The voltage dips and short interruptions immunity test simulates a sudden sag or brief loss of the AC mains supply. A device under test may flicker, reboot, lose settings, or switch to its battery and alarm; which reactions are acceptable must be settled before testing. This article covers how IEC 61000-4-11 (GB/T 17626.11 in China) relates to IEC 60601-1-2 and YY 9706.102, how to read levels, cycle counts and phase angles, acceptable degradation, battery-backed devices, and where SUNGO tests.
Key points
- IEC 61000-4-11 is a basic EMC test-method standard for equipment with a rated input current of up to 16 A per phase connected to 50 Hz or 60 Hz AC networks. The levels, durations and phase angles for a medical device come from IEC 60601-1-2 or YY 9706.102 (as modified by clause 202 of any applicable particular standard), not from the environment classes of IEC 61000-4-11.
- Know your edition. As of October 2026, the latest IEC consolidated version is IEC 60601-1-2 Edition 4.1 (the 2014 fourth edition plus Amendment 1:2020). China's YY 9706.102-2021 is a modified adoption of IEC 60601-1-2:2007, the third edition. A result under one cannot stand in for the other.
- "25/30 cycles" means 25 cycles at 50 Hz and 30 cycles at 60 Hz, both 0.5 s. "250/300 cycles" works the same way, both 5 s.
- Pass/fail is judged against basic safety and essential performance, with criteria written into the test plan beforehand. For battery-backed devices, the battery state and whether a switchover alarm is expected must also be defined in advance.
- IEC 61000-4-11, EN IEC 61000-4-11, GB/T 17626.11, IEC 60601-1-2, EN 60601-1-2 and YY 9706.102 are all within the IAS accreditation (TL-1119) of the Hefei laboratory, listed as complete standards with no edition year.
Standards and scope
IEC 61000-4-11: a method, not product levels
IEC 61000-4-11 Edition 3.0 was published on 2020-01-28 by IEC SC 77A (EMC – Low frequency phenomena). It replaces the 2004 second edition and its Amendment 1:2017, and the edition text includes the corrigenda of May 2020 and October 2022. The foreword lists two significant technical changes: rise time and fall time are now defined terms in Clause 3, and Clause 4 now states the origin of dips and short interruptions.
Clause 1 limits the scope to equipment with a rated input current not exceeding 16 A per phase, for connection to 50 Hz or 60 Hz AC networks; equipment for 400 Hz networks is excluded. The standard calls itself a basic EMC publication: product committees decide whether the test applies and at what levels.
GB/T 17626.11 and EN IEC 61000-4-11
According to China's national standards information platform, GB/T 17626.11-2023 was issued on 2023-05-23, took effect on 2024-06-01 and replaces GB/T 17626.11-2008. It is a modified adoption of IEC 61000-4-11:2020; refer to its foreword for the technical deviations from the IEC text. The European counterpart is EN IEC 61000-4-11; refer to the CENELEC text for the edition.
Which edition applies: IEC 60601-1-2 and YY 9706.102
IEC 60601-1-2 is a collateral standard to IEC 60601-1, covering the basic safety and essential performance of medical electrical equipment and systems in the presence of electromagnetic disturbances, and the disturbances they emit. As of October 2026, the latest IEC consolidated version is IEC 60601-1-2:2014+AMD1:2020 CSV (Edition 4.1, published 2020-09-01); the European number is EN 60601-1-2. The FDA Recognized Consensus Standards database lists Edition 4.1 under recognition number 19-36 as a partial recognition; the two unrecognized parts are unrelated to dips and interruptions.
The fourth edition specifies immunity levels by environment of intended use (professional healthcare facility, home healthcare, special environments) and by port, based on the reasonably foreseeable maximum disturbance; some levels are higher than in the third edition. In Edition 4.1 the input a.c. power port is Table 5 and the input d.c. power port is Table 6. The normative references, as updated by Amendment 1, cite IEC 61000-4-11:2004 with Amendment 1:2017.
YY 9706.102-2021 is a mandatory Chinese industry standard (in force as of October 2026), issued on 2021-03-09, effective 2023-05-01, replacing YY 0505-2012. A draft national standard for comment, GB 9706.284—202X, published on the Shanghai Medical Products Administration website, lists it as a modified adoption of IEC 60601-1-2:2007, and its foreword states that YY 9706.102-2021 is used in place of IEC 60601-1-2:2014. The third edition puts dips and interruptions in two separate tables (Tables 10 and 11), unlike the port tables of the fourth edition.
How the test works: levels, cycles and phase angles
Definitions
Clause 3 of IEC 61000-4-11 defines a voltage dip as a sudden reduction of the voltage below a specified dip threshold, followed by recovery after a brief interval; a short interruption as the same on all phases below an interruption threshold; and residual voltage as the minimum RMS voltage recorded during the event. Clause 4 attributes them to faults in the network or installations, or to sudden changes of large loads. Levels are expressed relative to the rated voltage, written UT.
Reading the preferred test levels
Under 5.2, the change between UT and the changed voltage is abrupt and can start and stop at any phase angle. The test levels (residual voltages) are 0 %, 40 %, 70 % and 80 % of UT. A 0 % level is a total interruption, and in practice 0 % to 20 % of rated voltage may be treated as one. Tables 1 and 2 give the preferred levels below. Classes are the IEC 61000-2-4 environment classes (see Annex B); Class 1 levels are case-by-case; Class X is set by the product committee but, for equipment connected directly or indirectly to the public network, shall not be less severe than Class 2.
| Class | Voltage dips (residual voltage, duration in cycles, 50 Hz/60 Hz) | Short interruptions |
|---|---|---|
| Class 2 | 0 %, ½ cycle; 0 %, 1 cycle; 70 %, 25/30 cycles | 0 %, 250/300 cycles |
| Class 3 | 0 %, ½ cycle; 0 %, 1 cycle; 40 %, 10/12 cycles; 70 %, 25/30 cycles; 80 %, 250/300 cycles | 0 %, 250/300 cycles |
Values are from Tables 1 and 2 of IEC 61000-4-11:2020. The standard notes that these levels represent many real-world dips but do not guarantee immunity to all of them. The levels, durations and phase angles a medical device is actually tested at are those in Table 5 of IEC 60601-1-2 Edition 4.1 or the corresponding tables of YY 9706.102; refer to the original text. Where a particular standard modifies them in its clause 202, follow the particular standard.
25/30 and 250/300: cycles at 50 Hz and 60 Hz
Duration is counted in mains cycles. The table footnotes state that "25/30 cycles" means 25 cycles for a 50 Hz test and 30 cycles for a 60 Hz test, and "250/300 cycles" likewise. A cycle lasts 20 ms at 50 Hz and about 16.7 ms at 60 Hz, so 25 and 30 cycles are both 0.5 s, and 250 and 300 cycles are both 5 s. The ½-cycle and 1-cycle dips are not converted, so they are shorter in absolute time at 60 Hz.
Table 1 of IEC 60601-1-2 Edition 4.1 allows the dip and short-interruption tests at any one frequency within the rated range, with no retest at other frequencies. For a 50/60 Hz device, the report should state the chosen frequency.
Phase angle
Clause 5.2 of IEC 61000-4-11 says the step can start and stop at any phase angle on the mains voltage; Figure 1 shows a 70 % dip starting at 0° and a 40 % dip starting at 90° as examples. Test execution is covered in 8.3.2 (Voltage dips and short interruptions); refer to the original text. For medical devices, the phase angles to apply, including those for the half-cycle dip, are given in the original text of the IEC 60601-1-2 or YY 9706.102 edition you test to.
Clause 5.2 says that when setting performance criteria for half-period disturbances, particular attention should be paid to effects that can result from inrush currents: in products with a mains transformer, core flux saturation after a dip can drive inrush currents of 10 to 40 times the rated current, and products with capacitors (for example EMC filters, or bridge rectifiers feeding DC capacitors) can also draw high inrush currents. Decide in advance how the criteria treat, say, a fuse or protective device operating (an engineering example, not one the standard lists).
Which supply voltage
Clause 5.1 of IEC 61000-4-11 allows a single voltage as UT when the rated range does not exceed 20 % of the lower voltage; otherwise the lowest and highest declared voltages are both tested. Medical devices follow Table 1 of IEC 60601-1-2 Edition 4.1, which differs slightly: dips are tested at minimum and maximum rated voltage; if the difference between them is less than 25 % of the highest rated input voltage, any one rated voltage may be used; equipment with transformer-tap voltage selection is tested at one tap; short interruptions may be tested at any one voltage. Equipment with multiple voltage settings or auto-ranging is tested at the voltage specified in Table 1 (footnote r of Table 5, as amended).
Test generator
In IEC 61000-4-11, 6.1.2 and Table 4 specify the generator, including voltage rise and fall times during abrupt changes, measured between the 10 % and 90 % points of the transition. Subclause 6.1.3 covers generator verification. Normative Annex A covers peak inrush current drive capability, the current monitor and the EUT peak inrush current requirement. Refer to the original text for the numerical values.
Pass/fail: what degradation is acceptable
Clause 9 of IEC 61000-4-11 covers the evaluation of test results, but for a medical device the verdict rests on IEC 60601-1-2: basic safety and essential performance must be maintained. The fourth-edition foreword says the edition harmonizes better with the risk concepts of basic safety and essential performance, including deletion of the defined term "life-supporting", and adds guidance on identifying immunity pass/fail criteria. In the Edition 4.1 table of contents, that guidance is informative Annex I; test plan guidance is informative Annex G.
The third edition took another approach. Its introduction states that life-supporting equipment needs higher immunity to establish a broader safety margin even in the general medical environment, so the standard specifies additional requirements for it. When testing to YY 9706.102-2021, which adopts that edition with modifications, take those additional requirements from the YY text.
For drafting criteria, the FDA guidance on EMC of medical devices issued on 2022-06-06 is a useful reference. It recommends acceptance criteria documented in the test plan before testing that are quantitative, device-specific and observable, noting that generic labels such as "Performance Criterion A" are not specific enough. It classes voltage dips and interruptions as transient phenomena, for which it might be acceptable that the device provides the specified performance after the disturbance; a recovery time can be acceptable but should be specified based on the risk analysis.
For this test, go through each dip and interruption level required by your standard and spell out whether the device may reset and whether outputs stay within limits during dips; and, for an interruption lasting seconds that may shut down a device without an internal power source, whether shutdown is acceptable, whether recovery is automatic or needs the operator, and whether settings are retained. Base these on the standard and the risk analysis, and write them into the plan in advance. IEC TS 60601-4-2:2024 gives guidance on assessing performance in the presence of EM disturbances and is a helpful reference.
Devices with an internal battery
Start with the scope
IEC 61000-4-11 addresses equipment connected to low-voltage AC networks, so a device that runs only on an internal power source with no AC mains connection is outside its scope. If the device can be used while connected to the mains, including through an external adapter or while charging, dips and interruptions must be assessed in mains-powered operation. For the EMC programme as a whole, the FDA guidance recommends that a device able to operate in battery power mode and in mains power mode be tested in both modes. The input d.c. power port (Table 6 of IEC 60601-1-2) is not covered here.
Battery state: full is not necessarily worst case
In a battery-backed device, the battery may take over during dips and interruptions. A full battery mainly exercises the switchover circuit; a low, removed or charging battery can give quite different results. Where the standard or an applicable particular standard specifies the battery state, follow it. Otherwise, we recommend choosing the worst case from the risk analysis and recording the battery state, whether the battery is removable, and whether it is charging, in the test plan and report.
Switchover alarms
Whether the device should raise an alarm or information signal on switching to the internal power source, and whether that alarm is expected behaviour or a failure, must be defined before testing, based on the applicable particular standard and the manufacturer's risk management. For example, the draft GB 9706.284—202X for emergency medical services ventilators (a modified adoption of ISO 80601-2-84:2020), published for comment on the Shanghai Medical Products Administration website, requires in 212.8.2 a technical alarm condition or information signal indicating the switch to the internal power source, and a technical alarm condition, before all power sources are exhausted, indicating that the internal power source is nearly depleted. The published text governs.
Common misconceptions
- "It passed with a full battery." A full battery can mask problems in the mains power path; choose and record the worst-case battery state.
- "Any alarm is a failure." It depends on the criteria defined beforehand; an expected switchover alarm can be designed behaviour.
Accreditation scope
All rows come from the IAS accreditation scope (TL-1119) of the Hefei laboratory. Each is listed as a complete standard, with no individual clauses and no limitations.
| Item | Clause | Notes / limitations |
|---|---|---|
| IEC 61000-4-11 Voltage dips, short interruptions and voltage variations immunity tests (input current up to 16 A per phase) | Complete standard | No limitation stated; no edition year |
| EN IEC 61000-4-11 (same) | Complete standard | No limitation stated; no edition year |
| GB/T 17626.11 (same) | Complete standard | No limitation stated; no edition year |
| IEC 60601-1-2 Electromagnetic disturbances – Requirements and tests | Complete standard | No limitation stated; no edition year |
| EN 60601-1-2 (same) | Complete standard | No limitation stated; no edition year |
| YY 9706.102 Electromagnetic compatibility – Requirements and tests | Complete standard | No limitation stated; no edition year |
| IEC TS 60601-4-2 Electromagnetic immunity: performance of ME equipment and ME systems (guidance and interpretation) | Complete standard | No limitation stated; no edition year |
Where SUNGO performs the test, and under which accreditation
- The test is performed by the Hefei laboratory of Hefei SUNGO Medical Technology Co., Ltd., in Zipeng Town, Feixi County, Hefei, Anhui Province. The items above are within the IAS accreditation (TL-1119) of the Hefei laboratory. The Hefei and Shanghai SUNGO companies are separate legal entities; the scope of the Shanghai laboratory (No. 958 Kangqiao East Road, Pudong, Shanghai) does not include EMC.
- The certificate does not state edition years. Before an IAS report is issued against a specific edition (for example IEC 60601-1-2 Edition 4.1, YY 9706.102-2021 or GB/T 17626.11-2023), please confirm the edition with the laboratory.
- Whether a report can be used for registration or certification in a given market should be confirmed in advance with the regulator or notified body.
- Service overview: medical EMC testing.
Preparing for the test
- Standards: the target market and the standards and editions to test against.
- Supply data: rated voltage range, rated frequency, rated input current per phase (to confirm the device is within the scope of IEC 61000-4-11), single- or three-phase, any voltage selector or transformer taps, and the external adapter model.
- Internal power source: battery type, whether removable, the battery and charging state for the test, the switchover logic, and any alarm or information signal expected at switchover.
- Essential performance and criteria: the essential performance list, how each item is observed and its acceptable range, the permitted recovery (automatic or by operator), and the settings that must be retained, all fixed before testing.
- Configuration: worst-case operating mode, alarm limit settings, any patient simulator or load, cables, and the frozen software version.
- Recovery instructions: if the operator must restart or reconfigure the device after an interruption, the corresponding steps from the instructions for use.
If problems appear during testing, see EMC Failures: Locating the Cause, Fixing It, Scoping the Retest.
FAQ
Does a battery-only device with no power cord need this test?
A device with no AC mains connection is outside the scope of IEC 61000-4-11. If it can be used on an adapter or while charging, assess it in mains-connected operation.
How many supply voltages for a device rated 100–240 V?
Under Table 1 of IEC 60601-1-2 Edition 4.1, one voltage is allowed only if the difference between maximum and minimum rated voltage is less than 25 % of the maximum. A 100–240 V range exceeds that, so dips are tested at both 100 V and 240 V; short interruptions may use any one voltage.
The device is rated 50/60 Hz. Do both frequencies need testing?
Edition 4.1 Table 1 allows any one frequency in the rated range, with no retest at other frequencies. The cycle count follows the chosen frequency; for example, "25/30 cycles" means 25 cycles at 50 Hz or 30 cycles at 60 Hz.
We tested to YY 9706.102. Must we retest to IEC 60601-1-2 Edition 4.1 for export?
The two correspond to different editions of IEC 60601-1-2, so one result cannot stand in for the other. Whether a retest is needed depends on the edition the target market requires; confirm with the regulator or notified body in advance.
When requesting a quote, please give the product and intended use, rated voltage, frequency and input current, whether it has a battery, the standards, editions and target market, and your essential performance and criteria. Reach us through the contact page, or see the medical EMC testing service page first.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.