EN 12183:2022 Manual Wheelchairs: Clause-by-Clause Tests and Report Scope

EN 12183:2022 Manual Wheelchairs: Clause-by-Clause Tests and Report Scope

If you sell manual wheelchairs into Europe, the relevant product standard is EN 12183:2022, Manual wheelchairs — Requirements and test methods. It covers general requirements, performance, components, braking, operations and manufacturer information, and lists several ISO 7176 parts as normative references. This guide explains each clause, the links to ISO 7176 and the changes from 2014, and which clauses the SUNGO Hefei laboratory reports under CNAS, which only under IAS, and which are subcontracted or need case-by-case confirmation.

Key points

  • As of October 2026, the current edition is EN 12183:2022, the fifth edition. Approved by CEN on 17 July 2022 and published in September 2022, it supersedes EN 12183:2014. Per the European foreword, national adoption was due by March 2023 and conflicting national standards were to be withdrawn by May 2024.
  • The scope has changed. It covers manual wheelchairs for one person of mass not less than 25 kg and not greater than 250 kg, now expressly including stand-up wheelchairs and wheelchairs whose intended use includes showering and/or toileting; see the comparison table below.
  • Own requirements plus ISO 7176 references. Clauses 5 to 13 set out its own requirements and methods; normative references include ISO 7176-1, -3, -8, -11, -13, -19, -22 and -26.
  • At SUNGO, testing is done by the Hefei laboratory. Its CNAS scope (L24279) lists 32 clause items of EN 12183:2022, and its IAS scope (TL-1119) covers EN 12183 except clause 5.6. EN 12183 is not on the CMA schedule.
  • Agree the report scope first. Clause 5.6 (biocompatibility and toxicity) is outside both scopes and is subcontracted by SUNGO to an appropriately qualified laboratory; under the accreditation rules, the subcontracting is agreed with the client in advance, and the report identifies the subcontracted items and the subcontractor. Clauses 5.1, 5.3 and 5.15 are not CNAS-listed and can only be reported under IAS. ISO 7176-19, a normative reference on the same topic as 7.4, is not within SUNGO's accreditation, so 7.4 needs case-by-case confirmation.

The standard and its scope

Edition and publication

EN 12183:2022 was prepared by CEN/TC 293, Assistive products and accessibility (secretariat: SIS, Sweden). The introduction calls it the fifth edition, after those of 1999, 2006, 2009 and 2014. The European foreword states that it supersedes EN 12183:2014, that Annex G details the significant technical changes, and that EN 12184 covers electrically powered wheelchairs.

CEN members must adopt the standard without alteration, so BS EN 12183:2022 (published 30 September 2022) and DIN EN 12183:2022-12 (the German-language version) carry the same technical content plus national front matter such as a national foreword. The introduction notes that manufacturers can consider using appropriate parts of it to assess products against the general safety and performance requirements of Regulation (EU) 2017/745.

Scope: 2014 edition compared with 2022 edition

TopicEN 12183:2014EN 12183:2022
Occupant massOne person, not greater than 250 kgOne person, not less than 25 kg and not greater than 250 kg
Stand-up manual wheelchairsNot covered in totalIncluded
Showering and toiletingSpecial-purpose wheelchairs not covered in totalIncluded if intended use includes them; excluded if intended only for them
Sports and custom-made wheelchairsNot covered in totalNot covered
Electrically powered ancillary equipmentCoveredCovered
Handrim-activated power-assisted propulsion; add-on power kits used for propulsionNot covered in totalNot mentioned in the scope; see full text

Clause 5.15 and normative Annex F (F.2 showering or toileting, F.3 stand-up wheelchairs and the standing configuration) say which clauses apply to these types.

What changed in 2022

The "Änderungen" (changes) section of DIN EN 12183:2022-12, published on the DIN Media product page, lists the changes from DIN EN 12183:2014-06. Those most relevant to testing:

  • Status of wheelchairs the previous edition covered only in part clarified; normative references updated; risk management requirements changed.
  • Above a maximum occupant mass of 100 kg, additional rearward static stability tests with a 100 kg test dummy are required.
  • Modifications to the ISO 7176-8 tests for occupant masses above 100 kg deleted, with former Table 2 (static load forces for user mass above 100 kg); wheel load distribution for ISO 7176-8 fatigue tests changed above 100 kg.
  • Tipping fatigue test now also required above a maximum occupant mass of 150 kg.
  • Requirements changed for wheelchairs used as seats in motor vehicles, anterior pelvic support, resistance to ignition and user information.
  • New normative Annex F for particular wheelchair types; Annex C (rail interoperability) updated; Annex ZA deleted.

This list follows DIN's change notes, some items of which do not match the published table of contents; Annex G and the clauses of EN 12183:2022 govern.

Clause by clause

Clause titles follow the BSI and DIN Media tables of contents; references follow BSI's list. Clauses 1 to 4 (scope, references, terms, test apparatus) are skipped; requirements and pass criteria are those in the standard.

Clause 5 General requirements (5.1 to 5.16)

Clause 5 has 16 subclauses. In DIN's published table of contents, only 5.9 has separate requirements and test methods; how the others are checked is set out in the standard.

  • 5.1 to 5.5: 5.1 Risk management, 5.2 Intended performance and technical documentation, and 5.3 Clinical evaluation and investigation are mainly documentation clauses (EN ISO 14971 and EN ISO 14155 are normative references); 5.4 Wheelchairs that can be dismantled and 5.5 Single-use fasteners are single clauses.
  • 5.6 Biocompatibility and toxicity: EN ISO 10993-1 is a normative reference. This clause is outside SUNGO's accreditation; see "Report scope" below.
  • 5.7 Contaminants and residues: general, and substances which can leak in intended use or in a fault condition.
  • 5.8 Infection and microbiological contamination: cleaning and disinfection, and animal tissue; EN ISO 22442-1 (medical devices utilizing animal tissues and their derivatives) is a normative reference.
  • 5.9 Overflow, spillage, leakage and ingress of liquids: four subclauses; overflow, spillage and ingress each have requirements and a test method, leakage has no further subdivision.
  • 5.10 to 5.12: safety of moving parts (squeezing, mechanical wear); prevention of traps for parts of the human body (holes and clearances, V-shaped openings); folding and adjusting mechanisms (general, locking mechanisms, guards).
  • 5.13 Surfaces, corners, edges and protruding parts; 5.14 Ergonomic principles: EN 614-1 (safety of machinery, ergonomic design principles) is a normative reference.
  • 5.15 Applicable provisions for specified types of wheelchair; 5.16 Recommendations: normative Annex F sets out applicable requirements for particular types of manual wheelchair, and informative Annex B lists recommended design features; see the full text for how they tie in.

Clause 6 Preparation for testing

6.1 General and 6.2 Test dummy. ISO 7176-22 (set-up procedures) and ISO 7176-11 (test dummies) are normative references. Configuration, adjustment and loading set the conditions for the clause 7 tests; state them when you order.

Clause 7 Wheelchair performance

  • 7.1 Static stability: requirements and test method; ISO 7176-1 (determination of static stability) is a normative reference. See the 2022 changes above.
  • 7.2 Static, impact and fatigue strength: requirements and test method; tests to ISO 7176-8 (see the 2022 changes above for occupant masses above 100 kg).
  • 7.3 Tipping fatigue strength: EN 12183 has its own method here: 7.3.1 General, 7.3.2 Requirement, 7.3.3 Test method. Which wheelchairs need it is set out in 7.3 itself; per the DIN change list it is now also required above a maximum occupant mass of 150 kg.
  • 7.4 Wheelchairs for use as seats in motor vehicles: ISO 7176-19 (wheeled mobility devices for use as seats in motor vehicles) is a normative reference; see "Report scope" below for how 7.4 is handled.
  • 7.5 Surface temperature: contact surfaces and requirements; EN ISO 13732-1 (assessment of human responses to contact with hot surfaces) is a normative reference.

Clause 8 Component properties

  • 8.1 Foot supports, lower leg support assemblies and arm supports: requirements; the test methods cover general performance and the gap between the foot supports.
  • 8.2 Component mass, 8.3 Pneumatic tyres, 8.4 Means for maintaining a sitting posture: single clauses without further subdivision.
  • 8.5 Resistance to ignition: general, then test methods: selection of method, a reference method and alternative methods. Ignition-related normative references include ISO 16840-10 (postural support devices), EN 1021-2 and ISO 8191-2 (upholstered furniture, match-flame equivalent); which applies is set out in the standard.

Clause 9 Propulsion and braking systems

  • 9.1 Means for operating brakes: requirement, and a test method for determining brake operating forces.
  • 9.2 Braking functions: requirements; the test methods cover parking brake effectiveness, protruding parts of the parking brake, and parking brake fatigue (per the DIN table of contents). ISO 7176-3 (determination of effectiveness of brakes) is a normative reference.
  • 9.3 Pushing force: requirements and test method.

Clause 10 Operations

10.1 Operations by the occupant and/or assistant; 10.2 Controls for the occupant; 10.3 Controls for an assistant; 10.4 Push handles and handgrips; 10.5 Operating forces; 10.6 Seating adjustments for tilting and reclining. 10.4 to 10.6 each have requirements and a test method; tilting or reclining wheelchairs need documentation for 10.6.

Clause 11 Electrically powered ancillary equipment

Clause 11, "Electrical systems — Electrically powered ancillary equipment", applies to manual wheelchairs fitted with such equipment. EN 12184 (electrically powered wheelchairs, scooters and their chargers) is a normative reference; how clause 11 uses it is set out in the standard. For electromagnetic compatibility, see "Report scope" below.

Clause 12 Information supplied by the manufacturer; clause 13 Test report

Clause 12 has five parts: general, pre-sale information, user information, service information and labelling. EN ISO 20417 (information to be supplied by the manufacturer) is a normative reference; per the DIN change list, user information requirements have changed. Clause 13 covers the test report and figures.

Annexes

AnnexStatusContent
AInformativeRecommendations for dimensions and manoeuvring space of manual wheelchairs
BInformativeRecommended design features
CInformativeEN 12183 and rail interoperability
DInformativeRecommendations for caster stem angle
EInformativeHazardous substances
FNormativeApplicable requirements for particular types of manual wheelchair
GInformativeTechnical changes from the previous edition

Annex E gives recommendations on substances of very high concern (CMR, PBT/vPvB and others) and on hazardous substances in textiles, plastics, metals and wood; which clauses it relates to is set out in the standard.

How EN 12183 ties in with ISO 7176

ISO 7176 parts among the normative references, by topic, with the Hefei laboratory's certificate entries. Which clause cites each part, and which edition, is set out in EN 12183:2022 itself. CNAS and CMA cover only the listed items and clauses (for example, clauses 8 to 11 of ISO 7176-1:2014 on CNAS); IAS lists each part as a whole standard with no year. See the certificate schedules.

TopicReferenced partHefei laboratory certificates
Preparation for testingISO 7176-22 set-up procedures; ISO 7176-11 test dummiesCNAS, CMA: ISO 7176-22:2014, ISO 7176-11:2012; IAS: whole standards
Static stabilityISO 7176-1CNAS, CMA: ISO 7176-1:2014; IAS: whole standard
StrengthISO 7176-8CNAS, CMA: ISO 7176-8:2014; IAS: whole standard
BrakesISO 7176-3CNAS, CMA: ISO 7176-3:2012; IAS: whole standard
Test surfacesISO 7176-13CNAS, CMA: ISO 7176-13:1989; IAS: whole standard
Motor vehicle seatingISO 7176-19Not within SUNGO's accreditation
TerminologyISO 7176-26 vocabularyTerminology standard, not a test method; not on the certificates

Report scope: which clauses go into which report

ClausesCNAS (L24279)IAS (TL-1119)CMA
The 32 CNAS-listed clauses: 5.2, 5.4, 5.5, 5.7 to 5.14, 6, 7.1, 7.2, 7.3, 7.5, 8.1 to 8.5, 9.1 to 9.3, 10.1 to 10.6, 11, 12In scope (EN 12183:2022)In scopeNot on the schedule
5.1 Risk management, 5.3 Clinical evaluation and investigation, 5.15 Applicable provisions for specified types of wheelchairNot listedIn scopeNot on the schedule
7.4 Wheelchairs for use as seats in motor vehiclesNot listedWithin the scope wording; ISO 7176-19 is not within SUNGO's accreditation, so confirm case by caseNot on the schedule
5.6 Biocompatibility and toxicityNot listedExplicitly excludedNot on the schedule

Clauses 1 to 4, 5.16 Recommendations and 13 Test report and figures are not CNAS-listed either; the IAS scope is EN 12183 except 5.6.

  • 5.6 is subcontracted: SUNGO subcontracts it to an appropriately qualified laboratory. Under the accreditation rules, the subcontracting is agreed with the client in advance, and the report identifies the subcontracted items and the subcontractor.
  • IAS-only clauses: 5.1, 5.3 and 5.15. A CNAS report covers only the 32 clauses above; one report covering every clause except 5.6 (any part of 7.4 involving ISO 7176-19 testing to be confirmed case by case) has to be an IAS report. The IAS scope gives no year, so confirm the edition before an IAS report is issued.
  • 7.4 needs a separate check: ISO 7176-19, a normative reference on the same topic, is not within SUNGO's accreditation. If your wheelchair is declared suitable as a seat in a motor vehicle, say so in your enquiry; the laboratory will confirm how 7.4 is handled.
  • Clause 11 and EMC: if clause 11 involves electromagnetic compatibility testing of the ancillary equipment, that part is within the Hefei laboratory's IAS accreditation (TL-1119); confirm the report arrangements with the laboratory.
  • No CMA: EN 12183 is not on the Hefei laboratory's CMA schedule, so the report cannot carry the CMA mark; CMA covers only scheduled items and is valid only within the People's Republic of China.

Accreditation scope

The 32 EN 12183:2022 rows on the CNAS scope (L24279) of Hefei SUNGO Medical Technology Co., Ltd., shown by English clause title, plus the undated EN 12183 row on its IAS scope (TL-1119). Every CNAS row has manual wheelchairs as the test object and no limitation.

ItemClauseNotes / limitations
Intended performance and technical documentation5.2CNAS
Wheelchairs that can be dismantled5.4CNAS
Single-use fasteners5.5CNAS
Contaminants and residues5.7CNAS
Infection and microbiological contamination5.8CNAS
Overflow, spillage, leakage and ingress of liquids5.9CNAS
Safety of moving parts5.10CNAS
Prevention of traps for parts of the human body5.11CNAS
Folding and adjusting mechanisms5.12CNAS
Surfaces, corners, edges and protruding parts5.13CNAS
Ergonomic principles5.14CNAS
Preparation for testing6CNAS
Static stability7.1CNAS
Static, impact and fatigue strength7.2CNAS
Tipping fatigue strength7.3CNAS
Surface temperature7.5CNAS
Foot supports, lower leg support assemblies and arm supports8.1CNAS
Component mass8.2CNAS
Pneumatic tyres8.3CNAS
Means for maintaining a sitting posture8.4CNAS
Resistance to ignition8.5CNAS
Means for operating brakes9.1CNAS
Braking functions9.2CNAS
Pushing force9.3CNAS
Operations intended to be carried out by the occupant and/or assistant10.1CNAS
Controls intended for operation by the occupant10.2CNAS
Controls intended for operation by an assistant10.3CNAS
Push handles and handgrips10.4CNAS
Operating forces10.5CNAS
Seating adjustments for tilting and reclining10.6CNAS
Electrical systems — Electrically powered ancillary equipment11CNAS
Information supplied by the manufacturer12CNAS
EN 12183 (except 5.6)EN 12183 (no year stated)IAS; limitation as written: "except for 5.6 Biocompatibility and toxicity"; confirm the edition with the laboratory before an IAS report is issued

Where the work is done and under which accreditation

  • Laboratory: Hefei SUNGO Medical Technology Co., Ltd., a separate legal entity, in Zipeng Town, Feixi County, Hefei, Anhui Province. CNAS (L24279) covers the 32 items above, IAS (TL-1119) covers EN 12183 except 5.6, and EN 12183 is not on the CMA schedule (260015342945). The scope of Shanghai SUNGO Medical Technology Co., Ltd. includes no wheelchair standards.
  • Referenced parts on their own: for ISO 7176, see the table above; three items of ISO 16840-10:2021/Amd 1:2024 are on the CNAS scope and ISO 16840-10 is on the IAS scope. See the certificate schedules for exact items.
  • Acceptance: whether a notified body or customer accepts a report is their decision; check the edition and accreditation mark they require first. CE certification procedures are not covered here.

Preparing your samples and documents

  • Model, intended use (stand-up, showering or toileting, use as a seat in a motor vehicle) and the declared maximum occupant mass; according to the DIN change list, the 2022 stability, strength and tipping fatigue arrangements depend on it.
  • Technical documentation describing intended performance; for an IAS report covering 5.1 and 5.3, also the risk management and clinical evaluation documents.
  • Instructions for use, labels, pre-sale information and service information, for clause 12.
  • Adjustment instructions, supplied tools and an options list (backrests, foot supports, arm supports, wheels, anti-tip devices, cushions; mark which are fitted to the sample), for clause 6.
  • Materials of cushions, fabrics and other upholstered parts, for 8.5 resistance to ignition.
  • Pneumatic tyre specification and recommended pressure.
  • For electrically powered ancillary equipment: model, battery, charger and instructions.
  • For subcontracted 5.6: a list of materials in contact with the user.
  • In your test request, state the edition (EN 12183:2022), the clauses you need and the report type (CNAS or IAS). The number of samples depends on the clauses and test sequence; agree it with the laboratory.

Frequently asked questions

We already have ISO 7176 reports. Do we still need EN 12183?

ISO 7176 reports cover only those parts. EN 12183 adds general requirements, components, operations, manufacturer information and tipping fatigue; whether existing data can be used depends on edition, configuration and maximum occupant mass, item by item.

Can an EN 12183 report carry the CMA mark?

No. EN 12183 is not on the Hefei laboratory's CMA schedule, and CMA covers only scheduled items and is valid only within the People's Republic of China.

Are reports to EN 12183:2014 still usable?

The 2014 edition is superseded; conflicting national standards were to be withdrawn by May 2024. Acceptance of an older report is up to the notified body or customer; new testing should follow the 2022 edition.

Can stand-up wheelchairs, or wheelchairs whose intended use includes showering or toileting, be tested to EN 12183:2022?

The 2022 scope covers both, with the applicable clauses in 5.15 and Annex F; wheelchairs intended only for showering and/or toileting are excluded. As 5.15 is not CNAS-listed, confirm the report type first.

For a quotation, send the documents listed above, the clauses you need and the report type through our contact page. Wheelchair and mobility aid testing services are described on our wheelchair testing page.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.