GB/T 4857.13 Low-Pressure Testing for Air-Shipped Medical Device Packages

GB/T 4857.13 Low-Pressure Testing for Air-Shipped Medical Device Packages

When sterile medical devices travel by air, or by road over high mountain passes, the outside pressure falls below the pressure at which the packages were sealed. The air trapped in paper-film pouches, thermoformed trays and foil pouches expands: the package balloons, the seals are pushed from the inside, and in bad cases a seal peels, a channel opens or the pouch bursts. A low-pressure test reproduces this in the laboratory. This guide covers the pressure and time set by ISO 2873:2000 (adopted with modifications as GB/T 4857.13-2005), where to look up the values for ASTM D4169-23e1 Schedule I and YY/T 0681.15-2019 Schedule I, the risks to a sterile barrier system and the follow-up integrity tests, and which accredited reports SUNGO's Hefei laboratory can issue.

Key Points

  • Low-pressure tests simulate the pressure drop in air and high-altitude ground transport. ISO 2873:2000 covers pressurized aircraft at any altitude and unpressurized aircraft flying at 3 500 m or less. ASTM D6653 notes that feeder aircraft shipments may reach 5 791 m (19 000 ft) and also names ground transport over high mountain passes.
  • GB/T 4857.13-2005 is a modified adoption of ISO 2873:2000. Per the ISO text, pressure is reduced at no more than 150 mbar/min to 650 mbar (±5 %), held for a predetermined period, then restored with dry air at laboratory temperature at no more than 150 mbar/min. For the hold time required by the GB standard, refer to the GB text.
  • ASTM D4169-23e1 Schedule I (Low Pressure Hazard, Section 16) appears in DC-12 and DC-13, the "air (intercity) and motor freight (local)" cycles; its test type is vacuum. For the Schedule I pressure and hold time, refer to Section 16 of D4169. D4169 lists ASTM D6653 (vacuum method for high-altitude effects) as a referenced document, but do not assume that Schedule I simply uses the D6653 defaults.
  • YY/T 0681.15-2019 Schedule I, "low pressure (high altitude) hazard", is one step in the transport sequence for sterile device packages. The full text is not public; refer to the standard for pressure and time.
  • For a sterile barrier system, the main risk is ballooning and the load it puts on the seals. Integrity tests (visual inspection, dye penetration, seal strength) should follow; they are within the Shanghai laboratory's accredited scope.
  • All three low-pressure tests are done in Hefei. GB/T 4857.13-2005: CNAS only. ASTM D4169-23e1 Schedule I: CNAS, for shipping units not exceeding 150 lb (68 kg); the IAS scope covers ASTM D4169 with no edition year, except Schedules G and H. YY/T 0681.15-2019 Schedule I: CNAS and CMA, for sterile medical device packages; the IAS scope covers YY/T 0681.15 with no edition year. Confirm the edition with the laboratory before an IAS report is issued. Sterile medical device packages that need a CMA mark should use YY/T 0681.15-2019 Schedule I.
  • The optional ISTA 3A "random vibration under low pressure" block is outside SUNGO's accredited scope.

Standards and Scope

Sources: the Chinese national and industry standards platforms and the first page of each ASTM standard, checked on 2026-10-22.

StandardPublished / effectiveType and roleLow-pressure content
GB/T 4857.13-2005 Packaging – Basic tests for transport packages – Part 13: Low pressure test2005-05-25 / 2005-11-01; reviewed 2023-12-28 and confirmed as still validRecommended national standard, test method, under TC49 (packaging); replaces GB/T 4857.13-1992; modified adoption of ISO 2873:2000The whole standard is a single low-pressure test method
ASTM D4169-23e1 Standard Practice for Performance Testing of Shipping Containers and SystemsApproved 2023-12-01, published January 2024; editorial corrections March 2024Practice that builds a test sequence from the hazards of a distribution cycleSchedule I, Low Pressure Hazard, Section 16
YY/T 0681.15-2019 (Chinese industry standard on test methods for sterile medical device packages, Part 15: performance testing of shipping containers and systems)2019-10-23 / 2020-10-01Recommended Chinese medical device industry standard, test method, issued by the NMPA, filing No. 83469-2021Schedule I, low pressure (high altitude) hazard

GB/T 4857.13 and ISO 2873. ISO 2873:2000 is the third edition, a technical revision of the 1985 edition, and was confirmed by ISO in 2022. It subjects complete, filled transport packages and unit loads to low pressure similar to that in aircraft, by reducing the chamber pressure to that at 3 500 m; the standard says this is roughly the pressure in pressurized aircraft flying at any higher altitude. GB/T 4857.13-2005 is not available for online reading because of ISO copyright, so the details below follow ISO 2873:2000; for GB clause numbers and modifications, refer to the GB text.

ASTM D4169. Per 1.1, it is a uniform laboratory basis for evaluating whether shipping units withstand the distribution environment through a sequence of expected hazards; per 1.2, ASTM D7386 may be considered for single parcel shipments. In Table 1, DC-12 (air intercity and motor freight local, units over 150 lb (68.1 kg) or unitized) runs A, D, I, E, J, A, and DC-13 (same route, single packages up to 150 lb (68.1 kg)) runs A, C, F, I, E, J, A. The user-defined DC-2 can also draw on Schedules A through I.

YY/T 0681.15. Its scope is a uniform laboratory method for evaluating whether sterile medical device shipping units withstand the transport environment, with a test plan built around the hazards of a given distribution cycle; single parcel shipments are excluded. Its full text is not published, for international copyright reasons. The schedules on SUNGO's certificates (A, C, E, F, I, J) use the same letters as, and similar names to, the D4169 schedules; for how the two standards relate, refer to the texts.

Pressure and Time: What We Could Verify

BasisTarget pressurePump-down / repressurizationHold timeTemperature
ISO 2873:2000 (adopted with modifications as GB/T 4857.13-2005)650 mbar (±5 %), i.e. 65 kPa, equivalent to 3 500 m altitudeBoth at no more than 150 mbar/min; repressurize with dry air at laboratory temperatureA predetermined period set in the test plan; for the GB requirement, refer to the GB textA note gives (−8 ± 1) °C in the chamber during the hold if the combined effects of temperature and pressure in pressurized aircraft are to be studied
ASTM D6653/D6653M-13(2021) (vacuum method referenced by D4169)Default: see D6653; 9.1.3 allows adjustment for shipping environment, product value, acceptable damage level and similar criteriaVacuum released at about 305 m (1 000 ft) of altitude per 30–60 s60 min by default (9.2); may be adjusted on the same groundsSee standard
ASTM D4169-23e1 Schedule IRefer to Section 16 of D4169See standardSee standardSee standard
YY/T 0681.15-2019 Schedule ISee standardSee standardSee standardSee standard

We could not verify the D4169 Schedule I and YY/T 0681.15 Schedule I values against publicly accessible text, so the test plan should state the values used and their source. ISO 2873 notes that 1 mbar = 1 hPa = 0.1 kPa; its Clause 9 requires the report to give the chamber temperature, pressure and hold time, the results and any observations that help interpret them.

ASTM D6653 defines a feeder aircraft as a small, potentially nonpressurized aircraft carrying express packages to remote areas, and says its results are for qualitative purposes. Under the ISO 2873 scope, 650 mbar covers pressurized aircraft at any altitude and unpressurized aircraft flying at 3 500 m or less. If the route may involve unpressurized feeder aircraft or high mountain passes, the plan should justify that the chosen pressure covers them. The values in the three standards are not interchangeable.

What Low Pressure Does to a Sterile Barrier System

At constant temperature, the volume of a trapped gas is inversely proportional to its pressure. As the outside pressure falls, the air inside either escapes through the material or inflates the package, and the force ends up on the seals and other weak points. Package types behave differently:

Paper-film pouches

One web is porous (medical paper or a nonwoven), the other a clear film. The porous web vents slowly, so if the pressure drops faster than it can vent, the pouch still balloons and the seals see a peeling force from inside. A footnote to Table 1 of ASTM D4169 says this high-altitude, non-pressurized transport simulation may be deleted from DC-12 and DC-13 when primary packages contain a porous material. Justify any deletion in the plan; for YY/T 0681.15, refer to the standard.

Thermoformed trays with lids

The rigid tray hardly deforms, so the force concentrates on the lid and the perimeter seal: the lid domes outward and the seal is peeled from its inner edge. If the device sits close to the lid, watch for contact and rubbing.

Foil pouches and other non-porous packages

Gas cannot vent through the material, so the pouch can only expand and ballooning is usually most obvious; the more residual air, the higher the load on the seals. For devices containing liquids or gels, also check whether the inner container leaks under the pressure differential. The porous-material footnote does not apply to these packages.

Multi-layer packaging

An inner pouch inside an outer pouch or carton has little room to expand, so the layers press on each other; focus the inspection on the layer that is the sterile barrier. If the destination itself is at high altitude, the package stays at low pressure longer than the flight or crossing; allow for that.

After the Test: Package Integrity Testing

The low-pressure test has no pass/fail criterion of its own. ISO 2873 only requires results and observations to be recorded. ASTM D6653 9.4 calls for examining the specimen for damage or deformation, and 10.1 says an intact distribution packaging system should be reassembled and tested to an industry-accepted performance test such as D4169, to help determine whether the pressure differential had an effect. In D4169 DC-13, low pressure is the fourth step, followed by vehicle vibration, concentrated impact and a second handling step, so a slight peel can grow later. D4169 7.1 requires acceptance criteria to be set before testing.

Common integrity tests for sterile barrier systems:

  • Visual inspection of seal integrity: YY/T 0681.11-2014 applies to terminally sterilized flexible and rigid packages with at least one transparent side; the seal inspection part of ASTM F1886/F1886M likewise requires at least one transparent side (the current F1886/F1886M-25 also includes inspection of material surfaces; refer to the standard). This is the first check after low pressure: peeling, channels, wrinkles, damage.
  • Dye penetration: YY/T 0681.4-2021 and ASTM F1929-23 detect and locate seal leaks in packages made of a transparent material and a porous material.
  • Seal strength: YY/T 0681.2-2010 and ASTM F88/F88M-23, useful for comparing seal strength before and after low pressure.
  • Bubble emission leak test: ASTM D3078-02(2021)e1 detects gross leaks in flexible packaging containing a headspace gas. ASTM D6653 4.4 also notes that vacuum leak tests such as D3078 are applicable for testing the effects of high altitude.

Dye penetration and seal strength are covered in separate articles.

Accreditation Scope

Hefei laboratory (low pressure and related conditioning). Item names are translated from the certificates.

ItemStandard and clauseNotes / limitations
Low pressure testGB/T 4857.13-2005 (no clause listed)CNAS; test object "transport packages"; not in CMA or IAS scope
Temperature and humidity conditioning testGB/T 4857.2-2005 (no clause listed)CNAS; −40 °C to 150 °C only
Schedule I low pressure (high altitude) testASTM D4169-23e1, Clause 16CNAS; test object "shipping units not exceeding 150 lb (68 kg)"; not in CMA scope
Temperature and humidity conditioningASTM D4169-23e1, Clause 6CNAS; test object "shipping units not exceeding 150 lb (68 kg)"; −40 °C to 150 °C only
ASTM D4169 (IAS, no edition year)PartialIAS; Except for Schedule G, H (Schedule I is not excluded)
Schedule I – low pressure (high altitude) hazardYY/T 0681.15-2019, Clause 14 (clause as listed by CNAS)CNAS and CMA; test object "sterile medical device packages"
ConditioningYY/T 0681.15-2019, Clause 6 (clause as listed by CNAS)CNAS and CMA; −40 °C to 150 °C only
YY/T 0681.15 (IAS, no edition year)Whole standardIAS
ISTA 3A (IAS, no edition year)PartialIAS: except for Random vibration under low pressure; none of the five CNAS-listed ISTA 3A(18-22) items is a low-pressure test

Shanghai laboratory (integrity tests after transport):

ItemStandardNotes / limitations
IntegrityYY/T 0681.11-2014CNAS and CMA; IAS: YY/T 0681.11, whole standard
ASTM F1886/F1886M (IAS, no edition year)Whole standardIAS only; not listed under CNAS or CMA
Dye penetrationYY/T 0681.4-2021CNAS and CMA; IAS: YY/T 0681.4, whole standard
Dye penetrationASTM F1929-23CNAS; IAS: ASTM F1929, whole standard; not in CMA scope
Seal strengthYY/T 0681.2-2010CNAS and CMA; IAS: YY/T 0681.2, whole standard
Seal strengthASTM F88/F88M-23CNAS; IAS: ASTM F88/F88M, whole standard; not in CMA scope
Vacuum leakASTM D3078-02(2021)e1CNAS; IAS: ASTM D3078, whole standard; not in CMA scope

Where SUNGO Does It and Under Which Accreditation

Low-pressure testing is carried out by Hefei SUNGO Medical Technology Co., Ltd. (Hefei laboratory, Zipeng Town, Feixi County, Hefei, Anhui; CNAS L24279, CMA 260015342945, IAS TL-1119). Choose the basis by the report you need:

  • A CMA mark is needed: for sterile medical device packages, use YY/T 0681.15-2019 Schedule I. YY/T 0681.15 is the only Hefei transport standard in the CMA scope, and its CMA test object is "sterile medical device packages". The CMA mark applies only to items in the CMA schedule, and the certificate states that it is valid within the People's Republic of China.
  • The test plan follows ASTM D4169: Schedule I can be reported under CNAS, whose test object is "shipping units not exceeding 150 lb (68 kg)". DC-12 is meant for units over 150 lb or unitized loads; units over 150 lb (68 kg) are outside the CNAS test object, so a CNAS report cannot cover them. Check with the laboratory before ordering whether a unitized load falls within it or another accreditation can apply. The IAS scope covers ASTM D4169 except Schedules G and H and lists no edition year, so confirm the edition with the laboratory before an IAS report is issued.
  • GB/T 4857.13 as a stand-alone test: CNAS report only; no CMA mark and no IAS report.

Integrity tests after transport are carried out by Shanghai SUNGO Medical Technology Co., Ltd. (Shanghai laboratory, No. 958 Kangqiao East Road, Pudong; CNAS L22083, CMA 240015340087, IAS TL-1019). The two companies are separate legal entities and each issues its own reports. Moving samples from Hefei to Shanghai is itself a journey, so agree on handover and protection when ordering. Whether a regulator accepts a report depends on its requirements; please confirm this yourself before ordering.

Preparing Samples and Information

  • Distribution route: air transport, possible unpressurized aircraft such as feeder aircraft, high-altitude roads, and the approximate destination altitude. These decide whether low pressure is needed and which basis to use.
  • Basis and parameters: GB/T 4857.13 on its own, or within a D4169 DC-12/DC-13 or YY/T 0681.15 sequence; the pressure, hold time and any temperature control, and the standard or plan each value comes from.
  • Sample condition: per ISO 2873 Clause 6, fill with the intended contents and close as for normal distribution; simulated contents may be used if they are as close as possible in dimensions and physical properties, with the same closure. D4169 5.1 calls for representative samples of complete shipping units including actual contents; it also allows dummy loads where testing the actual product might be hazardous, under conditions given in the standard. Samples used to support sterile barrier claims normally go through the same sterilization as the product; see Sterile Barrier Systems: Packaging and Sterilization Validation.
  • Package details: whether each layer is porous, whether there is a transparent side, how much residual air is in the pouch, and whether the contents include liquids.
  • Conditioning: ISO 2873 Clause 7 requires conditioning to one of the conditions in ISO 2233; GB/T 4857.2-2005 is a modified adoption of ISO 2233:2000. The Hefei laboratory's temperature and humidity conditioning covers −40 °C to 150 °C only.
  • Acceptance criteria and follow-up tests: set acceptance criteria before testing, and state which integrity tests follow the low-pressure step and the full sequence, and when samples are taken.
  • Report requirements: CNAS, CMA or IAS mark, and the edition of each standard.

FAQ

Our product only travels by road. Do we still need low-pressure testing?

It depends on the route. ASTM D6653 names ground transport over high mountain passes. If the route involves neither air transport nor high-altitude roads, the plan may leave low pressure out, with the reason stated.

Can we skip low pressure for paper-film pouches?

A footnote to Table 1 of ASTM D4169 allows deletion from DC-12 and DC-13 when the primary packages contain a porous material. Justify any deletion in the plan; for YY/T 0681.15, refer to the standard.

The pouches ballooned during the test. Is that a failure?

Ballooning is a common physical effect of reduced pressure and is not in itself a failure. The result is judged against the acceptance criteria set beforehand and the later integrity tests: peeling, channels or damage at the seals.

Can a GB/T 4857.13 report carry a CMA mark?

No. GB/T 4857.13-2005 is only in the Hefei laboratory's CNAS scope. For sterile medical device packages that need a CMA mark, consider YY/T 0681.15-2019 Schedule I.

Can SUNGO run the ISTA 3A "random vibration under low pressure" block?

It is not listed in the CNAS scope and is explicitly excluded from the IAS scope, so SUNGO cannot issue an accredited report for that block.

When requesting a quote, please tell us the device and packaging structure and materials, package mass and dimensions, the distribution route (air transport, high-altitude roads), the intended basis (GB/T 4857.13 alone, or a D4169 or YY/T 0681.15 sequence) with pressure and time values, the accreditation mark you need, and the follow-up integrity tests. The Hefei and Shanghai laboratories will then confirm whether each item is within their accredited scope. Send your request through our contact page; service details are on the transport simulation testing and sterile medical device packaging validation pages.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.