ISO 22609 Synthetic Blood Test for Masks: YY 0469 and EN 14683 Type IIR

ISO 22609 Synthetic Blood Test for Masks: YY 0469 and EN 14683 Type IIR

In China, synthetic blood penetration of a surgical mask is a YY 0469 requirement; in the EU, the same property is splash resistance under EN 14683. EN 14683:2025 and YY 0469-2023 both use ISO 22609:2004 (adopted identically in China as YY/T 0691-2008), while YY 0469-2011 clause 5.5 writes out its own procedure and cites neither. A shared method origin does not make the reports interchangeable: YY 0469 and EN 14683 each decide which masks are tested, how many specimens are used and what counts as a pass. This article explains the method, how each product standard applies it, and the matching scope of SUNGO's Shanghai laboratory.

Key points first

  • ISO 22609:2004 is a test method. 2 mL of synthetic blood is projected horizontally from a fixed distance at the target area of a complete mask. The face side is then inspected, and each specimen is reported as "pass" or "fail".
  • China: every surgical mask within the scope of YY 0469 is tested. Clause 4.5 is the same in YY 0469-2011 and YY 0469-2023: after 2 mL of synthetic blood is sprayed onto the outer side at 16.0 kPa (120 mmHg), no penetration shall appear on the inner side. Clause 5.5 of the 2023 edition requires at least 3 specimens, tested in accordance with YY/T 0691-2008; the 2011 edition uses 3 specimens and writes the procedure into the clause.
  • EU: only Type IIR masks are tested. Table 1 of EN 14683:2025 gives "Not required" for Type I and Type II and ≥ 16,0 kPa for Type IIR. Clause 5.2.4 requires the targeting plate for all tests, and clause 5.2.7 requires at least 29 of 32 specimens to pass.
  • The specimen plan is the biggest difference. Both standards use 16 kPa, but YY 0469-2023 says "at least 3 specimens" (3 in the 2011 edition) while EN 14683:2025 uses 29 out of 32. If one lot needs both reports, plan specimens for each.
  • SUNGO Shanghai scope. YY 0469-2011 and YY 0469-2023 clause 5.5 are on the CNAS schedule, and the CMA schedule lists the same items (no clause given). EN 14683:2025 and BS EN 14683:2025 clause 5.2.4, ISO 22609:2004 (no clause given), ASTM F1862/F1862M-24 (no clause given), and ASTM F2100-25 and F2100-23 clause 9.4 (the latter restricted to specific clients) are on the CNAS schedule but not on the CMA schedule. IAS covers these international standards and YY 0469 (with four clauses excluded). YY/T 0691 is not listed on any certificate.

The standards and their scope

ISO 22609:2004: the method itself

The full title is Clothing for protection against infectious agents — Medical face masks — Test method for resistance against penetration by synthetic blood (fixed volume, horizontally projected). It is the first edition, dated 2004-12-01, prepared by ISO/TC 94/SC 13; the foreword states that it is based on ASTM F1862-00a.

Clause 1 describes it as a laboratory method for the resistance of medical face masks to penetration by a splash of synthetic blood. It primarily addresses the mask materials or material constructions, not design, interfaces, filtration efficiency or pressure drop, nor protection against airborne or aerosolized exposure. The introduction says it models a fixed volume of high-velocity blood from a punctured vessel hitting the mask over 0 s to 2,5 s.

Three standard conditions are defined: stream velocities corresponding to blood pressures of 10,6, 16,0 and 21,3 kPa (80, 120 and 160 mmHg). Non-standard pressures, volumes and orientations are permitted for specific applications. A mask is rated at the highest pressure at which it demonstrates an acceptable quality limit (AQL) of 4,0 %.

YY/T 0691-2008: the Chinese identical adoption

The cover of YY/T 0691-2008 is marked "ISO 22609:2004, IDT". Its foreword states identical adoption and lists a few editorial changes, such as deleting the ISO foreword. It was issued by the former State Food and Drug Administration on 2008-10-17 and took effect on 2010-01-01. Mind the numbering: synthetic blood preparation is in Annex A (normative) of YY/T 0691 (Annex B in ISO 22609), and YY/T 0691 titles clause 7.3 "test using a targeting plate".

YY 0469: the Chinese surgical mask product standard

YY 0469-2023 Surgical mask was issued on 2023-11-22, takes effect on 2026-12-01 and replaces YY 0469-2011 in full. It covers single-use masks worn by medical staff where blood, body fluid or secretion contamination is possible and during invasive procedures, but not protective respirators. The editions compare as follows:

  • Requirement (clause 4.5). The wording is the same in both editions: 16.0 kPa, and no penetration on the inner side. The 2023 foreword lists a changed clause title: "synthetic blood penetration" in 2011 became "resistance to synthetic blood penetration".
  • Test method (clause 5.5). The 2011 edition writes the specimen number (3), pre-treatment, spraying distance, pressure and visual inspection into the clause, and its normative references include neither ISO 22609 nor YY/T 0691. Its distance, 30.5 cm from the specimen centre, is worded differently from the (300 ± 10) mm of ISO 22609. Its synthetic blood definition (3.2) cites ASTM F1862-00a, on which ISO 22609:2004 is also based, so the methods share an origin. The 2023 edition says to test at least 3 specimens with pre-treatment and method per YY/T 0691-2008, which it adds to its normative references.
  • Synthetic blood. Both editions prepare it according to their own Annex A. The 2023 foreword notes that Annex A now also allows other surfactants equivalent to Tween 20.

EN 14683:2025: the European medical face mask standard

EN 14683:2025 was approved by CEN on 2024-12-29 and supersedes EN 14683:2019+AC:2019. Clause 4 classifies masks as Type I or Type II by bacterial filtration efficiency; a splash-resistant Type II mask is marked "IIR", with "R" signifying splash resistance. Definition 3.13 describes splash resistance as the ability of a medical face mask to withstand penetration of synthetic blood projected at a given pressure.

Clause 5.2.4 requires that, tested to ISO 22609:2004, the mask meets the Type IIR minimum in Table 1 (≥ 16,0 kPa), and that all tests are undertaken with the targeting plate. Clause 5.2.7 requires every specimen to meet Table 1, except that for splash resistance at least 29 of 32 specimens shall pass. Neither the targeting-plate sentence nor the 29-of-32 wording appears in EN 14683:2019+AC:2019. Clause 5.2.1 requires all tests on finished products or samples cut from them.

How the three documents divide the work

PointYY 0469-2023EN 14683:2025ISO 22609:2004 (method)
Which masks are testedSurgical masks within the standard's scope; clause 4.5 is a requirementType IIR only; Not required for Type I and Type IIA method standard; does not say which products are tested
Test pressure16.0 kPa (120 mmHg)Type IIR ≥ 16,0 kPaThree standard pressures, 10,6, 16,0 and 21,3 kPa; non-standard pressures permitted
Method referenceClause 5.5: YY/T 0691-2008Clause 5.2.4: ISO 22609:2004—
Specimens and pass ruleAt least 3 specimens, judged against clause 4.5At least 29 of 32 specimens passAQL 4,0 % at each selected pressure; a single sampling plan uses 32 specimens, and 29 or more passes meet it
Targeting plateNot specified in YY 0469; in YY/T 0691-2008 it is a recommended addition to the apparatus, and when used the test pressure is set per clause 7.3 (test using a targeting plate)Required for all testsRecommended addition; when used, the test pressure is set per clause 7.3 (alternative set-up)
Synthetic bloodPrepared per Annex A of YY 0469-2023Tested to ISO 22609:2004, whose clause 5.2.1 points to its Annex BClause 5.2.1: prepared per Annex B (normative); Annex A in YY/T 0691-2008

How the ISO 22609 test is run

Apparatus and delivery

The apparatus comprises a specimen-holding fixture, fluid reservoir, pneumatic-controlled valve and valve controller. The convex fixture holds the mask gently stretched, with the valve (300 ± 10) mm from the target area. The canula is 12,7 mm long with a 0,84 mm inside diameter. Each spurt delivers 2 mL, checked with a graduated cylinder or balance and re-verified after every 16 specimens. The valve time is set from Table 1 of the standard for the selected pressure.

The targeting plate has a 0,5 cm hole and sits about 1 cm in front of the mask. The standard explains that the initial portion of the stream can reach two to three times the target pressure; the plate blocks it so only the steady-state stream hits the mask, making the impact velocity more accurate and repeatable. Clause 8 requires the report to state whether the plate was used, and for an EN 14683:2025 report the answer should be yes.

Conditioning and test environment

  • Each specimen is conditioned for at least 4 h at (21 ± 5) °C and (85 ± 5) % relative humidity. This simulates the humidity a wearer creates by breathing through the mask.
  • Testing is done at (21 ± 5) °C and (85 ± 10) % relative humidity.
  • Each specimen is tested within 60 s of removal from the conditioning chamber.

Where warranted, extra pre-treatments such as pre-wetting may be used to assess storage, shelf life or in-use degradation. Any pre-treatment is described in the report.

Specimens and target area

  • Complete masks are used as specimens, taken at random for each type, design or lot.
  • Areas of different material or thickness are tested separately, as are seams claimed to protect as well as the base material.
  • Pleats are spread out so the target area is a single layer; the specimen centre is the target area.

Visual inspection of the inner surface

  1. Before testing, place about 0,1 mL of synthetic blood on the inside of a spare mask to confirm that penetrating fluid will be visible. If it is not, talcum powder can be applied to the inside surface.
  2. (10 ± 1) s after the spurt, inspect the viewing side (the side that faces the wearer) under suitable lighting for synthetic blood or other evidence of wetness.
  3. If there is doubt about visible penetration, lightly daub the target area with a cotton absorbent swab.
  4. Any evidence of synthetic blood penetration on the side that contacts the wearer's face is a fail for that specimen.

AQL and report content

ISO 22609 asks for enough specimens to reach an AQL of 4,0 % (ISO 2859-1) at each selected pressure. A single sampling plan needs 32 specimens, and the AQL is met when 29 or more pass. The report gives pass or fail for each specimen at each pressure, the highest pressure meeting the 4,0 % AQL, conditioning and test climate, pre-treatment, target area, visibility aid and whether the targeting plate was used.

The product standards narrow this down: EN 14683:2025 fixes 16,0 kPa for Type IIR with the 29-of-32 rule, while YY 0469-2023 clause 5.5 sets at least 3 specimens judged against clause 4.5. The bench procedure is the same; the specimen plan and conclusion are not.

Relationship with ASTM F1862

ISO 22609:2004 is based on ASTM F1862-00a. F1862 has since been revised, and ASTM's website lists F1862/F1862M-24 as active; the US specification ASTM F2100 lists it as a referenced document. The scopes differ: ISO 22609 primarily addresses mask materials or material constructions, while clause 1.3 of F1862/F1862M-24 says it is primarily intended for finished masks and that results can differ between finished masks and materials taken from them. F1862/F1862M also gives SI and inch-pound values. US pressure levels and pass rules are outside this article.

SUNGO accreditation scope

The table reproduces the relevant rows from the CNAS, CMA and IAS scopes of Shanghai SUNGO Medical Technology Co., Ltd. On the CNAS schedule the two YY 0469 rows fall under "surgical mask" and the rest under "medical mask". All are effective from 2026-03-09; the ASTM F2100-23 clause 9.4 row is marked "restricted to specific clients", and the other notes fields are empty.

ItemStandard and clauseNotes / limitations
Synthetic blood penetration testYY 0469-2011 5.5CNAS; the CMA schedule lists the same item (no clause given)
Resistance to synthetic blood penetrationYY 0469-2023 5.5CNAS; the CMA schedule lists the same item (no clause given)
Splash resistanceEN 14683:2025 5.2.4CNAS; not on the CMA schedule
Splash resistanceBS EN 14683:2025 5.2.4CNAS; not on the CMA schedule
Splash resistanceISO 22609:2004 (no clause given on the schedule)CNAS; not on the CMA schedule
Splash resistanceASTM F1862/F1862M-24 (no clause given on the schedule)CNAS; not on the CMA schedule
Splash resistanceASTM F2100-25 9.4CNAS; not on the CMA schedule
Splash resistanceASTM F2100-23 9.4CNAS; restricted to specific clients; not on the CMA schedule
Whole standardISO 22609; EN 14683; BS EN 14683; ASTM F1862/F1862M; ASTM F2100IAS (TL-1019); scope does not state edition years
PartialYY 0469IAS (TL-1019); Except for 5.6.2, 5.11, 5.12, 5.13; clause 5.5 is not excluded; scope does not state edition years

Where the work is done and under which accreditation

All items in this article are tested at the Shanghai laboratory, 958 Kangqiao East Road, Pudong, which belongs to Shanghai SUNGO Medical Technology Co., Ltd., the company that issues the reports.

  • CNAS laboratory accreditation, CNAS L22083. All eight CNAS rows above are on the schedule. The CNAS certificate states that CNAS is a signatory to the ILAC and APAC mutual recognition arrangements.
  • CMA inspection body qualification, certificate 240015340087. The schedule lists the two YY 0469 items above; ISO 22609, EN 14683, BS EN 14683, ASTM F1862 and ASTM F2100 are not on it, so no CMA report can be issued for them. The CMA mark applies only to scheduled items and, as the certificate states, is valid within the People's Republic of China.
  • IAS accreditation, TL-1019. The scope lists ISO 22609, EN 14683, BS EN 14683, ASTM F1862/F1862M and ASTM F2100 as whole standards, and YY 0469 except clauses 5.6.2, 5.11, 5.12 and 5.13, all without edition years. Confirm the edition with the laboratory before an IAS report is issued. The IAS certificate carries the ilac-MRA mark.

A note on YY/T 0691-2008. YY 0469-2023 clause 5.5 refers to it, but it is not listed separately on any SUNGO certificate. Reports are issued against YY 0469 clause 5.5 (CNAS, CMA, IAS) or against ISO 22609 (CNAS, IAS), not against YY/T 0691 as a separate accredited item.

SUNGO's Hefei laboratory belongs to a separate legal entity, Hefei SUNGO Medical Technology Co., Ltd., whose scope does not include medical masks. Whether a regulator, notified body or customer accepts a given report depends on that party's requirements.

Preparing a submission

  1. State the standard and purpose. For Chinese registration testing, state the YY 0469 edition; YY 0469-2023 takes effect on 2026-12-01, and the edition follows the registration requirements. For the EU, state EN 14683:2025 or BS EN 14683:2025 and whether the product is Type IIR. Also state the report mark you need: CMA applies only to the two YY 0469 items in the table, so EN 14683, BS EN 14683, ISO 22609 and ASTM reports can be CNAS or IAS only.
  2. Send complete finished masks. ISO 22609 tests complete masks, and clause 5.2.1 of EN 14683:2025 requires finished products or samples cut from them. Take the samples from a single lot and keep them in their original packaging.
  3. Confirm quantities before ordering. EN 14683 Type IIR splash resistance is judged on 29 of 32 specimens, plus spare masks for the visibility check; YY 0469-2023 clause 5.5 needs at least 3 (3 in the 2011 edition). Samples for other items are counted separately, and the certificate scopes above decide which items SUNGO covers. The laboratory calculates the numbers from the test plan.
  4. Mark the inner and outer sides. The result is read on the face side. YY 0469-2023 clause 4.2 requires the inner and outer sides to be distinguishable, and EN 14683:2025 clause 7 requires the packaging to show which side contacts the face.
  5. Describe the construction. Tell us in advance about areas with different materials or thicknesses and seams claimed to protect as well as the base material, since these are tested separately. Also mention an inner layer whose colour could make synthetic blood harder to see.
  6. Specify any special pre-treatment. If testing follows ageing, pre-wetting or another treatment, give the conditions when ordering. Non-standard conditions are recorded in the report.

Frequently asked questions

Are 16 kPa and 120 mmHg two different requirements?

No. They are the same pressure in two units. YY 0469 writes 16.0 kPa (120 mmHg), ISO 22609 pairs 16,0 kPa with 120 mmHg, and Table 1 of EN 14683:2025 uses kPa only.

Do EN 14683 Type I and Type II masks need splash resistance testing?

No. Table 1 of EN 14683:2025 shows "Not required" for both. Only Type IIR must reach ≥ 16,0 kPa.

Can a YY 0469 synthetic blood result be used as an EN 14683 Type IIR result?

The methods share the same origin, but the specimen number and pass rule differ: at least 3 specimens under YY 0469-2023 (3 under the 2011 edition), versus 29 of 32 with the targeting plate under EN 14683:2025. Reuse depends on checking the raw records clause by clause, and the final decision rests with the notified body or customer.

Can the report cite YY/T 0691-2008 as the basis?

YY/T 0691 is not listed separately on SUNGO's certificates. Reports are issued against YY 0469 clause 5.5 (CNAS, CMA, IAS) or ISO 22609 (CNAS, IAS). Clause 5.5 of YY 0469-2023 itself refers to YY/T 0691-2008.

Can an ISO 22609 or EN 14683 report carry the CMA mark?

No. These two standards, ASTM F1862 and ASTM F2100 are not on the Shanghai CMA schedule. The CMA mark covers only scheduled items and is valid only within China.

Requesting a quote

For synthetic blood penetration testing to YY 0469, EN 14683 or ISO 22609, or to check whether an item can be reported under CNAS, CMA or IAS, see our medical mask testing service. To request a quote, use the contact page and send the mask construction, target market, standard and edition, and the report mark you need. We reply item by item against the standard clauses and our certificate scope.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.