At SUNGO, IEC 60601-1 (medical electrical safety) and IEC 60601-1-2 (medical EMC) testing is carried out by the laboratory of Hefei SUNGO Medical Technology Co., Ltd., accredited by IAS under certificate TL-1119. The scope lists whole standards without year designations. Besides these two, it covers IEC 60601-1-8, IEC 60601-1-11, ISO 80601-2-69 and a set of basic EMC standards; other particular standards, usability standards and software standards are not covered. This article lists what is in scope and what is not, so that regulatory, R&D and sourcing teams can decide before shipping a sample whether an accredited report can be issued here.
Short answer
- Where: medical electrical safety and medical EMC are tested only in Hefei, under IAS TL-1119. SUNGO Shanghai does not offer these two lines.
- Which report: these two lines are reported under IAS accreditation only. The CNAS and CMA certificates held by SUNGO Hefei do not cover them, so no CNAS or CMA report is issued for them; where a CMA report is required, SUNGO cannot issue one.
- Safety scope: the general standard IEC 60601-1, the collateral standards IEC 60601-1-8 (alarm systems) and IEC 60601-1-11 (home healthcare environment), and ISO 80601-2-69 (oxygen concentrators), the only particular standard on the schedule, each with the EN and GB/YY numbers listed on the schedule.
- EMC scope: IEC 60601-1-2 with its EN and YY 9706.102 numbers; the schedule also lists CISPR 11/GB 4824, ten basic standards from the IEC 61000-3 and IEC 61000-4 series (with the EN and GB numbers listed on the schedule), IEC TS 60601-4-2, IEEE/ANSI C63.18 and IEC 61326-2-6 for in vitro diagnostic equipment.
- Not in scope: all IEC 60601-2-x and ISO 80601-2-x particular standards other than ISO 80601-2-69, plus IEC 60601-1-6, IEC 62366-1, IEC 62304, IEC 61010-1, IEC 60335, IEC 62133, CISPR 14/25/32/35, FCC Part 15/18 and IEC 61326-1.
What IEC 60601-1 and IEC 60601-1-2 cover
IEC 60601-1 sets the general requirements for basic safety and essential performance of medical electrical equipment. The IEC abstract for the standard states that, for certain types of equipment, these requirements are supplemented or modified by a collateral or particular standard, and that where particular standards exist, IEC 60601-1 should not be used alone. This settles the common question of whether IEC 60601-1 testing alone is enough: if a particular standard applies to your product and that particular standard is not in the laboratory's accreditation scope, a report against the general standard alone does not cover the particular standard's requirements.
IEC 60601-1-2 is the collateral standard to IEC 60601-1 for electromagnetic disturbances. It covers basic safety and essential performance in the presence of electromagnetic disturbances (immunity) and the disturbances the equipment itself emits (emissions). The IEC abstract also notes that from the fourth edition (IEC 60601-1-2:2014), immunity test levels are specified according to the environment of intended use: the professional healthcare facility environment, the home healthcare environment and special environments. Stating the intended use environment clearly is therefore the starting point for an EMC test plan.
The two collateral standards also need a case-by-case decision. According to the IEC abstracts, IEC 60601-1-8 sets requirements, tests and guidance for alarm systems in medical electrical equipment and systems, so it comes into play for equipment with an alarm system. IEC 60601-1-11 applies to equipment and systems used in the home healthcare environment, whether the operator is a lay person or trained healthcare personnel. Whether either applies depends on the scope clause of the standard and the product's intended use, not on the product name.
On Chinese numbers: on the IAS schedule, GB 9706.1 carries the same title as IEC 60601-1, and the titles of YY 9706.102, YY 9706.108, YY 9706.111 and YY 9706.269 correspond to EMC, alarm systems, the home healthcare environment and oxygen concentrators. The edition and degree of adoption of each number are as stated in the standard text.
The schedule also lists ISO 7176-21 and GB/T 18029.21. These set EMC requirements and test methods for electrically powered wheelchairs, scooters and their battery chargers, and apply only to those products; EMC for general medical electrical equipment is still assessed against IEC 60601-1-2.
How to check whether a standard is in scope before you ship
Use these five steps. Limits and test conditions come from the standard text; this article does not list numerical values.
- Check for a particular standard. Decide whether an IEC 60601-2-x or ISO 80601-2-x particular standard applies to the product. The only particular standard on the Hefei IAS schedule is ISO 80601-2-69 (with EN ISO 80601-2-69 and YY 9706.269). Any other particular standard needs a laboratory whose scope lists that standard.
- List the collateral standards. EMC means IEC 60601-1-2; equipment with an alarm system means IEC 60601-1-8; equipment for the home healthcare environment means IEC 60601-1-11. Other collateral standards, such as IEC 60601-1-6 for usability, are not in scope.
- Match the numbering. The schedule lists IEC, EN and GB/YY numbers as separate entries. In the EN 61000 series, some parts are listed only as EN, some only as EN IEC and some under both; the table below and the schedule itself govern.
- Confirm the edition. The schedule does not state years. Before an IAS report is issued against a particular edition, confirm the edition with the laboratory and check which edition your target market requires.
- Confirm how the report will be used. Whether the receiving authority or buyer accepts an IAS-accredited report is for that party to decide. Where a CMA report is required, SUNGO cannot issue one for these two lines.
SUNGO Hefei IAS scope (TL-1119)
Medical electrical safety
| Item | Clause | Notes / limitations |
|---|---|---|
| IEC 60601-1 / EN 60601-1 / GB 9706.1 | Whole standard | Medical electrical equipment, Part 1: general requirements for basic safety and essential performance; no year stated on the schedule |
| IEC 60601-1-8 / EN 60601-1-8 / YY 9706.108 | Whole standard | Collateral standard: general requirements, tests and guidance for alarm systems in medical electrical equipment and systems |
| IEC 60601-1-11 / EN 60601-1-11 / YY 9706.111 | Whole standard | Collateral standard: requirements for medical electrical equipment and systems used in the home healthcare environment |
| ISO 80601-2-69 / EN ISO 80601-2-69 / YY 9706.269 | Whole standard | Particular standard: basic safety and essential performance of oxygen concentrator equipment; the only particular standard on the schedule |
Medical EMC
| Item | Clause | Notes / limitations |
|---|---|---|
| IEC 60601-1-2 / EN 60601-1-2 / YY 9706.102 | Whole standard | Collateral standard: electromagnetic disturbances, requirements and tests (YY 9706.102 is titled electromagnetic compatibility, requirements and tests on the schedule) |
| IEC TS 60601-4-2 | Whole standard | Guidance and interpretation: electromagnetic immunity performance of medical electrical equipment and systems |
| IEEE/ANSI C63.18 | Whole standard | Recommended practice for an on-site, ad hoc test method for estimating electromagnetic immunity of medical devices to radiated RF emissions from RF transmitters |
| CISPR 11 / GB 4824 | Whole standard | Industrial, scientific and medical equipment: radio-frequency disturbance characteristics, limits and methods of measurement |
| IEC 61000-3-2 / EN IEC 61000-3-2 / GB 17625.1 | Whole standard | Limits for harmonic current emissions |
| IEC 61000-3-3 / EN 61000-3-3 / GB/T 17625.2 | Whole standard | Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems |
| IEC 61000-4-2 / EN 61000-4-2 / EN IEC 61000-4-2 / GB/T 17626.2 | Whole standard | Electrostatic discharge immunity test |
| IEC 61000-4-3 / EN IEC 61000-4-3 / GB/T 17626.3 | Whole standard | Radiated, radio-frequency, electromagnetic field immunity test |
| IEC 61000-4-4 / EN 61000-4-4 / GB/T 17626.4 | Whole standard | Electrical fast transient/burst immunity test |
| IEC 61000-4-5 / EN 61000-4-5 / GB/T 17626.5 | Whole standard | Surge immunity test |
| IEC 61000-4-6 / EN 61000-4-6 / EN IEC 61000-4-6 / GB/T 17626.6 | Whole standard | Immunity to conducted disturbances induced by radio-frequency fields |
| IEC 61000-4-8 / EN 61000-4-8 / GB/T 17626.8 | Whole standard | Power frequency magnetic field immunity test |
| IEC 61000-4-11 / EN IEC 61000-4-11 / GB/T 17626.11 | Whole standard | Voltage dips, short interruptions and voltage variations immunity tests |
| IEC 61000-4-39 / EN 61000-4-39 / GB/T 17626.39 | Whole standard | Radiated fields in close proximity, immunity test |
| IEC 61326-2-6 / EN IEC 61326-2-6 / GB/T 18268.26 | Whole standard | EMC requirements for electrical equipment for measurement, control and laboratory use; particular requirements for in vitro diagnostic (IVD) medical equipment |
| ISO 7176-21 / GB/T 18029.21 | Whole standard | EMC requirements and test methods for electrically powered wheelchairs, scooters and battery chargers; for those products only |
Every row above is a whole standard on the schedule, with no clause limitations or exclusions listed. Numbers in the same row are grouped by the standard titles on the schedule; the edition correspondence between them is as stated on the certificate schedule and in the standard texts. The schedule shows an effective date of 4 June 2026. As of 13 November 2026, the IAS certificate search showed TL-1119 held by HEFEI SUNGO MEDICAL TECHNOLOGY CO., LTD., program Testing Laboratories, status Accredited. The live IAS search is the reference for current status.
Common standards that are not in SUNGO Hefei's scope
The table below lists standards that buyers and manufacturers often ask about but that are not in any SUNGO accreditation scope. The "Area" column follows the titles published by IEC or in the US Code of Federal Regulations; the original text governs actual applicability. If your product needs one of these, look for a body whose accreditation scope lists that standard.
| Standard or group | Area | Status at SUNGO |
|---|---|---|
| IEC 60601-2-x particular standards (e.g. IEC 60601-2-24, IEC 60601-2-27), ISO 80601-2-x other than ISO 80601-2-69, and the corresponding GB 9706.2xx and YY 9706.2xx | Particular requirements for basic safety and essential performance of specific equipment types; e.g. IEC 60601-2-24 covers infusion pumps and controllers, IEC 60601-2-27 covers electrocardiographic monitoring equipment | Not in scope; only ISO 80601-2-69 and its EN and YY numbers are on the schedule |
| IEC 60601-1-6 | Collateral standard: usability | Not in scope |
| IEC 62366-1 | Application of usability engineering to medical devices | Not in scope |
| IEC 62304 | Medical device software: software life cycle processes | Not in scope |
| IEC 61010-1 | Safety requirements for electrical equipment for measurement, control and laboratory use | Not in scope |
| IEC 60335 series | Safety of household and similar electrical appliances | Not in scope |
| IEC 62133 series (e.g. IEC 62133-2) | Safety requirements for portable sealed secondary cells and batteries made from them | Not in scope |
| CISPR 14-1, CISPR 14-2 | EMC for household appliances, electric tools and similar apparatus (emission, immunity) | Not in scope |
| CISPR 25 | Radio disturbance characteristics of vehicles, boats and internal combustion engines, for the protection of on-board receivers | Not in scope |
| CISPR 32, CISPR 35 | EMC of multimedia equipment: emission requirements, immunity requirements | Not in scope |
| IEC 61326-1 | EMC requirements for electrical equipment for measurement, control and laboratory use: general requirements | Not in scope; the schedule has only IEC 61326-2-6 for IVD equipment |
| FCC rules, 47 CFR Part 15 and Part 18 | US Code of Federal Regulations, Title 47: Part 15 (radio frequency devices) and Part 18 (industrial, scientific and medical equipment) | Not in scope |
| Biological tests (e.g. cytotoxicity, sensitization, irritation) | Test items in the biological evaluation of medical devices | Not performed in-house; when needed, subcontracted to a suitably qualified laboratory with the client's prior consent, and the subcontracted items and subcontractor are named in the report |
Where IEC 60601 testing is done at SUNGO and which accreditation applies
Location: medical electrical safety and medical EMC testing is carried out at Hefei SUNGO Medical Technology Co., Ltd., in Zipeng Town, Feixi County, Hefei, Anhui Province. Hefei SUNGO and Shanghai SUNGO Medical Technology Co., Ltd. are two separate legal entities; SUNGO Shanghai holds no accreditation for medical electrical safety or EMC, so these two lines are available only in Hefei. Service details are on the electrical safety testing and EMC testing pages.
Accreditation: IAS TL-1119 is the only accreditation for these two lines. The CNAS and CMA certificates held by SUNGO Hefei do not cover them, so these reports carry no CNAS or CMA mark. CMA designation covers only the items on its schedule, and the certificate states that it is valid within the People's Republic of China. If a Chinese registration or a buyer requires a CMA report, these two lines need a different body with that qualification.
Mutual recognition: IAS is a signatory to the ILAC MRA. According to IAS, the work of ILAC and IAF was brought together in the Global Accreditation Cooperation Incorporated (Global ACI), which began full operations on 1 January 2026, and IAS has been a signatory member since its launch. During the transition, accreditations issued under the ILAC MRA continue to be recognized and the ILAC MRA mark remains valid. Whether a given regulator or buyer accepts an IAS-accredited report is for that party to decide.
Reading the report: check two things. First, whether the standard numbers, editions and test items in the report match your registration or purchasing requirements. Second, which accreditation body and which certificate the accreditation mark on the report refers to. For a method to check a scope item by item, see How to Check an Accreditation Scope and Read the Mark on a Report. For ISO 80601-2-69, the only particular standard on the schedule, see Oxygen Concentrator Testing: ISO 80601-2-69, IEC 60601-1 and IEC 60601-1-2.
Preparing an IEC 60601 sample for submission
These are general points; the laboratory's confirmed list governs what is actually needed. Complete information reduces stops after testing has started.
- Intended use and intended use environment: state whether the product is for professional healthcare facilities, the home or special environments. This drives the IEC 60601-1-2 immunity test levels and decides whether IEC 60601-1-11 applies.
- List of applicable standards: general, collateral and particular standards, with the numbering (IEC, EN or GB/YY) and edition your target market needs, checked item by item against the schedule with the laboratory.
- Sample information: model, configuration, software version, operating modes to be tested and accessories supplied.
- Electrical documents: circuit diagrams, power supply and interface descriptions, and a list of critical components, which help define safety test points and EMC test ports.
- Draft labels and instructions for use: both the safety and the EMC assessment look at equipment marking and accompanying documents.
- Alarm information: for equipment with an alarm system, the alarm conditions, priorities and settings, so that IEC 60601-1-8 can be assessed.
- Battery information: for battery-powered equipment or equipment with a charger, the battery type and power arrangement; IEC 62133 testing of the battery itself is outside SUNGO's scope and needs separate arrangements.
Frequently asked questions
Where in China can IEC 60601-1 testing be done with an IAS-accredited report?
SUNGO Hefei can issue IAS-accredited reports for IEC 60601-1, IEC 60601-1-2 and the other standards on its TL-1119 schedule, which lists whole standards without years. For other laboratories, check the scope schedule published by their accreditation body.
Can an IEC 60601-1 test report carry a CMA mark?
Not at SUNGO. Medical electrical safety and EMC are accredited only by IAS, and SUNGO Hefei's CNAS and CMA certificates do not cover them. If you need a CMA report, use a body that holds that qualification.
If a particular standard applies, is IEC 60601-1 testing alone enough?
The IEC abstract for IEC 60601-1 states that where particular standards exist, the general standard should not be used alone. The only particular standard in SUNGO's scope is ISO 80601-2-69; for any other, you need a laboratory whose scope covers it.
The IAS scope has no years. Which edition will the report use?
The schedule accredits whole standards without stating years. Before an IAS report is issued against a particular edition, confirm the edition with the laboratory and check which edition your target market requires.
Can usability (IEC 62366-1) and software (IEC 62304) be done at SUNGO Hefei?
No. Neither is in SUNGO's accreditation scope. They set the manufacturer's usability engineering process and software life cycle processes, so ask a body with the relevant capability.
To confirm which standards apply to a specific product and whether an IAS-accredited report can be issued from Hefei, send your product information and target market through the contact page. The laboratory will check each item against the IAS schedule and reply.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.