Makers of medical absorbent cotton wool, cotton balls and swabs usually ask three questions when drafting a product technical requirement or preparing for China registration testing: what does YY/T 0330-2015 test, why do sinking time and water absorbency fail so often, and which accreditation mark can SUNGO put on the report. This article walks through each item, explains how results are judged, and maps every item to the CNAS and CMA scopes of Shanghai SUNGO Medical Technology Co., Ltd. Limits and test parameter values are not quoted; the standard prevails.
Key points first
- The standard. YY/T 0330-2015, Medical absorbent cotton, is a recommended (voluntary) pharmaceutical industry standard of China. It was issued on 2015-03-02, took effect on 2016-01-01 and fully replaced YY 0330-2002. As of 2026-10-22, the national standards information platform lists it as current.
- SUNGO's coverage. The CNAS scope of SUNGO Shanghai (L22083) lists 12 items of this standard (see the scope table below), including ethylene oxide (EO) residue, for which GB/T 14233.1-2022 Clause 9.4 is also listed. The CMA schedule (240015340087) lists 9 of these items; characters, foreign fibres and neps are not on it. The IAS scope (TL-1019) does not include YY/T 0330.
- Report marks. Characters, foreign fibres and neps can only be reported under CNAS. No item of this standard can be reported under IAS. The CMA mark applies only to items in the CMA schedule and is valid only within the People's Republic of China.
- Absorbency. Sinking time and water absorbency (Clause 4.6) are sensitive to sampling, packing density, water temperature and timing; on the product side, check first whether scouring and bleaching were sufficient, whether processing aids remain, and how tightly the cotton is compressed.
- A new national standard. GB/T 46936-2025 Medical absorbent cotton takes effect on 2027-01-01; the platform does not say it replaces YY/T 0330-2015, and it is not within SUNGO's accreditation scope.
The standard and its scope
Sources: the national standards information platform and the public preview of the original text, checked on 2026-10-22.
| Item | Details |
|---|---|
| Number and title | YY/T 0330-2015, Medical absorbent cotton (Chinese title: 医用脱脂棉) |
| Type | Recommended pharmaceutical industry standard; competent authority NMPA; issued by the former China Food and Drug Administration (cover) |
| Dates | Issued 2015-03-02, effective 2016-01-01 |
| Replaces | YY 0330-2002 in full (a mandatory industry standard); the foreword says the standard was first issued in April 2002 and the 2015 edition is its first revision |
Scope and what the introduction says
Clause 1 says the standard applies to medical absorbent cotton made from the hairs on the ripe seeds of plants of the genus Gossypium (Malvaceae), with impurities removed and then scoured, bleached and processed, and containing no coloured additives of any kind. Medical absorbent cotton is mainly used as a dressing in hospitals. The introduction adds a few points that matter for testing:
- Medical absorbent cotton is supplied either sterile or non-sterile.
- Absorbent cotton is highly adsorptive, and adsorbed EO would be a significant hazard to patients and healthcare staff, so the standard does not recommend EO sterilization.
- For non-sterile supply, the microbial limit provided by the manufacturer can serve as the basis on which hospitals set sterilization parameters.
- Absorbent cotton that meets this standard still does not meet the performance requirements of neurosurgical cotton patties (脑棉片).
What about cotton balls and swabs?
The scope covers medical absorbent cotton as such. Whether a cotton ball, swab or pad cites YY/T 0330-2015, and which clauses, depends on its product technical requirement. In SUNGO's scope the test object for this group is medical absorbent cotton; properties specific to the finished article, such as how a swab tip is fixed to its stick or its dimensions, are outside YY/T 0330 and outside this group of accredited items. State whether we test the finished article or the raw cotton, and how to sample it. Whether a report on finished balls or swabs can carry the CNAS or CMA mark is confirmed by the laboratory when the order is accepted.
What changed from the 2002 edition
The foreword lists the main changes: whiteness and readily oxidizable substances deleted; "absorption time" replaced by "sinking time"; identification, foreign fibres, neps and extractable colouring matter added; for non-sterile cotton, a stated microbial limit (3.16) and its test method (4.14) added; acidity or alkalinity and surface-active substances revised; residue on ignition changed to sulphated ash; inspection rules and marking and labelling removed. The standard refers to the Absorbent Cotton monograph of the British Pharmacopoeia 2010.
Clause 2 lists GB/T 14233.1, GB/T 19973.1, YY/T 0615.1 and the British Pharmacopoeia as normative references, all undated. Under that clause, the latest edition of an undated reference (including all amendments) applies. According to the platform, the current editions are GB/T 14233.1-2022 (effective 2023-11-01) and GB/T 19973.1-2023 (effective 2024-10-01).
The new national standard GB/T 46936-2025
The platform lists GB/T 46936-2025 Medical absorbent cotton as a recommended national standard under working group SWG30, issued on 2025-12-31 and effective from 2027-01-01 (status as of 2026-10-22: "to be implemented"). It names YY/T 0330-2015 as a related standard but states no replacement relationship. GB/T 46936-2025 is not within SUNGO's accreditation scope, and we have not compared its content. Whether to cite it is for the registrant to assess against review requirements.
Test items and method notes
Clause 3 sets the requirements and Clause 4 the test methods. SUNGO's scope lists characters, foreign fibres and neps against Clause 3 (3.1, 3.3, 3.4), because their examination is written into those requirement clauses; all other items are listed against Clause 4.
Characters, foreign fibres and neps: look, compare, count
- Characters (3.1). Visual examination: the cotton should be white or almost white, the standard sets a minimum mean fibre length, and there should be no leaf, shell or seed-coat residue or other impurities. It should offer some resistance when pulled apart and shed no dust when gently shaken.
- Foreign fibres (3.3). Under the microscope, only typical cotton fibres should be present; an occasional isolated foreign fibre is permitted. Synthetic fibres or other non-cotton fibres picked up from other lines or packaging show up here.
- Neps (3.4). A specified quantity of sample is spread evenly in colourless transparent dishes and examined for neps by transmitted light, against the neps reference standard specified in the British Pharmacopoeia. The sample should have no more neps than the reference. The judgement is by comparison. Nep levels can relate to raw cotton maturity and to opening and carding conditions; treat these as lines of investigation.
Identification and acidity or alkalinity
Identification (4.3) was added in 2015 to confirm that the fibre is cotton; the reagents and expected observations are as given in the standard. Acidity or alkalinity (4.5) had both its requirement and method revised in 2015. It reflects acid or alkali left behind after wet processing such as bleaching, acid washing and neutralization, and whether final rinsing was sufficient is the first thing to investigate.
Sinking time and water absorbency: results are very sensitive to handling
Both items sit under Clause 4.6; the sub-clause numbers listed in SUNGO's scope are shown in the scope table below. Readers who know the British Pharmacopoeia method will recognise the approach, but procedure and judgement follow the YY/T 0330-2015 text. As the names suggest, sinking time shows how quickly the sample is wetted and submerged, and water absorbency shows how much water it holds afterwards. Small handling deviations show up directly in the reading.
| Variable | Why it affects the result | What the client should do |
|---|---|---|
| Sample quantity and location | Parts of a roll or bale can differ in loft and impurities; the quantity taken affects sinking speed and water held | Supply enough unopened samples from one batch; the laboratory samples per Clause 4.6 |
| Packing and sample holder | Packing density changes how fast water enters between fibres; the holder specification also matters | Do not compress or tease out samples; holder specification per the standard |
| Immersion and timing | Timing runs from entry into the water until the sample is judged fully submerged, so start and end points set the reading | Check that your in-house method matches the standard before comparing data |
| Water temperature | Water temperature affects wetting speed and water held | Control water temperature in in-house testing to the conditions in the standard |
| Sample conditioning and test environment | The moisture state of the fibres may affect wetting and water uptake | If the standard specifies pre-test conditioning or a test environment, follow it in in-house testing |
| Draining and weighing (absorbency) | Different draining times and methods leave different amounts of water in the sample | Follow the standard in in-house testing and record the draining conditions |
The specific values and procedures for each variable are as given in Clause 4.6 of YY/T 0330-2015.
Where to look when sinking time or absorbency fails
- Incomplete scouring and bleaching. Raw cotton carries natural waxes and other water-repellent substances that scouring and bleaching are meant to remove. With too little time, temperature or chemical, residues lengthen sinking time and lower absorbency, and acidity or alkalinity may also be off.
- Residual processing aids. Carding oils or softeners not removed by final rinsing leave a water-repellent layer and slow sinking. Some are themselves surface-active and may also fail the surface-active substances test (4.11).
- Compressed too tightly. Over-compressed bales or tightly wound balls and rolls close the spaces between fibres and increase sinking time. If the article is tightly formed by design, state the test object and sampling approach when ordering.
Before retesting, fix sampling and handling first, then review the process.
Fluorescence and extractable colouring matter: how they are judged
Fluorescence (4.8) addresses optical brighteners and similar substances: brighteners added to raise whiteness, or fluorescent fibres and particles picked up in production, show up here. The sample is observed for fluorescence under specified lighting; light conditions and the permitted appearance are as given in the standard. Chasing "whiter" cotton only raises the risk of failing this test.
Extractable colouring matter (4.12) was added in 2015 and matches the "no coloured additives" wording in the scope. The sample is extracted with a specified solvent and the extract colour is compared with a specified reference. Coloured stray fibres, coloured aids or recycled material can cause failures. Solvent, quantities and reference preparation are as given in the standard.
Surface-active substances
The requirement for surface-active substances (4.11) was revised in 2015. The test looks for residual surfactant-type substances, mostly from detergents, wetting agents and softeners. Read it together with sinking time: adding a wetting agent to shorten sinking time can lead to a failure here. Method and judgement are as given in the standard.
Ethylene oxide residue
The introduction does not recommend EO sterilization for absorbent cotton, but if a product is in fact EO-sterilized, its residue has to be assessed; the accredited clauses are in the scope table below. Residue levels depend on aeration time and conditions, so the samples should be finished product that has completed sterilization and aeration, kept sealed, with the sterilization date stated.
Bioburden (microbial limit)
This item corresponds to the foreword's new requirement for non-sterile cotton to state a microbial limit (3.16) and its test method (4.14); SUNGO's scope names it "bioburden test". The microbiological method standard cited in Clause 2 is GB/T 19973.1 (microorganisms on products). When sampling:
- Samples must be unsterilized and represent normal production: one batch, routine process, final packaging.
- Keep the original packaging; do not open or repack. Repacking introduces extra microorganisms and the result no longer represents the product.
- State the production date and storage conditions, and agree delivery time in advance.
- If the result is to be judged against a stated microbial limit, provide the stated value from the label or the product technical requirement.
Accreditation scope table
Reproduced from the CNAS L22083 scope (test object: medical absorbent cotton) and CMA 240015340087 schedule of Shanghai SUNGO Medical Technology Co., Ltd.; the limitation field for these rows is empty in both.
| Item | Clause | Notes/limitations |
|---|---|---|
| Characters (性状) | YY/T 0330-2015 3.1 | CNAS; not in the CMA schedule, CNAS report only |
| Foreign fibres (外来纤维) | YY/T 0330-2015 3.3 | CNAS; not in the CMA schedule, CNAS report only |
| Neps (棉结) | YY/T 0330-2015 3.4 | CNAS; not in the CMA schedule, CNAS report only |
| Identification test (鉴别试验) | YY/T 0330-2015 4.3 | CNAS; CMA (no clause given in the schedule) |
| Acidity or alkalinity test (酸碱度试验) | YY/T 0330-2015 4.5 | CNAS; CMA (no clause given in the schedule) |
| Sinking time test (下沉时间试验) | YY/T 0330-2015 4.6.1 | CNAS; CMA (no clause given in the schedule) |
| Water absorbency test (吸水量试验) | YY/T 0330-2015 4.6.2 | CNAS; CMA (no clause given in the schedule) |
| Fluorescence test (荧光物试验) | YY/T 0330-2015 4.8 | CNAS; CMA (no clause given in the schedule) |
| Surface-active substances test (表面活性物质试验) | YY/T 0330-2015 4.11 | CNAS; CMA (no clause given in the schedule) |
| Extractable colouring matter test (可浸提着色物质试验) | YY/T 0330-2015 4.12 | CNAS; CMA (no clause given in the schedule) |
| Ethylene oxide residue test (环氧乙烷残留量试验) | YY/T 0330-2015 4.13; GB/T 14233.1-2022 9.4 | CNAS lists both standards; the CMA schedule lists YY/T 0330-2015 and GB/T 14233.1-2022 without clauses |
| Bioburden test (生物负载试验) | YY/T 0330-2015 4.14 | CNAS; CMA (no clause given in the schedule) |
Clause numbers are reproduced from the CNAS scope; for the exact sub-clauses under Clause 4.6, the standard text prevails. The Shanghai IAS scope (TL-1019) does not include YY/T 0330, so no IAS report can be issued against this standard. Clauses not listed in the certificates, such as sulphated ash (3.12, 4.10) mentioned in the foreword, are outside SUNGO's accreditation scope.
Where SUNGO tests it, and under which accreditation
All absorbent cotton testing is done by Shanghai SUNGO Medical Technology Co., Ltd., at No. 958 Kangqiao East Road, Pudong New Area, Shanghai; see mask, glove, dressing and gauze testing and sterilization-related testing. Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity whose certificates do not cover absorbent cotton. For both laboratories' certificates, see SUNGO Medical Device Testing Scope: CNAS, CMA and IAS in Shanghai and Hefei.
- CNAS reports: available for all 12 items. The CNAS certificate states that CNAS is a signatory to the ILAC and APAC mutual recognition arrangements; the signatory is the accreditation body, and whether an overseas regulator or buyer accepts a report depends on their requirements.
- CMA reports: only for the 9 items in the schedule, and the certificate states it is valid within the People's Republic of China. Characters, foreign fibres and neps are not in the schedule and cannot carry the CMA mark.
- IAS reports: not available, because YY/T 0330 is not in the Shanghai IAS scope.
- Biological tests: if registration of absorbent cotton as a dressing requires biological tests such as cytotoxicity, sensitization or irritation, SUNGO subcontracts them to laboratories holding the appropriate qualifications. Under accreditation rules, subcontracting is agreed with the client in advance, and the report identifies the subcontracted items and the subcontractor.
- Outside the scope: statements of compliance with the absorbent cotton monograph of the British Pharmacopoeia or other pharmacopoeias, GB/T 46936-2025, and any YY/T 0330 clause not listed in the certificates are outside SUNGO's accreditation scope.
For what each accreditation mark on a report actually proves, and how to check a scope yourself, see How to Check an Accreditation Scope and Read the Mark on a Report.
Preparing samples and the order
- State the standard and edition. Write YY/T 0330-2015 on the order form, list the items you need (see the table for item names), take clause numbers from your product technical requirement and the standard text, and attach the product technical requirement.
- State the purpose and the mark you need. Registration, buyer inspection or export.
- State the supply condition and sterilization method. Sterile or non-sterile; for EO-sterilized product, give the sterilization date and aeration details and keep samples sealed.
- Describe the product form. Roll, ball, swab or pad; finished article or raw cotton.
- Keep samples as shipped. Do not compress, tease out or repack absorbency samples; bioburden samples must be unsterilized and unopened, with the delivery time agreed in advance.
- Confirm sample quantities by item with the laboratory.
- Agree any items the standard does not specify. Whiteness was dropped in the 2015 edition and is not within SUNGO's accreditation scope; if your product technical requirement specifies it, mention this separately when you enquire.
Frequently asked questions
Can cotton balls and swabs be tested to YY/T 0330-2015?
The standard covers medical absorbent cotton itself, which is also the accredited test object. Order the clauses your product technical requirement cites and state how to sample the article; article-specific properties are outside this group of accredited items. Whether a report on finished balls or swabs can carry the CNAS or CMA mark is confirmed by the laboratory when the order is accepted.
Can characters, foreign fibres and neps carry the CMA mark, or be reported under IAS?
No to both. These three items are only in the CNAS scope, the CMA schedule lists the other 9 items only, and the Shanghai IAS scope (TL-1019) does not include YY/T 0330.
Will YY/T 0330-2015 reports still be usable after GB/T 46936-2025 takes effect?
YY/T 0330-2015 was still current as of 2026-10-22, and the platform states no replacement relationship. Applicability depends on which standard your product technical requirement cites and on review requirements.
Sinking time failed. What should we check before retesting?
First confirm that in-house sampling, sample conditioning, holder, water temperature and timing match Clause 4.6 of the standard, then check scouring and bleaching, processing-aid residues and compression. Do not tease samples out to pass; send what you normally ship.
When requesting a quotation, tell us the product name and form (absorbent cotton, balls, swabs and so on), the YY/T 0330-2015 items you need or your product technical requirement, the supply condition and sterilization method, and the report purpose and mark you need. Send them through our contact page, and the laboratory will reply item by item with the report types it can issue within its accreditation scope. A separate article covers YY/T 0331-2024 for absorbent cotton gauze.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.