GB/T 14233.1-2022 Clause 9: EO Residue by GC and Sampling for Masks and Gauze

GB/T 14233.1-2022 Clause 9: EO Residue by GC and Sampling for Masks and Gauze

Chinese product standards for medical masks, gauze, absorbent cotton and adhesive wound dressings rarely carry their own full method for ethylene oxide (EO) residue. Instead, they point to Clause 9 of GB/T 14233.1, "Test methods for infusion, transfusion, injection equipment for medical use – Part 1: Chemical analysis methods". This article covers the Clause 9 method, how product standards cite it, how sampling time and aeration affect results, and what SUNGO's Shanghai scope covers.

Key points first

  • GB/T 14233.1-2022 was issued in October 2022 and implemented on 2023-11-01, replacing GB/T 14233.1-2008 in full. Clause 9 is the determination of EO residue by gas chromatography (GC). The 2022 foreword lists two technical changes relating to EO residue determination: recommended GC test conditions were added (9.4.5), and the colorimetric method (Clause 10 of the 2008 edition) was deleted.
  • Clause 9.4 can be summed up as water extraction with headspace sampling: the body-contacting part with the highest relative EO residue is cut up and weighed, sealed with water and equilibrated at a set temperature, then gas from the top of the vessel is injected and quantified against a series of standard solutions.
  • GC conditions are not locked in: Clause 9.2 allows any GC method proven reliable, and 9.4.5 gives recommended conditions while allowing other validated ones.
  • Product standards cite different editions. YY 0469-2023 and YY/T 0969-2023 give a dated reference to the 2022 edition, and Clause 5.10 of YY 0469-2023 states that the test follows 9.4 of GB/T 14233.1-2022. YY/T 0330-2015 and YY/T 0331-2024 use an undated reference, so the latest edition applies. YY/T 1627-2018 gives a dated reference to the 2008 edition.
  • Results are expressed as relative content (μg/g or μg/cm²) or absolute content (mg per unit of product). Mask requirements are in μg/g; limit values are as stated in each product standard and are not reproduced here.
  • A result only represents the residue at the time of sampling and testing; sampling time and whether aeration is complete directly affect the number.
  • The EO residue items for the masks, gauze, absorbent cotton, gauze dressings and adhesive dressings discussed here are all within SUNGO Shanghai's CNAS (L22083) and CMA (240015340087) scope. IAS (TL-1019) covers GB/T 14233.1 Clause 9 and, among the product standards, only YY 0469, YY/T 0969 and YY/T 0331, all without edition years.

The standard and its scope

GB/T 14233.1-2022 is a recommended national method standard under technical committee TC106, with the NMPA as the competent authority. It applies to the chemical analysis of infusion, transfusion, injection and supporting equipment made of medical polymer materials, and other medical polymer products may use it as a reference. Masks, gauze and dressings use it only because their product standards cite it.

Its other clauses (general rules and other chemical analyses) are outside SUNGO's scope and not discussed here. Clause 9 is structured as 9.1 Instrument, 9.2 Analysis method, 9.3 Sample extraction methods, 9.4 Test procedure for exhaustive or exaggerated extraction (9.4.1 to 9.4.5), and 9.5 Calculation of results. SUNGO's certificate lists Clause 9.4 under the 2023 mask standards and under gauze and absorbent cotton.

According to the standard's revision explanation, the colorimetric method was dropped because of poor sensitivity and many interferences, while gas chromatography had become widely available. Which edition a product standard uses depends on the rule in its Clause 2: for dated references only the cited edition applies; for undated references the latest edition (including all amendments) applies. So "EO by GB/T 14233.1" can mean different editions and clauses.

Two bases also need to be kept apart: product standards give a mass fraction (μg/g), whereas GB/T 16886.7-2015 (identical to ISO 10993-7:2008) sets allowable limits as the dose a device delivers to the patient, grouped by contact duration, so the two should not be compared directly.

How each product standard cites Clause 9

The table shows only what was confirmed in standard texts or SUNGO's certificates; unchecked points are marked.

Product standardProductEO clauseGB/T 14233.1 reference
YY 0469-2023Surgical masksRequirement 4.10 (if applicable: masks sterilized or disinfected with EO); test method 5.10Dated reference to GB/T 14233.1-2022 in Clause 2; Clause 5.10 states the test follows 9.4 of GB/T 14233.1-2022
YY/T 0969-2023Single-use medical face masksRequirement 4.8 (if applicable); test method 5.8, which the foreword says was changed from the 2013 editionDated reference to GB/T 14233.1-2022 in Clause 2; SUNGO's certificate lists 9.4 under this test object
YY/T 0331-2024Absorbent cotton gauze and absorbent cotton and viscose gauzeRequirement 4.16; test method 5.17, both new in this edition according to the forewordUndated reference to GB/T 14233.1 in Clause 2 (latest edition applies); refer to the original text for the exact citation in 5.17
YY/T 0330-2015Medical absorbent cottonSUNGO's certificate lists 4.13 (EO residue test)Undated reference to GB/T 14233.1 in Clause 2; SUNGO's certificate lists GB/T 14233.1-2022 9.4 under this test object
YY/T 1627-2018Acute adhesive wound dressings and woundplastsNo clause number on this line of SUNGO's certificateDated reference to GB/T 14233.1-2008 in Clause 2; SUNGO's certificate lists Clause 9 of the 2008 edition, only as cited by YY/T 1627-2018

The method used in Clause 4.5 of YY 0594-2006 (surgical gauze dressings) cannot be confirmed from public sources; refer to the original text. SUNGO's certificate lists this item against YY 0594-2006 Clause 4.5.

The introductions of YY/T 0330-2015 and YY/T 0331-2024 note that cotton and cotton gauze are highly adsorbent and adsorbed EO is a significant hazard to patients and staff, so EO sterilization is not recommended. That is not a prohibition; for products still EO-sterilized, follow the product standard and technical requirements.

Key elements of the Clause 9 method

9.1 and 9.2: prove the method is reliable

Clause 9.1 simply specifies a gas chromatograph. Clause 9.2 requires the chosen method to have sufficient accuracy, precision, selectivity, linearity and sensitivity and to suit the device analysed; each product needs the necessary method evaluation, and other confirmed methods such as mass spectrometry may also be used. So ask whether the lab's method has been evaluated on similar materials; non-wovens, cotton and adhesives behave differently.

9.3: three extraction approaches, with a default

Clause 9.3.1 lists three basic extraction methods for EO-sterilized devices. Simulated-use extraction mimics product use as closely as possible, so the measured amount corresponds to the patient's actual intake. Exaggerated extraction is expected to release more than clinical use would. Exhaustive extraction is repeated until a further extraction yields only a small fraction of the first, or the cumulative residue no longer increases appreciably (see the original for the criterion). Where a citing document does not specify the extraction method, exhaustive or exaggerated extraction is used. Per 9.3.3, exhaustive extraction may be thermal or solvent-based, with water recommended.

Simulated-use conditions are to be set by the product standard under the most severe intended use (9.3.2). YY 0469-2023, by contrast, points straight to 9.4, the procedure for exhaustive or exaggerated extraction.

9.4.1: which part, and how it is prepared

Clause 9.4.1 takes the part that contacts the body and has the highest relative EO residue, cuts it into 5 mm pieces or small sheets, and places 1.0 g (or a specified area) in an extraction vessel with a precise volume of water. The vessel is sealed and equilibrated at 60 °C ± 1 °C for 40 min. See the original text for the vessel size and water volume.

Which part has the highest residue depends on the materials (mask body and ear loops; dressing adhesive, backing and pad). Tell the laboratory the component materials and whether the technical requirements name the part to test.

9.4.2 to 9.4.4: calibration and headspace injection

The stock solution can be prepared gravimetrically by injecting pure EO into a weighed volumetric flask containing water, taking the weight difference as the EO mass and making up to volume; certified commercial standard solutions diluted step by step may also be used. At least five concentration levels are then prepared and equilibrated in sealed vessels under the same conditions as the samples.

Gas is drawn quickly from the top of each equilibrated vessel and injected; EO peak height or area gives the calibration curve, and samples are measured the same way. If a sample falls outside the calibration range, the standard concentrations are changed and the curve redrawn. Equilibrating standards and samples under identical conditions keeps the quantification comparable.

9.4.5: recommended GC conditions

New in 2022, Clause 9.4.5 gives recommended values for the column stationary phase, inlet temperature, column flow, split ratio, oven temperature and detector temperature, with a flame ionization detector (FID), and states that other validated GC conditions may also be used. Per the revision explanation, it was added because the old standard gave none and unsuitable conditions could be used. See the original for the values.

9.5: calculation and units

Clause 9.5 allows EO residue to be expressed as absolute or relative content. Relative content is the extract concentration read from the curve, multiplied by the extract volume and divided by the sample amount, in μg/g or μg/cm². Absolute content scales this by the ratio of unit product mass to sample mass and is given in mg per unit of product. The mask requirements in μg/g correspond to relative content.

Duplicate determinations

Under the general rules in Clause 4 (4.1.1), every analysis in the standard is run as two parallel determinations. If they agree within the allowed relative deviation, the arithmetic mean is reported; if one passes and the other fails, no average is taken and the test is repeated. If a report shows one figure, you can ask for the parallel results.

Why sampling time and aeration decide the result

EO boils at about 10.6 °C at atmospheric pressure and is highly volatile at ambient temperature, so residue falls over time after sterilization; this is aeration. Clause 5 (Product release) of GB/T 16886.7-2015 uses dissipation curves to estimate how long a product needs to reach the residue limits, and its Annex D covers factors influencing product residuals. A reported figure corresponds to one sampling and testing point only.

GB/T 14233.1-2022 Clause 9.3.1 reflects this too: the extract should be prepared after sampling; otherwise the sample is to be kept sealed in a metal container closed with PTFE. These situations most often distort results:

  • Samples are sent soon after sterilization although the product is released only after aeration, so the result is high and unrepresentative.
  • Samples sit outside the lab for a long time or the pack has been opened, so the result may be low.
  • Samples come from only one position in the sterilizer load and do not represent the batch.
  • Samples cannot be matched to the sterilization batch or aeration end time.

Record the sterilization batch and date, aeration end time and sampling date, and ship samples sealed in original packaging. To study aeration, agree time points with the laboratory in advance.

Common reasons for failing at release

Release failures usually trace back to process or sampling:

  • Aeration time was too short, or aeration temperature and ventilation differed from the settings.
  • The load was packed too densely, or cartons and inner packs have poor gas permeability.
  • Highly adsorbent materials, such as cotton or thick multilayers, lose residue more slowly.
  • Packaging materials or loading pattern changed without the aeration time being reassessed.
  • Sampling took place before aeration ended, or from areas where residue tends to be higher.
  • The method had not been evaluated for that material, and matrix interference affected quantification.

The first five are for the manufacturer and its sterilization provider to resolve. EO sterilization process validation (ISO 11135 / GB 18279) is outside SUNGO's accredited scope; SUNGO provides EO residue test results to the standard method.

Accreditation scope table

Each line below is taken from the CNAS, CMA and IAS scopes of Shanghai SUNGO Medical Technology Co., Ltd. Item names, clauses and limitations are reproduced as listed. The CMA schedule lists standard numbers only, without clause numbers.

ItemBasis (clause)Notes / limitations
Surgical masks: EO residueYY 0469-2023 5.10; GB/T 14233.1-2022 9.4CNAS lists the clauses; CMA lists YY 0469-2023 and GB/T 14233.1-2022
Surgical masks: EO residueYY 0469-2011 5.10; GB/T 14233.1-2022 9CNAS lists the clauses; CMA lists YY 0469-2011 and GB/T 14233.1-2022
Single-use medical face masks: EO residueYY/T 0969-2023 5.8; GB/T 14233.1-2022 9.4CNAS lists the clauses; CMA lists YY/T 0969-2023 and GB/T 14233.1-2022
Single-use medical face masks: EO residueYY/T 0969-2013 5.8; GB/T 14233.1-2022 9CNAS lists the clauses; CMA lists YY/T 0969-2013 and GB/T 14233.1-2022
Absorbent cotton gauze (CMA test object: "gauze"): EO residueGB/T 14233.1-2022 9.4CNAS lists the clause; CMA lists GB/T 14233.1-2022; neither lists a separate YY/T 0331-2024 EO clause under this object
Medical absorbent cotton: EO residue testYY/T 0330-2015 4.13; GB/T 14233.1-2022 9.4CNAS lists the clauses; CMA lists YY/T 0330-2015 and GB/T 14233.1-2022
Gauze dressings: EO residueYY 0594-2006 4.5CNAS lists the clause; CMA lists YY 0594-2006
Adhesive wound dressings and woundplasts: EO residueYY/T 1627-2018 (no clause number on the certificate); GB/T 14233.1-2008 9Both CNAS and CMA; GB/T 14233.1-2008 is limited to use as cited by YY/T 1627-2018
IAS: GB/T 14233.1 Clause 9 (EO residue-GC)GB/T 14233.1 9IAS scope does not state an edition year
IAS: related product standardsYY 0469; YY/T 0969; YY/T 0331YY 0469 except 5.6.2, 5.11, 5.12, 5.13; YY/T 0969 except 5.9; YY/T 0331 full standard; no edition year stated. The EO clauses added in YY/T 0331-2024 (4.16/5.17) are not listed separately by CNAS or CMA; confirm the edition with the laboratory before relying on IAS for them. YY/T 0330, YY 0594 and YY/T 1627 are not in the IAS scope

Where the testing is done and under which accreditation

All EO residue items in this article are tested at the laboratory of Shanghai SUNGO Medical Technology Co., Ltd., No. 958 Kangqiao East Road, Pudong New Area, Shanghai, under CNAS L22083, CMA 240015340087 and IAS TL-1019. Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity, and its accreditation scope does not include these items.

  • CNAS reports: issued against the clauses listed above. The CNAS certificate states that CNAS is a signatory to the ILAC and APAC mutual recognition arrangements.
  • CMA reports: cover only the items in the CMA schedule, and the certificate states that it is valid within the People's Republic of China.
  • IAS reports: the scope states no edition years, so confirm the edition with the laboratory first. YY/T 0330, YY 0594 and YY/T 1627 are not in the IAS scope, so IAS reports cannot be issued against their clauses; for an IAS report to GB/T 14233.1 Clause 9, confirm the edition first.

Whether an overseas regulator or buyer accepts a report is for the client to confirm for the target market; SUNGO makes no commitment on this. Commissioning biological tests and EO residue reporting under different standards are covered in a separate article.

Preparing your submission

  • State the product standard and edition, for example YY 0469-2023 or YY 0469-2011, YY/T 0969-2023 or YY/T 0969-2013. For adhesive dressings tested to YY/T 1627-2018, the method corresponds to GB/T 14233.1-2008.
  • State the sterilization method. In the mask standards (YY 0469-2023 4.10, YY/T 0969-2023 4.8) the EO requirement is marked "if applicable" and applies to masks sterilized or disinfected with EO; for other products, follow the product standard and technical requirements.
  • If the product technical requirements specify an extraction approach (for example simulated use) or the part to be tested, provide them; otherwise exhaustive or exaggerated extraction applies under 9.3.1.
  • Provide the sterilization batch number, sterilization date, aeration conditions and end time, sampling date, and the planned release point.
  • Keep samples sealed in their original packaging and deliver them promptly; opening or repacking lets EO escape.
  • Describe the materials of each component and which parts contact the body.
  • Confirm sample quantity with the laboratory before ordering.
  • Specify the report type (CNAS, CMA or IAS) and whether it is for registration in China or for an overseas customer.

Frequently asked questions

What is the difference between citing "GB/T 14233.1-2022 Clause 9" and "Clause 9.4" on a report?

SUNGO's CNAS certificate lists 9.4 under YY 0469-2023, YY/T 0969-2023, gauze and absorbent cotton, and Clause 9 under YY 0469-2011 and YY/T 0969-2013. The report follows the clause the product standard cites; YY 0469-2023 5.10 explicitly calls for 9.4.

Is Clause 9.4 an extraction method or a headspace method?

Both: the sample is equilibrated with water in a sealed vessel at a set temperature, and gas from the top of the vessel is injected. Where the product standard does not specify the extraction method, exhaustive or exaggerated extraction applies.

The 2022 edition is in force, so why are adhesive dressings still tested to the 2008 edition?

Clause 2 of YY/T 1627-2018 gives a dated reference to GB/T 14233.1-2008, and under the referencing rule only that edition applies. SUNGO's certificate restricts Clause 9 of the 2008 edition to use as cited by YY/T 1627-2018.

An export customer wants an IAS report. Can it be issued against GB/T 14233.1?

Yes for Clause 9: SUNGO Shanghai's IAS scope covers it, plus YY 0469 (except 5.6.2, 5.11, 5.12, 5.13), YY/T 0969 (except 5.9) and YY/T 0331, all without edition years, so confirm the edition with the laboratory first. YY/T 0331 is listed as a full standard, but its 2024 EO clauses (4.16/5.17) are not listed separately by CNAS or CMA, so check with the laboratory whether they can be reported under IAS. Acceptance by the customer or regulator is for the client to confirm.

For a quotation, send the product name, product standard and edition, sterilization method and batch information, and the report type (CNAS, CMA or IAS), so the laboratory can confirm the applicable clauses and sample requirements. Mask, dressing, gauze and cotton testing, including EO residue, is described on our mask and dressing testing page; sterilization validation services are on the cleaning, disinfection and sterilization validation page. You can also use the contact page.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.