For vibration testing of medical device transport packages, the Chinese GB/T 4857 series offers three methods: GB/T 4857.7 sinusoidal vibration at constant frequency, GB/T 4857.10 sinusoidal vibration at variable frequency, and GB/T 4857.23 vertical random vibration. A request that only says "vibration" leaves excitation type, frequencies, spectrum, duration and orientation undefined, and the report may not match the registration file or customer specification. This guide explains what each method answers, resonance search and dwell, how the methods relate to ASTM D4169 Schedules D and E, what to inspect afterwards, and which SUNGO laboratory does the work under which accreditation.
Key points first
- To simulate sustained vibration during transport, start with random vibration (GB/T 4857.23-2021). ASTM D4728 states that random vibration excites many resonances at once, with less intense resonance build-up than sinusoidal dwell or sweep tests, so unrealistic fatigue damage is minimized.
- To find out which frequencies a package is sensitive to, run a sine sweep. The Chinese variable-frequency sine method is GB/T 4857.10-2005; whether it specifies a resonance search and dwell is as given in the standard text. ASTM D999 notes that resonant responses during shipment can be severe, and that identifying critical frequencies and the nature of package stresses can help minimize their effect.
- To assess how a package withstands vibration at a fixed low frequency, use constant-frequency sine vibration (GB/T 4857.7-2005). The abstract of the adopted ISO 2247:2000 describes fixed-frequency sinusoidal excitation for low-frequency vibration or repetitive-shock hazards; the frequency is as given in the standard text.
- There is no direct conversion between sine and random results. ASTM D4728 states that equivalent sine and random tests are generally difficult to establish, so one cannot replace the other.
- Report marks: all three methods are within the CNAS accreditation (L24279) of Hefei SUNGO Medical Technology Co., Ltd. Reports carry the CNAS mark only: no CMA mark and no IAS report. For sterile medical device packages that need a CMA-marked report, consider Schedules E and F of YY/T 0681.15-2019.
The standards and their scope
All three are part of GB/T 4857, “Packaging—Basic tests for transport packages”. They are recommended national standards under the packaging technical committee TC49. Data below are from China's National Public Service Platform for Standards Information.
| Standard | Official English title | Issued / effective | Relationship to ISO | Excitation |
|---|---|---|---|---|
| GB/T 4857.7-2005 | Part 7: Sinusoidal vibration test method at constant frequency | 2005-05-25 / 2005-11-01 | Modified adoption of ISO 2247:2000 | Sine wave, fixed frequency |
| GB/T 4857.10-2005 | Part 10: Sinusoidal vibration test method using at variable vibration frequency | 2005-05-25 / 2005-11-01 | Modified adoption of ISO 8318:2000 | Sine wave, frequency varied as specified |
| GB/T 4857.23-2021 | Part 23: Vertical random vibration test method | 2021-10-11 / 2022-05-01 | Modified adoption of ISO 13355:2016 | Random signal, vertical axis |
- GB/T 4857.7-2005 and GB/T 4857.10-2005 each replaced a 1992 edition; both were reviewed on 2023-12-28 as "remains valid".
- GB/T 4857.23-2021 fully replaced the 2012 edition. The 2012 edition, titled "Random vibration test method", was a modified adoption of ASTM D4728-2006 and was withdrawn on 2022-05-01. The 2021 edition is a modified adoption of ISO 13355:2016, and the word "vertical" was added to the title.
The test object is the complete, filled transport package (the adopted ISO standards also cover unit loads); full scope wording is as given in each standard, and SUNGO's certificate lists these three items for "transport packages". Test intensity, frequency range and duration must be set in a test plan or client specification.
What each method does
Constant-frequency sine vibration (GB/T 4857.7)
The table drives the package with a sine wave at one fixed frequency. Per the abstract of the adopted ISO 2247:2000, fixed-frequency sinusoidal excitation is used to assess strength or protection of the contents under low-frequency vibration, alone or within a test sequence, where distribution includes a low-frequency vibration hazard or repetitive shocks. In principle, with the frequency fixed, resonance amplification does not show if the package resonance lies far from it.
The plan usually states test frequency and amplitude (acceleration or displacement), duration, package orientation (axis as given in the standard), and any load. Frequency selection, methods and tolerances follow the standard text.
Variable-frequency sine vibration (GB/T 4857.10)
The frequency changes across a defined range: a sweep. Per the abstract of the adopted ISO 8318:2000, it specifies two variable-frequency sinusoidal test methods to assess strength or protection of the contents under vertical vibration, alone or within a test sequence. In principle, a frequency where the response is clearly amplified during a sweep is a resonance, so sweeps are also used to find sensitive frequencies; how the two methods are run, and whether they include a resonance search and dwell, follows the standard text.
The plan states frequency range, sweep mode and rate, amplitude, number of sweeps or total time, and package orientation (axis as given in the standard); if a dwell is required, also the dwell arrangement and how resonance is identified.
Vertical random vibration (GB/T 4857.23)
A random signal contains many frequency components at once, with randomly varying instantaneous amplitude. It is described by a power spectral density (PSD) curve showing energy over frequency; overall intensity is commonly given as overall RMS acceleration.
The plan states the PSD profile (breakpoints and levels), overall RMS acceleration, duration, orientation, any top load during vibration, and control tolerances. The profile source should be justified: a standard or specification, or measured transport data.
| Method | Question it answers | Main parameters in the test plan | Typical use |
|---|---|---|---|
| Constant-frequency sine, GB/T 4857.7 | Can the package withstand sine vibration at a fixed low frequency? | Frequency and amplitude, duration, package orientation | Assessing low-frequency vibration or repetitive-shock hazards |
| Variable-frequency sine, GB/T 4857.10 | Does it withstand vertical variable-frequency vibration? Which frequencies is it sensitive to? | Frequency range, sweep rate, amplitude, dwell arrangement (if any) | Sweep evaluation; package development diagnostics |
| Vertical random, GB/T 4857.23 | Is the package intact after a realistic stretch of transport vibration? | PSD profile, overall RMS acceleration, duration, orientation, top load | Conformity testing in transport validation |
Why random vibration is closer to real transport
The Significance and Use section of ASTM D4728-17(2026), “Standard Test Method for Random Vibration Testing of Shipping Containers”, makes three points:
- Approximating the actual damage, or lack of damage, experienced in real life may require random vibration tests, which excite many product and container resonances at once.
- Resonance build-up in random tests is less intense than in sinusoidal resonance dwell or sweep tests, so unrealistic fatigue damage due to resonance is minimized.
- There is no direct equivalence between random and sinusoidal vibration tests; because of nonlinearities, damping and product response characteristics, equivalent tests are difficult to establish in a general sense.
ASTM D999-08(2023), “Standard Test Methods for Vibration Testing of Shipping Containers”, describes its own methods this way: they do not simulate the shipping environment but are intended to create the damage-producing potential of the shipping environment. D999 includes both repetitive shock tests (Methods A1 and A2) and resonance tests (Methods B and C). One way to read this (our interpretation, not standard text): sine sweeps and dwells concentrate on resonance-related weak points, while random vibration is closer to transport vibration itself. Both standards recommend basing tests on representative field data.
D4728 allows random vibration in any axis, whereas GB/T 4857.23-2021 is limited to the vertical axis by its title; the abstract of ISO 8318:2000 (adopted by GB/T 4857.10) also refers to vertical vibration. If horizontal vibration is also a major hazard, the plan should state separately how it is addressed.
Resonance search and dwell
ASTM D999 notes that resonant responses during shipment can be severe and may lead to package or product failure. In principle, near the natural frequency of the package system (product, cushioning and container together), the response of the contents is amplified. D999 Method B (single container resonance test) and Method C (palletized load, unitized load or vertical stack resonance test) address this.
A resonance search and dwell is generally organized as follows. This is a general approach, not clause content of any one standard; the detailed procedure is as given in the applicable standard and the test plan:
- Define how response is monitored. Use accelerometers on the contents or critical locations, or agreed visual criteria, and define in advance what amplification counts as resonance.
- Sweep. Sweep across the agreed frequency range and amplitude, and record the frequencies where the response is clearly amplified.
- Dwell. Where required, vibrate at the selected resonance for the specified time. State in the plan whether the resonant frequency is tracked during the dwell.
- Record and use the results. Resonance data can guide cushioning design changes and help interpret damage found after random vibration.
ASTM D4728 groups resonance dwell and sweep as sinusoidal tests. Which standard a resonance search or dwell follows, and how, is as given in the standard text and agreed with the laboratory when ordering; anything that is not a procedure of the cited standard is outside SUNGO's accreditation scope.
How this maps to ASTM D4169 Schedules D and E
ASTM D4169 is a practice, not a single test method. Its scope describes a uniform laboratory basis for evaluating shipping units against the distribution environment through a sequence of anticipated hazard elements in a distribution cycle. In SUNGO Hefei's CNAS scope, "Schedule D stacked vibration and Schedule E vehicle vibration" cites clause 12 of ASTM D4169-23e1, and "Schedule F loose-load vibration" cites clause 13.
- Random vibration: ASTM D4728 states that recommended random vibration tests are given in Practice D4169. GB/T 4857.23-2021 adopts ISO 13355:2016, a different source from D4728 and D4169, so profiles, levels and durations cannot be transferred directly.
- Sine vibration: ASTM D999 states that its methods fulfill the requirements of ISO 8318 and ISO 2247, while cautioning that the ISO standards may not meet the requirements of D999.
- Test sequence: D4169 states that the tests should be performed sequentially on the same containers in the order given; the distribution cycle sets where vibration falls.
| Excitation | Chinese method | ISO basis | Related ASTM documents | SUNGO accreditation |
|---|---|---|---|---|
| Constant-frequency sine | GB/T 4857.7-2005 | ISO 2247:2000 | ASTM D999 (states it fulfills ISO 2247 and ISO 8318) | GB/T 4857.7: Hefei CNAS; D999 not in scope |
| Variable-frequency sine (sweep) | GB/T 4857.10-2005 | ISO 8318:2000 | ASTM D999 (Methods B and C are resonance tests) | GB/T 4857.10: Hefei CNAS (certificate item: "sinusoidal variable-frequency vibration"); D999 not in scope |
| Vertical random | GB/T 4857.23-2021 | ISO 13355:2016 | ASTM D4728 (refers to D4169 for recommended tests) | GB/T 4857.23: Hefei CNAS; D4728 not in scope |
| Vibration within a distribution sequence | — | — | ASTM D4169-23e1 Schedules D and E | Hefei CNAS (D4169-23e1, clause 12); Hefei IAS covers ASTM D4169 with no edition year, except Schedules G and H (confirm the edition before an IAS report); not in CMA scope |
Vibration type, profiles, levels and durations for Schedules D and E follow the D4169-23e1 text for the chosen distribution cycle and assurance level. When the FDA Recognized Consensus Standards database was checked on 2026-10-22, the recognized edition was ASTM D4169-22 (recognition number 14-576, complete standard), while SUNGO's certificate lists D4169-23e1. Which edition to cite depends on FDA's recognition list and review requirements at the time.
What to inspect after the test
ASTM D999 and D4728 both note that vibration affects the container, interior packing, closure and contents, and that these tests allow their interaction to be analyzed. Agree inspection items and failure criteria before testing. Common items:
- Outer package: damage, crushing or deformation of the container, loosened tape or straps, condition of labels and markings.
- Interior packaging and cushioning: broken or shifted cushions, signs of movement, impact or abrasion of the contents inside the box.
- Product: visible damage, loosened fasteners, detached parts; functional checks where agreed.
- Sterile barrier: for sterile devices, intact appearance does not prove an intact seal; post-transport integrity and strength tests are advisable, such as dye penetration (YY/T 0681.4-2021, ASTM F1929-23) and seal strength (YY/T 0681.2-2010, ASTM F88/F88M-23). All are in the Shanghai laboratory's scope (see the scope table).
Also agree in advance when packages are opened; the D4169 rule is summarized in the FAQ below.
Accreditation scope table
Each row is taken from SUNGO's accreditation scopes; item names are translated, clauses and limitations follow the certificate. "No clause listed" means the row gives no clause number. IAS entries carry no edition year; confirm the edition with the laboratory before an IAS report.
| Item (from certificate) | Basis and clause | Notes / limitations |
|---|---|---|
| Constant-frequency vibration | GB/T 4857.7-2005, no clause listed | Hefei CNAS; test object: transport packages; no CMA or IAS |
| Sinusoidal variable-frequency vibration | GB/T 4857.10-2005, no clause listed | Hefei CNAS; test object: transport packages; no CMA or IAS |
| Random vibration | GB/T 4857.23-2021, no clause listed | Hefei CNAS; test object: transport packages; no CMA or IAS |
| Schedule D stacked vibration and Schedule E vehicle vibration | ASTM D4169-23e1, clause 12 | Hefei CNAS; test object: shipping packages not exceeding 150 lb (68 kg); not in CMA scope |
| Schedule F loose-load vibration | ASTM D4169-23e1, clause 13 | Hefei CNAS; test object: shipping packages not exceeding 150 lb (68 kg); not in CMA scope |
| ASTM D4169 (partial) | No clause listed | Hefei IAS; limitation as written: "Except for Schedule G, H"; IAS scope lists no edition year |
| Schedule E: vehicle vibration | YY/T 0681.15-2019, clause 12 (CNAS row) | Hefei CNAS and CMA (CMA row: no clause); Hefei IAS covers YY/T 0681.15 (whole standard); test object: sterile medical device packages |
| Schedule F: unrestrained vibration | YY/T 0681.15-2019, clause 13 (CNAS row) | Hefei CNAS and CMA (CMA row: no clause); Hefei IAS covers YY/T 0681.15 (whole standard); test object: sterile medical device packages |
| Dye penetration | YY/T 0681.4-2021, no clause listed | Shanghai CNAS and CMA; Shanghai IAS covers YY/T 0681.4 (whole standard); test object: medical device packaging materials |
| Seal strength | YY/T 0681.2-2010, no clause listed | Shanghai CNAS and CMA; Shanghai IAS covers YY/T 0681.2 (whole standard); test object: medical device packaging materials |
| Dye penetration | ASTM F1929-23, no clause listed | Shanghai CNAS; Shanghai IAS covers ASTM F1929 (whole standard); not in CMA scope |
| Seal strength | ASTM F88/F88M-23, no clause listed | Shanghai CNAS; Shanghai IAS covers ASTM F88/F88M (whole standard); not in CMA scope |
Where the work is done and under which accreditation
- Vibration tests are performed in Hefei. All vibration items above are performed by Hefei SUNGO Medical Technology Co., Ltd., laboratory in Zipeng Town, Feixi County, Hefei, Anhui. Shanghai's certificates contain no transport simulation standards.
- GB/T 4857 reports carry CNAS only. Based on Hefei CNAS (L24279); neither Hefei CMA (260015342945) nor Hefei IAS (TL-1119) includes GB/T 4857.
- ASTM D4169-23e1: listed items are in Hefei's CNAS scope; Hefei's IAS scope covers ASTM D4169 except Schedules G and H, so D and E are included. The IAS scope lists no edition year, so confirm the edition with the laboratory before an IAS report is issued. D4169 is not in the CMA scope.
- YY/T 0681.15-2019: Schedules E and F are in Hefei's CNAS and CMA scopes; Hefei IAS covers YY/T 0681.15 as a whole standard with no edition year, so likewise confirm the edition before an IAS report. The CMA mark applies only to items in the CMA schedule and is valid only within the People's Republic of China.
- Post-transport sterile barrier tests are performed in Shanghai. Performed by Shanghai SUNGO Medical Technology Co., Ltd., laboratory at No. 958 Kangqiao East Road, Pudong, Shanghai (CNAS L22083, CMA 240015340087, IAS TL-1019).
- The two companies are separate legal entities, and each issues its own reports. If samples move from Hefei to Shanghai for integrity testing, say so when ordering so hand-over and custody records can be arranged. The transfer between the two cities is itself a transport exposure; agree the transfer method and protection in the test plan, record them, and consider them when evaluating results.
Preparing your samples and request
- State the purpose. Conformity testing for transport validation, or development work to find resonances or compare designs? The purpose changes the method and when packages are opened.
- State where the parameters come from. Name the standard, the customer specification, or the YY/T 0681.15 or D4169 distribution cycle and assurance level.
- Describe the distribution environment. Modes of transport (road, rail, air or combined), palletized or not, and stacking; include measured vibration data if you have it.
- Send complete transport packages. Container, inserts, cushioning and contents should match the shipping configuration, with the normal orientation marked.
- Agree on inspection items and failure criteria. Decide in advance what is checked for appearance, function and sterile barrier, and reserve samples for any integrity testing.
Frequently asked questions
We ran random vibration. Do we still need a sine sweep?
They answer different questions: random vibration shows whether the package survives realistic transport vibration, while a sweep finds sensitive frequencies. Whether both are needed depends on the test plan and the client's specification.
Is a report to GB/T 4857.23-2012 still usable?
The 2012 edition was withdrawn on 2022-05-01 and replaced by the 2021 edition, which has a different source standard. Acceptance of an old report is for the reviewer or customer to decide; SUNGO's certificate lists the 2021 edition.
Can a GB/T 4857 vibration report carry the CMA mark?
No. Hefei's CMA scope does not include GB/T 4857. For sterile medical device packages that need a CMA-marked report, consider Schedules E and F of YY/T 0681.15-2019, which are in Hefei's CMA schedule.
Should packages be opened right after vibration?
Under D4169, a performance test keeps the unit closed until the sequence is complete; for other purposes such as package development it may be opened and inspected mid-sequence, but this may prevent evaluating the influence of the closure on package performance. For a single GB/T 4857 test, timing is agreed in the plan.
Requesting a quote
For a quote, please send the product and package construction, transport route, test purpose, intended standard or customer specification, and required report marks. See Medical Device Simulated Transport Validation for vibration testing and Sterile Medical Device Packaging Validation for post-transport sterile barrier tests, and submit your request through the Contact Us page. How the GB/T 4857 parts combine into a transport test sequence is covered in a separate article.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.