ASTM F1140 Burst and Creep Testing of Sterile Packages, with YY/T 0681.3

ASTM F1140 Burst and Creep Testing of Sterile Packages, with YY/T 0681.3

Paper-film and film pouches can see a pressure difference across the package during sterilization and shipping. ASTM F1140/F1140M evaluates how much internal pressure a package can take, and how long it holds under constant pressure, by inflating it from the inside. A Chinese industry standard on the same subject is YY/T 0681.3-2010; refer to the standard texts for how the two relate. This article covers what the burst, creep and creep-to-failure tests measure, how Test Methods A and B relate to open and closed package testing, why burst pressure varies between testers, and how burst data sit alongside F88 seal strength and support process monitoring. Clause numbers are taken from the public sample of ASTM F1140/F1140M-13(2020)e1; the -13(2025) edition is the 2025 reapproval of -13, so refer to the standard itself for clause-level detail.

Key points

  • Three modes, two method letters. The burst test is Test Method A; its result is the maximum pressure detected before the package fails. The creep test and the creep-to-failure test both sit under Test Method B. The creep test holds a specified pressure for a specified time and gives a pass/fail result; creep to failure holds pressure until the package fails and reports the time to failure (Sections 4.1 to 4.3).
  • Open and closed are testers, not methods. The open package tester is for flexible packages with one open side; the closed package tester pressurizes a fully sealed package through a puncture (Section 6.1). Open package testing has procedure clauses under both Method A (9.1.1) and Method B (9.2.1). Describing Method A as the open-package method and Method B as the closed-package method does not match the standard.
  • Results depend on test parameters. Section 5.3 lists package size, material, seal configuration, equipment, air flow rate, sensitivity and test article position; to correlate testers, all must be exactly the same.
  • Burst pressure cannot be converted into seal strength. Section 5.4 states that these methods do not necessarily correlate with seal strength measured by ASTM F88. The two complement each other.
  • The method sets no pass/fail limit. Acceptable burst pressure, creep pressure and hold time are set by the manufacturer in its packaging specification, based on its own validation data.
  • The accreditation mark depends on method and standard. SUNGO Shanghai's CNAS schedule lists ASTM F1140/F1140M-13(2025) and YY/T 0681.3-2010. IAS covers F1140 Test Method A only, so creep-type tests to F1140 carry the CNAS mark only. F1140 is not on the CMA schedule; for a CMA mark, test to YY/T 0681.3-2010. The restraining-plate method, ASTM F2054, is outside SUNGO's accredited scope.

Standards and scope

Sources: the public sample of ASTM F1140/F1140M, version details from a standards sales platform, the US FDA Recognized Consensus Standards database and China's National Public Service Platform for Standards Information, checked on 2026-10-22.

ASTM F1140/F1140M

  • Title: Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages. It is the responsibility of Subcommittee F02.20 on Physical Properties, under ASTM Committee F02 on Primary Barrier Packaging, and runs to 5 pages.
  • History: originally approved in 1988; revised in 2013 as F1140/F1140M-13; reapproved in 2020 as -13(2020)e1, where the epsilon marks an editorial correction to Table 2 made in July 2020. It was reapproved again in 2025 as -13(2025), the most recent version as checked on 2026-10-22. Under ASTM's designation system, the year in parentheses is the year of last reapproval.
  • Scope (Section 1): procedures for determining the ability of packages to withstand internal pressurization. The burst test increasingly pressurizes the package until it fails; the creep test maintains a specified pressure for a specified time or until the package fails. SI and inch-pound units are used independently, not combined (Section 1.4).
  • Significance and use (Section 5.1): a rapid means of evaluating the tendency of a package to fail when exposed to a pressure differential, which may occur during processes such as sterilization and transportation.
  • Terminology (Section 3.2): "package failure" means rupture of seal or material; "restraint" means a mechanism that prevents package movement during inflation. "Unrestrained" in the title means no such mechanism is used.
  • In the US FDA Recognized Consensus Standards database, the entry for this standard lists F1140/F1140M-13 (Reapproved 2020)e1, recognition number 14-402, with the complete standard recognized. Recognition of a standard does not mean any given test report will be accepted.

YY/T 0681.3-2010

  • Test methods for sterile medical device package, Part 3: Internal pressurization failure resistance of unrestrained packages. A recommended (non-mandatory) Chinese medical device industry standard (YY/T), issued 2010-12-27 and implemented 2012-06-01, listed as current. The responsible technical committee is the national technical committee for medical infusion devices.
  • It was issued before F1140/F1140M-13. For differences in test modes, apparatus or reporting between the two, refer to the YY/T 0681.3 text; this article does not compare them clause by clause.

Restraining-plate testing is outside this article

Inflating a flexible package between restraining plates is covered by ASTM F2054/F2054M; a Chinese industry standard on the same subject is YY/T 0681.9-2011 (Part 9: burst testing of flexible package seals using internal air pressurization within restraining plates). Neither is in SUNGO's accredited scope, and neither is covered here. One point from the FDA database abstract for F2054 matters when reading F1140 results: without restraint, stress is highest at the middle of the pouch, where it inflates to its maximum diameter, so F1140 may not reliably detect the weakest area of the seal.

Burst, creep and creep to failure: what each one measures

ModeMethod in the standardWhat is doneResult
BurstTest Method A (Burst Test), Sections 4.1 and 9.1Internal pressure is increased until the package fails; the equipment must sustain increasing pressure until burstMaximum pressure detected before failure; position and type of failure (material or seal) and the pressure are also noted (9.1.1.4 for the open package test; for the closed package test, see the standard)
CreepTest Method B (Creep Test), Section 4.2The package is pressurized to a specified pressure and held for a specified timePass or fail
Creep to failureTest Method B (Creep to Failure), Section 4.3Set up like the creep test, but at a higher pressure so that the package fails in a reasonable time ("that is, about 15 s", in the standard's words)Time until failure

On creep pressure, Section 9.2 (Method B) says a common starting point is 90 % of the burst value, but the pressure may depend on the materials or sealing mechanism and may be raised or lowered so testing occurs in an appropriate time range, with time to failure as the test statistic. So burst data for the same package usually come first. The 90 % figure is a starting point, not an acceptance limit.

Open package and closed package: two testers, not two methods

Section 6.1 states that packages are tested unrestrained, using one of two testers:

  • Open package tester: for flexible packages with one open side. Section 9.1.1.2 requires the tester to be adjusted so it does not stop the package reaching its normal unrestricted height at burst. Per Note 2, the package is typically placed in one of the positions in Fig. 1; positions may give different values, so the report should state the one used.
  • Closed package tester: pressurizes a completely sealed package through a puncture. Section 9.1.2.2 calls for careful insertion of the entry device, with the package center as the preferred entry point. Entry location may affect results, so it is recorded and kept the same for all samples. Note 3 warns against damaging other surfaces with the needle; where material tears easily, the entry point is to be reinforced with tape or equivalent.

Both testers include a means of measuring package pressure, a timer and a pressure regulator (Section 6.2). Method A is split into 9.1.1 Open Package Test and 9.1.2 Closed Package Test, and Method B has its own open package procedure at 9.2.1. So "burst or creep" and "open or closed" are separate choices.

The choice depends on what you want to evaluate. A pouch with one side still unsealed goes on the open package tester, which evaluates the existing seals and material. A package with its final seal made goes on the closed package tester, so the final seal is also under pressure. Sections 9.1.1.1 and 9.1.2.1 allow testing with or without product, provided the preparation is recorded.

Inflation rate and scatter in results

The same package can give noticeably different burst pressures on different testers. Most reasons trace back to clauses in the standard:

Source of scatterBasis in the standardHow to control it
Rate of pressurization9.1.1.3: the rate of pressurization may affect results and must not exceed the response rate of the pressure indicatorFix the rate in the work instruction; when changing testers, confirm the pressure indicator can keep up
Equipment and sensitivity5.3: test equipment, rate of air flow into the package and sensitivity (machine response to pressure drop) are listed as parametersCheck each parameter before comparing data between testers; keep data from different parameter sets apart
Package size and seal configuration5.3; F2054 abstract: unrestrained, stress is highest where the package inflates to its maximum diameterBuild data for each package size separately; do not compare sizes directly
Position and entry pointNote 2; 9.1.2.2Fix the position and entry point, and state them in the report
Porous materialsNote 1: if the pressure regulator cannot supply the necessary volume, the porosity of the sample may need to be limited (see Appendix X1)For paper, nonwoven and other porous packages, state the material when ordering and agree the approach in advance
Temperature and humidity8.1: condition and test per ASTM E171; 8.2: if not tested under standard conditions, record temperature and relative humidityUse the same conditioning for data that will be compared
Sample selection7.1: indiscriminate elimination of defective samples can bias resultsFix the sampling rules in advance; do not select samples after testing

Failure location is easy to overlook. In the open package test, Section 9.1.1.4 requires noting whether failure occurred in the material or the seal. If some samples fail in the material and others in the seal, scatter is expected because the values reflect two different weak points; analyzing by failure type is more useful than looking only at the mean.

How burst testing relates to F88 seal strength

F1140 and ASTM F88 both concern seals but measure different things. F88 measures the force required to separate a test strip of material containing the seal and identifies the failure mode, which reflects that length of seal. F1140 inflates the whole package, and failure occurs wherever it first gives way under its own shape and loading, in a seal or in the material.

Section 5.4 states that these test methods do not necessarily correlate with seal strength measured by F88 or an equivalent method. In practice the two are often used together: F88 gives the strength and failure mode of each seal, and F1140 gives a quick picture of how the whole package resists a pressure differential. The scopes of F1140 and F88 (Section 1 of each) do not cover leak detection; package integrity needs dedicated tests. Clamping techniques, specimen width and result reporting under ASTM F88 are covered in a separate article.

Using burst and creep tests for process monitoring

Section 5.2 notes that these methods are frequently used to evaluate packages quickly during manufacturing and across the package life cycle. To use the data for monitoring:

  1. Fix the method. Write the tester, open or closed configuration, position or entry point, rate of pressurization, with or without product, and conditioning into the work instruction.
  2. Build a baseline from the validated process. Measure burst values on packages made at validated sealing parameters, record failure location and type, and use these as the reference.
  3. Define when to sample. Common points are the first article at start-up, after a roll change, after a tooling change and after equipment maintenance.
  4. Watch the trend and the failure location. A general downward shift in burst values, or failures moving from the material to the seal or clustering on one seal edge, often signals a process change earlier than a single value crossing a lower limit.
  5. Use creep for quick decisions. A pass/fail hold at a specified pressure and time suits quick checks on the line. The manufacturer sets the pressure and time from its own burst data and validation results; F1140 gives no values.
  6. Use the same parameters before and after sterilization, aging and shipping. Otherwise you cannot tell whether a difference comes from the package or the method. See Shelf Life Validation for accelerated aging and Medical Device Simulated Transport Validation for transport testing.

Accreditation scope table

Relevant lines from the CNAS and CMA schedules and IAS scope of Shanghai SUNGO Medical Technology Co., Ltd. On the CNAS and CMA schedules the test object is "medical device packaging materials" and the limitation field is empty; the IAS scope lists no test objects.

ItemStandardClauseNotes / limitations
Internal pressurization failure resistanceASTM F1140/F1140M-13(2025)No clause specified on the scheduleCNAS L22083; schedule effective 2026-03-09; no limitation stated; not on the CMA schedule
Internal pressurization failure resistanceYY/T 0681.3-2010No clause specified on the scheduleCNAS L22083; schedule effective 2026-03-09; no limitation stated
Internal pressurization failure resistanceYY/T 0681.3-2010No clause specified on the scheduleCMA 240015340087; schedule effective 2024-11-25; no limitation stated
Partial (see clause)ASTM F1140/F1140MTest Method AIAS TL-1019; Test Method A only; no year of issue stated; confirm the edition with the laboratory before an IAS report is issued
Whole standardYY/T 0681.3Whole standard, no clauses listedIAS TL-1019; no year of issue stated; confirm the edition with the laboratory before an IAS report is issued

Where SUNGO does this test and under which accreditation

This test is done in Shanghai. The laboratory is at 958 Kangqiao East Road, Pudong New Area, Shanghai, operated by Shanghai SUNGO Medical Technology Co., Ltd., which issues the report.

  • CNAS laboratory accreditation, CNAS L22083. The schedule lists "internal pressurization failure resistance" for ASTM F1140/F1140M-13(2025) and YY/T 0681.3-2010, so reports can carry the CNAS mark. Creep and creep-to-failure testing to F1140 can only be reported under CNAS. The CNAS certificate states that CNAS is a signatory to the ILAC and APAC mutual recognition arrangements.
  • CMA (China Inspection Body and Laboratory Mandatory Approval), certificate 240015340087. The schedule lists YY/T 0681.3-2010 only. The CMA mark applies only to items on the schedule, and the certificate states that it is valid within the People's Republic of China. Testing to ASTM F1140 cannot carry the CMA mark.
  • IAS accreditation, TL-1019. ASTM F1140/F1140M is accredited for Test Method A, the burst test, only; YY/T 0681.3 is accredited as a whole standard. The IAS scope does not state years of issue, so confirm the edition with the laboratory before a report is issued.

Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity with its laboratory in Zipeng Town, Feixi County, Hefei; its scopes do not include this test. Whether an overseas customer or regulator accepts a given report depends on their own requirements, so please check before ordering.

Choosing the standard: for a CMA mark, order to YY/T 0681.3-2010; for US-facing files citing F1140, a CNAS report is available, and the burst test can also carry the IAS mark. If one batch of samples needs conclusions under both standards, list both when ordering.

Preparing samples and information

  1. Standard, edition and accreditation mark: ASTM F1140/F1140M-13(2025) or YY/T 0681.3-2010, and whether you need the CNAS, CMA or IAS mark, matched against the scope table above.
  2. Test mode: burst, creep or creep to failure. For creep, give the pressure and hold time; for creep to failure, give the pressure setting or how it should be determined.
  3. Tester and position: state whether you need the open or closed configuration, and confirm the available tester configuration with the laboratory before ordering. For open testing, mark the open side; for closed testing, state the entry point (the standard prefers the center) and whether reinforcement is needed.
  4. Package details: package dimensions, materials on each side (porous or not) and seal configuration; and whether samples contain product. Section 9.1.1.1 allows either, but it must be recorded.
  5. Pre-treatment and conditioning: whether samples were sterilized, aged or transport-tested, each condition labeled separately, plus any conditioning requirements.
  6. Comparison needs: if results will be compared with earlier data, provide the earlier tester and test parameters (Section 5.3).
  7. Number of samples: Section 7.1 asks for enough specimens to determine representative performance; base it on the product standard or validation protocol and confirm with us before ordering.

FAQ

Can a creep test to F1140 carry the IAS mark?

No. The Shanghai IAS scope covers F1140 Test Method A (burst) only; creep and creep to failure are Test Method B, so they can only be reported under CNAS. YY/T 0681.3 is IAS-accredited as a whole standard; confirm its test modes against the standard text before ordering.

Is Test Method A the open package method?

No. In the standard, Method A is the burst test and Method B covers creep and creep to failure. The open package tester and the closed package tester are two kinds of apparatus and do not correspond to the method letters.

Can burst pressure be converted into seal strength?

No. Section 5.4 states that F1140 results do not necessarily correlate with seal strength measured by F88. If you need both kinds of data, test for each separately.

Is restraining-plate burst testing within SUNGO's accredited scope?

No. The restraining-plate methods ASTM F2054/F2054M and YY/T 0681.9 are both outside SUNGO's accredited scope. SUNGO reports unrestrained internal pressurization data to F1140 or YY/T 0681.3.

Requesting a quote

For internal pressurization testing to ASTM F1140/F1140M or YY/T 0681.3-2010, start with the Sterile Medical Device Packaging Validation service page. For how packaging validation links to sterilization validation, see Sterile Barrier Systems: Packaging and Sterilization Validation. To request a quote, use the Contact page with the package construction, materials and dimensions, test mode, open or closed configuration, standard and edition, accreditation mark needed, and any pre-treatment such as sterilization, aging or transport testing. We will check each item against the CNAS, CMA and IAS scopes and reply point by point.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.