When a surgical gauze pad, abdominal pad or gauze swab goes through registration testing in China, the product technical requirements almost always cite YY 0594-2006, General requirements for surgical gauze dressings. Manufacturers then ask: which standard covers the finished dressing and which covers the gauze? Why does YY 0594 repeat "fluorescent substances" and "minimum breaking force" when the gauze standard already has them? Can a 2006 standard still be used? This article answers those questions, lists the YY 0594 items in the certificate scopes of SUNGO's Shanghai laboratory, and explains what to prepare before sending samples.
Key points
- YY 0594-2006 is still a current mandatory industry standard. As of 22 October 2026, China's national standards information platform lists it as "current". It was issued on 19 April 2006 and took effect on 1 April 2007. Its Amendment No. 1, published by the former State Food and Drug Administration (SFDA) in Announcement No. 75 of 2010, rewrites clause 4.3 and deletes clause 5.4.4.
- YY 0594 covers the finished dressing; YY/T 0331 covers the gauze. YY 0594 applies to dressings made from absorbent cotton gauze or cotton and viscose gauze; the gauze itself is covered by YY/T 0331, and as of 22 October 2026 the current edition is YY/T 0331-2024.
- Same item names, different test objects. In YY 0594, fluorescent substances are checked on the sewing thread and the abdominal pad tape, and minimum breaking force applies to the tape. In YY/T 0331, the same item names apply to the gauze fabric. One set of data cannot stand in for the other.
- Seven YY 0594-2006 items are in the Shanghai laboratory's certificate scopes: dyeing (4.2), folding and stitching (4.3), ethylene oxide residue (4.5), fluorescent substances (5.3.1, 5.4.2, Annex D), material (5.4.1), minimum breaking force (5.4.3) and stitches (5.4.4). All seven are in the CNAS accreditation scope and on the CMA schedule. YY 0594 is not in the IAS scope.
- Other requirements, such as the X-ray detectable component and colour fastness, are not among these seven items and need to be commissioned separately from a suitably qualified body.
The standard and its scope
What YY 0594-2006 covers
YY 0594-2006 is a mandatory pharmaceutical industry standard issued by the former SFDA and administered by the Jinan Medical Device Quality Supervision and Testing Center. According to its scope, it sets general requirements, plus packaging and marking requirements, for surgical gauze dressings made from absorbent cotton gauze or absorbent cotton and viscose gauze. The dressings may be supplied sterile or non-sterile, and the main product types are abdominal pads, gauze swabs (rolls or balls) and gauze pads. The type and content of antibiotics in antibiotic-containing surgical gauze are outside its scope.
The NMPA Guideline for Registration Technical Review of Surgical Gauze Dressings (2018 Revision) (the "Guideline") describes these products. This article cites the 2018 revision; if NMPA issues a newer version, that version applies. In summary, an abdominal pad is a rectangular or square dressing made for abdominal surgery, with no cut edges exposed and all four sides stitched; a tape may be sewn to one corner. A gauze swab is a roll or ball of gauze. A gauze pad is gauze folded into a rectangle or square with no cut edges exposed. Some products also carry an X-ray detectable component. The Guideline applies to these products as Class II devices and states that the latest edition of each cited standard applies.
YY 0594 and YY/T 0331: finished product and raw material
YY/T 0331-2024, Performance requirements and test methods for absorbent cotton gauze and absorbent cotton and viscose gauze, sets requirements for the gauze itself. It was issued on 7 February 2024, took effect on 1 March 2025 and fully replaces YY/T 0331-2006. Its introduction states that the requirements for surgical gauze dressings made mainly from gauze conforming to it are given in YY 0594. The Guideline also requires the gauze used in surgical dressings to meet YY 0331. Its text cites YY 0331-2006, but it also says the latest edition applies.
A complete test plan for a surgical gauze dressing therefore usually has two layers. The raw gauze is tested to YY/T 0331 for items such as fibre identification, acidity or alkalinity, fluorescent substances, thread count, mass per unit area, minimum breaking force and sinking time. The finished dressing is tested to YY 0594 for product-level requirements such as folding and stitching, dyeing, the sewing thread and the tape. The Shanghai laboratory has YY/T 0331 items in its CNAS, CMA and IAS scopes, but the editions and items differ: the CNAS scope lists some YY/T 0331-2024 items, the CMA schedule lists items from both YY/T 0331-2006 and YY/T 0331-2024, and the IAS scope lists the whole standard without an edition. A separate article covers YY/T 0331-2024. State in the product technical requirements which edition the raw gauze is tested to, and confirm it with the laboratory.
What Amendment No. 1 changed
The former SFDA published Amendment No. 1 to YY 0594-2006 in Announcement No. 75 of 2010. It took effect on the date of publication and makes two changes:
- Clause 4.3, folding and stitching, is rewritten as two sub-clauses. Clause 4.3.1 requires the folding or stitching to leave no cut edge of the gauze exposed. Clause 4.3.2 requires at least 24 stitches per 10 cm of stitching.
- Clause 5.4.4 is deleted.
The Guideline includes the same cut-edge and stitch requirements. Product technical requirements that still copy the pre-amendment text will cite clause numbers that no longer match, so write "YY 0594-2006 (including Amendment No. 1)" and cite the amended clause numbers.
What comes next
On the YY 0594-2006 page of the national standards information platform, the "similar standards (plans)" section lists national standard plan 20250884-T-469, Medical gauze dressings. It is a plan to draft a new recommended national standard under the National Working Group for Standardization of Medical Protective Devices (SWG30). The plan was issued on 27 March 2025, was at the approval stage when we checked on 22 October 2026, and is due to take effect 12 months after publication. The platform does not say whether it will replace YY 0594; rely on the official text and announcement.
How the Guideline lists the YY 0594 requirements
Under "performance requirements in YY 0594-2006" the Guideline lists the gauze raw material, dyeing, and folding or stitching. EO residue is listed separately under chemical performance. The X-ray detectable component, colour fastness, the sewing thread (fluorescent substances and water-soluble substances) and the abdominal pad tape (material, fluorescent substances and minimum breaking force) are listed as special requirements, which must meet Chapter 5 of YY 0594. The certificate clause numbers place dyeing, folding and stitching, and EO residue at 4.2, 4.3 and 4.5. Whether each clause applies to a given product is governed by the text of the standard. Annexes B to E are normative. Annex D is the fluorescence test for sewing thread and abdominal pad tapes, and Annex E is the test for water-soluble substances in sewing thread and tapes.
The Guideline also requires the registration file to explain why any clause does not apply. The table below summarises coverage.
| Requirement | Products usually concerned | Coverage in SUNGO's scope |
|---|---|---|
| Gauze raw material | All products | Tested to YY/T 0331; some YY/T 0331 items are in the Shanghai CNAS and CMA scopes, and the IAS scope lists the standard without an edition (see separate article) |
| Dyeing (4.2) | Dyed products | Among the 7 YY 0594 items in the CNAS and CMA scopes |
| Folding and stitching (4.3) | All products | Among the 7 items |
| Ethylene oxide residue (4.5) | EO-sterilized products | Among the 7 items |
| Fluorescent substances in sewing thread (5.3.1) | Stitched products | Among the 7 items |
| Abdominal pad tape: material, fluorescent substances, minimum breaking force (5.4.1, 5.4.2, 5.4.3) | Abdominal pads with a tape | Among the 7 items |
The middle column summarises the Guideline's descriptions. Linking 5.3.1 to the sewing thread and 5.4.1 to 5.4.3 to the tape comes from comparing the Guideline's order with the certificate's clause numbers; applicability and exact wording follow the standard. The X-ray detectable component, colour fastness, and water-soluble substances in sewing thread and tapes (Annex E) are not among the seven certified items and need to be commissioned separately from a suitably qualified body.
How the seven in-scope items are tested
Fluorescent substances (5.3.1, 5.4.2, Annex D)
Annex D is a visual check under ultraviolet light, with two tests. In Test A, the sample is examined under a 365 nm UV lamp. It should show only faint brownish-violet fluorescence and a few yellow particles, and apart from a few isolated fibres it should not show intense blue fluorescence. Test B applies only to polymer sewing thread that does not meet Test A. The thread is soaked in water at 37 °C for 24 h, the liquid is sieved off, and a set volume of the extract is placed in a glass colorimetric tube. Under the 365 nm lamp, its fluorescence must not exceed that of the same volume of water. Sample mass, sieve size and tube volume are given in Annex D.
The certificate lists two clauses for this item, 5.3.1 and 5.4.2, together with Annex D; by its title, Annex D covers the sewing thread and the abdominal pad tape. Two practical points:
- The sewing thread (5.3.1) and the tape (5.4.2) are separate clauses with separate conclusions; a gauze fluorescence result under YY/T 0331 (clause 5.5) cannot replace them.
- Test A is a qualitative judgement based on the colour of the observed fluorescence. Annex D specifies a 365 nm UV lamp; other viewing conditions are set by the laboratory's work instructions.
Minimum breaking force (5.4.3)
This item applies to the abdominal pad tape. The specimen form, test conditions and acceptance value follow the text of clause 5.4.3.
YY/T 0331-2024 also has a "minimum breaking force" item (clause 5.8), but it measures the gauze itself, which is a different test. Clause 5.4.3 does not apply to gauze pads and swabs without a tape.
Material (5.4.1)
Clause 5.4.1 sets requirements for the material of the abdominal pad tape; what is checked follows the text of clause 5.4.1. Supplying the tape's raw material specification (yarn and fabric type) helps confirm the test basis.
Folding and stitching (4.3) and stitches (5.4.4)
After Amendment No. 1, clause 4.3 has two parts: clause 4.3.1 requires that no cut edge is exposed, and clause 4.3.2 requires at least 24 stitches per 10 cm for stitching on the dressing. For gauze pads and swabs the main question is whether any cut edge is exposed after folding; abdominal pads are stitched on all four sides, so the stitch requirement applies to them. The test method follows the standard.
The certificate still lists "stitches" under its original clause number, 5.4.4. Amendment No. 1 sets the stitch requirement in clause 4.3.2 and deletes clause 5.4.4, so confirm with the laboratory which clause the report will cite, so that it matches the product technical requirements.
Dyeing (4.2)
The dyeing clause applies only to dyed products, such as the green gauze pads and green abdominal pads shown in the Guideline's product images. The exact requirement is in clause 4.2. For undyed products, the product technical requirements can state that the clause does not apply and why. The Guideline lists colour fastness as a special requirement (to meet Chapter 5 of YY 0594), and colour fastness is not among the seven items on SUNGO's certificate.
Ethylene oxide residue (4.5)
Clause 4.5 applies only to products sterilized with ethylene oxide (EO). The Guideline sets an upper limit for EO residue in EO-sterilized surgical gauze dressings, expressed in μg/g. The introduction to YY/T 0331-2024 notes that gauze is highly adsorbent and does not recommend EO sterilization for gauze. If a dressing is still EO-sterilized, residue and aeration need close control.
Which method standard clause 4.5 cites should be checked against the text of YY 0594. The Guideline's list of related standards includes both GB/T 14233.1 and GB/T 16886.7. Under the test object "gauze dressings", the Shanghai CNAS and CMA scopes list YY 0594-2006 clause 4.5. In addition, EO residue to GB/T 14233.1-2022 is listed under "absorbent cotton gauze" in the CNAS scope (clause 9.4) and under "gauze" on the CMA schedule (no clause given). A separate article covers the GB/T 14233.1 Clause 9 GC method and the effect of sampling time and aeration.
Accreditation scope table
The table below reproduces every YY 0594-2006 entry under the test object "gauze dressings" in the CNAS scope (L22083) of Shanghai SUNGO Medical Technology Co., Ltd. The CMA schedule (240015340087) lists the same seven item names under the same object but gives no clause numbers.
| Item | Clause | Notes / limitations |
|---|---|---|
| Dyeing | YY 0594-2006 4.2 | CNAS, CMA; no limitation stated |
| Folding and stitching | YY 0594-2006 4.3 | CNAS, CMA; no limitation stated |
| Ethylene oxide residue | YY 0594-2006 4.5 | CNAS, CMA; no limitation stated |
| Fluorescent substances | YY 0594-2006 5.3.1, 5.4.2, Annex D | CNAS, CMA; no limitation stated |
| Material | YY 0594-2006 5.4.1 | CNAS, CMA; no limitation stated |
| Minimum breaking force | YY 0594-2006 5.4.3 | CNAS, CMA; no limitation stated |
| Stitches | YY 0594-2006 5.4.4 | CNAS, CMA; no limitation stated |
Note: the clause number for "stitches" is as printed on the certificate. Amendment No. 1 sets the stitch requirement in clause 4.3.2 and deletes clause 5.4.4; see above for how the report clause is handled.
Where the work is done and under which accreditation
The seven YY 0594-2006 items are tested by the laboratory of Shanghai SUNGO Medical Technology Co., Ltd., at No. 958 Kangqiao East Road, Pudong New Area, Shanghai. All seven are in the CNAS scope (L22083) and on the CMA schedule (240015340087), so reports can carry the CNAS or CMA mark. CMA covers only the items on its schedule, and the certificate states that it is valid within the People's Republic of China. YY 0594 is not in the Shanghai IAS scope (TL-1019), so no IAS report can be issued for it. Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity, and none of its CNAS, CMA or IAS scopes includes gauze or dressing tests. For what each mark on a report proves and how to check a scope, see How to Check an Accreditation Scope and Read the Mark on a Report. The full scopes of both laboratories are summarised in SUNGO Medical Device Testing Scope: CNAS, CMA and IAS in Shanghai and Hefei.
On biocompatibility, the Guideline requires surgical gauze dressings to undergo biological evaluation according to GB/T 16886.1. Biological tests such as cytotoxicity, sensitization and irritation are subcontracted by SUNGO to appropriately qualified laboratories. Under accreditation rules, the client's agreement is obtained before subcontracting, and the report identifies the subcontracted items and the subcontractor.
Preparing samples
- Product technical requirements (or a draft). State which clauses of YY 0594-2006 and Amendment No. 1 are cited, which do not apply, and why.
- Product description. Say whether the product is dyed, stitched, fitted with an abdominal pad tape or fitted with an X-ray detectable component, and give the materials of the sewing thread and the tape.
- Representative samples. Within one registration unit, follow the Guideline and choose the model with the most functions and the most complex construction. The Guideline's example is a sterile, dyed, stitched abdominal pad with an X-ray detectable component and a tape.
- Sterilization details. State the sterilization method. For EO sterilization, give the sterilization date and aeration conditions; samples should be finished products that have completed sterilization and aeration.
- Sample form and quantity. The laboratory works these out from the test items and product sizes. For example, the laboratory will specify the breaking-force specimen form according to clause 5.4.3, and fluorescence testing needs samples from which the sewing thread and the tape can be taken.
- Raw gauze data. If YY/T 0331 data on the raw gauze are also needed, give the gauze specification and the edition of the standard, and supply raw gauze samples separately if required.
Frequently asked questions
YY 0594-2006 dates from 2006. Can it still be used for registration testing?
Yes. As of 22 October 2026 it is listed as a current mandatory industry standard. The national standard plan Medical gauze dressings is at the approval stage; its relationship to YY 0594 will be set by the official announcement.
The raw gauze has already been tested for fluorescence and breaking force to YY/T 0331. Does the finished dressing need testing again?
It depends on the construction. YY 0594 tests fluorescence on the sewing thread and tape, and breaking force on the tape, not on the gauze. For products without sewing thread or a tape, these clauses may not apply, and the product technical requirements should say so.
The certificate says "stitches, 5.4.4", but the amendment deleted 5.4.4. What will the report say?
Amendment No. 1 sets the stitch requirement (at least 24 stitches per 10 cm) in clause 4.3.2 and deletes clause 5.4.4. When placing the order, confirm with the laboratory which clause number the report will cite, so that it matches your product technical requirements.
Which YY 0594 items apply to an undyed gauze pad with no stitching?
Decide clause by clause. Dyeing, sewing-thread and tape clauses generally do not apply, and the product technical requirements should say why. The no-exposed-cut-edge requirement still applies, and EO residue applies if the product is EO-sterilized. The raw gauze is tested separately to YY/T 0331.
Can a YY 0594 report be used for export?
YY 0594 is not in the IAS scope, so only CNAS or CMA reports can be issued, and CMA is valid only within China. The CNAS certificate states that CNAS is a signatory to the ILAC and APAC arrangements, but whether a particular overseas regulator accepts a report is for the client to confirm with that regulator.
If your surgical gauze dressing needs registration or commissioned testing, send us the product technical requirements, product description and sterilization method. We will check which clauses fall within the certificate scopes and which need separate arrangements, then propose a test plan and quotation. Our service scope is on the masks, gloves and dressings testing page, and you can request a quotation through the contact page.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.