Gurley air permeance appears on incoming inspection sheets, supplier data and regulatory files for medical packaging paper, adhesive coated paper and pouch paper, given as "Gurley seconds" (air resistance) or in µm/(Pa·s). This article explains how it is measured and converted, why both sides matter, how the ends of the range are handled, and how it relates to ethylene oxide (EO) and steam penetration and to microbial barrier. It works mainly from ISO 5636-5:2013, with GB/T 458-2008 alongside, and sets out which report types (CNAS, CMA, IAS) the SUNGO Shanghai laboratory can issue.
Key points
- The method measures time. It times how long a fixed volume of air takes to pass through a defined area of the test piece under a defined pressure difference. That time is the air resistance (s/100 mL); air permeance (µm/(Pa·s)) is calculated from it. More seconds means lower permeance.
- Conversion: P = 135.3/t. t is the mean time in seconds for 100 mL of air. The 1986 edition of ISO 5636-5 used a factor of 127; the 2003 edition changed it to 135.3. Check the factor before comparing with older data.
- Test both sides. Mark both sides and test half the pieces each way up. If the side means differ significantly (more than 10 %) and the report must show this, carry out 10 tests per side and report each side separately.
- There is a working range. ISO 5636-5 covers paper and board with air permeance between 0.1 and 100 µm/(Pa·s) and is unsuitable for rough-surfaced materials that cannot be clamped without leakage.
- Not a substitute for sterilization validation or barrier testing. Air permeance is one physical property. Sterilization validation shows whether the sterilant reaches the required conditions inside the package; microbial barrier testing shows whether microorganisms are kept out.
- Report types. The Shanghai laboratory can issue CNAS and CMA reports to GB/T 458-2008 and CNAS reports to ISO 5636-5:2013 (no CMA). Its IAS scope lists GB/T 458 and ISO 5636-5 with no edition year; confirm the edition with the laboratory before ordering an IAS report. The EN 868 series, ISO 5636-3 (Bendtsen method) and TAPPI T 460 are outside SUNGO's accreditation scope.
Standards and scope
| Standard | Edition and status | Role in this article |
|---|---|---|
| GB/T 458-2008, Paper and board – Determination of air permeance | Recommended Chinese national standard, issued 2008-08-19, effective 2009-05-01; reviewed 2025-03-27 and kept in force | Chinese reference; combines three formerly separate methods |
| ISO 5636-5:2013, Paper and board — Determination of air permeance (medium range) — Part 5: Gurley method | Third edition, 2013-11; confirmed at the 2019 and 2024 systematic reviews | Main source for the method details below |
| ISO 5636-3:2013 (Bendtsen method) | Third edition, 2013-11 | Air permeance method cited by EN 868-3:2017 and the EN 868-7 draft (refer to the 2025 editions); outside SUNGO's scope |
| TAPPI T 460, Air resistance of paper (Gurley method) | Edition as published by TAPPI | TAPPI Gurley method; outside SUNGO's scope |
GB/T 458-2008: three methods in one standard
According to China's national standards information platform, GB/T 458-2008 fully replaces GB/T 458-2002 (Schopper method), GB/T 5402-2003 (medium range, Gurley method) and GB/T 2679.13-1996 (medium range, Bendtsen method), and is a modified adoption (MOD) of ISO 5636-2:1984, ISO 5636-3:1992 and ISO 5636-5:2003. The three methods now share one document; see its text for the clause covering each.
This article covers only the Gurley method. GB/T 458-2008 adopted ISO 5636-5:2003. The foreword of ISO 5636-5:2013 says the third edition technically revises the 2003 edition, mainly through editorial changes, and adds an informative Annex C with precision data. The two editions also differ in apparatus details; for example, the timer is read to the nearest 0.2 s in 2003 and 0.1 s in 2013. GB/T 458-2008 is itself a modified adoption; refer to each text for the differences. SUNGO's certificate entry for GB/T 458-2008 carries the Gurley method title, and the reports discussed here cover the Gurley method only.
ISO 5636-5:2013: the international Gurley method
ISO 5636-5:2013 specifies the Gurley method for the air permeance of paper and board and may also be used for air resistance. Air permeance is the mean air flow rate through unit area under unit pressure difference in unit time, with 1 mL/(m²·Pa·s) = 1 µm/(Pa·s). Air resistance is the time for a specific volume of air under unit pressure to pass through unit area, in s/100 mL. The standard calls the property permeance rather than permeability because it is a sheet property, not normalized for thickness.
Where it sits in packaging standards
ISO 11607-1:2019, clause 5.1.7 e), requires packaging materials to conform to established physical properties such as tensile strength, thickness variation, tear resistance, air permeance and burst strength; 5.1.7 h) requires microbial barrier properties consistent with specified acceptance criteria unless the material meets the impermeability criterion of Annex C. ISO 11607-1:2019 lists ISO 5636-5 as a normative reference; its normative Annex C is a test method for resistance of impermeable materials to the passage of air (refer to the standard for the procedure).
ISO 11607-1:2019 also notes that conformity with the EN 868 series can demonstrate conformity with one or more of its requirements. As background: EN 868-3 covers paper for bags (EN 868-4) and for pouches and reels (EN 868-5), its 2025 edition being scoped to higher temperature sterilization such as moist heat in healthcare facilities; EN 868-7 covers adhesive coated paper for low temperature sterilization. EN 868-3:2017 (in its normative references) and the EN 868-7 draft text cite the ISO 5636-3 Bendtsen method, not Gurley, for air permeance; refer to the 2025 editions for whether this still applies and the values required. SUNGO does not issue EN 868 conformity reports.
Principle: a cylinder sinking under its own weight
Clause 4 of ISO 5636-5:2013: a vertical cylinder floating in liquid compresses air by its weight; as air passes through the test piece, the cylinder falls steadily; the time for a given volume to pass (the air resistance) is measured and air permeance calculated from it. Per clause 5, the apparatus comprises:
- Outer and inner cylinders: the outer cylinder is partly filled with sealing oil; the inner cylinder is graduated in 50 mL steps with a full scale of at least 300 mL, and the cylinder assembly weighs (567.0 ± 0.5) g, which provides the pressure.
- Clamping plates and gasket: a gasket inside diameter of (28.6 ± 0.1) mm defines a 642 mm² test area and prevents leakage at the paper surface. The recommended clamping force is (180 ± 30) N.
- Timing and zero check: timing is read to 0.1 s; with a non-porous plate clamped in place, the reading should be zero.
The time for the first two consecutive 50 mL graduations (100 mL in total) to pass the rim of the outer cylinder is recorded to the nearest 0.1 s. Vibration increases the rate of air displacement and must be avoided.
From seconds to µm/(Pa·s): conversion and reporting
Clause 11.1 of ISO 5636-5:2013 gives P = 135.3 / t, where P is air permeance in µm/(Pa·s), reported to three significant figures, and t is the mean time in seconds for 100 mL of air. The equation is based on a mean pressure difference of 1.22 kPa, a test area of 642 mm² and 100 mL of air.
- Worked example: a mean time of 20.0 s gives P = 135.3/20.0 ≈ 6.77 µm/(Pa·s); conversely, t = 135.3/P. On this basis, the ends of the range, 0.1 and 100 µm/(Pa·s), correspond to roughly 1353 s and 1.35 s per 100 mL.
- Air resistance: reported as "Air resistance (Gurley)" in seconds, i.e. the mean time t, to two significant figures.
- History of the factor: the foreword of ISO 5636-5:2003 states that the factor changed from 127 (1986 edition) to 135.3, raising results by about 7 %, and that the factor used should be reported. GB/T 458-2008 is based on the 2003 edition; refer to its text for its equation.
Conditioning and test pieces
Paper properties change with moisture content, which depends on ambient humidity and humidity history, so conditioning is essential. ISO 5636-5 calls for conditioning to ISO 187; the Chinese counterpart is GB/T 10739-2023 (issued 2023-12-28, effective 2024-07-01, a modified adoption of ISO 187:2022).
- Standard atmosphere: ISO 187:2022 specifies (23 ± 1) °C and (50 ± 2) % RH; tropical countries may use (27 ± 1) °C and (65 ± 2) % RH.
- Approach from the dry side: paper reaches a different equilibrium moisture content depending on whether it gains or loses moisture. For tests where this hysteresis can cause important errors, ISO 187:2022 requires 24 h preconditioning at 10 % to 35 % RH and not above 40 °C before conditioning.
- Same atmosphere throughout: test pieces are prepared and tested in the conditioning atmosphere.
- Test pieces: at least 10, both sides identified; test area free from folds, wrinkles, holes, watermarks or defects not inherent to the sample, and not handled; 100 mm × 100 mm is adequate. For the mean quality of a lot, sample to ISO 186.
Two sides: why both are tested
Paper coated or treated on one side may differ from side to side, so ISO 5636-5 requires both sides to be tested. Clause 10 requires at least 10 test pieces, five with side 1 up and five with side 1 down. Under clauses 8 and 11.2, if the side means differ significantly (more than 10 %) and the difference must be shown in the report, 10 tests are required per side and the side means are reported separately; otherwise the mean of both sides is reported, with standard deviations calculated to match. Naming the sides in your own terms ("coated / uncoated", "inside / outside") lets the results map onto your specification.
Very open and very tight materials
- Relatively impermeable paper and board: the reading may be taken at the end of the first 50 mL interval.
- Very open or porous paper: a larger volume of air may be timed.
- Cylinder not moving steadily: if steady movement is not reached before the zero mark, timing may start at the 50 mL mark.
- Back to 100 mL: any other timed volume is converted to a 100 mL basis, and the measured volume is stated in the report.
Agree two points beforehand. The method is unsuitable for rough-surfaced materials that cannot be clamped securely, and where surface leakage may occur the clamping force must be controlled and repeatable, so check whether strongly embossed or creped materials can be clamped leak-free. As the scope is paper and board, also confirm whether non-paper porous materials such as polyolefin nonwovens can be tested and how the report will describe this. Where air permeance is clearly outside 0.1–100 µm/(Pa·s), the Gurley method itself is not appropriate.
Air permeance, sterilant penetration and microbial barrier
ISO 11607-1:2019 states that a terminally sterilized medical device packaging system is meant to allow sterilization, provide physical protection, maintain sterility up to the point of use and allow aseptic presentation. It defines sterilization compatibility as the attributes that let packaging both withstand the sterilization process and attain the required conditions for sterilization inside it. Porous materials sit where these demands meet.
Sterilization: gas has to get in and out
- EO: the US FDA's sterilization page states that an EO-sterilized device must be sealed in a carefully designed gas-permeable package that enables the EO gas to enter. FDA's guidance on reprocessing devices in health care settings lists aeration time among critical EO cycle parameters, so residuals also have to get out.
- Steam: the same guidance lists cycle type (dynamic air removal or gravity), exposure time, temperature and drying time as critical steam parameters, and recommends labeling that identifies FDA-cleared packaging designed to allow adequate sterilant penetration as well as maintenance of sterility.
Higher air permeance (fewer Gurley seconds) means less resistance to air. Whether the sterilant reaches the required conditions inside the package, however, must be shown by sterilization validation for the specific product, package and load; it cannot be inferred from one permeance value. How the two validations fit together is covered in Sterile Barrier Systems: Packaging and Sterilization Validation.
Barrier: microorganisms have to be kept out
Air permeance describes how easily air passes through a material; whether microorganisms pass also depends on factors such as material structure, with no fixed conversion between the two. ISO 11607-1:2019 treats them as separate requirements: air permeance is a physical property under 5.1.7 e), while microbial barrier is required separately under 5.1.7 h) and 5.2. Material selection has to balance both, with barrier evaluated by dedicated methods. Two other articles on this site cover moist and dry microbial barrier testing to DIN 58953-6 and YY/T 0681.14, and microbial ranking to ASTM F1608.
How Gurley data are used
- Incoming inspection and lot consistency: specify air permeance and check each lot with the same method and factor.
- Before and after sterilization: within a sterilization compatibility study. Steam sterilization is itself a high-humidity treatment, so agree beforehand how sterilized samples are conditioned and reported.
- Before and after aging: to track physical properties in shelf-life studies, with both sets conditioned and tested the same way.
Accreditation scope table
Relevant rows from the scope annexes of the three certificates of Shanghai SUNGO Medical Technology Co., Ltd. Item, standard, clause and limitation follow the certificates; "Note" text is added.
| Certificate | Item | Standard and clause | Notes / limitations |
|---|---|---|---|
| CNAS L22083 | Medical device packaging materials: air permeance | GB/T 458-2008 (no clause listed) | No limitation stated |
| CNAS L22083 | Medical device packaging materials: air permeance | ISO 5636-5:2013 (no clause listed) | No limitation stated; Note: not in CMA scope |
| CMA 240015340087 | Medical device packaging materials: air permeance | GB/T 458-2008 (no clause listed) | No limitation stated; Note: CMA is valid within the People's Republic of China |
| IAS TL-1019 | Whole standard | GB/T 458 | No limitation stated; Note: no edition year in IAS scope |
| IAS TL-1019 | Whole standard | ISO 5636-5 | No limitation stated; Note: no edition year in IAS scope |
Where SUNGO does this test and under which accreditation
The test is performed at the Shanghai laboratory: Shanghai SUNGO Medical Technology Co., Ltd., laboratory at 958 Kangqiao East Road, Pudong, Shanghai, holding CNAS accreditation (L22083), CMA certification (240015340087) and IAS accreditation (TL-1019). Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity whose scope does not include air permeance testing.
| Certificate | Standards listed | Notes |
|---|---|---|
| CNAS L22083 | GB/T 458-2008; ISO 5636-5:2013 | CNAS reports to either |
| CMA 240015340087 | GB/T 458-2008 | ISO 5636-5 is not in the CMA annex, so no CMA report |
| IAS TL-1019 | GB/T 458; ISO 5636-5 (no edition year in scope) | Confirm the edition with the laboratory before ordering an IAS report |
| Not in any scope | EN 868 series, ISO 5636-3, TAPPI T 460 | None of these report types |
- CMA: covers only items in the CMA scope annex and is valid within the People's Republic of China. The only relevant CMA row here is GB/T 458-2008.
- IAS: GB/T 458 and ISO 5636-5 carry no edition year in the IAS scope; for an IAS report to a specific edition, confirm the edition with the laboratory before ordering.
- Method: GB/T 458-2008 contains more than one method; the certificate entry is registered under the Gurley method title, and the reports described here refer to the Gurley method only.
- Acceptance: an accreditation mark shows competence within that scope; whether a regulator, notified body or buyer accepts a report depends on their own requirements.
Preparing a submission
- Standard and report type: GB/T 458-2008 or ISO 5636-5:2013; CNAS, CMA or IAS report. CMA covers GB/T 458-2008 only; for IAS, confirm the edition first.
- Results: air permeance, air resistance or both; sides separately or combined; whether standard deviation or coefficient of variation is needed.
- Specification text: attach it so method, factor and timed volume can be checked; if it calls for TAPPI T 460 or the Bendtsen method, first confirm with your customer whether Gurley results are acceptable.
- Sides and condition: name the two sides, and say whether the sample is raw material, cut from finished pouches, or sterilized or aged.
- Protection: ship flat, not folded or rolled, dry and clean; check beforehand whether suitable test pieces can be cut from finished pouches.
- Quantity: depends on test-piece numbers, per-side reporting and any before/after comparison; confirm before ordering.
FAQ
Gurley seconds or µm/(Pa·s): which should the report show?
Both come from one measurement: seconds are the air resistance t, and P = 135.3/t. Report whichever your specification uses, or both.
My supplier's data are to TAPPI T 460. Can I compare them directly with ISO 5636-5 results?
TAPPI lists T 460 as related to ISO 5636-5 and notes that "related" does not imply "equivalent". T 460 is outside SUNGO's scope; whether results can be compared directly is for you to judge against your specification.
Can SUNGO test air permeance as required by EN 868-3?
EN 868-3:2017 (in its normative references) and the EN 868-7 draft text cite the ISO 5636-3 Bendtsen method for air permeance; refer to the 2025 editions. Neither EN 868 nor ISO 5636-3 is in SUNGO's scope. SUNGO can measure Gurley air permeance to GB/T 458-2008 or ISO 5636-5:2013, but that is not an EN 868 conformity report.
Can an ISO 5636-5 report carry the CMA mark?
No. The only air permeance standard in the Shanghai CMA annex is GB/T 458-2008. ISO 5636-5:2013 reports can be issued under CNAS; for an IAS report to the 2013 edition, confirm the edition first, as the IAS scope lists ISO 5636-5 without a year.
If air permeance is within specification, is microbial barrier testing still needed?
Yes. There is no fixed conversion between them; ISO 11607-1 treats them as separate requirements, and barrier needs its own test.
For a quote, give the material and grade, standard and edition, air permeance or air resistance, whether sides are reported separately, report type (CNAS, CMA or IAS), sample condition and intended use. Related services are on the sterile packaging validation service page, and requirements can be sent through the contact page; the laboratory will confirm which report types it can issue within its scope.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.