Paper/film pouches, film/film pouches and blister trays with lidding all have at least one transparent side, so the first check for channels or incomplete seals is usually by eye. The ASTM method is F1886/F1886M; the Chinese industry standard is YY/T 0681.11-2014, Test methods for sterile medical device package – Part 11: Determining integrity of seals for medical packaging by visual inspection. This article covers viewing conditions, the channel size visual inspection can find, how to judge seal features, the split between in-line inspection and laboratory testing, how it complements dye penetration, and which standard to choose at SUNGO for the mark you need. Unless stated otherwise, F1886 clause numbers refer to the public sample of the -16 text; for -25 we cite only the public scope and table of contents.
Key Points
- Detection capability: F1886-16 clause 1.1 covers channels down to a width of 75 µm (0.003 in.) with a 60–100 % probability; YY/T 0681.11-2014 states the same. This is a probability, not a guarantee; contrast, adhesive, reflecting light angle, materials and inspector experience all matter (1.1.1).
- Viewing conditions are specified: about 540 lumens/m² [50 fc] of white light or daylight on the specimens (6.1); visual acuity to inspect at 30 to 45 cm (7.1); magnification to characterize defects already identified (Note 1); a UV light box can enhance contrast for some materials (Note 3).
- Qualitative accept/reject (4.1): channels are recorded, marked and counted (7.3, 7.4); other seal characteristics follow user-defined criteria (Note 4).
- No substitute for leak testing: dye penetration (ASTM F1929, YY/T 0681.4) complements it for porous packages. The bubble test (ASTM F2096, YY/T 0681.5-2010) is not within SUNGO's accredited scope.
- Edition: as checked on 2026-10-22, the active ASTM edition is F1886/F1886M-25, retitled and extended to material surfaces; the FDA Recognized Consensus Standards database, searched the same day, lists -16.
- Mark: for CNAS or CMA, test to YY/T 0681.11-2014; a report to ASTM F1886/F1886M can only carry the IAS mark. The IAS scope gives no year and carries the seal-inspection title, so confirm the edition with the laboratory in advance. Both are done in Shanghai.
Standards and Scope
Sources: the ASTM website, the public sample of the standard text, the BSI table of contents, the FDA database and Chinese standards information platforms, checked on 2026-10-22.
ASTM F1886/F1886M
- The -09, -16 and -25 editions are under Committee F02, Subcommittee F02.40 (Package Integrity); ASTM shows the -98 edition as developed by Subcommittee F02.60. ASTM lists F1886-98, -98 (reapproved 2004), F1886/F1886M-09, -09 (reapproved 2013), -16 and -16 (reapproved 2024); as checked on 2026-10-22, the active edition is -25.
- Titles: -98 "…Seals for Medical Packaging by Visual Inspection"; -09 and -16 "…Seals for Flexible Packaging…"; -25 "Standard Test Method for Determining Integrity of Primary Packaging by Visual Inspection".
- -16 scope: channels down to 75 µm width with a 60–100 % probability; packages with at least one transparent side so that the seal area may be clearly viewed (1.2).
- -25 scope: a non-destructive visual evaluation of peelable seals, non-peelable seals and packaging materials, for bags/pouches or trays. Part 1 inspects seals for channels (at least one transparent side); Part 2 inspects opaque and transparent material surfaces for suspected integrity defects. Both parts make the user responsible for defining the detectable defect size during method validation (Part 2 references TMV Guide F3263). Note 1: observations are indications that there could be an issue; further testing may be required to confirm an integrity breach.
- The -25 table of contents (BSI) adds a Sampling section and keeps a separate Precision and Bias section for Part 1; refer to the text for content.
- FDA database: F1886/F1886M-16, recognition number 14-501, entered 2017-08-21, complete standard, Sterility task group; relevant guidance includes ISO 11607-1, ISO 11607-2 and ISO/TS 16775. A search for F1886 returned only this entry.
YY/T 0681.11-2014
- Recommended pharmaceutical industry standard (method standard); responsible authority NMPA; technical committee SAC/TC 106. Issued 2014-06-17, effective 2015-07-01, a new standard rather than a revision; filing number 85876-2022; current.
- Scope, translated: terminally sterilized flexible and rigid packages with at least one transparent side; a test method that can determine channels of 75 µm width and above with a 60%–100% probability.
- The National Library of Standards records it as non-equivalent (NEQ) to ASTM F1886-1998, so its specific requirements follow the YY/T text, not any F1886 edition. The industry standards platform does not publish the text ("arranged by the NMPA").
| Item | F1886/F1886M-16 | F1886/F1886M-25 | YY/T 0681.11-2014 |
|---|---|---|---|
| Inspected | Seals | Seals; material surfaces | Seals |
| Package requirement | At least one transparent side | Seals: one transparent side; surfaces: opaque or transparent | Terminally sterilized flexible and rigid packages, one transparent side |
| Stated capability | Down to 75 µm, 60–100 % | Defined by the user in method validation | 75 µm and above, 60%–100% |
| Lighting and distance | About 540 lumens/m² [50 fc]; 30 to 45 cm | Refer to the text | Refer to the text |
| SUNGO Shanghai certificates | IAS scope lists ASTM F1886/F1886M with no year, under the same title as -09 and -16 ("…Seals for Flexible Packaging…"); confirm the edition with the laboratory before an IAS report | The IAS scope carries the seal-inspection title, not the -25 title; whether an IAS report can follow -25, in particular Part 2 material surfaces, must be confirmed with the laboratory | CNAS and CMA schedules: 2014 edition; IAS scope lists YY/T 0681.11 with no year |
Viewing Conditions: Lighting, Distance, Magnification and UV
Sections 6 and 7 of F1886-16 are specific. For YY/T 0681.11-2014 and F1886-25, refer to their own texts.
| Element | F1886-16 text | In practice |
|---|---|---|
| Lighting | 6.1: about 540 lumens/m² [50 fc] of white light or daylight on the specimens | Measured at the specimen |
| Marking | 6.2: indelible marking pen | Marks channel locations |
| Distance | 7.1: visual acuity such that the seal can be inspected at 30 to 45 cm [12 to 18 in.] | A requirement on the inspector's vision |
| Magnification | Note 1: devices such as eyeloops may be used as an analytical tool to characterize identified seal defects | The -16 apparatus lists only the illuminant and pen |
| Coverage and time | 7.2: inspect the entire sealed area for completeness and uniformity; Note 2: any time requirement should allow complete seal inspection | A line time limit must allow the full perimeter |
| UV light | Note 3: some packaging materials and adhesives may fluoresce under ultraviolet light; viewing the seal area in a UV light box will enhance sealed-to-unsealed contrast | Material-dependent; check first |
| Peel confirmation | Note 5: peelable packages may be peeled apart and the transferred adhesive inspected; cool heat seals to ambient before peeling | Destructive; agree in advance |
How Small a Channel? Reading "75 µm" and "60–100 %"
ASTM F17-20 defines a channel as any unimpaired pathway across the entire width of the intended seal. A local defect that does not cross the full width is not a channel.
Section 8 of F1886-16 explains the 75 µm figure. A 1997 round robin involved ten laboratories and four package types, with channels randomly produced using 75 µm and 125 µm diameter wire and negative controls without channels. The packages were: open pouch, clear film/coated paper; open pouch, clear film/uncoated Tyvek; sealed blister, blue tinted blister/coated Tyvek; open end pouch, clear film/clear film. Results were pass or fail. Table 3 gives the percent incorrect by defect type:
| Defect type | Samples inspected | Incorrect | Percent incorrect |
|---|---|---|---|
| No channels (negative control) | 370 | 20 | 5.41% |
| 75 µm channel | 400 | 83 | 20.75% |
| 125 µm channel | 400 | 23 | 5.75% |
By material (Table 2), errors ranged from 4.67% for film/paper to 20.34% for PETG/Tyvek; by laboratory (Table 1), from 0.00% to 22.22%. Three practical points:
- 75 µm is near the lower edge, not a guaranteed threshold. Roughly one judgement in five was wrong at 75 µm; far fewer at 125 µm.
- Materials matter. The two Tyvek combinations had higher error rates than film/paper and film/film, consistent with clause 1.1.1.
- Train and validate. The -25 edition leaves detectable size to the user's method validation. Where in-line inspection relies on the eye, qualify inspectors and viewing conditions with samples containing known defects.
Classifying and Judging Seal Features
F1886-16 prescribes actions only for channels: record and mark them (7.3) and record their number and location on each package (7.4). Other seal characteristics are categorized against user-defined accept/reject criteria, with defined actions when found in production (Note 4). Note 4 refers to Appendix X1; see the original for its content. So apart from channels, acceptability is set by the manufacturer's protocol or procedure, not by the standard.
In the table, "channel", "unsealed/incomplete seal" and "completeness and uniformity" come from the F1886-16 text and F17. The other terms are common industry usage illustrating the judging logic, not a restatement of Appendix X1.
| Feature | What to look for | Judging logic | Basis |
|---|---|---|---|
| Channel | Unsealed pathway across the full seal width | Record, mark and count each; criteria must be stated in the protocol | F17; 7.3, 7.4, Note 5 |
| Unsealed / incomplete seal | A section not bonded, or a discontinuous seal | Not a channel unless it crosses the full width, but check the remaining effective seal width; peelable packages can be peeled to check adhesive transfer | 7.2, Note 5 |
| Non-uniform seal | Uneven shade or transparency | Often an early sign of sealing variation; classify and feed back to the process | 5.2, 7.2, Note 4 |
| Wrinkle | Creased or folded material in the seal | Treat as a channel if it crosses the seal width; otherwise apply the criteria | F17; Note 4 |
| Overseal | Visibly deformed or damaged seal material | Check for damage and effect on opening | Note 4 |
| Bubble | Trapped air or voids | Treat as a channel if they join across the seal; otherwise apply the criteria | F17; Note 4 |
| Foreign matter | Particles or fibres in the seal | Check for an unsealed pathway around it and contamination risk | F17; Note 4 |
| Narrow seal width | Effective width below the manufacturer's requirement | Judge against the manufacturer's minimum seal width | Note 4 |
Note 5 adds that incomplete adhesive transfer is not conclusive evidence of an unmade seal, because the coating may have a stronger affinity for its own substrate; a physical seal integrity test may be needed. Note 1 of -25 makes the same point: what the eye sees is an indication.
In-Line Inspection and Laboratory Testing
Visual inspection does not open or damage the package: F1886-16 clause 4.1 evaluates unopened, intact seals, and -25 clause 1.1 calls it non-destructive. It can be applied to each package on the line and also run as a laboratory method, but the purposes differ:
| Item | In-line inspection | Laboratory test to F1886 or YY/T 0681.11 |
|---|---|---|
| Purpose | Catch sealing variation early and remove defective packages | Controlled, documented results for validation and regulatory files |
| Items | Routine production; 100% or sampling per the manufacturer's quality system | Protocol specimens, for example after sterilization, aging or transport |
| Judgement | The manufacturer's defect classes and criteria, with defined actions (Note 4) | Protocol criteria; each channel recorded by location and number |
| Output | Production records, nonconforming product handling | A test report with the CNAS, CMA or IAS mark, depending on the standard |
F1886-16 clause 5.2 says visual seal defects will often be the first indication of heat-sealing process variation, and will also indicate a lack of, or potential compromise to, package integrity after physical package performance testing. The first half is the production line's job; the second covers re-inspection after aging or transport. A laboratory evaluates the specimens it receives and cannot replace the manufacturer's sealing process validation, which (ISO 11607-2, GB/T 19633.2) is not within SUNGO's accredited scope.
Visual Inspection, Dye Penetration and the Bubble Test
Visual inspection can miss defects and does not replace leak testing. A common approach is to inspect visually first, then run leak or strength tests on the same lot or separate specimens, as the protocol specifies.
| Method | Stated capability | Packages | Effect on package | SUNGO accreditation |
|---|---|---|---|---|
| Visual (ASTM F1886/F1886M, YY/T 0681.11-2014) | Channels of 75 µm and above, 60–100 % (-16 and YY/T) | At least one transparent side | Non-destructive | Shanghai: YY/T 0681.11-2014 on CNAS and CMA schedules; ASTM F1886/F1886M and YY/T 0681.11 on IAS scope |
| Dye penetration (ASTM F1929-23, YY/T 0681.4-2021) | A leak equal to or greater than a channel formed by a 50 µm wire (F1929 1.1) | Seals between a transparent material and a porous sheet | Dye applied locally to the seal edge under test (F1929 1.1); specimen disposition per protocol | Shanghai: CNAS (F1929-23, YY/T 0681.4-2021); CMA (YY/T 0681.4-2021); IAS (ASTM F1929, YY/T 0681.4) |
| Bubble test (ASTM F2096, YY/T 0681.5-2010) | Gross leaks (per the title) | Tray and pouch packages (F2096 1.1) | Destructive; requires entry to pressurize (F2096 1.3) | Not within SUNGO's accredited scope |
Visual inspection applies to packages with at least one transparent side, including film/film, and does not damage the package. Dye penetration applies only to transparent/porous seals and states a smaller detectable size, but dye is applied to the seal edge; whether the same specimen can go on to other tests is set by the protocol. Dye penetration is covered in a separate article.
Accreditation Scope Table
Rows are taken from the CNAS L22083 schedule, the CMA 240015340087 schedule and the IAS TL-1019 scope of Shanghai SUNGO Medical Technology Co., Ltd. None lists clauses or limitations for this item.
| Item | Standard | Clause | Notes / limitations |
|---|---|---|---|
| Integrity (product: medical device packaging materials) | YY/T 0681.11-2014 | No clause listed | CNAS L22083; no limitation stated; schedule effective 2026-03-09 |
| Integrity (product: medical device packaging materials) | YY/T 0681.11-2014 | No clause listed | CMA 240015340087; no limitation stated; schedule effective 2024-11-25; valid only within the People's Republic of China |
| Whole standard | ASTM F1886/F1886M | Whole standard, no clauses listed | IAS TL-1019; no year given; title on the scope: Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection; effective 2026-02-19; confirm the edition with the laboratory before an IAS report |
| Whole standard | YY/T 0681.11 | Whole standard, no clauses listed | IAS TL-1019; no year given; effective 2026-02-19; confirm the edition with the laboratory before an IAS report |
Where SUNGO Performs It and Under Which Accreditation
The test is performed by the laboratory of Shanghai SUNGO Medical Technology Co., Ltd., No. 958 Kangqiao East Road, Pudong New Area, Shanghai. Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity whose certificates do not include ASTM F1886 or YY/T 0681.11; it does not perform this test.
| Report need | Standard | Mark available |
|---|---|---|
| CMA mark (use in China) | YY/T 0681.11-2014 | CMA; this edition is also on the CNAS schedule, and the IAS scope lists YY/T 0681.11 |
| CNAS mark | YY/T 0681.11-2014 | CNAS |
| Report to ASTM F1886/F1886M | ASTM F1886/F1886M | IAS only, no CNAS or CMA mark. The IAS scope gives no year and carries the seal-inspection title ("…Seals for Flexible Packaging…"); confirm in advance whether -16 or -25 applies, and in particular whether -25 Part 2 material-surface inspection can be included |
The CMA mark applies only to items on the CMA schedule, and the certificate states it is valid within the People's Republic of China. The CNAS certificate states that CNAS is a member of the ILAC and APAC mutual recognition arrangements; the IAS certificate carries the ilac-MRA mark. Whether a report is accepted by a given regulator or customer is their decision; SUNGO makes no commitment on acceptance.
Preparing Samples
- Standard, edition and mark. State YY/T 0681.11-2014 or ASTM F1886/F1886M (which edition), the mark needed and the report's purpose.
- Package description. Format; materials on each side; peelable or non-peelable; seal positions and number; any seal-width requirement you have set.
- Specimen condition per the protocol. After sterilization, aging or transport, as the protocol defines; see Sterile Barrier Systems: Packaging and Sterilization Validation, Shelf Life Validation and Medical Device Simulated Transport Validation.
- Protect the seals. Do not crush or fold seal edges, or write on or label the seal area, so marks are not confused with defects.
- Criteria. Provide defect classes and acceptance criteria for features other than channels, and say whether peel confirmation is allowed.
- Follow-on tests. If the same specimens go on to dye penetration or seal strength, put the sequence in the protocol. Specimen numbers follow the protocol or sampling plan.
FAQ
How small a channel can visual inspection find?
F1886-16 and YY/T 0681.11-2014 both state channels of 75 µm width and above with a 60–100 % probability; F1886-25 leaves the detectable size to the user's method validation. In the -16 round robin, 75 µm channels were misjudged far more often than 125 µm channels.
Can a report to ASTM F1886 carry the CMA mark?
No. ASTM F1886/F1886M is only on SUNGO Shanghai's IAS scope. For CNAS or CMA, test to YY/T 0681.11-2014.
If seals pass visual inspection, is leak testing still needed?
Visual inspection can miss defects and does not replace leak testing. Dye penetration is a common complement for porous packages; the protocol decides which tests to run.
As of 2026-10-22, F1886 is at -25. Which edition for a US submission?
As searched on 2026-10-22, the FDA database lists F1886/F1886M-16 (recognition number 14-501). SUNGO's IAS scope gives no year and carries the seal-inspection title, so confirm with the laboratory which edition an IAS report will follow; acceptance is FDA's decision.
How is a suspected unsealed area on a peelable package confirmed?
Under F1886-16 Note 5, peel the package fully apart after heat seals have cooled and inspect the transferred adhesive. Incomplete transfer is not conclusive; a physical seal integrity test may be needed.
Requesting a Quote
Please tell us the package format and materials, the standard and edition, the mark needed (CNAS, CMA or IAS), the specimen condition, whether defect classes and acceptance criteria are defined, and whether dye penetration or seal strength testing will follow. The laboratory confirms the test plan before samples are sent. See Sterile Medical Device Packaging Validation for the service scope and Contact Us to get in touch.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.