Defibrillation-Proof Applied Parts: Testing to IEC 60601-1 Clause 8.5.5

Defibrillation-Proof Applied Parts: Testing to IEC 60601-1 Clause 8.5.5

ECG leads, temperature probes and SpO2 sensors stay attached to the patient, and in an emergency the patient may be defibrillated while they are connected, so the defibrillator's high voltage reaches the equipment. Will the enclosure, signal ports or other applied parts carry hazardous energy? Will the equipment still work afterwards? Will it absorb energy meant for the patient? Clause 8.5.5 of IEC 60601-1 (GB 9706.1 in China) answers these questions with two tests: 8.5.5.1 defibrillation protection and 8.5.5.2 the energy reduction test. This article covers only those two tests: marking, test set-ups, judging results, submission preparation, and which SUNGO laboratory performs the work under which accreditation. For how to classify applied parts as type B, BF or CF in the first place, see How to Classify ME Equipment and Identify Applied Parts.

Key points

  • Defibrillation-proof is an optional classification. The general standard requires every applied part to be classified as type B, BF or CF and allows the manufacturer to also classify it as defibrillation-proof. Once an applied part is classified and marked this way, Clause 8.5.5 applies, and the classification covers the applied part in its entirety.
  • The 8.5.5.1 defibrillation protection test has a common-mode and a differential-mode set-up. The equipment is powered and operating, and each test is repeated with the test voltage polarity reversed.
  • There are two pass criteria: no hazardous electrical energy on the enclosure, signal input/output parts, other applied parts' patient connections and the other listed points during the discharge; and basic safety and essential performance maintained after the recovery time stated in the accompanying documents.
  • The 8.5.5.2 energy reduction test checks that the equipment does not absorb too much of the defibrillation energy. The energy delivered to a 100 Ω load with the equipment connected must be at least 90 % of the energy delivered with the equipment disconnected. The cables, electrodes and transducers recommended in the instructions for use are used.
  • Any additional defibrillation requirements in particular standards for ECG and monitoring equipment are outside SUNGO's accreditation scope. This article covers only Clause 8.5.5 of the general standard.
  • At SUNGO, IEC 60601-1, EN 60601-1 and GB 9706.1 testing is performed by the Hefei laboratory, within its IAS accreditation (TL-1119). The certificate lists these as whole standards without edition years, so please confirm the edition for your report with the laboratory before placing the order.

Standards and scope

Three designations and their editions

IEC 60601-1 is the general standard for basic safety and essential performance of medical electrical equipment. As of 2026-10-22, the IEC Webstore lists edition 3.2 as current, the consolidated version IEC 60601-1:2005+AMD1:2012+AMD2:2020 CSV, published on 2020-08-20. GB 9706.1-2020 is the mandatory Chinese national standard. It was issued on 2020-04-09, took effect on 2023-05-01, is a modified adoption of IEC 60601-1:2012, and fully replaces GB 9706.1-2007 and GB 9706.15-2008. EN 60601-1, the European adoption, is as of the same date EN 60601-1:2006 with amendments A1, A12, A2 and A13.

The test steps and values here follow the official Korean text published by Korea's Ministry of Food and Drug Safety (MFDS) for IEC 60601-1:2012 (edition 3.1), which is the edition GB 9706.1-2020 adopts. For any edition, the original IEC or GB text you apply governs the exact wording and values. IEC TR 62348:2012 rates the Amendment 1 changes to 8.5.5.1 a) and 8.5.5.2 as having "significant impact". Products tested to IEC 60601-1:2005 (edition 3.0, before Amendment 1) should revisit both subclauses when moving to a later edition.

What a defibrillation-proof applied part is

Clause 3.20 defines a defibrillation-proof applied part as an applied part that is protected against the effects of a discharge of a cardiac defibrillator to the patient. Clause 6.2 requires type B, BF or CF classification and allows an additional defibrillation-proof classification. Both 8.5.2.1 and 8.5.5.1 state that this classification applies to an applied part in its entirety: some patient connections of one applied part cannot be protected while others are not.

The Annex A rationale makes two further points. First, a defibrillation-proof applied part is protected only against discharges from defibrillators that comply with IEC 60601-2-4, the particular standard for cardiac defibrillators. Higher-voltage defibrillators can damage it. IEC 60601-2-4 is outside SUNGO's accreditation scope. Second, a patient electrode may sit right next to a defibrillator paddle, so its exposure cannot be predicted. Without an applicable particular standard, such electrodes and the equipment they connect to must withstand the full no-load voltage of a fully charged defibrillator, so the general standard specifies 5 kV d.c. as the test voltage.

Which products are asked about it

The question usually comes up for equipment and accessories that stay connected to the patient where defibrillation may be performed. Examples are ECG equipment and its lead wires and electrodes, and monitoring equipment that connects to temperature or SpO2 probes. Start with the label: if an applied part carries a defibrillation-proof symbol, Clause 8.5.5 applies; if you do not intend to claim defibrillation protection, do not use the symbol. Whether a product type must be defibrillation-proof may also be set by its particular standard, and those particular standards are outside SUNGO's accreditation scope.

Marking and accompanying documents

  • Symbols: Under 7.2.10, defibrillation-proof applied parts use IEC 60417-5841 (defibrillation-proof type B applied part), IEC 60417-5334 (defibrillation-proof type BF applied part) or IEC 60417-5336 (defibrillation-proof type CF applied part). The symbol goes on or next to the connector, or on the applied part if there is no connector; where one connector serves applied parts with different classifications, each applied part is marked.
  • Protection partly in the patient cable: If part of the defibrillation protection is in the patient cable, safety sign ISO 7010-W001 goes next to the relevant outlet. The instructions for use must explain that the appropriate cable has to be used to keep the protection. Per the rationale, the wrong cable can reduce the defibrillation energy delivered to the patient, cause loss of essential performance through damage to the equipment, or give an electric shock to the operator or other persons.
  • Recovery time: 8.5.5.1 b) refers to the recovery time stated in the accompanying documents, and Table C.4 in Annex C lists it among the accompanying document items. According to the rationale, if no recovery time is stated, the equipment shall recover basic safety and essential performance immediately. The general standard itself does not set a recovery time value.

8.5.5.1 Defibrillation protection test

The requirement

The arrangement that isolates the patient connections of a defibrillation-proof applied part from the rest of the equipment shall meet two conditions.

a) While a defibrillator discharges into a patient connected to the defibrillation-proof applied part, hazardous electrical energy shall not appear at the points listed below. The check is the peak voltage measured between points Y1 and Y2 in Figures 9 and 10. A peak above 1 V means the requirement is not met.

  • the enclosure, including patient leads and cable connectors when connected to the equipment (this item does not apply to leads or connectors of the defibrillation-proof applied part that are not connected to the equipment);
  • signal input/output parts;
  • a test metal foil on which the equipment is placed, with an area at least equal to the base of the equipment;
  • patient connections of other applied parts, whether or not they are defibrillation-proof;
  • any unused or disconnected connections of any function of the same applied part, or of the applied part under test. Here the clause notes that wholly body-worn equipment (for example, a Holter recorder) is excluded from this requirement.

b) After the defibrillation voltage has been applied and the recovery time stated in the accompanying documents has passed, the equipment shall comply with the relevant requirements of the standard and maintain basic safety and essential performance.

Why test unused connectors? The rationale's example: ECG and invasive blood pressure (IBP) share one applied part, and only ECG is in use while the IBP connector is open. If the patient is defibrillated and the open connector's pins are not adequately recessed, a second operator adjusting the equipment could be shocked from it.

Common-mode test (Figure 9)

The equipment is connected to the circuit in Figure 9. The test voltage is applied to all patient connections of the defibrillation-proof applied part together, excluding any parts that are protectively or functionally earthed. If one applied part has several functions, the test voltage is applied to all patient connections of one function, with the connections of the other functions disconnected.

Differential-mode test (Figure 10)

The equipment is connected to the circuit in Figure 10. The test voltage is applied to each patient connection of the defibrillation-proof applied part in turn, with all remaining patient connections of the same applied part earthed. The differential-mode test does not apply when the applied part has only one patient connection.

Set-up and operating conditions

  • Except for permanently installed equipment, the equipment is tested with and without the protective earth conductor connected, as two separate tests.
  • Insulating surfaces of the applied part are covered with metal foil or, where appropriate, immersed in 0.9 % saline solution.
  • All external connections to the functional earth terminal are removed.
  • The non-protectively-earthed parts listed in a) are connected in turn to the oscilloscope.
  • The equipment is connected to its supply and operated according to the instructions for use.

After switch S is operated, the peak voltage between Y1 and Y2 is measured. Each test is repeated with the polarity of VT reversed. After the recovery time stated in the accompanying documents, the laboratory checks that basic safety and essential performance are maintained. Circuit component values are in Figures 9 and 10 and Table 5 of the standard. Per the rationale, inductance L gives a shorter-than-normal rise time so that integrated protection is adequately tested.

How results are judged

There are two layers. The first is energy: if the peak voltage between Y1 and Y2 exceeds 1 V at any measurement point, the requirement is not met. The rationale explains that this limits the shock to a person touching accessible parts during the discharge to a level that is unpleasant but not dangerous, corresponding to a charge of 100 μC. The second is recovery. The equipment is tested while operating because the test covers both basic safety and the ability to deliver essential performance after the stated recovery time. The laboratory checks function against the manufacturer's essential performance list and verification method, so prepare both.

The rationale also notes that a broken functional earth connection is fairly likely in practice, which is why external functional earth connections are removed.

8.5.5.2 Energy reduction test

The requirement: the defibrillation-proof applied part or its patient connections shall incorporate a means such that the defibrillator energy delivered to a 100 Ω load is at least 90 % of the energy delivered to that load with the equipment disconnected. Protection that diverts much of the defibrillation current leaves less energy for the patient; this test limits that loss.

The test uses the circuit in Figure 11 and the accessories recommended in the instructions for use (see 7.9.2.14), such as cables, electrodes and transducers. Where alternative accessories are specified, 5.5 e) calls for testing with the one that gives the least favourable condition. The test voltage is applied to each patient connection or applied part in turn, with the remaining patient connections of the same applied part earthed (differential mode). Any other defibrillation-proof applied parts are tested individually in turn. The procedure:

  1. Connect the applied part or patient connection to the test circuit, with the parts listed in 8.5.5.1 a) earthed.
  2. With switch S in position A, charge capacitor C to 5 kV d.c.
  3. Move switch S to position B to discharge C, and measure the energy E1 delivered to the 100 Ω load.
  4. Remove the equipment under test from the circuit, repeat steps 2 and 3, and measure the energy E2 delivered to the 100 Ω load.
  5. Check that E1 is at least 90 % of E2.
  6. Repeat with the test voltage polarity reversed.

Links to other clauses

  • Creepage distances and air clearances: 8.9.1.15 requires the creepage distances and air clearances needed to meet 8.5.5.1 to be at least 4 mm. Common problems are covered in Electrical Safety Failures: Creepage, Earthing and Leakage Current.
  • Working voltage and dielectric strength: Under 8.5.4, the working voltage of a defibrillation-proof applied part is determined without regard to the possible defibrillation voltage. The rationale explains that the special tests in 8.5.5 are considered to give adequate assurance against defibrillation pulses, so a separate dielectric strength test at defibrillation voltage is not needed.
  • Lower test voltage: According to the rationale, a manufacturer may use 4.5 to set a lower test voltage based on the intended use and where the applied part is attached to the patient. To do so, the manufacturer must show that the chosen voltage is the maximum that appears at the applied part when 5 kV is applied to the chest. Such parts may still be classified and marked as defibrillation-proof. Record the justification in the risk management file.
  • Test sequence: In the sequence recommended by informative Annex B, defibrillation protection (B.23) comes after humidity preconditioning (B.21) and after the cold leakage current and dielectric strength tests that follow it (B.22).
  • Particular standards: Particular standards for ECG, monitoring and similar equipment may modify or supplement the requirements of Clause 8.5.5. Those particular standards are outside SUNGO's accreditation scope.

Accreditation scope

Each row below is taken from the scope annex of the Hefei laboratory's IAS certificate (TL-1119). Item names, clauses and limitations are reproduced as listed:

StandardItemClauseNotes / limitations
IEC 60601-1Whole standardNo clause listedNo limitation listed; no edition year
EN 60601-1Whole standardNo clause listedNo limitation listed; no edition year
GB 9706.1Whole standardNo clause listedNo limitation listed; no edition year

Where SUNGO performs the work and under which accreditation

  • The 8.5.5 defibrillation protection and energy reduction tests to IEC 60601-1, EN 60601-1 and GB 9706.1, and the follow-up basic safety checks, are performed by the Hefei laboratory, within its IAS accreditation (TL-1119). The Hefei laboratory belongs to Hefei SUNGO Medical Technology Co., Ltd., an independent legal entity in Zipeng Town, Feixi County, Hefei, Anhui. The scope of Shanghai SUNGO Medical Technology Co., Ltd. does not include the standards discussed here.
  • The IAS certificate lists the standards without edition years. If you need a report to IEC edition 3.2, GB 9706.1-2020 or a particular EN version, please confirm the edition with the laboratory before placing the order.
  • The only particular standards on the certificate are ISO 80601-2-69, EN ISO 80601-2-69 and YY 9706.269 for oxygen concentrators. Any additional defibrillation requirements in ECG and monitoring particular standards are outside SUNGO's accreditation scope.
  • Acceptance of a report for registration in any country or region, including China, depends on the relevant regulator; SUNGO makes no commitment.

Preparing your submission

  1. Applied part list: For each applied part, list its patient connections, the functions they serve, which functions share an applied part, and the classification and marking location of each.
  2. Accessories: Supply every patient cable, lead set, electrode and transducer model recommended in the instructions for use. If part of the protection sits in a cable, supply that cable and check the ISO 7010-W001 sign and the cable warning in the instructions.
  3. Recovery time and essential performance: State the recovery time in the accompanying documents. Provide the essential performance list and explain how function is confirmed after the test, including any simulator, software screen or acceptance method needed.
  4. Accessible parts and interfaces: List conductive enclosure parts, signal input/output parts, functional earth terminals and other applied parts, so the laboratory can define the oscilloscope measurement points.
  5. Insulation diagram: Mark the defibrillation isolation barrier and attach a list of creepage distances and air clearances.
  6. Samples and sequence: Defibrillation pulses may damage protective components. Agree sample numbers with the laboratory, and whether samples are shared with leakage current, dielectric strength and other tests, following the Annex B sequence.
  7. Risk management file: If you use a lower test voltage or another alternative approach, record the justification in the risk management file.

FAQ

Our applied part has no defibrillation-proof symbol. Does Clause 8.5.5 still apply?

Under the general standard, defibrillation-proof is an optional classification, and Clause 8.5.5 applies to applied parts classified and marked as such. A particular standard may require some product types to be defibrillation-proof, and those standards are outside SUNGO's accreditation scope.

Are both the common-mode and the differential-mode tests needed?

Yes, except that the differential-mode test does not apply to an applied part with a single patient connection. Both set-ups are repeated with reversed polarity.

Which cable is used for the energy reduction test?

The accessories recommended in the instructions for use. Where several alternatives exist, the least favourable one is used, so bring all recommended accessories with the samples.

What recovery time should we state?

The general standard does not set a value. It requires basic safety and essential performance to be maintained after the recovery time stated in the accompanying documents, and immediate recovery if none is stated. Any recovery-time requirements in particular standards are outside SUNGO's accreditation scope.

Can we test to Clause 8.5.5 only?

The IAS certificate lists IEC 60601-1 as a whole standard. Whether Clause 8.5.5 can be evaluated alone, and how the report would state that scope, should be confirmed when you request a quote.

Requesting a quote

When requesting a quote, please provide: product name and intended use, models and configurations, the applied part list with classifications, recommended accessories, the stated recovery time, the essential performance list, target markets and intended standard editions, plus the instructions for use and the risk management file's table of contents. The Hefei laboratory will use these to confirm the test items, sample arrangement and test sequence. See medical electrical safety testing for the service overview, and send your request through the contact page.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.