Blood Collection Tube IVDR Classification: Class A and Using ISO 6710 Reports

Blood Collection Tube IVDR Classification: Class A and Using ISO 6710 Reports

Under the EU IVDR (Regulation (EU) 2017/746), single-use venous blood collection tubes, evacuated or non-evacuated, are specimen receptacles, and Annex VIII Rule 5(c) places them in Class A. For non-sterile Class A devices, the manufacturer draws up the technical documentation and issues the EU declaration of conformity itself; for devices placed on the market sterile, a notified body is involved, but only for the sterility aspects. As of 13 November 2026, ISO 6710:2017 is not on the list of harmonised standards under the IVDR, so an ISO 6710 test report can go into the technical documentation as evidence but does not give a presumption of conformity.

This article is for regulatory, quality and R&D staff at tube manufacturers and for buyers sourcing tubes for the EU. It covers the specimen receptacle definition and Class A, how MDCG 2020-16 rev.5 explains Rule 5(c), where an ISO 6710 report fits in the technical documentation and GSPR checklist, and which ISO 6710 clauses SUNGO Shanghai is accredited for. The test items, and how YY/T 0314-2021 relates to ISO 6710, are covered in Venous Blood Collection Tube Testing to YY/T 0314-2021 and ISO 6710:2017. The full IVDR classification rules and transitional arrangements are not covered here.

Key points

  • Classification: a blood collection tube is a specimen receptacle as defined in Article 2(3) of the IVDR and is Class A under Annex VIII Rule 5(c). Under implementing rule 1.1, classification is governed by the intended purpose.
  • Conformity route: under Article 48(10), for a non-sterile Class A device the manufacturer draws up the Annex II and III technical documentation and issues the EU declaration of conformity. If placed on the market sterile, the Annex IX or Annex XI procedure applies, with notified body involvement limited to establishing, securing and maintaining sterile conditions.
  • Status of ISO 6710: IVD harmonised standards are listed in Implementing Decision (EU) 2021/1195. As of 13 November 2026, the latest amendment listed by the European Commission is (EU) 2026/2481 of 9 November 2026, and ISO 6710 is not on the list. It can serve as an "other solution" under Section 4 of Annex II, justified in the technical documentation, but gives no presumption of conformity.
  • What SUNGO can test: the Shanghai laboratory has five items to ISO 6710:2017 and YY/T 0314-2021 within its CNAS (L22083) and CMA (240015340087) scopes: materials (4.1, 4.3), draw volume (Clause 5, Annex A, Annex B), design (Clause 6), construction (Clause 7), and marking and labelling (Clause 10). There is no IAS accreditation for them.
  • Not in SUNGO's certificates: the rest of Clause 4, Clause 8 (sterility and special microbiological states), Clause 9 (additives) and Clause 11 (container identification).
  • Report marks: reports for the EU can carry the CNAS mark on request; the CMA certificate states that it is valid within the People's Republic of China. Acceptance of a report is for the notified body or customer to decide.

What a blood collection tube is under the IVDR: the specimen receptacle definition

The IVDR was published in the Official Journal on 5 May 2017 and has applied since 26 May 2022. After defining an in vitro diagnostic medical device, Article 2(2) adds that specimen receptacles are also deemed to be in vitro diagnostic medical devices. So a tube is regulated under the IVDR even though it performs no test.

Under Article 2(3), a specimen receptacle is a device that:

  • may or may not be of a vacuum type;
  • is specifically intended by its manufacturer for the primary containment and preservation of specimens derived from the human body;
  • serves the purpose of in vitro diagnostic examination.

Whether a container qualifies depends on the intended purpose stated by the manufacturer. The examples in MDCG 2020-16 rev.5 under Rule 5(c) include a standalone urine collection cup, stool container, saliva collection tube and blood spot collection card, all Class A.

Article 1(3)(b) excludes invasive sampling products and products applied directly to the human body to obtain a specimen, and MDCG 2020-16 rev.5 repeats this under Rule 5(c). Within a blood collection system, a component applied directly to the body must be assessed against that provision separately; the tube being an IVD does not settle it.

Blood collection tube IVDR classification: why Rule 5(c) means Class A

Rule 5 of Annex VIII lists three groups of Class A devices: (a) products for general laboratory use, accessories with no critical characteristics, buffer and washing solutions, general culture media and histological stains, where the manufacturer intends them for in vitro diagnostic procedures relating to a specific examination; (b) instruments intended specifically for in vitro diagnostic procedures; and (c) specimen receptacles. Rule 5(c) says only "specimen receptacles", with no further conditions.

MDCG 2020-16 rev.5 (September 2026) is the MDCG-endorsed guidance on IVD classification. Its explanation of Rule 5(c), on pages 41 to 43, makes these points:

  • Specimen receptacles are specimen containers or evacuated or non-evacuated tubes, empty or prefilled with a fixative solution or other general reagent, intended for primary containment, preservation, transport and storage of biological specimens for in vitro diagnostic examination. They are Class A.
  • In line with Article 2(3), they are for primary containment and preservation, so they cannot have any integrated testing function.
  • They may be placed on the market standalone, or as part of a specimen collection kit or a kit intended for testing.
  • A specimen collection kit is classified by its intended purpose and implementing rule 1.9, and is typically Class A. Within a kit intended for testing, the whole kit is classified under implementing rules 1.8 and 1.9 by its intended purpose, while the specimen receptacle may be CE-marked and classified in its own right.
  • A receptacle placed on the market separately but intended for use with another IVD is classified independently under implementing rule 1.2.
  • Receptacles or collection kits for lay persons placed on the market separately from the testing device are not devices for self-testing.

The MDCG document states on its cover that it is endorsed by the Medical Device Coordination Group, is not a European Commission document and is not legally binding.

On tubes containing additives: the MDCG explanation of Rule 5(c) refers to empty tubes and tubes prefilled with a fixative solution or other general reagent. It does not separately address tubes containing additives intended for a specific examination. This article draws no conclusion on how such products are classified; check the intended purpose against Annex VIII and the MDCG text itself.

Blood collection tube CE marking route: self-declaration or notified body

Article 48(10) sets out conformity assessment for Class A devices other than devices for performance study:

  • Non-sterile: the manufacturer draws up the technical documentation in Annexes II and III and issues the EU declaration of conformity referred to in Article 17. This is a self-declaration.
  • Placed on the market sterile: the manufacturer applies the Annex IX or Annex XI procedure, and notified body involvement is limited to establishing, securing and maintaining sterile conditions.

Whether a tube is placed on the market sterile depends on its claims and labelling. Clause 8 of ISO 6710:2017 sets requirements for the sterility of the interior of evacuated containers, and for non-evacuated containers that claim sterility or a special microbiological state. Consider that clause and any sterility claim on the label together when deciding the route. For sterile devices, Section 6.5(a) of Annex II also requires the technical documentation to describe the packaging, sterilisation and sterility maintenance methods, with the validation reports.

On 16 December 2025 the European Commission proposed a regulation to simplify the MDR and IVDR (procedure 2025/0404(COD)). As of 13 November 2026, the European Parliament's Legislative Observatory shows it as awaiting committee decision. Provisions quoted here are from the IVDR text in force on 13 November 2026.

Is ISO 6710 a harmonised standard under the IVDR?

No. Under Article 8(1), only harmonised standards, or relevant parts of them, whose references are published in the Official Journal give a presumption of conformity with the requirements they cover. IVD harmonised standards were listed in Implementing Decision (EU) 2021/1195 of 19 July 2021, since amended several times. As of 13 November 2026, the latest amendment on the European Commission harmonised standards page is (EU) 2026/2481 of 9 November 2026, and ISO 6710 appears neither in the summary list on that page (generated on 11 November 2026) nor in the texts of (EU) 2025/679, 2026/197, 2026/1313 and 2026/2481. The list mainly covers horizontal standards on sterilisation, aseptic processing, packaging, quality management, risk management, symbols (EN ISO 15223-1:2021) and information supplied by the manufacturer (the EN ISO 18113 series).

Not harmonised does not mean unusable. Section 4 of Annex II requires a justification, validation and verification of the solutions adopted, and the technical documentation must set out:

  • the applicable GSPRs, and why the others do not apply;
  • the method or methods used to demonstrate conformity with each applicable GSPR;
  • the harmonised standards, common specifications (CS) or other solutions applied;
  • the precise identity of the controlled documents providing the evidence, cross-referenced to their location in the technical documentation.

ISO 6710 is an "other solution". An ISO 6710 report can serve as evidence, with its number in the controlled-document column of the GSPR checklist, but the manufacturer has to explain why the method is adequate for the requirement in question. It should not be presented as "conforms to a harmonised standard, therefore presumed to conform".

Writing the GSPR checklist: ISO 6710 clauses against IVDR Annex I

For each ISO 6710:2017 clause, the table gives a summary, Annex I sections that can be linked in a GSPR checklist, and whether SUNGO's certificates cover it. The links are one way to build the checklist, not an official correspondence from the standard or the regulation; the manufacturer's risk analysis and intended purpose decide the final mapping.

ISO 6710:2017 clauseSummaryAnnex I sections that can be linked (examples)SUNGO certificates
Clause 4 Materials: 4.1, 4.3Tube material allows visual inspection of the contents; container free from foreign matterSection 10, chemical, physical and biological propertiesWithin CNAS and CMA scope
Clause 4 Materials: 4.2Containers for determining a specific element or substance, with related information stated by the manufacturerSection 10.1Not in scope
Clause 5 Draw volume, Annex A, Annex BDraw volume tests for non-evacuated and evacuated containersSections 6 and 7, characteristics and performance not adversely affected over the lifetime or by transport and storageWithin CNAS and CMA scope
Clause 6 DesignClosure does not loosen during mixing; how the closure is removed; compatibility with transport systems and pre-analytical and analytical automationSection 11.1, safe handling, reduced microbial leakage and specimen contamination; Section 13.1, use with other devices or equipmentWithin CNAS and CMA scope
Clause 7 ConstructionNo breaking, cracking or leaking when centrifuged; no sharp edges, projections or rough surfaces that could injure the userSection 11.1; Section 18, protection against mechanical risksWithin CNAS and CMA scope
Clause 8 Sterility and special microbiological statesSterile interior of evacuated containers and the related validated processSections 11.2 to 11.4Not in scope
Clause 9 Additives, Annex ENominal amount of additive, mixing and related pointsTo be argued by additive and intended purposeNot in scope
Clause 10 Marking and labellingInformation on the container and primary pack, label adherence, fill indicator and related pointsSection 20, label and instructions for use (in part)Within CNAS and CMA scope
Clause 11 Container identificationIdentification of additives and accessories by letter code and similar meansSection 13.2(f), risk of incorrect identification and erroneous results from confusing colour or character codings on specimen receptaclesNot in scope

Note two points. First, Clause 10 of ISO 6710 checks the markings listed in the standard. Section 20.2 of Annex I asks for more on the label, such as an indication that the device is an in vitro diagnostic medical device, the authorised representative's name and address for manufacturers outside the Union, and the UDI carrier; the manufacturer has to check these against the regulation, and an ISO 6710 test does not demonstrate them. Second, Clause 11 links directly to Section 13.2(f) and Clause 8 to Sections 11.2 to 11.4, but neither clause is in SUNGO's certificates, so other evidence is needed.

ISO 6710 clauses in SUNGO's accreditation scope

The table is taken from the Shanghai CNAS schedule (test object: single-use containers for human venous blood specimen collection) and the Shanghai CMA schedule. Item names and clause numbers are as on the certificates; neither schedule states a limitation for these items.

ItemStandard and clauseNotes / limitations
MaterialsISO 6710:2017 4.1, 4.3; YY/T 0314-2021 4.1, 4.3Listed under CNAS and CMA; no limitation stated; only 4.1 and 4.3 of Clause 4 are listed
Draw volumeISO 6710:2017 Clause 5, Annex A, Annex B; YY/T 0314-2021 Clause 5, Annex A, Annex BListed under CNAS and CMA; no limitation stated
DesignISO 6710:2017 Clause 6; YY/T 0314-2021 Clause 6Listed under CNAS and CMA; no limitation stated
ConstructionISO 6710:2017 Clause 7; YY/T 0314-2021 Clause 7Listed under CNAS and CMA; no limitation stated
Marking and labellingISO 6710:2017 Clause 10; YY/T 0314-2021 Clause 10Listed under CNAS and CMA; no limitation stated
  • Clause numbers come from the CNAS schedule. The CMA schedule lists the standards without clause numbers; its item names match the CNAS schedule.
  • Annexes C, D, E and F of ISO 6710:2017 are not listed separately on the certificates, and Clauses 8, 9 and 11 are not included.
  • The certificates list only ISO 6710:2017 and YY/T 0314-2021.

For other items, send the standard and clause numbers; we check them against the certificates and will say plainly if they are outside the accredited scope. If the technical documentation also involves biological tests, SUNGO subcontracts them to a laboratory with the appropriate qualifications, with the client's prior agreement, and the report identifies the subcontracted items and the subcontractor.

Where SUNGO tests and which report marks apply

  • Location: the Shanghai laboratory, No. 958 Kangqiao East Road, Pudong New Area, Shanghai. Reports are issued by Shanghai SUNGO Medical Technology Co., Ltd.
  • CNAS (L22083): all five items, for both standards, are within the accredited scope. CNAS is a signatory to the ILAC MRA.
  • CMA (240015340087): the five items are in the CMA schedule. CMA covers only scheduled items, and its certificate states that it is valid within the People's Republic of China.
  • IAS (TL-1019): the Shanghai IAS scope does not include ISO 6710 or YY/T 0314, so no IAS reports are issued for these items.
  • Hefei: Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity whose certificates do not include these standards, so tube samples are tested in Shanghai.

For EU technical documentation you can request the CNAS mark; within one report, the CNAS mark applies only to items in the CNAS schedule. For what a mark on a report shows and how to check a scope, see How to Check an Accreditation Scope and Read the Mark on a Report. Whether a notified body or EU customer accepts a report depends on their requirements. For the United States, the FDA recognized consensus standards check is in the tube testing article above.

Preparing samples and documents

These are general points; details depend on the items confirmed when quoting.

  • Standard and purpose: state whether you need ISO 6710:2017, YY/T 0314-2021 or both, list the items and clauses, say the report is for EU IVDR technical documentation with the CNAS mark, and give any report language or format requirements.
  • Model list: for each model, the name, size (nominal capacity or draw volume), evacuated or non-evacuated, type of additive and accessory, and how it is identified on the closure or label.
  • Labels, primary packs and instructions for use: marking and labelling is checked against Clause 10, so provide the versions used on the EU market.
  • Sample condition: preferably one batch, intact packaging, within shelf life; state the batch number, manufacturing date, sterilisation status and storage conditions, and ship as the instructions require.
  • Accessories for testing: if tests need needles, holders or similar items, agree who supplies them and which models.
  • Client details: manufacturer name and address, product names and models should match the technical documentation and the EU declaration of conformity.
  • Sample quantity: depends on the standards and items; we give it clause by clause when quoting.

FAQ

What class is a blood collection tube under the IVDR?

It is a specimen receptacle as defined in Article 2(3) and is Class A under Annex VIII Rule 5(c). MDCG 2020-16 rev.5 notes that specimen receptacles cannot have an integrated testing function.

Do sterile evacuated blood collection tubes need a notified body under the IVDR?

Under Article 48(10), Class A devices placed on the market sterile follow the Annex IX or Annex XI procedure, with notified body involvement limited to establishing, securing and maintaining sterile conditions. Non-sterile Class A devices are self-declared.

Is ISO 6710 a harmonised standard under the IVDR?

Not as of 13 November 2026: it is not on the IVD harmonised standards list, including the amendment of 9 November 2026. An ISO 6710 report can serve as evidence in the technical documentation but gives no presumption of conformity.

How are blood collection tubes with additives classified?

The MDCG explanation of Rule 5(c) refers to empty tubes and tubes prefilled with a fixative solution or other general reagent, and does not separately address tubes with additives intended for a specific examination. Check the intended purpose against Annex VIII of the IVDR and the MDCG text.

Which accreditation marks can SUNGO's ISO 6710 reports carry?

Materials, draw volume, design, construction, and marking and labelling are within the Shanghai CNAS and CMA scopes; CMA is valid only within the People's Republic of China. The Shanghai IAS scope does not include ISO 6710, so no IAS report is issued.

Request a quotation

To have blood collection tubes tested to ISO 6710:2017 for EU technical documentation, see our other medical device testing services, then use the contact page to send your model list, the items and clauses needed, the purpose of the report and the accreditation mark required. We first confirm the scope item by item against the Shanghai CNAS and CMA schedules, then reply with a quotation and sample requirements.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.