Medical-grade paper, polyolefin nonwovens such as Tyvek, and nonwoven sterilization wraps are all porous packaging materials: they must let sterilant gas through while keeping airborne microorganisms out. When selecting a material, comparing suppliers or facing a grade change, the question is the same: how good a microbial barrier is this material, compared with the alternative? ASTM F1608 and its Chinese counterpart YY/T 0681.10 answer it by challenging the material with an aerosol of bacterial spores, calculating a log reduction value (LRV) and ranking materials by LRV. This article covers how the exposure chamber method works, how LRV is calculated and compared, why materials differ, why a supplier change calls for new data, and which accredited reports SUNGO's Shanghai laboratory can issue.
Key points
- It is a relative comparison between materials. Per ASTM F1608, the data help assess the relative potential of one porous material, versus another, to contribute to loss of sterility of the package contents; the method is not intended to predict performance in a specific sterile packaging application.
- Results are expressed as LRV, obtained by comparing the logarithm of the challenge with the logarithm of the spores passing through. Under the same conditions, a higher LRV means fewer spores got through. The standard sets no pass/fail limit.
- Compare only under identical conditions. ASTM's round-robin study advises caution when comparing data and ranking materials, especially with few replicates.
- The two standards are not the same edition. YY/T 0681.10-2011 is a technically modified adoption of ASTM F1608-00. ASTM has since revised the method to F1608-16 and F1608-21 (the current edition as of 2026-10-22).
- Report types. SUNGO Shanghai can issue CNAS or CMA reports to YY/T 0681.10-2011, and CNAS reports to ASTM F1608-21; ASTM F1608 is not in the CMA scope. Both standards are also in the IAS scope (whole standard, no edition year stated).
- Material ranking is not package validation. It does not replace seal and integrity evaluation, or validation that the packaging system maintains sterility.
The two standards: titles, editions and relationship
| Standard | Title | Edition details | Relationship |
|---|---|---|---|
| ASTM F1608-21 | Standard Test Method for Microbial Ranking of Porous Packaging Materials (Exposure Chamber Method) | Approved 2021-10-01, published November 2021; originally approved in 1995; previous edition F1608-16; under ASTM Committee F02 on Primary Barrier Packaging, Subcommittee F02.15 | Current ASTM edition (as of 2026-10-22) |
| YY/T 0681.10-2011 | Test methods for sterile medical device package — Part 10: Test for microbial barrier ranking of porous package material | Recommended Chinese pharmaceutical industry standard; issued 2011-12-31, implemented 2013-06-01, in force as of 2026-10-22; under SAC/TC 106, the national technical committee for medical infusion devices | The foreword states that it modifies ASTM F1608-00 in technical content |
Both standards cover porous materials used to package sterile medical devices (for the exact scope of YY/T 0681.10, refer to its Clause 1). ASTM F1608 defines a porous packaging material as a material used in medical packaging that is intended to provide an environmental and biological barrier while allowing sufficient air flow to be used in gaseous sterilization methods such as ethylene oxide, steam and gas plasma.
YY/T 0681.10 is based on the 2000 edition and ASTM has revised the method twice since, so the texts may differ in detail. Parameters and calculations follow the standard used, and the report should state the standard and edition.
The FDA Recognized Consensus Standards database lists ASTM F1608-21 (recognition number 14-574, date of entry 2022-05-30, extent of recognition: complete standard) and cites ISO 11607-1 and ISO 11607-2 among the related publications. Recognition of a standard does not mean that any particular report will be accepted; the reviewing authority's requirements apply.
How the exposure chamber method works
The points below follow ASTM F1608-21. For the corresponding content of YY/T 0681.10, refer to that standard.
- Apparatus. The chamber is built mainly from acrylic sheet in two sections. The upper section has a fan to disperse the aerosol, a nebulizer port and an exhaust port; the lower section has a manifold, connected to a vacuum source, feeding six flowmeters, each with a filter unit. Because spore aerosols are generated, the chamber is assembled and used inside a biological safety cabinet.
- Challenge organism. An aqueous spore suspension of Bacillus atrophaeus (ATCC 9372) is nebulized into the chamber and dispersed by the fan.
- Samples and controls. Random samples of the material are cut into disks that fit the filter holder and placed on top of a sterile membrane filter. The vacuum draws chamber air through each sample at the specified flow rate, and spores that pass through are collected on the membrane below. One of the six filter units carries a challenge control membrane to measure the challenge, with a sterile membrane beneath it serving as the negative control. When the materials to be tested are known to produce high LRVs, it may be advisable to include in each run a known material that allows substantial spore passage, as a positive control.
- Enumeration. After exposure the membranes are removed and enumerated as specified in the standard (typically elution, dilution and plate counting; for very low counts, the membrane may be placed challenge side up directly on an agar plate) to give the spores passing each sample and the challenge level. On first use of the chamber, the challenge to each port must be validated.
F1608-21 sets standard conditions so results can be compared between laboratories: 2.8 L/min through the sample, 15 min exposure, and a specified target challenge per port. The reasons given: a spore aerosol is hard to maintain for long; differences between materials only show when spores can pass; and the challenge must be large enough to cover the full range of commercial porous packaging materials. Users may modify challenge, time and flow rate after first testing at the standard conditions.
LRV: how it is calculated and compared
LRV stands for Log Reduction Value. According to Clause 4.1 of ASTM F1608-21, it is calculated by comparing the logarithm of the number of spores passing through the porous material with the logarithm of the microbial challenge, in other words the difference between two logarithms. The exact formula is in Section 12 of the standard. Three points help when reading the numbers:
- It is a log scale. A difference of 1 in LRV corresponds to roughly a tenfold difference in spores passing under the same challenge, so a small-looking gap can mean a substantial difference in penetration.
- Zero passage does not mean an absolute barrier. The standard notes that if a material allows no spores through, all that can be stated is that its resistance to penetration exceeds the severity of the challenge conditions. That is why the method is designed around conditions in which spores can pass.
- There is no pass line. F1608 is a ranking method. ASTM states in the scope that further collaborative work (such as described in ASTM E691) should be conducted before the method would be considered adequate for setting performance standards. If a manufacturer needs material acceptance criteria, it should set them based on the risks of its product and packaging system.
What to watch when comparing
ASTM ran a round-robin study in which eleven laboratories each tested duplicate samples of six commercially available porous materials. Under the standard conditions, the average LRVs ranged from 1.7 to 4.3. The study concluded that caution should be used when comparing data and ranking materials, especially with a small number of replicates. The figures show how widely commercial materials differ; they are not acceptance limits or typical values.
Annex B of YY/T 0681.10-2011 gives precision data in LRV. Tables B.6 and B.7 give 95% inter-laboratory reproducibility limits (critical ranges) for single and multiple materials at different values of n, narrowing as n increases; refer to Annex B for the definition of n and the repeatability data. Annex B states that these critical ranges should be regarded as a general description of repeatability or reproducibility precision. It also states that a statistical bias between two laboratories is meaningful only when based on recent data from randomly sampled specimens of the same material, and that the method's bias cannot be determined because the true LRV is unknown.
The safest comparison is therefore same laboratory, same standard and edition, same conditions, ideally side by side in one test series, with extra replicates. Ranking LRVs from different laboratories, dates or conditions rarely holds up.
Paper, Tyvek and nonwovens: where they differ
Common porous packaging materials fall into three broad groups:
- Medical packaging paper: used for paper bags, pouches and reels. Chinese standards YY/T 0698.3-2009, YY/T 0698.6-2009 and YY/T 0698.7-2009 (sealable adhesive-coated paper), among others, set requirements for papers for different uses.
- Polyolefin nonwovens: a typical example is Tyvek, a DuPont trademark. According to DuPont, it is made from high-density polyethylene (HDPE) fibers that are randomly laid and compressed. YY/T 0698.9-2009 and YY/T 0698.10-2009 cover uncoated and adhesive-coated polyolefin nonwoven materials, respectively, for the manufacture of sealable pouches, reels and lids.
- Sterilization wraps: such as nonwoven wrapping materials; YY/T 0698.2-2022 sets requirements and test methods for sterilization wraps.
The YY/T 0698 series are material requirement standards, not in SUNGO's accredited scope, cited here only to describe the categories. The materials differ in fiber type, fiber arrangement and pore structure, thickness, basis weight and coating. ASTM F1608 notes that the microstructure of a porous material influences its ability to adsorb or entrap microorganisms under different air-flow conditions. Hence different materials can show clearly different LRVs under the same conditions, and grades or basis weights within one type can differ too.
For that reason this article does not rank material types or quote typical LRVs; to compare, test the candidates under the same conditions. Note also that the exposure chamber method compares materials at a single specified flow rate. ASTM F2638 (aerosol filtration) instead challenges the material with 1.0 μm particles over a range of flow rates to find the point of maximum penetration and uses that to compare materials. The two methods measure different things, so their results should not be mixed in one ranking. ASTM F2638 and the Chinese YY/T 0681.17-2019 (aerosol filtration microbial barrier test for porous packaging materials) are not in SUNGO's accredited scope and are mentioned here for background only.
Why a supplier change needs new data
An LRV describes this material under these conditions, not a material type in general. A new supplier, grade, basis weight, coating or process means a different test item, and existing data no longer represent it, for three reasons:
- Structure governs barrier performance. As noted above, microstructure affects how a material adsorbs and entraps microorganisms. Two materials with the same name do not necessarily have the same structure.
- Data do not transfer across laboratories or time. Per Annex B of YY/T 0681.10, inter-laboratory bias is meaningful only for recent data from randomly sampled specimens of the same material. Comparing fresh results with old data from another laboratory mixes laboratory and time effects into the difference.
- The material is only one part of the packaging system. Per ASTM F1608, package design, adhesion between materials, and air exchange and pressure changes in distribution and storage also affect sterility maintenance. Whether system-level validation must be repeated is for the manufacturer to decide under its change control and risk assessment.
A practical approach is to test representative lots of the current and proposed materials side by side under the same conditions, with additional replicates, then decide together with the other packaging validation results. For how seal, integrity, aging and transport evaluations fit together, see Sterile Barrier Systems: Packaging and Sterilization Validation.
What a ranking test cannot tell you
It cannot tell you whether a specific package will stay sterile through distribution and storage. ASTM F1608 states that this depends on many factors: the number and kinds of microorganisms encountered; package design, such as adhesion between materials, secondary and tertiary packaging, and the device inside; air exchange during distribution and shelf life, influenced by the free-air volume in the package and pressure changes from transport, handling, weather, door closures and HVAC systems; and the microstructure of the material. Ranking data suit material selection, incoming comparison and change assessment, but do not replace packaging system validation.
Microbial barrier testing under moist conditions and with passage of air (DIN 58953-6, YY/T 0681.14) is covered in a separate article. The Shanghai IAS scope also includes ISO 11607-1 as a whole standard, which is not discussed here.
Certificate scope
The table lists the rows from the three certificates of Shanghai SUNGO Medical Technology Co., Ltd. that relate directly to this article. Item names and standard numbers are copied from the certificates; the test object on the CNAS and CMA rows is "medical device packaging materials"; none of the rows lists a clause or a limitation.
| Certificate | Item | Standard and clause | Notes / limitations |
|---|---|---|---|
| CNAS L22083 | Microbial barrier ranking (微生物屏障分等) | ASTM F1608-21 | No clause listed, no limitation; ASTM F1608 is not in the CMA scope |
| CNAS L22083 | Microbial barrier ranking (微生物屏障分等) | YY/T 0681.10-2011 | No clause listed, no limitation |
| CMA 240015340087 | Microbial barrier ranking (微生物屏障分等) | YY/T 0681.10-2011 | No clause listed, no limitation; CMA is valid within the People's Republic of China |
| IAS TL-1019 | Whole standard | ASTM F1608 | No edition year stated in the scope |
| IAS TL-1019 | Whole standard | YY/T 0681.10 | No edition year stated in the scope |
Where SUNGO performs it and under which accreditation
This test is performed at the Shanghai laboratory (No. 958 Kangqiao East Road, Pudong, Shanghai) of Shanghai SUNGO Medical Technology Co., Ltd., which holds CNAS accreditation (L22083), CMA qualification (240015340087) and IAS accreditation (TL-1019). Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity, and its certificate scopes do not include these tests.
| Standard | CNAS | CMA | IAS |
|---|---|---|---|
| ASTM F1608 | Available (ASTM F1608-21) | Not in scope | In scope (whole standard; no edition year stated in the scope) |
| YY/T 0681.10 | Available (YY/T 0681.10-2011) | Available (YY/T 0681.10-2011) | In scope (whole standard; no edition year stated in the scope) |
- CMA: covers only the items listed in the CMA schedule, and the certificate states that it is valid within the People's Republic of China. Tests to ASTM F1608 cannot be issued with the CMA mark.
- IAS: ASTM F1608 and YY/T 0681.10 are listed without edition years. If you need an IAS report to a specific edition, confirm the edition with the laboratory before ordering.
- Choosing the standard by purpose: YY/T 0681.10-2011 is an option where Chinese registration documents or a customer specify the industry standard; ASTM F1608-21 fits where a customer or overseas documentation specifies ASTM. Ultimately, the reviewer's requirements, product technical requirements or customer specification govern.
- Non-standard conditions: F1608 allows conditions to be modified after testing at the standard conditions. Whether a test under modified conditions can be reported within the accredited scope must be confirmed with the laboratory before the order.
- Acceptance: an accreditation mark shows the laboratory's competence within its scope. Whether a particular regulator, notified body or buyer accepts a given report depends on that party's requirements.
Preparing samples
- State the standard and edition: ASTM F1608-21 or YY/T 0681.10-2011; CNAS, CMA or IAS report; and, for IAS, which edition.
- State the purpose: material selection, incoming inspection or supplier change assessment. The purpose determines how samples are grouped and how many replicates are planned.
- Provide material information: material type (medical packaging paper, polyolefin nonwoven, nonwoven wrap, etc.), grade and specification, supplier and lot number, and whether it is adhesive-coated.
- Mark the orientation: if the material has distinct sides, mark the side that faces outward in use, and state any requirement for the challenge side in the order.
- Describe the sample condition: F1608-21 notes that sample disks are typically sterilized before testing by a method suited to the material, and that materials may also be tested before or after heat or cold, relative humidity, different sterilization processes, or real-time or accelerated aging. For sterilized or aged material, state the sterilization method or aging conditions in advance. Accelerated aging can be arranged through our shelf-life validation service.
- Send comparison sets together: when comparing materials, including those of current and new suppliers, send them together for side-by-side testing under the same conditions.
- Protect the samples: flat, uncreased, undamaged, clean, individually packed, no hand contact in the test area.
- Provide enough material: F1608-21 sets a minimum of two samples per material and strongly suggests more; it also suggests cutting extra samples to allow for errors during the procedure. Confirm the quantity with the laboratory before ordering.
FAQ
Can a test to ASTM F1608 carry the CMA mark?
No. In the Shanghai CMA schedule, microbial barrier ranking is listed only under YY/T 0681.10-2011; ASTM F1608 is not listed. Tests to ASTM F1608-21 can be issued as CNAS reports, and ASTM F1608 is also in the IAS scope.
What LRV counts as a pass?
The standard sets no pass line. F1608 is a ranking method: under the same conditions, a higher LRV means fewer spores passed. Acceptance criteria are for the manufacturer to set based on the risks of its product and packaging system.
If no spores passed, is the material an absolute barrier?
No. The standard states that when no spores pass, all that can be said is that the material's resistance to penetration exceeds the severity of the challenge conditions; it does not tell you how much better it is than another material.
Is this the same as dry microbial penetration testing for surgical gowns?
No. Dry microbial penetration of surgical drapes, gowns and clean air suits is tested to standards such as YY/T 0506.5-2009. The test object and method are different, and the results of one cannot substitute for the other.
When requesting a quotation, include the material type and specification, standard and edition, report type (CNAS, CMA or IAS), sample condition (sterilized or aged or not), the materials to be compared, and the intended use of the report. Related services are listed on the sterile packaging validation service page, and requests can be sent through the contact page. The laboratory will confirm which report types can be issued within its accredited scope.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.