YY 0469-2023 vs 2011: Surgical Mask Standard Changes, Clause by Clause

YY 0469-2023 vs 2011: Surgical Mask Standard Changes, Clause by Clause

YY 0469-2023, the Chinese standard for surgical masks, applies from 1 December 2026 and replaces YY 0469-2011 in full. The clause structure is essentially the same. The changes are in filtration efficiency, airflow resistance, microbiology, mask straps, marking and biocompatibility. This article compares the two editions clause by clause from the published texts. It also explains how to decide which items in a 2011-edition report need retesting, and which accreditation marks a new report can carry.

Key Points

  • Effective date: published 22 November 2023, applies from 1 December 2026, replaces YY 0469-2011 in full. It is a mandatory pharmaceutical industry standard.
  • Requirements: BFE rises from at least 95% to at least 98%, and PFE from at least 30% to at least 80%. "Pressure differential" becomes "airflow resistance": not more than 49 Pa becomes less than 60 Pa, with three test points per specimen averaged.
  • Methods: microbiology now follows General Chapters 1105 and 1101 of the Pharmacopoeia of the People's Republic of China (ChP), 2020 edition, Volume IV. EO residue follows GB/T 14233.1-2022 clause 9.4. Synthetic blood penetration cites YY/T 0691-2008. The 2011 clauses 5.11 to 5.13 are merged into 2023 clause 5.11 "Biocompatibility".
  • Not new: PFE (4.6.2/5.6.2) and packaging (7.1) were already in the 2011 edition.
  • Old reports: sort each item into "essentially unchanged", "method or acceptance criterion changed" or "not covered by the old report", and retest the last two groups to the 2023 edition. Change registration questions are for the drug regulatory authorities.
  • At SUNGO: the Shanghai CNAS and CMA scopes list both editions. 2023-edition PFE can be reported under CNAS and CMA, not IAS. Biocompatibility is not accredited and can only be subcontracted.

YY 0469-2023 Effective Date and Scope

The National Public Service Platform for Standards Information lists YY 0469-2023 "Surgical mask" as a mandatory pharmaceutical industry standard under the NMPA. It was published on 2023-11-22, applies from 2026-12-01 and replaces YY 0469-2011 in full. The cover of the standard gives the same dates and names the National Medical Products Administration as publisher. YY 0469-2011 was published on 31 December 2011 and applied from 1 June 2013.

The scope is revised. The 2011 edition covered single-use masks worn by clinical staff during invasive procedures. The 2023 edition covers single-use medical masks worn by healthcare workers in medical settings with a potential risk of contamination by blood, body fluids or secretions, and during invasive procedures. It states that it does not apply to protective masks for medical or other work settings.

Both editions have Clause 4 requirements, Clause 5 test methods, Clause 6 marking and Clause 7 packaging, transport and storage, with Clauses 4 and 5 numbered essentially in parallel (except the biological tests). SUNGO's scopes list items by Clause 5 numbers, so the comparison uses them too. The 2023 foreword lists eight main technical changes.

YY 0469-2023 vs 2011: Clause-by-Clause Comparison

Checked on 13 November 2026 against the YY 0469-2023 and YY 0469-2011 texts published by the NMPA Center for Medical Device Evaluation, and the scopes of SUNGO Shanghai CNAS L22083, CMA 240015340087 and IAS TL-1019. CMA entries carry no clause numbers and are matched by item name.

Clause2011 item2023 itemType of changeSUNGO, 2011 editionSUNGO, 2023 editionIAS (no year stated)
5.1AppearanceAppearanceEssentially unchangedCNAS (marked "for designated clients only"), CMACNAS, CMAIn scope
5.2Structure and dimensionsStructure and dimensionsRequirement changedCNAS, CMACNAS, CMAIn scope
5.3Nose clipNose clipRequirement changedCNAS, CMACNAS, CMAIn scope
5.4Mask strapsMask straps (5.4.1 ear-loop, 5.4.2 tie-on)Requirement and method changedCNAS, CMACNAS, CMAIn scope
5.5Synthetic blood penetration testResistance to penetration by synthetic bloodRenamed; method cites YY/T 0691-2008CNAS, CMACNAS, CMAIn scope
5.6.1BFE, Annex BBFE, Annex BRequirement raised; Annex B revisedCNAS, CMACNAS, CMAIn scope
5.6.2PFEPFERequirement raised; conditions essentially the sameNot on any scope lineCNAS, CMAExcluded
5.7Pressure differentialAirflow resistanceRenamed; requirement and method changedCNAS, CMACNAS, CMAIn scope
5.8Flame retardancyFlame retardancyEssentially unchangedCNAS, CMACNAS, CMAIn scope
5.9Microbiological indices5.9.1 Microbial limits; 5.9.2 SterilitySplit; methods moved to ChPCNAS, CMACNAS, CMA (plus ChP 1105 and 1101, "only when cited by YY 0469-2023")In scope
5.10EO residueEO residueMethod now GB/T 14233.1-2022 clause 9.4CNAS, CMACNAS, CMAIn scope
5.11 to 5.13 (2011) / 5.11 (2023)Skin irritation, cytotoxicity, delayed-type hypersensitivityBiocompatibility (5.11.1 to 5.11.4)Merged; requirements and methods changedNot accredited; subcontractedNot accredited; subcontractedExcluded
6MarkingMarkingRequirement changedCNAS, CMACNAS, CMAIn scope
7.1PackagingPackagingSimplified ("packed by quantity" removed)Not on any scope lineCNAS, CMAIn scope

IAS TL-1019 lists "YY 0469" with no year, excluding 5.6.2, 5.11, 5.12 and 5.13. For an IAS report to either edition, confirm the edition with the laboratory first.

BFE and PFE Requirements: BFE Up to 98%, PFE Up to 80%

BFE (5.6.1, Annex B): the requirement rises from at least 95% to at least 98%. Annex B still uses a Staphylococcus aureus (ATCC 6538) aerosol, a six-stage sampler and 28.3 L/min sampling flow. Two things changed. The positive control count moves from (2200 ± 500) CFU to 1700 CFU to 3000 CFU. A formula for mean particle size (MPS) is added in B.5.1, with MPS still (3.0 ± 0.3) μm. The 2023 text also says the inner side of the mask faces the aerosol.

PFE (5.6.2): already in the 2011 edition at a minimum of 30%; the 2023 edition requires at least 80%. Test conditions are essentially the same. Per the 2023 text:

  • Conditioning: (38 ± 2.5) °C and (85 ± 5)% RH for at least (25 ± 1) h, then sealed in an airtight container and tested within 10 h.
  • Test environment: (25 ± 5) °C and (30 ± 10)% RH.
  • Aerosol: sodium chloride, not more than 200 mg/m³, count median diameter (0.075 ± 0.020) μm, geometric standard deviation not more than 1.86.
  • Flow and area: (30 ± 2) L/min through 100 cm².

Old PFE data can be compared at the method level, but the pass line has moved from 30% to 80%. We recommend retesting to the 2023 edition and judging against the new requirement.

Airflow Resistance Replaces Pressure Differential: Less Than 60 Pa

Clauses 4.7 and 5.7 are retitled "airflow resistance", newly defined as the resistance of the mask over a specified area at a specified flow, expressed as pressure difference per unit area.

  • Requirement: not more than 49 Pa in 2011; less than 60 Pa in 2023.
  • Method: both editions use a 25 mm diameter, 4.9 cm² test area and ΔP = M/4.9. The 2023 edition sets the flow at (8.0 ± 0.2) L/min and uses three non-overlapping test areas at the centre of the mask (the mouth and nose area when worn). A specimen passes if the average of its three areas meets the requirement. At least five specimens are tested.

The limit is looser but the test points and calculation changed, so a 2011 pressure differential result is not a 2023 airflow resistance result. We recommend retesting.

Microbial Limits and Sterility: ChP General Chapters 1105 and 1101

In 2011, clause 5.9 had two routes. Non-sterile masks were tested to GB 15979-2002 Annex B against Table 1 (total bacterial count and several pathogens). Masks labelled sterilized or sterile had a sterility test to GB/T 14233.2-2005 Chapter 2. The 2023 edition splits this into two clauses:

  • 5.9.1 Microbial limits (products supplied non-sterile): membrane filtration to ChP General Chapter 1105, adding the total aerobic count and the total yeasts and moulds count; the total must not exceed 30 CFU/g. At least two smallest packages are sampled, not less than 10 g of product is taken, cut up and made into a 1:50 (g/mL) test solution with the specified diluent.
  • 5.9.2 Sterility (products supplied sterile): the mask must have been through a validated sterilization process and be sterile. The test follows ChP General Chapter 1101, using the direct inoculation provisions for dressing-type samples.

For how GB 15979 Annex B and ChP 1105 differ, see Non-Sterile Mask Microbial Limits: GB 15979 Annex B vs ChP Chapter 1105.

Synthetic Blood Penetration and EO Residue: Method Changes

Synthetic blood (5.5): the requirement is unchanged. After 2 mL of synthetic blood is sprayed at the outer surface at 16.0 kPa (120 mmHg), no penetration may appear on the inner surface. The 2011 edition set out the apparatus and procedure itself. The 2023 edition refers conditioning and testing to YY/T 0691-2008, a Chinese test method for medical face mask resistance to synthetic blood (fixed volume, horizontal projection). Synthetic blood is still made to Annex A, which now allows a surfactant equivalent to Tween 20. For how YY/T 0691-2008 relates to ISO 22609, refer to the foreword of YY/T 0691-2008. On the ISO 22609 method, see ISO 22609 Synthetic Blood Test for Masks: YY 0469 and EN 14683 Type IIR.

EO residue (5.10): still not more than 10 μg/g. The 2023 requirement title (4.10) adds "if applicable", and coverage widens from masks sterilized with ethylene oxide to masks sterilized or disinfected with it. The method moves from the GC method of GB/T 14233.1-2008 to GB/T 14233.1-2022 clause 9.4.

Changes to Straps, Nose Clip, Marking, Packaging and Biocompatibility

  • Structure and dimensions: dimensions follow the manufacturer's design dimensions and tolerances, and the inner and outer sides must be distinguishable.
  • Nose clip: an alternative design is allowed. The nose clip must be secure, without sharp protrusions, and at least 8.0 cm long; an alternative must hold the mask and give a basic seal.
  • Mask straps: 2011 required at least 10 N at each strap-to-body joint. In 2023, ear-loop masks take a 15 N static pull for 5 s across the strap and both attachment points, and tie-on masks take 10 N for 5 s, with a procedure for straps not fixed to the body.
  • Marking: "smallest package" becomes "smallest sales unit"; "size and tolerance" becomes "model, specification and size information"; "service life or expiry date" is added (2011 only required the sterilization expiry date for sterile products); "instructions for use" becomes "inner/outer side identification and how to wear"; "standard number" becomes "standard number or product technical requirement number".
  • Packaging (7.1): both editions require protection against mechanical damage and contamination before use; 2023 drops packing by quantity.
  • Biocompatibility (5.11): the three 2011 tests are merged. Strap and body materials are combined at a 1:3 area ratio and extracted per GB/T 16886.12, and cytotoxicity uses the MTT method of GB/T 16886.5. These tests are outside SUNGO's accreditation.

YY 0469-2011 Test Reports: What to Retest for the 2023 Edition

Check each item of the old report against the 2023 text and sort it into three groups:

  1. Essentially unchanged: appearance (5.1) and flame retardancy (5.8). They can be assessed line by line; whether old data can be reused is for reviewers and the drug regulatory authorities.
  2. Method or acceptance criterion changed: structure and dimensions, nose clip, straps, synthetic blood, BFE, PFE, airflow resistance, microbiology, EO residue, biocompatibility and marking. Retest to the 2023 edition, even where an old figure looks good enough (for example BFE above 98%), because the Annex B positive control range and specimen orientation changed.
  3. Not covered by the old report: PFE or packaging where the old report has no result, plus new 2023 requirements such as distinguishable sides, the alternative nose clip, the 15 N ear-loop test and the service life or expiry date marking.

Whether a registered product needs a change registration, when technical requirements must be updated, and from when reports must follow the 2023 edition are not addressed here. The NMPA and provincial drug regulatory authorities decide those.

YY 0469-2023 and YY/T 0969-2023: How They Relate

Both were published on 2023-11-22 but cover different products. YY/T 0969-2023 "Single-use medical face mask" is a recommended standard that has applied since 1 December 2025, replacing YY/T 0969-2013. It covers single-use masks for general medical settings and does not apply to medical protective masks or surgical masks. A surgical mask is tested to YY 0469, a single-use medical mask to YY/T 0969, and one cannot replace the other. See YY/T 0969-2023 Single-Use Medical Masks: Changes from 2013 and Testing Scope.

SUNGO Accreditation Scope for Both YY 0469 Editions

ItemClause basisNotes / limitations
AppearanceYY 0469-2023 5.1; YY 0469-2011 5.1CNAS and CMA, both editions; 2011 clause 5.1 marked "for designated clients only" on the CNAS scope
Structure and dimensions; nose clip; mask straps; flame retardancyYY 0469-2023 and YY 0469-2011, 5.2, 5.3, 5.4, 5.8CNAS and CMA, both editions
Bacterial filtration efficiency (BFE)YY 0469-2023 and YY 0469-2011, 5.6.1, Annex BCNAS and CMA, both editions
MarkingYY 0469-2023 and YY 0469-2011, Clause 6CNAS and CMA, both editions
Resistance to penetration by synthetic bloodYY 0469-2023 5.5CNAS, CMA; the 2011 line is "synthetic blood penetration test" 5.5
Particle filtration efficiency (PFE)YY 0469-2023 5.6.2CNAS, CMA; no 2011 line; IAS excludes 5.6.2
Airflow resistanceYY 0469-2023 5.7CNAS, CMA; the 2011 line is "pressure differential" 5.7
Microbial limitsYY 0469-2023 5.9.1; ChP 2020 edition Volume IV, General Chapter 1105CNAS, CMA; ChP line "only when cited by YY 0469-2023"
SterilityYY 0469-2023 5.9.2; ChP 2020 edition Volume IV, General Chapter 1101CNAS, CMA; ChP line "only when cited by YY 0469-2023"
Microbiological indicesYY 0469-2011 5.9CNAS, CMA
EO residueYY 0469-2023 5.10; YY 0469-2011 5.10; GB/T 14233.1-2022 Clause 9 and 9.4CNAS, CMA
PackagingYY 0469-2023 7.1CNAS, CMA; no 2011 line
YY 0469 (whole standard, no year stated)Except 5.6.2, 5.11, 5.12, 5.13IAS TL-1019; confirm the edition with the laboratory first
BiocompatibilityYY 0469-2023 5.11; YY 0469-2011 5.11 to 5.13Not on any scope; subcontracted (see below)

Where SUNGO Tests YY 0469 and Under Which Accreditation

Testing is done by Shanghai SUNGO Medical Technology Co., Ltd., at 958 Kangqiao East Road, Pudong New Area, Shanghai, under CNAS L22083, CMA 240015340087 and IAS TL-1019. Hefei SUNGO Medical Technology Co., Ltd. has no YY 0469 scope. SUNGO is accredited by CNAS and IAS, both signatories to the ILAC MRA.

  • CNAS: 11 lines for 2011 and 14 for 2023, plus ChP 1105 and 1101 limited to use when cited by YY 0469-2023. 2011-edition PFE and packaging are not on the scope.
  • CMA: 11 items for 2011 and 14 for 2023, matching CNAS by name. CMA covers only scheduled items and is valid only within the People's Republic of China. A 2023-edition report can carry the CMA mark for those 14 items, not for biocompatibility.
  • IAS: excludes 5.6.2, 5.11, 5.12 and 5.13, so 2023-edition PFE can be reported under CNAS and CMA but not IAS.
  • Biocompatibility: not on any SUNGO accreditation. It can be subcontracted to a laboratory with the appropriate qualifications, with the client's prior consent, and the report names the subcontracted items and the subcontractor.

See also Medical Mask Testing.

Preparing Samples for YY 0469-2023 Testing

  • State the edition (2023 or 2011) and the mark needed (CNAS, CMA or IAS). For IAS, confirm the edition first; PFE and biocompatibility are outside IAS.
  • State how the product is supplied: sterile products need 5.9.2, non-sterile products 5.9.1. EO-sterilized or EO-disinfected products also need 5.10; include the process details.
  • State the strap type and nose clip design, and provide the manufacturer's design dimensions and tolerances for 5.2 to 5.4.
  • Provide the smallest-sales-unit label, packaging and instructions for Clause 6 and 7.1.
  • Send finished products from one batch in intact packaging. Specimen counts are set clause by clause in Clause 5; confirm quantities with the laboratory.
  • Send any 2011-edition report so a retest list can be agreed.

FAQ

When does YY 0469-2023 take effect?

It was published on 22 November 2023 and applies from 1 December 2026, replacing YY 0469-2011 in full as a mandatory industry standard.

What are the main differences between YY 0469-2023 and 2011?

BFE 95% to 98%, PFE 30% to 80%, pressure differential replaced by airflow resistance (less than 60 Pa, three points per specimen), microbiology to ChP 1105 and 1101, ear-loop and tie-on straps tested separately, biological tests merged into 5.11, and revised marking items.

Is PFE new in YY 0469-2023?

No. The 2011 edition already had PFE at a minimum of 30%; 2023 requires at least 80%. SUNGO's PFE accreditation is on the 2023 lines only (CNAS, CMA), and IAS excludes 5.6.2.

Can a YY 0469-2011 report still be used after 1 December 2026?

Technically only appearance and flame retardancy are essentially unchanged; we recommend retesting the rest to the 2023 edition. Acceptance in a registration file is for the drug regulatory authorities to decide.

Can a YY 0469-2023 report carry the CMA mark?

Yes for the 14 YY 0469-2023 items in SUNGO Shanghai's CMA schedule, not for biocompatibility. CMA is valid only within the People's Republic of China.

To arrange YY 0469-2023 testing or a line-by-line review of a 2011-edition report, please use our contact page and tell us the product type, sterilization method, strap type and the mark you need. The laboratory will confirm sample requirements and quote by item.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.