IEC 61000-3-2 Harmonic Current Emissions: Class A or D for Medical Devices

IEC 61000-3-2 Harmonic Current Emissions: Class A or D for Medical Devices

Many mains-powered active medical devices have a "harmonic current emissions" line in their EMC results: IEC 61000-3-2 or EN IEC 61000-3-2 internationally, GB 17625.1 in China. It is often filled in as "Class A" without much thought. This article covers choosing the class, products below 75 W, test conditions, and where to start after a failure. RF disturbance classification and emission report structure are covered in EMC Emission Reports: GB 4824 Sits under a Higher Layer.

The short answer

  • Most medical devices are assessed as Class A. Under clause 5.1 of IEC 61000-3-2, anything not in Class B, C or D is Class A. Class D lists only PCs and their monitors, television receivers, and refrigerators and freezers with variable-speed compressor drives, up to 600 W. Medical devices are not on that list.
  • Non-lighting equipment rated at 75 W or less has no limits specified, but the control-method restrictions of clause 6.2 still apply.
  • Applicability depends on the supply and the use environment. IEC 60601-1-2 clause 7.2.1 covers equipment rated at 220 V line-to-neutral or more and 16 A per phase or less, intended for the public mains network; YY 9706.102-2021 clause 6.1.3.1 sets only the 16 A and public mains conditions.
  • Agree on the edition first. IEC 61000-3-2 consolidated edition 5.2 was published on 2024-03-04, GB 17625.1-2022 adopts IEC 61000-3-2:2020 with modifications, and IEC 60601-1-2 edition 4.1 cites IEC 61000-3-2:2005 with its two amendments.
  • Failures usually come from the power input stage; power factor correction (PFC) is the common fix.
  • SUNGO performs this test at the Hefei laboratory, within its IAS accreditation (TL-1119).

Standards and scope

StandardEdition informationRole
IEC 61000-3-2Edition 5.0, 2018-01; Amendment 1, 2020-07 (consolidated as edition 5.1); Amendment 2, 2024-03 (consolidated as edition 5.2). Prepared by IEC SC 77A; a product family standardClasses, test conditions, limits, assessment
EN IEC 61000-3-2EN IEC 61000-3-2:2019, based on IEC 61000-3-2:2018; A1:2021 and A2:2024 match the two IEC amendmentsEuropean version
GB 17625.1-2022Mandatory national standard, published 2022-12-29, effective 2024-07-01, replacing GB 17625.1-2012; modified adoption of IEC 61000-3-2:2020Chinese counterpart
IEC 60601-1-2Edition 4.0 published 2014-02-25; consolidated edition 4.1 published 2020-09-01Clause 7.2.1: when IEC 61000-3-2 applies
YY 9706.102-2021Published 2021-03-09, effective 2023-05-01, replacing YY 0505-2012; its foreword states it is a modified adoption of IEC 60601-1-2:2007Clause 6.1.3.1: when GB 17625.1 applies

Clause 1 of IEC 61000-3-2 covers equipment with a rated input current up to and including 16 A per phase, intended for public low-voltage distribution systems; limits have not yet been considered for systems below 220 V line-to-neutral. Two deviations listed in the GB 17625.1-2022 foreword matter here: "professional equipment" takes the GB/T 4365-2003 definition, and clause 6.1 covers only 220 V/380 V, 50 Hz systems. Clause numbers below apply to both standards; Amendment 2 changes are flagged.

An IEC interpretation sheet to Amendment 2 (April 2025) says that if the largest continuous input current exceeds 16 A, IEC 61000-3-12 applies instead. That standard is outside SUNGO's accreditation scope and is not covered here.

When the medical collateral standards call for the test

Clause 7.2.1 of IEC 60601-1-2:2014+AMD1:2020 says that ME equipment and ME systems rated at 220 V a.c. line-to-neutral or more and 16 A per phase or less, and intended for the public mains network, shall comply with IEC 61000-3-2. If both long-time and momentary current ratings exist, the higher one decides. Note b) to Table 2 says that in the professional healthcare facility environment the test does not apply unless the equipment will be connected to the public mains network and its power input is within the scope of the basic standard; the home healthcare column has no such note. Annex A adds that equipment in hospitals or similar locations whose mains is isolated from the public network by transformers or substations is not connected to the public mains network, and that systems rated below 220 V a.c. are exempt.

Clause 6.1.3.1 of YY 9706.102-2021 requires equipment rated at 16 A per phase or less and intended for the public mains network to comply with GB 17625.1, using its test methods and equipment. GB 17625.1 is an undated reference, so its latest edition applies. The emissions declaration in Table 1 has a harmonic row to be filled in as Class A, B, C, D or not applicable, so the class you choose also ends up in the accompanying documents.

In IEC 60601-1-2 edition 4.1, only the cited edition of a dated reference applies. The edition you test to depends on the target market; agree it with the laboratory.

The four classes, and where medical devices fall

ClassWhat clause 5.1 saysTypical medical device situation
Class AEquipment not specified as Class B, C or D. Examples: balanced three-phase equipment, household appliances not in other classes, tools other than portable tools, audio equipmentMost mains-powered active medical devices
Class BPortable tools; arc welding equipment that is not professional equipmentRarely applicable
Class CLighting equipment (clause 5.2 lists inclusions and exclusions)Only products whose primary function is lighting; built-in or indicator lighting is excluded
Class DPersonal computers and their monitors, television receivers, and refrigerators and freezers with variable-speed compressor drives, with a specified power (per 6.3.2) of 600 W or lessMedical devices are not listed. A medical display that works much like a computer monitor needs its own classification argument

Class D limits are reserved for equipment with a pronounced effect on the public supply, so a medical device cannot be put in Class D just because it "looks like a computer". Three other situations often cause confusion:

  • External power supplies. Under clause 5.3 and Annex B.17, a supply designated for specific models is tested together with that equipment under its conditions; a generic supply is classified by the equipment types named in its instructions for use.
  • Multifunction equipment. Under clause 6.5, independent functions that can run alone may be tested one by one. Otherwise all functions run together against the most stringent relevant limits, or against the limits of a function that is clearly the main one.
  • Control methods. Clause 6.2 restricts asymmetrical control (motors only) and half-wave rectification on the mains, and sets power or limit conditions for symmetrical control of heating elements. Check it first if a warming or humidifying device uses phase control on its heater.

The 75 W threshold and other cases with no limits

Clause 7.1 lists the categories with no limits specified. Those relevant to medical devices are:

  • non-lighting equipment rated at 75 W or less (Note 2 says this may drop to 50 W in the future);
  • professional equipment with a total rated power above 1 kW. IEC clause 3.9 defines this as equipment for trades, professions or industries, not intended for sale to the general public, as designated by the manufacturer; the GB text uses the GB/T 4365-2003 definition;
  • symmetrically controlled heating elements with a controlled active input power of 200 W or less.

Clause 6.1 extends the clause 6.2 restrictions to these categories, and in the Figure 1 flowchart such equipment "conforms" rather than being "out of scope". The 75 W threshold uses rated power, while the Class D ceiling and limit calculation use the 6.3.2 power: the maximum 1.5 s smoothed active power over the DFT windows. The manufacturer may specify a value within ±10 % of the measurement, and both values go into the report. If the nameplate is close to 75 W, measure the actual active power first.

Test conditions: supply, operating mode, observation period

Measurement circuit and supply source

Annex A gives the measurement circuits; the instrument shall comply with IEC 61000-4-7:2002 and its 2008 amendment. Under A.2 of the IEC text, the test voltage is the rated voltage, or 230 V single-phase or 400 V three-phase for a rated range, held within ±2.0 %, with frequency within ±0.5 % and capped supply voltage harmonics. The GB text uses 220 V and 380 V. Table 1 of IEC 60601-1-2, for equipment rated 220 V to 240 V or 380 V to 415 V, uses the single rated voltage, or 230 V single-phase or 400 V three-phase for a range, at 50 Hz or 60 Hz.

Operating mode

Annex B gives special conditions for television receivers, lighting, IT equipment, external power supplies and other products (Amendment 2 confirms they take precedence), but none for medical devices. Otherwise clause 6.3.1 applies. User controls or automatic programs are set to the normal operating mode expected to produce the maximum total harmonic current (THC). This defines the set-up; it does not require measuring THC or searching for the worst case. Limits apply to line currents (single-phase equipment may be measured in the neutral), and motor drives may need a preliminary run.

Measurement and observation period

Under clause 6.3.2, each harmonic is measured as a 1.5 s smoothed r.m.s. value in each discrete Fourier transform (DFT) window and averaged over the observation period. Under clause 6.3.3.3, the first 10 s after switching on or off are disregarded, and stand-by mode may not exceed 10 % of any observation period. Table 4 sets the period:

Equipment behaviourObservation period (Table 4, summarised)
Quasi-stationaryLong enough to meet the repeatability recommendation in 6.3.3.1
Short cyclic (cycle of 2.5 min or less)At least 10 cycles (reference method), or long enough or synchronised
RandomLong enough to meet the repeatability recommendation
Long cyclic (cycle longer than 2.5 min)Full program cycle (reference method), or a representative 2.5 min period expected to give the highest THC

The repeatability recommendation only helps set the period; it is not a pass/fail criterion. For cyclic devices, give the laboratory the cycle length and the highest-power phase.

Limits and how results are assessed

  • Class A: each harmonic must stay within the fixed values in amperes in Table 1.
  • Class B: Table 1 values multiplied by 1.5.
  • Class C: mostly a percentage of the fundamental current (Table 2). Lighting from 5 W to 25 W has three alternative sets of requirements.
  • Class D: Table 3 gives a per-watt limit (middle column, applied to the 6.3.2 power) and an absolute value (right column). Amendment 2 makes clear that the lower one applies.

Under clause 6.3.3.4, the average of each harmonic over the observation period must not exceed the limit, and every 1.5 s smoothed value must be at or below 150 % of it. Class A equipment may reach 200 % if the time above 150 % is less than 10 % of the observation period or 10 min in total, whichever is smaller, and the average is below 90 % of the limit. Harmonic currents below 0.6 % of the input current or 5 mA, whichever is greater, are disregarded; Amendment 2 limits this to the comparison of individual harmonics. Odd harmonics of order 21 and above have a separate exemption based on partial odd harmonic current (POHC), and the two exemptions cannot be combined. For the limit values themselves, refer to the standard.

Amendment 2 also added clause 8.2. With an IEC 61000-4-7 Class I instrument, results are compared directly with the limits and no further uncertainty calculation is needed. With a Class II instrument, a decision rule based on clause 4.3.4 of IEC Guide 115:2023 applies, along with a detailed uncertainty analysis. Clause 8 of GB 17625.1-2022 has no such provision, which is one more reason to agree on the edition before testing.

After a failure: fixing the input stage with PFC

Because Class A limits are fixed values in amperes (Table 1), it follows that a higher-power design with the same input circuit usually draws larger harmonic currents and comes closer to them. The points below come from power supply vendors' application notes, not from the standard: a rectifier feeding a bulk filter capacitor draws a pulsating current dominated by odd harmonics.

  • Passive PFC: a line-frequency inductor works with the filter capacitor to widen the conduction angle and cut the peak current. It is simple, but the inductor is bulky.
  • Active PFC: a boost PFC stage after the bridge rectifier makes the input current follow the voltage waveform. It is effective but more complex, and its high-frequency switching can add conducted and radiated EMI.
  • A supply or adapter with built-in PFC: a supply designated for the model is tested with the equipment under Annex B.17, so a change of supply needs reassessment.
  • Control method and test mode: if a heating load is regulated by phase control or half-wave rectification, check clause 6.2 first. Confirm the test uses the highest-THC mode and record it in the test plan.

Keep a record of the fix. In Table 10 of IEC 60601-1-2 (minimum test report contents), item 34 is the measured harmonic and flicker emissions, and item 35 is a description of modifications made to pass the tests, with a statement that they will be incorporated into production units. Input-stage changes often affect other emission tests, and the laboratory scopes the retest accordingly; see EMC Failures: Locating the Cause, Fixing It, Scoping the Retest. Voltage fluctuations and flicker (IEC 61000-3-3) are covered in a separate article.

Accreditation scope table

Every row comes from the IAS accreditation scope of Hefei SUNGO Medical Technology Co., Ltd. (TL-1119).

ItemClauseNotes / limitations
IEC 61000-3-2 Electromagnetic compatibility (EMC) - Part 3-2: Limits - Limits for harmonic current emissions (equipment input current ≤ 16 A per phase)Whole standardNo year listed; no limitation listed
EN IEC 61000-3-2 Electromagnetic compatibility (EMC) - Part 3-2: Limits - Limits for harmonic current emissions (equipment input current ≤ 16 A per phase)Whole standardNo year listed; no limitation listed
GB 17625.1 Electromagnetic compatibility - Limits - Part 1: Limits for harmonic current emissions (equipment input current ≤ 16 A per phase)Whole standardNo year listed; no limitation listed
IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and testsWhole standardNo year listed; no limitation listed
EN 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and testsWhole standardNo year listed; no limitation listed
YY 9706.102 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic compatibility - Requirements and testsWhole standardNo year listed; no limitation listed

Where SUNGO does this, and under which accreditation

The test is carried out by Hefei SUNGO Medical Technology Co., Ltd., a separate legal entity whose laboratory is in Zipeng Town, Feixi County, Hefei, Anhui Province. The standards above are covered at the Hefei laboratory within its IAS accreditation (TL-1119), and reports are IAS-accredited reports. The IAS scope lists these standards without a year, so please confirm the edition with the laboratory before an IAS report is issued against a specific one, such as GB 17625.1-2022 or a consolidated edition of IEC 61000-3-2.

Shanghai SUNGO Medical Technology Co., Ltd. (laboratory at No. 958 Kangqiao East Road, Pudong, Shanghai) is a separate legal entity whose scope does not include these standards. Whether a report is accepted for registration in China or by an overseas regulator or certification body depends on the receiving party; please check with them first. For an overview of our EMC services, see the EMC testing service page.

Preparing your submission

  • Ratings: rated voltage or range, frequency, current and power; single-phase or three-phase; both long-time and momentary current ratings if they exist.
  • Environment and supply: home healthcare or professional facility, public mains connection, internal power only or not.
  • Proposed class: the clause 5 class, and the reasoning for any clause 7.1 exception.
  • Operating mode: the highest-THC mode and how to set it, program cycle, stand-by state, preliminary run.
  • Power supply and functions: external supply model and whether it is designated for this device; whether functions can run alone; any specified power value.
  • Standards and editions: which editions of the harmonic standard and of IEC 60601-1-2 or YY 9706.102 apply, the target market, and the harmonic class you plan to declare.

FAQ

Does a device rated at 60 W still need harmonic testing?

Under clause 7.1, non-lighting equipment rated at 75 W or less has no limits and conforms per Figure 1, but clause 6.2 still applies. Agree the report wording with the laboratory.

Can a medical device be assessed as Class D?

Class D covers only the three listed product types, so medical devices are normally Class A. A medical display resembling a computer monitor needs its own argument.

Does a device used only in hospitals need the test?

Under note b) to Table 2 of IEC 60601-1-2, not unless the equipment connects to the public mains network and its power input is within the scope of the basic standard. YY 9706.102 likewise applies to equipment intended for the public mains network. Record the reasoning in the test plan.

What about a device rated at 110 V?

IEC 61000-3-2 has not yet considered limits below 220 V line-to-neutral, and Annex A of IEC 60601-1-2 exempts such systems from clause 7.2.1. A wide-range device that also covers 220 V class supplies is still assessed under 7.2.1.

We only changed the power board. Do we have to repeat the full test?

Clause 6.1 allows a simplified method for equipment that undergoes minor changes or updates, if the original full test showed emissions below 60 % of the limits and a supply current THD below 15 %: the updated device complies if its active input power is within ±20 % of the original and its THD is still below 15 %. If in doubt, the full test prevails. The laboratory decides from the original report and the change whether this route applies.

To request a quote, send the model, ratings, proposed class, target standards and editions, and target market through our contact page. The laboratory will confirm the edition, test conditions and report format before quoting.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.