Lab1 Shanghai SUNGO Lab Lab2 Lab3 Lab4 Hefei SUNGO Lab HefeiCMA HefeiCNAS HefeiIAS
WeChat WeChat QR Code
Contact
+86 132 4819 8029 (Mr. Gan)
+86 177 5560 4650 (Mr. Luo)

Hydrostatic Pressure for Gowns: GB/T 4744, EN ISO 811 and AATCC 127 Compared

Hydrostatic Pressure for Gowns: GB/T 4744, EN ISO 811 and AATCC 127 Compared

The hydrostatic-type water penetration test for surgical gowns and drapes uses a different method in each major market. In SUNGO's accreditation scope, YY/T 0506.1-2023 A 2.3 (China) is listed alongside GB/T 4744-2013; EN 13795-1:2025 (EU) gives a dated reference to EN ISO 811:2018; and the hydrostatic pressure item of AAMI PB70 and ASTM F2407 (US) uses AATCC TM127, with PB70 also using AATCC TM42 impact penetration. Manufacturers registering in several markets ask whether one test can serve all three, and whether cmH2O, kPa and mbar values can be compared once converted. Using publicly available standard previews and primary sources (the GB/T 4744 details come from a preview of an English translation; the AATCC TM127 procedure is not public, so refer to the standard), this article compares rate of pressure increase, test area, end point and units, explains why results are not interchangeable, shows how to specify the method, and explains how TM42 differs from the hydrostatic test. Performance levels and pass/fail values are outside its scope. Standard information was checked on 2026-10-22.

Key points first

  • Same principle, different parameters. Each method applies rising water pressure to one face and records the pressure when water comes through. They differ in rate, water temperature, direction of pressure, drop counting, clamp-edge rules and units.
  • ISO 811 itself has two rates. ISO 811:2018 allows 10 or 60 cmH2O/min and water at 20 °C or 27 °C. It notes that different rates may give different results and warmer water gives lower values; the choices must be reported.
  • GB/T 4744 modifies ISO 811:1981. Its main text specifies 6.0 kPa/min (60 cmH2O/min) and the pressure at the third drop; another rate must be reported. These points come from a public preview of the English translation and remain subject to the original text.
  • AATCC TM127 parameters: refer to the standard. AATCC's public summary gives only the scope: water resistance under hydrostatic pressure, tested without a restraint (AATCC TM208 if a restraint is needed; TM208 is not within SUNGO's accreditation scope), with results that may differ from rain or spray tests.
  • Unit conversion is not method equivalence. 1 cmH2O is about 0.098 kPa or 0.98 mbar. A converted result still reflects its own method's conditions. Test by the method the target product standard cites.
  • AATCC TM42 is not a hydrostatic test. It measures water penetrating under spray impact, in grams. Per the US NIOSH summary of the PB70:2012 table, Level 1 uses AATCC 42 only and Levels 2 and 3 use AATCC 42 and AATCC 127; for the 2022 edition, refer to the standard.
  • SUNGO Shanghai: GB/T 4744-2013 is under CNAS and CMA for surgical drapes, gowns and clean air suits; EN ISO 811:2018 is under CNAS for surgical gowns and drapes; AATCC TM127-2017(2018)e and TM42-2017e are under CNAS for surgical gowns only, and the IAS scope also lists both as whole standards without years. The last three are not in the CMA scope. The Hefei laboratory's scope does not include these tests.

Standards and scope

The three hydrostatic standards

ItemGB/T 4744-2013ISO 811:2018 / EN ISO 811:2018AATCC TM127-2017(2018)e
TitleTextiles — Testing and evaluation for water resistance — Hydrostatic pressure methodTextiles — Determination of resistance to water penetration — Hydrostatic pressure testTest Method for Water Resistance: Hydrostatic Pressure
Developed byChina's national textile standards committee (TC209), basic standards subcommitteeISO/TC 38/SC 2; EN version with CEN/TC 248, approved by CEN on 2018-03-15AATCC
DatesIssued 2013-10-10, effective 2014-05-01; reviewed 2022-01-21, confirmed validISO 2nd edition March 2018; EN May 2018The edition on sale at AATCC as checked on 2026-10-22, and the one in SUNGO's CNAS scope
HistoryReplaces GB/T 4744-1997; modified adoption of ISO 811:1981; revision plan 20264143-T-608: status "being drafted" as of 2026-10-22Replaces ISO 811:1981; EN ISO 811:2018 adopts it unchanged and supersedes EN 20811:1992AATCC TM208 is the separate method with a restraint
ScopeAll fabrics, including composites, and their products; also evaluates water resistanceFabrics intended to be water resistant, finished or notAll fabrics, including those with water-resistant or water-repellent finishes

ISO 811:2018's foreword lists the changes from 1981 as the first element of the title, a clarified scope and editorial updates. GB/T 4744-2013 shares the 1981 origin but was modified on adoption; for example, per the translation preview, its main text keeps only one rate.

Which product standard cites which method

Product standardMethodNotes
YY/T 0506.1-2023Annex A 2.3; SUNGO's CNAS and CMA scopes list it under the same item as GB/T 4744-2013Issued 2023-01-13, effective 2024-01-15; replaces YY/T 0506.1-2005, 0506.2-2016 and 0506.7-2014. Test conditions in A 2.3: refer to the standard
EN 13795-1:2025Annex A 2.3; Clause 2 cites EN ISO 811:2018 (dated)ISO 811 options chosen in Annex A: refer to the standard
AAMI PB70:2022, ASTM F2407/F2407M-23aHydrostatic pressure by AATCC TM127; impact penetration by AATCC TM42; Level 4 adds blood or viral penetrationThe F2407 abstract lists impact penetration, hydrostatic resistance and viral penetration of critical zones; PB70 Levels 1 to 3 use water only, Level 4 needs simulated body fluid tests

In the NIOSH PPE-Info summary of the PB70:2012 table, Level 4 uses ASTM F1670 for drapes and ASTM F1671 for surgical and isolation gowns. NIOSH adds that no correlation can be drawn between the AATCC water tests and F1671. FDA lists PB70:2022 as partially recognized (recognition number 6-487, entered 2023-05-29). For each level in the 2022 edition, refer to the standard text.

The three methods compared

Parameters

Entries marked * come from a public preview of the English translation of GB/T 4744-2013 and are subject to the original text. The ISO 811 column follows the ISO 811:2018 preview. No public primary source was found for the AATCC TM127 procedure.

ItemGB/T 4744-2013ISO 811:2018AATCC TM127-2017(2018)e
Rate of increase6.0 ± 0.3 kPa/min (60 ± 3 cmH2O/min); another rate, e.g. 1.0 kPa/min, must be reported *10 ± 0.5 or 60 ± 3 cmH2O/min (9.8 or 58.8 mbar/min), one chosen and reported; the higher rate suits highly water-resistant fabricsRefer to the standard
Test areaRefer to the standard100 cm²; specimen flat, not bulging, no leaks or slippingRefer to the standard
Direction, restraintRefer to the standardFrom below or above, reportedNo restraint (TM208 with restraint)
WaterClean distilled or deionized *ISO 3696 grade 3, fresh per specimen; 20 ± 2 °C or 27 ± 2 °C, reportedRefer to the standard
AtmosphereRefer to the standard (its normative references list GB/T 6529) *ISO 139; ambient if agreedRefer to the standard
Side facing waterFace; if unclear, the coated side of single-coated fabrics, otherwise both sides tested and reported separately *Face; report the side if not the faceRefer to the standard
End pointThird drop just appearing *Water at the third place; first drop may be reported by agreementRefer to the standard
Counting, invalid testsNon-growing droplets ignored; repeat seepage at one place not added; a third drop at the clamp edge below the sample's lowest other value is discarded and retested *Non-growing droplets ignored; later drops at one place count once; any test with clamp-edge penetration rejected, more specimens tested until reproducibleRefer to the standard
ResultsResult unit and grading: refer to the standard (clause 8.3 in the preview gives the rate in kPa/min with cmH2O/min in brackets) *Individual results and mean in cmH2O or mbar; Pa noted as unusualRefer to the standard; NIOSH's PB70 summary gives requirements in cm

GB/T 4744-2013 in brief

Being a "testing and evaluation" standard, it adds a hydrostatic pressure rating (clause 3.2) and an evaluation clause (clause 9). Per the translation preview, it dropped the 1.00 kPa/min rate of ISO 811:1981 and GB/T 4744-1997 and asks that bursting, jetting or swelling of composite fabrics be recorded with the pressure. Subject to the original text.

ISO 811:2018 in brief

Because rate, temperature and direction are options, clause 11 requires them in the report, with the side tested, individual and mean results, and any end point other than the third place. For quality control or protective clothing barrier claims, the standard advises larger data sets with statistical design, citing ISO 2859-1.

AATCC TM127 in brief

Only the purpose and scope are public (see above). Rate, area and drop counting could not be checked, so they are not quoted here. NIOSH's PB70 summary expresses the AATCC 127 requirement in centimetres.

Why the three results cannot simply be converted

  1. Rate. ISO 811 states that different rates may give different results. GB/T 4744 specifies only 6.0 kPa/min (per the translation preview); ISO 811 offers two. Which rate EN 13795-1 or YY/T 0506.1 selects is set in their own texts.
  2. Water temperature. ISO 811 allows 20 °C or 27 °C; warmer water gives lower values, by a fabric-dependent amount.
  3. End point and invalid tests. ISO 811 rejects any clamp-edge penetration; GB/T 4744 discards only a low third drop at the edge (per the translation preview). ISO 811 also allows first-drop reporting by agreement.
  4. Direction and restraint. ISO 811 allows pressure from below or above; AATCC TM127 is unrestrained, TM208 is the restrained method.
  5. Criteria belong to the product standard. Each sets requirements for the method it cites. PB70 classification also uses a sampling plan (an FDA 2016 webinar on PB70:2003 refers to an AQL), not an average of a few specimens. Applying one method's result to another standard needs justification, and acceptance is for the reviewer, notified body or customer to decide.

"Resistance to water penetration" can mean two tests

In SUNGO's CNAS scope, this item under YY/T 0506.1-2023 and EN 13795-1:2025 is a hydrostatic method (GB/T 4744-2013, EN ISO 811:2018). For the test object "surgical gown" under ASTM F2407 and AAMI PB70, the same item name is paired with AATCC TM42 impact penetration, and AATCC TM127 sits under "hydrostatic pressure". Always give the method number and edition on the request form.

AATCC TM42 versus hydrostatic pressure

ItemAATCC TM42Hydrostatic methods
Water actionSpray impactConstant contact, rising pressure
Original purposeAATCC: predicting rain penetration resistance of garment fabricsISO 811: articles under water pressure for short or moderate periods
ResultWater penetrated, in gramsPressure, in cmH2O, mbar or kPa
Main influencesPer the AATCC TM42 summary: fibre and yarn repellency; fabric constructionPer the AATCC TM127 summary: fibre and yarn repellency; fabric construction
PB70 use (NIOSH, 2012 edition)Levels 1, 2 and 3AATCC 127 for Levels 2 and 3

Per NIOSH, the two expose the fabric to water in different ways, and AATCC notes that TM127 results may differ from rain or spray tests. One cannot replace the other.

How to specify the method

  1. Choose the target standard: YY/T 0506.1-2023, EN 13795-1:2025, a level of AAMI PB70:2022, or ASTM F2407/F2407M-23a.
  2. Write clause and method edition, e.g. "EN 13795-1:2025 A 2.3, by EN ISO 811:2018" or "AAMI PB70:2022 Level 3, hydrostatic pressure by AATCC TM127-2017(2018)e", plus any method options (ISO 811 rate, temperature, direction) unless the product standard fixes them.
  3. For a fabric used in several markets, test by each cited method rather than converting.

Scope of accreditation

Relevant lines from the three scopes of Shanghai SUNGO Medical Technology Co., Ltd. CNAS and CMA editions are as listed; lines without a limitation carry none. IAS lists whole standards without years. Test objects: A, surgical drapes, gowns and clean air suits for patients, clinical staff and equipment; B, surgical gowns and drapes; C, surgical gown.

ItemStandard and clauseNotes / limitations
Water penetration resistanceYY/T 0506.1-2023 A 2.3CNAS L22083; object A
Water penetration resistanceGB/T 4744-2013CNAS L22083; object A; no clause listed
Water penetration resistanceEN 13795-1:2025 A 2.3CNAS L22083; object B
Water penetration resistanceEN ISO 811:2018CNAS L22083; object B; no clause listed; not in CMA; no separate IAS line
Hydrostatic pressureASTM F2407/F2407M-23a 7.1CNAS L22083; object C
Hydrostatic pressureAAMI PB70:2022CNAS L22083; object C; no clause listed
Hydrostatic pressureAATCC TM127-2017(2018)eCNAS L22083; object C; no clause listed
Water penetration resistanceASTM F2407/F2407M-23a 7.1CNAS L22083; object C
Water penetration resistanceAAMI PB70:2022CNAS L22083; object C; no clause listed
Water penetration resistanceAATCC TM42-2017eCNAS L22083; object C; no clause listed
Synthetic blood penetrationAAMI PB70:2022CNAS L22083; object C; same item: ASTM F1670/F1670M-24. PB70:2022 appears in the CNAS scope only under synthetic blood penetration, hydrostatic pressure and water penetration resistance
Water penetration resistanceYY/T 0506.1-2023CMA 240015340087; object A; no clause listed
Water penetration resistanceGB/T 4744-2013CMA 240015340087; object A
Whole standardAATCC TM 127IAS TL-1019; no year
Whole standardAATCC TM 42IAS TL-1019; no year
Whole standardANSI/AAMI PB70IAS TL-1019; no year
Whole standardASTM F2407/F2407MIAS TL-1019; no year
Whole standardASTM F1671/F1671MIAS TL-1019; no year; viral penetration for PB70 Level 4 gowns, not covered here
Whole standardEN 13795-1IAS TL-1019; no year
PartYY/T 0506.1 Annex A,BIAS TL-1019; as written: Accredited only for Annex A,B

Where SUNGO performs these tests and under which accreditation

The tests listed in the scope table above are performed by Shanghai SUNGO Medical Technology Co., Ltd. at 958 Kangqiao East Road, Pudong, Shanghai (CNAS L22083, CMA 240015340087, IAS TL-1019). Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity whose scopes do not include these tests. See our surgical gown and drape testing service.

  • YY/T 0506.1-2023 A 2.3 and GB/T 4744-2013: CNAS and CMA, test object surgical drapes, gowns and clean air suits. CMA covers only items in its scope and is valid within the People's Republic of China. IAS lists YY/T 0506.1 Annexes A and B (no year) but not GB/T 4744 separately; confirm with the laboratory before requesting an IAS report.
  • EN ISO 811:2018: CNAS, test object surgical gowns and drapes; not in the CMA scope, and no separate line in the IAS scope.
  • EN 13795-1:2025 A 2.3: CNAS, test object surgical gowns and drapes; not in the CMA scope. IAS lists EN 13795-1 as a whole standard without a year; confirm the edition with the laboratory before an IAS report is issued.
  • AATCC TM127-2017(2018)e, AATCC TM42-2017e, and hydrostatic pressure and water penetration resistance under ASTM F2407/F2407M-23a 7.1 and AAMI PB70:2022: CNAS, test object surgical gown only, limited to these editions. IAS lists all four as whole standards without years; confirm the edition before an IAS report is issued. None is in the CMA scope.
  • Full PB70 classification: CNAS covers only three PB70:2022 items, for surgical gowns only, and ASTM F1671 for Level 4 gowns is in the IAS scope only, so full classification can be reported only under IAS. As IAS lists no years, confirm the edition with the laboratory before an IAS report is issued.
  • Acceptance: the CNAS certificate states CNAS is an ILAC and APAC mutual recognition member, and the IAS certificate bears the ilac-MRA mark. Whether a regulator, notified body or customer accepts a report depends on their requirements.

Preparing samples

  • Request form: standard, edition and options as above, plus the report type (CNAS, CMA or IAS).
  • Keep material flat: ISO 811 asks for minimal handling, no sharp folds and no treatment such as ironing other than conditioning; deeply creased areas are not tested. Ship flat or rolled.
  • Mark the face: ISO 811 and GB/T 4744 put the outer face in contact with water. If the face is unclear, describe the construction, e.g. single-side coating.
  • Seams: when seams are tested under GB/T 4744-2013, the seam sits mid-specimen (per the translation preview; subject to the original text), so mark seam positions on the sample. For other methods, check with the laboratory first whether seams can be tested.
  • Zones: label critical and less critical zone materials separately.
  • Reusable products: EN 13795-1:2025 requires compliance throughout the useful life; for post-reprocessing tests, state the process and cycles and send samples in that state.
  • Quantities: ISO 811 calls for at least five specimens from different places, and PB70 classification involves sampling; agree quantities with the laboratory beforehand.

Frequently asked questions

We have an EN ISO 811 report. Do we need GB/T 4744 testing for China?

In SUNGO's scope, YY/T 0506.1-2023 A 2.3 is listed alongside GB/T 4744-2013, and the two methods differ in rate options and clamp-edge rules (the GB/T 4744 points come from the translation preview and are subject to the original text). We recommend testing by the cited method; whether an existing report is usable is for the reviewing authority to decide.

How do cmH2O, mbar and kPa convert?

1 cmH2O is about 0.098 kPa or 0.98 mbar; ISO 811 equates 60 cmH2O/min with 58.8 mbar/min. Conversion does not change test conditions.

Is AATCC TM127 alone enough for a PB70 Level 2 or 3 claim?

No. Per the NIOSH summary of the PB70:2012 table, Levels 2 and 3 need both AATCC 42 and AATCC 127; for the 2022 edition, refer to the standard.

Two labs got very different results on one fabric. Is that normal?

Compare each report's method edition, rate, water temperature, direction, side tested and end point (third place or first drop). Results under different conditions should not be compared directly.

Can SUNGO report all PB70 levels?

Only under IAS: the Shanghai IAS scope lists ANSI/AAMI PB70 and ASTM F1671 as whole standards without years, so the edition is confirmed first. The CNAS scope covers only three PB70:2022 items.

For a quotation, tell us the product and material construction, the target market and standard (YY/T 0506.1, EN 13795-1, a PB70 level or ASTM F2407), the method edition and options, the report type (CNAS, CMA or IAS), whether the samples include seams (for methods other than GB/T 4744-2013, check first), and the processing state of samples of reusable products. Please use our contact page.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.