FDA recognizes AAMI PB70:2022 only in part, under recognition number 6-487. Three parts are not recognized: the "procedure gown" definition in Clause 3, Clause 4.1 Labeling, and informative Annex B. The Level 1 to Level 4 liquid barrier classification and its test methods remain within the recognized portion. For companies preparing a surgical gown 510(k), this means the classification tests are planned as usual, but the product name, the labeling claims and the critical zone drawing should follow 21 CFR 878.4040 and FDA guidance rather than the excluded parts of the standard. This article walks through the FDA database entries and guidance text and explains how to plan testing.
Key Takeaways
- As of November 13, 2026, a search of the FDA Recognized Consensus Standards database by designation "PB70" returns a single record: 6-487, ANSI/AAMI PB70:2022, extent of recognition "Partial", date of entry May 29, 2023.
- Three parts are not recognized: the "procedure gown" definition in Clause 3, Clause 4.1 Labeling, and informative Annex B (barrier classification examples for gowns and drapes, with Figure B.2 singled out by FDA).
- The Level 1 to Level 4 classification requirements and test methods are not excluded. ASTM F1671/F1671M-22 (viral penetration), commonly found in the barrier data for Level 4 surgical gowns, is fully recognized by FDA (6-488).
- ASTM F1670/F1670M has two versions in the FDA database: -24a (6-518, complete recognition) and -17a (6-491). FDA will accept declarations of conformity to -17a until December 19, 2027.
- Labeling should follow FDA's 2015 final guidance and clearly identify the PB70 liquid barrier level. FDA treats gowns claiming moderate or high barrier protection, such as Level 3 or Level 4, as "surgical gowns" under 21 CFR 878.4040: Class II, 510(k) required.
- SUNGO Shanghai's CNAS scope covers only three AAMI PB70:2022 items for surgical gowns: synthetic blood penetration, hydrostatic pressure and water resistance (impact penetration). ASTM F1671 is covered only by SUNGO Shanghai's IAS scope, so a Level 4 surgical gown classification that includes F1671 can only be reported under IAS.
What AAMI PB70:2022 Covers and What FDA Recognition 6-487 Says
ANSI/AAMI PB70 is titled "Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities." The FDA 6-487 record describes its scope as establishing minimum barrier performance requirements, a classification system and associated labeling requirements for protective apparel, surgical drapes and drape accessories used in health care facilities.
Key fields on the 6-487 Supplementary Information Sheet:
| Field | Entry |
|---|---|
| Recognition number | 6-487 |
| Standard | ANSI AAMI PB70:2022 |
| Date of entry | May 29, 2023 (Recognition List 060) |
| Extent of recognition | Partial recognition |
| Not recognized | Clause 3 Definitions, Procedure gown; 4.1 Labeling; Annex B (informative) |
| Specialty task group | General Plastic Surgery/General Hospital |
Regulation numbers and product codes listed on the record are shown below. FDA notes that these are examples and others may apply.
| Regulation | Device name | Class | Product code |
|---|---|---|---|
| 21 CFR 878.4040 | Gown, Surgical | Class II | FYA |
| 21 CFR 878.4040 | Gown, Isolation, Surgical | Class II | FYC |
| 21 CFR 878.4040 | Non-Surgical Gown | Class I | OEA |
| 21 CFR 878.4040 | Gown, Non-Sterile, Non-Isolation, Intended To Provide Moderate Or High Barrier Protection | Class II | QPC |
| 21 CFR 878.4040 | Medical Gowns With Chemotherapy Labeling Claims - Tested For Use With Chemotherapy Drugs | Class II | QSO |
| 21 CFR 878.4370 | Surgical drapes and accessories (e.g., Drape, Surgical, Ent; Drape, Surgical, Antimicrobial; General Surgery Tray) | Class II | ERY, EYX, EYY, FNW, HMT, HMW, KGW, KKX, LRO |
| 21 CFR 878.4380 | Drape, Adhesive, Aerosol | Class I | KGT |
Partial recognition rests on section 514(c) of the Federal Food, Drug, and Cosmetic Act, under which FDA may recognize all or part of a standard, and a submitter may file a Declaration of Conformity (DOC) to what is recognized. FDA's September 2018 guidance "Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices" states that a DOC certifies conformity to all requirements of the recognized standard, with no deviations. If a submitter deviates from the standard, or relies on a standard FDA has not recognized, a DOC cannot be submitted, and the standard can only be cited as "general use" with supporting information. In practice, the three excluded parts of PB70:2022 cannot serve as a basis for declaring conformity to FDA; those topics go back to the regulation and FDA guidance.
Which Parts of PB70:2022 FDA Does Not Recognize
Clause 3: the "procedure gown" definition
FDA's rationale: the definition conflicts with 21 CFR 878.4040 and the December 2015 final guidance "Premarket Notification Requirements Concerning Gowns Intended for Use in Health Care Settings." The record states that FDA considers procedure gowns claiming moderate to high barrier protection to be "surgical gowns" within the meaning of 21 CFR 878.4040.
What this means: calling a product a "procedure gown" does not, by itself, change its regulatory category at FDA. The 2015 guidance lists the factors FDA considers: whether the gown is labeled as a surgical gown, whether its labeling describes it as one, whether it carries moderate or high barrier protection claims, and whether it is stated to be for use during sterile procedures. Gowns that fall within these factors are regulated as Class II "surgical gowns" and require a 510(k). The guidance also explains that a gown is not treated as a surgical gown only when all three of the following apply: it is labeled as a gown other than a surgical gown (for example, an isolation gown); its labeling does not describe it as a surgical gown; and any barrier claims are limited to minimal or low protection (PB70 Level 1 or Level 2). FDA decides how a specific product is classified. Manufacturers can assess their product against the guidance and, where needed, consult FDA through a Pre-Submission.
Clause 4.1: Labeling
FDA's rationale: Clause 4.1 conflicts with FDA regulations and the 2015 final guidance, which states that the level of liquid barrier protection must be clearly identified in labeling.
What this means: do not base labeling claims on PB70:2022 Clause 4.1 alone; check them item by item against Section IV of the 2015 guidance. The guidance lists what a surgical gown 510(k) should contain:
- Evidence that the gown complies with the claimed barrier performance criteria of the FDA-recognized version of PB70, or an equivalent standard;
- Performance test data showing the gown is an effective barrier per the PB70 barrier performance specifications;
- Engineering drawings, schematics, illustrations or figures labeled with the barrier protection levels, dimensions and the location of critical and non-critical zones;
- Sample labeling that clearly identifies the level of liquid barrier protection per PB70;
- Sample labeling that includes directions for use and indications for use.
The guidance also discourages claims such as "Fluid-Resistant," "Splash Resistant," "Impervious" and "Impermeable," because they lack specificity regarding performance characteristics and test methods.
Annex B: informative classification examples
Annex B gives barrier performance classification examples for surgical gowns, surgical gowns-E, isolation gowns, other protective gowns, surgical drapes, hoods, togas and other protective apparel. FDA singles out Figure B.2, "Example of a surgical gown-E intended for surgical applications," as conflicting with Figure 1, "Critical Zones for Surgical Gowns," on the FDA Medical Gowns web page; Annex B as a whole is listed as not recognized.
What this means: draw surgical gown critical zones according to the FDA web page. It states that the critical zones include the front of the body from the top of the shoulders to the knees and the arms from the wrist cuff to above the elbow, comprising at least areas A and B; the entire front of the gown (areas A, B and C) must have a barrier performance of at least Level 1; and the back (area D) may be non-protective. For surgical isolation gowns and non-surgical isolation gowns, all areas except bindings, cuffs and hems are critical zones that must meet the highest level claimed, and all seams must have the same liquid barrier protection as the rest of the gown.
How FDA Classifies Surgical and Isolation Gowns
21 CFR 878.4040 (surgical apparel) places surgical gowns and surgical masks in Class II (special controls) and other surgical apparel in Class I, exempt from 510(k) subject to the limitations in 878.9. The FDA Medical Gowns page describes the gown types as follows (FDA decides the classification of a specific product):
| Type | FDA description | Regulatory status |
|---|---|---|
| Surgical gown | Worn by health care personnel during surgical procedures; must be provided sterile and labeled as a surgical gown | Class II, 510(k) required |
| Surgical isolation gown | Used when there is a medium to high risk of contamination and a need for larger critical zones than traditional surgical gowns | Class II, 510(k) required |
| Non-surgical isolation gown | Low or minimal risk patient isolation; not worn during surgical or invasive procedures or at medium to high risk of contamination | Class I, exempt from premarket review |
| Non-sterile, non-isolation gown with moderate or high barrier protection | Non-sterile and non-patient-isolation situations, moderate or high barrier protection | Class II, 510(k) required |
| Non-surgical, non-isolation gown | Non-surgical and non-patient-isolation situations, minimal or low barrier protection | Class I, exempt from premarket review |
In short, FDA looks at the name, the labeling description, the barrier level claimed and whether the gown is for sterile procedures; renaming a gown does not change its class. 6-487 excludes the PB70 "procedure gown" definition precisely because it conflicts with 21 CFR 878.4040 and the 2015 guidance.
PB70 Level 1 to 4 Test Methods and FDA-Recognized Versions
6-487 does not exclude the classification requirements or test methods. In its background section, FDA's 2015 guidance also states that FDA has recognized PB70's barrier performance levels (Levels 1 to 4) and the associated test methods, while noting that PB70's definitions and terminology are inconsistent with FDA's historical definitions. Level 1 to Level 4 testing therefore remains the core of the barrier performance data in a 510(k); what is excluded is the definition, the labeling clause and the examples.
The scope of ASTM F2407 (the surgical gown specification) notes that PB70 evaluates the barrier properties of surgical gown fabrics using water only in Levels 1, 2 and 3, and that, because surgical gowns are exposed to blood and other fluids with different surface tensions, additional testing with simulated biological fluids is required for a Level 4 gown. The water-based tests are AATCC 42 impact penetration (water resistance) and AATCC 127 hydrostatic pressure; the simulated-fluid tests are ASTM F1670 synthetic blood penetration and ASTM F1671 viral penetration (using Phi-X174 bacteriophage as the test system). Public 510(k) summaries on the FDA website for surgical gowns claiming Level 4 under PB70:2022 list AATCC TM42, AATCC TM127 and ASTM F1671/F1671M-22 among the barrier tests. Which tests apply at each level, how gowns and drapes differ, and the acceptance and sampling requirements should be taken from the PB70:2022 text. This article does not list pass/fail values.
Status of these methods in the FDA database as of November 13, 2026:
| Method | FDA recognition number and version | Status |
|---|---|---|
| ASTM F1671/F1671M (viral penetration) | 6-488, F1671/F1671M-22 | Complete recognition, entered May 29, 2023 |
| ASTM F1670/F1670M (synthetic blood penetration) | 6-518, F1670/F1670M-24a | Complete recognition, entered December 22, 2025 |
| ASTM F1670/F1670M | 6-491, F1670/F1670M-17a | Superseded by 6-518; FDA accepts declarations of conformity to 6-491 until December 19, 2027, and not after |
| AATCC TM42, AATCC TM127 | No separate recognition record found by designation search | Referenced in PB70 as classification test methods; check the edition in the PB70:2022 text |
The 6-488 record lists product codes FYA, FYC, OEA, QPC and QSO, while 6-518 lists only KKX (antimicrobial surgical drape). A search by product code FYA also returns ASTM F2407/F2407M-23a (6-516, partial recognition) among other standards; F2407 physical property testing is covered in a separate article on this site. How the AATCC 127 hydrostatic pressure method compares with GB/T 4744 and EN ISO 811 is also covered in a separate article on this site.
SUNGO Accreditation Scope for PB70-Related Testing
The table reproduces the PB70-related rows from SUNGO Shanghai's certificates. The three CNAS rows all list "surgical gown" as the test object, with no clause number and no limitation note. The IAS rows cover whole standards and carry no edition year.
| Item | Standard / clause | Notes / limitations |
|---|---|---|
| Synthetic blood penetration | AAMI PB70:2022; ASTM F1670/F1670M-24 | Shanghai CNAS (L22083), test object: surgical gown; no clause listed |
| Hydrostatic pressure | AAMI PB70:2022; AATCC TM127-2017(2018)e | Shanghai CNAS (L22083), test object: surgical gown; no clause listed |
| Water resistance (impact penetration) | AAMI PB70:2022; AATCC TM42-2017e | Shanghai CNAS (L22083), test object: surgical gown; no clause listed |
| Whole standard | ANSI/AAMI PB70 | Shanghai IAS (TL-1019), no edition year; confirm the edition with the lab before reporting to the 2022 edition |
| Whole standard | ASTM F1670/F1670M | Shanghai IAS (TL-1019), no edition year; confirm with the lab before reporting to -24a or -17a |
| Whole standard | ASTM F1671/F1671M | Shanghai IAS (TL-1019), no edition year; not covered by CNAS or CMA; confirm with the lab before reporting to -22 |
| Whole standard | AATCC TM 42 | Shanghai IAS (TL-1019), no edition year |
| Whole standard | AATCC TM 127 | Shanghai IAS (TL-1019), no edition year |
Where SUNGO Performs It and Under Which Accreditation
All of the above is performed by Shanghai SUNGO Medical Technology Co., Ltd. (laboratory at 958 Kangqiao East Road, Pudong, Shanghai) under CNAS L22083 and IAS TL-1019. SUNGO is accredited by CNAS and IAS, both signatories to the ILAC MRA. Hefei SUNGO Medical Technology Co., Ltd. does not offer these tests. CMA does not cover PB70, ASTM or AATCC standards; CMA covers only the items listed in its schedule and is valid only within China.
- When the claimed level involves only water resistance (AATCC TM42) and hydrostatic pressure (AATCC TM127) (check the PB70:2022 text): both are among the three CNAS items, with surgical gowns as the test object, so either a CNAS or an IAS report can be chosen.
- Level 4 surgical gown claims: the barrier data usually include ASTM F1671 viral penetration. At SUNGO, F1671 is covered only by the Shanghai IAS scope, so a full classification that includes F1671 can only be reported under IAS.
- ASTM F1670 edition: CNAS accreditation is for F1670/F1670M-24, while the FDA database lists -24a and -17a; the IAS scope carries no year. For an FDA submission, confirm with the lab which edition will be tested and reported before shipping samples.
- SUNGO's certificates do not list isolation gowns as a test object, and this article does not address isolation gown testing capability.
Whether a report is accepted is decided by FDA during review. For other surgical gown tests, see Surgical Gown Testing; for the full scope at both sites, see SUNGO Medical Device Testing Scope: CNAS, CMA and IAS in Shanghai and Hefei.
Preparing Samples and Information
- Settle the claims first: product name, intended PB70 level, and whether the intended use involves surgery or sterile procedures. These determine the gown's FDA category and the tests needed, and are decided by the manufacturer and its regulatory advisers.
- Prepare a critical zone drawing: mark areas A, B, C and D as in Figure 1 of the FDA Medical Gowns page, with the materials, laminate layers and seam locations in each area. Do not use the PB70:2022 Annex B examples.
- Sample condition: the 2015 FDA guidance recommends barrier performance testing on the final, finished, pre-shipment gown at the end of its stated shelf life; for reusable gowns, also at the end of the labeled use-life (maximum reprocessing cycles). Plan aged or reprocessed samples in advance.
- State the editions: on the test request, specify the PB70 edition and each test method edition (for example F1670/F1670M-24a or -17a, F1671/F1671M-22), and whether a CNAS or IAS report is needed.
- Sample quantities and specimen locations follow PB70:2022 and the referenced methods; the lab confirms them based on the claimed level and the critical zone layout.
FAQ
Does FDA recognize AAMI PB70:2022?
Yes, partially (6-487). Only the Clause 3 "procedure gown" definition, Clause 4.1 Labeling and informative Annex B are excluded; the Level 1 to 4 classification and test methods remain recognized.
Does a PB70 Level 4 surgical gown require ASTM F1671 testing?
According to the ASTM F2407 scope, a PB70 Level 4 gown needs simulated-fluid testing in addition to the water tests, and public 510(k) summaries on the FDA website for gowns claiming Level 4 under PB70:2022 list ASTM F1671/F1671M-22 viral penetration; refer to the PB70:2022 text for the exact requirements. F1671/F1671M-22 is fully recognized by FDA (6-488), and at SUNGO it can only be reported under IAS.
Can a "procedure gown" label avoid a 510(k)?
The name alone does not decide it. FDA considers procedure gowns claiming moderate to high barrier protection to be surgical gowns, regulated as Class II with a 510(k); FDA decides the classification of a specific product.
Are test reports to PB70:2012 still usable?
As of November 13, 2026, a designation search for PB70 in the FDA database returns only the 2022 edition (6-487). FDA's 2018 guidance states that superseded standards withdrawn from recognition cannot be used in declarations of conformity. Whether an older report supports a new submission is up to FDA; the safer course is to plan testing to the 2022 edition and the FDA-recognized method versions.
Can SUNGO's CNAS report cover all PB70 levels?
No. CNAS covers only three AAMI PB70:2022 items for surgical gowns (synthetic blood penetration, hydrostatic pressure, water resistance) and does not include ASTM F1671. A Level 4 surgical gown classification that includes F1671 can only be reported under IAS.
To arrange PB70 classification testing, send us the product name, intended barrier level, material construction and critical zone drawing, target standard editions, and whether you need a CNAS or IAS report. We will first confirm which report types are available within our accreditation scope, then provide a test plan and quotation. Please submit your inquiry through our contact page.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.