In the pharmacopoeial test for sterility (term 3.12, quoted above), positives are handled under the result interpretation rules of the pharmacopoeia edition cited.
Where ISO 11737-2 fits in radiation dose setting and dose audits
The first two examples in the ISO 11737-2 introduction are establishing a radiation sterilization dose and demonstrating the continued validity of an established sterilization dose. ISO 11137-2:2013 lists ISO 11737-1 and ISO 11737-2 as normative references, and both its dose setting methods and Method VDmax involve tests of sterility on items irradiated below the sterilization dose. SUNGO Shanghai's accreditation here is as follows. CNAS covers GB 18280.2-2015 Clause 7, Method 1 only, using bioburden information. The CMA scope lists it as GB18280.2-2015/7, Method 1 only, using bioburden information. Under IAS, ISO 11137-2 and GB/T 18280.2 are limited to the items on the certificate: Radiation sterilization dose setting (VDmax15or25), Sterilization Dose Review (Method 1) and Sterilization dose review (VDmax15or25). ISO 11137-1 and ISO 11137-3 are outside SUNGO's accredited scope. For how dose setting works, see Setting Radiation Sterilization Dose: ISO 11137-2 and Dose Audits.
Note also that ISO 11137-2:2013 has an amendment, ISO 11137-2:2013/Amd 1:2022. According to the ISO website, the standard is at the "to be revised" stage and the next edition, ISO/DIS 11137-2, is under development; this article cites the 2013 base text. According to China's national standards platform, GB/T 18280.2-2025 was published on 2025-12-02, takes effect on 2027-07-01, is a modified adoption of ISO 11137-2:2013, and will replace GB 18280.2-2015 in full. SUNGO's CNAS and CMA scopes list GB 18280.2-2015. The IAS scope lists GB/T 18280.2 without a year; confirm the edition with the laboratory before an IAS report is issued.
Accredited scope
The table lists the Shanghai scope entries behind the capabilities described in this article. Item names, clauses and limitations follow the certificates. IAS entries carry no year.
| Item | Standard and clause | Notes and limitations |
|---|---|---|
| Sterility test | ISO 11737-2:2019 | CNAS; test object: medical devices. IAS separately accredits the whole of ISO 11737-2 (no year on the certificate; marked Third edition). |
| Sterility test | GB/T 19973.2-2025 | CMA only; test object: medical devices. No CNAS or IAS report. |
| Sterility test | ChP 2020 edition, Volume IV, General Chapter 1101 | CNAS; test object: medical devices. |
| Sterility test | ChP 2025 edition, Volume IV, General Chapter 1101 | CNAS; test object: medical devices. |
| Sterility test | ChP 2020 edition, Volume IV, 1101 | CMA; test object: sterile medical devices. |
| Sterility test | ChP 2025 edition, General Chapter 1101 | CMA; test object: medical devices. |
| Sterile test method | Pharmacopoeia of the People's Republic of China, Volume 4, Section 1101 | IAS; limitation: Only accredited for medical device testing. No edition stated. |
| Radiation dose establishment test | GB 18280.2-2015 Clause 7 | CNAS; test object: medical devices; limitation: Method 1 only, using bioburden information. |
| Radiation dose confirmation | GB 18280.2-2015 (written on the certificate as GB18280.2-2015/7) | CMA; test object: health care products; limitation: Method 1 only, using bioburden information. |
| Establishing the sterilization dose (partial) | ISO 11137-2; GB/T 18280.2 | IAS; limitation: Only accredited for Radiation sterilization dose setting (VDmax15or25), Sterilization Dose Review (Method 1), Sterilization dose review (VDmax15or25). |
| Determination of a population of microorganisms on products | ISO 11737-1 | IAS, whole standard (no year). |
Where the work is done and under which accreditation
- Location: the laboratory of Shanghai SUNGO Medical Technology Co., Ltd., No. 958 Kangqiao East Road, Pudong, Shanghai, holding CNAS L22083, CMA 240015340087 and IAS TL-1019. Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity, and its certificates do not cover these tests.
- CMA report for use in China: use GB/T 19973.2-2025 for validation tests of sterility, and the 2020 or 2025 edition of ChP 1101 for the pharmacopoeial test. The CMA mark covers only the items in the CMA schedule and, as the certificate states, is valid within the People's Republic of China.
- CNAS or IAS report: for validation tests of sterility, a CNAS report is issued against ISO 11737-2:2019 and an IAS report against ISO 11737-2, with the edition confirmed with the laboratory first; GB/T 19973.2-2025 cannot carry a CNAS or IAS report. CNAS is a signatory to the ILAC and APAC arrangements, and the IAS certificate carries the ilac-MRA mark. Whether a particular overseas regulator accepts a particular report is that regulator's decision.
- Edition check: IAS scopes carry no year; confirm the edition with the laboratory before any IAS report is issued.
Preparing your submission
- State the purpose clearly: sterilization process validation (dose setting or dose audit) or the pharmacopoeial sterility test, with the standard number and edition. Avoid a generic "sterility test" request.
- For validation work, give the sterilization standard and method (for example ISO 11137-2 Method 1 or VDmax) and the rationale for the number of items and batches. If products are grouped into a family, include the family rationale and why the test product represents it.
- Say whether whole items or SIPs will be tested. For SIPs, provide bioburden distribution information, the basis for calculating the SIP and evidence of its adequacy.
- Package samples as in routine production where possible and flag any differences in advance. In line with the 2019 edition, keep the time between manufacture and exposure as short as possible and plan sample dispatch accordingly.
- Provide materials, construction and the parts labelled sterile for method design and suitability assessment, and flag anything that may inhibit microbial growth.
- Using Annex B of ISO 11737-2 as a reference, agree in advance which tasks sit with the manufacturer and which with the laboratory.
- For the pharmacopoeial test, state which ChP edition (2020 or 2025) your product standard or technical requirements cite.
- State the report type you need (CMA, CNAS or IAS) so the right standard number is used.
FAQ
Can a pharmacopoeial sterility test on finished product replace sterilization validation?
No. The pharmacopoeial test for sterility is outside the scope of ISO 11737-2 and, per a note in the standard, is not a requirement of sterilization process standards such as ISO 11137-1 and ISO 11137-2. Validation uses tests of sterility after a reduced exposure, as set out in the relevant sterilization standard. The ISO 11737-2 introduction adds that performing a test of sterility on product exposed to the complete sterilization process provides no scientifically usable data and is not recommended.
Should I cite GB/T 19973.2-2025 or ISO 11737-2:2019?
The two are an identical adoption pair. At SUNGO, GB/T 19973.2-2025 can only carry a CMA report and ISO 11737-2:2019 can carry a CNAS report; if you need an IAS report, it is issued against ISO 11737-2, with the edition confirmed with the laboratory first. Choose by how the report will be used.
Can sterility tests for shelf-life or package integrity studies follow ISO 11737-2?
Not directly. ISO 11737-2 excludes tests used to demonstrate shelf life, stability and/or package integrity. The manufacturer sets the method in the study protocol; for related packaging and ageing tests, see our packaging validation and shelf-life validation services.
Does a positive in a dose setting experiment mean the product fails?
No. Those items received less than the sterilization dose. The number of positives is data for predicting or substantiating the dose and is judged by what ISO 11137-2 specifies for the chosen method. ISO 11137-2 also defines false positives; see the standard for how they are handled.
Does the US FDA recognize ISO 11737-2:2019?
The FDA Recognized Consensus Standards database lists ISO 11737-2 Third edition 2019-12 under recognition number 14-540, with recognition of the complete standard. Whether a given submission is accepted is for FDA to decide.
When requesting a quote, tell us the product name, construction and materials, sterilization method, the purpose of testing (dose setting, dose audit or pharmacopoeial sterility test), the standard and edition, and the report type you need, through our contact page. For an overview of sterilization-related testing, see our sterilization validation service.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.