ASTM D5250 and D6977: Testing Vinyl (PVC) and Polychloroprene Exam Gloves

ASTM D5250 and D6977: Testing Vinyl (PVC) and Polychloroprene Exam Gloves

For vinyl (PVC) and polychloroprene examination gloves sold in the US, the ASTM product specifications are ASTM D5250 and ASTM D6977, which cover sterility, freedom from holes, dimensions and thickness, tensile strength and elongation before and after accelerated aging, and powder. FDA recognizes both only in part and excludes all text related to powdered gloves.

This guide goes through both specifications clause by clause against their 2019 texts, compares them with D6319 and D3578, explains the FDA product codes and 510(k) position, and lists the clauses within the CNAS and IAS scopes of Shanghai SUNGO Medical Technology Co., Ltd. External information was checked as of 13 November 2026.

Key Points

  • Two specifications, two materials. D5250 covers PVC gloves for medical examinations and diagnostic and therapeutic procedures, and PVC gloves for handling contaminated medical material, but not two-dimensional heat-sealed PVC gloves. D6977 covers polychloroprene rubber examination gloves. As of 13 November 2026, ASTM lists the active editions as D5250-19(2023) and D6977-19(2023): the 2019 editions reapproved in 2023.
  • Same clause numbers. Section 7 of both: 7.2 sterility (U.S. Pharmacopeia), 7.3 freedom from holes (D5151), 7.4 dimensions, 7.5 physical requirements before and after aging (D412 tensile, D573 aging), 7.6 powder-free residue and 7.7 powder amount (both D6124).
  • The main difference is aging and tensile. D5250 has one aging condition and one set of tensile minima for unaged and aged gloves. D6977, like D6319 and D3578, has two aging conditions, one designated for referee tests, and separate before- and after-aging minima.
  • FDA partial recognition. D5250 is recognition number 6-445 and D6977 is 6-442. Both exclude all text related to powdered surgeon's gloves, powdered patient examination gloves and absorbable powder for lubricating a surgeon's glove, because FDA banned these devices effective 18 January 2017.
  • Regulatory location. Both fall under 21 CFR 880.6250 (non-powdered patient examination glove), Class I. The D5250 record links product codes LYZ and LZA; the D6977 record links LZA only. Both codes show submission type 510(k), and 880.6250 exempts only finger cots from premarket notification.
  • What SUNGO Shanghai covers. CNAS (L22083) lists 7.3, 7.4/7.5 and 7.6/7.7 for each glove type. IAS (TL-1019) lists D5250, D6977, D5151 and D6124 as whole standards; D412 and D3767 have no IAS. No ASTM standard is in the CMA schedule.

What ASTM D5250 and D6977 Cover

ASTM D5250-19(2023): PVC gloves for medical application

  • Full title: Standard Specification for Poly(vinyl chloride) Gloves for Medical Application; Subcommittee D11.40, Volume 09.02, 4 pages. Originally approved in 1992; previous edition D5250-06(2015).
  • Scope: PVC gloves for medical examinations and diagnostic and therapeutic procedures, plus PVC gloves for handling contaminated medical material (1.1); gloves that fit either hand, paired gloves and gloves by size, packaged sterile, packaged nonsterile or bulk nonsterile (1.2); two-dimensional heat-sealed PVC gloves excluded (1.3); similar to D3578 (1.4).
  • Materials (Section 3): any PVC compound that meets the specification; a lubricant meeting U.S. Pharmacopeia requirements for absorbable dusting powder, or another lubricant of established safety and efficacy, may be applied; both surfaces shall be free of talc.
  • Purpose (Section 4): a referee procedure for evaluating the performance and safety of PVC examination gloves.

ASTM D6977-19(2023): polychloroprene examination gloves

  • Full title: Standard Specification for Polychloroprene Examination Gloves for Medical Application; Subcommittee D11.40, Volume 09.02, 4 pages. Originally approved in 2004; previous edition D6977-04(2016).
  • Scope: polychloroprene rubber gloves for medical examinations and diagnostic and therapeutic procedures (1.1); the same glove and supply forms (1.2); similar to D3578 and to nitrile specification D6319 (1.3).
  • Purpose (Section 3): a referee procedure, not intended for testing prior to routine lot release.
  • Material (Section 4): any polychloroprene compound that meets the requirements; lubricant and talc-free provisions as in D5250.

Documents referenced by both

Section 2 of both lists these documents without dates; the last column shows where the text cites each.

Referenced documentTitleUsed in
ASTM D412Vulcanized rubber and thermoplastic elastomers, tension7.5 tensile strength and ultimate elongation; 7.4.4 thickness gauge
ASTM D573Rubber, deterioration in an air oven7.5 accelerated aging
ASTM D3767Rubber, measurement of dimensions7.4.4 thickness measurement
ASTM D5151Detection of holes in medical gloves7.3 freedom from holes
ASTM D6124Residual powder on medical gloves7.6 and 7.7 powder
ASTM D3578Rubber examination glovesSimilar specification; 7.7.2 glove area calculation
ISO 2859Sampling procedures for inspection by attributesSampling
U.S. PharmacopeiaLatest editionLubricant; 7.2 sterility test

ASTM D5250 and D6977 Test Requirements, Clause by Clause

The performance requirements (D5250 Section 6, D6977 Section 5) map one-to-one onto the referee test methods in Section 7. Clauses and values below were checked against the D5250-19 and D6977-19 texts held by the laboratory; the active editions are their 2023 reapprovals, so work from the edition you have purchased. Dimension and sampling tables are not reproduced here.

Sterility: 7.2

Sterility testing follows the latest U.S. Pharmacopeia; D6977 5.1.1 applies it to products labeled sterile, which are sampled after sterilization.

Freedom from holes: 7.3

Both call for ASTM D5151, with the inspection level and AQL in Table 1. FDA's own sampling uses 21 CFR 800.20, covered below.

Dimensions and thickness: 7.4

  • Length runs from the tip of the middle finger (D5250 says "second finger") to the outside edge of the cuff; palm width is measured between the base of the index finger and the base of the thumb.
  • Thickness is measured at the locations in Fig. 1. To avoid cutting the glove, double-wall thickness may be measured and halved.
  • D6977 Table 2 adds cuff thickness to finger and palm; D5250 lists finger and palm only. A note to D6977 Table 2 lets gloves within the tolerance overlap of two sizes be labeled with a size range, such as small/medium.

Tensile before and after aging: 7.5

Both require the gloves to meet Table 3 before and after accelerated aging, with tensile testing per D412 (Die C recommended) and aging per D573. The aging set-up differs:

  • D5250: Under 7.5.2, specimens cut from the glove are exposed to 70 ± 2 °C for 72 ± 2 h, and the glove shall withstand this without tackiness, exudation or other deterioration. Table 3 gives a single set of values: tensile strength at least 11 MPa and ultimate elongation at least 300%, applied both before and after aging.
  • D6977: Under 7.5.2, either of two methods may be used: 70 ± 2 °C for 166 ± 2 h (the referee condition) or 100 ± 2 °C for 22 ± 0.3 h. Table 3 lists before-aging values (tensile strength at least 14 MPa, ultimate elongation at least 500%) and after-aging values (tensile strength at least 14 MPa, ultimate elongation at least 400%).
  • Common rule: gloves more than 6 months past their date of manufacture, or whose date of manufacture cannot be determined, are not given accelerated aging (D5250 7.5.3, D6977 7.5.2.3). D6977 adds that such gloves are judged against the after-aging values. State the manufacturing date when you submit samples.

Powder: 7.6 and 7.7

  • 7.6 (powder-free gloves): residual powder by D6124. D5250 6.1.5 sets a limit of 2.0 mg; D6977 Annex A1.1.1 recommends not more than 2 mg per glove.
  • 7.7 (powdered gloves): powder amount by D6124, with glove area calculated as in D3578. The recommended maximum in D5250 6.1.6 and D6977 Annex A1.2.1 is 10 mg/dm².
  • Examination gloves for the US must be powder-free (see the FDA section below), so 7.6 is the clause that matters in practice. For the filtration and weighing details of D6124, see Residual Powder on Medical Gloves: ASTM D6124, GB/T 21869 and EN ISO 21171.

Sampling, acceptance and package marking

  • Sampling (D5250 Section 5, D6977 Section 6): referee samples come from finished product and are inspected per ISO 2859 at the levels and AQLs in Table 1, or a more comprehensive plan agreed between purchaser and seller.
  • Acceptance (Section 8 in both): gloves meet the referee requirements when results conform to Table 1; retests are permitted under the U.S. Pharmacopeia and ISO 2859.
  • Package marking (Section 9): every packaging level carries the glove size and manufacturing lot number; sterile packages also carry opening instructions and the word "sterile".

D5250 and D6977 vs D6319 and D3578

The four ASTM glove specifications share a structure. Differences, based on the 2019 texts (recognition numbers from FDA's database):

ItemD5250D6977D6319D3578
MaterialPoly(vinyl chloride)PolychloropreneNitrile rubberNatural rubber
Scope notesAlso gloves for handling contaminated medical material; excludes two-dimensional heat-sealed glovesExamination glovesExamination glovesExamination gloves, Type I and Type II
Test clauses7.2 to 7.77.2 to 7.77.2 to 7.78.2 to 8.9 (extra classification section)
Accelerated agingOne conditionTwo conditions, one refereeTwo conditions, one refereeTwo conditions, one referee
Tensile minimaOne set for unaged and agedSeparate before/afterSeparate before/afterSeparate before/after
Protein requirementsNoneNoneNoneAqueous extractable and antigenic protein
FDA recognition number6-4456-4426-4466-440

D5250 and D6977 address PVC and polychloroprene and contain no protein tests. Because D3578 has an extra classification section, its CNAS clauses are 8.3, 8.4/8.5, 8.6/8.8 and 8.7; the other three use 7.3, 7.4/7.5 and 7.6/7.7. For the nitrile specification, see ASTM D6319 Nitrile Examination Gloves: Dimensions, Tensile and Leak Testing. Latex examination gloves under D3578 are covered in a separate article.

Why FDA Recognizes D5250 and D6977 Only in Part

As of 13 November 2026, FDA's Recognized Consensus Standards database lists ASTM D5250-19(2023) (6-445) and ASTM D6977-19(2023) (6-442), both as partial recognitions excluding the same content: all text related to powdered surgeon's gloves, powdered patient examination gloves and absorbable powder for lubricating a surgeon's glove. The stated reason is that these devices present an unreasonable and substantial risk of illness or injury that labeling cannot correct, so FDA banned them.

  • The ban: FDA published the final rule in the Federal Register on 19 December 2016 (81 FR 91722), effective 18 January 2017. Powdered patient examination gloves are listed at 21 CFR 895.103. 21 CFR 880.6250 is titled "Non-powdered patient examination glove", and it states that the finished glove includes only residual powder from manufacturing.
  • Mapping to the specification text: FDA's records list no clause numbers. In the specifications, the text on powdered gloves and dusting powder is mainly the lubricant clause (D5250 3.2, D6977 4.2, allowing absorbable dusting powder) and 7.7 on powdered-glove powder amount, with its performance requirement.
  • Effect on powder-free claims: FDA's Medical Glove Guidance Manual (issued 22 January 2008) recommends no more than 2 mg of residual powder and debris per glove for a "powder-free" claim, measured by D6124 or an equivalent method. It also recommends describing any powder used in manufacturing, how it is removed and the release specification, and certifying that the finished glove meets D5250 or another appropriate or equivalent standard.
  • Status of the manual: its cover page says it has not been updated for the ban, so its powdered vinyl glove label example no longer applies to the US market.

FDA Product Codes and 510(k) for Vinyl and Polychloroprene Exam Gloves

  • Regulation: 21 CFR 880.6250 classifies the device as Class I (general controls) and exempts only finger cots from the premarket notification (510(k)) procedures.
  • LYZ: in FDA's product classification database, LYZ is "vinyl patient examination glove", defined as a disposable glove made of poly(vinyl chloride) that may bear a trace amount of residual powder; Class 1, submission type 510(k), not GMP exempt.
  • LZA: "polymer patient examination glove", made of nitrile rubber or synthetic polymers; Class 1, submission type 510(k). The D6977 recognition record links LZA only, and the guidance manual's suggested 510(k) format also assigns synthetic polymer gloves to LZA.
  • Parameters in the 510(k): the manual's suggested format (not required by FDA) asks for minimum length, width (medium), palm and finger thickness, tensile strength and elongation before and after aging, and the pinhole AQL. Vinyl gloves are asked whether they meet D5250 or an equivalent standard; synthetic or polychloroprene gloves, whether they meet D3578, D6977 or another applicable standard.

21 CFR 800.20: How FDA Tests Exam Gloves for Leaks

Apart from the manufacturer's 7.3 hole test, FDA's own sampling follows 21 CFR 800.20: visual examination, then a 1,000 mL water leak test. Plans are based on ISO 2859: single normal sampling for lots of 1,200 gloves or fewer, multiple normal sampling for larger lots, general inspection level II. AQL 2.5 is the minimum acceptable quality level for patient examination gloves (1.5 for surgeons' gloves); a lot is adulterated when defectives in the sample meet or exceed the rejection number. For how D5151 fills the glove, what counts as a leak, and how it compares with 800.20 and EN 455-1, see ASTM D5151 Glove Hole Detection: EN 455-1 Water Leak Test and AQL Compared.

Accreditation Scope Table

The table reproduces the relevant lines of the CNAS (L22083) schedule and IAS (TL-1019) scope of Shanghai SUNGO Medical Technology Co., Ltd. All CNAS lines are effective from 2026-03-09 and carry no limitation. The IAS scope lists whole standards without edition years. The Shanghai CMA (240015340087) schedule contains no ASTM standard.

ItemStandard and clauseNotes / limitations
PVC gloves for medical use: water-tightness (hole) testASTM D5250-19(2023) 7.3; ASTM D5151-19(2023)CNAS; D5151 line lists no clause
PVC gloves for medical use: physical and dimensional propertiesASTM D5250-19(2023) 7.4, 7.5; ASTM D412-16(2021); ASTM D3767-03(2020)CNAS; D412 and D3767 lines list no clause; neither is in the IAS scope
PVC gloves for medical use: residual powderASTM D5250-19(2023) 7.6, 7.7; ASTM D6124-06(2022)CNAS; D6124 line lists no clause
Medical polychloroprene examination gloves: water-tightness (hole) testASTM D6977-19(2023) 7.3; ASTM D5151-19(2023)CNAS; D5151 line lists no clause
Medical polychloroprene examination gloves: physical and dimensional propertiesASTM D6977-19(2023) 7.4, 7.5; ASTM D412-16(2021); ASTM D3767-03(2020)CNAS; D412 and D3767 lines list no clause; neither is in the IAS scope
Medical polychloroprene examination gloves: residual powderASTM D6977-19(2023) 7.6, 7.7; ASTM D6124-06(2022)CNAS; D6124 line lists no clause
Whole standardASTM D5250IAS
Whole standardASTM D6977IAS
Whole standardASTM D5151IAS
Whole standardASTM D6124IAS

Where SUNGO Does the Testing and Under Which Accreditation

Testing is done at the laboratory of Shanghai SUNGO Medical Technology Co., Ltd., No. 958 Kangqiao East Road, Pudong New Area, Shanghai. The company holds CNAS accreditation (L22083), CMA certification (240015340087) and IAS accreditation (TL-1019). Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity, and its accreditation and certification scopes do not include medical gloves.

  • CNAS reports: against D5250-19(2023) or D6977-19(2023) clauses 7.3, 7.4/7.5 and 7.6/7.7, using D5151-19(2023), D412-16(2021), D3767-03(2020) and D6124-06(2022). The 7.2 sterility test is not in the schedule; if you need it, confirm with the laboratory when requesting a quote.
  • IAS reports: D5250, D6977, D5151 and D6124 are accredited as whole standards without edition years; confirm the edition with the laboratory before an IAS report is issued. D412 and D3767 are CNAS only, so a report under either method number alone can only be CNAS.
  • Aged and unaged properties: part of clause 7.5, reported under the CNAS item "physical and dimensional properties" (7.4, 7.5) with aging as set in the specification. D573 is not listed separately, so no standalone aging report is issued.
  • CMA: no ASTM standard is in the Shanghai CMA schedule, so these results cannot carry the CMA mark. CMA covers only scheduled items and is valid within the People's Republic of China.
  • Biocompatibility: FDA's manual recommends biocompatibility testing on finished gloves, noting that primary skin irritation and dermal sensitization studies are appropriate. SUNGO subcontracts biocompatibility tests to laboratories with the relevant qualifications, with the client's prior consent, and the report identifies the subcontracted items and the subcontractor.

CNAS and IAS are both signatories to the ILAC MRA. Whether an overseas regulator or customer accepts a given report depends on that party's own requirements.

Preparing Samples

  • State the standard and clauses. Name ASTM D5250-19(2023) or D6977-19(2023), the clauses (7.3, 7.4/7.5, 7.6/7.7) and the purpose: 510(k), customer inspection or internal verification.
  • Specify the report mark. Say whether you need CNAS or IAS before testing starts.
  • Declare the material. PVC or polychloroprene; for a polymer blend, FDA's manual recommends identifying each component in the identity statement, so tell the laboratory too. Note any formulation differences between colors or models.
  • Powdered or powder-free. US examination gloves should be powder-free; tell us whether any former-release or other powder is used in manufacturing.
  • Sizes. List the size range and whether you label size ranges such as small/medium.
  • Sterile or not. For gloves labeled sterile, send finished product after sterilization.
  • Lot number and manufacturing date. Send one lot in the original packaging without repacking. The manufacturing date decides whether accelerated aging is performed.
  • Quantity and criteria. Sample quantity depends on the sampling plan and tests and will be confirmed with your quote. To judge against your own specification, such as 510(k) values, provide it.

Frequently Asked Questions

Does ASTM D5250 apply only to examination gloves?

No. Clause 1.1 also covers PVC gloves used in handling contaminated medical material, and clause 1.3 excludes two-dimensional heat-sealed PVC gloves.

Are vinyl exam gloves exempt from 510(k)?

No. FDA's product classification database shows submission type 510(k) for LYZ, and 21 CFR 880.6250 exempts only finger cots from premarket notification.

Does passing the aged tensile test prove the glove's shelf life?

No. FDA's manual says expiration dating is voluntary but, if claimed, should rest on real-time stability studies; a provisional date from accelerated aging is accepted if real-time testing starts immediately. The Shanghai CNAS schedule lists ASTM D7160-16(2023) under "medical gloves" (item: shelf life); it is not in the IAS scope.

Can ASTM D5250 or D6977 test reports carry the CMA mark?

No. The Shanghai CMA schedule contains no ASTM standard. The options are CNAS or IAS reports, as set out in the scope table above.

Request a Quote

For ASTM D5250 or D6977 testing, see our Medical Mask / Glove / Dressing Testing service, then send the product details (material, powdered or powder-free, sterile or not, size range, manufacturing date), clauses, target market and report mark through our contact page. We will reply item by item against our CNAS and IAS scopes.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.