ASTM D6319 Nitrile Examination Gloves: Dimensions, Tensile and Leak Testing

ASTM D6319 Nitrile Examination Gloves: Dimensions, Tensile and Leak Testing

For nitrile examination gloves exported to the US, ASTM D6319 is the standard most often cited in purchase contracts, FDA 510(k) files and inspection checklists. Buyers and manufacturers ask what it requires, how aged and unaged tensile testing works, how holes and powder are judged, and what to send for testing. Using public information from ASTM, the US FDA and the US Centers for Disease Control and Prevention (CDC), with clause references checked against the D6319-19 text, this guide explains what D6319 covers, the key points of tensile and thickness measurement, and which clauses fall within the CNAS and IAS scopes of Shanghai SUNGO Medical Technology Co., Ltd. We do not reproduce the numerical requirements; the ASTM text governs. External information was checked on 2026-10-22.

Key points

  • D6319 is a specification; the methods are referenced. The current edition is ASTM D6319-19(2023), the 2019 edition reapproved in 2023. The abstract on the ASTM website lists the required degree of sterility, freedom from holes, physical requirements before and after ageing, amount of powder, tensile strength and ultimate elongation. The test methods referenced are D412, D573, D3767, D5151 and D6124, with ISO 2859 for sampling.
  • FDA recognition is partial. FDA lists D6319-19(2023) as recognized consensus standard 6-446, with all text related to powdered gloves not recognized. Powdered patient examination gloves have been banned in the US since 2017-01-18, and 21 CFR 880.6250 states that the finished glove includes only residual powder from manufacturing.
  • FDA has a suggested 510(k) format. The suggested 510(k) format for examination gloves in FDA's Medical Glove Guidance Manual asks, under Device Description and Specifications, for minimum length, width, palm and finger thickness, tensile strength and ultimate elongation before and after aging, and the pinhole AQL, and whether the glove meets D6319 or an equivalent standard.
  • What SUNGO Shanghai covers. Under the test object "medical nitrile gloves", the CNAS scope lists D6319-19(2023) clause 7.3 (water-tightness test), clauses 7.4 and 7.5 (physical and dimensional properties) and clauses 7.6 and 7.7 (residual powder), with D5151-19(2023), D412-16(2021) and D3767-03(2020) as supporting methods. The IAS scope lists ASTM D6319, D5151 and D6124 as whole standards. D412 and D3767 are CNAS only, and no ASTM standard is in the CMA scope.
  • Aging sits inside clause 7.5. D6319 clause 7.5 sets the physical requirements before and after aging and calls for accelerated aging to D573, so tensile after aging is reported under clauses 7.4 and 7.5. D573 is not listed separately in the certificates and is not offered as a stand-alone aging test.

The standard and its scope

ASTM D6319-19(2023), Standard Specification for Nitrile Examination Gloves for Medical Application

  • Under ASTM Committee D11, Subcommittee D11.40; Annual Book of ASTM Standards Volume 09.02; 4 pages. The ASTM website shows it as Active, last updated 2023-09-28, and lists D6319-00a as the earliest edition.
  • Scope: nitrile rubber gloves used in medical examinations and diagnostic and therapeutic procedures (1.1); gloves that fit either hand, paired gloves and gloves by size, packaged sterile, packaged nonsterile or bulk nonsterile (1.2); similar to D3578 for rubber examination gloves (1.3); SI units are standard (1.4).
  • Material (Section 4): a lubricant meeting the U.S. Pharmacopeia requirements for absorbable dusting powder may be applied, other lubricants may be used if their safety and efficacy have been established (FDA does not recognize the powdered-glove text), and both glove surfaces shall be free of talc.

Documents referenced by D6319

Section 2 of D6319-19 lists ASTM D412, D573, D3578, D3767, D5151 and D6124, ISO 2859 and the U.S. Pharmacopeia, all undated; the NIOSH PPE standards database of the CDC lists the same references for D6319-19(2023). The method editions in the table are those shown as current on the ASTM website on 2026-10-22, and the last column gives where D6319 cites each document.

StandardTitleUse in D6319
ASTM D412-16(2021)Vulcanized Rubber and Thermoplastic Elastomers—TensionClause 7.5 tensile strength and ultimate elongation; clause 7.4.4 thickness gauge
ASTM D573-04(2025)Rubber—Deterioration in an Air OvenClause 7.5.2 accelerated aging
ASTM D3767-03(2020)Rubber—Measurement of DimensionsClause 7.4.4 thickness measurement
ASTM D5151-19(2023)Detection of Holes in Medical GlovesClause 7.3 freedom from holes
ASTM D6124-06(2022)Residual Powder on Medical GlovesClauses 7.6 and 7.7 powder
ASTM D3578Specification for Rubber Examination GlovesClause 1.3 (similar specification); clause 7.7.2 (glove area)
ISO 2859Sampling procedures for inspection by attributesSection 5 sampling
U.S. PharmacopeiaLatest editionClause 4.2 lubricant; clause 7.2 sterility test

FDA recognition and product code

  • Recognition: number 6-446, partial. All text related to powdered surgeon's gloves, powdered patient examination gloves and absorbable powder for lubricating a surgeon's glove is not recognized, because FDA has banned these devices.
  • Product code: the database links D6319 to LZA, defined in the FDA product classification database as a disposable patient examination glove made of nitrile rubber or synthetic polymers that may bear a trace amount of residual powder; 21 CFR 880.6250, Class I, 510(k).
  • Powder ban: the final rule published on 2016-12-19 (81 FR 91722) banned these three devices, effective 2017-01-18. 21 CFR 880.6250 is titled "Non-powdered patient examination glove".

For the 510(k) route and the full test list, see Exporting Medical Gloves to the US: FDA Route and Test Items.

D6319 requirement by requirement

Below, each requirement is explained by clause, with FDA's reporting expectations; values are in the specification. FDA's manual also says manufacturers selling gloves in the US should keep an original copy of each referenced standard at the manufacturing site, and FDA investigators may ask to see it and the conformance data during an inspection. Hold your own copy of D6319.

Material and supply forms

D6319 covers nitrile rubber examination gloves only. Natural rubber examination gloves fall under D3578, poly(vinyl chloride) gloves under D5250, polychloroprene examination gloves under D6977, and packaged sterile rubber surgical gloves under D3577. The D3578 abstract defines Type I and Type II gloves; D6319 has no types. Gloves labeled sterile must also pass the sterility test of clause 7.2, run to the U.S. Pharmacopeia (clause 6.1.1).

Dimensions and thickness

The FDA suggested 510(k) format asks for minimum overall length, minimum width (for a medium glove), minimum palm thickness and minimum finger thickness, all in mm. D6319 clause 7.4 defines where length and palm width are measured, takes thickness at the locations in its Fig. 1, and allows single-wall thickness to be taken as half the double-wall reading so the glove need not be cut. Values for each size are in the specification.

The FDA manual also says labeling may state the measured thickness but should not use ambiguous terms such as "extra thick", and that all labeling claims should be supported by data in the 510(k). If your label states a thickness, the data must support that number, not just the D6319 minimum.

Tensile strength and ultimate elongation before and after aging

D6319 clause 7.5 requires the gloves to meet the tensile strength and ultimate elongation in its physical requirements table both before and after accelerated aging, tested to D412 with Die C recommended, and aged to ASTM D573 (air oven). The FDA suggested format likewise asks for minimum tensile strength (MPa) and ultimate elongation (%), each before and after aging. Tensile testing is therefore done on two sets: one as received, one after aging.

D6319 gives two accelerated aging conditions and names one of them as the referee condition. Gloves older than a set age from manufacture, or with an unknown manufacturing date, are not aged and are judged directly against the after-aging values. Temperatures, durations and the age limit are in the specification. D573 itself notes that its results may not correlate exactly with service performance.

Tensile after aging is a physical property check within the specification, not a shelf-life study. FDA's manual says expiration dating of medical gloves is voluntary; if claimed, the gloves should meet the 510(k) specifications over the whole claimed shelf life, and the date should be determined by real-time stability studies (a provisional date based on accelerated aging is allowed if real-time testing starts immediately). The LZA page also lists ASTM D7160-16(2023), Standard Practice for Determination of Expiration Dating for Medical Gloves (recognition number 6-401). The SUNGO Shanghai CNAS scope lists it under the test object "medical gloves" (item: shelf life); it is not in the IAS scope. See our shelf-life validation service.

Freedom from holes and AQL

D6319 clause 7.3 requires freedom from holes to be tested by ASTM D5151; the inspection level and AQL are in its Table 1.

For its own sampling, FDA follows 21 CFR 800.20: plans are based on ISO 2859 sample sizes, inspection levels and AQLs; lots of 1,200 gloves or less use single normal sampling and larger lots multiple normal sampling; general inspection level II applies to any lot size; the minimum acceptable quality level is AQL 2.5 for patient examination gloves (AQL 1.5 for surgeons' gloves); and a lot is considered adulterated when the defectives in the sample meet or exceed the rejection number. The FDA suggested format also asks for your own pinhole AQL.

Powder

D6319 clause 4.3 requires both surfaces to be free of talc; clause 7.6 (residual powder on powder-free gloves) and clause 7.7 (recommended maximum powder on powdered gloves) both use ASTM D6124, and the limits are in the specification. For the US, FDA's manual recommends that a "powder-free" claim requires no more than 2 mg of residual or trace powder and debris per glove by D6124 or an equivalent method, and its suggested 510(k) format says trace powder should not exceed 2 mg per glove or the limit in the ASTM standard. The manual notes it has not been updated for the 2016 powdered glove ban.

Sampling and inspection level

D6319 clause 5.1 says that for referee purposes gloves are sampled from finished product, after sterilization when labeled sterile, and inspected to ISO 2859, with the inspection levels and AQLs of Table 1 or a more comprehensive plan agreed between purchaser and seller. Under product code LZA, FDA lists ANSI/ASQ Z1.4-2003 (R2018), recognition number 5-62, as the recognized sampling standard. Attribute sample sizes depend on lot size and inspection level, and tensile needs unaged and aged sets, so there is no fixed total quantity.

Method notes: tensile, thickness, holes, powder

Tensile: ASTM D412

D412-16(2021) covers the tensile properties of vulcanized thermoset rubbers and thermoplastic elastomers. Test Method A uses dumbbell and straight section specimens and Test Method B uses cut ring specimens; the standard notes the two do not produce identical results. D6319 recommends Die C, and the report should state the die and specimen used.

D412 stresses that tensile results depend on test conditions such as extension rate, temperature, humidity, specimen geometry and pretest conditioning, so materials should be compared only under the same conditions. The unaged and aged sets should therefore be cut with the same die and tested with the same conditioning and extension rate, or the change after aging cannot be interpreted. Tensile strength is calculated from specimen thickness, so thickness measurement directly affects the result.

Thickness and dimensions: ASTM D3767

D3767-03(2020) is a practice for measuring length, width, thickness, diameter and circumference of rubber products and test specimens. Three points matter for gloves:

  • Its uses include checking product size, with rubber gloves as an example, and measuring tensile specimen thickness.
  • It does not cover sampling or where samples are taken. The thickness locations on the glove are set by D6319 clause 7.4.4 and Fig. 1.
  • Instrument pressure strongly affects the observed thickness of soft materials, so the foot pressure must be specified for comparable results. If factory and laboratory thickness data disagree, check this first.

Holes: ASTM D5151

D5151-19(2023) detects holes in medical gloves that leak water under the test conditions; the smallest leaking hole size has not been determined and is outside its scope. Filling, what counts as a leak, and comparison with 21 CFR 800.20 and EN 455-1 are covered separately.

Powder: ASTM D6124

D6124-06(2022) determines the average powder or filter-retained mass on a sample of medical gloves, reported in milligrams per glove. Its FDA recognition number is 6-178, which also excludes text related to powdered gloves. Gravimetric control points and comparison with GB/T 21869 and EN ISO 21171 are covered separately.

Scope table

These rows are taken from the CNAS (L22083) and IAS (TL-1019) scopes of Shanghai SUNGO Medical Technology Co., Ltd. The first three CNAS rows are under the test object "medical nitrile gloves" and the fourth under "medical gloves", all effective 2026-03-09. The IAS scope lists whole standards without years. The Shanghai CMA scope (240015340087) contains no ASTM standard.

ItemClauseNotes / limitations
Water-tightness testASTM D6319-19(2023) 7.3CNAS; ASTM D5151-19(2023) also listed under the same item, no clause given; no limitation
Physical and dimensional propertiesASTM D6319-19(2023) 7.4, 7.5CNAS; ASTM D412-16(2021) and ASTM D3767-03(2020) also listed under the same item, no clause given; no limitation. D412 and D3767 are not in the IAS scope
Residual powderASTM D6319-19(2023) 7.6, 7.7CNAS; no limitation. No ASTM D6124 row under this test object
Shelf lifeASTM D7160-16(2023)CNAS; test object "medical gloves", no clause given; no limitation. Not in the IAS scope
ASTM D6319Whole standardIAS
ASTM D5151Whole standardIAS
ASTM D6124Whole standardIAS

Where SUNGO tests it and under which accreditation

D6319 testing is performed at the laboratory of Shanghai SUNGO Medical Technology Co., Ltd., No. 958 Kangqiao East Road, Pudong, Shanghai, which holds CNAS accreditation (L22083), CMA qualification (240015340087) and IAS accreditation (TL-1019). Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity whose scopes do not include medical glove items.

  • CNAS reports: against D6319-19(2023) clauses 7.3, 7.4/7.5 and 7.6/7.7, using D5151-19(2023), D412-16(2021) and D3767-03(2020). The CNAS scope lists only these three items; the clause 7.2 sterility test is not among them, so confirm it with the laboratory when requesting a quote.
  • IAS reports: D6319, D5151 and D6124 are accredited as whole standards with no year given; confirm the edition with the laboratory before an IAS report is issued. D412 and D3767 are CNAS only, so results reported under those method numbers alone can only be CNAS reports.
  • Residual powder: a CNAS report under D6319 clauses 7.6 and 7.7; an IAS report when issued under the D6124 method number alone, because the CNAS D6124 rows sit under other glove test objects, not medical nitrile gloves.
  • Physical properties before and after aging: part of D6319 clause 7.5, reported under the CNAS item "physical and dimensional properties" (clauses 7.4 and 7.5); no stand-alone D573 aging report is issued. Expiration dating under ASTM D7160-16(2023) can be reported under CNAS, not IAS.
  • CMA: no ASTM standard is in the Shanghai CMA scope, so D6319 results cannot carry the CMA mark. CMA covers only the items in its scope and, per the certificate, is valid within the People's Republic of China.
  • Biocompatibility: FDA's manual recommends that the 510(k) summary include data showing conformance with biocompatibility and other specifications. Biocompatibility tests are subcontracted by SUNGO to appropriately qualified laboratories; under accreditation rules, the client's agreement is obtained in advance, and the report identifies the subcontracted items and the subcontractor.

The CNAS certificate states that CNAS is a signatory to the ILAC and APAC mutual recognition arrangements, and the IAS certificate carries the ILAC-MRA mark. Whether an overseas regulator or customer accepts a given report is for them to decide.

Preparing a submission

  • State the standard and clauses. Give ASTM D6319-19(2023), the clauses you need (7.3, 7.4/7.5, 7.6/7.7) and the purpose: a 510(k) file, customer inspection or internal verification.
  • Specify the report mark. Say before ordering whether you need a CNAS or IAS mark, so each item can be checked against the right scope.
  • Describe the product. Material, powdered or powder-free, sterile or nonsterile, packaging (single, paired, boxed or bulk), size range, and whether colours or models differ in formulation.
  • One lot, original packaging. Give the lot number and manufacturing date; the date decides whether tensile samples are aged (clause 7.5.2). Send gloves labeled sterile as sterilized finished product (clause 5.1). Do not repack; opening and transferring gloves can introduce contamination or damage that affects powder and hole results.
  • Sizes and quantities. Which sizes and how many gloves per size follow D6319 and the sampling plan, and the unaged and aged tensile sets need separate samples. The laboratory will confirm quantities when quoting.
  • Provide the acceptance basis. If results are to be judged against your own specification (such as values in your 510(k) or a labeled thickness), send that document.

FAQ

How does D6319 differ from D3578?

D3578 covers natural rubber examination gloves and D6319 covers nitrile rubber examination gloves; D6319 clause 1.3 says they are similar. The D3578 abstract also lists protein content and antigenic protein content tests, which D6319 does not have.

Can tensile after aging be reported separately under D573?

No; D573 is not in SUNGO's certificates. Tensile before and after aging is part of D6319 clause 7.5 (clause 7.5.2 calls for aging to D573) and is reported under the CNAS item "physical and dimensional properties", with aging conditions as specified in D6319.

Do gloves sterilized inside a kit need to be re-verified?

FDA's manual says kit manufacturers and assemblers should make sure the gloves are cleared for marketing and can meet the appropriate FDA and ASTM standard (such as D6319) or an equivalent test method after the kit is sterilized. Agree with the laboratory which items to verify.

Will FDA accept a SUNGO D6319 report?

Acceptance is for FDA to decide in its review, and we make no commitment about any regulator's conclusion.

Request a quote

For D6319 testing of nitrile examination gloves, start with our medical mask and medical glove testing service, then use our contact page to send the product details (material, powdered or powder-free, sterile or nonsterile, size range), the clauses you need, the target market and the report mark required. We will reply item by item against the CNAS and IAS scopes. If the same gloves are also going to the EU, see Medical Gloves for the EU: What EN 455 Actually Requires.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.