Lab1 Shanghai SUNGO Lab Lab2 Lab3 Lab4 Hefei SUNGO Lab HefeiCMA HefeiCNAS HefeiIAS
WeChat WeChat QR Code
Contact
+86 132 4819 8029 (Mr. Gan)
+86 177 5560 4650 (Mr. Luo)

EN 12184 Class A, B and C: How Powered Wheelchair Classes Change Testing

EN 12184 Class A, B and C: How Powered Wheelchair Classes Change Testing

Clause 5 of EN 12184:2022 sorts electrically powered wheelchairs, including scooters, into Class A, Class B and Class C according to the surfaces and slopes they are intended to drive on. Class A is for flat horizontal surfaces and gentle slopes; Class B adds moderately uneven surfaces and moderate slopes; Class C adds uneven terrain and steeper slopes. Once a class is declared, slope climbing, obstacle climbing, stability, maximum speed, distance range and parking brake tests are run and judged against that class. This guide explains how the three classes differ, which tests change with the class, what has to be retested if the class was declared wrongly, and what the Hefei laboratory is accredited for under EN 12184.

Short answer

  • Clause 5 of EN 12184:2022, "Type classes", requires each wheelchair to be classified in one or more of three classes, A, B and C, according to its intended use. The 2022 edition describes the classes by driving surface and slope; the 2014 edition described them by compactness, manoeuvrability and the ability to negotiate outdoor obstacles. Older descriptions should not be carried over to the 2022 edition.
  • The class sets the conditions and requirements for the clause 8.1 driving tests: rated slope climbing, ground unevenness, maximum downhill speed, dynamic stability, obstacle climbing, static stability, maximum speed and distance range. The parking brake test in clause 10.2 also takes its slope from the class. Class A wheelchairs are not required to undergo the ISO 7176-8 drop test.
  • The class must be stated in the pre-sale information (clause 13.2), and a Class A wheelchair not intended for outdoor use must carry a warning label to that effect (clause 13.5). If the class changes, the tests, the user documentation and the labels all have to follow.
  • The Chinese standard GB/T 12996-2024 divides powered wheelchairs into indoor, outdoor and road types. Its basis and requirements differ from EN 12184, so one cannot be converted directly into the other.
  • Hefei SUNGO Medical Technology Co., Ltd. holds CNAS accreditation L24279 for 47 clause items of EN 12184:2022, including 8.1.2 to 8.1.9. Its IAS accreditation TL-1119 covers EN 12184 except clauses 6.6 and 12.3.1. CMA does not cover EN 12184.

EN 12184:2022 scope and edition

EN 12184:2022, Electrically powered wheelchairs, scooters and their chargers – Requirements and test methods, was approved by the European Committee for Standardization (CEN) on 17 July 2022, published in September 2022 and supersedes EN 12184:2014. According to its European foreword, national standards bodies had to give it national-standard status by March 2023 at the latest and withdraw conflicting national standards by May 2024 at the latest. As of 13 November 2026, the 2022 edition is the current edition.

The 2022 edition applies to electrically powered wheelchairs with a maximum speed not exceeding 20 km/h, intended to carry one person with a mass of 25 kg to 300 kg. Its scope explicitly includes electrically powered scooters with three or more wheels, manual wheelchairs with an add-on electrically powered drive system, handrim-activated power-assisted wheelchairs, electrically powered stand-up wheelchairs, wheelchairs with a pivot drive wheel unit and push-assist wheelchairs. It does not apply to balancing wheelchairs, custom-made electrically powered wheelchairs or electrically powered wheelchairs intended for sports. It also contains requirements for manual wheelchairs with electrically powered ancillary equipment.

The 2014 edition was limited to a maximum speed of 15 km/h and listed handrim-activated power-assisted wheelchairs, electrically powered stand-up wheelchairs and manual wheelchairs with add-on power kits among the products it did not apply to in total. A product tested to the 2014 edition should therefore first be checked against the new scope, and only then against the class definitions. The introduction to the standard also notes that manufacturers can consider using relevant parts of it to assess performance against the general safety and performance requirements of Regulation (EU) 2017/745 on medical devices. Manual wheelchairs are covered by EN 12183; see EN 12183:2022 manual wheelchairs, clause by clause.

EN 12184 Class A vs Class B vs Class C

The table below paraphrases how clause 5 of the 2022 and 2014 editions describes the three classes. The wording of the standard itself prevails.

ClassEN 12184:2022 descriptionEN 12184:2014 description
Class AIntended for driving on flat horizontal surfaces and gentle slopesCompact, manoeuvrable wheelchairs not necessarily capable of negotiating outdoor obstacles
Class BIn addition to Class A use, intended for driving on moderately uneven surfaces and moderate slopesCompact and manoeuvrable enough for some indoor environments and capable of negotiating some outdoor obstacles
Class CIn addition to Class A and B use, intended for driving on uneven terrain and steeper slopesUsually large, not necessarily intended for indoor use, but capable of travelling longer distances and negotiating outdoor obstacles

The key difference is that the 2022 edition no longer uses "compact", "indoor" and "outdoor" to separate the classes. It classifies by the surfaces and slopes the product is intended to drive on, and each class builds on the one below it. The class is decided by comparing the intended use and environment that the manufacturer states in the pre-sale information and user manual with the wording of clause 5. Speed, size and price are not classification criteria, and rules of thumb such as "indoor chairs are Class A, outdoor chairs are Class C" are no substitute for checking the 2022 text.

Scooters are classified too. Clause 5 of the 2014 edition stated explicitly that scooters are included within the classes, and the 2022 definition of an electrically powered wheelchair also includes scooters. The standard does not assign scooters to a fixed class; like wheelchairs, they are declared as Class A, B or C according to intended use.

Which EN 12184 tests change with the class

Clause 8.1 of the 2022 edition requires the loaded wheelchair to meet the driving performance requirements in Tables 3 and 4 for its type class. The class-dependent items are listed below. Slope angles, heights, distances and speeds are given in Table 3 of the standard and are not reproduced here.

ClauseItemHow the class affects it
8.1.2Ability to climb rated slopeThe rated slope declared by the manufacturer must not be less than the Table 3 value for the class; the test uses the class slope or the declared slope if greater
8.1.3Ground unevennessThe height a wheel is raised to is taken from Table 3
8.1.4Maximum downhill speedMeasured on the rated slope for the class, or the declared slope if greater
8.1.5Dynamic stabilityTested to ISO 7176-2:2017 on the rated slope for the class, and also on the declared slope if greater
8.1.6Obstacle climbing and descendingTested to ISO 7176-10:2008 with the higher of the class obstacle height and the manufacturer's declared maximum
8.1.7Static stabilityTested to ISO 7176-1:2014; each direction must meet the class requirement
8.1.8Maximum speedTested to ISO 7176-6:2018; forward and reverse speed on the horizontal must not exceed the class limit
8.1.9Distance rangeTested to ISO 7176-4:2008; theoretical continuous range must not be less than the class requirement
8.2Static, impact and fatigue strengthTested to ISO 7176-8:2014; Class A wheelchairs are not required to undergo its clause 10.4 drop test
10.2Braking functions (parking brake)Tested to ISO 7176-3:2012 at the angle for the class, or the declared rated slope if greater
13.2, 13.5Information supplied by the manufacturerPre-sale information states Class A, B or C; a Class A wheelchair not intended for outdoor use carries a warning label

Two points are easy to miss. First, maximum speed is a "shall not exceed" requirement, so when a class is moved downwards the measured speed has to be judged again against the new class limit. Second, once the declared rated slope or maximum obstacle height exceeds the class requirement, the tests are run at the declared value, so declared values are part of the test conditions. In addition, clause 12.7 requires wheelchairs intended by the manufacturer for outdoor use to have suitable integral lighting where its absence is likely to cause a risk; if the class and intended use move towards outdoor use, this clause has to be checked again.

EN 12184 class declared wrongly: what has to be retested

If the declared class does not match the basis of the tests, the report cannot support the declaration. Three common situations:

  • Moving to a higher class (for example A to B or C): the driving items in 8.1.2 to 8.1.9 and the 10.2 parking brake test need to be reassessed against the new class, and items whose test conditions are more demanding than before have to be retested; the ISO 7176-8 drop test skipped for Class A has to be added; pre-sale information, user manual and labels must be updated.
  • Moving to a lower class (for example C to B or A): slope and obstacle conditions become easier, but the maximum speed limit may also differ, so results have to be judged again against the new class; a Class A wheelchair not intended for outdoor use also needs the warning label.
  • Same class, but a higher declared rated slope or maximum obstacle height: the slope climbing, downhill speed, stability, obstacle and parking brake tests have to be repeated at the new declared value.

Which existing data can still be used should be assessed by the laboratory item by item against the new class, and the change recorded in the technical documentation. After changing key parts such as motors, batteries, tyres or the speed limit in the controller, the class and declared values should also be checked again.

How to determine and declare the EN 12184 class

  1. Define the intended operator, intended use and intended environment. Clause 13.2 requires these in the pre-sale information, and the user manual and labels must match.
  2. Compare the intended surfaces and slopes with the clause 5 wording for Classes A, B and C, and select the class.
  3. Fix the declared values, such as rated slope, maximum obstacle height and maximum occupant mass. The rated slope must not be lower than the requirement for the declared class.
  4. Record the class, the reasoning and the declared values in the technical documentation (clause 6.2), state them when ordering the tests, and build the test plan on the same class.

EN 12184 classes vs GB/T 12996-2024 indoor, outdoor and road types

GB/T 12996-2024, Electrically powered wheelchair, is a recommended Chinese national standard. It was issued on 26 October 2024, has applied since 1 February 2025 and fully replaces GB/T 12996-2012. It divides powered wheelchairs into indoor, outdoor and road types by intended use scenario, and scooters are divided into the same three types (the text of the standard prevails).

EN 12184:2022 classifies by surface and slope into Classes A, B and C, while GB/T 12996-2024 uses indoor, outdoor and road use scenarios, and the two standards also differ in test items, methods and requirements. The Chinese type cannot be used to derive the EU class, or the other way round. For a product sold in the EU and also registered in China, determine the class under each standard from its own text and explain the relationship and differences in the technical documentation. At the Hefei laboratory, GB/T 12996-2024 is within the CNAS L24279 and CMA scopes (19 items in each) and is not covered by IAS.

Accreditation scope: Hefei CNAS L24279 items for EN 12184:2022

The table lists all 47 EN 12184:2022 items on the Hefei laboratory's CNAS L24279 scope. Clause numbers follow the certificate; item names are English renderings of the certificate entries. The "Notes/limitations" column reflects the certificate; "—" means no limitation is stated.

ItemClauseNotes/limitations
Intended performance and technical documentation6.2—
Wheelchairs that can be dismantled6.4—
Single-use fasteners6.5—
Contaminants and residues6.7—
Infection and microbiological contamination6.8—
Overflow, spillage, leakage and ingress of liquids6.9—
Safety of moving parts6.10—
Prevention of traps for parts of the human body6.11—
Folding and adjusting mechanisms6.12—
Surfaces, corners, edges and protruding parts6.13—
Ergonomic principles6.14—
Applicable provisions for specified types of wheelchair6.16—
Recommendations6.17—
Preparation for testing7—
Slope climbing ability8.1.2—
Driving on uneven ground8.1.3—
Maximum downhill speed8.1.4—
Dynamic stability8.1.5—
Obstacle climbing ability8.1.6—
Static stability8.1.7—
Maximum speed8.1.8—
Distance range8.1.9—
Static, impact and fatigue strength8.2—
Climatic test8.3—
Foot supports, lower leg support assemblies and arm supports9.1—
Component mass9.2—
Pneumatic tyres9.3—
Means for maintaining a sitting posture9.4—
Resistance to ignition9.5—
Means for operating brakes10.1—
Braking performance10.2—
Freewheel device10.3—
Operations intended to be carried out by the occupant and/or assistant11.1—
Controls intended for operation by the occupant11.2—
Controls intended for operation by an assistant11.3—
Assistant control unit, push handles and handgrips11.4—
Operating forces11.5—
Occupied seating adjustments11.6—
General requirements for electrical systems12.1—
Circuit protection12.2—
Battery chargers12.3—
Charging connector12.4—
Battery enclosures and containers12.5—
Emergency stop12.6—
Lighting12.7—
Switching off while driving12.8—
Information supplied by the manufacturer13—

Three points when reading this table. First, the "Climatic test" row carries clause 8.3 as listed on the certificate; in the EN 12184:2022 table of contents, 8.3 is "Wheelchairs for use as seats in motor vehicles" and climatic performance is 8.4. The ISO 7176-19 crash testing involved in motor-vehicle seating use is not within SUNGO's accredited scope; please confirm with the laboratory which clause the climatic test will be reported under. Second, for the ISO 7176-21 electromagnetic compatibility testing referenced in clause 12, SUNGO holds only the Hefei laboratory's IAS TL-1119 accreditation; neither CNAS nor CMA covers EMC. Third, clause 6.6, biocompatibility and toxicity, is not in the table; see the next section.

Where SUNGO tests EN 12184 and under which accreditation

EN 12184 testing is performed by Hefei SUNGO Medical Technology Co., Ltd., whose laboratory is in Zipeng Town, Feixi County, Hefei, Anhui Province. Accreditation status:

  • CNAS L24279: CNAS-accredited reports can be issued for the 47 items above; 8.1.2 to 8.1.9 are the driving performance items that change with the class.
  • IAS TL-1119: covers EN 12184, with the scope noting exceptions for clause 6.6, biocompatibility and toxicity, and clause 12.3.1. The IAS scope does not state an edition year, so please confirm the edition with the laboratory before an IAS report is issued to the 2022 edition.
  • CMA: does not cover EN 12184. CMA applies only to items listed in its schedule and is valid only within China.
  • Clause 6.6, biocompatibility and toxicity: not performed by SUNGO. It can only be subcontracted to a laboratory with the appropriate qualifications, with the client's prior consent, and the report identifies the subcontracted items and the subcontractor.

The Hefei laboratory is accredited by CNAS and IAS, both signatories to the ILAC MRA. For an overview of wheelchair testing, see wheelchair testing services.

Preparing samples for EN 12184 testing

  • Class declaration and statement of intended use: Class A, B or C, plus the intended operator and environment, consistent with the pre-sale information, user manual and labels.
  • Declared values: rated slope, maximum obstacle height, maximum occupant mass, maximum speed and similar values, which feed directly into the test conditions.
  • Technical documentation corresponding to clause 6.2, intended performance and technical documentation.
  • Information supplied by the manufacturer under clause 13: pre-sale information, user information and draft labels.
  • Charger and battery information, so that clauses 12.3 to 12.5 can be checked.
  • Set-up before testing: clause 7 of the 2022 edition requires the wheelchair to be prepared as specified in ISO 7176-22:2014 unless otherwise stated; see ISO 7176-22 wheelchair set-up procedures.

Sample quantity, test sequence and specific parameters are agreed with the laboratory in the test plan.

FAQ

What is the difference between EN 12184 Class A, B and C?

Under clause 5 of the 2022 edition, Class A is intended for flat horizontal surfaces and gentle slopes, Class B also for moderately uneven surfaces and moderate slopes, and Class C also for uneven terrain and steeper slopes. Each class has its own Table 3 requirements for slope climbing, obstacles, stability, speed and distance range.

Which EN 12184 class is a mobility scooter?

The standard does not assign scooters to a fixed class. Scooters are within the scope of EN 12184:2022 and, like powered wheelchairs, are declared as Class A, B or C according to their intended use.

What changed in the classes between EN 12184:2014 and EN 12184:2022?

Both editions use Classes A, B and C. The 2014 edition described them by compactness, manoeuvrability and outdoor obstacles; the 2022 edition describes them by driving surface and slope, and its scope is also wider, so a class chosen under the old edition should be rechecked against the new text.

Can one powered wheelchair be declared in more than one class?

Clause 5 states that a wheelchair is classified in one or more classes. How the tests and judgements are arranged for more than one class should be agreed with the laboratory in the test plan, and the standard text prevails.

Do the GB/T 12996-2024 indoor and outdoor types match Classes A, B and C?

Not directly. The two standards classify on different bases and have different test requirements, so determine the class under each standard from its own text.

EN 12184 testing quotation

If you need an EN 12184 report, please tell us the product type (powered wheelchair or scooter), the class you intend to declare and the intended environment, declared values such as rated slope and maximum occupant mass, the accreditation you need (CNAS or IAS) and your target market. We will use this to confirm the applicable tests, the accreditation scope and the documents to submit. Send your enquiry through the contact page.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.