Phones, two-way radios, Wi-Fi tablets and Bluetooth devices are routinely used right next to patient monitors, infusion equipment and hand-held instruments. IEC 60601-1-2 covers this in subclause 8.10 and Table 9 (immunity to proximity fields from RF wireless communications equipment), and much of what circulates about its frequencies, levels and distance does not match the standard. This article walks through Table 9 as amended, how it differs from general radiated immunity, why hand-held and bedside equipment struggles with it, IEEE/ANSI C63.18, and where SUNGO does this work.
Key points
- Table 9 was not introduced in Edition 4.1. Edition 4.0 (February 2014) already had subclause 8.10 and Table 9. Amendment 1 (September 2020) replaced Table 9 in full and added subclause 8.11 and Table 11 for proximity magnetic fields.
- The amended Table 9 has 15 test frequencies from 385 MHz to 5 785 MHz and only three test levels: 9, 27 and 28 V/m. Most bands use 18 Hz or 217 Hz square-wave pulse modulation; 450 MHz uses FM.
- The amended Table 9 has no "0.3 m" column. A statement within the table allows the antenna-to-equipment distance to be reduced to 1 m if needed to reach the test level, as IEC 61000-4-3 permits. The field comes from a transmitting antenna, not a probe swept over the enclosure.
- IEEE/ANSI C63.18 is an on-site, ad hoc estimation method for common transmitters and cannot replace Table 9 type testing.
- The seven standards in the accreditation scope table below (IEC/EN 60601-1-2, YY 9706.102, IEC/EN IEC 61000-4-3, GB/T 17626.3, IEEE/ANSI C63.18) are within the Hefei laboratory's IAS accreditation scope (TL-1119). The scope states no edition years, so the edition for an IAS report is confirmed with the laboratory beforehand.
Standards and scope
IEC 60601-1-2, EN 60601-1-2 and YY 9706.102
IEC 60601-1-2 is the collateral standard on electromagnetic disturbances for medical electrical equipment. Edition 4.0 was published on 2014-02-25; with Amendment 1 (September 2020) it forms consolidated Edition 4.1 (2020-09-01), still the latest version on the IEC Webstore as of 2026-10-22. In the Edition 4.0 contents, subclause 8.10 is "Immunity to proximity fields from RF wireless communications equipment" and Table 9 gives its test specifications. The introduction notes that such equipment can no longer be prohibited from most patient environments.
Amendment 1 replaces the existing Table 9 with a new table, adds IEC 61000-4-39:2017 to the normative references, and adds subclause 8.11 (proximity magnetic fields, 9 kHz to 13.56 MHz) with Table 11, which has its own article.
EN 60601-1-2:2015 is based on IEC 60601-1-2:2014, and EN 60601-1-2:2015/A1:2021 on IEC Amendment 1. The Chinese counterpart is YY 9706.102-2021, a mandatory industry standard published on 2021-03-09, in force since 2023-05-01, replacing YY 0505-2012. Check the standard itself for its IEC correspondence and proximity field wording.
IEC 61000-4-3
IEC 61000-4-3 is the basic EMC publication for radiated RF field immunity. Edition 4.0 (2020-09-08) replaced the 2006 edition and its 2007 and 2010 amendments. The 2020 edition's scope covers RF sources not in close proximity to the equipment under test and refers close-proximity testing to IEC 61000-4-39. Its European adoption is EN IEC 61000-4-3:2020. The Chinese national standard number is GB/T 17626.3; check its text for the IEC edition it follows.
Note that IEC 60601-1-2 Edition 4.1 cites IEC 61000-4-3:2006 with Amendments 1:2007 and 2:2010. For dated references only the cited edition applies, so the 2020 edition is not substituted automatically. The scope of that cited edition (consolidated Edition 3.2) covers RF electromagnetic fields from any source, without the "not in close proximity" limitation.
IEEE/ANSI C63.18
IEEE/ANSI C63.18-2014 (full title in the scope table below) was approved by ANSI on 2014-05-16 and published on 2014-06-20, superseding C63.18-1997. The FDA recognized consensus standards database lists it with complete recognition (recognition number 19-17).
Table 9 frequency by frequency
This is the Table 9 introduced by Amendment 1; for Edition 4.0 testing, use the original Edition 4.0 table.
| Test frequency (MHz) | Band (MHz) | Service | Modulation | Immunity test level (V/m) |
|---|---|---|---|---|
| 385 | 380 to 390 | TETRA 400 | Pulse modulation, 18 Hz | 27 |
| 450 | 430 to 470 | GMRS 460, FRS 460 | FM, ±5 kHz deviation, 1 kHz sine | 28 |
| 710, 745, 780 | 704 to 787 | LTE Band 13, 17 | Pulse modulation, 217 Hz | 9 |
| 810, 870, 930 | 800 to 960 | GSM 800/900, TETRA 800, iDEN 820, CDMA 850, LTE Band 5 | Pulse modulation, 18 Hz | 28 |
| 1 720, 1 845, 1 970 | 1 700 to 1 990 | GSM 1800; CDMA 1900; GSM 1900; DECT; LTE Band 1, 3, 4, 25; UMTS | Pulse modulation, 217 Hz | 28 |
| 2 450 | 2 400 to 2 570 | Bluetooth, WLAN, 802.11 b/g/n, RFID 2450, LTE Band 7 | Pulse modulation, 217 Hz | 28 |
| 5 240, 5 500, 5 785 | 5 100 to 5 800 | WLAN 802.11 a/n | Pulse modulation, 217 Hz | 9 |
The statement in the table and its footnotes a) to c):
- Statement: if necessary to achieve the immunity test level, the distance between the transmitting antenna and the ME equipment or ME system may be reduced to 1 m; the 1 m test distance is permitted by IEC 61000-4-3.
- Footnote a): for some services, only the uplink frequencies (those the terminal transmits on) are included.
- Footnote b): the carrier is modulated with a 50 % duty cycle square wave, so "18 Hz pulse modulation" means the carrier is switched on and off by an 18 Hz square wave.
- Footnote c): instead of FM, the 450 MHz carrier may be pulse modulated with a 50 % duty cycle square wave at 18 Hz; the standard notes this does not represent actual modulation but would be worst case.
5G does not appear in the service column. The June 2022 FDA EMC guidance lists cellular 5G among common emitters not adequately addressed by FDA-recognized consensus standards at the time, to be considered in risk management.
Test distance, antenna and exposed faces
The amended Table 9 has no 0.3 m column
Some material describes Table 9 as "applied at 0.3 m", or as sweeping a near-field probe over the enclosure. The table as replaced by Amendment 1 has five columns only (test frequency, band, service, modulation, test level); the only distance is the 1 m in the statement within the table, which concerns the distance between the transmitting antenna and the equipment. The test setup follows subclause 8.10 and the cited edition of IEC 61000-4-3. For Edition 4.0 testing, take column and distance meanings from the Edition 4.0 table and subclause 8.10.
Uniform field area, setup and exposed faces
IEC 61000-4-3 sets the test field over a uniform field area (UFA). In the 2020 edition, subclause 6.3 covers the UFA and two level-setting methods (constant field strength, constant power), Clause 7 the test setup, and Annex G cable layout. Exposed faces and cable routing follow subclause 8.10 of IEC 60601-1-2 and the cited edition of IEC 61000-4-3, whose clause numbering may differ.
For hand-held products, also check subclause 8.4 of IEC 60601-1-2, "Hand-held ME equipment and parts intended to be hand-held". Confirm before submission which parts are intended to be hand-held.
How Table 9 differs from the general radiated immunity test
Both are enclosure port radiated RF immunity tests within the IEC 61000-4-3 framework, so they are often confused. They differ in three ways:
- Frequencies: the general test sweeps a range in steps; Table 9 applies only 15 fixed frequencies, each tied to a real class of wireless service.
- Modulation: Table 9 specifies modulation per service. Informative Annex A of IEC 61000-4-3 discusses the rationale for choosing modulation for digital radio services.
- Scenario: Table 9 uses the IEC 61000-4-3 method for proximity fields from RF wireless communications equipment (subclause 8.10), listing two-way radio, cellular, WLAN and Bluetooth services. The scope of IEC 61000-4-3 itself depends on the edition: the 2020 edition excludes close-proximity sources, the cited 2006 edition covers any source (see above).
Take the range, levels and modulation of the general radiated test from IEC 60601-1-2 itself (a separate article covers it). The two tests are specified separately; neither substitutes for the other.
Why hand-held and bedside equipment has more trouble with this test
The Table 9 scenario is a wireless device in use next to the equipment: staff radios, phones and tablets around a bed, or a phone in the same hand as a hand-held device. Common engineering reasons include:
- The frequencies sit on common transmitter bands: the 27 and 28 V/m points cover the usual bands of two-way radios, mobile phones, DECT, Bluetooth and 2.4 GHz WLAN.
- Low-frequency pulse modulation is easily "demodulated": semiconductor junctions at amplifier inputs or sensor interfaces can rectify the 18 Hz or 217 Hz on/off envelope into the signal path. Circuits measuring slowly varying quantities (physiological signals, pressure, flow, temperature) and audible alert circuits are often sensitive to this.
- Cables and leads act as antennas: long leads, sensor, power and communication cables on bedside equipment readily pick up RF energy from a few hundred MHz to several GHz. Subclauses 8.2 and 8.3 of IEC 60601-1-2 cover patient physiological simulation and termination of patient-coupled parts, so lead arrangement matters.
- Hand-held construction: small plastic housings leave little room for shielding and filtering, and touch screens, display drivers and battery management circuits sit close to the enclosure.
- Built-in radios: the 2 450 MHz and 5 GHz points fall in the operating bands of built-in Bluetooth or Wi-Fi, so the link may drop; whether that is a failure depends on whether the link carries essential performance.
For troubleshooting a failure, see EMC Failures: Locating the Cause, Fixing It, Scoping the Retest.
Pass/fail logic: back to basic safety and essential performance
Immunity testing under IEC 60601-1-2 is judged against basic safety and essential performance. The June 2022 FDA EMC guidance recommends acceptance criteria that are quantitative, specific to the device and its functions, observable, and documented in the test plan before testing. For continuous phenomena such as radiated RF, specified performance should be provided during and after the disturbance. Wireless technology used for the intended use should be on and communicating, and monitoring should observe performance quantitatively without significantly perturbing the test.
For Table 9: at each frequency, do displayed values, outputs or alarms exceed the deviation defined in advance; are there false alarms, shutdowns, resets or altered settings; and does the device recover by itself afterwards?
IEEE/ANSI C63.18: ad hoc assessment on site
Table 9 tests a model at fixed frequencies and levels. A hospital wants to know whether its own radios, phones and RFID readers will disturb a particular device; C63.18 addresses that.
Scope
- What it evaluates: immunity of medical devices to radiated RF emissions from common transmitters, such as two-way radios, mobile phones, wireless-enabled tablets and laptops, RFID readers, two-way pagers and wireless PDAs.
- Where it is used: existing, newly purchased or prepurchase devices in health-care facilities; it can be adapted to home or mobile health-care settings.
- Where it does not apply: implantable devices such as pacemakers and defibrillators; transport environments such as ambulances and helicopters; and transmitters rated above 8 W output. Testing above 8 W in health-care facilities is not recommended because of possible adverse effects on critical-care devices in use nearby.
- What it does not address: in-band interference where a transmitter's fundamental frequency overlaps frequencies used by a hospital wireless network, monitoring system or other medical device wireless links.
Transmitters, distances, steps and records
For transmitter selection, distances, steps and recording, follow the standard text; they are not covered here. For any on-site assessment, we suggest recording at least the following, for review and for facility policy:
- Transmitter: type, model, band, rated output power, and whether it was transmitting.
- Device: name, model, software version, operating mode, connected patient accessories or simulators.
- Site: location, other equipment in use nearby, date and personnel.
- Result: distance, orientation and symptoms when an anomaly occurred; otherwise the closest distance tried.
It does not replace type testing
The C63.18 title says "estimating", and its scope states that it is not a comprehensive test and offers no guarantee. Results depend on the transmitters, site and device state and cannot replace Table 9 type testing. Equally, passing Table 9 does not guarantee freedom from interference everywhere; the FDA EMC guidance notes that consensus standards may not keep up with a rapidly changing EM environment.
Accreditation scope: Hefei laboratory, IAS (TL-1119)
Each row corresponds to one of the seven relevant entries in the IAS scope (TL-1119) of Hefei SUNGO Medical Technology Co., Ltd. For these seven entries the certificate lists only the standard number and title (reproduced in the "Item" column), with no separate clauses, no limitations and no edition year.
| Item | Basis | Notes / limitations |
|---|---|---|
| Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | IEC 60601-1-2 (no clauses listed separately) | No limitation stated; no edition year |
| Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | EN 60601-1-2 (no clauses listed separately) | No limitation stated; no edition year |
| Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic compatibility - Requirements and tests | YY 9706.102 (no clauses listed separately) | No limitation stated; no edition year |
| Electromagnetic compatibility (EMC) - Part 4-3: Testing and measurement techniques - Radiated, radio-frequency, electromagnetic field immunity test | IEC 61000-4-3 (no clauses listed separately) | No limitation stated; no edition year |
| Electromagnetic compatibility (EMC) - Part 4-3: Testing and measurement techniques - Radiated, radio-frequency, electromagnetic field immunity test | EN IEC 61000-4-3 (no clauses listed separately) | No limitation stated; no edition year |
| Electromagnetic compatibility - Testing and measurement techniques - Part 3: Radiated, radio-frequency, electromagnetic field immunity test | GB/T 17626.3 (no clauses listed separately) | No limitation stated; no edition year |
| American National Standard Recommended Practice for an On-Site, Ad Hoc Test Method for Estimating Electromagnetic Immunity of Medical Devices to Radiated Radio-Frequency (RF) Emissions from RF Transmitters | IEEE/ANSI C63.18 (no clauses listed separately) | No limitation stated; no edition year |
Where SUNGO does this work and under which accreditation
IEC 60601-1-2 (including Table 9), IEC 61000-4-3 radiated immunity and IEEE/ANSI C63.18 are handled by the laboratory of Hefei SUNGO Medical Technology Co., Ltd. (an independent legal entity; laboratory in Zipeng Town, Feixi County, Hefei, Anhui Province), with reports issued within its IAS accreditation scope (TL-1119). The address on the certificate is the Hefei laboratory; where and how a C63.18 assessment is carried out should be confirmed with the laboratory before the job is accepted. Shanghai SUNGO Medical Technology Co., Ltd. is a separate legal entity with its laboratory at 958 Kangqiao East Road, Pudong, Shanghai; its accreditation scope does not include the EMC standards discussed here.
The IAS scope states no edition years, while IEC 60601-1-2 exists as Editions 4.0 and 4.1 and IEC 61000-4-3 as the 2006 (amended) and 2020 editions; confirm the edition with the laboratory before acceptance. SUNGO makes no commitment that any regulator will accept a report; whether it can support registration or certification in a market, including NMPA registration testing, should be confirmed with the relevant authority or notified body in advance.
Preparing for testing
- Standard and edition: IEC 60601-1-2 Edition 4.0 or 4.1, EN 60601-1-2 or YY 9706.102.
- Use environment: professional, home healthcare or special, and the wireless devices expected nearby.
- Emitters outside Table 9: whether site emitters absent from Table 9 (5G, for example) need separate evaluation or extra testing is the manufacturer's risk management decision, recorded in the test plan.
- Essential performance and acceptance criteria: quantitative and observable, with permitted deviations, alarm behaviour and recovery requirements.
- Monitoring: how values are read during the test, in real time and without disturbing it.
- Operating modes: all modes, which are tested and why.
- Wireless functions: bands, power and other parameters plus the communication partner; keep the link on and active during testing.
- Accessories and cables: cables, leads, sensors and patient-simulating loads specified in the accompanying documents, with lengths and layout.
- Hand-held parts: which parts are intended to be hand-held.
Frequently asked questions
Was Table 9 added in IEC 60601-1-2 Edition 4.1?
No. Edition 4.0 (2014) already had subclause 8.10 and Table 9. Amendment 1 (2020) replaced Table 9 and added subclause 8.11 and Table 11 for proximity magnetic fields.
Is Table 9 proximity field immunity tested with a near-field probe against the enclosure?
No. The statement in the amended Table 9 concerns the transmitting antenna-to-equipment distance, which may be reduced to 1 m if necessary, as IEC 61000-4-3 permits. Proximity magnetic fields (subclause 8.11) differ: where testing is required, the field is applied with IEC 61000-4-39 methods to enclosure or accessory surfaces accessible during intended use (separate article).
My device has Bluetooth. If the link drops at 2 450 MHz, is that a failure?
It depends on whether the link carries essential performance and on the criteria set before testing. For non-critical data, specify an allowed interruption and recovery; for critical control or monitoring data, whether a drop means loss of essential performance depends on the risk analysis and pre-defined criteria, e.g. whether the device detects the loss and alarms or enters a safe state.
The hospital has done a C63.18 assessment. Do we still need Table 9?
Yes. By its own scope, C63.18 is not a comprehensive test and offers no guarantee, so it cannot replace the type testing required by IEC 60601-1-2.
Which markets can the report be used for?
SUNGO issues reports under the Hefei laboratory's IAS scope and makes no commitment about acceptance by any regulator, including NMPA; confirm with the authority in advance. The US FDA recognizes IEC 60601-1-2 Edition 4.1 in part (recognition number 19-36); the two parts not recognized do not concern Table 9.
To arrange Table 9 testing or a full IEC 60601-1-2 programme, send the product and intended use, standard and edition, use environment, essential performance and acceptance criteria, wireless parameters, operating modes and accessory list through the contact page; we will confirm the test plan and edition. See also EMC testing.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.