ISO 10993-7:2026: Can EO Residual Reports to the 2008 Edition Still Be Used?

ISO 10993-7:2026: Can EO Residual Reports to the 2008 Edition Still Be Used?

ISO 10993-7:2026 was published on 20 April 2026. As the third edition, it replaces ISO 10993-7:2008 together with its 2009 Technical Corrigendum (Cor 1) and its 2019 Amendment 1 (Amd 1). Whether an ethylene oxide (EO) residual report written to the 2008 edition can still be used depends on the target market: the EU listed EN ISO 10993-7:2026 as a harmonised standard under the MDR with effect from 10 November 2026, while as of 13 November 2026 the FDA-recognized and UK-designated versions were still the 2008 edition. Whether an older report needs retesting is for the manufacturer to decide in its biological evaluation documentation, after comparing contact category, patient population, source of the limits and measurement method.

This article covers the facts about the new edition, which edition each market refers to, how to assess older reports, and which edition the SUNGO accreditation scope covers. It does not quote limit values; take them from the standard itself.

Key Takeaways

  • Edition: ISO 10993-7:2026 is the third edition, published 2026-04-20, 122 pages in English, owned by ISO/TC 194. ISO's published information shows that it replaces ISO 10993-7:2008, ISO 10993-7:2008/Cor 1:2009 and ISO 10993-7:2008/Amd 1:2019.
  • EU: Commission Implementing Decision (EU) 2026/2479 was adopted on 2026-11-09 and published in the Official Journal on 2026-11-10, entering into force that day. EN ISO 10993-7:2026 is entry No 68 in the Annex to Implementing Decision (EU) 2021/1182. A harmonised standard gives a presumption of conformity; using it is not mandatory.
  • US: The FDA Recognized Consensus Standards database has one entry for ISO 10993-7, recognition number 2-275: the 2008 edition including Cor 1:2009 and Amd 1:2019, complete standard (as of 2026-11-13).
  • UK: Version 2 of the GOV.UK consolidated list of designated standards for medical devices (2026-01-17) lists EN ISO 10993-7:2008 and EN ISO 10993-7:2008/AC:2009.
  • China: GB/T 16886.7-2015 is the current recommended national standard, an identical adoption of ISO 10993-7:2008.
  • Older reports: As of 13 November 2026, none of the official sources above sets a single cut-off date for reports to the 2008 edition. Keeping, supplementing or repeating older data is for the manufacturer to justify in its biological evaluation and risk management files. The laboratory supplies test data.
  • SUNGO: The Shanghai laboratory's CNAS accreditation scope and CMA schedule list ISO 10993-7:2008+A1:2019 and GB/T 16886.7-2015 (ECH limitations in the table below). The IAS scope covers Annex B of both, without an edition year. None of the three certificates names the 2026 edition. The Hefei laboratory does not have this capability.

What ISO 10993-7:2026 Covers: Scope and Exclusions

ISO 10993-7 is Part 7 of the biological evaluation of medical devices series, titled "Ethylene oxide sterilization residuals". According to the abstract ISO publishes for the 2026 edition, it specifies:

  • allowable limits (AL) for residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) in EO-sterilized medical devices;
  • procedures for the measurement of EO and ECH;
  • methods for determining conformity so that devices can be released.

Annexes A to K give further background, including guidance and a flowchart showing how the standard is applied.

Three scope boundaries, paraphrased from the abstract:

  • EO-sterilized devices or components with neither direct nor indirect body or user contact (the abstract cites in vitro diagnostic devices) are out of scope;
  • the standard does not apply to devices shown not to absorb or retain EO or its degradation product ECH, for example devices made exclusively of metal alloys and glass (see Clause C.5);
  • no limits are specified for ethylene glycol (EG), because the risk assessment in Annex F indicates calculated allowable levels are higher than those likely to occur in a medical device.

The Chinese counterpart is GB/T 16886.7-2015. The national standards information platform shows it was issued on 2015-12-10, took effect on 2017-01-01, is an identical adoption of ISO 10993-7:2008, and was confirmed as still valid at review on 2023-12-28. It therefore predates the 2019 Amd 1. Whether and when China adopts the 2026 edition depends on the text formally issued by the national standards authorities.

A related method standard is GB/T 14233.1-2022 (chemical analysis methods for infusion, transfusion and injection equipment), issued 2022-10-12 and effective 2023-11-01. Its Clause 9 covers EO residues and is often cited by product standards for masks, gauze and absorbent cotton. Sampling and the GC method under Clause 9 are covered in a separate article.

ISO 10993-7:2026 vs 2008: What Changed

The clause-level list of main changes is in the Foreword of the 2026 edition, which we could not check while writing, so we do not go through it item by item. The table compares only the abstracts and edition details ISO publishes for the two editions. Use it as a checklist when reading the new text.

Point of comparison2008 edition (abstract and edition data)2026 edition (abstract and edition data)
Edition and lengthSecond edition, published 2008, 86 pagesThird edition, published 2026-04-20, 122 pages
What it specifiesAllowable limits for EO and ECH in individual EO-sterilized devices, measurement procedures, methods for determining compliance for releaseAllowable limits (AL) for EO and ECH in EO-sterilized devices, measurement procedures, methods for determining conformity for release
AnnexesBackground, guidance and a flowchart in informative annexesBackground, guidance and a flowchart in Annexes A to K
Devices out of scopeEO-sterilized devices with no patient contact (e.g. IVD devices)Devices or components with neither direct nor indirect body or user contact (e.g. IVD devices)
Devices that do not absorb or retain EONot mentioned in the abstractNot applicable if shown not to absorb or retain EO and ECH (e.g. exclusively metal alloys and glass); see C.5
Ethylene glycol (EG)Not mentioned in the abstractNo EG limits specified, based on the risk assessment in Annex F
Neonates and infantsAddressed by Amd 1:2019, "Applicability of allowable limits for neonates and infants"Amd 1 is replaced by the third edition; check the new text for how the topic is handled

Two cautions. First, this compares abstracts only; something missing from an abstract may still be in the full text, so check the 2008 standard itself for similar provisions. Second, recital (5) of Implementing Decision (EU) 2026/2479 states that CEN and CENELEC revised these standards to take account of the latest technical and scientific progress. Do not treat the 2026 edition as a mere consolidation of 2008 plus its amendment; re-read the limits, categories and conformity methods in the new text.

ISO 10993-7 Edition by Market: EU, FDA, UK and China

The same EO residual data maps to different editions in different markets. Compiled from official sources; status as of 13 November 2026.

Market / bodyEdition referencedSource and status
ISOISO 10993-7:2026 (third edition)Published 2026-04-20; replaces 2008, Cor 1:2009 and Amd 1:2019
EU MDREN ISO 10993-7:2026Entry No 68 in the Annex to (EU) 2021/1182, added by (EU) 2026/2479; in force since 2026-11-10
US FDAISO 10993-7:2008 with Cor 1:2009 and Amd 1:2019Recognition number 2-275, complete standard, date of entry 2020-07-06; 2026 edition not in the database
UKEN ISO 10993-7:2008, EN ISO 10993-7:2008/AC:2009Designated standards list version 2 (2026-01-17), under S.I. 2002/618; presumption of conformity from 2021-01-01
ChinaGB/T 16886.7-2015 (identical to ISO 10993-7:2008)Current recommended national standard; confirmed at review on 2023-12-28

In the EU, EN ISO 10993-7:2026 Is a New Listing, Not a Replacement

Recitals (5) and (6) of (EU) 2026/2479 state that the references of EN ISO 10993-7:2008, its amendment EN ISO 10993-7:2008/A1:2022 and corrigendum EN ISO 10993-7:2008/AC:2009 had not been published under the MDR, and that the revised EN ISO 10993-7:2026 is listed as a new harmonised standard. There is therefore no older MDR reference to withdraw. The 10 May 2028 date in Article 2 applies only to points (1) and (3) of the Annex, which delete the older entries for EN ISO 11137-1 and EN ISO 14155.

Under Article 8(1) of the MDR, devices conforming to harmonised standards whose references are published in the Official Journal are presumed to conform to the requirements those standards cover. Article 2(1) of Regulation (EU) No 1025/2012 defines a standard as one with which compliance is not compulsory. A manufacturer may still show conformity by other means, but EN ISO 10993-7:2026 gives a direct presumption of conformity. Discuss with the notified body in advance how older reports will be used.

FDA Still Recognizes the 2008 Edition as of November 2026

Entry 2-275 carries the 2008 abstract and is recognized as a complete standard. Recognition List Number 066, published in the Federal Register on 24 August 2026, did not include ISO 10993-7. Guidance listed under 2-275 includes "Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices" (September 2018) and the guidance on sterility information in 510(k) submissions for devices labeled as sterile (January 2016). Decide which edition to cite in an FDA submission, and how to declare it, based on the database and these documents. Whether and when FDA recognizes the 2026 edition will show in future database updates.

The UK Still Lists the 2008 Edition with Its Corrigendum

Version 2 of the UK designated standards list includes only EN ISO 10993-7:2008 and AC:2009, without A1. As of 13 November 2026, the GOV.UK page was last updated on 17 January 2026.

How to Assess EO Residual Reports Written to the 2008 Edition

A new edition does not automatically invalidate older reports, nor can you assume they remain valid. Compare point by point and record the conclusion and rationale in the biological evaluation report and risk management file. The manufacturer owns this comparison; the laboratory supplies test data and a method description.

  1. Contact category: Which nature and duration of contact did the older report use to select limits, and does the 2026 edition categorize the device the same way? If not, confirm the category against the new text.
  2. Patient population: Amd 1:2019 dealt with whether the allowable limits apply to neonates and infants. If an older report relied only on the 2008 edition or GB/T 16886.7-2015 (neither includes Amd 1) and the intended population includes neonates or infants, re-check against the 2026 text.
  3. Source of limits: Identify which edition and clause the older report used for its decision and compare it with the 2026 allowable limits one by one. The closer old data are to the new limits, the more careful the justification or retesting needs to be. Take values from the standard itself.
  4. Measurement method: Are the older report's extraction approach, analytical method (such as gas chromatography), detection limit and quantitation limit adequate for a decision under the new edition? If only EO was measured, decide from the new text whether ECH must be determined.

Also confirm whether the product has changed since the older report, for example materials, packaging, sterilization process or aeration conditions; if so, the file must explain whether old data still represent the current product. EO sterilization validation (ISO 11135) and aeration process validation are outside the SUNGO accreditation scope, and this article draws no conclusions on either.

For devices made exclusively of metal alloys and glass, the 2026 abstract excludes devices shown not to absorb or retain EO and ECH. Whether a device qualifies depends on what Clause C.5 requires as demonstration; check the standard itself.

SUNGO Scope for ISO 10993-7 and GB/T 16886.7 (from the Certificates)

The table reproduces the relevant rows from the Shanghai laboratory's three certificates. Item names, clauses, editions and limitations follow the certificates (Chinese item names translated).

ItemStandard and clauseNotes / limitations
Ethylene oxide / ethylene chlorohydrin residualsISO 10993-7:2008+A1:2019CNAS L22083, test object "medical devices", no limitation stated; CMA 240015340087 schedule item "ethylene oxide / ethylene chlorohydrin residuals", no limitation stated
Ethylene oxide / ethylene chlorohydrin residualsGB/T 16886.7-2015CNAS, limitation: ECH (2-chloroethanol) residuals not tested
Ethylene oxide / ethylene chlorohydrin residualsGB/T 16886.7-2015CNAS, limitation: ECH residuals only
EO/ECH residuals (two CMA rows)GB/T 16886.7-2015CMA schedule: one row ("ethylene oxide / ethylene chlorohydrin residuals") with no limitation; one row ("EO/ECH residuals") with the limitation "ECH not tested"
Gas chromatographic determination of EOGB/T 16886.7-2015 Annex BCNAS, listed under "single-use medical masks"
Gas chromatographic determination for EO and ECHISO 10993-7 Annex B; GB/T 16886.7 Annex BIAS TL-1019, Annex B only, no edition year on the certificate
Ethylene oxide residualsGB/T 14233.1-2022 Clause 9 (CNAS clause 9 or 9.4)CNAS: listed under surgical masks, single-use medical masks, absorbent cotton gauze and medical absorbent cotton; CMA: listed under surgical masks, single-use medical masks, gauze and medical absorbent cotton; IAS covers GB/T 14233.1 Clause 9 (EO residue-GC), no edition year
ISO 10993-7:2026—CNAS and CMA list 2008+A1:2019; IAS lists only ISO 10993-7 Annex B, without an edition year; none of the three certificates names the 2026 edition
  • ECH coverage differs by row. For GB/T 16886.7-2015, CNAS has one row "ECH not tested" and one "ECH only"; CMA has one row without a limitation and one "ECH not tested". If you need ECH data in an accredited report, confirm at quotation which row and report type apply.
  • IAS covers Annex B only. The certificate names Annex B (gas chromatographic determination for EO and ECH) without an edition year; in the CNAS schedule, Annex B of GB/T 16886.7-2015 is likewise the gas chromatographic determination of EO. The 2026 edition has Annexes A to K; check the new annex numbering in the standard itself. Before an IAS report is issued to a particular edition, confirm the edition and annex mapping with the laboratory.
  • The certificates do not name the 2026 edition. Before a report to ISO 10993-7:2026 carries a CNAS or CMA mark, it must be confirmed that the scope has been updated. The IAS certificate carries no edition year, so confirm the edition and annex mapping with the laboratory before an IAS report to the 2026 edition.

Where SUNGO Performs EO Residual Testing and Under Which Accreditation

EO residual testing is performed at the laboratory of Shanghai SUNGO Medical Technology Co., Ltd. (958 Kangqiao East Road, Pudong, Shanghai), under CNAS L22083, CMA 240015340087 and IAS TL-1019. The scope of Hefei SUNGO Medical Technology Co., Ltd. does not include EO residuals. See our sterilization-related testing services.

  • CMA: covers only the items in its schedule; the certificate states it is valid within the People's Republic of China.
  • CNAS and IAS: both are signatories to the ILAC MRA. Whether an overseas regulator or notified body accepts a given report is for that body to decide; SUNGO makes no commitment on acceptance.
  • Other biological tests: cytotoxicity, sensitization, irritation, implantation and similar tests are outside the SUNGO accreditation scope. SUNGO subcontracts them to laboratories with the relevant qualifications; the client's consent is obtained in advance, and the report identifies the subcontracted items and the subcontractor. See Medical Device Biological Tests (ISO 10993): Subcontracting and EO Residuals.

Preparing Samples for EO Residual Testing

  • Target market and edition: state whether the report is for China, the EU, the US or the UK, which edition it should follow (GB/T 16886.7-2015, ISO 10993-7:2008+A1:2019 or other), and whether an accreditation mark is required.
  • ECH or not: EO only, or EO and ECH. This determines which certificate row applies.
  • Product information: materials, nature and duration of contact, intended patient population (including neonates or infants) and how a single device is used. These affect sampling and interpretation; the conformity decision is made by the client in its evaluation documentation.
  • Sterilization information: lot number, completion date and aeration conditions. Residual levels change with aeration time, so record how the shipping date relates to the sterilization date.
  • Original packaging: send samples in their final post-sterilization packaging, unopened. Confirm the number of samples with the laboratory based on the method and product.
  • Older report: if assessing whether older data can be kept, provide its edition, method and detection limit so any additional testing can be scoped.

FAQ

When was ISO 10993-7:2026 published, and what does it replace?

It was published on 20 April 2026 as the third edition and replaces ISO 10993-7:2008, ISO 10993-7:2008/Cor 1:2009 and ISO 10993-7:2008/Amd 1:2019.

Can EO residual reports to ISO 10993-7:2008 still be used?

It depends on the market. As of 13 November 2026, FDA and the UK still refer to the 2008 edition and the EU MDR lists the 2026 edition; none of them sets a single expiry date for older reports. The manufacturer justifies reuse after comparing contact category, patient population, source of limits and measurement method.

Is EN ISO 10993-7:2026 harmonised under the EU MDR, and is it mandatory?

It is harmonised: entry No 68 in the Annex to Implementing Decision (EU) 2021/1182 under (EU) 2026/2479, in force since 10 November 2026. It provides a presumption of conformity, but compliance with the standard itself is not compulsory.

Does FDA recognize ISO 10993-7:2026?

As of 13 November 2026, the FDA database lists only recognition number 2-275 for ISO 10993-7, covering the 2008 edition with Cor 1:2009 and Amd 1:2019. The 2026 edition is not listed.

Can SUNGO issue accredited reports to ISO 10993-7:2026?

None of the Shanghai laboratory's three certificates names the 2026 edition. The CNAS scope and CMA schedule list ISO 10993-7:2008+A1:2019 and GB/T 16886.7-2015; IAS covers Annex B only, without an edition year. A report to the new edition can carry a CNAS or CMA mark only after the scope is confirmed as updated; for an IAS report, confirm the edition with the laboratory first.

Request a Quote

For EO or ECH residual testing, or to work out what data an older report is missing, use our contact page to send the product name, materials, contact category, target market and the edition of the standard you need. The laboratory will confirm, against its certificate scope, which report type and accreditation mark can be issued.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.