Conducted Emission Failure on Medical Devices: How to Fix the Mains Port

Conducted Emission Failure on Medical Devices: How to Fix the Mains Port

How do you fix a conducted emission failure on a medical device? Work in this order: read the spectrum to decide whether the exceedance is a narrowband peak at the switching harmonics or a continuous broadband rise; separate differential-mode from common-mode; then change the filter, earthing, cables or external adapter that the diagnosis points to; and finally retest according to the scope of the change, rechecking leakage current whenever the Y capacitors or earthing were touched. This article covers mains-port conducted emissions (also called conducted disturbance) only. Radiated emissions and immunity are outside its scope.

Short answer

  • Diagnose before you change anything. Narrowband peaks at the switching frequency and its multiples usually come from switching action. A continuous broadband rise usually points to ringing, earth loops or cables.
  • Treat common-mode and differential-mode separately. For common-mode exceedances, start with the Y capacitors, the common-mode choke, PE routing, heatsink earthing and any external adapter. For differential-mode exceedances, start with the X capacitors, the differential-mode inductor and the area of the switching loop.
  • A bigger Y capacitor is not automatically better. It raises earth leakage current (in Class II equipment without protective earth, it mainly affects touch current, and patient leakage current where there is an applied part), so EMC and electrical safety have to be assessed together.
  • Change one thing at a time and keep records. Retest according to the scope of the change, not just the single frequency point that failed.
  • On the standards side, GB 4824-2025 is an identical adoption of CISPR 11:2024 and has been in force since 1 March 2026. EMC requirements for medical electrical equipment also have to be checked against IEC 60601-1-2 and YY 9706.102-2021.

Which standards apply? GB 4824-2025 and CISPR 11

GB 4824-2025, Industrial, scientific and medical equipment - Radio-frequency disturbance characteristics - Limits and methods of measurement, is a mandatory Chinese national standard. It was published on 28 February 2025, has been in force since 1 March 2026, and fully replaces GB 4824-2019. The adoption record on China's national standards information platform states that it is an identical adoption of the IEC standard CISPR 11:2024. As of 13 November 2026, the platform lists GB 4824-2025 as current and GB 4824-2019 as withdrawn.

For medical electrical equipment, the international EMC requirements are in IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests. The Chinese industry standard is YY 9706.102-2021, with a corresponding title on electromagnetic compatibility. It was published on 9 March 2021, has been in force since 1 May 2023, and replaces YY 0505-2012. According to its registered scope, YY 9706.102 sets out the general EMC requirements and tests for medical electrical equipment and medical electrical systems, and serves as the basis for particular standards.

In practice, for a medical device you first use these collateral standards (and any applicable particular standard) to establish the emission test basis and how the equipment is categorised, and then judge the result with the methods and limit tables of GB 4824 / CISPR 11. Which set of limits applies depends on how the equipment is categorised in the standard; take the definitions and tables from the standard text. This article does not quote any limit or frequency values.

How is conducted emission measured?

Mains-port conducted emission is the radio-frequency disturbance that the equipment sends back into the supply along its power cord. During the measurement, the power cord is connected through an artificial mains network (often called a LISN), and the receiver reads the disturbance voltage of each supply conductor relative to the reference ground. That reading contains both differential-mode and common-mode components, so a single failing trace does not tell you directly what to change.

Because the measurement is a conductor-to-ground voltage, the stray capacitance between the equipment and the reference ground, the way PE is connected, and the length and routing of the power cord all change the reading. If the pre-scan layout differs from the formal test layout, the results may not match. Fix the layout before you start changing the design; otherwise you cannot tell whether a change actually worked.

What to look at first: reading the spectrum

Plot the measured trace and the limit line on the same chart. If the exceedances sit at integer multiples of the switching frequency as a series of narrow peaks, the source is usually switching action, typically an internal DC/DC converter, a motor driver or the switching node of an external adapter. If the exceedance is a continuous broadband rise, ringing on switching edges, resonance or an earth loop is more likely, and the way PE and cable shields are connected has a large effect. The two cases call for different fixes, so identify which one you have before changing anything.

Pay attention to the detector and the measurement bandwidth. Pre-scans often use a peak detector across the whole band for speed. A peak reading is generally not lower than readings with other detectors, so a peak exceedance is not the same as a formal failure. The formal verdict uses the detectors and limits the standard specifies. Another practical check is whether the exceedance changes with operating mode. If it moves noticeably with load or mode, switching action is the likely cause. If it barely changes, earthing, routing or the enclosure are more likely.

The following is a hypothetical case, used only to show the order of diagnosis. During a pre-scan, an active device with an external adapter shows narrowband exceedances at several multiples of the switching frequency. A current-probe measurement shows that the exceedance is mainly common-mode, while the differential-mode component has ample margin. Increasing the X capacitors would do little here. The Y capacitors, PE routing, heatsink earthing and the adapter should be checked first.

How do you separate common-mode from differential-mode?

Differential-mode current flows out on the line conductor and back on the neutral, in opposite directions on the two conductors. Common-mode current flows in the same direction on both supply conductors and returns through PE, the enclosure, or the stray capacitance between the equipment and the reference ground. Once you know which one dominates, the entry point for the fix is clear.

Three methods are common in engineering practice:

  1. Current probe. Pass the line and neutral conductors through the probe together. The differential-mode currents cancel, so the reading is mainly common-mode current. Clamping a single conductor gives the sum of both components. Comparing readings before and after a change shows whether a measure is effective against common-mode.
  2. DM/CM separator. Connected to the outputs of the artificial mains network, it delivers the differential-mode and common-mode components separately, which you then compare with the original trace.
  3. Rule of thumb. Exceedances at the lower end of the band are often differential-mode; those at the upper end are often common-mode. This is experience only and does not replace the two measurements above.

Once the mode is known: for differential-mode exceedances, start with the X capacitors, the differential-mode inductor and the switching-loop area. For common-mode exceedances, start with the Y capacitors, the common-mode choke, PE routing, coupling between the heatsink and the enclosure, and any external adapter.

Conducted emission fixes and what each one costs

The usual fixes fall into the groups below. Each one has to be weighed against safety and performance:

  • Larger or additional Y capacitors. These suppress common-mode interference directly, but a larger Y capacitor increases leakage current. Leakage current is an electrical safety item, so the two must be assessed together. For the electrical safety side, see Electrical Safety Failures: Creepage, Earthing and Leakage Current.
  • An additional common-mode choke or a revised winding. Effective against common-mode interference. Check size, temperature rise, core saturation and the effect of leakage inductance on differential-mode behaviour.
  • Larger X capacitors or a differential-mode inductor. Aimed at differential-mode exceedances. After component values change, temperature rise and the related electrical safety checks need to be revisited as well.
  • Revised PE routing and earthing. Shorten the earth loop, sort out how the heatsink, cable shields and enclosure are connected, and avoid moving earth points without a reason. Changes to earthing affect electrical safety and should be assessed by the design team before they are made.
  • External adapter. If the exceedance comes mainly from the adapter, change the model or add filtering. After a change of model, the adapter's own electrical safety evaluation also has to be confirmed, and the submission documents should state the adapter model used.
  • Cables and routing. Cable length, shielding and routing change the common-mode return path. The same device with a different cable can give noticeably different results, so the cable configuration submitted should match the product as it is actually used.

Why does a larger Y capacitor increase leakage current?

In Class I equipment with protective earth, the Y capacitors are connected between the supply conductors and PE. They give common-mode noise a short, low-impedance return path so that this current no longer flows through the measuring network, and the reading drops. The same path, however, also carries a mains-frequency current into PE, which forms part of the earth leakage current. The larger the capacitor, the larger that current. In Class II equipment without protective earth, the Y capacitor often bridges the primary and secondary sides, and what it mainly affects is touch current; where the device has an applied part, patient leakage current also has to be considered.

Medical electrical equipment has explicit safety requirements for these leakage currents, so the Y capacitor value cannot be chosen on EMC results alone. In practice, a common approach is to choose the smallest capacitor that gives adequate EMC margin and then confirm the choice with electrical safety testing. Because a Y capacitor bridges insulation, it is a safety-related component, and any change of type or value should be assessed by the design team against the safety standard. The applicable limits and measurement conditions are those in the text of IEC 60601-1 / GB 9706.1.

In what order should fixes be made?

The main risk in a fix is changing several things at once. If the Y capacitors, a common-mode choke and the earthing are all changed and the exceedance disappears, you cannot tell which change worked, or whether one of them introduced a new problem. A safer approach is to change one thing at a time, record the spectrum before and after each change, and then decide the next step.

Start with low-cost, low-side-effect measures such as cable routing and the earth loop; then consider adding or replacing components; and only then change the PCB layout or the mechanical design. Any safety-related change, such as the Y capacitors or earthing, should be risk-assessed before it is made, and the assessment record should be kept. For a general approach to locating any kind of EMC failure, see EMC Failures: Locating the Cause, Fixing It, Scoping the Retest.

Which items need retesting after a fix?

Retest what you changed. A change to filter components or earthing normally means repeating the mains-port conducted emission measurement. If the Y capacitors or the earthing arrangement changed, leakage current and the related electrical safety items should also be rechecked. If the change affects the enclosure, PE or cable routing, radiated emissions and immunity may also need to be re-assessed. Deciding which items must be fully repeated and which can be retested partially is covered in a separate article on this site, available in Chinese only.

Related standards in Sungo's accreditation scope

The table below lists the entries in the Hefei SUNGO IAS accreditation scope (certificate TL-1119) that relate to this article, as recorded on the certificate. Each entry is recorded as a whole standard, with no clauses listed and no year stated.

ItemClause basisNotes and limits
CISPR 11 (Industrial, scientific and medical equipment - Radio-frequency disturbance characteristics - Limits and methods of measurement)Whole standardHefei, IAS. No year stated; no limitation recorded.
GB 4824 (same title, Chinese national standard)Whole standardHefei, IAS. No year stated; no limitation recorded.
IEC 60601-1-2 (medical electrical equipment, collateral standard on electromagnetic disturbances)Whole standardHefei, IAS. No year stated; no limitation recorded.
EN 60601-1-2Whole standardHefei, IAS. No year stated; no limitation recorded.
YY 9706.102 (medical electrical equipment, collateral standard on EMC)Whole standardHefei, IAS. No year stated; no limitation recorded.
IEC 60601-1 / EN 60601-1 / GB 9706.1 (medical electrical equipment, general requirements for basic safety and essential performance)Whole standardHefei, IAS. No year stated; no limitation recorded. The safety standards behind a leakage current recheck after a fix.

The IAS scope does not state years. Before an IAS report is issued against a particular edition, confirm the edition with the laboratory. The same IAS scope also includes harmonic current emissions (IEC 61000-3-2, GB 17625.1) and voltage fluctuations and flicker (IEC 61000-3-3, GB/T 17625.2). These are separate tests from the mains-port conducted emission discussed here, and one cannot stand in for the other. CISPR 14, CISPR 32, FCC Part 15 and FCC Part 18 are not within Sungo's accreditation scope.

Where does Sungo test this, and under which accreditation?

The medical electrical equipment EMC and electrical safety items in the table above are performed by Hefei SUNGO Medical Technology Co., Ltd., in Zipeng Town, Feixi County, Hefei, within its IAS accreditation scope, certificate IAS TL-1119. These items are not within Sungo's CNAS accreditation or CMA scope, and the Shanghai laboratory's scope does not include EMC. CNAS and IAS are both signatories to the ILAC MRA. The scope of EMC testing services is described on the EMC service page.

What to prepare before submitting samples

The point of the submission package is that the result can be reproduced. The sample state before and after the fix, the cable configuration and the operating modes should all be recorded in writing, with photographs where useful. If the design has been revised, state what differs between the old and new versions, and keep the original measurement data for comparison.

  • A description of the sample's operating modes, including which modes switch on the power converter and which draw the most power.
  • The input voltage and frequency, the model of any external accessory or adapter, and the firmware or software version.
  • The protection class against electric shock (Class I or Class II) and how PE is connected, which helps identify the common-mode return path.
  • If filtering or earthing was changed, a before-and-after record of each difference.
  • Cable length, shielding and routing, consistent with the product as actually used and written into the test plan.
  • For a device with several operating modes, identify the mode that gives the worst case for the exceedance, and write that mode into the test configuration.

Frequently asked questions

Does a conducted emission failure always need a filter change?

No. Confirm the nature of the exceedance with the spectrum and a common-mode/differential-mode separation first. Adjusting earthing or cable routing, or changing the adapter, sometimes resolves it without adding components.

How do I tell quickly whether a conducted emission failure is common-mode or differential-mode?

Pass the line and neutral conductors through a current probe together; the reading is mainly common-mode current. Then compare the original trace with the outputs of a DM/CM separator. Where in the band the failure sits is only a rule of thumb.

Will a larger Y capacitor fix every common-mode exceedance?

Not necessarily. A larger Y capacitor increases leakage current, so it must be assessed together with electrical safety, and the final result must be confirmed by retest.

If I only change the mains-port filter, do I still need to retest radiated emissions?

It depends on the scope of the change. A change confined to the mains-port filter normally means repeating conducted emissions first. If the change affects the enclosure, earthing or cable routing, radiated emissions should also be re-assessed.

Are the limits in GB 4824-2025 and CISPR 11 the same?

GB 4824-2025 is an identical adoption of CISPR 11:2024, so its technical content matches that edition. If you are comparing against another edition of CISPR 11, take any differences from the text of each edition.

If you want to confirm whether a mains-port conducted emission test for your device falls within the Hefei IAS scope, or want to know what configuration information to prepare before submission, please contact us for a quotation.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.