ASTM F2407 Surgical Gown Physical Tests: D5034, D5587 and D1683 Methods

ASTM F2407 Surgical Gown Physical Tests: D5034, D5587 and D1683 Methods

For US-bound surgical gowns, the first question is usually the PB70 barrier level. ASTM F2407 also sets tensile, tear and seam strength requirements that apply to every gown regardless of barrier level, mainly tested by ASTM D5034, D5587 and D1683. This guide covers specimen size, clamping, gauge length, speed and direction for each method, seam failure, tear direction, lint reporting (EN ISO 9073-10), and the items and report types in SUNGO Shanghai's accreditation.

Key points

  • PB70 for barrier level, F2407 for physical strength. F2407 clause 7.1 requires barrier testing and reporting per ANSI/AAMI PB70 (the -23 text cites PB70-2012 there; clause 10.1 gives no year); clause 7.2 requires physical testing per Table 1. Under clause 4.5.1, every barrier level has the same strength minimums.
  • Three properties and their methods. Tensile by D5034 (grab), seam by D1683/D1683M. For tear, Table 1 of F2407/F2407M-23 lists D5587 or D5733; clause 10.2.2 assigns D5587 to woven and D5733, withdrawn in 2008, to nonwoven fabrics. Table 1 sets ≥30 N tensile, ≥10 N tear and ≥30 N seam strength for barrier Levels 1 to 4. As listed by ASTM on 2026-10-22, the active edition is -23a; check the -23a text before ordering.
  • Judged by direction and seam type. Tensile and tear are judged on the machine direction (MD) and cross-machine direction (CD) averages separately; either failing fails the material. Seams are judged per seam type; any failing type fails the seam.
  • Lint is documented only. F2407 lists lint as an optional documentation test with no minimum.
  • SUNGO Shanghai scope. The CNAS schedule lists tensile, tear (D5587), seam and lint under F2407/F2407M-23a clause 7.2 and three liquid barrier items under clause 7.1. The IAS scope lists F2407/F2407M, D5034, D5587, D1683/D1683M, D3776/D3776M, ANSI/AAMI PB70 and other whole standards. None is on the CMA schedule.

Standards and scope

ASTM F2407/F2407M

F2407, Standard Specification for Surgical Gowns Intended for Use in Healthcare Facilities, is under ASTM Committee F23 on Personal Protective Clothing and Equipment. As listed by ASTM on 2026-10-22, the active edition is F2407/F2407M-23a, preceded by -06, -20, -22, -22a and -23. It covers the performance, documentation and labeling of surgical gowns (clause 1.1). Clause 4.2 excludes nonsurgical protective clothing such as isolation and decontamination gowns; isolation gowns fall under ASTM F3352/F3352M, outside SUNGO's accredited scope.

PB70 assigns the barrier level (Level 1 to 4), with barrier testing covering impact penetration, hydrostatic resistance and viral penetration. F2407 references that classification and adds design criteria, minimum physical performance, labeling and documentation (clause 4.5). The strength minimums rest on an analysis of workplace requirements and do not vary by level (4.5.1).

The FDA Recognized Consensus Standards database entered F2407/F2407M-23a on 2025-12-22 with partial recognition; subclause 1.1 including Notes 1 and 2 is not recognized. Listed product codes include FYA (surgical gown, 21 CFR 878.4040). Per the entry, declarations of conformity to F2407-20 are accepted until 2027-12-19.

The three methods

Editions below are those listed as active by ASTM on 2026-10-22.

  • ASTM D5034 (breaking strength and elongation, grab test), D5034-21(2025). The grab procedure applies to woven, nonwoven and felted fabrics; it is not recommended for glass fabrics or high-stretch fabrics over 11 % (1.1.1, 1.2).
  • ASTM D5587 (tearing strength, trapezoid procedure), D5587-15(2024). It applies to most fabrics, including woven, knitted, layered and nonwoven; because the tear is started before testing, the result is not directly related to tear initiation (1.2, 1.3).
  • ASTM D1683/D1683M (failure in sewn seams of woven fabrics), D1683/D1683M-22. Force is applied perpendicular to the seam; the method does not predict actual wear performance (1.3).

Although D1683 is titled for woven fabrics, F2407 clause 10.2.3 applies it to gown woven and nonwoven materials and materials incorporating such layers. Seams in knit or stretch woven materials are tested by ASTM D751 with a ring clamp, outside SUNGO's accredited scope.

How F2407 clause 7.2 uses the methods

Checked against the F2407/F2407M-23 text (Table 1 values are the same in -22a). SUNGO's CNAS schedule cites -23a, a second 2023 revision, so confirm against -23a.

  • 7.2: single-use and multiple-use gowns are tested per Table 1; multiple-use gowns must meet the minimums after one washing/drying and sterilization cycle and after the manufacturer's maximum number of cycles.
  • Note 9: tensile and tear requirements apply to both MD and CD; seam strength is tested perpendicular to the seam, in locations where material orientations differ.
  • 8.3: material, seam and attachment specimens are cut from manufactured gowns; each seam type is tested separately (8.3.1).
  • 9.2: conditioning at (21 ± 3) °C and (65 ± 5) % RH for at least 24 h per ASTM D1776/D1776M, unless the test method specifies otherwise.
  • 10.2.1 and 10.2.2: tensile and tear are run on each material, excluding fastening elements such as hook-and-loop tape, snaps, belts, ties and cuffs; separable layers are tested combined. Results are reported to the nearest 0.5 N and judged on the MD and CD averages separately. Tear: D5587 Option 1 for woven fabrics; D5733 (withdrawn 2008 per the -23 reference list) for nonwovens, which Table X1.1 says report the maximum recorded force. For nonwoven gowns, agree the method and calculation with the laboratory first.
  • 10.2.3: seam results are reported to the nearest 0.5 N, averaged and judged per seam type.
  • 7.3 and Table 2: lint is an optional documentation test per ISO 9073 Part 10 (10.3.1).

Test conditions compared

Values from the D5034-21, D5587-15 and D1683/D1683M-22 texts; D5034-21(2025) and D5587-15(2024) are reapprovals.

MethodSpecimenClamping and gauge lengthSpeedResult
D5034 grab (G)(100 ± 1) mm wide, at least 150 mm long, long side parallel to test direction (9.2.1)Front jaw face (25 ± 1) mm wide; gauge (75 ± 1) mm (6.2.1, 10.1.2)(300 ± 10) mm/min unless otherwise specified (10.1.4)Breaking force; apparent elongation if requested
D5587 trapezoid75 mm × 150 mm with an isosceles trapezoid; 15 mm cut at the center of the 25 mm edge (Fig. 2, 7.3.2.3)Clamp faces at least 50 mm × 75 mm; gauge (25 ± 1) mm (6.2, 8.1)(300 ± 10) mm/min (8.3)Average of five highest peaks (Option 1) or single peak (Option 2)
D1683 seam from manufactured items(350 ± 3) mm across the seam (250 mm one side, 100 mm the other), (100 ± 3) mm along the stitch line (7.4)Grab-type jaws; gauge (75 ± 3) mm, seam centered (10.3.2, 10.6)(305 ± 10) mm/min (6.1)Maximum seam strength and failure type; seam efficiency optional

D5034 grab test: "effective strength"

A 100 mm wide specimen is gripped centrally by a 25 mm front jaw, so fabric beside the gripped yarns also carries load. Clause 5.7 calls this effective strength and notes no simple relationship with strip tests, so D5034 results cannot be compared directly with strip methods such as EN ISO 9073-3.

  • Direction: long side parallel to MD or CD, specimens spread along a diagonal, no nearer the edge than one tenth of the width (9.1.1); five MD and eight CD specimens per laboratory sample and condition (7.3).
  • Alignment: a line drawn (37 ± 1) mm from one edge is aligned with the jaw edge when mounting (9.2.2, 11.1) so both clamps grip the same yarns.
  • Machine: CRE is preferred; in disputes, unless agreed otherwise, use a constant time-to-break of (20 ± 3) s (5.6).
  • Jaw breaks: Note 10 suggests discarding breaks within 5 mm of the jaws that fall below 50 % of the average of the other breaks.

D5587 trapezoid tear: the direction is easy to misread

A trapezoid is marked on the specimen, the short base is cut at its center, and the nonparallel sides are clamped; as the clamps separate, the tear runs across the specimen width (4.1, 5.3). Clause 5.3 notes the method suits comparing fabrics, or directions within one fabric.

The long direction is the test direction (7.3.1), and MD specimens have the long side parallel to MD (7.3.2). For woven fabrics the tear runs between the short-direction yarns rather than across them (7.3.2.1). For a woven fabric, then, "MD tear strength" breaks MD yarns one by one as the tear travels in CD; the reported direction is not the tear path.

Nonwoven fibers are often oriented, so MD and CD tear strength can differ markedly. F2407 judges each direction separately, so look at the weaker direction when selecting materials.

Clause 11.1 uses the average of the five highest peaks after the initial peak when there are five or more peaks (Option 1), otherwise the highest peak (Option 2); F2407 clause 10.2.2 specifies Option 1 for woven fabrics (nonwovens: see above). The test stops after about 75 mm of tear or complete tearing (10.4). If 25 % or more of specimens still break within 5 mm of the clamp edge after clamp modifications, the fabric is considered untearable by this method (10.4.2). The report states the calculation option and any yarn distortion (12.2.3).

D1683 seam strength: how did it fail?

Force is applied across the stitching until the stitching fails without fabric damage, or yarn displacement causes fabric failure (4.1). The cause is recorded as fabric yarn rupture, thread rupture, seam slippage or a combination (10.7.2), and the report gives each specimen's maximum seam strength, the average and the failure type (11.1, 12.2).

D1683 Table 1 defines default seams (Procedure A, stitch type 301 lockstitch; Procedure B, 401 chainstitch) for use when production seams are not available or when specified (8.2.2.1). Under F2407, seams are cut from finished gowns, so the production seam itself is tested. Where 350 mm cannot be cut, for example near cuffs or shoulders, purchaser and supplier agree a modification in advance (7.4.1); put it on the test request. Seam efficiency is 100 × seam strength ÷ fabric breaking force by D5034 (11.2); F2407 judges on newtons only, but efficiency shows whether the weak point is the seam or the material.

Why seams fail: stitches, welds and tape

D1683 clause 5.1 names four seam engineering variables: sewing thread, stitch type, seam type and stitch density. The failure mode points to what to change:

  • Thread breaks, fabric intact: the seam is weaker than the material; consider thread size or stitch density. D1683 Procedures A/B likewise retest with a changed thread tex size or stitch density when yarn displacement occurs (10.1.3.1 and 10.2.3.1). Stitch density is counted per centimetre (10.3.1).
  • Fabric breaks along the stitch line: denser stitching means more needle holes, where material can fail first, so more stitches do not always mean a stronger seam.
  • Seam slippage: mainly in woven fabrics; clause 11.3 gives a method based on seam opening.
  • Ultrasonic welds: D1683 covers sewn seams in woven fabrics (1.1) and defines failure types for them; F2407 clause 10.2.3 cites it for gown seams without separating sewn and welded seams. A weld has no thread, so agree with the laboratory how failures are described (for example bond separation or material breaking beside the weld).
  • Sewn and taped seams: needle holes can become liquid paths, so barrier-area seams are often heat-taped. Taped and untaped seams are generally treated as different seam types, tested separately under F2407 clause 8.3.1.

Note 9 does not name locations with differing material orientations; sleeve-to-body and back panel seams are common places to check (our examples, not the standard's text). Clause 8.3.2 exempts seams or attachment points less than 2 in. from the edge from barrier testing only; it does not mention strength testing.

Lint: EN ISO 9073-10 and F2407 reporting

ISO 9073-10:2003 (nonwovens, Part 10: lint and other particles generation in the dry state), first edition 2003-05-15 by ISO/TC 38, is adopted in Europe as EN ISO 9073-10:2004. It measures dry-state linting of nonwovens and may be applied to other textiles (clause 1). In a modified Gelbo Flex method, the specimen is twisted and compressed in a chamber while air is drawn off and particles are counted and classified, from 0.3 µm or 0.5 µm to 25 µm depending on the counter (clause 4).

F2407 clause 10.3.1 requires each side of each material to be tested for 5 min, or one side if both are the same; Table 2 asks for the particle count per side. A footnote warns that results depend heavily on operator and equipment and recommends one instrument, one operator and one day; compare across laboratories with care.

SUNGO's CNAS schedule lists lint under F2407/F2407M-23a clause 7.2 with EN ISO 9073-10:2004. In the F2407 text, lint is an optional clause 7.3 / Table 2 item; the clause cited on a CNAS report follows the schedule as written. EN ISO 9073-10 is not in the IAS scope.

Fabric weight: ASTM D3776

F2407 Tables 1 and 2 omit mass per unit area, though it is often used to describe materials and compare lots. ASTM D3776/D3776M has four options: A, full piece, roll or bolt; B, full-width sample; C, small swatch; D, narrow fabrics. Option C results apply to the sample only (ASTM abstract). SUNGO's D3776/D3776M accreditation is IAS only.

Accreditation scope table

Rows are taken from the CNAS schedule (L22083) and IAS scope (TL-1019) of Shanghai SUNGO Medical Technology Co., Ltd., with items, clauses and editions as listed. All CNAS rows are under the test object "surgical gown".

ItemStandard and clauseNotes / limitation
Synthetic blood penetrationASTM F2407/F2407M-23a clause 7.1; AAMI PB70:2022; ASTM F1670/F1670M-24CNAS; no limitation stated
Hydrostatic pressureASTM F2407/F2407M-23a clause 7.1; AAMI PB70:2022; AATCC TM127-2017(2018)eCNAS; no limitation stated
Resistance to water penetrationASTM F2407/F2407M-23a clause 7.1; AAMI PB70:2022; AATCC TM42-2017eCNAS; no limitation stated
Tensile strengthASTM F2407/F2407M-23a clause 7.2; ASTM D5034-21(2025)CNAS; no limitation stated
Tear strengthASTM F2407/F2407M-23a clause 7.2; ASTM D5587-15(2024)CNAS; no limitation stated
Seam strengthASTM F2407/F2407M-23a clause 7.2; ASTM D1683/D1683M-22CNAS; no limitation stated
LintASTM F2407/F2407M-23a clause 7.2; EN ISO 9073-10:2004CNAS; no limitation stated
Whole standardASTM F2407/F2407MIAS; no edition year
Whole standardANSI/AAMI PB70IAS; no edition year
Whole standardAATCC TM 42; AATCC TM 127; ASTM F1670/F1670MIAS; no edition year
Whole standardASTM D5034; ASTM D5587; ASTM D1683/D1683MIAS; no edition year
Whole standardASTM D3776/D3776MIAS; no edition year; not on CNAS or CMA
Whole standardASTM F1671/F1671MIAS; no edition year; not on CNAS or CMA

Where the testing is done and under which accreditation

Testing is at the Shanghai laboratory of Shanghai SUNGO Medical Technology Co., Ltd., 958 Kangqiao East Road, Pudong, Shanghai (CNAS L22083, CMA 240015340087, IAS TL-1019). Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity whose scope does not include surgical gowns. Report types:

  • Clause 7.1 barrier items (synthetic blood penetration, hydrostatic pressure, resistance to water penetration): CNAS under F2407/F2407M-23a clause 7.1 or AAMI PB70:2022 (methods F1670/F1670M-24, AATCC TM127 and AATCC TM42 respectively; editions in the table), or IAS under F2407/F2407M and ANSI/AAMI PB70. CNAS covers only these three PB70 items, and the viral penetration method referenced by PB70:2022, ASTM F1671/F1671M, is IAS only, so a full Level 1 to 4 PB70 classification report can only be issued under IAS.
  • The four items listed under clause 7.2 on the CNAS schedule (tensile, tear, seam, lint; in the F2407 text lint belongs to clause 7.3 / Table 2): CNAS. Tensile, tear and seam can also be reported under IAS against F2407/F2407M, D5034, D5587 and D1683/D1683M; EN ISO 9073-10 (lint) is not in the IAS scope.
  • Fabric weight, D3776/D3776M: IAS only.
  • CMA: not available for these standards; it covers only listed items and is valid only within the People's Republic of China.
  • IAS edition: the IAS scope carries no edition years; confirm the edition (for example F2407-20 or -23a) with the laboratory before ordering.
  • Biocompatibility (F2407 clause 6.1): biological tests such as cytotoxicity, sensitization and irritation are subcontracted by SUNGO to suitably qualified laboratories. Under accreditation rules, subcontracting requires the client's prior consent, and the report identifies the subcontracted items and the subcontractor.

CNAS is a signatory to the ILAC and APAC mutual recognition arrangements, and the IAS certificate carries the ilac-MRA mark. Whether an overseas regulator accepts a given report depends on that regulator's requirements.

Preparing samples

  • Send finished gowns. F2407 clause 8.3 requires specimens cut from manufactured gowns; roll goods alone cannot be used for seams.
  • Describe the construction: materials in critical and noncritical zones, any separable layers, and which parts are fasteners and cuffs.
  • List the seams: each seam type (sewn, ultrasonic, sewn and taped) and its location, ideally on a marked-up drawing.
  • Mark material direction on the sample or matching fabric if MD and CD cannot be identified on the gown.
  • Multiple-use gowns: per clause 7.2, prepare separate samples after one washing/drying and sterilization cycle and after the manufacturer's maximum number of cycles, and state the procedure and cycle counts.
  • Sterile gowns: send sterilized finished product and state the sterilization method.
  • Edition, report type and quantity: state the F2407 edition and CNAS or IAS. Quantities follow the method standards and F2407 sampling; F2407/F2407M-23 clause 8.1 takes physical property samples from different garments and at least three nonconsecutive lots. Check -23a and confirm quantities with the laboratory.

FAQ

We already have PB70 Level 3 data. Do we still need F2407 physical testing?

Yes. PB70 covers barrier classification only; F2407 clause 4.5.1 applies the same strength minimums to every level, so an F2407 claim needs the three Table 1 results.

MD passes and CD fails. Can they be averaged?

No. MD and CD averages are judged separately, and failing either direction fails the material.

Can D1683 be used for nonwoven gown seams?

Yes. F2407 clause 10.2.3 applies D1683/D1683M to woven and nonwoven gown materials. Knit or stretch woven materials use ASTM D751, outside SUNGO's accredited scope.

Can these items be issued under CMA?

No. None of these standards is on SUNGO Shanghai's CMA schedule; reports are CNAS or IAS as listed above.

Requesting a quote

Please include the product name and model, single-use or multiple-use, target barrier level, the items needed (F2407 clause 7.1, clause 7.2 or full PB70 classification), the F2407 edition, the report type, and the seam list and construction details. See our surgical gown testing service and send your request through the contact page.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.