If you export sterile surgical gloves to the United States, ASTM D3577 is a product specification commonly cited in 510(k) files, purchase contracts and inspection checklists. This guide covers the D3577 test items and the methods it calls up, what FDA expects for surgeon's gloves in a 510(k), the practical points for dimensions, physical properties, leaks, powder and protein, and exactly what Shanghai SUNGO Medical Technology Co., Ltd. can and cannot report under D3577. External information was checked on 2026-10-22. Numerical requirements are not reproduced here; the ASTM text governs.
Key points
- D3577 covers packaged sterile rubber surgical gloves only. The current edition is ASTM D3577-19(2023), Standard Specification for Rubber Surgical Gloves, with two types: natural rubber latex (Type 1) and synthetic rubber latex (Type 2). Inside and outside surfaces must be free of talc.
- Six groups of test items: sterility, freedom from holes, dimensions and tensile properties before and after aging, powder, water-extractable protein, and antigenic protein. The methods cited are ASTM D5151, D3767, D412, D573, D6124, D5712 and D6499, with sampling per ISO 2859.
- How FDA uses it: D3577-19(2023) is FDA-recognized (recognition number 6-443, partial recognition) and linked to product code KGO (surgeon's gloves, 21 CFR 878.4460, Class I, 510(k)). FDA's Medical Glove Guidance Manual (2008) states that FDA had not cleared any 510(k) for non-sterile surgeon's gloves.
- SUNGO Shanghai reports D3577 under CNAS only. The CNAS scope lists, for "rubber surgical gloves", D3577-19(2023) clause 8.3 (water leak test), 8.4 and 8.5 (physical and dimensional properties), 8.6 and 8.8 (residual powder) and 8.7 (water-soluble protein content), plus D5151, D412, D3767 and D6124. D3577 is not in the IAS scope, and no ASTM standard is in the CMA schedule.
- Outside the scope: the D3577 sterility item, D573 aging as a stand-alone test, D6499 antigenic protein, GB/T 7543 and ISO 10282. Biocompatibility testing is subcontracted by SUNGO to a laboratory with the appropriate qualifications.
The standard and its scope
ASTM D3577-19(2023)
- Subcommittee D11.40; Book of Standards Volume 09.02; 4 pages. The 2019 edition was reapproved in 2023 and is marked Active on the ASTM site (page last updated 2023-09-26).
- Scope 1.1: requirements for packaged sterile rubber surgical gloves used in surgical procedures. Scope 1.3 identifies Section 8 as the test method portion.
- Abstract: the gloves may be made from any rubber polymer compound; Type 1 is the natural rubber latex type and Type 2 the synthetic rubber latex type. FDA's sample 510(k) format describes Type 1 as compounded primarily from natural rubber latex, and Type 2 as compounded primarily from rubber cement or synthetic rubber latex.
- Tests listed in the abstract: sterility, freedom from holes, physical dimensions, tensile strength, ultimate elongation, stress at elongation, powder-free residue, powder amount, protein content and antigenic protein content.
- Values for each type and size, and the sampling plans, are not shown on the ASTM page. Refer to the specification.
How D3577 relates to examination glove specifications
Examination gloves have their own specifications: D3578 for natural rubber examination gloves (sterile and non-sterile) and D6319 for nitrile examination gloves, each covered in a separate article.
How FDA references D3577
- The FDA Recognized Consensus Standards database lists D3577-19(2023) under recognition number 6-443, with partial recognition: text on powdered surgeon's gloves, powdered patient examination gloves and absorbable powder for lubricating a surgeon's glove is not recognized, because FDA has banned those devices. The final rule was published on 2016-12-19 (81 FR 91722) and took effect on 2017-01-18.
- FDA's classification database defines KGO as a disposable device made of natural rubber latex or synthetic polymers, worn on the hands usually in surgical settings to provide a barrier against potentially infectious materials and other contaminants; submission type 510(k), Class 1. The current title of 21 CFR 878.4460 is "Non-powdered surgeon's glove".
- Other recognized standards on the KGO page include D5151 (6-424), D6124 (6-178), D7160 (6-401), D6499 and ANSI/ASQ Z1.4.
- FDA's Medical Glove Guidance Manual (issued 2008-01-22; its cover note says it has not been updated for the powdered glove ban) states that surgeon's gloves should be sterile when offered for sale and that FDA had not cleared any 510(k) for non-sterile surgeon's gloves. The 510(k) should include the sterilization method and cycle, and the sample format states that the sterility assurance level (SAL) should be 10⁻⁶ or better.
D3577 test items, methods and the SUNGO scope
| Item group | Method cited (current ASTM edition) | What is measured | SUNGO CNAS scope |
|---|---|---|---|
| Sterility | Refer to the specification | Sterility of packaged sterile gloves | Not in the D3577 rows; can be ordered separately under Chinese Pharmacopoeia General Chapter 1101 |
| Freedom from holes | D5151-19(2023) | Holes that leak water under the test conditions | D3577-19(2023) 8.3; D5151-19(2023) also listed |
| Dimensions and tensile before/after aging | D3767-03(2020), D412-16(2021); aging per D573-04(2025) | Length, width, thickness; tensile strength, ultimate elongation, stress at elongation | Scope item "physical and dimensional properties": D3577-19(2023) 8.4, 8.5; D412-16(2021) and D3767-03(2020) also listed; D573 not listed separately; whether post-aging properties are included is confirmed at contract review |
| Powder | D6124-06(2022) | Powder-free residue, powder amount | D3577-19(2023) 8.6, 8.8; D6124-06(2022) also listed |
| Water-extractable protein | D5712-15(2025) | Aqueous extractable protein (modified Lowry) | D3577-19(2023) 8.7; D5712 not listed separately |
| Antigenic protein | D6499-24 | Antigenic latex protein (immunoassay) | Not listed; outside the SUNGO accredited scope |
The method list comes from the CDC NIOSH PPE-Info record for D3577-19(2023). Which edition of each method applies is governed by the specification and the agreed test request.
Type 1 and Type 2: dimensional and physical requirements
D3577 sets dimensional and physical requirements; the requirements and values for each type are in the specification. The sample 510(k) format for surgeon's gloves in FDA's manual (FDA states that the format is not mandatory) asks the submitter to tick Type 1 or Type 2 and state minimum overall length, minimum width (medium size), minimum palm and finger thickness, minimum tensile strength and ultimate elongation before and after aging, and the pinhole AQL. It then asks whether latex gloves meet all current specifications in D3577 or an equivalent consensus standard, and whether synthetic polymer gloves meet D3577 or an equivalent standard for that polymer. A "no" answer needs an explanation and the equivalent standard.
The manual also addresses special surgeon's gloves:
- Microsurgery gloves are thinner in some areas, but when the manual was written FDA had not cleared any thinner than the specifications in the FDA-recognized consensus standard.
- Colored, flavored or scented gloves may share one 510(k), but each glove type should be tested separately as a finished glove for biocompatibility and all physical characteristics.
Points that commonly cause discrepancies in the laboratory:
- D412 Method A (dumbbell and straight specimens) and Method B (cut rings) do not produce identical results; D3577 sets which applies.
- Tensile properties depend on test conditions (extension rate, temperature, humidity, specimen geometry, conditioning), so compare results only under the same conditions.
- D3767 covers length, width and thickness measurement, including the specimen thickness used to calculate tensile results. Instrument pressure strongly affects thickness readings on soft materials, so the presser foot pressure must be specified.
- D573 assesses the effect of elevated temperature on vulcanized rubber; results may not correlate exactly with service performance but serve for laboratory comparison. Aging conditions are set in D3577. In the SUNGO scope, the corresponding item is "physical and dimensional properties" (clauses 8.4 and 8.5); whether it includes post-aging properties, and under which aging conditions, depends on the D3577 text and is confirmed at contract review. D573 is not listed separately in the scope.
Leaks: D5151 and FDA's water leak test
D5151 detects holes that leak water under the test conditions and is designed as a reference method. The smallest hole size that will leak has not been determined and is outside its scope. The inspection level and AQL that D3577 applies are given in the specification.
For its own regulatory sampling, FDA follows 21 CFR 800.20: visual examination plus a water leak test with 1,000 mL of water; sampling based on ISO 2859, single normal sampling for lots of 1,200 gloves or less and multiple normal sampling for larger lots, general inspection level II; and AQL 1.5 as the minimum acceptable quality level for surgeons' gloves (AQL 2.5 for patient examination gloves). 21 CFR 800.20 is not in the SUNGO accredited scope and is cited for comparison only. A separate article covers D5151 in more detail.
Powder: residue on powder-free gloves and powder amount
- D6124 determines the average powder or filter-retained mass on a sample of medical gloves, reported in milligrams per glove. It is a reference method and includes a Procedure II for powdered gloves.
- Powdered surgeon's gloves have been banned in the US since 2017-01-18, so for US-bound surgical gloves the powder item is in practice about residue on powder-free gloves.
- The sample 510(k) format asks for the residual powder weight per powder-free glove determined by D6124, and states that trace powder should not exceed 2 mg per glove or the limit in the ASTM standard. It also asks for the powders used in manufacturing, the removal process and the release specification.
- Powder-free surgeon's gloves may be lubricated with small amounts of silicone or other suitable lubricant, or coated with a non-tacky polymer; the 510(k) should characterize it, including chemical identity, specifications and biocompatibility.
Residual powder testing is covered in a separate article.
Protein: water-extractable and antigenic protein
FDA's manual notes that latex proteins can cause Type I sensitivity in some individuals. In the sample 510(k) format, the section on protein level of natural rubber latex gloves asks the submitter to:
- State the protein level (µg/dm²) measured by an FDA-recognized standard method, with the method and its year; a non-standard method needs a full description and correlation data against a recognized method.
- Confirm whether testing was done on finished gloves after real-time aging or accelerated aging per D3577, because water-soluble proteins may migrate to the surface during shipping.
- Summarize results from at least one lot and state the routine production control method.
D3577 references D5712: water-soluble proteins are extracted in buffer, precipitated to concentrate them and separate interfering substances, redissolved, and quantified colorimetrically by the modified Lowry method against a protein standard, with results recorded in micrograms per dm². Other methods may be used for routine quality control once validated and correlated with D5712. Extraction conditions and specimen area are as set in D5712.
The abstract also lists antigenic protein; the method in the NIOSH record is ASTM D6499, an inhibition immunoassay using rabbit antisera specific for natural rubber latex proteins. Its scope states that it measures antigenic protein and should not be taken as a measure of allergenic proteins. D6499 is not in the SUNGO accredited scope.
At SUNGO, water-extractable protein is reported under D3577 clause 8.7; D5712 is not listed separately. Process causes of high protein results are discussed in the D3578 article.
Accreditation scope table
Below are all rows in the CNAS (L22083) scope of Shanghai SUNGO Medical Technology Co., Ltd. for the test object "rubber surgical gloves". All took effect on 2026-03-09. The limitation column in the scope is empty for each, so the notes column below shows only the accreditation type.
| Item | Standard and clause | Notes / limitations |
|---|---|---|
| Water leak test | ASTM D3577-19(2023) 8.3 | CNAS |
| Water leak test | ASTM D5151-19(2023) (no clause listed) | CNAS |
| Physical and dimensional properties | ASTM D3577-19(2023) 8.4, 8.5 | CNAS |
| Physical and dimensional properties | ASTM D412-16(2021) (no clause listed) | CNAS |
| Physical and dimensional properties | ASTM D3767-03(2020) (no clause listed) | CNAS |
| Residual powder | ASTM D3577-19(2023) 8.6, 8.8 | CNAS |
| Residual powder | ASTM D6124-06(2022) (no clause listed) | CNAS |
| Water-soluble protein content | ASTM D3577-19(2023) 8.7 | CNAS |
Note: ASTM D573 (aging) and D5712 (water-extractable protein method) are not listed separately in the scope.
These items are not part of a D3577 report but can be ordered separately and reported separately:
| Item | Standard and clause | Notes / limitations |
|---|---|---|
| Sterility test | Chinese Pharmacopoeia 2020, Volume IV, General Chapter 1101 | CNAS; test object "medical devices" |
| Sterility test | Chinese Pharmacopoeia 2025, Volume IV, General Chapter 1101 | CNAS; test object "medical devices" |
| Sterility test | Chinese Pharmacopoeia 2020, Volume IV, 1101 | CMA; test object "sterile medical devices" |
| Sterility test | Chinese Pharmacopoeia 2025, General Chapter 1101 | CMA; test object "medical devices" |
| Sterile test method | Pharmacopoeia of the People's Republic of China, Volume 4, Section 1101 | IAS; only accredited for medical device testing; edition not stated |
| Shelf life | ASTM D7160-16(2023) (no clause listed) | CNAS; test object "gloves for medical use" |
Where the work is done and under which accreditation
The items in the scope tables above are tested at the laboratory of Shanghai SUNGO Medical Technology Co., Ltd., No. 958 Kangqiao East Road, Pudong, Shanghai; biocompatibility is subcontracted (see below) and is not performed in this laboratory. Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity whose scopes do not include medical gloves.
- CNAS (L22083): covers every row in the first table; D3577 reports carry the CNAS mark.
- IAS (TL-1019): does not include ASTM D3577, D412 or D3767, so a D3577 report cannot carry the IAS mark. The IAS scope lists D5151 and D6124 as whole standards, but these are test methods and do not replace a conformity statement against D3577.
- CMA (240015340087): contains no ASTM standards, so D3577 items cannot carry the CMA mark. CMA covers only the items in its schedule and is valid only within the People's Republic of China.
What we can and cannot do
Can do: the four D3577-19(2023) item groups in the scope (water leak test, physical and dimensional properties, residual powder, water-soluble protein content), together with D5151, D412, D3767 and D6124, reported under CNAS.
Cannot do, or not part of a D3577 report: an IAS report for D3577; D573 aging as a stand-alone test; D6499 antigenic protein; 21 CFR 800.20. GB/T 7543-2020 (single-use sterile rubber surgical gloves) and ISO 10282, the Chinese and international specifications for sterile rubber surgical gloves (GB/T 7543-2020 is identical to ISO 10282:2014), are not in any SUNGO accreditation.
Biocompatibility: FDA's manual recommends biocompatibility testing on finished gloves processed, packaged and sterilized like the distributed product. Biocompatibility testing is subcontracted by SUNGO to a laboratory with the appropriate qualifications. In line with accreditation rules, the client's consent is obtained before subcontracting, and the report identifies the subcontracted items and the subcontractor.
The CNAS certificate states that CNAS is a signatory to the ILAC and APAC mutual recognition arrangements. Whether FDA or a buyer accepts a given report depends on their own requirements; we make no commitment about any regulator's decision.
Preparing your samples
- State the basis: ASTM D3577-19(2023), Type 1 or Type 2, the clauses required, and whether a pass/fail statement is needed.
- Describe the product: natural or synthetic rubber (name the polymer); any coating or lubricant; color or scent; size range; special use such as microsurgery or orthopedic.
- Sample condition: in its biocompatibility chapter, FDA's manual states that data for sterile devices should include results from finished sterilized devices, and the 510(k) protein section refers to finished gloves. We recommend finished sterilized gloves identical to the marketed product, from one lot, in unopened original packaging, labeled with the lot and sterilization method.
- Aging groups: label unaged and aged samples separately. If post-aging data are needed, state whether aging is done by you or by the laboratory, on what basis and under what conditions; whether it can be included in the CNAS report is confirmed at contract review. For protein, state whether samples are aged.
- Lot information: leak testing is sampled per ISO 2859 and depends on lot size, so please give it. The laboratory confirms sample quantities per item, size and lot.
- Reports: D3577 items under CNAS; a sterility test (General Chapter 1101) ordered and reported separately; biocompatibility through the subcontracting procedure.
FAQ
Can SUNGO issue an IAS report for D3577?
No. D3577 is not in the Shanghai IAS (TL-1019) scope, so D3577 reports are issued under CNAS.
Does the D3577 report include a sterility test?
No. The D3577 rows in the CNAS scope do not include sterility. A sterility test can be ordered separately under Chinese Pharmacopoeia General Chapter 1101; whether it satisfies D3577 or your buyer depends on the specification and their requirements.
Which specification applies to nitrile surgical gloves?
Type 2 in D3577 is the synthetic rubber latex type, and FDA asks whether synthetic polymer surgeon's gloves meet D3577 or an equivalent standard for that polymer. Whether a material falls under Type 2 is set by the specification. D6319 covers nitrile examination gloves, not surgical gloves.
Can non-sterile surgical gloves be assessed against D3577?
D3577 covers packaged sterile gloves, and FDA's Medical Glove Guidance Manual (2008) states that FDA had not cleared any 510(k) for non-sterile surgeon's gloves. The Chinese standard GB/T 24787-2009 for single-use non-sterile rubber surgical gloves is covered in a separate article.
Can you test glove shelf life?
The Shanghai CNAS scope lists ASTM D7160-16(2023) for "gloves for medical use", item "shelf life"; the standard sets methods for testing medical gloves and analyzing the data to determine shelf life. It is not in the IAS or CMA scope. FDA's manual notes that if a sterile glove's sterility-based and physical-integrity-based expiration dates differ, only the earlier one should be shown. See our shelf-life validation page.
Requesting a quote
For D3577 testing, start with our mask, glove, dressing and gauze testing services, then send product details through the contact page: Type 1 or Type 2, material and surface treatment, clauses, sizes, lot size, sterilization method, aging status and the report mark you need. We will answer item by item against the CNAS scope and confirm sample quantities. For the overall US route, see Exporting Medical Gloves to the US: FDA Route and Test Items; for the EU, see Medical Gloves for the EU: What EN 455 Actually Requires.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.