UK Designated Standards v2: EN 455, EN 14683 and EN 13795 Editions vs EU

UK Designated Standards v2: EN 455, EN 14683 and EN 13795 Editions vs EU

Which EN edition should medical gloves, face masks and surgical gowns be tested to for Great Britain? As checked on 13 November 2026, the UK designated standards list for medical devices, version 2 (17 January 2026), uses the EU editions of EN 455-1, -2 and -3 but still designates EN 14683:2019+AC:2019 for masks and EN 13795-1/-2:2019 for gowns, drapes and clean air suits, while the EU MDR has harmonised the 2025 editions. Below: a line-by-line UK/EU comparison, the CE marking deadlines for Great Britain, and which editions SUNGO's Shanghai laboratory can report against under CNAS and IAS.

Key points

  • Gloves are aligned. From 17 January 2026, list v2 designates EN 455-1:2020+A2:2024, EN 455-2:2024 and EN 455-3:2023, the EU MDR harmonised editions; EN 455-1:2000, EN 455-2:2009+A2:2013 and EN 455-3:2006 were removed the same day. EN 455-4:2009, designated since 1 January 2021, is still listed.
  • Masks and gowns are out of step. The UK still designates EN 14683:2019+AC:2019, EN 13795-1:2019 and EN 13795-2:2019 (since 1 January 2021); the EU published EN 14683:2025, EN 13795-1:2025 and EN 13795-2:2025 as MDR harmonised standards on 20 October 2025.
  • Packaging and microbiology are aligned. From 17 January 2026, v2 designates EN ISO 11607-1:2020 (with A1:2023), EN ISO 11737-1:2018 (with A1:2021) and EN ISO 11737-2:2020, the EU MDR harmonised editions.
  • CE marking deadlines. Under the gov.uk guidance (updated 20 February 2026), EU MDR-compliant general medical devices can be placed on the Great Britain market until 30 June 2030, and MDD-compliant devices until the sooner of certificate expiry or 30 June 2028. On 21 October 2026 the government said in a consultation response that it will legislate for indefinite recognition of EU MDR and IVDR CE-marked devices, subject to parliamentary approval.
  • How SUNGO reports. The Shanghai CNAS schedule lists only the 2025 editions of EN 14683 and EN 13795, so there is no 2019-edition CNAS report; a 2019 edition can only be considered under IAS (TL-1019, no edition years), confirmed at quotation. The BS EN 14683:2025 CNAS lines exclude BFE and cleanliness, which go under the EN 14683:2025 CNAS lines or, once the edition is confirmed, IAS.

What UK designated standards are and where they apply

Great Britain (England, Scotland and Wales) regulates medical devices under the Medical Devices Regulations 2002 (S.I. 2002/618, UK MDR 2002); the regulator is the Medicines and Healthcare products Regulatory Agency (MHRA). Regulation 3A defines a designated standard as a technical specification designated by the Secretary of State by publishing a reference to it, and compliance with it is not compulsory. MHRA's UKCA guidance says that complying with designated standards means conforming with the requirements they cover, and that using standards is not mandatory, though most manufacturers do.

The list is published as Department of Health and Social Care (DHSC) notices. The gov.uk page "Designated standards: medical devices" applies to England, Scotland and Wales and holds notices 0034/21 (1 January 2021) and 0129/26 (17 January 2026) plus consolidated list version 2 (17 January 2026). Notice 0129/26 states that the references are published for the purposes of regulation 3A of S.I. 2002/618 and designated in relation to England and Wales and Scotland. The consolidated list states it has no legal effect of its own; the notices are the formal source.

Northern Ireland is different: the MHRA guidance states that the EU MDR has applied there since 26 May 2021 and that CE marking is required, and the BSI national foreword to BS EN 14683:2025 says Annex ZA/ZZ and references to EU legislation remain valid for that market. The UK list discussed here covers Great Britain only.

How long can CE-marked medical devices be placed on the Great Britain market?

The gov.uk guidance "Regulating medical devices in the UK" (updated 20 February 2026) gives these timelines:

  • general medical devices compliant with the EU MDD or AIMDD, with a valid declaration and CE marking: until the sooner of certificate expiry or 30 June 2028;
  • IVDs compliant with the EU IVDD: until the sooner of certificate expiry or 30 June 2030;
  • general medical devices, including custom-made devices, compliant with the EU MDR, and IVDs compliant with the EU IVDR: until 30 June 2030.

On 21 October 2026 the government published its response to the targeted consultation on indefinite recognition of CE-marked medical devices. It intends to legislate to recognise CE-marked devices certified under the EU MDR and IVDR indefinitely and to align MDD transitional arrangements with EU timelines; the response notes that the EU has extended the MDR transition for certain devices to 31 December 2028. Subject to parliamentary approval, the changes are to be made before the earliest relevant transition period ends on 30 June 2028, and CE-marked devices must still meet domestic obligations such as MHRA registration and post-market surveillance. As checked on 13 November 2026, the gov.uk guidance was still the 20 February 2026 version; exact deadlines follow the legislation and MHRA's published text.

Why does this matter for editions? UK designated standards serve regulation 3A of the UK MDR 2002; EU harmonised standards serve Article 8(1) of the EU MDR. A CE-marked device in Great Britain relies on the EU MDR; the UK list gives presumption of conformity for declarations under the UK MDR 2002. Products sold in both markets may be asked about both sets of editions; how they are cited is for the assessing body to decide.

UK designated standards v2 vs EU harmonised editions

Sources: UK column, consolidated list version 2 (ODS, 17 January 2026) and notice 0129/26; EU column, the European Commission's MDR summary list (generated 11 November 2026) and each implementing decision, with the Official Journal publication date; SUNGO columns, the Shanghai CNAS schedule (L22083) and IAS scope (TL-1019).

StandardUK v2 designated edition and start of presumptionEU MDR harmonised edition and decisionSUNGO CNAS linesSUNGO IAS
EN 455-1 gloves, freedom from holesEN 455-1:2020+A2:2024, from 17 Jan 2026; EN 455-1:2000 removed same dayEN 455-1:2020+A2:2024, (EU) 2025/681, 9 Apr 2025EN 455-1:2020+A2:2024, watertightness testEN and BS EN 455-1, whole standard
EN 455-2 physical propertiesEN 455-2:2024, from 17 Jan 2026; EN 455-2:2009+A2:2013 removed same dayEN 455-2:2024, (EU) 2025/681, 9 Apr 2025EN 455-2:2024, physical propertiesEN and BS EN 455-2, whole standard
EN 455-3 biological evaluationEN 455-3:2023, from 17 Jan 2026; EN 455-3:2006 removed same dayEN 455-3:2023, (EU) 2024/815, 8 Mar 2024EN 455-3:2023 clause 4.4, powder residue (only 4.4 powder residue is tested)EN and BS EN 455-3, only 5.2
EN 455-4 shelf lifeEN 455-4:2009, from 1 Jan 2021, still listedNot in the MDR summary listEN 455-4:2009, shelf lifeEN and BS EN 455-4, whole standard
EN 14683 medical face masksEN 14683:2019+AC:2019, from 1 Jan 2021, still listedEN 14683:2025, (EU) 2025/2078, 20 Oct 2025EN 14683:2025, seven items; BS EN 14683:2025, five items (no BFE, no cleanliness); no 2019 editionEN and BS EN 14683, whole standard
EN 13795-1 surgical gowns and drapesEN 13795-1:2019, from 1 Jan 2021, still listedEN 13795-1:2025, (EU) 2025/2078, 20 Oct 2025EN 13795-1:2025, seven items; no 2019 editionEN 13795-1, whole standard
EN 13795-2 clean air suitsEN 13795-2:2019, from 1 Jan 2021, still listedEN 13795-2:2025, (EU) 2025/2078, 20 Oct 2025EN 13795-2:2025, five items; no 2019 editionEN 13795-2, whole standard
EN ISO 11607-1 sterile barrier systemsEN ISO 11607-1:2020 with A1:2023, from 17 Jan 2026; EN ISO 11607-1:2009 removed same dayEN ISO 11607-1:2020 with A1:2023, (EU) 2024/815, 8 Mar 2024NoneISO 11607-1, whole standard
EN ISO 11737-1 bioburdenEN ISO 11737-1:2018 with A1:2021, from 17 Jan 2026; 2006 edition removed same dayEN ISO 11737-1:2018 with A1:2021, (EU) 2022/6, 5 Jan 2022EN ISO 11737-1:2018+A1:2021, surgical gowns and drapes, microbial cleanlinessISO 11737-1, whole standard
EN ISO 11737-2 sterility testsEN ISO 11737-2:2020, from 17 Jan 2026; 2009 edition removed same dayEN ISO 11737-2:2020, (EU) 2021/1182, 19 Jul 2021ISO 11737-2:2019, sterility testISO 11737-2, whole standard

IAS entries carry no edition years; confirm the edition with the laboratory before an IAS report is issued. "Removed same day" means the old edition's removal date (end of presumption of conformity) on the list is 17 January 2026, the new edition's start date. The (EU) 2021/1182 annex shows ISO 11737-2:2019 in brackets after EN ISO 11737-2:2020.

Gloves: UK and EU editions match

Version 2 replaced the first three glove parts in one step, with no overlap. Old reports to EN 455-1:2000, EN 455-2:2009+A2:2013 or EN 455-3:2006 can be gap-assessed against the new editions for Great Britain; the manufacturer records the conclusion in the technical documentation. SUNGO's CNAS schedule has the same EN 455-1, -2 and -3 editions as v2, and EN 455-4 is the 2009 edition on both.

Masks and gowns: the UK still lists 2019

The recitals of Implementing Decision (EU) 2025/2078 state that CEN and CENELEC revised EN 13795-1:2019, EN 13795-2:2019 and EN 14683:2019+AC:2019 (whose references are not published in the Official Journal), resulting in the 2025 editions; the decision entered into force on publication on 20 October 2025. The UK has designated the 2019 editions since 1 January 2021; v2 did not change them. For clause-level changes, see our articles on EN 14683:2025 vs 2019 for medical face masks and EN 13795-1:2025 vs 2019 for surgical gowns and drapes.

Packaging and microbiology: v2 catches up

The UK previously designated EN ISO 11607-1:2009, EN ISO 11737-1:2006 (with AC:2009) and EN ISO 11737-2:2009. Since 17 January 2026, v2 designates the newer editions shown above, matching the EU MDR, so packaging and microbiology data for sterile gloves and gowns can follow the same editions for both markets.

BS EN 14683:2025 or EN 14683:2019: which one does the UK recognise?

BSI publishes British Standards; the UK government designates standards. A BS EN is BSI's UK adoption of a European standard. The national foreword to BS EN 14683:2025 states that it is the UK implementation of EN 14683:2025 and supersedes BS EN 14683:2019, which is withdrawn; BSI's product record marks it "Identical" to EN 14683. The BS EN 13795-1:2025 foreword likewise supersedes and withdraws BS EN 13795-1:2019.

BSI withdrawing an edition is not the government removing a designation. The BS EN 14683:2025 foreword also says that, for Great Britain, if the government has designated the publication for UKCA marking (or similar) legislation, it may contain a National Annex showing its correlation with UK legislation. As checked on 13 November 2026, the gov.uk list of 17 January 2026 still names EN 14683:2019+AC:2019 and EN 13795-1/-2:2019; any change of designated edition is a matter for a new DHSC notice.

The BS EN or EN number on the report also decides which SUNGO CNAS lines apply, as the two sets cover different items (see the scope table below).

Handling the edition gap for mask and gown reports used in the UK

  • Confirm the route and market first. CE marking in Great Britain, a UK MDR 2002 declaration and the Northern Ireland market each point to a different standards list; for both the EU and Great Britain, state both when ordering.
  • A 2025-edition report for a UK MDR 2002 declaration. The manufacturer explains the 2025 vs 2019 differences in the technical documentation and justifies how the 2025 results cover the 2019 requirements. Acceptance depends on the assessing body or the MHRA.
  • Designated standards are voluntary. Regulation 3A and the MHRA guidance both say so. Testing to the designated edition gives the presumption of conformity; otherwise the technical documentation must show separately how the requirements are met.
  • A report to the 2019 edition. Not under CNAS, whose schedule lists only the 2025 editions of EN 14683, BS EN 14683, EN 13795-1 and EN 13795-2. IAS lists these as whole standards without edition years; whether a 2019-edition IAS report can be issued is confirmed at quotation.

Accreditation scope table

Reproduced from the Shanghai CNAS schedule (L22083), with item names, clauses and limitations as listed; for EN 14683 and EN 13795 only the 2025 editions are listed.

ItemStandard and clauseNotes/limitations
Single-use medical gloves: watertightness testEN 455-1:2020+A2:2024No clause listed
Single-use medical gloves: physical propertiesEN 455-2:2024No clause listed
Single-use medical gloves: powder residueEN 455-3:2023 4.4Only 4.4 powder residue is tested
Single-use medical gloves: shelf lifeEN 455-4:2009No clause listed
Medical face masks: materials and constructionEN 14683:2025 5.1.1; BS EN 14683:2025 5.1.1—
Medical face masks: designEN 14683:2025 5.1.2; BS EN 14683:2025 5.1.2—
Medical face masks: bacterial filtration efficiency (BFE)EN 14683:2025 5.2.2, Annex BNot listed for BS EN 14683:2025
Medical face masks: differential pressureEN 14683:2025 5.2.3, Annex C; BS EN 14683:2025 5.2.3, Annex C—
Medical face masks: splash resistanceEN 14683:2025 5.2.4; BS EN 14683:2025 5.2.4—
Medical face masks: cleanlinessEN 14683:2025 5.2.5, Annex DNot listed for BS EN 14683:2025
Medical face masks: marking, labelling and packagingEN 14683:2025 6; BS EN 14683:2025 6—
Surgical gowns and drapes: resistance to microbial penetration, dryEN 13795-1:2025 A 2.6—
Surgical gowns and drapes: resistance to microbial penetration, wetEN 13795-1:2025 A 2.7—
Surgical gowns and drapes: microbial cleanlinessEN 13795-1:2025 A 2.1; EN ISO 11737-1:2018+A1:2021—
Surgical gowns and drapes: lintingEN 13795-1:2025 A 2.2—
Surgical gowns and drapes: resistance to liquid penetrationEN 13795-1:2025 A 2.3—
Surgical gowns and drapes: bursting strength, dry and wetEN 13795-1:2025 A 2.4—
Surgical gowns and drapes: tensile strength, dry and wetEN 13795-1:2025 A 2.5—
Clean air suits: resistance to microbial penetration, dryEN 13795-2:2025 A 2.5—
Clean air suits: microbial cleanlinessEN 13795-2:2025 A 2.1—
Clean air suits: lintingEN 13795-2:2025 A 2.2—
Clean air suits: bursting strength, dryEN 13795-2:2025 A 2.3—
Clean air suits: tensile strength, dryEN 13795-2:2025 A 2.4—
Medical devices: sterility testISO 11737-2:2019No clause listed

The Shanghai IAS scope (TL-1019) lines are in the SUNGO IAS column of the comparison table: no edition years, all whole standards except EN 455-3 and BS EN 455-3 (5.2 only).

Where SUNGO does this work and under which accreditation

The accredited items above are all performed at the laboratory of Shanghai SUNGO Medical Technology Co., Ltd. (958 Kangqiao East Road, Pudong New Area, Shanghai) under CNAS L22083 and IAS TL-1019; CNAS and IAS are both signatories to the ILAC MRA. None of these standards is on the Shanghai CMA (240015340087) schedule, so no CMA report is possible; CMA covers only items on its schedule and is valid only within China. The certificates of Hefei SUNGO Medical Technology Co., Ltd. do not include these standards. See also the medical face mask testing and surgical gown testing pages.

Which line to use:

  • Mask reports with the BS EN number. The five BS EN 14683:2025 CNAS lines omit BFE and cleanliness (see the table); those go under the EN 14683:2025 CNAS lines, or IAS once the edition is confirmed.
  • Gloves, EN 455-3. CNAS covers only 4.4 powder residue and IAS only 5.2; other clauses of that part are outside SUNGO's accredited scope.
  • EN ISO 11607-1 and EN ISO 11737-1. ISO 11607-1 is IAS only. The EN ISO 11737-1 CNAS line covers surgical gowns and drapes (microbial cleanliness); bioburden for other products is confirmed under IAS.
  • Biological tests. SUNGO does not perform biological tests such as cytotoxicity, sensitisation or irritation itself. If needed, they are subcontracted to an appropriately qualified laboratory with the client's prior consent, and the report names the subcontracted items and subcontractor.

Preparing a test request

  • Target market and route: Great Britain (CE marking or UK MDR 2002), Northern Ireland, the EU, or several.
  • Standard number and edition: EN or BS EN, 2025 or 2019. A 2019 edition is only possible under IAS and must be confirmed first.
  • For masks, the claimed type; for gowns, drapes and clean air suits, the product category and claimed performance level.
  • Materials and construction, sterile or not, intended use, and draft labels, packaging and instructions for use (EN 14683 clause 6 is within the CNAS scope).
  • Existing reports to older editions, to judge which items need retesting; justifying edition differences remains the manufacturer's job.
  • Sample quantities and sampling follow the standard text and are confirmed with the laboratory.

Frequently asked questions

When was version 2 of the UK designated standards list for medical devices published?

Consolidated list version 2 and notice 0129/26 are both dated 17 January 2026; they were issued by DHSC and apply to England, Scotland and Wales.

Should medical face masks for the UK be tested to EN 14683:2019 or 2025?

As of 13 November 2026, the UK list designates EN 14683:2019+AC:2019 and the EU MDR has harmonised EN 14683:2025. It depends on the route (CE marking or UK MDR 2002) and the assessing body; if a 2025-edition report supports a UK MDR 2002 declaration, the manufacturer explains the edition differences.

Is BS EN 14683:2025 the same as EN 14683:2025?

BSI marks them as identical: BS EN 14683:2025 is the UK adoption of EN 14683:2025. On SUNGO's CNAS schedule, however, the BS EN lines exclude BFE and cleanliness, which go under the EN 14683:2025 lines or IAS.

How long can CE-marked medical devices be sold in Great Britain?

The gov.uk guidance (updated 20 February 2026) gives 30 June 2030 for EU MDR-compliant devices and the sooner of certificate expiry or 30 June 2028 for MDD-compliant devices. On 21 October 2026 the government said it will legislate for indefinite recognition of EU MDR and IVDR CE-marked devices, subject to parliamentary approval.

Can a test report still be issued to EN 13795-1:2019?

At SUNGO, not under CNAS, which lists only EN 13795-1:2025 and EN 13795-2:2025. IAS lists EN 13795-1 and EN 13795-2 without edition years; whether a 2019-edition report can be issued is confirmed at quotation.

To test gloves, masks, gowns or related packaging and microbiology items to UK or EU editions, use the contact page to give the product category, target market, standard number and edition, and any existing reports; the laboratory will confirm the applicable accreditation and how the report can be issued.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.