For the same sterile medical device, a US submission usually cites ASTM F1980-21, while Chinese registration testing usually cites YY/T 0681.1-2018. Where do the two accelerated aging guides differ, and can one protocol and one batch of samples satisfy both? This article compares the texts clause by clause (scope, humidity, Q10 and the temperature ceiling, real-time aging, reporting) and shows how to write one protocol for both. Choosing Q10 is not covered here.
Key points
- The core framework is the same. Both treat Q10 = 2 as the common, conservative choice, keep the aging temperature at or below 60 °C unless a higher one is shown to be appropriate, require real-time aging in parallel, and list essentially the same report content.
- The scope differs. The F1980-21 title adds "and Medical Devices", and clause 1.1 says its protocol guidance may also be used for medical devices and device materials. YY/T 0681.1-2018 covers only sterile barrier systems as defined in GB/T 19633.1-2015 and their packaging material components.
- Humidity is the biggest difference. F1980-21 clause 7.4.6 has the protocol define humidity conditions and says the rationale for not controlling humidity should be documented; Appendix X3 recommends targeting 45 % to 55 % RH when controlling. YY/T 0681.1-2018 only asks you to decide whether to use humidity conditions and, if so, to set a target and tolerance; it asks for no rationale if humidity is uncontrolled.
- Edition lineage. The foreword of YY/T 0681.1-2018 states no adoption relationship, and its main clauses correspond to ASTM F1980-16, not F1980-21.
- One protocol can meet the text of both if it follows the more detailed requirement: state humidity conditions, with a reason if humidity is not controlled; use the upper limit of any labeled long-term storage range as the ambient temperature; run real-time aging in parallel with the same evaluation methods.
- Accreditation marks. At both SUNGO laboratories (Shanghai and Hefei), CNAS covers both standards and CMA covers only YY/T 0681.1-2018, in each case under packaging materials for medical devices; whether aging of the device itself can be reported within accreditation should be confirmed when ordering. Both IAS scopes list the two as whole standards without edition years, so confirm the edition before an IAS-accredited report is issued.
The two standards and their scope
| Item | ASTM F1980-21 | YY/T 0681.1-2018 |
|---|---|---|
| Title | Standard Guide for Accelerated Aging of Sterile Barrier Systems and Medical Devices | Test methods for sterile medical device package — Part 1: Test guide for accelerated aging |
| Publication | Under ASTM Subcommittee F02.50; approved 2021-12-15, published December 2021; previous edition F1980-16 | Issued by the National Medical Products Administration on 2018-12-20, effective 2020-01-01; recommended industry standard replacing YY/T 0681.1-2009 |
| Packaging standard referenced | ANSI/AAMI/ISO 11607-1:2019 | GB/T 19633.1-2015 |
| Scope | Sterile integrity of the sterile barrier system and physical properties of its packaging materials; the protocol guidance may also be used for medical devices and device materials | Sterile integrity of the sterile barrier system and physical properties of its packaging material components |
| Real-time aging protocols | Not addressed, but real-time aging must confirm accelerated results using the same evaluation methods | Not applicable to real-time aging protocols; Note 3 describes real-time aging as the basic study that confirms accelerated results |
| Label storage conditions | Clause 1.7: accelerated aging data are not to be used to establish label storage conditions | No corresponding clause |
Both are guides: they show how to design an aging protocol but do not prescribe temperature, duration or acceptance criteria for a product; the manufacturer sets those and documents the rationale.
Which ASTM edition does YY/T 0681.1-2018 follow?
The 2009 edition is marked "ASTM F 1980:02, MOD" on its cover, a modified adoption of F1980-02. The 2018 foreword states no adoption relationship, and its bibliography lists ASTM F 1980 without a date. Clause by clause, 2018 clauses 6.5, 7.3.4 and 7.4.5 and Annex C match the same-numbered clauses and Appendix X3 of ASTM F1980-16, and Table C.1 matches Table X3.1. F1980-21 rewrote, moved or deleted each of these: clause 6.5 was rewritten, clause 7.3.4 is gone, the humidity step moved to 7.4.6, and Appendix X3 was rewritten without Table X3.1. The differences therefore come mostly from the ASTM revision between the 2016 and 2021 editions.
The FDA Recognized Consensus Standards database lists ASTM F1980-21 under recognition number 14-575, entered on 2023-02-20, with complete-standard recognition. Whether a given report is accepted is decided by FDA during review.
Clause-by-clause comparison
Scope: can the device itself be aged under these guides?
F1980-21 clause 1.2 says information obtained using the guide may be regarded as sufficient evidence for expiration date claims for medical devices and sterile barrier systems until real-time data are available. Clause 7.1 adds moisture absorption and known degradation mechanisms (hydrolysis, oxidation, corrosion, environmental stress cracking and others) to material characterization, and Note 2 explains that most flexible packaging materials do not degrade through moisture, whereas device materials are far more varied and some may hydrolyze.
Clause 1.3, however, still addresses sterile barrier systems as a whole, with or without devices; clause 7.4.1 applies the requirements to the sterile barrier system; and Section 8 is written around it. F1980-21 thus gives device aging protocols a framework to cite but does not say how to evaluate device performance after aging; that comes from the product standard or technical requirements.
Humidity: the largest difference
| Topic | ASTM F1980-21 | YY/T 0681.1-2018 |
|---|---|---|
| Theory | 6.5 and 6.6: consider RH as well as temperature; humidity control offsets the low RH at elevated temperature, so materials are not dried out and moisture-driven degradation seen on the shelf is not missed | 6.5: no humidity factor in the aging time calculation; effects of unrealistic extreme temperature and humidity are a separate study, unrelated to material aging |
| Protocol step | 7.4.6: define the humidity conditions; if controlled, define the RH target and tolerance; if not controlled, the rationale should be documented | 7.4.5: decide whether humidity conditions will be used; if so, set the RH target and tolerance |
| Appendix | X3 (nonmandatory): target 45 % to 55 % RH when controlling humidity; a rationale should be documented for other conditions; maintain comparable relative, not absolute, humidity | Annex C (informative): set the RH range from material knowledge, supplier recommendations and history; Table C.1 gives the RH for constant moisture content at several temperatures; assess humidity effects on materials separately from the aging protocol |
| Tolerances | Note 5: ±2 °C and ±5 % humidity are acceptable | 7.2.3, Note 3: ±2 °C and ±5 % RH |
YY/T 0681.1-2018 follows F1980-16 and sees humidity mainly as an extra condition that protects materials from damage; its Table C.1 assumes constant moisture content, so the higher the temperature, the lower the corresponding RH. F1980-21 Appendix X3.4 says that at elevated aging temperatures (for example 50 °C to 60 °C), low RH (for example below 30 % RH) typically leaves less moisture in the material and can slow hydrolysis inherent in real-time storage. Above 50 °C, RH over 55 % may cause unnatural physical changes, such as delamination of laminates and coextrusions that incorporate water-based or water-soluble primers or adhesives, particularly hygroscopic polymers. The 45 % to 55 % RH target sits between those two risks.
Under F1980-21 clause 7.4.6, the protocol defines humidity conditions, and a rationale should be documented if humidity is not controlled. Under YY/T 0681.1-2018 you may decide not to use humidity conditions (clause 7.4.5) without documenting a reason, although Section 9 a) still requires the protocol to state the aging conditions, including humidity. For moisture-absorbing, hydrolysis-prone materials in the package or device (F1980-21 cites polyamides), justify the humidity condition carefully.
Q10 and the temperature ceiling: essentially the same
| Topic | ASTM F1980-21 | YY/T 0681.1-2018 |
|---|---|---|
| Q10 | 7.3.1: Q10 = 2 is a common and conservative choice; Note 7: a more aggressive coefficient such as 2.2 to 2.5 may be used if the system is well characterized in the literature, at the manufacturer's responsibility | 7.3.1 and note: same content |
| Temperature ceiling | 7.2.3.2: keep TAA at or below 60 °C (for example 50 °C) unless a higher temperature has been shown to be appropriate | 7.2.3 b): not above 60 °C unless a higher temperature is shown to be appropriate |
| Material transitions | 7.2.3.1: below material transitions or the temperature at which the sterile barrier system distorts | 7.2.3 a): same content |
| Ambient temperature TRT | Note 3: typically 20 °C to 25 °C, with 25 °C considered conservative; 7.2.2: for a labeled long-term storage temperature range, aligning TRT with its upper limit is recommended | 7.2.2 note: typically 20 °C to 25 °C, with 25 °C considered conservative; no upper-limit recommendation |
The aging factor equations are identical, and both guides state that determining aging factors is outside their scope. One point to watch: for a labeled long-term storage temperature range, F1980-21 recommends its upper limit as the calculation's ambient temperature; per clause 7.2.2, a higher TRT gives a longer, more conservative aging time.
Real-time aging: both require it in parallel
- Clause 4.3 is the same in both: real-time aging must run in parallel with accelerated aging and continue to the claimed shelf life until complete.
- Clause 4.5 is the same in both: an expiry date based on a conservative aging factor is tentative until real-time results are complete.
- F1980-21 clause 1.4 adds that real-time aging must confirm accelerated results using the same evaluation methods, and that real-time aging (stability) is required by ANSI/AAMI/ISO 11607-1:2019.
- If real-time results fail, both guides require cutting the shelf life to the longest period that successful real-time testing supports, and reviewing, documenting and acting on product released on accelerated data (YY/T 0681.1-2018 Annex B.8.2; F1980-21 Appendix X2.7.2).
Neither laboratory's accreditation lists a standard dedicated to real-time aging, so agree storage location, condition records and testing at each time point separately in the protocol. On organizing both files, see Accelerated and Real-Time Aging in Parallel: Organizing the File.
Post-aging testing and the report
Section 8 is essentially the same in both: evaluate physical properties and integrity, with acceptance criteria set before aging starts. F1980-21 adds that new sealing parameters and equipment do not require a new aging study (Note 11), and that ISO 11607-1:2019 does not require aging and performance testing to be combined (clause 8.6).
Section 9 lists the same five report items in both:
- a written protocol prepared before testing, giving the aging conditions (test temperature, humidity, cycle, ambient temperature), time frame, sample sizes, sterile barrier system description, sampling intervals and the tests at each interval;
- the chamber temperature and relative humidity, and the calibrated instruments used to measure and monitor them;
- the test method standards used to evaluate the sterile barrier system;
- the equipment used for physical and microbial testing, with calibration dates;
- the post-aging results, including any statistical methods used to judge conformance.
Item 1 in both guides lists humidity among the aging conditions; F1980-21 clause 7.4.6 adds that if humidity is not controlled, the rationale should be documented.
Writing one protocol that satisfies both
The two texts do not conflict; following the more detailed requirement in each area covers both.
- Samples and configuration. Build samples with the validated production process and sterilize them with the validated process, exposing them before aging to the most severe sterilization conditions at the expected number of cycles (F1980-21 Appendix X2.4 allows one more than expected). State whether protective packaging is included and whether the sterile barrier systems contain product or simulated product. If the study also covers the device itself, the samples must contain the device.
- Temperature. Choose the aging temperature from material characterization, below material transitions and at or below 60 °C. Set the ambient temperature to represent actual storage; for a labeled long-term storage range, F1980-21 recommends the upper limit.
- Q10 and duration. Use one Q10 and one pair of temperatures, calculate the aging time with the shared equations, and record the rationale.
- Humidity. Base the decision on the materials' moisture sensitivity and degradation mechanisms. If controlling, state target and tolerance and justify any target outside 45 % to 55 % RH; if not, state why. This meets F1980-21 clause 7.4.6 and YY/T 0681.1-2018 clause 7.4.5 together.
- Time points. Include time zero and the point equivalent to the desired shelf life. Both guides' example appendixes recommend considering at least three time points when trending.
- Conditioning. The example in YY/T 0681.1-2018 Annex B.5 conditions samples per GB/T 4857.2, a step absent from F1980-21 Appendix X2; including it does not affect conformance with F1980-21.
- Real-time aging. Start real-time samples from the same batch at the same time, use the same evaluation methods, and record actual storage temperature.
- Post-aging tests. Fix the tests (for example seal strength and package integrity), sample sizes, acceptance criteria and statistical methods before aging starts, and name the test method standards; confirm each test's accreditation status when ordering.
- Report. Agree at ordering how the report cites the standards and whether each standard gets its own report. Any CMA-marked part can only cite YY/T 0681.1-2018.
Accreditation scope
The table reproduces every entry for these two standards in both laboratories' schedules. The CNAS and CMA schedules list no clauses and no limitations; the IAS scopes give no edition year, so confirm the edition with the laboratory before an IAS-accredited report is issued against a particular edition.
| Laboratory and accreditation | Item | Standard and clause | Notes and limitations |
|---|---|---|---|
| Shanghai, CNAS L22083 | Accelerated aging | ASTM F1980-21 (no clause listed) | Test object: packaging materials for medical devices |
| Shanghai, CNAS L22083 | Accelerated aging | YY/T 0681.1-2018 (no clause listed) | Test object: packaging materials for medical devices |
| Shanghai, CMA 240015340087 | Accelerated aging | YY/T 0681.1-2018 (no clause listed) | Test object: packaging materials for medical devices; ASTM F1980 not on the CMA schedule |
| Shanghai, IAS TL-1019 | Whole standard | ASTM F1980 | No edition year |
| Shanghai, IAS TL-1019 | Whole standard | YY/T 0681.1 | No edition year |
| Hefei, CNAS L24279 | Accelerated aging | ASTM F1980-21 (no clause listed) | Category: packaging materials for medical devices |
| Hefei, CNAS L24279 | Accelerated aging | YY/T 0681.1-2018 (no clause listed) | Category: packaging materials for medical devices |
| Hefei, CMA 260015342945 | Accelerated aging | YY/T 0681.1-2018 (no clause listed) | Category: packaging materials for medical devices; ASTM F1980 not on the CMA schedule |
| Hefei, IAS TL-1119 | Whole standard | ASTM F1980 | No edition year |
| Hefei, IAS TL-1119 | Whole standard | YY/T 0681.1 | No edition year |
On the CNAS and CMA schedules, these entries cover accelerated aging itself, under packaging materials for medical devices. Post-aging tests such as seal strength and integrity, and any device tests, need their own accreditation check under their own methods.
Where SUNGO does it and under which accreditation
The Shanghai laboratory, at 958 Kangqiao East Road, Pudong, belongs to Shanghai SUNGO Medical Technology Co., Ltd. The Hefei laboratory, in Zipeng Town, Feixi County, belongs to Hefei SUNGO Medical Technology Co., Ltd. The two companies are separate legal entities, each issuing its own certificates and reports.
- Which site: both can perform accelerated aging of packaging materials for medical devices to these two standards. If the same samples need other tests, aging usually goes to the laboratory doing them, avoiding transfers between sites.
- CNAS: both sites list ASTM F1980-21 and YY/T 0681.1-2018.
- CMA: both sites list only YY/T 0681.1-2018. CMA covers only the items on its schedule and is valid only within the People's Republic of China.
- Test object: on both sites' CNAS and CMA schedules, accelerated aging is listed under packaging materials for medical devices; whether aging the device itself as the test object falls within accreditation should be confirmed when ordering.
- IAS: both sites list ASTM F1980 and YY/T 0681.1 as whole standards, without edition years; confirm the edition with the laboratory before an IAS-accredited report is issued against a particular edition.
See the shelf-life validation and sterile packaging validation service pages.
Preparing your submission
- Sterile barrier system and sample configuration: package format, protective packaging, product or simulated product.
- Materials: composition (including adhesives, primers and coatings), thermal transitions, additives, moisture absorption and known degradation mechanisms.
- Sterilization method, conditions and expected cycles; samples already sterilized by the validated process.
- Desired shelf life and labeled storage conditions, including any temperature range.
- Q10, ambient and aging temperatures with rationale; humidity control and why.
- Time points, tests, sample sizes and acceptance criteria, finalized before aging begins.
- Real-time aging: start date, storage location and test methods.
- Report: standard and edition, target market, and accreditation mark needed.
FAQ
Can accelerated aging done to YY/T 0681.1-2018 go straight into a US submission?
The temperature, Q10 and reporting frameworks are close; the differences are in humidity and scope. F1980-21 clause 7.4.6 says a rationale for not controlling humidity should be documented, which YY/T 0681.1-2018 does not require, so a protocol lacking one may need a supplementary explanation; FDA decides acceptance during review.
Can one report cite both standards?
One protocol can meet both texts; how the report cites them and whether separate reports are issued is agreed at ordering. The CMA mark applies only to YY/T 0681.1-2018.
Is it acceptable not to control humidity?
Neither guide makes humidity control mandatory. F1980-21 clause 7.4.6 says the rationale for not controlling it should be documented; for moisture-absorbing or hydrolysis-prone materials, set humidity conditions following its Appendix X3.
Once accelerated aging passes, is the shelf life established?
Both guides treat a shelf life from accelerated aging as tentative until real-time aging is complete; if real-time results fail, it must be cut to the longest period that successful real-time testing supports.
Can the device itself be aged under YY/T 0681.1-2018?
Its scope is the sterile barrier system and its packaging material components. F1980-21 allows its protocol guidance to be used for devices but does not define how device performance is evaluated; that comes from the product standard or technical requirements. SUNGO's CNAS and CMA schedules list accelerated aging under packaging materials for medical devices, so confirm when ordering whether device aging can be reported within accreditation.
When requesting a quotation, send the product and package format, material list, sterilization method, desired shelf life and labeled storage conditions, target market, and the standards and accreditation marks you need. The laboratory will then confirm whether the work runs in Shanghai or Hefei and which standard and edition the report will cite. Please use our contact page.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.