Before testing assistive products for personal hygiene, such as commode chairs, shower chairs, raised toilet seats, bath stools and bathroom grab bars, four questions need answers: how ISO 17966 and the Chinese standard GB/T 43386 relate; which clauses apply to a given product; whether the Chinese commode chair (GB/T 24434) and shower chair (MZ/T 119) standards are older editions of GB/T 43386 or separate documents; and which accreditation mark the report can carry. This article answers them against the standard and the SUNGO Hefei laboratory's current scope.
Key points first
- Testing is performed at the Hefei laboratory (Hefei SUNGO Medical Technology Co., Ltd.). The Shanghai company's scope does not include these standards.
- Within SUNGO's accreditations, ISO 17966, GB/T 43386, GB/T 24434 and MZ/T 119 appear only in the Hefei IAS scope (TL-1119). They are not in the CNAS or CMA scopes, so reports to these standards can be issued under IAS accreditation only.
- The IAS scope lists ISO 17966 and GB/T 43386 "except for 4.3 Clinical evaluation and investigation, 5.3 Biocompatibility and toxicity", GB/T 24434 with no exclusions, and MZ/T 119 "except for 7 Materials". No edition years are given.
- GB/T 43386-2023 is an identical adoption of ISO 17966:2016. The ISO website shows the second edition, ISO 17966:2026 (retitled Assistive products for toileting, bathing and showering — Requirements and test methods), published in October 2026, with the 2016 edition withdrawn. Confirm the edition with the laboratory before ordering an IAS report.
- GB/T 24434 (commode chairs) and MZ/T 119 (shower chairs) are Chinese single-product standards. They are not earlier editions of GB/T 43386, and all three were in force as of 2026-10-22. GB/T 24434-2025 took effect on 2026-01-01 and fully replaces the 2009 edition.
- Biocompatibility (biological tests such as cytotoxicity, sensitization, irritation and implantation) is subcontracted by SUNGO to a laboratory with the appropriate qualifications. Under accreditation rules, the customer's consent is obtained in advance and the report identifies the subcontracted items and the subcontractor.
Standards and scope
ISO 17966:2016 and GB/T 43386-2023
ISO 17966:2016, Assistive products for personal hygiene that support users — Requirements and test methods, is the first edition, dated 2016-01-15, prepared by ISO/TC 173. Clause 1 states that it specifies requirements and test methods for assistive products for personal hygiene (APPHs) that support users and are intended to alleviate or compensate for disability, including the working environment and safety of assistants, during normal use and foreseeable misuse and failure.
GB/T 43386-2023, with the same title in Chinese, was published and took effect on 2023-11-27 under the Ministry of Civil Affairs and the national committee for rehabilitation and special equipment for persons with disabilities. The national standards platform records it as an identical adoption (IDT) of ISO 17966:2016 and classifies it as a "method" standard. Identical adoption implies the same technical content and structure, so clause numbers should match; check the GB/T 43386 text. The SUNGO IAS scope lists the same exclusions (4.3 and 5.3) for both.
Which products are covered
ISO 17966:2016 defines its scope by ISO 9999:2011 codes. The table maps them to common trade names; the standard text governs.
| ISO 9999:2011 code | Category in the standard | Common trade names |
|---|---|---|
| 09 12 03 | Commode chairs (mobile and static) | Commode chairs, wheeled commodes |
| 09 12 06 | Toilets with built-in raising and height-adjustable mechanism (excluding toilets with built-in douche and air dryer) | Height-adjustable toilets |
| 09 12 09, 09 12 12, 09 12 15, 09 12 18 | Toilet seats; raised toilet seats mounted on frame; toilet seat inserts; raised toilet seats fixed to toilet | Raised toilet seats, toilet seat risers, over-toilet frames |
| 09 12 21 | Toilet seats with built-in raising mechanism to help standing up and sitting down | Powered toilet lifts |
| 09 12 24, 09 12 25 | Toilet arm supports and back supports, mounted on the toilet or free-standing | Toilet safety frames |
| 09 12 36 | Douches and air dryers for attachment to a toilet | Bidet attachments |
| 09 33 03 | Bath/shower chairs (with and without wheels), bath boards, stools, back supports and seats | Shower chairs, bath stools, bath boards |
| 09 33 12 | Bathing stretchers, shower tables and diaper-changing tables | Shower trolleys, changing tables |
| 18 15 06 | Height-adjustable plinths and brackets, when used as an APPH | Height-adjustable basin brackets |
| 18 18 03, 18 18 06, 18 18 10, 18 18 11 | Handrails and support rails; fixed grab bars and handgrips; removable grab rails and handgrips (excluding those that are static as defined in 3.27); hinged rails and arm supports | Bathroom grab bars, suction grab rails, fold-up rails |
The standard expressly does not cover: safe mounting in building structures; fixed building installations such as water and electricity; bathtub hoists covered by ISO 10535; bathtubs (ISO 9999 code 09 33 21); stability and friction issues related to surfaces made slippery by soap; and products customised or custom-made for an individual user.
The new edition, ISO 17966:2026
The ISO website, as checked on 2026-10-22, shows ISO 17966:2026 as edition 2, published in October 2026, and ISO 17966:2016 as withdrawn. According to the ISO abstract, the new scope uses ISO 9999:2022 and lists bath boards (09 33 04), bath seats (09 33 05), shower chairs with and without wheels (09 33 07) and back supports for bath or shower (09 33 08); code 09 33 03 of the 2016 edition no longer appears. It also lists four categories absent from the 2016 edition: toilet splash guards (09 12 10), wash basins (09 33 15), bidets (09 33 18) and drying aids (09 33 36). The abstract states that toilets (09 12 06), toilet seats (09 12 09), douches and air dryers (09 12 36) and these four categories have no specific requirements or test methods; only the relevant general requirements apply. It also notes that some of these products can be classified as medical devices.
As of 2026-10-22, the Chinese national standards platform still lists GB/T 43386-2023, adopting ISO 17966:2016, as current. Whether clause numbers, including 4.3 and 5.3, changed in the 2026 edition should be checked against the standard text.
Structure of the standard and main tests
The introduction to ISO 17966:2016 explains that Clauses 1 to 21 and Clause 25 apply to all products in the scope, while Clauses 22 to 24 add requirements for mobile, fixed and static products. To claim compliance, a product must meet all clauses relevant to its type; clauses on electrical components, for example, may not apply to a product without them.
Classify the product first
- Mobile: moved from one location to another on its own wheels or equivalent means, such as a wheeled commode or wheeled shower chair. Clause 22 applies.
- Fixed: designed to be fastened to a support or secured in a specific location, for example by screws, nuts or vacuum. Bath boards and toilet seats held in place by friction are not fixed products. Clause 23 applies.
- Static: stationary with the occupant in place during use, but not fastened to a support; it may be portable. Bath boards and toilet seats secured by friction belong here. Clause 24 applies.
Clause topics
| Clause | Topic | Main content |
|---|---|---|
| 4 | General requirements and test methods | Risk analysis, intended performance, clinical evaluation (4.3), fasteners, user mass and load limits |
| 5 | Materials | Flammability, biocompatibility and toxicity (5.3), infection control and cleaning, resistance to corrosion (5.5) |
| 6 | Emitted sound and vibration | Noise, vibration, audible warning devices |
| 7 and 8 | Electromagnetic compatibility; electrical safety | Emissions and immunity; single fault safety, emergency stopping, batteries, circuit protection |
| 9 and 10 | Overflow, leakage and ingress of liquids; temperatures of parts in contact with skin | Liquid ingress and leakage; skin-contact temperatures |
| 11 to 13 | Moving and folding parts; traps for parts of the body; folding and adjusting mechanisms | Squeezing, powered lifting speed, trapping; holes, clearances and V-shaped openings; locking |
| 14 and 15 | Lifting and carrying means; portable and hand-held products | Lifting and carrying test |
| 16 | Static strength, impact and durability | Force formulae, durability cycles; static strength of supports, seats and backs; durability; impact |
| 17 | Stability | Static stability requirements and test method |
| 18 to 21 | Surfaces, corners, edges and protruding parts; small parts; forces in soft tissues; ergonomic principles | General safety and ergonomic requirements |
| 22 | Mobile APPHs | Immobilizing means and brakes, threshold and door-frame shock tests, foot supports and push handles, turning diameter, moving forces |
| 23 | Fixed APPHs | Strength and durability tests for shower seats, bathing stretchers, fixed raised toilet seats, grab bars, hinged rails and plinths |
| 24 | Static APPHs | Toilet seat insert impact; raising seat durability; bath chair and bath board stability (including on a bathtub rim), bracket strength, board friction, handle forces |
| 25 to 27 | Information supplied by the manufacturer; packaging; test report | Pre-sale, user and service information; labelling |
How the mechanical tests work
Strength, durability and stability are covered in Clauses 16, 17 and 22 to 24. Clause 4.2 requires sufficient strength and durability for all loads expected in intended use, described in the manufacturer's information. Clause 4.7 requires the manufacturer to declare the maximum user mass, and the maximum load too where it differs. It states that products intended to support an occupant in a seated or lying position shall be capable of supporting a person with at least 100 kg body mass. If a product is intended for users under 100 kg, the design shall limit misuse by heavier users. The relevant test methods use the maximum user mass and maximum load specified by the manufacturer. Clause 16 gives the formulae for calculating forces (16.2) and the number of durability test cycles (16.3), and Clauses 23 and 24 add tests specific to each product category.
Pass/fail criteria and values are as given in the standard text.
Products with electrical functions
Products with powered lifting, douche and dryer functions or battery power also fall under Clause 7 (electromagnetic compatibility) and Clause 8 (electrical safety). The normative references of ISO 17966:2016 include IEC 60601-1, IEC 60601-1-2, CISPR 11, IEC 61000-3-2, IEC 61000-3-3, IEC 61000-4-3 and IEC 61000-4-8, as well as IEC 60335-1, IEC 60529 and IEC 60601-2-35. IEC 60601-1 and IEC 60601-1-2 are dated references (IEC 60601-1:2005+A1:2012, IEC 60601-1-2:2014); per Clause 2, only the edition cited applies.
Medical electrical safety and medical EMC testing are done at the Hefei laboratory. IEC 60601-1, IEC 60601-1-2, CISPR 11 and the IEC 61000 parts above are within its IAS accreditation (TL-1119), without edition years; IEC 60335-1, IEC 60529 and IEC 60601-2-35 are not within SUNGO's accreditation scope. For powered products, confirm the approach and editions with the laboratory before submission.
Commode chairs to GB/T 24434 and shower chairs to MZ/T 119
GB/T 24434-2025, Commode chairs, was published on 2025-06-30 and took effect on 2026-01-01, fully replacing GB/T 24434-2009. It has the same ministry and committee as GB/T 43386, is classified as a "product" standard, and is not recorded as adopting an international standard. According to the State Administration for Market Regulation, it specifies the classification, technical requirements, inspection rules, marking, packaging, transport and storage of commode chairs.
MZ/T 119-2018, Assistive products for showering — Shower chairs, is a civil affairs industry standard. It was published and took effect on 2018-11-05, was in force as of 2026-10-22, and is classified as a product standard. Whether products sold as bath chairs fall within its scope should be checked against the standard text.
How the three relate: GB/T 43386 is a general standard for a family of products, covering both commode chairs (09 12 03) and shower chairs (09 33 03); GB/T 24434 and MZ/T 119 are single-product standards. None replaces another. Whether a product must meet one or both depends on the registration route, target market or buyer.
Both appear only in the Hefei IAS scope (TL-1119), without edition years: GB/T 24434 with no exclusions, MZ/T 119 except for Clause 7 Materials. Confirm whether a GB/T 24434 report is to the 2009 or 2025 edition before ordering.
Accreditation scope table
Each row below is taken from the Hefei laboratory's IAS scope (TL-1119), with numbers, titles and limitations transcribed from the certificate. None of these standards is in SUNGO's CNAS or CMA scope.
| Item | Clauses | Notes / limitations |
|---|---|---|
| ISO 17966 Assistive products for personal hygiene that support users - Requirements and test methods | Whole standard except 4.3 and 5.3 | IAS: except for 4.3 Clinical evaluation and investigation, 5.3 Biocompatibility and toxicity; no edition year |
| GB/T 43386 Assistive products for personal hygiene that support users - Requirements and test methods | Whole standard except 4.3 and 5.3 | IAS: except for 4.3 Clinical evaluation and investigation, 5.3 Biocompatibility and toxicity; no edition year |
| GB/T 24434 Commode chairs | Whole standard | IAS; no exclusions listed; no edition year |
| MZ/T 119 Assistive products for showering-shower chairs | Whole standard except Clause 7 | IAS: except for 7 Materials; no edition year |
| ISO 9227 Corrosion tests in artificial atmospheres - Salt spray tests | Whole standard | IAS; no exclusions listed; no edition year. A normative reference of ISO 17966:2016; corrosion resistance is Clause 5.5, applied as stated in the standard text |
Handling the excluded clauses
- 4.3 Clinical evaluation and investigation: required where risk management shows a need, with clinical investigations conforming to ISO 14155. This is not a laboratory test, is outside SUNGO's scope, and is arranged by the manufacturer.
- 5.3 Biocompatibility and toxicity: outside SUNGO's scope. Where needed, it is handled through the subcontracting arrangement described above.
- MZ/T 119 Clause 7 Materials: the IAS scope excludes the whole clause, so it is outside SUNGO's accreditation scope and related conclusions cannot carry the IAS mark. If you need this clause covered, raise it when ordering and the laboratory will explain the possible arrangement; any biological testing involved follows the subcontracting arrangement above.
Where the work is done and under which accreditation
Testing takes place at the Hefei laboratory in Zipeng Town, Feixi County, Hefei, Anhui Province. The legal entity, Hefei SUNGO Medical Technology Co., Ltd., is an independent company holding CNAS L24279, CMA 260015342945 and IAS TL-1119; the standards here fall only under IAS TL-1119. Shanghai SUNGO Medical Technology Co., Ltd. (laboratory at 958 Kangqiao East Road, Pudong, Shanghai; CNAS L22083, CMA 240015340087, IAS TL-1019) is a separate legal entity whose scope does not include them. Both scopes are compared in our overview of SUNGO's testing scope.
When choosing a report type, note the following:
- Reports to these four standards carry the IAS mark only, not CNAS or CMA. If a domestic project needs a CMA stamp, note that these standards are not in SUNGO's CMA scope; the CMA certificate is valid within the People's Republic of China.
- The IAS certificate carries the ilac-MRA mark. Whether a regulator or buyer accepts a given report depends on their own requirements. See CNAS, CMA and IAS compared and how to check an accreditation scope.
- The IAS scope gives no edition years; confirm the edition with the laboratory before ordering, particularly for ISO 17966 (2016 or 2026) and GB/T 24434 (2009 or 2025).
Preparing for testing
- Product classification: state the ISO 9999 category; whether the product is mobile, fixed or static; whether it folds or dismantles; and whether it has electrical functions (if so, include ratings and battery details). This decides which of Clauses 22 to 24, 7 and 8 apply.
- Maximum user mass and maximum load: Clause 4.7 requires the manufacturer to declare them, and the relevant test methods use them. Make sure the values on the label, in the instructions and in the technical file match.
- Complete configuration: submit samples as marketed or registered, including arm and back supports, buckets, foot supports, castors and brakes. List optional accessories separately.
- Mounting hardware for fixed products: the standard does not cover mounting in building structures; supply the hardware and installation instructions so the sample can be installed as specified.
- Instructions, labels and packaging: Clauses 25 and 26 cover these; provide the documents used for the marketed product. Cleaning and disinfection instructions also relate to Clause 5.4.
- Standard, edition and report type: state the target market, standards and editions, and that an IAS report is required. Sample quantity depends on the tests selected and is confirmed at quotation.
- Biological testing: if biocompatibility testing of materials is needed, raise it when ordering so the subcontracting arrangement can be agreed in advance.
Frequently asked questions
Should a commode chair be tested to GB/T 24434 or GB/T 43386?
They are not two editions of one standard. GB/T 43386 is a general standard covering many personal hygiene products, including commode chairs; GB/T 24434 is the commode chair product standard. Whether you need one or both depends on your registration file, target market or buyer's specification.
ISO 17966:2026 has been published. Can SUNGO issue an IAS report to the new edition?
The Hefei IAS scope lists ISO 17966 without an edition year, and the title on the certificate is still the 2016 title. Confirm with the laboratory before ordering whether the report will be to the 2016 or the 2026 edition.
How is biocompatibility handled?
Clause 5.3 of ISO 17966 and GB/T 43386, and Clause 7 of MZ/T 119, are outside SUNGO's scope. Biological tests are subcontracted by SUNGO to a laboratory with the appropriate qualifications; the customer's consent is obtained in advance, and the report identifies the subcontracted items and the subcontractor.
Does a powered toilet lift also need electrical safety and EMC testing?
Clauses 7 and 8 of ISO 17966 apply to products with electrical functions. Medical electrical safety and medical EMC are performed by the Hefei laboratory within its IAS accreditation (TL-1119); confirm the approach and editions of referenced standards before submission. If electrical safety must be assessed to IEC 60335-1 or IEC 60601-2-35, those standards are not within SUNGO's accreditation scope.
Requesting a quotation
To arrange testing, start with the wheelchair and assistive products testing service page, then use the contact page to send your product list (product category, whether it has electrical functions, maximum user mass), target market, and the standards and editions you need. We will confirm which reports the Hefei laboratory can issue within its accreditation scope and which items need to be subcontracted.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.