Lab1 Shanghai SUNGO Lab Lab2 Lab3 Lab4 Hefei SUNGO Lab HefeiCMA HefeiCNAS HefeiIAS
WeChat WeChat QR Code
Contact
+86 132 4819 8029 (Mr. Gan)
+86 177 5560 4650 (Mr. Luo)

EN 14079 Absorbent Cotton Gauze for the EU: 15 Test Items vs YY/T 0331

EN 14079 Absorbent Cotton Gauze for the EU: 15 Test Items vs YY/T 0331

Gauze shipped to the European Union may need testing to EN 14079:2003, which sets 15 requirements and test methods for absorbent cotton gauze and absorbent cotton and viscose gauze. Its items look almost identical to China's YY/T 0331, but scope, definitions and additional requirements differ. Based on the public preview of the standard, this article explains each test and its acceptance criteria, the rules for cotton and viscose gauze and the comparison with YY/T 0331-2024, then sets out the accredited scope of Shanghai SUNGO Medical Technology Co., Ltd. and sample preparation. The US FDA route is covered in a separate article.

Key points

  • What the standard is. EN 14079:2003 was prepared by CEN/TC 205, Non-active medical devices, of the European Committee for Standardization (CEN), approved on 21 February 2003 and dated April 2003. It covers absorbent cotton gauze and absorbent cotton and viscose gauze, but not gauze impregnated with a pharmaceutical substance. On 22 October 2026 the BSI catalogue listed BS EN 14079:2003 as current.
  • The 15 requirements. Clauses 4.1 to 4.15, from fibre identification to sulphated ash (listed in the table below), with test methods in clauses 5.2 to 5.16.
  • How results are judged. Sinking time, ether-soluble substances, water-soluble substances, loss on drying and surface active substances each have one limit. Sulphated ash has four limits by material. Thread count, mass per square metre and minimum breaking load are read from Tables 1 and 2 by gauze type.
  • Relationship with YY/T 0331-2024. Both standards use the same item names and clause numbers for 4.1 to 4.15 and 5.2 to 5.16. YY/T 0331-2024 adds ethylene oxide (EO) residue, bioburden, sterility, and packaging and storage; EN 14079 has none of these. Scope and definitions also differ.
  • SUNGO Shanghai's accreditation. CNAS (L22083) accredits all 15 items of EN 14079:2003, clauses 5.2 to 5.16, with the test object listed as absorbent cotton gauze. IAS (TL-1019) lists EN 14079 as a whole standard, with no edition year. EN 14079 is not in the CMA schedule. Ether-soluble substances, water-soluble substances, loss on drying and sulphated ash are in SUNGO's CNAS rows for EN 14079:2003 but not in its CNAS rows for YY/T 0331-2024.

The standard and its scope

Sources: the public preview of SIST EN 14079:2003 (the Slovenian adoption, stated to be identical to EN 14079:2003) and the BSI catalogue, both checked on 22 October 2026.

Basic facts

  • Title: Non-active medical devices - Performance requirements and test methods for absorbent cotton gauze and absorbent cotton and viscose gauze.
  • Prepared by CEN/TC 205 (secretariat: BSI); approved by CEN on 21 February 2003; EN dated April 2003. The EN cover gives ICS 11.120.20 (wound dressings and compresses). The UK adoption, BS EN 14079:2003, was published on 13 May 2003.
  • Clause 2 states there are no normative references. Normative Annex A gives the reagents for judging the coloration of liquids, and normative Annex B the preparation of test solution S.
  • Clause 6 covers the test report; informative Annex ZA relates the clauses to the essential requirements of EU Directives.

Scope and the three types of gauze

Clause 1 covers absorbent cotton gauze and absorbent cotton and viscose gauzes and excludes gauzes impregnated with a pharmaceutical substance. Clause 3 defines three products:

  • Absorbent cotton gauze: plain-weave cotton cloth, bleached and purified, white and practically odourless, with no more than slight traces of leaf, seed-coat or similar impurities, and reasonably free from weaving defects.
  • Absorbent cotton ribbon gauze: woven cloth supplied in continuous ribbons of various widths with fast selvedges; the cotton threads may be purified, bleached and made absorbent before or after weaving.
  • Absorbent cotton and viscose ribbon gauze: also continuous ribbon with fast selvedges, with cotton threads in the warp and viscose, or combined cotton and viscose, threads in the weft.

Its link to the European Pharmacopoeia

The introduction explains that these gauzes were once described in European Pharmacopoeia monographs, removed when the Medical Device Directive 93/42/EEC came in. The foreword calls the standard a reformatting of those monographs, in line with the essential requirements in Annex I of the Directive, and says it does not describe the full state of the art. Hence its pharmacopoeia style: microscopy, colour reactions and reference solutions.

The 15 tests, one by one

The last column is our own summary. Verdicts follow the text of the standard.

ItemRequirementMethodWhat the requirement says
Fibre identification4.15.2Cotton fibres meet IDENT tests A, B and C. In cotton and viscose gauze, cotton fibres meet A and C and viscose fibres B; IDENT D separates lustrous from matt viscose where needed
Acidity and alkalinity4.25.3Test solution S with phenolphthalein, and a second portion with methyl orange: neither may be pink
Foreign fibres4.35.4Only occasionally a few isolated foreign fibres allowed
Fluorescence4.45.5Shall conform to 5.5
Thread count4.55.6Warp and weft threads per 100 mm per Tables 1 and 2
Mass per square metre4.65.7In g/m², per Tables 1 and 2
Minimum breaking load4.75.8In N per 50 mm, per Tables 1 and 2
Sinking time4.85.9Not more than 10 s
Ether-soluble substances4.95.10Not more than 0.50 %
Surface active substances4.105.11Froth height above the liquid after 300 s not more than 2 mm
Water-soluble substances4.115.12Not more than 0.50 %
Starch and dextrin4.125.13No blue, violet, reddish or brownish colour
Extractable colouring matter4.135.14Not more intensely coloured than reference solution Y5, GY6, or a reference solution prepared from the blue primary solution as set out in 4.13; reference solutions in Annex A
Loss on drying4.145.15Loss of mass not more than 8.0 %
Sulphated ash4.155.16Per Table 3; the limit depends on the material

Clause 5.1 matters for sample submission: all tests are performed on the material in its final form, sterile or non-sterile, and all reagents are of analytical grade.

Identity and purity

For fibre identification, fibres freed from untwisted warp and weft threads are examined under a microscope and treated with reagents such as iodinated zinc chloride solution. For cotton and viscose gauze, under 5.2.2.2 warp fibres must meet that clause's IDENT A and C and weft fibres IDENT B and C; where the weft is a cotton/viscose mixture, some weft fibres must also meet IDENT A for cotton. Follow the wording of 5.2.2.2. IDENT D uses the sulphated ash residue: the solution from a product with matt viscose turns orange-yellow, with an intensity depending on the titanium dioxide content, while that from lustrous viscose does not change colour.

Acidity and alkalinity uses test solution S (Annex B): one portion with phenolphthalein, one with methyl orange, and neither may turn pink. Clause 5.3.3 adds sensitivity checks on the indicators. The methods for foreign fibres and fluorescence (5.4, 5.5) are outside the public preview; refer to the full text.

Fabric construction

Thread count, mass per square metre and minimum breaking load are read from tables by gauze type. Table 1 covers absorbent cotton gauze, with nine types named by threads per cm²: 12, 13 light, 13 heavy, 17, 18, 20, 22, 24 a and 24 b. Each type has warp and weft threads per 100 mm with tolerances, warp and weft minimum breaking load (N per 50 mm) and minimum mass (g/m²). For types 12 and 13 light the breaking load columns show a dash, so no breaking load is specified.

Table 2 covers the two ribbon gauzes, with types 22 a, 22 b and 24 a, and gives only the warp-way breaking load. A footnote widens the warp thread tolerance for ribbon gauze 25 mm, 50 mm and 12.5 mm wide. So tell us the product's type, and the width for ribbon gauze; otherwise these three items cannot be given a verdict.

Absorbency: sinking time

Sinking time is the only absorbency item. The limit is 10 s and the method is in clause 5.9. Keep samples dry and free from oil in transit and storage.

Chemical items: sulphated ash by material

Sulphated ash has four limits in Table 3: absorbent cotton gauze, not more than 0.40 %; type 13 light, not more than 0.75 %; absorbent cotton and matt viscose ribbon gauze, not more than 1.20 %; absorbent cotton and lustrous viscose ribbon gauze, not more than 0.45 %. State the material and type when submitting samples. Procedures from clause 5.4 onwards are outside the public preview; refer to the full text.

Cotton and viscose gauze

Cotton and viscose gauze is in the title of EN 14079, but by definition (clause 3.3) it is a ribbon gauze with a cotton warp and a viscose or combined cotton and viscose weft. The standard treats it separately in three places:

  • Split fibre identification. Clause 4.1.2 requires cotton and viscose fibres each to meet their own IDENT tests, with IDENT D for lustrous versus matt viscose.
  • Table 2, not Table 1. Type, thread count, breaking load and minimum mass come from Table 2.
  • Sulphated ash by viscose type. Matt and lustrous viscose have different limits.

China's YY/T 0331-2024 covers absorbent cotton gauze and absorbent cotton and viscose blended gauze, but excludes gauze whose yarn is entirely viscose and gauze containing drugs. Its definition of the blended gauze (clause 3.2) is not in the public preview; refer to the full text. If one product is to be judged against both standards, first check that it meets each standard's definition.

How EN 14079 maps to YY/T 0331-2024

YY/T 0331-2024 (Chinese title, in our translation: performance requirements and test methods for absorbent cotton gauze and absorbent cotton and viscose blended gauze) was published on 7 February 2024 and took effect on 1 March 2025, replacing the 2006 edition. In the two tables of contents, the first 15 requirements and test methods 5.2 to 5.16 have the same item names and clause numbers. The table also shows SUNGO Shanghai's CNAS accredited rows.

ItemEN 14079:2003YY/T 0331-2024SUNGO CNAS: EN 14079:2003SUNGO CNAS: YY/T 0331-2024
Fibre identification to sinking time (8 items)4.1-4.8 / 5.2-5.94.1-4.8 / 5.2-5.95.2-5.9 accredited5.2-5.9 accredited
Ether-soluble substances4.9 / 5.104.9 / 5.105.10 accreditedNot listed
Surface active substances4.10 / 5.114.10 / 5.115.11 accredited5.11 accredited
Water-soluble substances4.11 / 5.124.11 / 5.125.12 accreditedNot listed
Starch and dextrin; extractable colouring matter4.12, 4.13 / 5.13, 5.144.12, 4.13 / 5.13, 5.145.13, 5.14 accredited5.13, 5.14 accredited
Loss on drying4.14 / 5.154.14 / 5.155.15 accreditedNot listed
Sulphated ash4.15 / 5.164.15 / 5.165.16 accreditedNot listed
EO residueNo requirement4.16 / 5.17-Accredited row cites GB/T 14233.1-2022 clause 9.4
BioburdenNo requirement4.17 / 5.18, Annex C-Annex C accredited
SterilityNo requirement4.18-Not listed
Clause 6Test reportPackaging and storage-Not listed

Other differences

  • Scope. EN 14079 excludes only gauze impregnated with a pharmaceutical substance; YY/T 0331-2024 also excludes gauze whose yarn is entirely viscose.
  • Definitions. EN 14079 defines three products, including two ribbon gauzes, and describes the gauze as bleached to a good white. YY/T 0331-2024 defines absorbent cotton gauze as a continuous woven fabric of cotton yarn that has been scoured, bleached or dyed, and purified.
  • References. EN 14079 has none. YY/T 0331-2024 references GB/T 14233.1, GB/T 19633.1, GB/T 19973.1, YY/T 0615.1 and the Chinese Pharmacopoeia, 2020 edition, Volume II.
  • Three items changed in China. The YY/T 0331-2024 foreword says that, compared with 2006, the surface active substances requirement and the test methods for extractable colouring matter and loss on drying were changed. Whether the 2024 wording matches EN 14079, and how the YY/T 0331-2024 limits compare, cannot be checked from the public previews; compare the full texts clause by clause.

In practice, a YY/T 0331 report cannot simply be used as an EN 14079 report, or the other way round: a report is judged against the standard it names. If you need both, say so when ordering. The changes between the 2006 and 2024 editions of YY/T 0331 are covered in a separate article on this site.

Four chemical items: in the CNAS rows for EN 14079, not for YY/T 0331-2024

Ether-soluble substances, water-soluble substances, loss on drying and sulphated ash are still YY/T 0331-2024 requirements, but they are not in SUNGO Shanghai's CNAS or CMA rows for YY/T 0331-2024; its CNAS rows for EN 14079:2003 include all four (5.10, 5.12, 5.15, 5.16). The IAS scope lists YY/T 0331 as a whole standard with no edition year; confirm the edition before an IAS report is issued. For absorbent cotton gauze ordered to EN 14079, these four can be reported under the CNAS mark; for YY/T 0331-2024 orders, check with the laboratory first which type of report can be issued.

What EN 14079 does not cover

Judging by its table of contents, EN 14079 covers only the physical, chemical and fabric properties of the gauze, with no requirements for EO residue, bioburden, sterility, packaging or biological evaluation. These need other standards, chosen by how the product is supplied and sterilized and its intended use.

  • Sterilization and packaging. SUNGO Shanghai offers testing services in both areas, with scope as shown in the respective accredited rows. See cleaning, disinfection and sterilization validation and sterile medical device packaging validation.
  • Biological tests. Tests such as cytotoxicity, sensitization and irritation are subcontracted by SUNGO to laboratories with the appropriate qualifications. Under accreditation rules, the client's agreement is obtained before subcontracting, and the report identifies the subcontracted items and the subcontractor.

Accredited scope

Rows are taken from the EN 14079:2003 entries for the test object "absorbent cotton gauze" in the CNAS (L22083) scope of Shanghai SUNGO Medical Technology Co., Ltd., plus the EN 14079 entry in its IAS (TL-1019) scope. The limitation field is blank in every CNAS row.

ItemStandard and clauseNotes / limitations
Fibre identificationEN 14079:2003 5.2CNAS; none stated
Acidity and alkalinityEN 14079:2003 5.3CNAS; none stated
Foreign fibresEN 14079:2003 5.4CNAS; none stated
FluorescenceEN 14079:2003 5.5CNAS; none stated
Thread countEN 14079:2003 5.6CNAS; none stated
Mass per square metreEN 14079:2003 5.7CNAS; none stated
Minimum breaking loadEN 14079:2003 5.8CNAS; none stated
Sinking timeEN 14079:2003 5.9CNAS; none stated
Ether-soluble substancesEN 14079:2003 5.10CNAS; none stated
Surface active substancesEN 14079:2003 5.11CNAS; none stated
Water-soluble substancesEN 14079:2003 5.12CNAS; none stated
Starch and dextrinEN 14079:2003 5.13CNAS; none stated
Extractable colouring matterEN 14079:2003 5.14CNAS; none stated
Loss on dryingEN 14079:2003 5.15CNAS; none stated
Sulphated ashEN 14079:2003 5.16CNAS; none stated
Whole standardEN 14079 (no edition year)IAS; confirm the edition with the laboratory before an IAS report is issued

EN 14079 is not in SUNGO Shanghai's CMA (240015340087) schedule.

Where the testing is done and under which accreditation

EN 14079 testing is carried out by Shanghai SUNGO Medical Technology Co., Ltd. at its laboratory at 958 Kangqiao East Road, Pudong New Area, Shanghai. Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity whose accreditation does not cover gauze. See masks, gloves, dressings and gauze testing.

  • CNAS. All 15 items are within the CNAS scope, with the test object listed as absorbent cotton gauze, so a CNAS-marked report can be issued for that product; for cotton and viscose ribbon gauze, confirm with the laboratory before ordering. The CNAS certificate states that CNAS is a member of the ILAC and APAC mutual recognition arrangements.
  • IAS. The IAS scope lists EN 14079 as a whole standard with no edition year; confirm the edition with the laboratory before an IAS report is issued. The IAS certificate carries the ilac-MRA mark.
  • CMA. EN 14079 is not in the CMA schedule, so no CMA-marked report can be issued for it. The CMA covers only scheduled items and is valid only within the People's Republic of China.

Acceptance of a report by any regulator, notified body or buyer outside China is for that party to decide. For what each mark proves, see How to Check an Accreditation Scope and Read the Mark on a Report.

Preparing samples

  • Standard and items. Write EN 14079:2003 on the order form, list all 15 items or the ones you need, and say whether you need a CNAS or IAS mark.
  • Product category. Absorbent cotton gauze, cotton ribbon gauze, or cotton and viscose ribbon gauze; for the last, state whether the viscose is lustrous or matt.
  • Type and size. The Table 1 or Table 2 type, and the width for ribbon gauze.
  • Final form. Under clause 5.1, send sterile products after sterilization and non-sterile products as supplied, in original packaging.
  • YY/T 0331 as well? Say so in advance.
  • Other items. Sterilization, packaging and biological items are not part of EN 14079; list them separately when requesting a quote.
  • Sample quantity. The laboratory will set this from the standard and your item list. Keep samples dry and protected from contamination.

Frequently asked questions

We have a YY/T 0331-2024 report. Do we still need EN 14079 testing?

The 15 items and clause numbers correspond, but scope and definitions differ and some clauses may differ in content, so one report cannot stand in for the other. If the target market asks for EN 14079, order testing to EN 14079.

Can cotton and viscose gauze be tested to EN 14079?

The standard applies if the product meets its definition: cotton in the warp, viscose or combined cotton and viscose in the weft. SUNGO's CNAS rows list the test object as absorbent cotton gauze, so confirm with the laboratory before ordering whether a CNAS-marked report can be issued for this product; the IAS scope lists EN 14079 as a whole standard with no edition year. State the viscose type and the Table 2 type.

Can an EN 14079 report carry the CMA mark?

No. EN 14079 is not in SUNGO Shanghai's CMA schedule. For absorbent cotton gauze a CNAS-marked report can be issued, and an IAS-marked report once the edition is confirmed.

Does EN 14079 cover sterility or biocompatibility?

No. Biological tests are subcontracted by SUNGO to laboratories with the appropriate qualifications, with the client's prior agreement, and the report identifies the subcontracted items and the subcontractor.

Is EN 14079:2003 the current edition?

On 22 October 2026 the BSI catalogue listed BS EN 14079:2003 as current. Its foreword says it does not describe the full state of the art, so choose test items with your technical documentation and your buyer's requirements in mind.

To request a quote, use our contact page to send the standard (EN 14079:2003), the items you need, the product category and type (with the viscose type for cotton and viscose products), how the product is supplied, what the report is for and which mark you need. The laboratory will reply item by item on the report types it can issue within its accredited scope.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.