At release, registration and export, every powder-free glove faces the same question: how much powder is left on each glove? Manufacturers and buyers usually ask how ASTM D6124, GB/T 21869 and EN ISO 21171 differ, what makes a glove powder-free or powdered, and which mark the report can carry (CNAS accreditation, CMA qualification or IAS accreditation). This guide answers from published standards and regulatory sources and the scopes of Shanghai SUNGO Medical Technology Co., Ltd.
Key points
- The three methods are really two. GB/T 21869-2008 is an identical adoption of ISO 21171:2006, and EN ISO 21171:2006 is the European adoption of the same ISO standard. ASTM D6124 is a separate ASTM method that the US FDA lists as a recognized consensus standard (recognition number 6-178, partial recognition).
- The pass/fail line is not in the test method. The ISO 21171 abstract says it prescribes no powder limits, which also holds for its identical adoption GB/T 21869; the public abstract of ASTM D6124 shows no limit either (see the full text). The criteria come from product standards and regulators: EN 455-3:2023 clause 4.4 allows a powder-free glove no more than 2 mg of powder per glove (method in 5.2) and treats anything above 2 mg as powdered; FDA's Medical Glove Guidance Manual recommends no more than 2 mg of residual powder and debris per glove, by ASTM D6124 or an equivalent method, for a "powder-free" claim.
- Not only donning powder is measured. EN 455-3:2023 counts any water-insoluble material washed off the glove surface, including processing aids and loosely attached material.
- Each method carries a different mark. At SUNGO Shanghai, the CNAS scope lists GB/T 21869-2008, ASTM D6124-06(2022) and EN 455-3:2023 (clause 4.4 only); the CMA scope lists GB/T 21869-2008 only; the IAS scope lists GB/T 21869, ASTM D6124 and EN ISO 21171 as whole standards and EN 455-3 (only 5.2), with no edition given for any IAS entry. No ASTM or EN standard is in the CMA scope.
- Glove testing is done in Shanghai only. The scopes of Hefei SUNGO Medical Technology Co., Ltd. do not include medical glove items.
The standards and their scope
Checked on 2026-10-22 against ASTM and ISO abstract pages, China's national standards platform, the FDA Recognized Consensus Standards database and the public preview of EN 455-3:2023. For test steps, sample numbers and calculations, see the full text.
ASTM D6124-06(2022) Standard Test Method for Residual Powder on Medical Gloves
- Under ASTM Subcommittee D11.40. The 2006 edition was reapproved in 2022 and is shown as Active.
- Scope: the average powder or filter-retained mass on a sample of medical gloves, reported in milligrams per glove.
- Significance and use: a reference method for residual powder and non-powder solids on medical gloves. For powdered gloves, the mass found by Procedure II is taken to combine water-insoluble manufacturing residue, former release agents and donning powder.
- FDA recognition 6-178 is partial: all text on powdered surgeon's gloves, powdered patient examination gloves and absorbable powder for lubricating a surgeon's glove is not recognized, because FDA has banned these devices.
GB/T 21869-2008 Medical gloves - Determination of removable surface powder
- A recommended national method standard, shown as current. Published 2008-05-14, implemented 2008-10-01, and confirmed as remaining valid at its review on 2016-12-31.
- Identical adoption of ISO 21171:2006.
ISO 21171:2006 and EN ISO 21171:2006
- Prepared by ISO/TC 45/SC 4 and published in May 2006 as edition 1; confirmed in 2024, and the ISO website states that this version remains current.
- The ISO abstract describes three methods for removable surface powder: method A for powdered gloves, and methods B and C for powder-free gloves. It does not address powder-related safety issues or prescribe limits, and its applicability to medical gloves not made from rubber has not been established.
- EN 455-3:2023 lists EN ISO 21171:2006 as a normative reference.
The powder clauses of EN 455-3:2023
- EN 455-3:2023, Medical gloves for single use - Part 3: Requirements and testing for biological evaluation, was prepared by CEN/TC 205, approved by CEN on 2023-10-29, and supersedes EN 455-3:2015. National adoption was due by May 2024 and withdrawal of conflicting national standards by November 2025.
- In the table of contents, 4.4 is "Powder-free gloves" (a requirement) and 5.2 is "Powder" (a test method).
- Clause 4.4: powder residue on a powder-free glove, determined by the method in 5.2, shall not exceed 2 mg per glove; a glove with more than 2 mg is a powdered glove. The 2015 preview has no notes after 4.4; the 2023 edition adds Notes 1 and 2. Note 1 strongly recommends powder-free gloves; Note 2 is covered under "Powdered gloves" below.
- In the 2015 preview, clause 5.2 calls for the method in Clauses 7 and 9 of EN ISO 21171:2006. The 2023 text of 5.2 is not in the public preview; refer to the full standard.
Chinese limit standards
In China, limits for residual surface powder on medical gloves are set by the mandatory national standard GB 24788-2009 (current, implemented 2010-10-01). GB 24788-2025, Safety technical requirements for medical gloves, was published on 2025-05-30 and will fully replace it on 2028-06-01. Refer to the full text for their scope and limits. Neither is in SUNGO's scopes; this guide covers only the measurement method, GB/T 21869.
Comparing the methods and choosing one
| Aspect | ASTM D6124-06(2022) | GB/T 21869-2008 | EN ISO 21171:2006 |
|---|---|---|---|
| Origin | ASTM Subcommittee D11.40 | Identical adoption of ISO 21171:2006 | European adoption of ISO 21171:2006 |
| Procedures | Abstract mentions Procedure II for powdered gloves; others: see the full text | Same as ISO 21171 | Method A for powdered gloves; methods B and C for powder-free gloves |
| Result | Average powder mass per glove, in mg | See the full text | See the full text |
| Limits | None in the public abstract (see the full text); criteria come from product specifications or regulators | None (identical to ISO 21171) | None, per the ISO abstract |
| Referenced by | FDA Medical Glove Guidance Manual for powder-free claims; FDA recognition 6-178 | See the relevant Chinese product standards | EN 455-3:2023 (normative reference) |
| SUNGO Shanghai accreditation / CMA qualification | CNAS: ASTM D6124-06(2022), four test objects only; IAS: ASTM D6124, whole standard, no edition given | CNAS and CMA: GB/T 21869-2008; IAS: GB/T 21869, whole standard, no edition given | Not in the CNAS or CMA scope; IAS: EN ISO 21171, whole standard, no edition given |
The choice mainly depends on the target market and what your customer specifies:
- United States: the FDA guidance manual uses ASTM D6124 or an equivalent method for powder-free claims, and the ASTM glove specifications have their own powder items (see below).
- European Union: the criterion is EN 455-3:2023 clause 4.4, the test method is in 5.2, and EN ISO 21171:2006 is a normative reference.
- China: the national method standard is GB/T 21869-2008 (identical to ISO 21171) and the limits are in GB 24788; which test method applies is set by the full text of the applicable product standard and your product technical requirements.
The two families were developed separately and are divided differently (D6124 into Procedures, ISO 21171 into methods A, B and C), so do not assume a result from one can replace the other unless your customer or regulator accepts it. The ISO 21171 abstract also says its applicability to gloves not made from rubber has not been established, so confirm the method with your customer or reviewer before testing PVC gloves to ISO 21171 or GB/T 21869.
Powder-free versus powdered: how gloves are judged
Powder-free gloves
EN 455-3:2023 clause 3.4 treats as powder any water-insoluble material washed off the glove surface under the test conditions. Its Note 1 adds that a glove with 2 mg of powder or less is powder-free and one with more than 2 mg is powdered, with the requirement in 4.4.
FDA's Medical Glove Guidance Manual, issued on 2008-01-22, recommends no more than 2 mg of residual or trace powder and debris per glove, measured by ASTM D6124 or an equivalent method, to support a "powder-free" claim. Its 510(k) format section adds that trace powder should not exceed 2 mg per glove or the limit in the ASTM standard. The cover page notes that the manual has not been updated to reflect the 2016 ban on powdered gloves.
Why do powder-free gloves still carry residue? The FDA manual explains that most manufacturers dip the former into a calcium carbonate and calcium nitrate solution as a release agent; a very small amount of calcium carbonate remains, most of it removed by leaching and washing. So a powder-free glove need not test at zero; what matters is whether it meets the criterion.
Powdered gloves
Note 2 to EN 455-3:2023 clause 4.4 says experts do not agree on how much powder a powdered medical glove may carry, or whether an upper limit is needed; sterile gloves must be washed free of donning powder before invasive procedures (see 4.6 f)). The standard also prohibits dressing gloves with talcum powder (magnesium silicate).
In the United States, an FDA final rule published on 2016-12-19 (81 FR 91722) banned powdered surgeon's gloves, powdered patient examination gloves and absorbable powder for lubricating a surgeon's glove, effective 2017-01-18. FDA's recognition of D6124 excludes the related text. For the US market, residual powder testing therefore serves to verify powder-free claims.
Powder items in the ASTM glove specifications
The abstracts of ASTM D3577 (rubber surgical gloves), D3578 (rubber examination gloves) and D6977 (polychloroprene examination gloves) list powder-free residue and powder amount among the required tests; those of D5250 (PVC gloves) and D6319 (nitrile examination gloves) list the amount of powder on the glove. All five abstracts call for gloves free of talc on the inside and outside surfaces. For the limits, refer to each specification.
SUNGO Shanghai's CNAS scope has a "residual powder" item under the 19(2023) edition of all five (clauses in the scope table below); the IAS scope lists D3578, D5250, D6319 and D6977 as whole standards (no edition given), but not D3577.
Filtration and weighing: key control points
The scope of D6124 states that it determines filter-retained mass; EN 455-3 defines powder as water-insoluble material washed off the glove surface; for the detailed steps of EN ISO 21171 (GB/T 21869), see the full text. Where a method washes the gloves, then filters and weighs the residue, these general laboratory controls apply. They do not restate any standard; filter type, washing, drying and calculations follow the chosen method.
- Filters at constant mass. The powder-free criterion is around 2 mg per glove, so filter moisture, static and weighing conditions go straight into the result. Dry, equilibrate and weigh filters to constant mass under the same conditions before and after filtration, on a balance suited to milligram-level weighing.
- Wash water and blanks. Insoluble particles from wash water or glassware are weighed as "powder" too. Laboratories commonly run a glove-free blank to detect this background; whether the method requires one, and how it is subtracted, follows the full text.
- Inside and outside surfaces. Powder can sit on the inner (donning) or outer surface. Which surfaces are washed, whether gloves are turned, and how many gloves form one specimen must follow the chosen method exactly, or results cannot be compared.
- Extraneous particles. Fibres and dust picked up during unpacking or transfer are weighed too, so send samples in original packaging and let the laboratory open them.
- Reading the result. D6124 reports the average mass per glove, and EN 455-3 clause 4.4 also judges per glove. Before comparing a result with a limit, check whether the report gives single-glove values or an average.
Accreditation scope table
All "residual powder" rows in the CNAS (L22083) scope of Shanghai SUNGO Medical Technology Co., Ltd., the matching CMA (240015340087) row, and the related IAS (TL-1019) entries. The CNAS rows took effect on 2026-03-09. The only powder row in the CMA scope is GB/T 21869-2008, with no clause given; none of the ASTM or EN rows is in the CMA scope. IAS entries give no edition, so confirm the edition with the laboratory before an IAS report is issued.
| Item | Clause | Notes / limitations |
|---|---|---|
| Residual powder (GB/T 21869-2008; test object: medical gloves) | No clause given in the scope | CNAS and CMA, with no limitation |
| Residual powder (ASTM D6124-06(2022)) | No clause given in the scope | CNAS, under four test objects: rubber surgical gloves, rubber examination gloves, PVC medical gloves and polychloroprene medical examination gloves, with no limitation |
| Residual powder (EN 455-3:2023; test object: medical gloves for single use) | 4.4 | CNAS, limited to clause 4.4 (residual powder) only |
| Residual powder (ASTM D3577-19(2023); test object: rubber surgical gloves) | 8.6, 8.8 | CNAS, no limitation; ASTM D3577 is not in the IAS scope |
| Residual powder (ASTM D3578-19(2023); test object: rubber examination gloves) | 8.6, 8.8 | CNAS, no limitation |
| Residual powder (ASTM D6319-19(2023); test object: medical nitrile gloves) | 7.6, 7.7 | CNAS, no limitation |
| Residual powder (ASTM D5250-19(2023); test object: PVC medical gloves) | 7.6, 7.7 | CNAS, no limitation |
| Residual powder (ASTM D6977-19(2023); test object: polychloroprene medical examination gloves) | 7.6, 7.7 | CNAS, no limitation |
| IAS: GB/T 21869, ASTM D6124, EN ISO 21171, ASTM D3578, ASTM D5250, ASTM D6319, ASTM D6977 (no edition given) | Whole standard | IAS, whole standards; the ASTM D6124 entry names no test object; EN ISO 21171 is not in the CNAS or CMA scope |
| IAS: EN 455-3 and BS EN 455-3 (no edition given) | 5.2 | IAS, partial, with the limitation "only 5.2" |
For nitrile gloves, residual powder is listed in the CNAS scope under ASTM D6319-19(2023) 7.6, 7.7, with no separate D6124 row for that test object. The other clauses of EN 455-3 are outside SUNGO's accreditation scopes.
Where the testing is done and under which accreditation
Testing is performed at the laboratory of Shanghai SUNGO Medical Technology Co., Ltd., No. 958 Kangqiao East Road, Pudong, Shanghai, which holds CNAS accreditation (L22083), CMA qualification (240015340087) and IAS accreditation (TL-1019). Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity whose scopes do not include medical glove items.
- GB/T 21869: GB/T 21869-2008 is in the CNAS and CMA scopes, so reports can carry either mark; the IAS entry GB/T 21869 is a whole standard with no edition given.
- ASTM D6124: CNAS covers ASTM D6124-06(2022) for the four test objects in the table; the IAS entry ASTM D6124 is a whole standard with no edition given; no CMA report.
- EN 455-3: CNAS covers EN 455-3:2023 clause 4.4 (residual powder) only; the IAS entries EN 455-3 and BS EN 455-3 are limited to "only 5.2" (the powder test method), with no edition given. Because they cover different clauses, agree with the laboratory when ordering which clause the report addresses and under which accreditation.
- EN ISO 21171: in the IAS scope only (no edition given), so IAS reports only.
General rules: CMA qualification covers only the items in its scope and, per the certificate, is valid within the People's Republic of China. Items in the CNAS or CMA scope but not in the IAS scope (for example ASTM D3577) cannot be reported under IAS. The CNAS certificate states that CNAS is a signatory to the ILAC and APAC mutual recognition arrangements, and the IAS certificate carries the ILAC-MRA mark. Whether an overseas regulator or customer accepts a given report is for them to decide.
Preparing a submission
- State the target market and governing document. Chinese registration, the EU, a US 510(k) and customer contracts may call for different methods and editions; give the standard number, edition and clause on the order form.
- Describe the glove. Examination or surgical, powdered or powder-free, material (natural rubber latex, nitrile, polychloroprene or PVC), and sterile or not. Method procedures and the test objects in SUNGO's scopes depend on this.
- Provide the acceptance criterion. ISO 21171 (GB/T 21869) sets no limits, and the public abstract of D6124 shows none. For a conformity statement, supply the limit clause from your product technical requirements or the applicable product standard.
- Keep samples in their original packaging. One lot, unopened, with the lot number and production date; do not repack.
- State the size. Results are per glove, and sizes differ in surface area. Choose the size according to the standard, or agree it with the laboratory.
- Sample quantity. This depends on the method and glove type; the laboratory will confirm it when quoting.
- Specify the report mark. Say before ordering whether you need a CNAS, CMA or IAS mark, so the item can be checked against that scope.
FAQ
Are GB/T 21869 and EN ISO 21171 the same method?
Yes. GB/T 21869-2008 is an identical adoption of ISO 21171:2006, and EN ISO 21171:2006 is its European adoption. At SUNGO, however, GB/T 21869-2008 is in the CNAS and CMA scopes and the IAS scope also lists GB/T 21869 (no edition given), while EN ISO 21171 is in the IAS scope only.
What is the powder-free criterion?
EN 455-3:2023 clause 4.4 sets no more than 2 mg per glove, and FDA's guidance manual recommends the same figure for a powder-free claim. In China, gloves are judged against GB 24788 and other applicable standards; refer to their full text for the limits.
Do powdered gloves still need a powder test?
Method A of ISO 21171 and Procedure II of D6124 are for powdered gloves. Whether you need the test depends on the market: the US has banned powdered surgeon's and examination gloves since 2017-01-18, and EN 455-3:2023 notes there is no agreed upper limit for powdered gloves.
Can a residual powder report carry the CMA mark?
Only GB/T 21869-2008 is in the Shanghai CMA scope. ASTM D6124, EN 455-3 and EN ISO 21171 are not, so their results cannot carry the CMA mark.
Can SUNGO perform the other parts of EN 455-3?
The Shanghai CNAS scope covers EN 455-3 clause 4.4 only and the IAS scope covers clause 5.2 only; the other clauses are outside SUNGO's accreditation scopes. Where biological tests such as cytotoxicity or sensitization are needed, SUNGO subcontracts them to appropriately qualified laboratories. Under accreditation rules, the client's agreement is obtained in advance, and the report identifies the subcontracted items and the subcontractor.
Request a quote
For residual powder testing, or to check which mark a method can carry, start with our medical mask and medical glove testing service, then use our contact page to send the glove type (material, powdered or powder-free, examination or surgical), target market, required standard and edition, and the report mark you need; we will reply item by item against the CNAS, CMA and IAS scopes. See also what EN 455 requires for the EU and the FDA route and test items for the US. Testing to the new Chinese examination glove standard GB/T 10213-2025 is covered in a separate article.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.