EN 455-2:2024 is the European standard for the physical properties of single-use medical gloves, with two groups of requirements: dimensions (length and width) and strength (force at break, including force at break after challenge testing). It applies to surgical and examination/procedure gloves, was published in May 2024 to replace EN 455-2:2015, and has been a harmonised standard under the EU MDR since 9 April 2025. This article covers what it tests, what changed, its status in the EU and Great Britain, and which report marks SUNGO can issue.
Key points
- What is tested: Clause 4 of EN 455-2:2024 covers dimensions (4.2 length, 4.3 width), Clause 5 covers strength (5.2 force at break, 5.3 force at break after challenge testing), Clause 6 is the test report and Clause 7 is labelling. The table of contents has no separate subclause for thickness.
- What changed: The foreword lists five main changes from the 2015 edition: revised normative references; 4.2 updated on recording the measured length ("median" removed); Clauses 5 and 6 updated; and Annex ZA updated for the EU MDR.
- EU: Commission Implementing Decision (EU) 2025/681 added EN 455-2:2024 to the MDR harmonised standards. It was published in the Official Journal on 9 April 2025 and entered into force that day.
- Great Britain: Version 2 of the UK designated standards list for medical devices designates EN 455-2:2024 from 17 January 2026 and removes EN 455-2:2009+A2:2013 on the same date.
- SUNGO reports: The Shanghai laboratory holds CNAS accreditation (L22083) for EN 455-2:2024 "physical properties", and IAS accreditation (TL-1019) for EN 455-2 and BS EN 455-2 as whole standards, with no edition year stated. No CMA report is available, and the Hefei laboratory's accreditation does not include the EN 455 series.
EN 455-2:2024 scope: which gloves it covers
Clause 1 states that the document specifies requirements and gives test methods for the physical properties of single-use medical gloves (surgical gloves and examination/procedure gloves). The aim is to make sure the gloves provide, and keep providing in use, adequate protection against cross-contamination for both patient and user.
Clause 3 gives three definitions; check which applies before submitting samples:
- Medical gloves for single use: gloves intended for use in the medical field to protect patient and user from cross-contamination.
- Surgical gloves: sterile, anatomically shaped medical gloves with the thumb positioned towards the palmar surface of the index finger rather than lying flat, intended for use in invasive surgery.
- Examination/procedure gloves: sterile or non-sterile medical gloves, which can be anatomically shaped, intended for medical examinations, diagnostic and therapeutic procedures, and handling contaminated medical material.
Clause 1 does not specify a lot size, but notes that very large lots can be hard to distribute and control and recommends a maximum individual production lot size of 500 000. The scope of EN 455-4:2009 says the same.
The 2024 edition has four normative references: EN ISO 15223-1:2021 and EN ISO 20417:2021 on symbols and information supplied by the manufacturer, ISO 23529:2016 on preparing and conditioning rubber test pieces for physical tests, and ISO 188:2023 on accelerated ageing and heat resistance tests for rubber.
EN 455-2:2024 clause structure
| Clause | Title | Notes |
|---|---|---|
| 1 | Scope | Surgical and examination/procedure gloves; lot size not specified |
| 2 | Normative references | EN ISO 15223-1:2021, EN ISO 20417:2021, ISO 23529:2016, ISO 188:2023 |
| 3 | Terms and definitions | Medical gloves for single use; surgical gloves; examination/procedure gloves |
| 4 | Dimensions (4.1 General, 4.2 Length, 4.3 Width) | Recording of measured length in 4.2 updated in 2024 |
| 5 | Strength (5.1 General, 5.2 Force at break, 5.3 Force at break after challenge testing) | Clause updated in 2024 |
| 6 | Test report | Clause updated in 2024 |
| 7 | Labelling | Labelling requirements; refer to the standard text |
| Annex ZA (informative) | Relationship with the general safety and performance requirements of Regulation (EU) 2017/745 | Updated for the MDR in 2024 |
EN 455-2 dimension requirements: length and width
Clause 4 of EN 455-2:2024 has two dimensional subclauses, 4.2 length and 4.3 width, after 4.1 General. The introduction to EN 455-4:2009 explains why: to work as a barrier, a medical glove must fit the hand properly, be free from holes and be strong enough not to fail in use. Put simply, dimensions relate to fit, Part 1 covers holes, and Clause 5 of Part 2 covers strength.
The only stated change to dimensions in 2024 is in 4.2: the foreword says the recording of the measured length has been updated and "median" removed. Length results given as a median in 2015-edition reports should not simply be carried over; refer to the 2024 text for how length is recorded.
On thickness: in the 2024 table of contents, Clause 4 has subclauses for length and width only; check the text for any other mention of thickness. Strength is expressed as force at break, so the result needs no thickness conversion. If a contract or another market's standard requires thickness, name that standard when you ask for a quote.
Measuring positions, instruments, number of gloves and acceptance criteria are set out in the standard; this article does not quote those values.
EN 455-2 force at break: as received and after challenge testing
Clause 5 "Strength" has three subclauses: 5.1 General, 5.2 Force at break, and 5.3 Force at break after challenge testing. Force at break is the force at which a test piece breaks under tension, expressed in newtons (N).
Subclause 5.2 measures force at break on unchallenged gloves. Normative reference ISO 23529:2016 covers preparing and conditioning rubber test pieces; where and how it applies is set out in the EN 455-2:2024 text.
Subclause 5.3 measures force at break after "challenge testing". ISO 188:2023 (accelerated ageing and heat resistance tests) is also a normative reference. For the challenge conditions, how ISO 188 is applied, the number of test pieces and the acceptance criteria, refer to the EN 455-2:2024 text.
Note that 5.3 is a test within EN 455-2, not a shelf-life determination; shelf life is covered by EN 455-4:2009. Its clause 4.1 requires gloves stored as the manufacturer instructs to comply with the EN 455 series until the end of their stated shelf life, and requires manufacturers to test the properties expected to change over time, at least force at break, freedom from holes and, for sterile gloves, pack integrity. Under clause 4.2, before a new or significantly modified product is placed on the market, a real-time study must have been completed, or started with an accelerated ageing study completed.
Annex A of EN 455-4:2009 (real-time studies) takes gloves from a minimum of three lots and, at intervals of one year or less, assesses the median force at break (13 gloves per interval) according to EN 455-2, freedom from holes according to EN 455-1, whether the glove is still fit for its intended purpose and, for sterile gloves, pack integrity. Each lot is tested and reported separately; lots may not be combined. EN 455-4 cites "EN 455 (all parts)" undated, so the latest edition applies. A new edition of EN 455-2 therefore affects both release testing and shelf-life studies.
The 2024 foreword does not itemise the Clause 5 changes. If force at break was tested to the 2015 edition, compare against the full 2024 text before deciding whether to retest.
EN 455-2:2024 vs EN 455-2:2015: what changed
EN 455-2:2024 was prepared by CEN/TC 205 "Non-active medical devices". CEN approved it on 15 April 2024, it was published in May 2024, and it supersedes EN 455-2:2015. The foreword required national adoption, by identical text or endorsement, and withdrawal of conflicting national standards by November 2024 at the latest. The main changes it lists are:
| Item | Change (per the 2024 foreword) | Impact on testing and reporting |
|---|---|---|
| a | Normative references revised | The 2024 edition cites EN ISO 15223-1:2021, EN ISO 20417:2021, ISO 23529:2016 and ISO 188:2023; apply the editions cited in the 2024 text |
| b | 4.2 updated on recording the measured length ("median" removed) | Length results may be expressed differently from 2015 reports |
| c | Clause 5 (Strength) updated | Check the full text for changes to force at break and force at break after challenge testing |
| d | Clause 6 (Test report) updated | Required report content updated; reports follow Clause 6 of the 2024 edition |
| e | Annex ZA updated for the MDR, Regulation (EU) 2017/745 | Shows how the standard relates to the MDR general safety and performance requirements; the scope of the presumption of conformity is set out in Annex ZA |
The foreword gives no detail on items c and d; base any gap assessment on the full 2024 text. National standards bodies record the old edition as withdrawn, for example: the Estonian standards body EVS shows EVS-EN 455-2:2015 as withdrawn from 15 May 2024, and BSI states that BS EN 455-2:2024 supersedes BS EN 455-2:2015, which is withdrawn.
EN 455-2:2024 status in the EU and Great Britain
European Union. Commission Implementing Decision (EU) 2025/681 of 8 April 2025 amends Implementing Decision (EU) 2021/1182, adding EN 455-1:2020+A2:2024, EN 455-2:2024 and other standards to the MDR harmonised list. Its recitals state that CEN and Cenelec revised EN 455-2:2015 under the Commission's standardisation request. The decision was published in the Official Journal on 9 April 2025 and entered into force that day. Under Article 8(1) of the MDR, devices that conform to harmonised standards whose references have been published in the Official Journal are presumed to conform to the requirements those standards cover.
The Commission's summary list of MDR harmonised standards, generated on 11 November 2026, shows three EN 455 entries: EN 455-1:2020+A2:2024 and EN 455-2:2024 (both from 9 April 2025) and EN 455-3:2023 (from 8 March 2024, under Decision (EU) 2024/815). EN 455-4:2009 is not on that list.
Great Britain. On 17 January 2026 the Department of Health and Social Care issued notice of publication 0129/26, and the UK government's designated standards page published version 2 of the consolidated list for medical devices the same day. The list supports the Medical Devices Regulations 2002 (S.I. 2002/618) and applies to England, Scotland and Wales. It gives EN 455-2:2024 a presumption of conformity from 17 January 2026, and EN 455-2:2009+A2:2013, designated since 1 January 2021, was removed on the same date. EN 455-2 appears in this list only as 2009+A2:2013 and 2024.
BSI publishes the UK implementation as BS EN 455-2:2024. Whether a report should cite EN 455-2 or BS EN 455-2 depends on the target market and customer; at SUNGO, BS EN 455-2 reports are issued under IAS only (see below).
This status is as of 13 November 2026 and may change; check the Commission and UK government websites before submitting technical documentation.
Force at break vs ASTM tensile strength
EN 455-2 expresses strength as force at break: a force in newtons that depends on both the strength of the material and the film thickness. US examination glove specifications work differently. For ASTM D6319 (nitrile examination gloves), the abstract on ASTM's website lists physical requirements before and after ageing, including tensile strength and ultimate elongation. Tensile strength is a stress (force per original cross-sectional area of the test piece, in MPa); calculating it needs the thickness, and it mainly reflects the material itself. The exact definition is in the relevant ASTM standard.
So EN 455-2 force at break and ASTM tensile strength results cannot be converted into each other or used interchangeably: the concepts, calculations and acceptance criteria all differ. A glove sold in both the EU and the US generally needs separate testing to each standard. For ASTM D6319 dimensions and tensile testing before and after ageing, see ASTM D6319 Nitrile Examination Gloves: Dimensions, Tensile and Leak Testing. For changes to China's national standard for examination gloves, see GB/T 10213-2025 and GB/T 24786-2025 Medical Examination Glove Testing Guide.
How EN 455-2 fits with EN 455-1, EN 455-3 and EN 455-4
According to the EN 455-2:2024 foreword, the EN 455 series has four parts: Part 1, freedom from holes; Part 2, physical properties; Part 3, biological evaluation; and Part 4, shelf life determination. Medical gloves exported to the EU usually need the parts planned together:
- EN 455-1 (freedom from holes): a water leak test checks gloves for holes. For how it compares with ASTM D5151, see ASTM D5151 Glove Hole Detection: EN 455-1 Water Leak Test and AQL Compared.
- EN 455-2 (physical properties): dimensions and force at break, as covered in this article.
- EN 455-3 (biological evaluation): SUNGO's accreditation covers only clause 4.4 (residual powder) under CNAS and clause 5.2 under IAS. Other biological evaluation tests are outside SUNGO's accreditation. Where they are needed, they are subcontracted to a suitably accredited laboratory with the client's prior agreement, and the report identifies the subcontracted items and the subcontractor. For a comparison of residual powder methods, see Residual Powder on Medical Gloves: ASTM D6124, GB/T 21869 and EN ISO 21171.
- EN 455-4 (shelf life): shelf-life studies repeat the force at break and freedom-from-holes tests, using the methods from Parts 2 and 1.
In practice, decide up front which lots go to release testing and which to ageing studies, to avoid running short of samples or mismatching lots.
SUNGO accreditation scope: EN 455 rows
The table reproduces the rows relating to the EN 455 series in the Shanghai laboratory's accreditation schedules.
| Item | Standard / clause | Notes / limitations |
|---|---|---|
| Medical gloves for single use: physical properties | EN 455-2:2024 (no clause breakdown in the schedule) | Shanghai, CNAS L22083 |
| Whole standard | EN 455-2 (no edition year) | Shanghai, IAS TL-1019; confirm the edition at quotation before a 2024-edition report is issued |
| Whole standard | BS EN 455-2 (no edition year) | Shanghai, IAS TL-1019; confirm the edition at quotation, as above |
| Medical gloves for single use: water leak test | EN 455-1:2020+A2:2024 | Shanghai, CNAS L22083 |
| Whole standard | EN 455-1, BS EN 455-1 (no edition year) | Shanghai, IAS TL-1019 |
| Medical gloves for single use: residual powder | EN 455-3:2023 clause 4.4 | Shanghai, CNAS L22083; 4.4 residual powder only |
| Partial | EN 455-3, BS EN 455-3 clause 5.2 (no edition year) | Shanghai, IAS TL-1019; only 5.2 |
| Medical gloves for single use: shelf life | EN 455-4:2009 | Shanghai, CNAS L22083 |
| Whole standard | EN 455-4, BS EN 455-4 (no edition year) | Shanghai, IAS TL-1019 |
No CMA report can be issued for the EN 455 series (see below), and the Hefei laboratory's accreditation does not include it.
Where SUNGO tests EN 455-2 and under which accreditation
EN 455-2 testing is carried out by the laboratory of Shanghai SUNGO Medical Technology Co., Ltd. at 958 Kangqiao East Road, Pudong, Shanghai. You can choose between these report marks:
- CNAS (L22083): EN 455-2:2024, item "physical properties".
- IAS (TL-1019): EN 455-2 or BS EN 455-2 as whole standards. The IAS scope gives no edition year, so confirm the edition at quotation before a 2024-edition report is issued. Reports citing BS EN 455-2 are IAS only.
- CMA: does not cover the EN 455 series. CMA applies only to the items in its schedule and is valid only within China.
SUNGO is accredited by CNAS and IAS, both signatories to the ILAC MRA. Whether a given overseas regulator or notified body accepts a particular report is for that body to decide, and SUNGO makes no guarantee.
EN 455-2 sample submission checklist
- Glove type: state whether the gloves are surgical or examination/procedure gloves, sterile or non-sterile, and the material (for example natural rubber latex, nitrile or PVC). The standard defines the two glove types separately, so state the type clearly to avoid applying the wrong criteria.
- Sizes: list the sizes and nominal dimensions, and which sizes need testing.
- Lot number and date of manufacture: for ageing-related tests, submit samples in their original packaging; EN 455-4 Annex A real-time studies condition gloves in their consumer packages.
- Edition and report mark: state whether you need EN 455-2:2024 (CNAS or IAS) or BS EN 455-2 (IAS).
- Combined planning: mention any EN 455-1 water leak test or EN 455-4 shelf-life study you also need, so samples and lots can be planned together.
- Sample quantity: depends on the sampling requirements of the standard and the tests selected; the laboratory will confirm it at quotation.
- Previous reports: send any 2015-edition reports so the retest scope can be set against the changes listed in the 2024 foreword.
FAQ
What is the difference between EN 455-2:2024 and EN 455-2:2015?
Per the 2024 foreword: revised normative references, 4.2 updated on recording the measured length ("median" removed), Clauses 5 and 6 updated, and Annex ZA updated for the MDR. Details of the Clause 5 and 6 changes are in the full text.
Can I still use an EN 455-2:2015 test report?
The 2015 edition has been superseded. Under the EU MDR, presumption of conformity rests on EN 455-2:2024 (from 9 April 2025). Whether an old report can still support the technical documentation is for the manufacturer to assess against the edition differences and agree with its notified body; SUNGO does not make that call for regulators.
Which edition of EN 455-2 does the UK recognise?
Version 2 of the UK designated standards list designates EN 455-2:2024 from 17 January 2026 and removes EN 455-2:2009+A2:2013 on the same date. It applies to England, Scotland and Wales (as of 13 November 2026).
Does EN 455-2 require glove thickness measurement?
In the 2024 table of contents, Clause 4 has subclauses for length and width only, and strength is expressed as force at break (N); check the text for any other mention of thickness. If another standard or a contract requires thickness, state it separately.
Can SUNGO issue a CMA report for EN 455-2?
No. CMA does not cover EN standards. The Shanghai laboratory can issue reports carrying the CNAS mark (EN 455-2:2024 "physical properties") or the IAS mark (EN 455-2 or BS EN 455-2 as whole standards).
If you need EN 455-2:2024 physical property testing, or want to plan it together with EN 455-1 and EN 455-4, send an enquiry through our contact page. Include the glove type, material, sizes, target market and the report mark you need (CNAS or IAS), and the laboratory will confirm the edition, tests and sample requirements. Our glove testing services are listed under medical masks, gloves and dressings testing.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.