Active medical devices need a stated service life, and real-time aging takes years, so many teams borrow the sterile packaging approach: pick an aging temperature from YY/T 0681.1 or ASTM F1980, apply Q10 and calculate the duration. For circuit boards and components, that acceleration factor usually lacks a defensible basis. This guide uses IEC 62506, Methods for product accelerated testing, and GB/T 34986-2017, the identical Chinese adoption of its 2013 edition, to explain how a Type B quantitative temperature-humidity test is designed: model choice, sample size and confidence, stop rules, acceleration factor evidence, and which SUNGO laboratory runs the test under which accreditation.
Key points
- Do not simply reuse the packaging Q10 calculation for an active device. ASTM F1980 addresses sterile barrier system integrity and packaging material properties, converts time on temperature using a conservative aging factor, and states that determining aging factors is outside its scope. Electronic parts often degrade under temperature and humidity together.
- IEC 62506 has three test types: Type A qualitative tests (HALT, HAST, HASS/HASA and similar), Type B quantitative accelerated tests, and Type C quantitative time and event compressed tests. This guide covers only Type B tests with temperature, humidity or both.
- A Type B test rests on three things: the failure mechanism, an acceleration model (Arrhenius for temperature; Peck's relationship is common in the literature for temperature and humidity, and the form used in IEC 62506 should be taken from the standard itself), and a statistical test plan for sample size, duration and confidence.
- The acceleration factor needs evidence: data at several stress levels, an activation energy estimate, failure analysis showing the mechanism is unchanged, and comparison with real-time data.
- SUNGO's scope: IEC 62506 and GB/T 34986 are within the IAS accreditation (TL-1019) of Shanghai SUNGO Medical Technology Co., Ltd., limited to Type B quantitative accelerated tests for temperature, humidity and their combinations. Only IAS reports are issued for them. HALT, HASS/HASA and other Type A qualitative tests, and acceleration by vibration or electrical stress, are outside this accreditation; confirm with the laboratory whether thermal cycling counts as temperature stress under the certificate.
The standards and their scope
| Standard | Title | Edition and dates | Notes |
|---|---|---|---|
| IEC 62506:2013 | Methods for product accelerated testing | Edition 1.0, published 2013-06-18, withdrawn 2023-11-07 | Prepared by IEC TC 56, Dependability |
| IEC 62506:2023 | Methods for product accelerated testing | Edition 2.0, published 2023-11-07 | Replaces edition 1.0 as a technical revision |
| GB/T 34986-2017 | Methods for product accelerated testing | Issued 2017-11-01, effective 2018-05-01; reviewed 2026-03-18 and confirmed | Recommended national standard, identical adoption of IEC 62506:2013, under SAC/TC 24 |
Clause 1 of IEC 62506 says it gives guidance on applying accelerated test techniques to measure or improve product reliability, so that design weaknesses can be found, dependability information obtained or reliability improved within a compressed time. It covers non-repairable and repairable systems, probability ratio sequential tests, fixed duration tests and reliability growth tests, and also production screening. The 2023 edition adds that services and people are not covered. Three points come first:
- It is a methods guide, not a medical device product standard. It sets no life target or pass criterion. The Chinese national standards portal classifies GB/T 34986-2017 as a methods standard.
- It is used with statistical test plan standards. The introduction names IEC 61123 (success ratio), IEC 61124 (constant failure rate and constant failure intensity), IEC 61649 (Weibull analysis) and IEC 61710 (power law model).
- The Chinese standard follows edition 1.0. The 2023 foreword lists updated references, revised symbols, corrected errors (mainly in 5.7.2.3 and Annex B) and corrected calculation errors in the examples of Annexes B and F; annex lettering also changed. For differences noted in the GB/T 34986 foreword, refer to the standard itself.
For registration in China, service-life studies for active devices can refer to the NMPA Guideline for Registration Technical Review of the Service Life of Active Medical Devices, using the current NMPA version. IEC 62506 supplies methods for test design and data analysis; how results are used in a submission is for the applicant to justify and the reviewer to judge. Some review reports published by the NMPA Center for Medical Device Evaluation show applicants using combinations of key component life analysis, accelerated aging of the complete device and real-time shelf aging; some tested appearance, performance and electrical safety after aging.
Why the YY/T 0681.1 packaging approach does not transfer
YY/T 0681.1-2018, "Test methods for sterile medical device package, Part 1: Test guide for accelerated aging", was written for sterile barrier systems, like ASTM F1980. Four differences matter for an active device:
| Aspect | Packaging aging (ASTM F1980-21 as an example) | Type B test of an active device (IEC 62506) |
|---|---|---|
| What is evaluated | Sterile barrier system integrity and packaging material properties (clause 1.1) | Degradation of device function, performance and safety-related characteristics |
| Accelerating variable | Temperature, with time converted using a conservative aging factor (Q10) | Chosen from the failure mechanism; temperature, humidity and electrical bias may act together |
| Source of the acceleration factor | F1980 states that determining aging factors is outside its scope | Needs a physical basis (5.2.2) and a design built on the failure mechanism (5.3); parameters can be estimated by testing |
| Form of the conclusion | Requirements still met after the equivalent time; expiry is tentative until real-time aging is complete | Reliability over a stated time with its confidence, or a life measure such as B10 |
The mechanisms differ. Escobar and Meeker's review of accelerated test models notes that relative humidity combined with temperature can accelerate corrosion and other chemical reactions, with applications including electronic devices, semiconductor parts and circuit boards. Higher temperature or humidity can speed up degradation of an adhesive bond, or growth of a conducting filament across an insulator that eventually causes a short circuit. A temperature-only Q10 misses these humidity effects, and since components differ in dominant mechanism, one factor for the whole device needs justification.
The operating state differs. Packaging aging represents storage; the service life of an active device covers storage and powered use. The IEC 62506 symbol list includes THB, temperature humidity bias test, which usually means applying electrical bias under temperature and humidity stress. Whether and how the device runs during the test belongs in the plan.
F1980-21 does allow its guidance to be used for medical devices; the question is whether the borrowed mechanism and parameters can be justified.
The three test types in IEC 62506
Clause 4.1 explains accelerated testing with a cumulative damage model: higher stress produces, in less time, the damage expected over the product's life. A test that only checks the margin between destruct and specification limits, without a probability of survival, is qualitative; one that relates stress to the probability of surviving beyond the predetermined life is quantitative.
| Type | Clauses | Purpose | SUNGO accreditation |
|---|---|---|---|
| Type A: qualitative accelerated tests | 4.2.2, 5.1 (HALT, HAST, HASS/HASA) | Find the most likely failure modes and design limits | Outside the accredited scope |
| Type B: quantitative accelerated tests | 4.2.3, 5.2, 5.5, 5.6 | Relate stress to life on the physical basis of the failure mechanism and estimate reliability | Temperature, humidity and their combinations only (Shanghai, IAS) |
| Type C: quantitative time and event compressed tests | 4.2.4, 5.2.3 | Remove exposure time at low or non-damaging stress (C1), or raise the stress repetition rate well above field levels (C2) | Outside the accredited scope |
The Purpose column is our summary; see clause 4.2 of the standard for the definitions. A note to the HALT definition says HALT is a non-measurable test that gives the stress representing the design limit, not information on life duration. HALT results alone cannot support a service-life claim.
Designing a Type B temperature-humidity test
Step 1: use profile and failure mechanisms
Clauses 5.3, 5.4 and 5.7.1 cover failure mechanisms and test design, stress levels and profiles in use and test, and reliability requirements and the use profile. For an active device, first answer:
- Where is the product used and stored, and what temperature and relative humidity does it see?
- How much of the time is it powered, how often is it switched, and how much does it self-heat?
- What service life is targeted, and what reliability must be shown?
- Which mechanisms in the FMEA and key component list are driven by temperature or humidity?
Fatigue from repeated stresses such as thermal cycling or vibration is a different mechanism; clause 5.6.3 discusses fatigue models (see the standard for details). Vibration acceleration is outside SUNGO's accreditation for IEC 62506 / GB/T 34986; confirm with the laboratory whether thermal cycling counts as temperature stress under the certificate.
Step 2: choose the acceleration model
- Temperature: Arrhenius. Clause 5.6.1 covers single-stress acceleration; Figure 5 is a line plot for the Arrhenius reaction model and Figure 6 a plot for determining activation energy. The standard defines activation energy, Ea, as an empirical factor for estimating the acceleration caused by a change in absolute temperature; an informative annex covers its estimation.
- Temperature plus humidity: Clause 5.5 covers multiple-stress acceleration for Type B tests. A model common in the literature is Peck's relationship, where humidity enters as a power-law term alongside the Arrhenius term, with no interaction term; Peck used it in 1986 to study failures of epoxy packaging. For the model form and parameters in IEC 62506, refer to the standard.
- State parameter sources. Activation energy and humidity parameters may be estimated from tests at several stress levels or taken from component data or literature, with the source and applicability stated.
- Keep the mechanism unchanged. A note to the step stress definition says the test should not alter the basic failure modes, failure mechanisms or their relative prevalence; 5.10.2 discusses the level of accelerated stresses. Excessive temperature can cause material changes never seen in use, and humidity conditions should avoid condensation that would not occur in use.
- The accreditation covers temperature and humidity stress only. SUNGO's accreditation for these standards is limited to temperature, humidity and their combinations; voltage, current or other electrical stress used as the accelerating stress (for example raised bias) is outside the accredited scope. Powering samples in their normal operating mode is a test condition; agree it with the laboratory before ordering.
Step 3: sample size and confidence
Sample size and duration come from the statistical test plan. In the 2023 edition, 5.7.2 includes Weibull-based test planning, a chart of reliability against the life time ratio Lv and the number of test items, and a planning nomogram; the symbol list includes confidence C and B10, the time by which 10 % of items have failed. Clause 6.1 discusses the sampling plan. Fix in advance:
- The target: reliability and confidence over the stated use time, or a life measure such as B10.
- The distribution: constant failure rate or Weibull, with the shape parameter and its basis.
- The plan type: zero-failure fixed duration, fixed duration allowing some failures, test to failure, or sequential probability ratio test.
- The trade-off: sample count, equivalent time per unit and allowed failures constrain one another. In a zero-failure plan, fewer samples need more equivalent time each; when the achievable acceleration caps the duration, more samples make up the difference.
Calculations follow the chosen statistical standard and the text of IEC 62506; this guide gives no generic figures.
Step 4: failure criteria and stop rules
Clause 4.1 states that when degradation reduces performance below the requirement specifications, the product can be declared failed if that behaviour is defined as a failure. So define failure before testing: the function, performance and safety-related parameters monitored, with limits from the product technical requirements or design specification; intermediate and final inspections, methods and equipment; and how intermittent faults are recorded and counted.
A test usually ends when the equivalent time for the target life is reached (fixed duration), when a specified number of failures occurs or all units fail (test to failure), or when a sequential boundary gives an early accept or reject decision. If a failure mode appears that could not occur in use, stop and check whether stress is too high. Keep failed units for failure analysis.
Step 5: supporting the acceleration factor
The 2013 edition defines the acceleration factor as the ratio of failure distribution characteristics or reliability measures under expected use stresses to those under the higher test stresses, greater than 1 for an effective test. The 2023 edition uses "test acceleration factor": the ratio of the stress response rate under accelerated conditions to that under specified operational conditions, such as operating time to failure, failure intensity or rate of wear. Evidence can come from:
- Several stress levels: plot failure data from different temperatures or temperature-humidity combinations, fit the model and derive activation energy (Annex D in 2013, Annex C in 2023).
- Testing to failure: an informative annex (Annex G in 2013, Annex F in 2023) shows acceleration factors determined by comparing Weibull plots of data sets.
- Mechanism consistency: failure analysis confirms that modes and mechanisms at each stress level match those in use.
- Calibrated accelerated life testing (CALT): described in an informative annex (Annex E in 2013, Annex D in 2023).
- Real-time comparison: recheck the result when real-time or field data arrive. See Accelerated and Real-Time Aging in Parallel: Organizing the File.
Clause 7 discusses the limitations of accelerated testing, which helps bound a conclusion.
Accredited scope
The table reproduces the relevant lines of SUNGO's scope documents. IAS scopes do not show edition years.
| Item | Clause | Notes / limitation |
|---|---|---|
| IEC 62506 (Methods for product accelerated testing) | No clause listed; partial scope | Shanghai, IAS TL-1019. Limitation as written: Only accredited for B type quantitative accelerated tests for temperature, humidity and their combinations |
| GB/T 34986 (Methods for product accelerated testing) | No clause listed; partial scope | Shanghai, IAS TL-1019. Limitation as written: Only accredited for B type quantitative accelerated tests for temperature, humidity and their combinations |
| IEC 60601-1 (Medical electrical equipment - Part 1: General requirements for basic safety and essential performance) | Whole standard | Hefei, IAS TL-1119. For electrical safety checks after aging, where needed |
| GB 9706.1 (Medical electrical equipment - Part 1: General requirements for basic safety and essential performance) | Whole standard | Hefei, IAS TL-1119. For electrical safety checks after aging, where needed |
Where the test is done and under which accreditation
- IEC 62506 and GB/T 34986 are run in Shanghai by Shanghai SUNGO Medical Technology Co., Ltd., laboratory at No. 958 Kangqiao East Road, Pudong, Shanghai, under IAS accreditation TL-1019; the scope gives 2026-02-19 as the effective date for these lines. The Hefei laboratory does not hold them.
- IAS reports only. The two standards are not in the Shanghai CNAS (L22083) or CMA (240015340087) scope, so no CNAS report or CMA mark is available. The CMA mark applies only to items in the CMA schedule and is valid only within China.
- Agree the edition. State whether IEC 62506:2013, IEC 62506:2023 or GB/T 34986-2017 applies, and confirm the edition with the laboratory before an IAS report is issued.
- Electrical safety after aging is done in Hefei. IEC 60601-1 and GB 9706.1 are within the scope of the Hefei laboratory under IAS accreditation (TL-1119). Hefei SUNGO Medical Technology Co., Ltd. (Zipeng Town, Feixi County, Hefei) and the Shanghai company are separate legal entities issuing their own reports, so plan sample transfers in advance.
- Report acceptance. The IAS certificate carries the ilac-MRA mark, but acceptance is for each regulator or customer to decide. See How to Read a Test Report: CNAS, CMA and IAS Compared and SUNGO's full testing scope.
Accelerated aging of sterile packaging and accelerated life testing of the device should be designed separately. A separate article compares ASTM F1980 and YY/T 0681.1.
Preparing for the test
- Purpose and target: the service life or storage period to support, and the reliability, confidence or life measure to show.
- Use profile and mechanisms: temperature and humidity in use and storage, powered time, the FMEA and the key component list.
- Model basis: the intended model and parameter sources; without a reliable source, consider testing at several stress levels.
- Samples: hardware and software matching the marketed product, each unit identified, with the power supplies, loads, software and fixtures needed to run them.
- Inspections and criteria: intermediate and final checks, methods, limits and the equipment needed, written into the plan. Function and performance checks not covered by a certificate line above are outside the accredited scope.
- Powering, edition and use: whether and how the device runs during the test, the edition of the standard, and whether the report is for China or elsewhere.
Frequently asked questions
Can the ASTM F1980 Q10 approach set the service life of an active device?
F1980 allows its guidance to be used for medical devices, but it converts time on temperature alone and does not cover how aging factors are determined. Where humidity and powered operation contribute to failure, a Type B test with a mechanism-based model and test-verified parameters is easier to defend.
Should we cite GB/T 34986 or IEC 62506?
GB/T 34986-2017 is identical to IEC 62506:2013, and IEC published edition 2.0 on 2023-11-07. State the edition in the order and confirm it with the laboratory before an IAS report is issued.
Can HALT, thermal cycling or vibration acceleration be done at SUNGO under IEC 62506?
HALT and other Type A tests, and vibration acceleration, are not within SUNGO's accreditation for these two standards, which is limited to Type B quantitative accelerated tests for temperature, humidity and their combinations. Confirm with the laboratory whether thermal cycling counts as temperature stress under the certificate.
After an accelerated test, is real-time aging still needed?
The accelerated result depends on model and parameter assumptions. Published NMPA review reports include an applicant who combined accelerated aging with real-time shelf aging; whether and how much real-time data is needed is for the reviewer to decide.
Requesting a quote
Please describe the product and key components, the use profile, the target service life and reliability, the intended model, the inspections and the edition of the standard. For the service, see Shelf Life Validation; for electrical safety after aging, see Medical Electrical Safety. To send a request, use the contact page.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.