Active medical devices with compressors, heaters or frequently cycling motors usually have a "voltage fluctuations and flicker" line in their EMC emission tests: IEC 61000-3-3 / EN 61000-3-3 internationally, GB/T 17625.2 in China. The test asks whether changes in input current, acting on the supply impedance, make lighting on the same network flicker visibly. This article covers the parameters and limits, reference impedance and observation time, when medical EMC standards call for the test, which products tend to fail, and fixes. Harmonic current emissions (IEC 61000-3-2) are covered in a separate article.
Key points
- All limits apply together, except where the standard excludes them (such as manual switching, where Pst and Plt are not applied, and specific Annex A equipment). Clause 5 of IEC 61000-3-3: Pst not greater than 1.0; Plt not greater than 0.65; Tmax, the accumulated time during a single voltage change for which d(t) deviates by more than 3.3 %, not more than 500 ms; steady-state change dc not more than 3.3 %; maximum change dmax limited to 4 %, 6 % or 7 % depending on how the equipment is switched.
- The scope has edges. Equipment drawing no more than 16 A per phase, intended for public low-voltage systems of 220 V to 250 V line to neutral at 50 Hz, and not subject to conditional connection.
- For medical devices, start with public mains connection. YY 9706.102-2021 subclause 6.1.3.2 applies to ME equipment and ME systems rated at no more than 16 A per phase and intended for connection to the public mains network. In IEC 60601-1-2 Edition 4.1 the topic is subclause 7.2.2 "Voltage fluctuations and flicker"; check its text for the exact conditions.
- Align editions first. As of October 2026, the IEC webstore lists consolidated Edition 3.2 (2013 third edition plus two amendments). GB/T 17625.2-2007 is an identical adoption of IEC 61000-3-3:2005.
- Failures usually trace to inrush current and repeated switching. Typical fixes are soft starting, inrush limiting, staggered loads and finer heater control.
- SUNGO performs this test at the Hefei laboratory, within its IAS accreditation (TL-1119).
Standards and scope
| Standard | Edition information | Role |
|---|---|---|
| IEC 61000-3-3 | Edition 3.0 published 2013-05-14; Amendment 1 2017-05; Amendment 2 2021-03-25, consolidated as Edition 3.2, which includes corrigendum 1 (2022-01) and interpretation sheet 1 (2025-04). A product family standard | Limits, assessment methods, test conditions |
| EN 61000-3-3 | EN 61000-3-3:2013 with A1:2019 and A2:2021 (including the 2022 corrigendum) | European adoption |
| GB/T 17625.2-2007 | Recommended national standard, issued 2007-04-30, effective 2008-01-01, replacing GB 17625.2-1999; identical adoption of IEC 61000-3-3:2005 | Chinese counterpart |
| IEC 60601-1-2 | Edition 4.1 consolidated version, published 2020-09-01 | Subclause 7.2.2 "Voltage fluctuations and flicker"; dated reference to IEC 61000-3-3:2013 |
| YY 9706.102-2021 | Issued 2021-03-09, effective 2023-05-01; modified adoption of IEC 60601-1-2:2007 | Subclause 6.1.3.2 says when GB/T 17625.2 applies |
Under Clause 1, IEC 61000-3-3 specifies limits on the voltage changes equipment may produce under specified test conditions and gives assessment methods. It applies to equipment with an input current of no more than 16 A per phase, intended for public low-voltage distribution systems of 220 V to 250 V line to neutral at 50 Hz, and not subject to conditional connection; the tests are type tests. Note 2 says the limits are based mainly on the subjective severity of flicker from 230 V 60 W coiled-coil filament lamps, and that for systems below 220 V line to neutral and/or at 60 Hz the limits and reference circuit values are still under consideration. The IEC webstore gives a stability date of 2026 and, as of October 2026, shows Edition 4.0 under development; confirm the edition in force when ordering.
"Conditional connection" (definition 3.13) means equipment that needs a supply impedance at the user's interface point lower than the reference impedance Zref to meet the limits. Clause 1 says equipment that cannot be declared compliant when tested with Zref may be retested or evaluated against IEC 61000-3-11 (rated input current up to 75 A per phase, subject to conditional connection). That standard is not discussed here.
After the 2005 text adopted by GB/T 17625.2-2007, IEC published a second edition in 2008 and the third in 2013, whose significant technical change is to take account of IEC 61000-4-15:2010 (the flickermeter specification). For China, GB/T 17625.2-2007 itself governs.
When the medical collateral standards require it
YY 9706.102-2021, 6.1.3.2.1: ME equipment and ME systems rated at no more than 16 A per phase and intended for connection to the public mains network shall comply with GB/T 17625.2; where there are long-time and momentary current ratings, the higher one decides applicability. 6.1.3.2.2 applies the test methods and equipment of GB/T 17625.2. The Annex A rationale says the requirement applies only to equipment rated at 220 V or more and intended for public mains connection. Homes, doctors' offices and small clinics are examples of public mains locations, while sites such as hospitals, isolated from the public low-voltage network by transformers or substations, are not considered connected. Table 1, the emissions declaration, has a flicker row completed as "complies" or "not applicable", so the conclusion appears in the accompanying documents.
IEC 60601-1-2 Edition 4.1 places voltage fluctuations and flicker in 7.2.2, next to harmonic distortion under 7.2 "Protection of the public mains network". Its normative references cite IEC 61000-3-3:2013 as a dated reference, and Clause 2 says only the cited edition applies to dated references. Table 2 summarises emission requirements by environment. For the exact conditions, read 7.2.2 and Table 2; confirm the IEC 61000-3-3 edition for the target market when ordering.
What each parameter measures
Subclause 4.1: the voltage change at the terminals of the equipment under test comes from the change in its fundamental input current acting on the reference impedance. For single-phase and symmetrical three-phase equipment, ΔU ≈ ΔIp·R + ΔIq·X, the active and reactive parts of the current change times the resistance and reactance of the reference impedance, and d = ΔU/Un. Larger, faster and more frequent current changes give worse results.
| Parameter | Definition (Clause 3) | Limit (Clause 5) |
|---|---|---|
| d(t) | Time function of the relative r.m.s. voltage change, one value per half period between zero crossings, except while the voltage is steady for at least 1 s | No limit of its own; used for Tmax |
| Tmax | Maximum time in the observation period that d(t) exceeds the dc limit; mainly intended to evaluate the inrush current pattern | Accumulated time with d(t) deviating by more than 3.3 % during a single voltage change: not more than 500 ms |
| dc | Maximum steady-state voltage change during the observation period | Not more than 3.3 % |
| dmax | Maximum absolute voltage change during the observation period | 4 %, 6 % or 7 %, by switching type |
| Pst | Short-term flicker severity; evaluated over 10 min unless specified otherwise | Not greater than 1.0 |
| Plt | Long-term flicker severity: cube root of the mean of the cubes of N consecutive Pst values | Not greater than 0.65 |
Percentages are relative to the nominal test voltage Un; if no steady state is reached in the observation period, Un is used for dmax and Tmax.
The three dmax limits
- a) 4 %: without additional conditions.
- b) 6 %: equipment switched manually, or switched automatically more than twice per day with a delayed restart (not less than a few tens of seconds) or manual restart after a supply interruption. A note warns that Pst and Plt still limit the cycling rate: a 6 % rectangular change twice per hour gives a Plt of about 0.65.
- c) 7 %: equipment attended while in use (for example hair dryers, vacuum cleaners, mixers, electric drills), or switched on automatically, or intended to be switched on manually, no more than twice per day, with a delayed or manual restart after a supply interruption.
The next paragraph of Clause 5, as amended by Amendment 1, says: for equipment with several separately controlled circuits, b) and c) apply only with a delayed or manual restart after an interruption; equipment with automatic switching that is energised immediately when supply returns falls under a); manually switched equipment takes b) or c) according to the switching rate typical of normal operation (Amendment 1 added "typical of normal operation"). Pst and Plt are not applied to voltage changes from manual switching, and the limits do not apply to emergency switching or emergency interruptions.
How Pst and Plt are obtained
Subclause 4.2 lists four methods: a flickermeter on line, for any fluctuation; simulation when d(t) is known; an analytical method for the shapes in Figures 3 to 5 occurring less than once per second; and the Pst = 1 curve of Figure 2 for equal rectangular changes at equal intervals. Direct measurement with a flickermeter to IEC 61000-4-15:2010 is the reference method and decides in case of doubt.
The analytical method converts each change into a flicker impression time, tf = 2.3(F·dmax)^3.2, with shape factor F (1.0 for a step; from Figures 3 to 5 for double-step, ramp, rectangular, triangular and motor-start characteristics that match a figure, as extrapolation outside the figures gives unacceptable errors), then Pst = (Σtf/Tp)^(1/3.2). Results are expected within ±10 % of direct measurement, and the method is not used when the gap between changes is under 1 s. Plt uses N = 12 consecutive Pst values, which at 10 min each covers 2 h; 4.3 notes that Plt generally needs assessing for equipment normally operated for more than 30 min at a time.
Test conditions: impedance, observation time, operating state
Under 6.1, the test circuit is the test supply, the reference impedance, the equipment under test and, if needed, a flickermeter, as in Figure 1 (reference network for single-phase and three-phase supplies derived from a three-phase, four-wire supply).
- Reference impedance Zref: represents the supply impedance at the interface between the public network and the user's installation, made up of resistance and reactance; values are in 6.4 and Figure 1. IEC TR 60725:2012 notes that residential reference impedances were based on the 90 % column of a 1980 survey, and that, with supply impedances reduced overall, those values had become more relevant to the 95 % column. Note 1 of Clause 1 warns that a higher actual impedance inside the user's installation can still lead to disturbance above the limits.
- Observation time: 10 min for Pst and 12 Pst periods for Plt; how Tp is chosen is set out in 6.5. The test plan should state the operating mode and the on/off sequence within the observation period.
- Whether testing is needed: under 6.1, equipment unlikely to produce significant fluctuations or flicker need not be tested, judged from the circuit diagram, specification and a short functional test. For manual switching, equipment is deemed to comply if the maximum r.m.s. input current including inrush, evaluated over each 10 ms half period between zero crossings, does not exceed 20 A and the current after inrush stays within a 1.5 A band. If measurement is used, dmax from manual switching is measured per Annex B, which Amendment 1 says may also be used for dc and Tmax.
- Operating state: Annex A gives test conditions for equipment such as cookers, washing machines, refrigerators, water heaters, air conditioners and heat pumps, fan heaters and arc welders. Medical devices are not listed, so the mode and observation time go into the test plan under Clause 6; Annex A is only a reference for similar loads. For example, after Amendment 2, refrigerators are evaluated over at least three compressor switch-on and switch-off events without Pst or Plt, and air conditioners and heat pumps use the declared cycles per hour.
Which medical devices tend to fail
This draws on the principle in 4.1 and the load types in Annex A; it is not text from the standard, and test results decide.
- Compressor devices such as oxygen concentrators: starting current far exceeds running current, giving a large, sustained voltage drop that mainly affects dmax and Tmax; if input current rises and falls with the operating cycle, Pst builds up too. Other test topics are in Oxygen Concentrator Testing: ISO 80601-2-69, IEC 60601-1 and IEC 60601-1-2.
- Heating devices: high-power elements switched by a thermostat cause a near-rectangular voltage change each time. Step size drives dmax and dc, repetition rate drives Pst, and long operation brings in Plt.
- Motors that start and stop often: pumps and motorised mechanisms cycling under program control draw inrush at every start, and Pst builds with cycling rate.
- Large input capacitance: capacitor charging at power-up is also inrush; for manual switching, check the 20 A / 1.5 A condition first.
After a failure: inrush and switching first
These are common engineering approaches, not requirements of the standard; retesting decides whether a change works.
- Soft start: drive compressors, fans and pumps with a soft starter or variable-frequency drive so current ramps up, lowering dmax and shortening the time d(t) stays above 3.3 %.
- Inrush limiting: a limiting resistor or NTC thermistor at the input, bypassed by a relay after start-up, or an active circuit controlling capacitor charging.
- Staggered loads: do not switch several motors or heaters on at once, so one large change becomes several small ones; recheck Pst, as there are now more changes.
- Finer heater control: switch large heaters in stages to reduce the power switched per step, or adjust the switching period. Changing the power control method (for example to phase-angle control) affects harmonics, so recheck them.
- Restart logic: for automatically switched equipment, b) and c) require a delayed or manual restart after a supply interruption; equipment energised immediately when supply returns is held to a), 4 %. Manually switched and attended equipment fall under their own branches of b) and c). Whether to add a restart delay must be weighed against clinical use.
Keep records. The laboratory sets the retest scope from what changed; see EMC Failures: Locating the Cause, Fixing It, Scoping the Retest. For how emission reports are layered, see EMC Emission Reports: GB 4824 Sits under a Higher Layer.
Accreditation scope
Every row comes from the IAS accreditation scope (TL-1119) of Hefei SUNGO Medical Technology Co., Ltd.; titles are as listed on the certificate.
| Item | Clause | Notes / limitations |
|---|---|---|
| IEC 61000-3-3 Electromagnetic compatibility (EMC) - Part 3-3: Limits - Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems, for equipment with rated current ≤ 16 A per phase and not subject to conditional connection | Entire standard | No edition year on the certificate; no limitation listed |
| EN 61000-3-3 (same title as above) | Entire standard | No edition year on the certificate; no limitation listed |
| GB/T 17625.2 Electromagnetic compatibility (EMC) - Limits - Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems, for equipment with rated current ≤ 16 A per phase and not subject to conditional connection | Entire standard | No edition year on the certificate; no limitation listed |
| IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | Entire standard | No edition year on the certificate; no limitation listed |
| EN 60601-1-2 (same title as above) | Entire standard | No edition year on the certificate; no limitation listed |
| YY 9706.102 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic compatibility - Requirements and tests | Entire standard | No edition year on the certificate; no limitation listed |
Where SUNGO tests and under which accreditation
The test is carried out by Hefei SUNGO Medical Technology Co., Ltd., a separate legal entity with its laboratory in Zipeng Town, Feixi County, Hefei, Anhui Province. The standards above are tested at the Hefei laboratory, within its IAS accreditation (TL-1119), and IAS-accredited reports are issued. The IAS scope gives no edition years for these standards; before an IAS report is issued against a specific edition (for example IEC 61000-3-3 Edition 3.2 or GB/T 17625.2-2007), please confirm the edition with the laboratory.
The accredited items cover equipment rated at no more than 16 A per phase and not subject to conditional connection; evaluating conditionally connected equipment under IEC 61000-3-11 is outside SUNGO's accreditation. Shanghai SUNGO Medical Technology Co., Ltd. (laboratory at Kangqiao East Road 958, Pudong, Shanghai) is another separate legal entity whose scope does not include these standards. Whether a report can be used for registration in China or accepted by an overseas regulator or certification body depends on that body; please confirm with it before ordering. Service overview: medical EMC testing.
Preparing for the test
- Ratings: rated voltage or range, frequency, rated input current (long-time and momentary if both exist), rated power, single- or three-phase.
- Use and supply: home, clinic or hospital; intended public mains connection; internal power source and use while charging.
- Loads and control: number and power of compressors, motors and heaters; starting method; heater control method; and whether the device restarts automatically, after a delay, or manually when supply returns.
- Operating cycle: typical modes, switching frequency, normal continuous run time, and the mode expected to give the largest voltage changes.
- Power-up inrush: design values or measurements, to judge whether the 6.1 deemed-to-comply condition applies.
- Standards and editions: which edition of IEC 61000-3-3, EN 61000-3-3 or GB/T 17625.2, the matching IEC 60601-1-2 or YY 9706.102 edition, the target market, and how the flicker row in the accompanying documents will be completed.
FAQ
Does a device rated 100–240 V need flicker testing?
IEC 61000-3-3 covers equipment intended for 220 V to 250 V line-to-neutral, 50 Hz public networks, so a wide-range device intended for such a network is in scope. The test supply voltage follows 6.3; confirm it with the laboratory when ordering.
Does a device used only in hospitals need the test?
Under YY 9706.102 Annex A, sites such as hospitals isolated from the public low-voltage network are not considered connected; equipment not intended for public mains connection is outside the requirement. For IEC 60601-1-2, follow 7.2.2 and Table 2. Record the conclusion and reasoning in the test plan and accompanying documents.
We have a GB/T 17625.2 report. Must we retest to IEC 61000-3-3 for export?
GB/T 17625.2-2007 adopts IEC 61000-3-3:2005, not Edition 3.2 of IEC 61000-3-3. Whether the report can be reused depends on the edition the target market requires; confirm with the receiving body in advance.
To request a quote for flicker testing, please send the product name and model, ratings, main loads, operating cycle and switching pattern, standards and editions, and target market through the contact page. The laboratory will confirm the applicable edition, test conditions and report format before quoting.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.