For natural rubber latex examination gloves sold into the US, ASTM D3578 is the specification commonly cited in 510(k) files, purchase contracts and inspection checklists. A common question is about water-extractable protein: which method measures it, how to prepare samples, and why results come back high. This guide lists the D3578 test items and referenced methods, focuses on protein, briefly covers aged and unaged tensile, leaks and powder, and sets out SUNGO Shanghai's CNAS and IAS scope. It covers latex gloves only; nitrile gloves fall under ASTM D6319, covered in a separate article. External information was checked on 2026-10-22. Numerical requirements are not reproduced here; the ASTM text governs.
Key points
- D3578 covers natural rubber gloves only. The current edition is ASTM D3578-19(2023), Standard Specification for Rubber Examination Gloves. Gloves are Type I or Type II, and both surfaces must be free of talc.
- The test items fall into a few groups. Apart from sterility for sterile products: freedom from holes, dimensions, tensile before and after aging, powder and protein, using ASTM D5151, D3767, D412, D573, D6124 and D5712, with ISO 2859 for sampling.
- Three things decide a protein result: the condition of the gloves (FDA suggests recently made finished gloves after accelerated or real-time aging); whether the method is D5712 or one correlated with it; and whether leaching, washing, curing and chlorination are controlled. Results are in µg/dm²; the D3578 requirement is in the specification text.
- SUNGO Shanghai's scope: under the test object "rubber examination gloves", the CNAS schedule lists D3578-19(2023) clause 8.3 (water impermeability test), clauses 8.4 and 8.5 (physical and dimensional properties), clauses 8.6 and 8.8 (residual powder) and clause 8.7 (water-soluble protein content). D5151, D412, D3767 and D6124 are also listed. IAS lists ASTM D3578, D5151 and D6124 as whole standards. No ASTM standard is on the CMA schedule.
- Two boundaries: protein is reported under D3578 clause 8.7; D5712, which D3578 references, is not listed separately on any certificate. Nor is D573, so air-oven aging is not a stand-alone accredited test. Confirm before ordering whether tensile after aging is reported under "physical and dimensional properties" (clauses 8.4 and 8.5) and who does the aging.
The standard and its scope
ASTM D3578-19(2023)
- Title: Standard Specification for Rubber Examination Gloves. Developed by ASTM Subcommittee D11.40, published in Book of Standards Volume 09.02, 5 pages. The 2019 edition was reapproved in 2023; the ASTM website marks it Active, last updated 2023-09-27.
- Scope 1.1: natural rubber gloves used in conducting medical examinations and diagnostic and therapeutic procedures, and natural rubber gloves used in handling contaminated medical material.
- Scope 1.2: gloves that fit either hand, paired gloves and gloves by size; packaged sterile gloves and packaged or bulk nonsterile gloves.
- The Type I/Type II distinction and the numerical requirements for dimensions, tensile, powder and protein are not on the ASTM page; refer to the specification text.
FDA recognition and product code
- The FDA Recognized Consensus Standards database lists ASTM D3578-19(2023) as recognition number 6-440, with partial recognition: all text related to powdered surgeon's gloves, powdered patient examination gloves and absorbable powder for lubricating a surgeon's glove is not recognized, because FDA has banned these devices. The final rule was published in the Federal Register on 2016-12-19 (81 FR 91722) and took effect on 2017-01-18.
- Linked product codes include LYY (latex patient examination glove, 21 CFR 880.6250, Class I, 510(k)), defined as a disposable natural rubber latex glove that may bear a trace amount of residual powder. 21 CFR 880.6250 is now titled "Non-powdered patient examination glove".
- ASTM D5712, the reference method for water-extractable protein, is also on the FDA list: recognition number 6-356, recognized edition D5712-15(2020), complete standard.
- The 510(k) format recommended in the FDA Medical Glove Guidance Manual (issued 2008-01-22) asks whether latex glove data meet D3578 or an equivalent standard, with a separate section on protein level. Its cover note says it has not been updated for the powdered glove ban.
Main methods referenced by D3578
| Method | Title | What it measures | Current edition on the ASTM website |
|---|---|---|---|
| ASTM D5151 | Test Method for Detection of Holes in Medical Gloves | Holes that leak water | D5151-19(2023) |
| ASTM D3767 | Practice for Rubber—Measurement of Dimensions | Dimensions and thickness of rubber products | D3767-03(2020) |
| ASTM D412 | Test Methods for Vulcanized Rubber and Thermoplastic Elastomers—Tension | Tensile properties | D412-16(2021) |
| ASTM D573 | Test Method for Rubber—Deterioration in an Air Oven | Air-oven aging | D573-04(2025) |
| ASTM D6124 | Test Method for Residual Powder on Medical Gloves | Residual powder on gloves | D6124-06(2022) |
| ASTM D5712 | Test Method for Analysis of Aqueous Extractable Protein … Using the Modified Lowry Method | Aqueous extractable protein (modified Lowry) | D5712-15(2025) |
The table lists the referenced methods relevant to this article, taken from the CDC NIOSH PPE-Info record for D3578-19(2023); that record also lists ISO 2859 for sampling. The edition of each method applied follows the specification text and the client agreement.
D3578 test items and sample quantities
| Item group | Method | Basis on SUNGO's CNAS schedule | Sample quantity depends on |
|---|---|---|---|
| Freedom from holes (leaks) | D5151 | D3578-19(2023) 8.3; also D5151-19(2023) | Lot size; ISO 2859 sampling, with inspection level and AQL per the specification text |
| Dimensions, thickness, tensile before and after aging | D3767, D412; aging per D573 | Schedule item "physical and dimensional properties": D3578-19(2023) 8.4, 8.5; also D412-16(2021), D3767-03(2020). Whether aged tensile is reported under these clauses: confirm before ordering | Number of sizes; one set of specimens before aging and one after |
| Residual powder | D6124 | D3578-19(2023) 8.6, 8.8; also D6124-06(2022) | Number of gloves set by the method |
| Water-soluble protein content | D5712 (referenced by D3578) | D3578-19(2023) 8.7 | Specimen area; whether aged gloves are to be tested |
We do not quote a fixed sample count because it depends on the test combination. Leak-test sample size depends on lot size (see Leaks below for FDA's own sampling). Physical properties are sampled per size, before and after aging. Powder and protein specimens follow the methods, with a margin for retesting. The laboratory gives quantities item by item when quoting.
Water-extractable protein: method, sample handling and high results
The FDA manual notes that latex proteins can cause Type I reactions in some individuals, and that since May 1991 FDA has recommended reducing water-extractable protein on latex devices. It treats these proteins as concomitant constituents of manufacturing materials: not bound into the film and adding nothing to the barrier, they leach out readily, so manufacturers should establish and validate processes to remove or limit them.
How D5712 works
- It measures total aqueous extractable protein: proteins are extracted in a buffer solution, then precipitated to concentrate them and separate them from water-soluble substances that may interfere, then redissolved and quantified colorimetrically by the modified Lowry method against a protein standard.
- Results are recorded in micrograms per dm² of test specimen.
- Other methods for leachable protein may be used for routine quality control once validated and correlated against D5712 as the reference method.
- The scope notes that the method has not been validated for lubricated products such as condoms.
- Extraction buffer, extraction time and temperature, specimen area and the standard curve are set out in the D5712 text.
Handling the samples
- Which glove condition to test. The FDA manual suggests testing recently manufactured finished gloves after accelerated aging per D3578 or real-time aging, because water-soluble proteins may migrate to the surface during shipping and aging should approximate that period. The manual's recommended 510(k) format also asks whether protein was tested on such aged finished gloves.
- State the condition. Say whether samples are unaged, accelerated-aged or real-time aged, and who aged them under what conditions; compare results only between samples in the same condition.
- Keep the original packaging. The FDA manual recommends making low-protein gloves in a separate area, and not in tanks or dryers used for regular latex gloves unless these are thoroughly cleaned, to avoid protein cross-contamination. By the same logic, send samples sealed in original packaging, not packed with other latex products.
- Declare surface treatments. State whether the gloves are chlorinated and whether they carry a polymer coating or lubricant, so the laboratory can judge whether the method applies.
- Document the lot. The manual's recommended 510(k) format suggests a summary of results from at least one lot of gloves to support the stated protein level. Name the test method with its year; for a non-standard method, attach data correlating it to a recognized method.
Common causes of high results
These points come from the FDA manual's production-control recommendations and are a reasonable starting point:
- Not enough leaching and washing. FDA recommends removing water-soluble proteins by reducing protein in the raw latex, controlling pre-cure leaching and post-cure washing, and leaching and washing for an appropriate time.
- Tank water not refreshed. Leaching tanks and spray or washing tanks should use flow-controlled, continually refreshed water to avoid protein saturation.
- Washing only before curing. During heat curing, proteins become more water-soluble and/or move to the glove surface, so washing should be done both before and after curing, and completed before any final donning powder or lubricant is applied.
- Chlorination and neutralization out of control. Treating the cured glove surface with chlorine or similar agents denatures surface proteins and washes them away, but chlorine itself can degrade the glove and has to be removed by washing, neutralization or other means. The chlorination process needs process control and validation.
- Cross-contamination. Low-protein gloves processed in tanks (especially starch slurry tanks) or dryers previously used for regular latex gloves, without thorough cleaning, can pick up protein.
- Different sample condition or method. Shipped or aged gloves may carry more surface protein than fresh ones, and in-house routine data not correlated with D5712 should not be compared directly with reference-method results.
The manual adds that FDA usually does not require a new 510(k) when a manufacturer only does more of an existing process, such as extra leaching or washing, and does not mention it on the labeling. FDA's current requirements govern.
Protein claims on the label
According to the FDA manual, manufacturers who can reliably reduce latex protein to a known level may make a labeling claim with supporting data in the 510(k). FDA does not allow a protein claim below the 50 µg/dm² sensitivity limit of the D5712 Lowry method cited in the manual, and gloves labeled as containing 50 µg/dm² or less of extractable protein should also carry a caution that safe use by latex-sensitized individuals has not been established. Unquantified terms such as "low protein" are listed as inappropriate claims.
Tensile properties before and after aging
The D3578 abstract lists tensile strength, ultimate elongation and stress at elongation. The 510(k) format recommended in the FDA manual lists, under Device Description and Specifications, minimum tensile strength (MPa) and ultimate elongation (%) before and after aging, plus length, width, palm and finger thickness. Points to watch:
- D412 has two methods: Method A uses dumbbell and straight section specimens, Method B uses cut ring specimens, and the standard notes that they do not give identical results. Which applies to gloves is set by D3578.
- Tensile properties depend on the material and on the test conditions (extension rate, temperature, humidity, specimen geometry, pretest conditioning, etc.), so results are comparable only under the same conditions.
- Tensile strength is calculated from specimen thickness. D3767 notes that instrument pressure significantly affects thickness readings on soft materials, so the presser foot pressure must be specified.
- D573 assesses the effect of elevated temperature on vulcanized rubber; its scope notes results may not correlate exactly with service performance but can be used for laboratory comparison. Aging temperature and time are set by D3578.
SUNGO's CNAS schedule lists D3578-19(2023) clauses 8.4 and 8.5 under "physical and dimensional properties", with D412-16(2021) and D3767-03(2020). Confirm with the laboratory before ordering whether tensile after aging is reported under these clauses and who does the aging. ASTM D573 is not listed separately. Our separate D6319 nitrile glove article covers tensile and thickness measurement in more detail.
Leaks and residual powder
Leaks. D5151 detects holes that leak water under the conditions of the test; the smallest hole that will allow leakage has not been determined and is outside the scope of the method. For its own regulatory sampling under 21 CFR 800.20, FDA uses a 1,000 mL water leak test; ISO 2859 sampling, single normal for lots of 1,200 gloves or fewer and multiple normal for larger lots, at general inspection level II; and AQL 2.5 as the minimum acceptable quality level for patient examination gloves. D5151 practice and AQL acceptance are covered in a separate article.
Residual powder. The D3578 abstract lists powder-free residue and powder amount. The method referenced is D6124, which reports the average powder mass per glove in milligrams. For a "powder-free" claim, the FDA manual recommends no more than 2 mg of residual or trace powder and debris per glove, as determined by D6124 or an equivalent method. Method details for residual powder are covered in a separate article.
Scope table
The table reproduces every row of the CNAS (L22083) schedule of Shanghai SUNGO Medical Technology Co., Ltd. under the test object "rubber examination gloves" (all effective 2026-03-09), plus the relevant rows of the IAS (TL-1019) scope (effective 2026-02-19). The limitation column is blank for all of these CNAS rows. Item names are translated from the Chinese schedule.
| Item | Basis (standard and clause) | Notes / limitations |
|---|---|---|
| Water impermeability test | ASTM D3578-19(2023) 8.3 | CNAS |
| Water impermeability test | ASTM D5151-19(2023) (no clause listed) | CNAS |
| Physical and dimensional properties | ASTM D3578-19(2023) 8.4, 8.5 | CNAS |
| Physical and dimensional properties | ASTM D412-16(2021) (no clause listed) | CNAS; not on the IAS scope |
| Physical and dimensional properties | ASTM D3767-03(2020) (no clause listed) | CNAS; not on the IAS scope |
| Residual powder | ASTM D3578-19(2023) 8.6, 8.8 | CNAS |
| Residual powder | ASTM D6124-06(2022) (no clause listed) | CNAS |
| Water-soluble protein content | ASTM D3578-19(2023) 8.7 | CNAS; reported under this clause; the protein method it references is not listed separately |
| Whole standard | ASTM D3578 | IAS; no edition year |
| Whole standard | ASTM D5151 | IAS; no edition year |
| Whole standard | ASTM D6124 | IAS; no edition year |
The Shanghai CMA (240015340087) schedule contains no ASTM standards. ASTM D5712 and D573 do not appear separately on any of the three Shanghai certificates.
Where SUNGO tests it and under which accreditation
The items in the scope table above are tested at the laboratory of Shanghai SUNGO Medical Technology Co., Ltd., 958 Kangqiao East Road, Pudong, Shanghai. Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity, and its accreditation and CMA scopes do not include medical gloves.
- CNAS accreditation (L22083): covers the rows marked CNAS above. None of the D3578 rows on the schedule is a sterility test.
- IAS accreditation (TL-1019): lists ASTM D3578, D5151 and D6124 as whole standards without an edition year. Please confirm the edition with the laboratory before an IAS report is issued against it, and confirm before ordering whether a given item can be reported under IAS. D412 and D3767 are not listed separately on the IAS scope.
- CMA (240015340087): the schedule contains no ASTM standards, so none of these items can be reported with the CMA mark. CMA covers only the items on its schedule, and the certificate states that it is valid within the People's Republic of China.
The CNAS certificate states that CNAS is a signatory to the ILAC and APAC mutual recognition arrangements, and the IAS certificate carries the ILAC-MRA mark. Whether FDA or a buyer accepts a report depends on their own requirements; we make no commitment on any regulator's decision. The FDA manual also recommends biocompatibility data in a 510(k) summary, which is outside this article. Where biological tests are needed, SUNGO subcontracts them to an appropriately qualified laboratory; under accreditation rules, the client's consent is obtained in advance, and the report identifies the subcontracted items and the subcontractor.
Preparing a submission
- State the basis. Standard number and edition (for example ASTM D3578-19(2023)), the clauses required, whether the gloves are Type I or Type II, and whether a conformity statement is needed.
- Describe the product. Natural rubber latex; powdered or powder-free; sterile or nonsterile; chlorinated or not; any polymer coating or lubricant; size range.
- Sample condition. One lot, original sealed packaging, with lot number and manufacturing date. Label unaged and aged samples as separate groups and state who aged them, to what basis and under what conditions.
- Lot information. Leak-test sample size under ISO 2859 depends on lot size, so please give it.
- Protein testing. State which sample condition is to be tested and whether results should be compared with data from your in-house routine method.
- Report mark. Say before ordering whether you need the CNAS or the IAS mark. ASTM items cannot be reported under CMA.
FAQ
Can nitrile examination gloves be tested to D3578?
D3578 covers natural rubber gloves. Nitrile examination gloves fall under ASTM D6319, which we cover in a separate article.
Is SUNGO's water-soluble protein test accredited to D5712?
Not as a separate listing. The Shanghai CNAS schedule lists ASTM D3578-19(2023) clause 8.7, "water-soluble protein content", and reports are issued under that clause; D5712 is not listed separately on the certificates.
Is air-oven aging (ASTM D573) a separate accredited test?
No. D573 is not listed separately on SUNGO's certificates. Confirm with the laboratory before ordering whether tensile after aging is reported under the CNAS item "physical and dimensional properties" (D3578 clauses 8.4 and 8.5) and who does the aging.
Can the label say "low protein"?
The FDA manual lists unquantified "low protein" wording as an inappropriate claim. Any protein claim needs supporting data in the 510(k), subject to FDA's current requirements.
Request a quote
For D3578 testing of latex examination gloves, see our mask, glove, dressing and gauze testing services, then use the contact page to send product details, required clauses, sizes, lot size, sample aging condition and the report mark you need. We will reply item by item against the CNAS and IAS scopes, with sample quantities. For the overall US route for gloves, see Exporting Medical Gloves to the US: FDA Route and Test Items; for the EU, see Medical Gloves for the EU: What EN 455 Actually Requires.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.