IEC 60601-1 Clause 15.3 Enclosure Strength: Push, Impact, Drop, Mould Stress

IEC 60601-1 Clause 15.3 Enclosure Strength: Push, Impact, Drop, Mould Stress

A hand-held pulse oximeter slides off a table, a portable oxygen concentrator is picked up and put down, a trolley-mounted unit hits a door frame. Will the enclosure crack, and will live parts become accessible? Clause 15.3 "Mechanical strength" of IEC 60601-1 (GB 9706.1 in China) deals with these questions. This guide explains which tests apply to which equipment type, how the test conditions are defined and results judged, how to prepare samples for three typical products, and which SUNGO laboratory performs the work under which accreditation.

Key points first

  • The equipment type decides the test programme. Table 28 lists the tests for hand-held, body-worn, portable, mobile, and fixed or stationary equipment: push and mould stress relief are listed for every type (mould stress relief concerns only moulded or formed thermoplastic enclosures and may be confirmed by inspecting the construction and available data); the drop test applies to hand-held, body-worn and portable equipment; rough handling applies only to mobile equipment; the impact test does not apply to hand-held equipment.
  • The criterion is not "no visible damage" but whether damage results in an unacceptable risk, that is, a loss of basic safety or essential performance. A note in the standard lists evaluation means that may be useful: measuring creepage distances and air clearances, checking accessible parts, the 8.8.3 dielectric strength test, and the criteria of Clause 8 and 11.6. Which ones are applied depends on the damage found.
  • Drop and rough handling can damage samples, and if mould stress relief is tested in an oven, the temperature depends on the maximum enclosure temperature from the temperature test. Plan samples and sequence with the laboratory before ordering.
  • At SUNGO, IEC 60601-1, EN 60601-1 and GB 9706.1 are tested by the Hefei laboratory, within its IAS accreditation (TL-1119). The scope lists each as a whole standard without a year; confirm the edition with the laboratory before commissioning.

Standards and scope

Three designations and their editions

IEC 60601-1 is the general standard for basic safety and essential performance of medical electrical equipment. As of 2026-10-22, the IEC Webstore lists consolidated edition 3.2 (IEC 60601-1:2005+AMD1:2012+AMD2:2020 CSV) as current; it was published on 2020-08-20. GB 9706.1-2020 is the mandatory Chinese national standard, issued on 2020-04-09 and in force since 2023-05-01; it is a modified adoption of IEC 60601-1:2012 and replaces GB 9706.1-2007 and GB 9706.15-2008. EN 60601-1 is the European adoption; as of the same date it is EN 60601-1:2006 with amendments A1, A12, A2 and A13.

The test parameters below were compiled from the Korean-language regulatory text that the Korean Ministry of Food and Drug Safety (MFDS) issued on the basis of IEC 60601-1:2012 (edition 3.1), not from the IEC English original; GB 9706.1-2020 is also a modified adoption of edition 3.1. For the exact wording and values of edition 3.2 and GB 9706.1-2020, refer to the edition you test against. IEC TR 62348:2012 rates the Amendment 1 changes to 15.3.1 (general) and 15.3.5 (rough handling) as "moderate impact", so products last tested to the original 2005 text should check these two subclauses.

Structure of Clause 15.3 and Table 28

Clause 15.3 sits in Clause 15 "Construction of ME equipment": 15.3.1 General, 15.3.2 Push test, 15.3.3 Impact test, 15.3.4 Drop test (15.3.4.1 hand-held, 15.3.4.2 portable), 15.3.5 Rough handling test, 15.3.6 Mould stress relief and 15.3.7 Environmental influences. Table 28 in 15.3.1 sets out the applicable tests:

Equipment typePush 15.3.2Impact 15.3.3Drop 15.3.4Rough handling 15.3.5Mould stress relief 15.3.6
Hand-heldYesNo15.3.4.1NoYes
Body-wornYesYes15.3.4.1NoYes
PortableYesYes15.3.4.2NoYes
MobileYesYesNoYesYes
Fixed or stationaryYesYesNoNoYes

"Yes" only means Table 28 lists the test; conditions are in each subclause below (for example, the push test excludes the enclosure bottom of equipment over 18 kg, and mould stress relief concerns only moulded or formed thermoplastic enclosures). The type follows the Clause 3 definitions: hand-held equipment is supported by the hand in use; portable equipment is carried by one or more persons between locations; mobile equipment moves on its own wheels or equivalent means; stationary equipment is not intended to be moved once installed; body-worn equipment operates while worn by the patient or attached to the patient's clothing. The hand-held and portable definitions note that "equipment" can mean an accessory or a part, so a main unit and its accessories may be tested as different types.

Test by test: conditions and acceptance

15.3.2 Push test

The enclosure must be sufficiently rigid. A steady force of 250 N ± 10 N is applied for 5 s to the outside of the enclosure through a test tool contacting a circular plane surface 30 mm in diameter. The bottom of the enclosure of equipment over 18 kg is excluded. Lasting damage that could result in an unacceptable risk is a failure. Annex A notes that the force depends on the person handling the equipment, not on its mass. Marking thin walls, large spans and areas with live parts close behind on the enclosure drawing helps the laboratory choose test points.

15.3.3 Impact test

The test covers enclosures and other external insulating parts whose deterioration could result in an unacceptable risk. It is not applied to hand-held equipment or hand-held parts, flat panel displays, flat glass surfaces of equipment (for example film scanners) or cathode ray tubes. The complete enclosure, or a portion representing the largest unreinforced area, is supported in its normal position, and a smooth solid steel ball of about 50 mm diameter and 500 g ± 25 g is dropped once from 1.3 m onto each relevant part. For vertical surfaces the ball is suspended on a cord and released as a pendulum from a vertical height of 1.3 m. Annex A explains that this represents accidental strikes, for example by an object carried by a passer-by or a mop handle; where the manufacturer considers the requirement unnecessary, the rationale and an alternative requirement are documented in the risk management file in accordance with 4.5.

15.3.4 Drop test

Hand-held equipment (15.3.4.1). Hand-held equipment, accessories and parts of equipment shall not result in an unacceptable risk from a free fall. With the safe working load in place, the sample falls freely once from each of three different attitudes encountered in normal use, from the use height specified in the accompanying documents or 1 m, whichever is greater, onto a hardwood board 50 mm ± 5 mm thick (density above 600 kg/m³) lying flat on concrete or a similarly rigid base.

Portable equipment (15.3.4.2). With the safe working load in place, the sample is lifted to the Table 29 height and dropped onto the same type of board, which shall be at least as large as the sample, three times from each attitude in which it can be placed in normal use. Table 29 sets the height by mass:

Mass of the portable equipment or partDrop height
Up to and including 10 kg5 cm
Over 10 kg up to and including 50 kg3 cm
Over 50 kg2 cm

Lasting damage that results in an unacceptable risk is a failure. The safe working load is the maximum external mechanical load (mass) permitted on the equipment or part in normal use. Annex A explains that the portable drop represents normal use, not reasonably foreseeable misuse; if the risk management process concludes that a more severe test is appropriate, it should be performed.

15.3.5 Rough handling test

Mobile equipment and mobile parts must withstand the stresses of rough handling and movement. The sample is tested in its transport position, with the safe working load in place, in the worst case permitted in normal use, with precautions against overbalancing. The speed is 0.8 m/s ± 0.1 m/s, or the maximum speed that can be maintained for motor-driven equipment, three times in the normal direction of travel, perpendicular to the obstacle:

  • Ascending step shock: pushed against a hardwood step with a 40 mm vertical face fixed to a flat floor; it does not need to climb over it.
  • Descending step shock: pushed off a 40 mm vertical step fixed flat on a rigid base; if a part other than the castors touches the obstruction first, pushing continues until the equipment has fully descended.
  • Door frame shock: pushed against a vertical hardwood obstacle 40 mm wide and 40 mm thick on a rigid support, taller than the contact point of the equipment.

After each test, lasting damage that results in an unacceptable risk is a failure. If the equipment is not damaged, stability is assessed separately under 9.4. Annex A notes that equipment such as beds, which moves forwards and backwards at normal speed, should be considered in both directions.

15.3.6 Mould stress relief

Moulded or formed thermoplastic enclosures shall not shrink or distort from released internal stresses to the point of unacceptable risk. Compliance is checked by inspecting the construction and available data, or by test: the complete equipment, or the enclosure with any supporting frame, is placed in a circulating-air oven for 7 h at 10 °C above the maximum enclosure temperature observed during the temperature test (11.1.3 measurements), but not less than 70 °C, then cooled to room temperature. Relative humidity need not be controlled. For large equipment, a portion of the enclosure representative in thickness and shape may be used. Annex A recommends, where possible, a written comparison of the maximum temperature in normal use with the service temperature range recommended by the polymer manufacturer.

15.3.7 Environmental influences

Construction materials must be selected and treated with regard to intended use, expected service life and transport and storage conditions, so that corrosion, ageing, mechanical wear or degradation of biological materials does not reduce mechanical properties to the point of unacceptable risk over the expected service life. Compliance is checked by inspecting the equipment, accompanying documents, material and processing specifications, and the manufacturer's relevant tests or calculations, so this is largely a document review.

How results are judged

15.3.1 requires that equipment and parts do not lose basic safety or essential performance through moulding stress or mechanical stress from pushing, impact, dropping or rough handling. The note gives examples of damage affecting basic safety: creepage distances and air clearances reduced below 8.9, access to parts exceeding the 8.4 limits, and access to moving parts that could cause harm. The note also lists criteria that may be useful:

  • the criteria of Clause 8 (including 8.4 limitation of voltage, current or energy and 8.7 leakage currents) and of 11.6;
  • the dielectric strength test of 8.8.3 for solid insulation serving as a means of protection;
  • measurement of creepage distances and air clearances against the minimum values of 8.9.

Small chips that do not affect protection against electric shock or moisture can normally be ignored. For equipment with essential performance, confirm afterwards that it is maintained. The tests are not applied to handles, levers, knobs, cathode ray tubes or transparent or translucent covers of indicating or measuring devices, unless removing them would create an unacceptable risk of electric shock. For troubleshooting, see Electrical Safety Failures: Creepage, Earthing and Leakage Current.

Sample preparation for three typical products

Hand-held pulse oximeter

  • The main unit is hand-held: Table 28 lists push, drop (15.3.4.1) and mould stress relief (for a moulded or formed thermoplastic enclosure); no impact test. The drop height is the use height in the instructions or 1 m, whichever is greater, once from each of three attitudes. Whether probes and other accessories are dropped as hand-held parts depends on how they are used.
  • Dropped units may go on to dielectric strength and distance evaluations and an essential performance check, so they are best not reused for other tests; provide extra units for parallel testing.
  • Pulse oximeter equipment has its own particular standard (in China, YY 9706.261-2023), which may modify Clause 15.3. It is not on SUNGO's certificates; this guide covers the general standard only.

Portable oxygen concentrator

  • Portable equipment: Table 28 lists push, impact, drop (15.3.4.2) and mould stress relief (see the next point). Provide the mass of each configuration (for example with and without battery) to set the Table 29 band; three drops from each attitude in which it can be placed.
  • Where the enclosure is moulded or formed thermoplastic, schedule the temperature test first, then set the mould stress relief temperature from the maximum enclosure temperature.
  • The oxygen concentrator particular standards ISO 80601-2-69, EN ISO 80601-2-69 and YY 9706.269 are within the Hefei laboratory's IAS accreditation (TL-1119); their modifications apply according to the edition used. See Oxygen Concentrator Testing: ISO 80601-2-69, IEC 60601-1 and IEC 60601-1-2.

Table-top patient monitor

  • The type depends on design and accompanying documents: a unit with a carrying handle meant to be carried between locations is dropped as portable; a trolley-mounted unit moved on wheels is tested as a mobile combination under rough handling; a unit not intended to be moved is stationary (push, impact and, for a moulded or formed thermoplastic enclosure, mould stress relief).
  • A flat panel display is excluded from the impact test, but its surrounding enclosure is not. Supply trolleys, stands and mounts with the sample and state the configuration.

Certificate scope

The rows below are taken from the scope of the Hefei laboratory's IAS certificate (TL-1119); items, clauses and limitations are reproduced as listed:

StandardItemClauseNotes / limitations
IEC 60601-1Whole standardNo clause listedNo limitation listed; no year given; Clause 15.3 is part of the whole standard
EN 60601-1Whole standardNo clause listedNo limitation listed; no year given
GB 9706.1Whole standardNo clause listedNo limitation listed; no year given
IEC 60601-1-11Whole standardNo clause listedHome healthcare collateral standard; no limitation listed; no year given
EN 60601-1-11Whole standardNo clause listedNo limitation listed; no year given
YY 9706.111Whole standardNo clause listedNo limitation listed; no year given
ISO 80601-2-69Whole standardNo clause listedOxygen concentrator particular standard; no limitation listed; no year given
EN ISO 80601-2-69Whole standardNo clause listedNo limitation listed; no year given
YY 9706.269Whole standardNo clause listedNo limitation listed; no year given

Where SUNGO performs the work, and under which accreditation

  • Clause 15.3 testing to IEC 60601-1, EN 60601-1 and GB 9706.1, and the electrical safety checks that follow, are performed by the Hefei laboratory, within its IAS accreditation (TL-1119). The laboratory belongs to Hefei SUNGO Medical Technology Co., Ltd., a separate legal entity in Zipeng Town, Feixi County, Hefei, Anhui Province. The accredited scope of Shanghai SUNGO Medical Technology Co., Ltd. does not include these standards.
  • The IAS scope lists these standards without a year. For a report to IEC edition 3.2, GB 9706.1-2020 or a specific EN version, confirm the edition with the laboratory before commissioning.
  • For home healthcare equipment, Clause 10.1 of IEC 60601-1-11 adds mechanical strength requirements, split by whether the equipment is transit-operable. The collateral standard normatively references IEC 60068-2 test methods for shock, rough handling shocks and broadband random vibration, applied as the standard text specifies; the IEC 60068-2 series is not listed separately on SUNGO's certificate. A separate article on IEC 60601-1-11 covers this.
  • The only particular standards on the certificate are ISO 80601-2-69, EN ISO 80601-2-69 and YY 9706.269 for oxygen concentrators. Modifications to Clause 15.3 by any other particular standard are outside SUNGO's accredited scope.
  • Whether a report is accepted for registration in a given country or region, including China, depends on the relevant regulatory requirements; SUNGO makes no commitment on acceptance.

Preparing your submission

  1. Equipment type: state whether the main unit and each accessory is hand-held, body-worn, portable, mobile or stationary, based on how the instructions describe use and handling.
  2. Samples: match the design intended for market, with production enclosure material, wall thickness and fastening. Drop, rough handling and mould stress relief (for moulded or formed thermoplastic enclosures) can damage samples, so provide spares and agree with the laboratory which tests can share samples. Subclause 5.2 requires type tests on a representative sample, and its note allows several samples to be used simultaneously if the validity of the results is not significantly affected. See Planning Sample Quantities: Destructive and Parallel Tests.
  3. Test sequence: Subclause 5.8 requires a sequence in which one result does not influence later ones. In the recommended sequence of informative Annex B, mechanical strength comes after the temperature test, humidity preconditioning and the leakage current and dielectric strength tests around it.
  4. Accompanying documents: state the use height, handling method, transport position, safe working load and the attitudes in which the equipment can be placed.
  5. Technical data: enclosure material grade and supplier data (including the recommended service temperature range), processing specifications, mass of each configuration, insulation diagram with creepage and clearance list, and the essential performance list with its verification method.
  6. Risk management file: record the rationale and alternative requirement wherever an alternative method is used or a test is considered not applicable.

Frequently asked questions

Can we test Clause 15.3 only?

The IAS scope lists IEC 60601-1 as a whole standard. When requesting a quote, confirm with the laboratory whether Clause 15.3 can be assessed alone and how the report would state the scope.

Is a crack after the drop test automatically a failure?

No. The question is whether the damage results in an unacceptable risk: reduced distances, access to hazardous parts, a failed dielectric strength test or loss of essential performance.

Does mould stress relief always require oven testing?

No. The standard allows inspection of the construction and available data. Whether the data are sufficient is decided by the laboratory from the material specifications and temperature data you submit.

Is Clause 15.3 of GB 9706.1-2020 the same as in IEC 60601-1 edition 3.2?

GB 9706.1-2020 is a modified adoption of IEC 60601-1:2012 and predates Amendment 2, so some subclauses may differ. Testing follows the edition confirmed when the order is placed.

Requesting a quote

Please provide the product name and intended use, models and configurations, equipment type and mass of each configuration, enclosure materials, use environment (home use, use during transport), target markets and the intended edition of the standard. The Hefei laboratory will use this to confirm test items, sample plan and sequence. Service details are on the medical electrical safety testing page; send your request through the contact page.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.