Radiated RF immunity testing asks whether a medical electrical device keeps its basic safety and essential performance in an RF field from a distant transmitter such as a hand-held radio or broadcast station. This article explains how the test is run: calibrating the uniform field area, placing the device and its cables, how many faces the antenna sees, and how step size, dwell time and modulation are set. A table then compares the immunity levels for the professional healthcare facility environment and the home healthcare environment, and shows how a home-use product picks its column from the instructions for use.
The short answer
- Division of roles: IEC 61000-4-3 (adopted identically in China as GB/T 17626.3) defines the method and recommended levels only. Field strength and pass/fail logic for a medical device come from IEC 60601-1-2 or YY 9706.102.
- IEC 60601-1-2 Editions 4.0/4.1 grade by environment: enclosure-port radiated immunity is 3 V/m for professional healthcare and 10 V/m for home healthcare, both 80 MHz to 2.7 GHz with 80 % AM at 1 kHz. Other immunity differences are few: conducted RF immunity adds the amateur radio bands at home, and vehicle-mounted equipment has an ISO 7637-2 requirement on its d.c. power port in the home column. In Edition 4.1 (Amendment 1 added 8.11, proximity magnetic fields) the 30 kHz test is home-only.
- YY 9706.102-2021 splits differently: 3 V/m or 10 V/m by whether the equipment is life-supporting, up to 2.5 GHz, with 2 Hz modulation for equipment that controls, monitors or measures a physiological parameter.
- Hard method rules: at least 75 % of 16 uniform-field points within 0 dB to +6 dB; steps no larger than 1 % of the preceding frequency; dwell at least 0.5 s; both polarizations on every face; at least 1 m of cable in the field.
- Choosing the environment: if the instructions for use allow home use, test to home healthcare levels. Tables 4 to 8 tested at home levels need not be repeated at professional levels.
Standards and scope
IEC 61000-4-3 / EN IEC 61000-4-3 / GB/T 17626.3
IEC 61000-4-3:2020 is Edition 4.0, published on 2020-09-08 by IEC TC 77/SC 77B as a basic EMC publication. It covers RF sources not in close proximity to the equipment (close-proximity sources are in IEC 61000-4-39) and defines testing above 80 MHz, limited only by the test instrumentation. It replaces the 2006 edition and its 2007 and 2010 amendments. The IEC webstore abstract lists changes including multiple-signal testing, removal of the upper frequency limit, and specified amplifier linearity checks.
EN IEC 61000-4-3:2020 is based on IEC 61000-4-3:2020. In China, GB/T 17626.3-2023 is a recommended national standard, published 2023-12-28 and implemented 2024-07-01, identical to IEC 61000-4-3:2020 and replacing GB/T 17626.3-2016. Clause 5 gives recommended levels 1 to 4 plus level X.
IEC 60601-1-2 / EN 60601-1-2
Edition 4.0 of this EMC collateral standard for medical electrical equipment was published on 2014-02-25 and Edition 4.1 (consolidated with Amendment 1:2020) on 2020-09-01. The fourth edition sets immunity levels by environment of intended use, with locations harmonized with IEC 60601-1-11. Radiated immunity is in Table 4 (enclosure port) of subclause 8.9. EN 60601-1-2:2015/A1:2021 is based on IEC 60601-1-2:2014/A1:2020.
Clause 2 of Editions 4.0/4.1 cites IEC 61000-4-3:2006 with its two amendments as a dated reference, for which only the cited edition applies, and under 1.3.2 a particular standard takes priority over the collateral standard. The method figures below are from IEC 61000-4-3:2020; check differences from the 2006 edition in the standard text, and state the edition in the test plan.
YY 9706.102-2021
YY 9706.102-2021 is a mandatory Chinese industry standard, published 2021-03-09 and implemented 2023-05-01, replacing YY 0505-2012 and marked as a modified adoption of IEC 60601-1-2:2007. Radiated immunity is in 6.2.3, graded by whether the equipment is life-supporting. GB/T 17626.3 is cited undated, so the latest edition applies.
How the test is run
Test site, antennas and the uniform field area
IEC 61000-4-3 Clause 6 requires a shielded enclosure, to avoid interfering with radio services and to keep instruments out of the field. Typically this is an anechoic or modified semi-anechoic chamber, with a signal generator, power amplifier, linearly polarized antennas (biconical, log-periodic, horn), an isotropic field probe and forward-power measurement.
The field is calibrated beforehand without the EUT. The uniform field area (UFA) is a vertical plane, preferably 1.5 m × 1.5 m, with a 0.5 m grid of 16 points. At each frequency the field is uniform if at least 75 % of the points (12 of 16) are within 0 dB to +6 dB of nominal, so the field does not fall below nominal. Up to 1 GHz, +10 dB is allowed for at most 3 % of frequencies if stated in the report; in a dispute, 0 dB to +6 dB prevails.
If the EUT and cabling fit a smaller area, the UFA can be as small as 0.5 m × 0.5 m with a fifth centre point, and all five points must comply. If the EUT face is larger than 1.5 m × 1.5 m and a large enough UFA (the preferred method) cannot be realised, partial illumination may be used, moving the antenna or EUT, with a full level setting for each antenna position. The preferred antenna distance is 3 m, minimum 1 m, the same for calibration and test. Calibration uses an unmodulated carrier in both polarizations, at a field at least 1.8 times the test field to show the amplifier will not saturate under 80 % AM. The standard intends full level setting at least annually and after chamber changes.
Placing the EUT
- Set up as installed in practice, wired per the instructions for use, covers fitted. No metallic ground plane; supports bulk non-conductive, and low-permittivity material such as rigid polystyrene above 1 GHz.
- Table-top equipment on a non-conductive table (0.8 ± 0.05) m high; floor-standing equipment on a non-conductive support 0.05 m or more above the floor, with any deviation recorded.
- The illuminated face coincides with the UFA plane.
Four faces, two polarizations
Under 8.3 the antenna normally faces each side of the EUT. All sides are exposed if the equipment can be used in several orientations; fewer faces need technical justification, and depending on EUT type and size or the test frequencies, more than four azimuths may need to be exposed. Each face is tested with the antenna vertical and again horizontal, with the EUT exercised through its critical modes.
Step size and dwell time
Under 8.4 each frequency step is at most 1 % of the preceding frequency, for calibration and test. Under 8.3 the dwell time is no less than the time the EUT needs to be exercised and respond, and never less than 0.5 s. The 2020 edition allows multiple simultaneous frequencies, but care is needed to attribute any degradation to a single signal.
IEC 60601-1-2 8.1 bases dwell time on test-system settling and device response. YY 9706.102-2021 sets at least 3 s for 2 Hz modulation, at least 1 s otherwise, and at least 1.2 times the averaging period for averaging devices. For slow devices (YY Annex A cites pulse oximeters, infusion pumps and ventilators) the product sets the dwell time, not the minimum.
Modulation: 80 % AM at 1 kHz, and 2 Hz
IEC 61000-4-3 Table 1 levels are unmodulated carrier values; the test uses a 1 kHz sine at 80 % AM, with (80 ± 10) % depth and 1 kHz ± 0.1 kHz at the generator. IEC 60601-1-2 Edition 4.1 Table 4 gives "80 % AM at 1 kHz" for both environments, and footnote c) allows other modulation frequencies identified by risk management; Annex F explains these can replace or supplement the Table 4 modulation where a device is particularly sensitive. Table 4 has no separate modulation for physiological-signal equipment.
YY 9706.102-2021 6.2.3.2 c) and Table 9 differ: equipment that controls, monitors or measures a physiological parameter uses 2 Hz modulation, with physiological simulation or operating frequencies below 1 Hz or above 3 Hz; other equipment uses 1 kHz; 2 Hz testing needs no extra 1 kHz run. Annex A places 2 Hz between the typical passbands of respiratory and cardiac equipment. A monitoring product may need different settings under each standard.
Cable layout
- Use specified cables and connectors; otherwise unshielded parallel conductors.
- Specified lengths of 1 m or less are used as specified; longer or unspecified lengths follow typical installation.
- At least 1 m of cable is exposed vertically or horizontally, close to the UFA; runs orthogonal to the UFA do not count.
- Excess cable is bundled low-inductively near its middle; each cable is in the UFA in at least one orientation.
- Cables leaving the test area may be decoupled with common mode absorption devices, recorded in the report.
For medical devices, IEC 60601-1-2 Table 4 footnote a) puts any patient physiological simulator interface within 0.1 m of the UFA plane in one orientation. YY 9706.102-2021 6.2.3.2 h) repeats the 0.1 m rule and requires patient-coupled cables at maximum permitted length, no intentional conductive or capacitive connection from the patient coupling point to ground, and unintended stray capacitance preferably not above 250 pF.
Back to essential performance
IEC 61000-4-3 Clause 9 classifies results a) to d), from normal performance to non-recoverable loss. For medical devices IEC 60601-1-2 applies: before testing, the manufacturer sets specific pass/fail criteria for basic safety and essential performance, from the particular standard or risk management, and decides how to monitor them. Radiated immunity is a continuous phenomenon, so criteria apply during and after the test. IEC TS 60601-4-2 gives guidance on assessing performance under electromagnetic disturbances.
Equipment that intentionally receives RF energy, such as a product with a Bluetooth module, is also tested at its receive frequency (Table 4 footnote b)); the test assesses basic safety and essential performance, and normal reception might not be achieved. If a device fails, see EMC Failures: Locating the Cause, Fixing It, Scoping the Retest.
Professional healthcare or home healthcare?
IEC 60601-1-2 5.2.1.1 a) requires the instructions for use to state the suitable environments and any exclusions from risk analysis. Figure 3 examples: professional healthcare covers physician and dental offices, clinics, limited care facilities, freestanding surgical and birthing centres, and hospital emergency rooms, patient rooms, intensive care and surgery rooms (except near HF surgical equipment and inside an MRI RF shielded room). Home healthcare covers residences and nursing homes, vehicles, stations and airports, hotels, shops, schools and outdoors. Military areas, heavy industry and high-power treatment areas are special environments (Annex E).
Under 8.1, the most stringent level applies when several environments are intended; Tables 4 to 8 tested at home levels need no repeat at professional levels; and the emergency medical services environment (an ambulance, for example) uses home levels. For the US, FDA partially recognizes Edition 4.1 (FR Recognition Number 19-36) and does not recognize "nursing homes" in Figure 3 as a home healthcare example.
So under Editions 4.0/4.1, an oxygen concentrator, nebulizer or blood pressure monitor whose instructions allow home use is tested at 10 V/m; one limited to hospitals and clinics, with no special environment, at 3 V/m. The instructions, risk file and test plan must agree. A particular standard that modifies EMC requirements governs; see Oxygen Concentrator Testing: ISO 80601-2-69, IEC 60601-1 and IEC 60601-1-2. Safety requirements of IEC 60601-1-11 are covered in a separate article.
Immunity levels: professional vs home healthcare
From IEC 60601-1-2 Edition 4.1 Tables 4 to 6 and 8.10/8.11; RF levels are before modulation. Conducted RF on d.c. power, patient coupling and SIP/SOP ports splits the same way. Applicability is as stated in the table footnotes.
| Phenomenon (port) | Basic standard | Professional healthcare | Home healthcare | Notes |
|---|---|---|---|---|
| Electrostatic discharge (enclosure) | IEC 61000-4-2 | ±8 kV contact; ±2, ±4, ±8, ±15 kV air | Same | Same |
| Radiated RF EM fields (enclosure) | IEC 61000-4-3 | 3 V/m; 80 MHz to 2.7 GHz; 80 % AM at 1 kHz | 10 V/m; 80 MHz to 2.7 GHz; 80 % AM at 1 kHz | This article |
| Proximity fields from RF wireless equipment (enclosure) | IEC 61000-4-3 | Per 8.10, Table 9 | Same | One table; separate article |
| Rated power frequency magnetic fields (enclosure) | IEC 61000-4-8 | 30 A/m; 50 Hz or 60 Hz | Same | Only with magnetically sensitive parts |
| Proximity magnetic fields (enclosure) | IEC 61000-4-39 | Per 8.11, Table 11 | Same, plus 30 kHz | Added in Edition 4.1; 30 kHz home only; separate article |
| Electrical fast transients/bursts (a.c. power) | IEC 61000-4-4 | ±2 kV; 100 kHz repetition | Same | Same; separate article |
| Surges (a.c. power) | IEC 61000-4-5 | Line-to-line ±0.5, ±1 kV; line-to-ground ±0.5, ±1, ±2 kV | Same | Same |
| Conducted RF disturbances (a.c. power) | IEC 61000-4-6 | 3 V, 0.15 MHz to 80 MHz; 6 V in ISM bands; 80 % AM at 1 kHz | 3 V, 0.15 MHz to 80 MHz; 6 V in ISM and amateur radio bands; 80 % AM at 1 kHz | Home adds amateur bands |
| Voltage dips and interruptions (a.c. power) | IEC 61000-4-11 | 0 % UT 0.5 cycle (8 phase angles); 0 % UT 1 cycle; 70 % UT 25/30 cycles; interruption 0 % UT 250/300 cycles | Same | Same; separate article |
| Transient conduction along supply lines (d.c. power) | ISO 7637-2 | Not applicable | As specified in ISO 7637-2 | Only for equipment in vehicles (including ambulances) with 12 V or 24 V systems; ISO 7637-2 is not in SUNGO's accredited scope |
By contrast, YY 9706.102-2021 sets radiated immunity at 3 V/m (non-life-supporting) or 10 V/m (life-supporting) over 80 MHz to 2.5 GHz, and conducted immunity at 3 V with an extra 10 V in ISM bands for life-supporting equipment. A home-use, non-life-supporting product is 10 V/m under IEC 60601-1-2 Edition 4.1 but 3 V/m under YY 9706.102-2021, so settle the market and standard first.
Accredited scope: Hefei laboratory, IAS (TL-1119)
Seven relevant entries from the IAS scope (TL-1119) of Hefei SUNGO Medical Technology Co., Ltd.; Item reproduces the certificate's English title. No clauses, limitations or years are listed.
| Item | Basis | Notes / limitations |
|---|---|---|
| Electromagnetic compatibility (EMC) - Part 4-3: Testing and measurement techniques - Radiated, radio-frequency, electromagnetic field immunity test | IEC 61000-4-3 (no clauses listed) | None listed; no year |
| Electromagnetic compatibility (EMC) - Part 4-3: Testing and measurement techniques - Radiated, radio-frequency, electromagnetic field immunity test | EN IEC 61000-4-3 (no clauses listed) | None listed; no year |
| Electromagnetic compatibility - Testing and measurement techniques - Part 3: Radiated, radio-frequency, electromagnetic field immunity test | GB/T 17626.3 (no clauses listed) | None listed; no year |
| Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | IEC 60601-1-2 (no clauses listed) | None listed; no year |
| Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | EN 60601-1-2 (no clauses listed) | None listed; no year |
| Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic compatibility - Requirements and tests | YY 9706.102 (no clauses listed) | None listed; no year |
| Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems | IEC TS 60601-4-2 (no clauses listed) | None listed; no year |
Where SUNGO does it, and under which accreditation
At SUNGO, radiated immunity is tested by Hefei SUNGO Medical Technology Co., Ltd.: Hefei laboratory, within its IAS accreditation scope (TL-1119), with reports issued within that scope. It is an independent legal entity with its laboratory in Zipeng Town, Feixi County, Hefei, Anhui. Shanghai SUNGO Medical Technology Co., Ltd., a separate legal entity with its laboratory at 958 Kangqiao East Road, Pudong, Shanghai, does not have these EMC standards in its scope. The basic standards for the other phenomena in the table (IEC 61000-4-2, -4-4, -4-5, -4-6, -4-8, -4-11, -4-39 and their European and Chinese designations) are also in the Hefei IAS scope.
The IAS scope gives no years. IEC 61000-4-3, IEC 60601-1-2 and GB/T 17626.3 each have several editions; confirm the edition with the laboratory before an IAS report is issued. SUNGO makes no commitment that any regulator will accept a report; whether it can support registration or certification in a given market, including Chinese registration testing, should be confirmed in advance with the regulator or notified body.
Preparing for testing
- Standard and edition: IEC 60601-1-2 Edition 4.0 or 4.1, EN 60601-1-2 or YY 9706.102, and the IEC 61000-4-3 or GB/T 17626.3 edition.
- Environment: the environment statement and exclusions from the instructions for use; for YY 9706.102, whether the equipment is life-supporting.
- Particular standard: clauses and edition of any Part 2 standard that modify EMC requirements.
- Criteria and monitoring: quantified, observable essential-performance criteria, with deviations, alarms, recovery and monitoring method.
- Modes and modulation: modes to be tested and why; whether a physiological parameter is measured, monitored or controlled; any extra modulation from risk management.
- Wireless: receive bands and partner equipment, so the link runs during the test.
- Cables and accessories: specified cables, sensors, patient simulator and loads, with types, lengths and layout.
- Placement: table-top or floor-standing; usable orientations.
FAQ
We tested a home-use product at 10 V/m. Do we also need 3 V/m?
No. IEC 60601-1-2 8.1 states that Tables 4 to 8 tested at home healthcare levels need not be repeated at professional levels.
IEC 60601-1-2 and YY 9706.102 give different levels. Which applies?
That depends on the target market and applicable standard. If both apply, the test plan lists field strength, upper frequency and modulation for each.
Our Bluetooth link drops at its own receive frequency. Is that a fail?
Table 4 footnote b) says the test assesses basic safety and essential performance, and normal reception might not be achieved. The verdict depends on whether the link carries essential performance and on the criteria set beforehand.
Is 10 V/m measured after modulation?
No. IEC 61000-4-3 Table 1 and IEC 60601-1-2 Table 4 give unmodulated carrier values; 80 % AM is applied on top.
For radiated immunity or other IEC 60601-1-2 / YY 9706.102 EMC tests, see our EMC testing service and send an enquiry via the contact page, with the product model, target market, intended standards and editions, the environment statement, draft pass/fail criteria and a cable list, so the laboratory can confirm the plan and report edition.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.