In China, sticking plasters, IV dressings and incision dressings built from an adhesive layer, an absorbent pad and a release liner usually take YY/T 1627-2018, General requirements of acute adhesive wound dressing and woundplast, as the main reference. Manufacturers writing product technical requirements usually ask three things: which items the standard calls for, how holding power and peel strength are tested and why in-house passes turn into lab failures, and which items SUNGO Shanghai's CNAS scope and CMA schedule each cover. This article follows YY/T 1627 as the product standard; YY/T 0148 and YY/T 0471 are discussed only where YY/T 1627 cites them.

The short answer

  • YY/T 1627-2018 is a recommended pharmaceutical industry standard, issued 2018-11-07, implemented 2019-11-01, status current. It applies to acute wound dressings and woundplasts whose main structure is an adhesive layer, an absorbent pad and a release liner. Hydrocolloid products and products containing antibacterial or anti-inflammatory substances such as silver or chitosan are excluded.
  • It sets general requirements only. Water resistance, water vapour permeability, comfort (conformability), elasticity and fluid absorption use methods from YY/T 0471.3, YY/T 0148 Annex C, YY/T 0471.4, YY/T 0148 Annex F and YY/T 0471.1.
  • Holding power and peel strength are tested to Annex B of YY/T 1627 itself (B.1.2 and B.1.3), which is what SUNGO's CNAS scope records, not YY/T 0148. Both are sensitive to the adhesive layer, the release liner, temperature and humidity, and dwell times. Common causes of disagreement between in-house and lab results include sample condition and timing.
  • EO residue is a dated reference to GB/T 14233.1-2008. That edition has been replaced by the 2022 edition, but testing to YY/T 1627 still uses the 2008 edition; SUNGO's CNAS scope records its Clause 9 (gas chromatography).
  • SUNGO Shanghai's CNAS scope (L22083) and CMA schedule (240015340087) each list 15 YY/T 1627-2018 items, but not the same 15: CNAS has water-soluble substances and not bioburden; CMA has bioburden and not water-soluble substances. YY/T 1627 is not in the IAS scope. The Hefei laboratory does not test these products.
  • YY/T 1627 cites GB/T 16886.1 (biological evaluation); biological tests such as cytotoxicity, sensitization and irritation are outside the scope of SUNGO's certificates. SUNGO subcontracts them to appropriately qualified laboratories; under accreditation rules, the client's consent is obtained before subcontracting, and the report identifies the subcontracted items and the subcontractor.

The standard and its scope

China's national standards information platform lists YY/T 1627-2018 as a current recommended pharmaceutical industry standard (ICS 11.120.20), issued 2018-11-07 and implemented 2019-11-01, administered by the national technical body for medical hygiene materials and dressings.

The scope published by the National Medical Products Administration (NMPA) covers dressings and woundplasts for acute body-surface wounds such as surgical incisions, puncture sites and mechanical wounds, where the main structural feature is an adhesive layer, an absorbent pad and a release liner. It does not apply to hydrocolloid dressings or woundplasts, or to products containing specific antibacterial or anti-inflammatory substances (silver, chitosan and the like).

Dressing versus woundplast

The standard defines an acute wound as one caused by external force that, with proper treatment, is expected to heal within two weeks, including sutured surgical wounds, mechanical trauma, catheter insertion sites, abrasions, stab wounds and cuts. The introduction splits the products by setting. A dressing is used in healthcare facilities on wounds that have been cleaned, disinfected or sutured, to keep the wound sterile and support healing. A woundplast is used outside healthcare facilities as a temporary covering for accidental wounds that have not been disinfected or sutured, providing temporary closure, haemostasis and isolation. The two terms apply only within this standard, which sets no separate specific requirements for either type.

Checking your own product

Check three things: the type of wound, whether the structure is mainly adhesive layer, absorbent pad and release liner, and whether either exclusion applies. IV dressings, incision dressings and sticking plasters that match and are not excluded usually use YY/T 1627 as the main reference. Products without an absorbent pad do not match the main structural feature, and plain medical tape generally falls under YY/T 0148. The technical requirements and the reviewer's opinion decide the final position.

Normative references

Dated references are GB/T 14233.1-2008, YY/T 0148-2006, YY/T 0471.1-2004, YY/T 0471.3-2004 and YY/T 0471.4-2004; under Clause 2, only the dated edition applies. Undated references are GB/T 16886.1 (biological evaluation), GB/T 19633.1 (packaging for terminally sterilized devices), YY/T 0466.1 (labelling symbols) and YY/T 0615.1 (devices labelled "sterile"), for which the latest edition applies.

How the test items group together

Using the clause numbers on SUNGO Shanghai's CNAS scope (the CMA schedule lists items without clauses):

  • Appearance, dimensions, design: 6.1, 6.2, 6.3.
  • Adhesion: holding power 6.4.2 (Annex B.1.2); peel strength 6.4.3 (Annex B.1.3).
  • Fluid handling and barrier: water resistance 6.4.4; water vapour permeability 6.4.5; fluid absorption 6.4.8.
  • Comfort (conformability) and elasticity: 6.4.6 and 6.4.7.
  • Chemical properties: water-soluble substances 6.5.1 (Annex C.1); acidity/alkalinity 6.5.2 (Annex C.2); surface-active substances 6.5.3 (Annex C.3).
  • EO residue: no YY/T 1627 clause number on the certificate; the method recorded on the CNAS scope is GB/T 14233.1-2008 Clause 9.
  • Marking: Clause 7.

Water resistance, water vapour permeability, comfort and elasticity are often included only when the product claims them. The Beijing Municipal Medical Products Administration's 2017 technical review specification for IV dressings and sterile woundplasts says these properties are assessed when claimed. For how YY/T 1627 words the applicability of each item, refer to the standard text.

Holding power (6.4.2, Annex B.1.2)

Holding power measures how well the adhesive resists sliding under a sustained load, mainly reflecting its cohesion and its bond to the test surface. The basic form of the test can be seen in YY/T 0148-2006, which YY/T 1627 cites: its Clause 5.2.1 refers to test B.2 in its Annex B and looks at how far the top of the tape, applied to a stainless steel plate, slides down during the oven test. YY/T 1627 has its own method in Annex B.1.2, and SUNGO's CNAS scope records "6.4.2, Annex B.1.2". Specimen size, load, temperature, time and acceptance values should be taken from the YY/T 1627 text; this article does not reproduce them.

The laboratory controls these points:

  1. Specimens: cut from the finished adhesive layer as the standard requires, without stretching or creasing edges or touching the adhesive face.
  2. Test plate: cleaned to procedure before each use and checked for residue and oil, since cleanliness directly affects the bond.
  3. Application: after liner removal, pressed down as the standard specifies, the same way every time, with no trapped air.
  4. Dwell and loading: dwell for the specified time before loading and placing in the oven; actual dwell time and oven temperature are recorded.
  5. Reading and failure mode: besides slide distance, the lab notes whether the strip slid as a whole, left adhesive on the plate or the backing stretched.

Peel strength (6.4.3, Annex B.1.3)

Peel strength is the force needed to peel the specimen off the test plate. When the 2017 Beijing specification cites YY/T 0148-2006, it describes peel strength as the average force required per 1 cm width of tape, that is, average force per unit width. The YY/T 1627 method is in Annex B.1.3; peel angle, speed, specimen length, how readings are taken and how the result is expressed should be taken from the standard text.

A peel trace is rarely flat. An elastic backing stretches first and the force rises and falls with it. The area under the absorbent pad does not bond to the plate, so specimens avoid it or are handled as the standard specifies. The lab also records adhesive left on the plate, which points to weak cohesion, and backing breaks, which mean the reading no longer represents peel strength.

Why peel results do not repeat between laboratories or over time is discussed from the medical tape (YY/T 0148) angle in a separate article on this site. Below are four common causes from the YY/T 1627 product perspective.

Why holding power and peel strength fail

The adhesive layer

Uneven or thin coat weight lowers peel strength. A soft formulation or too little crosslinking means poor cohesion: the strip slides in the holding power test and leaves residue after peeling. An adhesive that is too hard or overcrosslinked may hold well but fail on peel. Temperature, humidity or radiation during sterilization can also change the adhesive, so a pass before sterilization does not guarantee a pass afterwards. Submit sterilized finished product in its marketed condition.

The release liner

The liner is not graded on its own, but its effects show in adhesion results. High release force can roughen or distort the adhesive face; release agent migrating into the adhesive reduces bonding; release force can change after long contact. The scope of YY/T 0148-2006 expressly excludes requirements for the protective layer on the adhesive face, so liner quality is mainly controlled at incoming inspection. The 2017 Beijing specification notes that IV dressings and sterile woundplasts are usually cut and assembled from purchased backing, adhesive layer, absorbent layer and release paper or film. Re-evaluate adhesion when changing liner supplier.

Temperature and humidity

Pressure-sensitive adhesives are viscoelastic: as temperature rises they soften, and slide in the holding power test usually increases. Humidity affects absorbent backings, the pad and paper liners. Clause 4 of YY/T 0148-2006 sends conditioning and test atmosphere to its Annex A; for YY/T 1627, follow the standard text. Heat in transit or storage can also shift results.

Dwell and elapsed time

Three periods matter: manufacture to test (adhesive properties may still be changing), liner removal to application, and application to loading or peeling. As a general rule for pressure-sensitive adhesives, longer dwell lets the adhesive wet the surface more fully, and peel force tends to rise. If in-house tests skip the specified dwell, or use product fresh off the line while the lab receives product stored for months, results will be hard to reconcile.

Items that borrow methods from YY/T 0471 and YY/T 0148

  • Water resistance (6.4.4): method from YY/T 0471.3-2004.
  • Water vapour permeability (6.4.5): method from YY/T 0148-2006 Annex C (a normative annex of YY/T 0148); the test specimen and requirement are as stated in YY/T 1627.
  • Fluid absorption (6.4.8): method from YY/T 0471.1-2004; the CNAS scope records Clause 3.2 (free swell absorptive capacity).
  • Comfort (conformability) (6.4.6): method from YY/T 0471.4-2004.
  • Elasticity (6.4.7): method from YY/T 0148-2006 Annex F.

Because these references are dated, testing to YY/T 1627 uses the 2004 and 2006 editions named.

Chemical properties and EO residue

Water-soluble substances, acidity/alkalinity, surface-active substances

These are Clauses 6.5.1 to 6.5.3, with methods in Annex C of YY/T 1627 (C.1 to C.3). Test conditions and acceptance criteria should be taken from the standard text.

EO residue

Products sterilized with ethylene oxide need an EO residue test. YY/T 1627 makes a dated reference to GB/T 14233.1-2008. The national standards platform shows GB/T 14233.1-2022 was issued 2022-10-12, took effect 2023-11-01 and fully replaces the 2008 edition. Clause 2 of YY/T 1627 says only the dated edition applies, so testing to YY/T 1627 still uses the 2008 edition, in which Clause 9 is the gas chromatography method and Clause 10 a colorimetric method. SUNGO's CNAS scope records Clause 9 with the limitation "only where cited by YY/T 1627-2018"; the CMA schedule lists GB/T 14233.1-2008 without the clause, with the same restriction in its remarks.

The result reflects the level at the time of sampling and testing; sterilization date, completion of aeration and transit time all affect it. The Clause 9 method itself is covered in a separate article.

Where the CNAS scope and CMA schedule differ

Under the test object "adhesive dressings and woundplasts", 14 of the 15 YY/T 1627-2018 items are on both the CNAS scope and the CMA schedule. Two differ:

  • Water-soluble substances: on CNAS (6.5.1, Annex C.1), not on CMA, so it can only be reported under CNAS.
  • Bioburden: on the CMA schedule under YY/T 1627-2018, not on CNAS. CMA gives no clause number; refer to the standard text.

The CNAS scope gives clause numbers for most items (not for the YY/T 1627 EO residue item); the CMA schedule lists only items and standard numbers. Which mark a report can carry depends on whether each item is within that certificate's scope. For what each certificate means and how to verify it, see SUNGO Medical Device Testing Scope: CNAS, CMA and IAS in Shanghai and Hefei.

Scope table: CNAS scope and CMA schedule

All rows come from the CNAS L22083 accreditation scope and the CMA 240015340087 qualification schedule of Shanghai SUNGO Medical Technology Co., Ltd., test object "adhesive dressings and woundplasts". Clause references follow the CNAS scope.

ItemStandard and clauseNotes / limitations
AppearanceYY/T 1627-2018 6.1CNAS, CMA
DimensionsYY/T 1627-2018 6.2CNAS, CMA
DesignYY/T 1627-2018 6.3CNAS, CMA
Holding powerYY/T 1627-2018 6.4.2, Annex B.1.2CNAS, CMA
Peel strengthYY/T 1627-2018 6.4.3, Annex B.1.3CNAS, CMA
Water resistanceYY/T 1627-2018 6.4.4; YY/T 0471.3-2004CNAS, CMA
Water vapour permeabilityYY/T 1627-2018 6.4.5; YY/T 0148-2006 Annex CCNAS, CMA (CMA lists YY/T 0148-2006 without the annex)
Comfort (conformability)YY/T 1627-2018 6.4.6; YY/T 0471.4-2004CNAS, CMA
ElasticityYY/T 1627-2018 6.4.7; YY/T 0148-2006 Annex FCNAS, CMA (CMA lists YY/T 0148-2006 without the annex)
Fluid absorptionYY/T 1627-2018 6.4.8; YY/T 0471.1-2004 3.2CNAS, CMA (CMA lists YY/T 0471.1-2004 without the clause)
Water-soluble substancesYY/T 1627-2018 6.5.1, Annex C.1CNAS only; not on the CMA schedule
Acidity/alkalinityYY/T 1627-2018 6.5.2, Annex C.2CNAS, CMA
Surface-active substancesYY/T 1627-2018 6.5.3, Annex C.3CNAS, CMA
EO residueYY/T 1627-2018 (no clause number on the certificate); GB/T 14233.1-2008 9CNAS, CMA (CMA lists GB/T 14233.1-2008 without the clause); GB/T 14233.1-2008 only where cited by YY/T 1627-2018
BioburdenYY/T 1627-2018 (no clause number on the CMA schedule)CMA only; not on CNAS
MarkingYY/T 1627-2018 7CNAS, CMA

Biological tests such as cytotoxicity, sensitization and irritation are not in this table and are not within the scope of any SUNGO certificate (of the GB/T 16886 / ISO 10993 biological evaluation series, SUNGO's certificates cover only Part 7, EO sterilization residuals).

Where the work is done and which certificate applies

  • Laboratory: Shanghai SUNGO Medical Technology Co., Ltd., at No. 958 Kangqiao East Road, Pudong, Shanghai. Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity whose scope does not include dressings.
  • CNAS (L22083): every item in the table except bioburden. CNAS is an ILAC and APAC MRA signatory.
  • CMA (240015340087): every item except water-soluble substances. CMA covers only items on its schedule and is valid only within China.
  • IAS (TL-1019): does not include YY/T 1627, so no IAS report can be issued against it. The IAS scope separately lists YY/T 0471.1, YY/T 0471.4 and GB/T 14233.1 Clause 9, without edition years; for an IAS report to one of these methods on its own, confirm the edition with the laboratory first.
  • Biological tests (cytotoxicity, sensitization, irritation and so on): subcontracted by SUNGO to appropriately qualified laboratories; under accreditation rules, the client's consent is obtained before subcontracting, and the report identifies the subcontracted items and the subcontractor.

See the mask, glove and dressing testing service. Sterile barrier packaging tests and accelerated aging are separate service lines: sterile packaging validation and shelf-life validation.

Preparing samples

  • Item list: send the technical requirements or intended clause list, mark claimed properties (water resistance, breathability, comfort, elasticity), and say whether each item needs a CNAS or CMA report.
  • Sample condition: finished product in marketed condition, with liner and unit packaging; for sterile products, sterilized product with the sterilization method and date.
  • Sizes and representativeness: small woundplasts may not have enough adhesive area for the Annex B specimens; agree the sampling approach in advance, and it will be stated in the report. The 2017 Beijing specification says products with different backing, absorbent layer or adhesive materials should be tested separately, a useful guide when choosing models.
  • Transport and storage: original packaging, no heat, pressure or damp; tell the lab the manufacturing date and storage conditions.
  • EO residue: give the sterilization date and aeration status, and keep transit short.
  • Quantity: depends on items and product size; the laboratory confirms after review.

Frequently asked questions

Are IV dressings and sticking plasters all tested to YY/T 1627?

If the main structure is adhesive layer, absorbent pad and release liner, the product is for acute body-surface wounds, and it is neither hydrocolloid nor contains silver, chitosan or similar substances, YY/T 1627 is usually the main reference. How it is cited is set in the technical requirements.

Holding power passes in-house but fails at the lab. What should we check first?

Align the samples first: sterilized or not, storage time, heat in transit. Then the procedure: conditioning, application, dwell time and test temperature, all as the standard specifies.

Why not use GB/T 14233.1-2022 for EO residue?

YY/T 1627 makes a dated reference to GB/T 14233.1-2008, and its Clause 2 says only the dated edition applies. SUNGO's CNAS and CMA entries for GB/T 14233.1-2008 are also restricted to use where cited by YY/T 1627-2018.

Can the report be used for registration outside China?

CNAS is an ILAC and APAC MRA signatory, CMA is valid only within China, and YY/T 1627 is not in the IAS scope. Whether an overseas authority accepts the report is for the client to confirm with that authority; SUNGO makes no commitment on acceptance.

Quotes and sample submission

When asking for a quote, tell us the product structure (materials of the adhesive layer, pad and liner), the sterilization method, the items and report type you need, and which properties the product claims. The laboratory will check each item against the CNAS scope and CMA schedule and explain sample requirements. Please use the contact page.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.