The labels look perfect on the prototype, yet at the last step of the safety test programme, "verification of markings", the print smears under isopropyl alcohol, a label edge lifts, or a warning cannot be read under the specified conditions. Fixing that usually means new label stock or printing and another set of labelled samples. Clause 7.1 of IEC 60601-1 (GB 9706.1 in China) controls this with two tests: legibility (7.1.2) and durability (7.1.3), which includes the rub test. This guide explains how both are performed and judged, what Clause 7.2 requires as a minimum on the outside of the equipment, and which SUNGO laboratory does the work under which accreditation.
Key points first
- Legibility (7.1.2): markings are viewed from the operator's intended position or, if that is not specified and not obvious, from within a 30° cone at 1 m from the marking. Ambient illuminance is the least favourable level between 100 lx and 1 500 lx, and an observer with the specified visual acuity must read the marking correctly.
- Durability (7.1.3): markings must be removable only with a tool or appreciable force and stay legible for the expected service life. a) After all tests, markings are re-checked to 7.1.2 and adhesive labels must not have loosened or curled; b) markings required by 7.2, 7.4, 7.5 and 7.6 are rubbed by hand, without undue pressure, for 15 s each with distilled water, ethanol 96 % and isopropyl alcohol.
- External marking (7.2): if equipment is too small for all markings, at least those of 7.2.2, 7.2.5, 7.2.6 (not for permanently installed equipment), 7.2.10 and 7.2.13 (if applicable) are applied, and the rest go in the accompanying documents.
- Edition: Amendment 2 (AMD2:2020) made only a terminology change in 7.1.2 and did not touch 7.1.3. Within 7.2, subclauses 7.2.3 and 7.2.9 were substantively rewritten.
- At SUNGO: IEC 60601-1, EN 60601-1 and GB 9706.1 are tested by the Hefei laboratory, within the scope of its IAS accreditation (TL-1119). The certificate gives no year; confirm the edition before ordering.
Standards and scope
Designations and editions
IEC 60601-1 is the general standard for basic safety and essential performance of medical electrical equipment. As of 2026-10-22, the IEC Webstore lists edition 3.2 (IEC 60601-1:2005+AMD1:2012+AMD2:2020 CSV) as current, published 2020-08-20; Amendment 2 itself, IEC 60601-1:2005/AMD2:2020, was published the same day. GB 9706.1-2020 is the mandatory Chinese national standard, published 2020-04-09 and in force since 2023-05-01; it is a modified adoption of IEC 60601-1:2012. EN 60601-1 is the European adoption, which on the same date consists of EN 60601-1:2006 with amendments A1, A12, A2 and A13.
The test conditions here were taken from a regulator-published translation corresponding to IEC 60601-1:2012 (edition 3.1); refer to the IEC text for exact wording. They were checked against Amendment 2, which in Clause 7.1 only changes "safety signs" to the defined term "SAFETY SIGNS" in 7.1.2 and leaves 7.1.3 unchanged. Both tests are therefore the same in editions 3.1 and 3.2. GB 9706.1-2020 also specifies the wet rub test in 7.1.3; for its exact wording, refer to the standard.
How Clause 7 is organised, and which markings each test covers
Clause 7 "ME EQUIPMENT identification, marking and documents" covers general requirements (7.1), outside and inside marking (7.2, 7.3), controls and instruments (7.4), safety signs (7.5), symbols (7.6), conductor colours (7.7), indicator lights and controls (7.8) and accompanying documents (7.9). The two tests do not cover exactly the same markings:
- The legibility requirement of 7.1.2 and the final check in 7.1.3 a) cover all markings required by 7.2 to 7.6, including markings on the inside (7.3).
- The rub test in 7.1.3 b) applies only to markings required by 7.2, 7.4, 7.5 and 7.6. Internal markings under 7.3 are not listed.
Annex C (informative), Table C.1, collects external marking requirements found in other clauses, such as the warning about instability during transport (9.4.2.2), the overflow marking for reservoirs (11.6.2), the potential equalization conductor terminal (8.6.7) and AP/APG category markings (Annex G). Check that table too. Subclause 7.1.1 brings markings and documents into the usability evaluation (12.2); it is not covered here, and the usability engineering standard IEC 62366-1 is not within SUNGO's accreditation scope.
7.1.2 Legibility: viewpoint, light level and observer
Viewing conditions by equipment and marking location
Markings required by 7.2 to 7.6 shall be clearly legible under these conditions:
- warnings, instructive statements, safety signs and symbols on the outside: from the intended position of the person performing the related function;
- fixed equipment: when mounted in its position of normal use;
- transportable equipment and stationary equipment that is not fixed: in normal use, or after moving it away from a wall against which it is positioned, or after turning it from its normal position of use, and for rack-mounted equipment, after removal from the rack;
- markings on the inside of the equipment or its parts: from the intended position of the person performing the related function.
Test method
| Element | What the text specifies |
|---|---|
| Viewpoint | Equipment or part positioned so that the viewpoint is the intended position of the operator; if not specified and not obvious, any point within the base of a cone subtended by an angle of 30° to the axis normal to the centre of the plane of the marking, at a distance of 1 m |
| Ambient illuminance | Least favourable level in the range 100 lx to 1 500 lx |
| Observer | Visual acuity of 0 on the log MAR scale or 6/6 (20/20), corrected if necessary, and able to read the Jaeger N6 test card in normal room lighting (about 500 lx) |
| Criterion | The observer correctly reads the marking from the viewpoint |
How results are judged
- "Least favourable" is not always the darkest setting. Small, low-contrast text suffers at low illuminance; glossy surfaces, reflective laminates or markings on transparent windows can suffer at high illuminance. Consider both ends.
- Where the accompanying or design documents state the intended operator position, the test uses it; otherwise the default 1 m / 30° condition applies.
- The criterion is reading correctly, not merely seeing. Shadowed, obstructed or easily confused characters can change the result.
- Annex A explains that the illuminance levels come from interior lighting design recommendations, that Jaeger N6 corresponds roughly to reading at 0.87 m, and that 500 lx is recommended for measuring visual acuity.
7.1.3 Durability: the rub test and label adhesion
The requirement
Markings required by 7.2 to 7.6 shall be removable only with a tool or by appreciable force, and sufficiently durable to remain clearly legible during the expected service life of the equipment. The effect of normal use shall be taken into account. The text gives no figure for "appreciable force"; it is checked by inspection.
Compliance: parts a) and b)
| Part | Markings covered | When and how | Criterion |
|---|---|---|---|
| a) Final check | Markings required by 7.2 to 7.6 | After all tests of the standard (see the test sequence recommended in Annex B) | Legibility tested to 7.1.2; adhesive labels have not worked loose or become curled at the edges |
| b) Additional rub test | Markings required by 7.2, 7.4, 7.5 and 7.6 | Rubbed by hand without undue pressure: 15 s with a cloth soaked in distilled water, then 15 s with ethanol 96 %, then 15 s with isopropyl alcohol | Part b) specifies the method only; whether the marking is still clearly legible and whether labels have loosened or curled is judged against 7.1.3 and the criteria of part a) |
Annex A notes that ethanol 96 % and isopropyl alcohol are defined as reagents in the European Pharmacopoeia. No stroke count or force value is given.
How results are judged
- The rub test applies by clause, not by process. Screen printing, pad printing, inkjet coding, thermal transfer, laser marking, moulded characters and rating plates are all included if they carry markings required by 7.2, 7.4, 7.5 or 7.6.
- Possible findings include dissolved or faded ink, smeared thermal transfer print and label edges lifting after solvent reaches the adhesive. The result depends on the print material and substrate combination, so the same label may behave differently on another enclosure material or finish.
- Curling is not only a rub test issue. In the Annex B sequence, humidity preconditioning (B.21), mechanical strength (B.27) and overflow, ingress, cleaning and disinfection (B.34, Clause 11.6) come before verification of markings (B.36). Labels on curved or textured surfaces or near heat sources deserve early attention.
Expected service life and the effect of normal use
Annex A, in its rationale for 4.4, explains that the manufacturer determines the expected service life within the risk management process, and that this is a precondition for assessing several requirements, including 7.1.3. Subclause 11.6.6 requires equipment to withstand the cleaning or disinfection processes in the instructions for use without damage to safety provisions, with the manufacturer evaluating the effect of repeated cleaning over the expected service life. Where the instructions specify other agents, see the FAQ.
Clause 7.2: the minimum external marking
Subclause 7.2.1 sets the floor. If the size of the equipment, part or accessory, or the nature of its enclosure, does not permit all markings of 7.2.2 to 7.2.20, at least the items in the first five rows of the table below shall be marked, and all remaining markings shall be recorded in the accompanying documents. Where no marking of the equipment is practicable, these markings may be applied to the individual packaging. Single-use materials, components, accessories or equipment, or their packaging, shall be marked "Do Not Reuse", "Single Use Only" or with symbol ISO 7000-1051.
| Subclause | External marking | Notes |
|---|---|---|
| 7.2.2 Identification | Manufacturer's name or trademark and contact information; model or type reference; serial number, lot or batch identifier; date of manufacture or use-by date, if applicable | Minimum set. Serial, lot or batch number and date of manufacture may be human-readable or provided by automatic identification such as bar code or RFID; software forming part of a PEMS needs a unique identifier available to designated persons, not necessarily on the outside |
| 7.2.5 Equipment powered from other equipment | One of three: manufacturer's name or trademark and model of the specified equipment near the connection; or ISO 7010-M002 near the connection with details in the instructions for use; or a special connector not commonly available, with details in the instructions for use | Minimum set. Applies where compliance depends on the equipment supplying power |
| 7.2.6 Connection to the supply mains | Rated supply voltage or voltage range; nature of supply (number of phases, except single-phase, and type of current); rated frequency or frequency range; symbol IEC 60417-5172 for Class II equipment | Minimum set (not for permanently installed equipment). Marked on the outside of the part containing the supply mains connection, preferably near the connection point |
| 7.2.10 Applied parts | Type B, BF or CF applied part symbol; the corresponding symbol for defibrillation-proof applied parts | Minimum set. Adjacent to or on the applied part connector, or on the applied part if there is no connector; where defibrillation protection is partly in the patient cable, ISO 7010-W001 near the relevant outlet |
| 7.2.13 Physiological effects | An appropriate safety sign | Minimum set (if applicable). For physiological effects that are not obvious to the operator and could cause harm to the patient or operator |
| 7.2.3 Consult accompanying documents | Safety sign ISO 7010-M002, or symbol ISO 7000-1641 | Rewritten in edition 3.2: M002 when consulting the documents is a primary risk control measure and the usability engineering process finds marking necessary; otherwise 1641 may be used |
| 7.2.9 IP classification | IP code to IEC 60529 according to the classification in 6.3 | Edition 3.2: no marking needed if the IP classification is not specified (IPXX) or is IP00, IPX0 or IP0X |
The other subclauses of 7.2 (7.2.4 accessories, 7.2.7 power input, 7.2.8 output connectors, 7.2.11 mode of operation, 7.2.12 fuses, 7.2.14 high-voltage terminal devices, 7.2.15 cooling, 7.2.16 mechanical stability, 7.2.17 protective packaging, 7.2.18 external pressure source, 7.2.19 functional earth terminals, 7.2.20 removable protective means, 7.2.21 mass of mobile equipment) apply depending on the design. For 7.2 to 7.4, compliance is checked by inspection and the tests and criteria of 7.1.2 and 7.1.3; 7.6 also requires symbols used in markings to be explained in the instructions for use. IEC 60417, ISO 7000, ISO 7010 and IEC 60529 are not listed separately on SUNGO's certificates and are mentioned only as sources referenced by Clause 7.
The accompanying documents (7.9) are themselves inspected; see IFU and Label Errors: What to Check Before You Ship Samples. For equipment intended for the home healthcare environment, Clause 7.2 of IEC 60601-1-11 adds requirements for marking the IP classification; a separate article on this site covers that standard.
Accreditation scope
Each row below is taken from the schedule of the Hefei laboratory's IAS certificate (TL-1119), with item, clause and limitation as listed:
| Standard | Item | Clause | Notes / limitations |
|---|---|---|---|
| IEC 60601-1 | Whole standard | No clause listed | No limitation listed; no year |
| EN 60601-1 | Whole standard | No clause listed | No limitation listed; no year |
| GB 9706.1 | Whole standard | No clause listed | No limitation listed; no year |
| IEC 60601-1-11 | Whole standard | No clause listed | No limitation listed; no year |
| EN 60601-1-11 | Whole standard | No clause listed | No limitation listed; no year |
| YY 9706.111 | Whole standard | No clause listed | No limitation listed; no year |
| ISO 80601-2-69 | Whole standard | No clause listed | No limitation listed; no year |
| EN ISO 80601-2-69 | Whole standard | No clause listed | No limitation listed; no year |
| YY 9706.269 | Whole standard | No clause listed | No limitation listed; no year |
Our note, not certificate text: Clauses 7.1 and 7.2 are not listed separately; they are part of IEC 60601-1 as a whole standard. Per the standard titles on the certificate, IEC 60601-1-11 is the home healthcare collateral standard and ISO 80601-2-69 is the particular standard for oxygen concentrators.
Where the work is done at SUNGO, and under which accreditation
- The legibility test, rub test and inspection of markings under 7.2 to 7.6 are performed as part of IEC 60601-1, EN 60601-1 and GB 9706.1 by the Hefei laboratory, within the scope of its IAS accreditation (TL-1119). It belongs to Hefei SUNGO Medical Technology Co., Ltd., a separate legal entity in Zipeng Town, Feixi County, Hefei, Anhui Province. The scope of Shanghai SUNGO Medical Technology Co., Ltd. does not include these standards.
- The IAS certificate lists these standards without a year. For a report to IEC edition 3.2, GB 9706.1-2020 or a particular EN version, confirm the edition with the laboratory before ordering.
- For particular standards, the certificate lists only ISO 80601-2-69, EN ISO 80601-2-69 and YY 9706.269 for oxygen concentrators. Modifications or additions to Clause 7 made by other particular standards are not within SUNGO's accreditation scope.
- Whether a report is accepted for registration in a given country or region, including China, depends on the relevant regulatory requirements. SUNGO makes no commitment on acceptance.
Preparing for testing
- Production-equivalent labels: use production face stock, ink or ribbon, adhesive and laminate, applied with the production process, on the production enclosure material and finish.
- Marking list: give each marking's location, content, subclause, process and material, with label artwork. Annex B places the inspection of marking and documents (B.6) early, so content problems surface early.
- Space limitations: if not all markings fit, state which ones were kept under 7.2.1 and where the others appear in the accompanying documents.
- Operator position: state the intended operator position in the accompanying documents or design documents; it defines the viewpoint for 7.1.2.
- Service life and cleaning: provide the expected service life statement, the cleaning and disinfection methods and agents in the instructions for use, and the risk management assessment of marking durability.
- Sample planning: the 7.1.3 a) check comes last, so the unit that goes through the full programme keeps its original labels to the end. Agree with the laboratory which tests can share samples; see Planning Sample Quantities: Destructive and Parallel Tests.
- Pre-check (suggested): rub same-batch labels on a panel of the same enclosure material and finish with the specified reagents and order. This does not replace the formal test.
FAQ
How many strokes and how much force does the rub test use?
The text only says by hand, without undue pressure, 15 s per reagent in sequence. It gives no stroke count or force value; follow the edition used.
Do markings inside the equipment get the rub test?
Part b) of 7.1.3 covers markings required by 7.2, 7.4, 7.5 and 7.6; internal markings under 7.3 are not listed. Internal markings still have to meet 7.1.2 and are checked under 7.1.3 a) after all tests.
Are laser-marked or moulded markings exempt from rubbing?
No. The rub test follows the clause requiring the marking, not the process. Rub resistance is not legibility: laser marking contrast and moulded character depth and shadowing are still judged under 7.1.2.
Our instructions for use specify a disinfectant other than ethanol or isopropyl alcohol. What then?
The three reagents of 7.1.3 b) are fixed and do not change with your instructions for use. Your cleaning and disinfection processes are evaluated separately under 11.6.6, which precedes verification of markings in Annex B; the manufacturer assesses the effect of repeated cleaning on markings in the risk management file. Rubbing with your own agent outside the standard methods is not the 7.1.3 method and is outside the IAS accreditation scope.
Can only Clause 7 be evaluated?
The IAS certificate lists IEC 60601-1 as a whole standard, and the 7.1.3 a) check assumes all tests are complete. Confirm whether a partial evaluation is possible, and how the report would state its scope, when you request a quote.
Requesting a quote
Please provide the product name and intended use, models and configurations, target markets, the intended edition of the standard and whether home healthcare use is intended, with the marking list, label artwork and draft instructions for use. The Hefei laboratory uses this to confirm the scope, sample plan and test sequence. Service details are on the medical electrical safety testing page; send your request through the contact page.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.