ISO 17665:2024 is the moist heat sterilization standard ISO published in March 2024 as a first edition. This one document replaces ISO 17665-1:2006, ISO/TS 17665-2:2009 and ISO/TS 17665-3:2013, bringing requirements and guidance together. Adoption differs by market: the EU lists EN ISO 17665:2024 as an MDR harmonised standard, the FDA recognizes both the old and new editions during a transition period, and the UK designated standards list and China's GB 18278.1-2015 still correspond to the 2006 edition. Below: what changed, which edition each market recognizes, and which edition to cite when validating steam sterilization parameters in a reusable device's instructions for use (IFU).
Key points
- One document, no more parts. The title of ISO 17665:2024 no longer carries "Part 1". The foreword states that it cancels and replaces the three documents above, and it lists only one main change: ISO 17665-1, ISO/TS 17665-2 and ISO/TS 17665-3 were combined into a single standard.
- United States: the FDA Recognized Consensus Standards database entered ISO 17665:2024 on 2024-05-29 (recognition number 14-601, complete standard). As of November 13, 2026, the older entries 14-333 (ISO 17665-1:2006) and 14-277 (ISO/TS 17665-2:2009) remain in the database, and the FDA accepts declarations of conformity to them in premarket submissions until July 4, 2027.
- European Union: Commission Implementing Decision (EU) 2026/193 was adopted on January 28, 2026 and published in the Official Journal on January 30, 2026, entering into force that day. It lists EN ISO 17665:2024 as an MDR harmonised standard (new entry 41).
- United Kingdom: in version 2 of the consolidated designated standards list, published on January 17, 2026, moist heat sterilization is still covered by EN ISO 17665-1:2006. The 2024 edition is not listed.
- China: GB 18278.1-2015 is a mandatory national standard that is an identical adoption of ISO 17665-1:2006. As of November 13, 2026, the revision project that adopts ISO 17665:2024 with modifications (plan 20252031-Q-464) is at the approval stage.
- SUNGO: Shanghai laboratory only. IAS (TL-1019) accredits ISO 17665 and GB 18278.1 as whole standards, with no edition year. CMA (240015340087) covers only the GB 18278.1-2015 moist heat sterilization effectiveness test, limited to Annex C 2.2a). The Shanghai CNAS scope excludes moist heat sterilization, and the Hefei laboratory's certificates do not cover these standards.
ISO 17665:2024 scope: which moist heat processes are covered
The full title is Sterilization of health care products — Moist heat — Requirements for the development, validation and routine control of a sterilization process for medical devices. According to the foreword, it was prepared by ISO/TC 198, Sterilization of health care products, in collaboration with CEN/TC 204, Sterilization of medical devices, under the Vienna Agreement between ISO and CEN.
Clause 1 states that the document provides requirements and also contains guidance intended to explain the requirements in the normative sections, and that it considers application in both industrial and health care settings. Covered processes include, but are not limited to:
- saturated steam sterilization in which air is removed by passive purging (gravity displacement principle);
- saturated steam sterilization in which air is removed actively (dynamic air removal, pre-vacuum or fractionated vacuum principle);
- contained product sterilization in which heat transfer is achieved by steam or steam-air mixtures;
- contained product sterilization in which heat transfer is achieved by water sprays;
- contained product sterilization in which heat transfer is achieved by water immersion.
Four exclusions apply: processes for inactivating the agents of spongiform encephalopathies such as scrapie, BSE and Creutzfeldt-Jakob disease; processes combining moist heat with other biocidal agents such as formaldehyde; a specific requirement for designating a device as "sterile", which is left to national or regional rules (EN 556-1 and ANSI/AAMI ST67 are cited as examples); and occupational safety in sterilization facilities.
The scope is organized differently from 2006. The FDA database abstract of ISO 17665-1:2006 lists five systems: saturated steam venting, saturated steam active air removal, air-steam mixtures, water spray and water immersion. The new edition groups them into two families, saturated steam and contained product sterilization. Clause 6 (process and equipment characterization) and Clause 10 (evaluation of additional data in routine control) have separate subclauses for each family.
ISO 17665:2024 vs ISO 17665-1:2006: the three-in-one structure
The foreword states that this first edition cancels and replaces ISO 17665-1:2006, ISO/TS 17665-2:2009 and ISO/TS 17665-3:2013, which have been technically revised. The only main change it lists is the merger into a single standard. The table below compares the role of each old document with where its content sits in the 2024 edition.
| Former document | Original role | In ISO 17665:2024 |
|---|---|---|
| ISO 17665-1:2006 | International Standard with requirements for development, validation and routine control of a moist heat sterilization process | Requirements sit in normative Clauses 4 to 12, and conformance is claimed against the normative parts |
| ISO/TS 17665-2:2009 | Technical Specification giving guidance on the application of ISO 17665-1 | Guidance appears as informative annexes in the same document, for example Annex A (principles and rationales for requirements), Annex F (health care facilities) and Annex H (industrial settings) |
| ISO/TS 17665-3:2013 | Technical Specification giving guidance on designating a medical device to a product family and processing category for steam sterilization | Annex G gives guidance on designating a medical device to a product family and processing category for moist heat sterilization, the same subject as the former TS 17665-3 |
The normative clauses run as follows: Clause 4 General; Clause 5 Sterilizing agent characterization (sterilizing agent, microbicidal effectiveness, effects on materials, environmental consideration); Clause 6 Process and equipment characterization (saturated steam processes, contained product processes, equipment); Clause 7 Product definition; Clause 8 Process definition; Clause 9 Validation (installation qualification, operational qualification, performance qualification, and review and approval of validation); Clause 10 Routine monitoring and control; Clause 11 Product release from sterilization; and Clause 12 Maintaining process effectiveness (including recalibration, equipment maintenance, requalification and assessment of change).
Annexes A to H are all informative: A, principles and rationales for requirements; B, a process primarily based on microbiological inactivation; C, a process primarily based on measuring physical parameters; D, example cycles; E, temperature and pressure of saturated steam; F, guidance for health care facilities; G, product family and processing category designation; H, guidance for industrial settings.
A few points in the Introduction matter in practice for manufacturers and health care facilities:
- Conformance is claimed against the normative parts; the informative guidance is not a conformance checklist, and other effective methods can be used.
- Annex F and Annex H mirror the numbering and structure of the normative clauses.
- Table A.1 summarizes the purpose of each normative section and suggests roles and responsibilities for the organizations and personnel involved in development, validation and routine control.
- Products used in health care and presented for sterilization according to the IFU are best regarded as special cases (the Introduction cites ISO 17664-1): even after cleaning they can carry varied microorganisms and organic or inorganic residues, so validation and control of cleaning and disinfection get particular attention.
Most definitions in Clause 3 are taken from ISO 11139:2018 and its 2024 amendment; some, such as contained product sterilization, D value and equilibration time, were modified or given added notes. The normative references are ISO 11138-1:2017 and ISO 11138-3:2017 (biological indicators), all parts of ISO 11140 (chemical indicators), ISO 11607-1 and ISO 11607-2 (packaging for terminally sterilized medical devices), and ISO 11737-1 and ISO 11737-2 (microbiological methods). This article does not paraphrase how specific approaches, such as biological indicator, overkill or half-cycle methods, are named and developed in the new edition. Refer to the text of Annexes B and C.
What it means for manufacturers: which edition to cite for IFU sterilization validation
The IFU of a reusable device must set out the processing steps, and ISO 17664-1 and ISO 17664-2 specify what processing information the manufacturer has to provide. If the IFU gives steam sterilization parameters, validation data must support them. For the cleaning and disinfection steps, see our article Manual Cleaning Validation of Reusable Medical Devices: AAMI ST98 and ISO 17664. This section covers only the moist heat sterilization step.
New projects. For EU technical documentation, EN ISO 17665:2024 is the MDR harmonised standard. For US submissions, the FDA recognizes ISO 17665:2024 as a complete standard and accepts declarations of conformity to the old edition only until July 4, 2027. Both point to the 2024 edition. For the UK and China, the designated and mandatory national standards still correspond to the 2006 edition, so your file can map the 2024 edition to EN ISO 17665-1:2006 or GB 18278.1-2015 to head off review questions.
Existing validation reports. If earlier reports were written against ISO 17665-1:2006 and ISO/TS 17665-2:2009, consider a gap assessment recording what is restated under the new edition, which test data carry over, and why. Since the foreword lists only the merger as a main change, focus on structure and references: a product family rationale following Annex G, each process clearly identified as saturated steam or contained product, and referenced standards for indicators, packaging and microbiological methods matching the new normative references.
Sterility testing. ISO 11737-2 is a normative reference of the new edition. For how sterility tests performed in defining and validating a sterilization process differ from pharmacopoeial sterility testing, see ISO 11737-2 vs ChP 1101: Sterility Tests for Validation vs Product Testing.
Whether a regulator or notified body accepts a test report is for that body to decide.
ISO 17665 edition by market: FDA, EU MDR, UK and China
| Market | Edition recognized or designated | Key dates and status |
|---|---|---|
| US FDA | 14-601: ISO 17665 First edition 2024-03 (complete standard); 14-333: ISO 17665-1:2006; 14-277: ISO/TS 17665-2:2009 | 14-601 entered 2024-05-29; declarations of conformity to 14-333 and 14-277 accepted until 2027-07-04 |
| EU MDR | EN ISO 17665:2024 (entry 41 in the Annex to (EU) 2021/1182) | (EU) 2026/193 adopted 2026-01-28, published and in force 2026-01-30 |
| UK (England, Scotland and Wales) | EN ISO 17665-1:2006 | Designated since 2021-01-01; version 2 of the consolidated list (2026-01-17) does not include the 2024 edition |
| China | GB 18278.1-2015 (mandatory, identical to ISO 17665-1:2006) | In force since 2017-01-01; revision plan 20252031-Q-464 at the approval stage as of 2026-11-13 |
US FDA. The 14-601 record sets out the transition. Recognition of 14-277 and 14-333 will be superseded by 14-601. The FDA accepts declarations of conformity to those two entries in support of premarket submissions until July 4, 2027, and will not accept them after the transition period. The 14-601 record lists related regulations including 21 CFR 880.2800 (sterilization process indicators), 880.6850 (sterilization wrap, containers, trays and other accessories) and 880.6880 (steam sterilizers). Recognition applies to the edition listed in each entry.
EU MDR. Recitals (4) and (5) of Decision 2026/193 state that CEN and Cenelec, acting on the Commission's standardisation request, revised a group of standards including EN ISO 17665-1:2006, resulting in EN ISO 17665:2024, which the Decision adds as entry 41 in the Annex to Implementing Decision (EU) 2021/1182. The consolidated text of 2021/1182 dated October 20, 2025 on EUR-Lex contains no 17665 entry, so this is the first time an ISO 17665 standard has appeared on the MDR harmonised list. Recital (11) explains that the presumption of conformity applies from the date the reference is published in the Official Journal. Which MDR requirements the presumption covers is defined in the EN text itself. If you have validations completed to the old edition, discuss them with your notified body early.
United Kingdom. The GOV.UK page "Designated standards: medical devices" was updated on January 17, 2026. In version 2 of the consolidated list, which supports the Medical Devices Regulations 2002, EN ISO 17665-1:2006 has been designated since January 1, 2021 and has no removal date. EN ISO 17665:2024 is not on the list. The same update designated EN ISO 17664-1:2021 and EN ISO 17664-2:2023 and removed EN ISO 17664:2017, but did not update the moist heat sterilization entry. If your documentation follows the 2024 edition, confirm with your conformity assessment body how to demonstrate conformity with UK requirements.
China. According to the National Public Service Platform for Standards Information, GB 18278.1-2015 (Sterilization of health care products — Moist heat — Part 1) is a mandatory national standard that is an identical adoption of ISO 17665-1:2006. It was published on 2015-12-10, took effect on 2017-01-01, and was confirmed as still valid at its review on 2022-10-27. Revision plan 20252031-Q-464 was issued on 2025-06-30 as a mandatory national standard plan. It adopts ISO 17665:2024 with modifications and will fully replace GB 18278.1-2015. The project period is 16 months, and the planned implementation date is 36 months after publication. As of November 13, 2026, the platform shows the plan at the approval stage. China also has a recommended industry standard, YY/T 1600-2018, which sets requirements for grouping devices into product families and process categories by their attributes and selecting the moist heat sterilization program accordingly. Until the new national standard is published and takes effect, GB 18278.1-2015 remains the mandatory national standard in force.
SUNGO accreditation scope: ISO 17665, GB 18278.1 and ISO 17664
| Standard | Accreditation | Item / clause | Notes / limitations |
|---|---|---|---|
| ISO 17665 | IAS TL-1019 (Shanghai) | Whole standard | No edition year on the scope document; to receive an IAS report to the 2024 edition, confirm the edition with the laboratory when requesting a quote |
| GB 18278.1 | IAS TL-1019 (Shanghai) | Whole standard | No edition year on the scope document |
| GB 18278.1-2015 | CMA 240015340087 (Shanghai) | Moist heat sterilization effectiveness test | Annex C 2.2a) only; the CMA certificate is valid only within the People's Republic of China |
| ISO 17664-1 | IAS TL-1019 (Shanghai) | Partial | Only accredited for manual cleaning validation, manual disinfection (disinfectant), half cycle sterilization method |
| ISO 17664-2 | IAS TL-1019 (Shanghai) | Partial | Same limitation as ISO 17664-1 |
| ISO 11737-1 | IAS TL-1019 (Shanghai) | Whole standard | No edition year on the scope document; a normative reference of ISO 17665:2024 |
| ISO 11737-2 | IAS TL-1019 (Shanghai) | Whole standard | No edition year on the scope document; a normative reference of ISO 17665:2024 |
The IAS rows follow the TL-1019 scope document, which carries no edition years; the CMA row follows the CMA schedule. Standards, clauses and methods not in this table are outside SUNGO's accreditation scope.
Where SUNGO does this work and under which accreditation
Moist heat sterilization work is handled by the laboratory of Shanghai SUNGO Medical Technology Co., Ltd., located at 958 Kangqiao East Road, Pudong New Area, Shanghai. Hefei SUNGO Medical Technology Co., Ltd. is a separate company, and its certificates do not cover the standards in this article. For an overview of the service, see cleaning, disinfection and sterilization validation.
- IAS-accredited reports: TL-1019 reports to ISO 17665 or GB 18278.1. IAS is a signatory to the ILAC MRA. Because the IAS scope carries no edition year, if you need a report to ISO 17665:2024 or to ISO 17665-1:2006 specifically, state the edition when you request a quote so the laboratory can confirm it.
- Domestic use with the CMA mark: only the moist heat sterilization effectiveness test of GB 18278.1-2015, limited to Annex C 2.2a). Anything outside the CMA schedule cannot carry the CMA mark.
- CNAS: the Shanghai CNAS scope does not include moist heat sterilization, so these reports do not carry the CNAS mark.
- IFU processing validation: for ISO 17664-1 and ISO 17664-2, the Shanghai IAS scope covers only manual cleaning validation, manual disinfection (disinfectant) and the half cycle sterilization method; other cleaning, disinfection and sterilization validation methods are outside the scope.
Which tests are performed and who provides the sterilizer are confirmed item by item in the project protocol.
Preparing a submission
- Target markets and the edition to cite (ISO 17665:2024, ISO 17665-1:2006 or GB 18278.1-2015), plus the report type needed (IAS or CMA).
- Device information: drawings and a materials list, marking lumens, blind holes, hinges, detachable parts and other areas steam may not reach easily.
- Process information: whether it is saturated steam sterilization (passive or active air removal) or contained product sterilization, and the sterilization parameters specified in the IFU or process documents.
- Your rationale for product families and processing categories, and why the representative product was chosen.
- Packaging information: the sterile barrier system or wrapping method (ISO 11607-1 and ISO 11607-2 are normative references of the new edition).
- Cleaning and disinfection steps before sterilization; for reusable devices, as given in the IFU.
- Earlier validation reports, if completed to the 2006 edition, to support a gap assessment.
FAQ
Can ISO 17665-1:2006 still be used?
ISO has cancelled and replaced it with ISO 17665:2024, and the EU MDR harmonised list carries EN ISO 17665:2024. The FDA accepts declarations of conformity to it (14-333) until July 4, 2027, the UK designated list still includes EN ISO 17665-1:2006, and GB 18278.1-2015 remains an identical adoption of it.
Does ISO 17665:2024 still have Part 2 and Part 3?
No. The guidance from ISO/TS 17665-2 and the product family content from ISO/TS 17665-3 are informative annexes in the same document. Product families and processing categories are in Annex G.
When did EN ISO 17665:2024 become an MDR harmonised standard?
Decision (EU) 2026/193 was published in the Official Journal on January 30, 2026 and entered into force that day. From that date, EN ISO 17665:2024 gives presumption of conformity as an MDR harmonised standard.
Will GB 18278.1 be revised to ISO 17665:2024?
Revision plan 20252031-Q-464 adopts ISO 17665:2024 with modifications. As of November 13, 2026 it is at the approval stage, with implementation planned 36 months after publication. Official announcements will set the actual publication and implementation dates.
Can SUNGO issue a report to ISO 17665:2024?
Shanghai IAS (TL-1019) accredits ISO 17665 as a whole standard with no edition year, so confirm the 2024 edition with the laboratory when requesting a quote. CMA covers only GB 18278.1-2015 Annex C 2.2a), and CNAS does not include moist heat sterilization.
To arrange moist heat sterilization testing to ISO 17665 or GB 18278.1, or to plan manual cleaning, manual disinfection and half-cycle sterilization validation for a reusable device IFU, send your product information, target markets and the edition you need to cite through our contact page. The laboratory will confirm the accreditation scope and report type before quoting.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.