Hydrogel Dressing Testing: YY/T 0471.1 Fluid Affinity, Dispersibility and pH

Hydrogel Dressing Testing: YY/T 0471.1 Fluid Affinity, Dispersibility and pH

Hydrogel dressing testing in China commonly relies on YY/T 0471.1-2004, Test methods for primary wound dressing, Part 1: Aspects of absorbency. Amorphous hydrogels are assessed with clause 3.4 (fluid affinity) and clause 3.7 (dispersibility/solubility), while fibrous dressings that gel on contact with exudate, such as alginates, are assessed with clause 3.5 (gelling) and clause 3.6 (dispersion characteristics). pH is usually written into the product technical requirements and measured with a pH meter under General Chapter 0631 of the Pharmacopoeia of the People's Republic of China (ChP), 2025 Edition, Volume IV. Below we cover each method, how results are read, the EN 13726:2023 Annex F/G mapping and the accreditation marks available from SUNGO's Shanghai laboratory.

Key Points

  • As of November 13, 2026, YY/T 0471.1-2004 is a current, recommended pharmaceutical industry standard that identically adopts EN 13726-1:2002. It gives test methods only, with no performance requirements; acceptance ranges have to be set in the product technical requirements.
  • Clauses 3.4 and 3.7 apply to amorphous hydrogels; clauses 3.5 and 3.6 are aimed at fibrous dressings, with alginates given as the example in the notes, not at gel paste.
  • Fluid affinity looks in two directions: fluid absorbed from an agar substrate and fluid donated to a gelatin substrate. Results are expressed as percentage change in gel mass.
  • EN 13726:2023 supersedes EN 13726-1:2002 to EN 13726-4:2003 and merges them into one document. The relevant annexes are Annex F, Fluid donation of amorphous hydrogel dressings, and Annex G, Dispersion characteristics of gelling dressings; EN 13726 is not on SUNGO's CMA scope.
  • ChP General Chapter 0631 measures the pH of a solution; whether a device sample is measured directly, dispersed or extracted must be specified in the product technical requirements.
  • All of these items are on the scope of Shanghai SUNGO Medical Technology Co., Ltd.: YY/T 0471.1-2004 clauses 3.2 to 3.7 under CNAS and CMA; EN 13726:2023 Annexes F and G under CNAS; YY/T 0471.1 and EN 13726 on the IAS scope as whole standards without a year; and General Chapter 0631 under CNAS and CMA.

Which Standard Applies to Hydrogel Dressings: Scope of YY/T 0471.1

As of November 13, 2026, China's national standards information platform lists YY/T 0471.1-2004 as current. It is a recommended pharmaceutical industry standard, published on March 23, 2004 and implemented on January 1, 2005. Its scope: recommended test methods for evaluating the absorbency aspects of primary wound dressings. The foreword states that it identically adopts EN 13726-1:2002. Other parts of the series cover moisture vapour transmission, waterproofness, conformability and more; this article deals with Part 1 only.

Clause 2 defines an alginate dressing (contains inorganic salts of alginic acid and gels on contact with physiological fluid), an amorphous hydrogel (a semi-solid gel of hydrophilic polymers and water) and fluid affinity (the ability to absorb fluid from, or donate fluid to, a simulated wound).

Products sold as "hydrogel" or "gel dressings" come in very different forms, so identify the form before choosing tests:

Product formTypical appearanceDirectly relevant YY/T 0471.1 clauses
Amorphous hydrogelPaste or tube gel that is spread onto the wound3.4 fluid affinity; 3.7 dispersibility/solubility of hydrogel dressings
Fibrous dressing that gels on contact with fluidAlginate dressing in sheet or rope (packing) form3.5 gelling; 3.6 dispersion characteristics; 3.2 for absorbency by comparison
Sheet hydrogel or backed gel patchGel with a fixed shape3.4 and 3.7 name amorphous hydrogels; for sheet products, set the method in the product technical requirements based on construction and intended use

How Fluid Affinity Is Tested: YY/T 0471.1 Clause 3.4

Clause 3.4 covers the fluid affinity of amorphous hydrogel dressings. The test determines the ability of a hydrogel to absorb fluid from, or donate fluid to, gelatin or agar. In Table 1 of the standard, the agar column corresponds to absorption and the gelatin column to donation.

The main elements of the method in the original text are:

  • Container: syringes with the nozzle cut off and a low-profile plunger, held vertically during the test.
  • Test solution A: a sodium chloride and calcium chloride solution that the standard describes as having an ion content comparable to human serum or wound exudate. Both substrates are prepared with it.
  • For each test sample, five syringes are filled with agar and five with gelatin. After the substrate sets and is weighed, a specified mass of gel is spread on it, and the syringe is weighed, sealed and held upright in an incubator for a specified time.
  • The assembly is then weighed, the gel is removed with the substrate left intact, and the syringe and substrate are weighed again.
  • Percentage change in gel mass is calculated with Formula (1). If the overall mass difference before and after incubation exceeds the value in the original text, the test is repeated.

Substrate composition, temperature, time and weighing resolution should follow the original text. The standard requires the report to state any deviation from the method.

How are the results read? Table 1 of the standard is an example of how to present results. The agar column is graded 1 to 5 by percentage increase in gel mass, and the gelatin column is graded a to e by percentage loss in gel mass, giving a "number plus letter" type code. The note to the table gives examples: a dressing that absorbs a large amount of fluid from agar and donates little to gelatin could be classified as type 3a, and one that donates a lot and absorbs little could be classified as type 1c.

When drafting the product technical requirements, we suggest deciding first whether the gel is mainly meant to donate or take up moisture, then specifying the direction and range of mass change on each substrate. Two common mistakes: treating Table 1, which is explicitly an example, as an acceptance limit; and looking at only the agar or only the gelatin result.

Hydrogel Dispersibility and Solubility: YY/T 0471.1 Clause 3.7

Clause 3.7, dispersibility/solubility of hydrogel dressings, determines the physical behaviour of amorphous hydrogels when exudate volume is high. A specified amount of hydrogel is added to test solution A in a stoppered graduated cylinder, shaken at room temperature so that it can disperse or dissolve, left to stand for a specified time, and then examined visually.

If the sample dissolves, it is soluble; if two distinct phases appear, or it disperses unevenly and then separates into two layers, it is dispersible; if it keeps its structure, it is non-dispersible. The report gives a category, not a number.

Gelling and Dispersion Characteristics: Which Dressings Clauses 3.5 and 3.6 Cover

Clause 3.5 is often mistaken for a required hydrogel test. In fact it distinguishes dressings that gel quickly or slowly on contact with excess fluid and applies only to dressings that can be broken up as the procedure requires; the note gives fibrous alginate-type dressings as the example. Clause 3.6 identifies whether a fibrous wound dressing disperses when gently swirled in excess fluid; its note also uses alginates.

Clause 3.5 procedure: the fibrous sample is broken up through a sieve, a specified mass is shaken with test solution A in a conical flask and vacuum-filtered, and these steps are repeated, leaving a residue on the filter paper. A non-gelling control is prepared the same way with a calcium chloride standard solution. Comparing the two residues shows whether gelling occurred, which is the reported result.

Clause 3.6 procedure: a sample of specified size is placed in a wide-neck conical flask with test solution A, swirled without forming a vortex, and inspected visually. If the fibres separate and lose their original structure, the dressing is dispersible; if the original fibrous structure remains, it is non-dispersible. The standard notes that dressings which lose their integrity under these conditions are best removed from the wound by irrigation.

Absorbency Tests in Clauses 3.2 and 3.3: For Comparison

Clause 3.2, free swell absorptive capacity, applies to dressings in static physical contact that reach maximum absorption within 30 min under the test conditions; the standard notes that it suits most sheet or rope alginate dressings. Clause 3.3, fluid handling capacity, is the sum of fluid absorbed by the dressing and moisture vapour lost through it, and is used for water-resistant wound dressings applied for more than 24 h whose main roles are absorbing exudate and controlling the microbial environment.

Amorphous hydrogels are usually assessed with 3.4 and 3.7; products with both a gel layer and an absorbent layer can combine them in the product technical requirements.

EN 13726:2023 Annex F and Annex G vs YY/T 0471.1

YY/T 0471.1-2004 identically adopts EN 13726-1:2002. EN 13726:2023, Test methods for wound dressings - Aspects of absorption, moisture vapour transmission, waterproofness and extensibility, was prepared by CEN/TC 205. Its European foreword states that it supersedes EN 13726-1:2002, EN 13726-2:2002, EN 13726-3:2003 and EN 13726-4:2003 and merges the previously separate parts. Its scope covers absorption test methods for wound dressings, moisture vapour transmission rate of permeable film wound and fixation dressings, and waterproofness and extensibility. As of November 13, 2026, YY/T 0471.1 in China still corresponds to the 2002 Part 1.

Comparing the annex titles in the contents of EN 13726:2023 with the item names on SUNGO's CNAS scope, the two line up as follows:

YY/T 0471.1-2004EN 13726:2023 annex (title in the standard's contents)Note
3.4 Fluid affinity of amorphous hydrogel dressingsAnnex F Fluid donation of amorphous hydrogel dressingsSimilar subject; the 2023 title says fluid donation (SUNGO's CNAS scope uses a Chinese item name meaning fluid affinity); follow the original text for method details
3.6 Dispersion characteristicsAnnex G Dispersion characteristics of gelling dressingsSimilar subject
3.5 GellingNo annex titled for gelling in the contentsReport under YY/T 0471.1 when needed
3.7 Dispersibility/solubility of hydrogel dressingsNo annex titled for hydrogel dispersibility/solubility in the contentsReport under YY/T 0471.1 when needed
3.2 Free swell absorptive capacityAnnex B Free swell absorptive capacityFor comparison
3.3 Fluid handling capacityAnnex E Fluid handling capacityFor comparison

Do not treat the two standards as interchangeable editions of one method. EN 13726 was merged and revised, so whether the apparatus, conditions and result wording of Annexes F and G match the 2002 edition must be checked against the EN 13726:2023 text. If a registration or customer document names an edition, test to that edition.

One Sample, Two Reports: How the Accreditation Marks Differ

Typically YY/T 0471.1 is used for Chinese registration and EN 13726 for an EU technical file; the marks available differ:

Test basisAccreditation marks available (SUNGO Shanghai)Notes
YY/T 0471.1-2004 clauses 3.2 to 3.7CNAS, CMA, IASThe IAS scope lists YY/T 0471.1 as a whole standard without a year, so confirm the edition with the laboratory before an IAS report is issued; CMA covers only items in its schedule and is valid only within the People's Republic of China
EN 13726:2023 Annex F and Annex GCNAS, IASNo CMA; the IAS scope lists EN 13726 as a whole standard without a year, so confirm the edition with the laboratory before an IAS report is issued
ChP 2025 Edition, Volume IV, General Chapter 0631, pHCNAS, CMANot on the IAS scope

SUNGO is accredited by CNAS and IAS, both signatories to the ILAC MRA. Whether an overseas regulator or notified body accepts a given report depends on its requirements; confirm with them before ordering.

How pH Is Measured: ChP 2025 Edition, Volume IV, General Chapter 0631

The ChP 2025 Edition was issued by the National Medical Products Administration and the National Health Commission in Announcement No. 29 of 2025, dated March 20, 2025, and came into force on October 1, 2025. General Chapter 0631 in Volume IV is the pH determination method.

General Chapter 0631 is a potentiometric method. The pH of a solution is measured with a pH meter, and aqueous solutions are normally measured with a glass indicator electrode together with a reference electrode. Before measurement, the instrument is calibrated with standard buffer solutions: two or three suitable buffers are chosen as specified so that the pH of the test solution falls between them, and the test solution is measured once calibration meets the requirements. The chapter also lists precautions on electrode rinsing and buffer preparation and storage; follow the pharmacopoeia text for details.

General Chapter 0631 does not say how a device sample becomes the test solution. A hydrogel is semi-solid, so the product technical requirements need to state whether it is measured directly, dispersed in water at a set ratio, or extracted and the extract measured, including sample amount, medium, ratio, temperature, time and measurement temperature. Different sample preparation can give different readings.

The chapter also gives no pH range for device products; the acceptable range is set by the product technical requirements.

SUNGO Accreditation Scope: Hydrogel and Gelling Dressing Items

The table follows the item names and clauses on the Shanghai CNAS scope (L22083) and shows CMA (240015340087) and IAS (TL-1019) coverage. The certificates list no limitations for these items.

ItemStandard and clauseNotes / limitations
Fluid affinity of amorphous hydrogel dressingsYY/T 0471.1-2004 3.4CNAS, product category "dressings"; the CMA schedule lists the same item; IAS covers YY/T 0471.1 as a whole standard
Gelling characteristicsYY/T 0471.1-2004 3.5As above
Dispersion characteristicsYY/T 0471.1-2004 3.6As above
Dispersibility/solubility of hydrogel dressingsYY/T 0471.1-2004 3.7As above
Free swell absorptive capacityYY/T 0471.1-2004 3.2As above; the product category "adhesive dressings and plasters" also lists "fluid absorption" under 3.2
Fluid handling capacityYY/T 0471.1-2004 3.3As above
Fluid affinity of amorphous hydrogel dressings (CNAS item name, translated)EN 13726:2023 Annex FCNAS, product category "wound dressings"; no CMA; IAS covers EN 13726 as a whole standard (EN 13726-1 and EN 13726-4 are also listed), all without a year
Dispersion characteristics of gelling dressingsEN 13726:2023 Annex GAs above
pH valuePharmacopoeia of the People's Republic of China, 2025 Edition, Volume IV, General Chapter 0631CNAS and CMA, product category "medical devices"; not on the IAS scope

Where SUNGO Performs These Tests and Under Which Accreditation

All of the items above are performed by Shanghai SUNGO Medical Technology Co., Ltd. at its laboratory at No. 958 Kangqiao East Road, Pudong New Area, Shanghai, under CNAS L22083, CMA 240015340087 and IAS TL-1019. Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity, and these items are not on its scope.

For an overview of our dressing testing services, see the medical mask, glove and dressing testing service page. Biological tests required for registration (cytotoxicity, sensitization, irritation and so on) are outside SUNGO's accreditation scope. Where they are needed, they can be subcontracted to a laboratory with the appropriate qualifications, with the client's prior consent, and the report will identify the subcontracted items and the subcontractor.

Preparing Samples for Testing

  • State the product form (amorphous gel, sheet gel or gelling fibrous dressing) and describe its construction and intended use so the applicable clauses can be confirmed.
  • Provide the product technical requirements or a draft: list the clauses to be tested, how results are to be expressed and the acceptance ranges, and for pH, the sample preparation method.
  • When ordering, state the test basis and marks in one go: YY/T 0471.1-2004 or EN 13726:2023, and whether CNAS, CMA or IAS is needed.
  • Send samples from one batch in unopened original packaging, labelled with batch number and storage conditions. Clause 3.4 needs separate agar and gelatin syringe sets and clause 3.7 a separate portion, so confirm the total quantity with the laboratory in advance.
  • For gel patches and other products that also have adhesion requirements, ChP General Chapter 0952 adhesion testing can be considered at the same time (SUNGO's accreditation covers Method 1 and Method 3 only); SUNGO has a separate article on it.

FAQ

Which YY/T 0471.1 tests does a hydrogel dressing need for registration?

Amorphous hydrogels generally use clauses 3.4 and 3.7; fibrous alginate-type gelling dressings use clauses 3.5 and 3.6. The exact clauses depend on the product technical requirements.

Is there a pass/fail limit for fluid affinity?

No. YY/T 0471.1 gives test methods only, and Table 1 is just an example of presenting results. The manufacturer sets the acceptance range in the product technical requirements.

Can a YY/T 0471.1 report replace an EN 13726 report?

Not directly. EN 13726-1:2002, which YY/T 0471.1-2004 adopts, has been superseded by EN 13726:2023, and Annex F is now titled fluid donation; when the 2023 edition is required, test and report under Annexes F and G.

Is hydrogel pH measured directly or on an extract?

General Chapter 0631 covers only the measurement of a solution; sample preparation is specified in the product technical requirements, and testing follows them.

Should the pH of gel patches and cooling patches follow ChP 0631 or GB 15979?

It depends on how the product is regulated; take the category from the original NMPA Medical Device Classification Catalogue and related classification determinations. For medical devices, General Chapter 0631 can be cited in the product technical requirements. For disposable sanitary products under GB 15979-2024, pH is covered by clause 7.2.2 of that standard, which SUNGO holds under CNAS only; see GB 15979-2024 Disposable Sanitary Products: What Changed and What Is Tested.

To request a quote, leave a message through our contact page with the product form, the clauses you need (for example YY/T 0471.1 clauses 3.4 and 3.7, and ChP General Chapter 0631), the test basis and marks required, and the sample preparation for pH, so the test plan can be confirmed in one round.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.