When medical-grade paper, crepe paper or nonwovens are used as sterilization packaging, regulatory files, customer audits and buyers often ask for wet and dry microbial barrier data on the material. In China three documents are commonly cited: the German standard DIN 58953-6, the industry standard YY/T 0681.14-2018, and clause 2.1.7.5.2 of the Technical Standard for Disinfection (2002 edition) issued by China's Ministry of Health. Their test logic is similar, but their editions, texts and acceptance wording differ, and so do the accredited reports SUNGO's Shanghai laboratory can issue against each. This article covers how the three relate, how the moist and dry tests work and are judged, what changed in the 2025-03 edition of DIN 58953-6, and which accreditation mark applies to which test.
Key points first
- Two conditions, tested separately. All three documents test porous packaging materials under moist conditions (the material is wetted by a liquid carrying microorganisms) and dry conditions (the material stays dry, and the question is whether microorganisms carried by air passing through it get through). DIN Media's English summary of DIN 58953-6 calls the dry condition "air passage" (German: Luftdurchgang).
- Moist method: CNAS for all three, no CMA for DIN. At the Shanghai laboratory, the moist method to YY/T 0681.14-2018 or the Technical Standard for Disinfection (2002 edition) can be reported under CNAS or CMA. The moist method to DIN 58953-6:2025 can be reported under CNAS, but DIN 58953-6 is not in the CMA scope. All three documents are also in the Shanghai IAS scope.
- Dry method: IAS only. CNAS (all three documents) and CMA (YY/T 0681.14-2018 and the Technical Standard for Disinfection only) accredit only "microbial barrier (moist method)". The dry test can be reported only under IAS: DIN 58953-6 and YY/T 0681.14 as whole standards, the Technical Standard for Disinfection for clause 2.1.7.5.2 only.
- DIN 2025-03 edition. As of 2026-10-22, DIN Media lists the 2025-03 edition of DIN 58953-6 as current; it is a corrected edition of the 2023-11 edition. Compared with 2016-12, it adds a terms and definitions clause, renumbers the following clauses and adds controls to the air-passage test, among other changes.
- A material test is not a system validation. These tests ask whether the material itself holds back microorganisms. They do not validate sterility maintenance of the packaging system.
The three documents: editions and scope
| Document | Edition information | Applies to | Methods |
|---|---|---|---|
| DIN 58953-6:2025 | German standard, published 2025-03 in German; corrected edition of 2023-11 | Packaging materials for medical devices which are to be sterilized | Microbial barrier determination in the case of humidity and air passage |
| YY/T 0681.14-2018 | Recommended Chinese pharmaceutical industry standard, issued 2018-11-07, implemented 2019-11-01; listed as current as of 2026-10-22 | Porous packaging materials for sterile medical device packaging (see clause 1 of the standard for the exact scope) | Moist and dry conditions |
| Technical Standard for Disinfection (2002 edition), 2.1.7.5.2 | Issued by the Ministry of Health (Wei Fa Jian Fa [2002] No. 282), took effect 2003-04-01 | Porous materials used to package medical supplies that are sterilized after packaging | (1) moist conditions; (2) dry conditions |
DIN 58953-6: the 2025-03 edition
The English title of DIN 58953-6 is Sterilization – Sterile supply – Part 6: Microbial barrier testing of packaging materials for medical devices which are to be sterilized, part of the DIN 58953 series. According to the scope published by DIN Media, it specifies test methods for microbial barrier determination in the case of humidity and air passage, applicable to packaging materials for medical devices which are to be sterilized.
DIN Media shows this edition history: the 2016-12 edition replaced 2010-05, mainly adding an Annex A on reproducibility of the method; the 2023-11 edition followed; and 2025-03 is a corrected edition of 2023-11, supplied free to buyers of 2023-11. The 2016-12 edition is marked as withdrawn. The SUNGO Shanghai CNAS scope lists DIN 58953-6:2025.
YY/T 0681.14-2018: moist and dry tests, relationship per its foreword
YY/T 0681.14-2018 is a recommended Chinese pharmaceutical industry standard whose Chinese title covers wet and dry microbial barrier tests for porous packaging materials. It was issued on 2018-11-07 and implemented on 2019-11-01; as of 2026-10-22 China's national standards information platform lists it as current. In the English title listed in SUNGO's IAS scope, the two tests appear as "under moist conditions" and "with passage of air".
Judging by its title, it follows the same two-method framework as DIN 58953-6. How far it adopts DIN 58953-6, and which DIN edition it refers to, should be taken from the foreword of YY/T 0681.14-2018 itself. One point is clear from the dates: YY/T 0681.14-2018 was issued in November 2018, before the 2023-11 and 2025-03 DIN editions, so their changes are not reflected in it.
Technical Standard for Disinfection, 2.1.7.5.2
The Technical Standard for Disinfection (2002 edition) was issued by the Ministry of Health and took effect on 1 April 2003. The microbial barrier tests sit in Part 2, the testing specifications for disinfection products, under 2.1.7, evaluation tests for packaging materials of medical supplies that require terminal sterilization (sterilization after packaging). Clause 2.1.7.5 has two sub-clauses: 2.1.7.5.1, an impermeability (dye penetration) test, and 2.1.7.5.2, the microbial barrier test for porous materials, split into (1) moist conditions and (2) dry conditions.
Clause 2.1.7.7 on sterile shelf life also refers back to 2.1.7.5: samples after natural storage or accelerated aging are tested for microbial barrier by the 2.1.7.5 methods. SUNGO's CNAS and IAS scopes list the 2.1.7.5.2 test itself (the CMA schedule gives no clause number; its item is microbial barrier, moist method); the 2.1.7.5.1 dye penetration test and the storage and aging arrangements of 2.1.7.7 are outside SUNGO's accredited scope.
The document matters mainly for products regulated as disinfection products. China's 2014 regulation on hygiene and safety evaluation of disinfection products (Guo Wei Jian Du Fa [2014] No. 36) classes "packaging for sterilized items bearing a sterilization label" as a Class II disinfection product, which needs a hygiene and safety evaluation before first being placed on the market. The tests required are set by that regulation's annexes and the relevant standards, and the authority's current requirements prevail. For a medical device manufacturer, the right document depends on how the product is regulated and which document the reviewer, the product technical requirements or the customer cites.
Moist-condition test: inoculate one side, look for growth on the other
This test looks at the outside of a package being wetted by a contaminated liquid and asks whether, once dry, microorganisms reach the other side. In the Technical Standard for Disinfection, drops of suspension are placed on the outer surface of the sample, allowed to dry, and the reverse side is then brought into contact with agar, incubated and checked for colonies.
In 2.1.7.5.2 (1) of the Technical Standard for Disinfection, the main steps are:
- Sample pieces are steam sterilized, then vacuum dried.
- The test organism is Staphylococcus aureus (ATCC 6538); a viable count is made on the cultured suspension.
- Each piece is laid outer surface up in a sterile Petri dish, and five separate, non-touching drops of suspension are placed on it.
- The pieces are left to dry under the specified temperature and humidity.
- Each piece is laid on a blood agar plate, inoculated side up so the reverse side is in full contact with the agar, and removed after a few seconds.
- The plates are incubated and colonies counted. Organisms that penetrated to the reverse side grow on the agar.
The moist-condition tests in YY/T 0681.14 and DIN 58953-6 also ask whether microorganisms pass through the wetted material, but test organism, inoculation and drying, contact with the medium, controls and number of pieces follow each document's own text. Do not carry parameters from one document to another, and do not substitute the disinfection standard's steps.
Dry-condition test: contaminated powder plus temperature cycling
This test looks at air moving into and out of a package and asks whether airborne microorganisms get through the material. The Technical Standard for Disinfection seals the sample over a test bottle containing culture medium, spreads contaminated quartz powder on top, and makes the air inside expand and contract so that air is forced through the sample.
In 2.1.7.5.2 (2) of the Technical Standard for Disinfection, the main steps are:
- The test organism is spores of Bacillus subtilis var. niger (ATCC 9372). A spore suspension is mixed with sterile quartz powder and dried.
- Nutrient agar is poured into glass test bottles and allowed to set.
- A circular sample piece is clamped between two PTFE or PTFE-coated sealing rings and held against the bottle rim by a screw cap with an opening.
- The bottles are wrapped in aluminium foil and sterilized; after cooling, the contaminated quartz powder is spread over the sample.
- The bottles are warmed in an incubator, then cooled in a refrigerator. That counts as one cycle; the standard says to repeat it five times.
- The bottles are incubated and colonies on the agar counted.
Warming and cooling make the air in the bottle expand and contract, so outside air is drawn in through the sample as the bottle cools. If the material does not hold, spores from the powder are carried in and grow on the agar. In DIN 58953-6:2025 the air-passage test is clause 5; in YY/T 0681.14-2018 it is the dry-condition test. Apparatus and detailed parameters follow each document's own text.
How results are judged
The Technical Standard for Disinfection gives a pass or fail evaluation; it does not rank the material into classes. Its result reporting and evaluation are:
- Moist conditions: report the colonies on each plate and the total for five plates. No growth on any of the five plates means the test organism did not penetrate. If the five-plate total is not more than five colonies, retest with 20 pieces; not more than five colonies on the 20 plates is a pass.
- Dry conditions: report the colonies passing each piece and the total for ten pieces. Each piece shall not exceed 5 cfu, and the ten-piece total shall not exceed 15 cfu.
For DIN 58953-6 and YY/T 0681.14, the way results are evaluated, the number of pieces, acceptance figures and retest rules come from their own texts, so do not read a report made to either standard against the disinfection standard's figures. Compared with 2016-12, DIN 58953-6:2025 also extends guidance on evaluating results and adds controls to the air-passage test; when reading a report made to DIN 58953-6:2025, check that the controls are valid as well as the sample results.
Ranking the microbial barrier of porous materials in an exposure chamber is a separate type of method with a different form of result. It is outside this article and covered in a separate one.
What changed in DIN 58953-6:2025
DIN Media's change note lists these changes against DIN 58953-6:2016-12:
- normative references updated;
- new clause 3, "Terms and definitions", with following clauses renumbered;
- conditioning no longer required (clause 4);
- controls added for the air-passage test as well (clause 5);
- aseptic technique for performing the tests clarified;
- guidance on evaluating results extended;
- values and tolerances harmonized;
- information on culture media adjusted;
- bibliography updated;
- editorial revision.
Against 2023-11, the 2025-03 edition makes two corrections: in 4.6 and 4.7.3, "0.9 % NaCl solution" becomes "distilled water"; in 4.7.1, "light rolling with the Drigalski spatula" becomes "light pressing with the Drigalski spatula".
Three practical consequences:
- Clause numbers have changed. Documents, protocols or technical requirements citing 2016-12 clause numbers need checking clause by clause.
- State the edition precisely. 2023-11 and 2025-03 differ only by corrections, but reports and files should name the edition actually used.
- Keep the Chinese standard and the 2025 DIN apart. YY/T 0681.14-2018 predates these changes, so results under each should cite their own basis in a regulatory file.
What the test shows and what it does not
As their scopes show, all three documents test packaging materials. So:
- It shows whether the porous material itself, under the specified moist or dry conditions, holds back the test organism.
- It does not show seal quality, whether formed pouches or trays are intact, or whether the packaging system stays sterile after sterilization, transport and storage.
A sterile barrier system validation also needs evidence on seals and integrity, plus evaluation after aging and transport. The Shanghai CNAS and CMA scopes also include YY/T 0681.2-2010 (seal strength), YY/T 0681.4-2021 (dye penetration) and YY/T 0681.11-2014 (integrity), which can be combined as a validation plan requires. For the overall approach, see Sterile Barrier Systems: Packaging and Sterilization Validation.
Accreditation scope table
Rows relevant to this article from the three certificate schedules of Shanghai SUNGO Medical Technology Co., Ltd. Items, standard numbers, clauses and limitations are taken from the schedules; notes are added explanations.
| Certificate | Item | Standard and clause | Notes / limitation |
|---|---|---|---|
| CNAS L22083 | Medical device packaging materials: microbial barrier (moist method) | DIN 58953-6:2025 | No clause restriction or limitation; DIN 58953-6 is not in the CMA scope |
| CNAS L22083 | Medical device packaging materials: microbial barrier (moist method) | YY/T 0681.14-2018 | No clause restriction or limitation |
| CNAS L22083 | Medical device packaging materials: microbial barrier (moist method) | Technical Standard for Disinfection (2002 edition), 2.1.7.5.2 | Clause 2.1.7.5.2; no limitation |
| CMA 240015340087 | Medical device packaging materials: microbial barrier (moist method) | YY/T 0681.14-2018 | No clause restriction or limitation; CMA is valid within the People's Republic of China |
| CMA 240015340087 | Medical device packaging materials: microbial barrier (moist method) | Technical Standard for Disinfection (2002 edition) | No clause number in the schedule; CMA is valid within the People's Republic of China |
| IAS TL-1019 | Whole standard | DIN 58953-6 | Listed without year of edition |
| IAS TL-1019 | Whole standard | YY/T 0681.14 | Listed without year of edition |
| IAS TL-1019 | Partial (2.1.7.5.2 only) | Technical Standard for Disinfection issued by Ministry of Health of China (2002), 2.1.7.5.2 | Limitation as listed: Accredited only for 2.1.7.5.2 Microbial barrier test for breathable materials |
Where the test is done and under which accreditation
The test is performed at the Shanghai laboratory of Shanghai SUNGO Medical Technology Co., Ltd., No. 958 Kangqiao East Road, Pudong, Shanghai, which holds CNAS (L22083), CMA (240015340087) and IAS (TL-1019) accreditation. Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity whose scope does not include these tests.
| Basis | Moist conditions | Dry (air-passage) conditions |
|---|---|---|
| DIN 58953-6 (CNAS scope: 2025 edition; IAS scope: no year listed) | CNAS, IAS; no CMA | IAS only |
| YY/T 0681.14 (CNAS and CMA scopes: 2018 edition; IAS scope: no year listed) | CNAS, CMA, IAS | IAS only |
| Technical Standard for Disinfection, 2002 edition (CNAS and IAS scopes: 2.1.7.5.2; CMA schedule: no clause number) | CNAS, CMA, IAS | IAS only |
- CMA covers only items in its schedule and is valid within the People's Republic of China. The relevant CMA rows cover the moist method only and exclude DIN 58953-6.
- CNAS accredits DIN 58953-6:2025, YY/T 0681.14-2018 and the Technical Standard for Disinfection (2002 edition) 2.1.7.5.2, all for the moist method.
- IAS lists DIN 58953-6 and YY/T 0681.14 without a year of edition. For an IAS report to a specific edition, confirm the edition with the laboratory before ordering.
- Moist and dry together: the two carry different marks, so state which report you need for each; the laboratory will confirm how reports are issued.
- Acceptance: an accreditation mark shows competence within the scope. Whether a reviewer, notified body or buyer accepts a given report depends on that party's requirements.
Preparing samples
- Basis and edition: DIN 58953-6:2025, YY/T 0681.14-2018 or the Technical Standard for Disinfection 2.1.7.5.2; moist, dry or both; CNAS, CMA or IAS report.
- Recipient's requirements: for registration or audits, confirm the expected document and edition first, so the test is not repeated.
- Material details: name and type (for example medical-grade paper, crepe paper or nonwoven), specification and batch number.
- Exposed side: mark the side that faces outward in use. The disinfection standard's moist test inoculates the outer surface; for the other documents follow their text. If unsure which side faces outward, say so.
- Sample condition: raw, sterilized or aged material. The disinfection standard's moist test itself specifies steam sterilization of the pieces; if you want to evaluate material after a particular sterilization method or aging, agree the pretreatment in advance and write it into the request.
- Handling: flat, uncreased, undamaged and clean, packed separately, test area not touched by bare hands.
- Quantity: allow for the pieces required by the chosen document and a possible retest (the disinfection standard's moist test includes one). Confirm the quantity with the laboratory before ordering.
Frequently asked questions
Can DIN 58953-6:2025 and YY/T 0681.14-2018 replace each other?
Not directly. Judging by their titles, both cover moist and dry tests, but the editions and texts differ. Use the one required by the regulator, the product technical requirements or the customer. DIN 58953-6 also cannot be reported under CMA.
Can the dry (air-passage) test be reported under CNAS or CMA?
No. At the Shanghai laboratory, CNAS (all three documents) and CMA (YY/T 0681.14-2018 and the Technical Standard for Disinfection) cover the moist method only. The dry test can be reported only under IAS.
If the material passes, does the package still need other validation?
Yes. This is a material test. It does not replace seal and integrity evaluation, or evaluation of sterility maintenance after sterilization, transport and storage.
Is this the same as the wet bacterial penetration test for surgical gowns?
No. YY/T 0506.6-2009 applies to surgical drapes, gowns and clean air suits for patients, clinical staff and equipment. It is a different standard and method, and the reports cannot stand in for each other.
When requesting a quotation, include the material type and specification, the document and edition, moist, dry or both, the report type (CNAS, CMA or IAS), the sample condition and the intended use of the report. Related services are on the sterile packaging validation service page, and requests can be sent through the contact page. The laboratory will confirm which report types can be issued within its accredited scope.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.