Paper-film pouches, film pouches made with Tyvek-type material, and clear thermoformed trays with a porous lid all raise one question during validation: does any seal contain a channel that runs right through it? The dye penetration test is one of the most common ways to find out. The ASTM method is F1929, which is listed in the US FDA Recognized Consensus Standards database; the Chinese industry standard is YY/T 0681.4-2021. This article covers how small a channel the method is designed to find, what the injection, edge dip and eyedropper methods named in F1929 clause 1.2 mean, which packages it suits, how to tell a real leak from wicking, and what to prepare before sending samples. Clause numbers refer to the publicly available parts of ASTM F1929-23: the Scope (Section 1) and Significance and Use (Section 4). The dye formulation, contact times and other procedural details sit in the body of the standard; they are not repeated here, and the published text governs.
Key points
- It looks for individual channels in a seal. Under clause 1.1, the method detects and locates leaks at least as large as a channel made by a 50 µm (0.002 in) wire, in edge seals between a transparent material and a porous sheet. YY/T 0681.4-2021 also refers to channels of 50 μm or larger.
- Three ways to apply the dye. Clause 1.2 lists injection, edge dip and eyedropper. In each case dye is applied locally to the seal edge under test, left for a specified time, and the package is examined visually for penetration.
- The material has to qualify. The porous material must hold the dye and keep it from discolouring the seal area for at least 5 seconds; uncoated papers are singled out as especially prone to leakage and need careful evaluation (clause 1.3). The dye must contrast well with the opaque material (clause 1.4).
- The result is qualitative. The method does not measure leak size and is normally run as pass/fail; components showing any sign of leakage are normally rejected (clauses 4.4 and 4.5).
- Rule out wicking before calling a leak. Given time, dye wicks through any porous material, though usually not within the maximum suggested time; staining on the porous face confirms it (clause 4.6 and Appendix X1).
- It complements seal strength. Seal strength measures force; dye penetration asks whether a channel exists.
- A CMA mark is only available under YY/T 0681.4-2021. At SUNGO Shanghai, the CNAS schedule lists ASTM F1929-23 and YY/T 0681.4-2021, the CMA schedule lists only YY/T 0681.4-2021, and the IAS scope lists ASTM F1929 and YY/T 0681.4 without edition years (confirm the edition with the laboratory before an IAS report is issued). The Hefei laboratory does not offer this test.
The standards and their scope
Sources: the ASTM product page, the US FDA Recognized Consensus Standards database and China's national standards information platform, checked on 22 October 2026.
ASTM F1929-23
- Title: Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration. Committee F02, Subcommittee F02.40.
- ASTM shows F1929-23 as the active edition; earlier editions listed are F1929-15, F1929-12, F1929-98 (reapproved 2004) and F1929-98. The changes from -15 to -23 are not visible in the public material, so none are described here.
- Scope in brief (clauses 1.1 to 1.4): edge seals between a transparent material and a porous sheet; dye applied locally to the seal edge, a specified contact time, then visual inspection; the material prerequisite and contrast requirement are compared in the table below.
- Significance and use in brief: leaks are often found at seals between package components and can also come from pinholes (4.1); the method addresses only individual leaks in a seal (4.3); it verifies and locates leak sites but is not quantitative (4.5).
- FDA recognition: F1929-23 is recognition number 14-600, entered on 29 May 2024, complete standard, Sterility specialty task group. Declarations of conformity to F1929-15 (recognition number 14-484) in support of premarket submissions are accepted until 4 July 2027. ISO 11607-1, ISO 11607-2 and ISO/TS 16775 are listed as examples under Relevant FDA Guidance and/or Supportive Publications.
YY/T 0681.4-2021
- Title: Test methods for sterile medical device package, Part 4: Detecting seal leaks in porous packages by dye penetration. A recommended industry standard, classified as a test method standard, issued by the National Medical Products Administration (NMPA).
- Published 9 March 2021, effective 1 April 2022, fully replacing YY/T 0681.4-2010. Filing number 81788-2021.
- Scope: applies to channels of 50 μm or larger in seals made from a transparent material and a porous material; not applicable to porous materials stained by capillary action within 5 seconds, or where the dye contrasts poorly with the opaque material.
- The national platform does not publish the text, citing an NMPA arrangement. The published text governs clause-level details.
| Aspect | ASTM F1929-23 | YY/T 0681.4-2021 |
|---|---|---|
| Detection capability as stated | Channels at least as large as one formed by a 50 µm (0.002 in) wire | Channels of 50 μm or larger |
| Package construction | Edge seals between a transparent material and a porous sheet | Seals made from a transparent material and a porous material |
| Material prerequisite | Holds the dye for at least 5 s without discolouring the seal area; uncoated papers need careful evaluation | Not applicable to porous materials stained by capillary action within 5 s |
| Contrast | Good contrast with the opaque material required | Not applicable where contrast is poor |
| Marks available at SUNGO Shanghai | CNAS (F1929-23); IAS (ASTM F1929, no edition year; confirm the edition before an IAS report is issued); not on the CMA schedule | CNAS and CMA (2021 edition); IAS (YY/T 0681.4, no edition year; confirm the edition before an IAS report is issued) |
The two scopes read in much the same way. Whether they match clause for clause, and what the adoption relationship is, cannot be read from the public information. Whichever standard you choose, the test is run and reported to that standard's own text.
Principle and the three application methods
Dye is applied locally to the seal edge under test, left for the specified time, and the seal is checked for dye that has crossed it (clause 1.1). Clause 4.2 puts the purpose plainly: to make channel defects visible through the dye that leaks along them. The public scope limits the method to edge seals between a transparent material and a porous sheet and requires good dye contrast with the opaque packaging material (clauses 1.1, 1.3 and 1.4); the viewing procedure is set out in the standard text.
The dye solution
The body of the standard specifies the composition and preparation of the dye penetrant; see the published text. The only dye requirement in the public scope is good contrast with the opaque packaging material (clause 1.4). In practice, check whether the colour of the porous material, or any printing, covers the seal area; YY/T 0681.4-2021 excludes cases of poor contrast. Whether a different dye or formulation is permitted is a question for the standard text, and any deviation from the standard method should be written into the protocol and the report and agreed with the client beforehand.
Three application methods
| Method | How the dye reaches the seal edge under test | Points to agree before testing |
|---|---|---|
| Injection | Dye is injected into the package with a syringe so that it reaches the seal edge under test | Needle entry point away from the seal under test; whether contents stay in the package |
| Edge dip | The seal edge under test is dipped into the dye | Which seals are tested; any opening cut away from the seal under test |
| Eyedropper | Dye is applied along the seal edge under test with a dropper | Which seals are tested; any opening cut away from the seal under test |
The table describes each method only as far as its name implies. Dye volume, package orientation, contact and viewing times, and the package formats each method suits are set out in the standard text. The protocol states which method is used; naming it in the report helps when comparing lots.
Contact and viewing time
The test is controlled by a specified contact time: apply the dye, wait, then inspect (clause 1.1). The time for each method is in the standard text. The only time value in the public text is the 5-second material prerequisite described above (clause 1.3). Clause 4.6 warns that dye eventually wicks through porous material, so keeping to the specified viewing time is the basic safeguard against false positives.
Which packages it suits: paper-film pouches, Tyvek and the exclusions
- Start with construction. The method covers edge seals between a transparent material and a porous sheet. Paper-film pouches (clear plastic film sealed to medical paper), clear film pouches made with Tyvek-type material, and clear trays with a porous lid fit that construction; the material prerequisite still has to be confirmed.
- For paper, coating matters. Clause 1.3 names uncoated papers as especially prone to leakage. For pouches made with uncoated paper, confirm the 5-second prerequisite on material from the same lot beforehand.
- Tyvek-type materials are judged the same way. The public scope names no branded material. Suitability depends on whether the material holds the dye for at least 5 seconds and whether the contrast is clear, checked on the actual material and seal.
- Where it does not apply. Packages whose seals are not formed between a transparent material and a porous sheet (outside the construction defined in clauses 1.1 and 1.3); porous materials that wick within 5 seconds; poor dye contrast.
- It does not test the material itself. Clause 4.3 notes that the procedure applies only to individual leaks in a package seal; a number of small leaks, as found in porous packaging material, will not be indicated, and such leaks could be detected by other techniques.
Interpreting results: channels, wicking and false positives
What a positive looks like
Under clause 4.2, a positive is dye that has crossed the seal along a channel, leaving a stain within the seal whose position can be pinned down. Clause 4.5 says the method locates leak sites but gives no indication of leak size. Clause 4.4 notes there is no general agreement on what level of leakage harms a given package, and that components with any indication of leakage are normally rejected. The client sets acceptance criteria in the validation protocol.
Wicking and false positives
Clause 4.6 explains that dye wicks through any porous material over time, though usually not within the maximum suggested time, and that wicking can be confirmed by looking at the porous side of the seal area, where the material surface will be stained. Appendix X1 gives further guidance on false positives. Three practical checks:
- Timing: did the discolouration appear within the specified viewing time, or only after it?
- Pattern: a channel leak is a localised path through the seal band; wicking stains the surface of the porous material.
- Porous face: turn the package over and check whether the porous surface itself is stained.
Avoid handling artefacts such as a needle entry point too close to the seal under test, or dye on the outer surface. Record the seal and position of every positive, with photographs.
How it relates to seal strength testing
Seal strength testing (ASTM F88/F88M or YY/T 0681.2-2010, both for flexible barrier materials) measures the force required to separate a test strip containing the seal (F88/F88M clause 1.4) and is a quantitative measure (clause 4.1). Dye penetration asks whether a channel runs through the seal, a qualitative result. A seal with a local channel can still reach its strength target, so neither data set replaces the other. Seal strength testing is covered in a separate article.
On SUNGO Shanghai's certificates, seal strength appears on the CNAS schedule as ASTM F88/F88M-23 and YY/T 0681.2-2010, on the CMA schedule as YY/T 0681.2-2010, and on the IAS scope as ASTM F88/F88M and YY/T 0681.2 (no edition years). Visual inspection of seal integrity is YY/T 0681.11-2014, an integrity item on the CNAS and CMA schedules, with YY/T 0681.11 on the IAS scope.
Accreditation scope table
Each row comes from the CNAS L22083 schedule, the CMA 240015340087 schedule or the IAS TL-1019 scope of Shanghai SUNGO Medical Technology Co., Ltd. None lists clauses or limitations for this test.
| Item | Standard | Clause basis | Notes / limitations |
|---|---|---|---|
| Dye penetration (product: medical device packaging materials) | ASTM F1929-23 | No clause listed on the schedule | CNAS L22083; no limitation stated; schedule effective date 2026-03-09; not on the CMA schedule |
| Dye penetration (product: medical device packaging materials) | YY/T 0681.4-2021 | No clause listed on the schedule | CNAS L22083; no limitation stated; schedule effective date 2026-03-09 |
| Dye penetration (product: medical device packaging materials) | YY/T 0681.4-2021 | No clause listed on the schedule | CMA 240015340087; no limitation stated; schedule effective date 2024-11-25; valid within the People's Republic of China only |
| Whole standard | ASTM F1929 | Whole standard, no clauses listed | IAS TL-1019; no edition year on the scope; confirm the edition with the laboratory before an IAS report is issued against it |
| Whole standard | YY/T 0681.4 | Whole standard, no clauses listed | IAS TL-1019; no edition year on the scope; confirm the edition with the laboratory before an IAS report is issued against it |
Where the test is done and under which accreditation
The test is performed at the laboratory of Shanghai SUNGO Medical Technology Co., Ltd., 958 Kangqiao East Road, Pudong New Area, Shanghai. Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity; its certificates do not include ASTM F1929 or YY/T 0681.4, and it does not offer this test.
| Report requirement | Possible basis | Available marks |
|---|---|---|
| CMA mark needed (for use within China) | YY/T 0681.4-2021 | CMA; this edition is also on the CNAS schedule; YY/T 0681.4 is on the IAS scope (no edition year; confirm the edition with the laboratory before an IAS report is issued) |
| Report to ASTM F1929-23 | ASTM F1929-23 | CNAS; IAS (ASTM F1929, edition confirmed first); no CMA mark |
| Report to another F1929 edition (for example -15) | To be confirmed with the laboratory | The CNAS schedule lists only -23; the IAS scope carries no edition year, so confirm the edition first |
A CMA mark covers only items on the CMA schedule, and the certificate states that it is valid within the People's Republic of China. The CNAS certificate states that CNAS is a signatory to the ILAC and APAC mutual recognition arrangements, and the IAS certificate carries the ilac-MRA mark. Whether a report is accepted by a given regulator or customer is for that party to decide; SUNGO makes no promise on acceptance.
Preparing a submission
- Standard, edition and mark. State whether the test is to ASTM F1929-23 or YY/T 0681.4-2021, which mark you need (CNAS, CMA or IAS), and what the report is for.
- The package. The transparent material, the porous material (coated or uncoated paper, grade of any Tyvek-type material), the seal locations and number of seal edges, and the format (pouch, tray or other).
- Sample condition. Whether samples are taken after sterilization, accelerated aging or simulated transport is set by the validation protocol. On sterilization and sampling, see Sterile Barrier Systems: Packaging and Sterilization Validation; for accelerated aging, see shelf-life validation. Any pre-test conditioning follows the standard text and the protocol.
- Protect the seals. Avoid crushing or folding seal edges in transit and keep the seal area clean.
- Method and seals. The client specifies the application method and the seals to be tested, or the laboratory proposes them and they are confirmed in the protocol, including whether contents stay in the package.
- Check suitability first if in doubt. For uncoated paper or dark or printed porous material, send extra material from the same lot or blank sealed specimens so the 5-second prerequisite and the contrast can be checked.
- Acceptance criteria. The client writes them into the protocol. Tested packages have been in contact with dye and cannot be used as product; sample numbers follow the protocol or sampling plan.
FAQ
How small a leak can dye penetration detect?
F1929 states that it detects and locates leaks at least as large as a channel formed by a 50 µm (0.002 in) wire; YY/T 0681.4-2021 refers to channels of 50 μm or larger. The method is not quantitative, so a positive says nothing about channel size.
Can a report to ASTM F1929 carry a CMA mark?
No. SUNGO Shanghai's CMA schedule lists only YY/T 0681.4-2021. ASTM F1929-23 can be reported under CNAS, and ASTM F1929 under IAS once the edition has been confirmed with the laboratory.
The whole porous side has changed colour. Is that a leak?
Not necessarily. Under clause 4.6, wicking stains the surface of the porous material and can be confirmed from the porous side, whereas a channel leak is a localised path through the seal band. A material that wicks within 5 seconds is not suitable for this method at all.
Our US file cites F1929-15. Can we still use it?
The FDA database states that declarations of conformity to F1929-15 in support of premarket submissions are accepted until 4 July 2027, and not after the transition period. SUNGO Shanghai's CNAS schedule lists only F1929-23; the IAS scope carries no edition year, so confirm the edition with the laboratory before an IAS report is issued against it.
If we run dye penetration, do we still need seal strength?
One looks for channels, the other measures the force needed to separate the seal, and neither replaces the other. Which tests are run is set by the packaging validation protocol.
Enquiries
When requesting a quotation, please give the package format and both materials, the standard and edition, the mark required (CNAS, CMA or IAS), the sample condition (after sterilization, aging or transport), whether the application method and seals to be tested are specified, and whether acceptance criteria have been set. The laboratory confirms the protocol on that basis before samples are sent. For the service scope, see Sterile Medical Device Packaging Validation; to get in touch, see Contact us.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.