IEC 60601-1 Locked Rotor Test and Motor Overload Test (Clause 13.2)

IEC 60601-1 Locked Rotor Test and Motor Overload Test (Clause 13.2)

The IEC 60601-1 locked rotor test and motor overload test both sit in clause 13.2, single fault conditions: locking of moving parts is 13.2.8, motor capacitor open and short circuit is 13.2.9, run times and winding temperature for both are in 13.2.10 with Table 26, and running overload is 13.2.13.3 with Table 27 (Edition 3.1 numbering; check the edition you test to). A device passes when winding temperatures stay within the table limits, capacitor voltage stays within its rating and no hazardous situation listed in clause 13.1 occurs. Using breast pumps, air compression therapy devices and oxygen concentrator compressors as examples, this article covers when the tests apply, how they are run and judged, and what to prepare.

Short answer

  • The locked rotor test applies when any of the three conditions in clause 13.2.8 is met.
  • A motor with a capacitor in its auxiliary winding circuit is also run locked with the capacitor short circuited, then open circuited; at most the short-circuit test can be omitted.
  • Both tests start from cold, run for a period set by equipment type, and winding temperatures must not exceed Table 26.
  • Remotely or automatically controlled motors (single control device, no redundant protection), or motors likely to run continuously unattended, get a running overload test against Table 27.
  • After the overload tests the device must still be safe once cooled. No limit values or run times are given here; the standard governs.

IEC 60601-1 and GB 9706.1-2020: edition and scope

IEC 60601-1 is the general standard for the basic safety and essential performance of medical electrical equipment. As of 13 November 2026, the IEC webstore lists the consolidated Edition 3.2, IEC 60601-1:2005+AMD1:2012+AMD2:2020 (publication date 20 August 2020), as the current publication, with Edition 4 still under development. The webstore entry "IEC 60601-1:2026 SER" is a bundle of the general and collateral standards that still contains Edition 3.2; it is not a new edition.

The Chinese counterpart, GB 9706.1-2020, has been in force since 1 May 2023. It replaces GB 9706.1-2007 and GB 9706.15-2008 in full and is a modified adoption of IEC 60601-1:2012. The clause numbers in this article follow the Edition 3.1 structure (IEC 60601-1:2005+AMD1:2012). In the Edition 3.2 table of contents, clause 13.2 is still titled "Single fault conditions", and Table 26 (temperature limits of motor windings) and Table 27 (maximum motor winding steady-state temperature) are still there. The Edition 3.2 sub-clause text was not compared word for word here, so rely on the edition you test to.

The standard's own abstract notes that its requirements are supplemented or modified by collateral and particular standards. Breast pumps and air compression therapy devices intended for the home healthcare environment also fall under IEC 60601-1-11 (YY 9706.111-2021 in China, in force since 1 May 2023), covered in Home Healthcare Medical Devices: What IEC 60601-1-11 Adds to IEC 60601-1. Oxygen concentrators have their own particular standard, ISO 80601-2-69. This article covers only the motor-related single fault tests in the general standard.

Motor-related sub-clauses in clause 13.2

Sub-clause (Edition 3.1 numbering)ContentBasis for judgement
13.2.7Impairment of cooling, for example locking ventilation fans one at a time, blocking vents, simulating clogged filtersTemperature and other requirements of 13.1.2
13.2.8Locking of moving parts (locked rotor)Run per 13.2.10; winding temperature against Table 26
13.2.9Interruption and short circuiting of motor capacitorsCapacitor voltage within rating; winding temperature per 13.2.10
13.2.10Additional test criteria for motor-operated equipment (run times, when to measure)Table 26
13.2.13.1General overload test conditionsEquipment remains safe after cooling
13.2.13.3Equipment with motors: running overload protection, three-phase motors with one phase disconnectedTable 27; run times for the phase-loss test per 13.2.10
13.2.13.4Equipment rated for non-continuous operationWinding temperature against the values in 13.2.10

When these single faults are applied, clause 13.2.1 requires the normal conditions identified in 8.1 a) to be applied at the same time in the least favourable combination.

Locked rotor test: when clause 13.2.8 applies

Clause 13.2.8 requires the equipment to remain single fault safe if moving parts become jammed. Moving parts are locked if the equipment meets any of the following:

  • it has accessible moving parts, including applied parts, that are liable to be jammed;
  • it is liable to be operated while unattended, including equipment that is automatically or remotely controlled;
  • it has one or more motors with a locked rotor torque smaller than the full load torque.

If the equipment has more than one such moving part, only one is locked at a time. If a single fault can lock several motors, all of them are locked simultaneously. The run conditions are in clause 13.2.10.

In practice the second condition is the one most often missed: a product that can run unattended, or is started by automatic or remote control, needs the test even if users cannot reach the motor. The third condition needs the torque data from the motor datasheet; without it, do not simply rule the test out.

Cooling fans are a separate case. Locking ventilation fans one at a time is one of the simulations listed under clause 13.2.7, impairment of cooling, judged against the temperature requirements of 13.1.2. For a fan-cooled oxygen concentrator, a locked compressor (13.2.8) and a stalled fan (13.2.7) are two different faults.

Motor capacitor open and short circuit: clause 13.2.9

A motor with a capacitor in the circuit of an auxiliary winding is run with the rotor locked, with the capacitor short circuited and then open circuited in turn, under the conditions of clause 13.2.10. The capacitor voltage is measured with one side disconnected (open circuit). A measured voltage above the capacitor's rated voltage is a failure.

The exemption covers the short-circuit test only. It is not performed if the motor's capacitor complies with IEC 60252-1 and the equipment is not intended for unattended use (including automatic or remote control); the open-circuit test still applies. Both conditions must hold, so provide evidence of IEC 60252-1 compliance and instructions for use that state clearly how the device is used. A motor with no capacitor in its auxiliary winding circuit is outside this sub-clause.

Run times and winding temperature: clause 13.2.10 and Table 26

Clause 13.2.10 applies to every test under clauses 13.2.8 and 13.2.9, taking into account the exemptions in 13.1.2. Motor-operated equipment is started from cold and run at rated voltage or at the upper limit of the rated voltage range for one of four periods:

  1. hand-held equipment, equipment that has to be kept switched on by hand, and equipment that has to be kept under physical load by hand run for a short specified period;
  2. other equipment not intended for unattended use runs for another specified period;
  3. equipment not covered above whose operation is ended by a timer runs for the maximum period of the timer;
  4. all other equipment runs until thermal stability is reached.

Winding temperatures are determined at the end of the specified period, or at the instant a fuse, thermal cut-out, motor protective device or similar operates. Measurement follows 11.1.3 d), using the resistance method or thermocouples and other methods. A temperature above Table 26 is a failure. Table 26 gives limits by insulation class (A, B, E, F, H) and equipment type: equipment not intended for unattended use, with timers, or run only for the specified short periods forms one group; other equipment has separate limits for impedance protection and for protective devices operating within or after the first hour. The values are in the standard.

Small motors have an alternative, set out in clause 13.2.13.3 a). Where a motor sits in a circuit whose voltage does not exceed the limits given in the standard, and its small size or design makes accurate temperature measurement impractical, compliance with 13.2.9 and 13.2.10 can be determined without measuring temperature: the motor is covered with a single layer of bleached cotton cloth of the specified type, and ignition of the cloth during or at the end of the test is a failure. The voltage limits and cloth specification are in the standard.

Motor overload test: clause 13.2.13.3 and Table 27

Under clause 13.2.13.3 a), the motor part of equipment is checked, where applicable, by the tests of 13.2.8 to 13.2.10, 13.2.13.3 b) and c), and 13.2.13.4. Equipment that also has heating parts is tested with motor and heating parts operating together in the least favourable condition, and multiple applicable tests are run one after another.

The running overload protection test in clause 13.2.13.3 b) applies to motors that are either intended to be remotely or automatically controlled by a single control device without redundant protection, or likely to be subjected to continuous operation while unattended. The equipment runs under normal load at rated voltage, or the maximum of the rated voltage range, until thermal stability. With the supply voltage held, the load is then increased in steps, reaching thermal stability after each, until the overload protection operates or no further temperature rise is noted. The motor winding temperature is determined during each steady period, and a maximum above Table 27 is a failure. Table 27 gives maximum steady-state temperatures by insulation class only; the values are in the standard.

Two practical details: if the load cannot be changed in steps within the equipment, the motor is removed for the test; and for a motor in a circuit whose voltage does not exceed the limits given in the standard, the running overload test is performed only if inspection or design review shows that an overload could occur; it need not be performed, for example, where electronic drive circuits maintain a substantially constant drive current. This matters for small low-voltage DC pumps; document the reasoning for the laboratory to review.

In addition, equipment with three-phase motors is run with normal load on a three-phase supply with one phase disconnected (13.2.13.3 c)), for the periods in 13.2.10. Under clause 13.2.13.4, equipment rated for non-continuous operation (except hand-held equipment, equipment kept on or loaded by hand, and equipment with a timer and back-up timer) runs under normal load until temperatures stabilise or a protective device operates, with winding temperatures within the values in 13.2.10.

Pass/fail criteria: temperature, capacitor voltage, hazardous situations and safety after cooling

Taken together, the judgement for motor fault tests falls into four areas:

  • Winding temperature: Table 26 for locked rotor and capacitor faults, Table 27 for running overload.
  • Capacitor voltage: during the 13.2.9 test it must not exceed the capacitor's rated voltage.
  • Hazardous situations in clause 13.1: for example, emission of flames, molten metal or poisonous or ignitable substances in hazardous quantities from the enclosure; enclosure deformation impairing compliance with 15.3.1; excessive temperatures of applied parts and parts likely to be touched; leakage current or accessible-part voltage beyond single fault limits; and the mechanical hazards of 9.1 to 9.8. With the cotton cloth method, ignition of the cloth is a failure.
  • Safety after cooling: under clause 13.2.13.1, equipment must still be safe after the overload tests once cooled to near laboratory temperature, checked by inspection or an appropriate test such as a dielectric strength test of motor insulation under 8.8.3. Thermoplastic insulation used as a means of protection also gets a ball pressure test at a raised temperature set in the standard.

Keep clause 11.1 and clause 13.2 apart. Clause 11.1 assesses temperatures in normal use, with motor-operated equipment at normal load and duty cycle and the least favourable voltage in the specified range. Clause 13.2 assumes a fault has occurred and asks whether the equipment is still safe. Both use the measurement methods of 11.1.3, but conditions and limit tables differ, so passing the normal temperature test cannot replace the fault tests. Clause 11.1 is covered in a separate article.

Applying this to breast pumps, air compression therapy devices and oxygen concentrator compressors

Here the motor drives a suction pump, an inflation pump or a compressor, usually through a control circuit, and some products have timers, automatic operation or long continuous running. Work through applicability in this order:

  1. Use: unattended operation, automatic or remote control, timer, and continuous or non-continuous rating. These decide whether 13.2.8, 13.2.13.3 b) and 13.2.13.4 apply and which run period 13.2.10 uses.
  2. Motor: AC or DC, whether the circuit voltage is within the low-voltage limits, capacitor or not, insulation class, locked rotor and full load torque.
  3. Protection: how thermal protectors, fuses and electronic current limiting operate, which decides when winding temperature is measured and whether the low-voltage exemption applies.
  4. Cooling: consider stalled fans under 13.2.7, separately from a locked pump or compressor.

The particular standard for oxygen concentrators, ISO 80601-2-69, can modify clauses of the general standard, so for these devices the two should be read together, with the particular standard governing. For background on oxygen concentrator testing, see Oxygen Concentrator Testing: ISO 80601-2-69, IEC 60601-1 and IEC 60601-1-2. Particular standards for other products, such as infusion pumps or medical electric beds (the IEC 60601-2-x series), are outside Sungo's accreditation scope and are not covered here.

Accreditation scope table

ItemStandardNotes / limits
Whole standardIEC 60601-1Hefei, IAS TL-1119; no clauses itemised, no limitation stated, no year stated
Whole standardGB 9706.1Hefei, IAS TL-1119; no clauses itemised, no limitation stated, no year stated
Whole standardIEC 60601-1-11Hefei, IAS TL-1119; no clauses itemised, no limitation stated, no year stated
Whole standardYY 9706.111Hefei, IAS TL-1119; no clauses itemised, no limitation stated, no year stated
Whole standardISO 80601-2-69Hefei, IAS TL-1119; no clauses itemised, no limitation stated, no year stated

Where the testing is done and under which accreditation

Medical electrical safety testing is carried out by Hefei SUNGO Medical Technology Co., Ltd. under IAS accreditation, certificate TL-1119. The scope covers the whole standards IEC 60601-1, GB 9706.1, IEC 60601-1-11, YY 9706.111 and ISO 80601-2-69, and the motor fault tests in clause 13.2 are part of the general standard. IAS is a signatory to the ILAC MRA.

Three points apply: medical electrical safety is only within the Hefei IAS scope, not the CNAS or CMA certificates, so these reports cannot be presented as CNAS or CMA reports, and the Shanghai laboratory does not offer this testing; the IAS scope states no year, so confirm the edition with the laboratory before testing; among particular standards, only ISO 80601-2-69 is in scope. Acceptance of a report by a regulator depends on that regulator's requirements. For the service overview, see the medical electrical safety service page.

Preparing samples and documents

  • Motor datasheet: rated voltage and frequency, rated power, insulation class, locked rotor and full load torque (where available), and the motor circuit voltage.
  • Capacitor datasheet and evidence of IEC 60252-1 compliance (where relevant), to assess the short-circuit exemption.
  • Model, rating, certification and operating mode (automatic or non-automatic reset) of thermal protectors, fuses and motor protective devices.
  • Circuit diagram and motor control description: remote or automatic control, single control device, redundant protection, and whether the electronic drive holds a constant current. If you claim a low-voltage motor does not need the running overload test, include the design review behind it.
  • Moving parts and the consequences if each jams, cross-referenced to the risk management file; location and role of cooling fans.
  • Instructions for use and labelling on unattended use, continuous operation, timers and rated mode of operation.
  • Sample planning: fault tests can damage samples; if the load cannot be stepped within the device, a separate motor or loading fixture is needed; the resistance method needs access to the winding leads. Agree sample number and condition with the laboratory in advance.

Frequently asked questions

Is the IEC 60601-1 locked rotor test always required?

No. Under clause 13.2.8 it is required when any one of three conditions is met: accessible moving parts (including applied parts) liable to be jammed, equipment liable to be operated while unattended (including automatic or remote control), or a motor whose locked rotor torque is smaller than its full load torque. If none applies, record the reasoning in the technical documentation for the laboratory to review.

Can the motor capacitor short-circuit test be skipped?

Only the short-circuit test can be omitted, and only when both conditions hold: the capacitor complies with IEC 60252-1, and the equipment is not intended for unattended use (including automatic or remote control). The open-circuit test is still required.

Do low-voltage DC motors need the overload test?

Only if the equipment falls under clause 13.2.13.3 b). Even then, for a motor in a circuit within the voltage limits given in the standard, the test is performed only if inspection or design review shows an overload could occur, and not where the electronic drive maintains a substantially constant current.

Which IEC 60601-1 table gives the motor winding temperature limits?

Locked rotor and capacitor fault tests use Table 26, temperature limits of motor windings; the running overload test uses Table 27, maximum motor winding steady-state temperature. Both are organised by insulation class, and the Edition 3.2 table of contents keeps both table titles. Take the values from the edition you are testing to.

Are the motor tests in GB 9706.1-2020 the same as in IEC 60601-1?

GB 9706.1-2020 is a modified adoption of IEC 60601-1:2012, so the framework described here is a useful guide. A modified adoption can contain differences from the international standard, so when testing to GB 9706.1-2020 the Chinese text governs.

If you have a motor-driven product and want to confirm which clause 13.2 tests apply and how to prepare, request a quotation through the contact page. Include the product type, target standard and edition, motor and capacitor specifications, control method and use scenario, so the laboratory can confirm the applicable clauses and propose a test plan.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.