IEC 61326-2-6 EMC for IVD Equipment: When It Applies Instead of IEC 60601-1-2

IEC 61326-2-6 EMC for IVD Equipment: When It Applies Instead of IEC 60601-1-2

For in vitro diagnostic (IVD) instruments such as chemistry analyzers, immunoassay analyzers and point-of-care testing (POCT) devices, the opening EMC question is which standard applies: IEC 60601-1-2 or IEC 61326-1? Chinese registration files add GB/T 18268.1 and GB/T 18268.26. This article explains why IVD instruments normally fall under IEC 61326-2-6, how Part 2-6 works with Part 1, what Part 2-6 adds, where GB/T 18268.26 stands in China, and which SUNGO laboratory tests to it under which accreditation.

Key points

  • Per Clause 1.1 of IEC 60601-1, the IEC 60601 series does not apply to IVD equipment outside the definition of medical electrical (ME) equipment; the IEC 61010 series covers it. IVD EMC therefore sits in the IEC 61326 family: IEC 61326-1:2020 covers IVD medical equipment, and IEC 61326-2-6 adds the IVD particular requirements. A product that is ME equipment goes back to IEC 60601-1-2.
  • IEC 61326-2-6 is used together with IEC 61326-1:2020. It follows the Part 1 numbering and only states additions, modifications and replacements; Part 1 subclauses it does not mention apply as far as is reasonable.
  • What Part 2-6 adds, taking Edition 3 as the reference: the most severe power option per operating mode, chosen by risk analysis; immunity tables split by professional healthcare facility and home healthcare environments; and pass/fail criteria grounded in ISO 14971, focusing on incorrect results and direct hazards to the operator. For emissions, IEC 61326-1:2020 classifies equipment as Class A or B under CISPR 11.
  • The current IEC edition is Edition 4.0 (2025-06-18). China's GB/T 18268.26-2010 (identical to IEC 61326-2-6:2005) is still current; a revision adopting IEC 61326-2-6:2025 with modifications was "being approved" on 2026-10-22.
  • IEC 61326-2-6, EN IEC 61326-2-6 and GB/T 18268.26 are within the IAS accreditation (TL-1119) of the Hefei laboratory, listed as complete standards with no edition year. A stand-alone report to IEC 61326-1 or GB/T 18268.1 is not within SUNGO's accreditation.

Standards and scope

Why IVD instruments usually do not follow IEC 60601-1-2

Under its Clause 1.1, IEC 60601-1-2 applies to the basic safety and essential performance of ME equipment and ME systems in the presence of electromagnetic disturbances, and to the disturbances they emit.

Clause 1.1 of IEC 60601-1 (Edition 3.2) says the IEC 60601 series does not apply to IVD equipment outside the ME equipment definition. A note to entry 3.101 of IEC 61326-2-6:2020 explains that IVD medical equipment examines specimens taken from the human body, without direct or wired patient connection. Clause 1 of IEC 61326-1:2020 covers "electrical laboratory equipment, including in vitro diagnostic (IVD) medical equipment", and Edition 3 of IEC 61326-2-6 (Clause 1) adds minimum immunity and emission requirements reflecting IVD equipment and its electromagnetic environment; Edition 4 rewords the scope (see Editions below).

So start by deciding whether the product is ME equipment (defined in subclause 3.63 of IEC 60601-1). An instrument that only handles specimens and has no patient connection normally goes to IEC 61326-1 plus IEC 61326-2-6; a product with a patient-connected part needs reassessing against the ME definition.

The FDA guidance "Electromagnetic Compatibility (EMC) of Medical Devices" (issued 2022-06-06) says most laboratory equipment and IVDs are outside the scope of IEC 60601-1-2. FDA then partially recognized IEC 61326-1:2020 and IEC 61326-2-6:2020 and recommended their test methods, but with acceptance criteria specific to the device's functions and intended use, and test levels from IEC 60601-1-2 or the reasonably foreseeable maximum levels in the intended use environments.

How Part 2-6 relates to Part 1

The forewords of Editions 3 and 4 both state that Part 2-6 is used in conjunction with IEC 61326-1:2020 and follows its numbering. Where Part 2-6 says "addition", "modification" or "replacement", the Part 1 text is adapted, and items numbered from 101 are additional. The Edition 4 foreword adds that if an IEC 61326-2-6 report is available, the IEC 61326-1 report is integrated.

IEC 61326-1:2020 sets immunity requirements by three electromagnetic environments (basic, industrial and controlled). Part 2-6 instead splits immunity requirements between the professional healthcare facility environment and the home healthcare environment, described below.

Editions: IEC Edition 3, Edition 4 and the European/UK versions

  • IEC 61326-2-6:2020 (Edition 3.0): published 2020-10-28 by IEC SC 65A, replacing the 2012 second edition. Its listed technical change is an update to IEC 61326-1:2020; the IEC Webstore gives a withdrawal date of 2025-06-18.
  • IEC 61326-2-6:2025 (Edition 4.0): published 2025-06-18, replacing Edition 3. The listed technical change is an update of test levels and documentation. The title now reads "IVD medical electrical equipment", the scope is worded around basic safety and essential performance in terms similar to IEC 60601-1-2, and IEC 61010 (all parts) joins the normative references.
  • European/UK versions: CENELEC approved EN IEC 61326-2-6:2025 on 2025-07-23, endorsing IEC 61326-2-6:2025 without modification and superseding EN IEC 61326-2-6:2021. BSI published BS EN IEC 61326-2-6:2025 on 2025-08-31; its national foreword says the superseded BS EN IEC 61326-2-6:2021 will be withdrawn on 2028-08-31.
  • FDA recognition: Edition 4.0 is recognition number 19-57, entered on 2025-07-28 and recognized as a complete standard. FDA will accept declarations of conformity to Edition 3.0 (recognition number 19-43) until 2028-07-02.

China: GB/T 18268.26 and GB/T 18268.1

According to China's national standards platform (std.samr.gov.cn), GB/T 18268.26-2010 (Part 26: particular requirements for IVD medical equipment) is a recommended national standard, published 2011-01-14, implemented 2011-05-01, identical to IEC 61326-2-6:2005 and still current; a 2025-07-01 review concluded it should be revised. Revision plan 20252556-T-604 modifies and adopts IEC 61326-2-6:2025 and is planned to take effect 24 months after publication; on 2026-10-22 it was listed as "being approved". GB/T 18268.1-2025 (Part 1), published 2025-06-30 and implemented 2026-01-01, is identical to IEC 61326-1:2020 and replaces the 2010 edition.

So as of 2026-10-22, the two Chinese companion standards rest on different IEC editions: Part 1 on IEC 61326-1:2020, Part 26 on IEC 61326-2-6:2005. Which editions the product technical requirements cite is for the reviewers to accept; state them on the test order.

Registration testing in China

Two guidance documents issued with NMPA Notice No. 80 of 2020 address EMC for IVD instruments directly:

  • The guideline for specific protein immunoassay analyzers says EMC indicators should include all indicators of GB/T 18268.1, GB/T 18268.26 and other applicable standards, and that EMC shall meet the Group I Class A requirements of GB/T 18268.1 and GB/T 18268.26.
  • The guideline for home-use IVD medical devices says the instructions for an IVD instrument should state, under warnings and precautions, its EMC (emission and immunity) and its classification as Group 1 Class B under GB 4824.

Under the Provisions for Medical Device Registration and Filing (SAMR Order No. 47), Article 30 requires testing against the product technical requirements, and Article 32 accepts either the applicant's self-test report or a report from a qualified medical device testing institution it engages. Whether a third-party report can be used in a given application depends on the route and the reviewers; confirm before ordering. SUNGO makes no commitment on acceptance.

What IEC 61326-2-6 requires

Test plan and EUT configuration

Subclause 5.1 of IEC 61326-1:2020 requires an EMC test plan before testing, covering at least subclauses 5.2 to 5.5: EUT configuration, operating conditions, functional performance and test description. Clause 4 allows tests to be split among several EUTs if each result is traceable to the EUT tested.

Edition 3 of Part 2-6 adds subclause 5.3.101: the device is set as the user documentation specifies for the intended use, and for each mode of operation the power option (battery, AC or DC supply) representing the most severe condition is specified from the product risk analysis. Subclause 6.101 requires the configuration and modes to be precisely noted in the report, tests to be run one at a time per the basic standards, and any additional method and rationale to be documented.

Edition 4 adds subclause 5.2.101 on subsystems, informative Annex D (test plan guidance, with Table D.1 on recommended minimum contents) and Annex E (power input voltages and frequencies). Note 2 to its scope says IT equipment can be part of IVD medical electrical equipment if needed to maintain basic safety or essential performance, so for a system with a main unit, sample modules and an external PC, settle in the test plan which parts form the EUT.

Immunity: choose the table by use environment

  • Professional healthcare facility environment (Edition 3, entry 3.102): where professional healthcare is administered. The note lists hospitals, diagnostic laboratories, blood banks, clinics and patient rooms, among others; equipment that might be used in ambulances, other vehicles or aircraft can need higher immunity.
  • Home healthcare environment (entry 3.103): any environment other than a professional healthcare facility, with less controlled disturbances. Locations include the home and public places such as shops, offices, transport stations and airports.

The selection rule is in the replaced subclause 6.2 of Edition 3 (in Edition 4 the risk-assessment subclause is 6.201). Professional-environment equipment follows Table 101, with Tables 102 and 103 to be considered depending on the risk assessment of the electromagnetic environment; that assessment must cover point-of-care and other equipment used close to patients, who may carry mobile phones. Equipment for the home healthcare environment follows Tables 102 and 103. Table 103 covers RF wireless communications equipment; its frequencies are representative examples, so risk management should consider country-specific services and add missing frequencies.

In Edition 3, Tables 101 and 102 are organized by port (enclosure, AC power, DC power, I/O signal/control) and cover ESD, radiated RF, power-frequency magnetic fields, fast transients/bursts, surges, conducted RF, and voltage dips and interruptions under IEC 61000-4-2, -4-3, -4-4, -4-5, -4-6, -4-8 and -4-11, each with an assigned performance criterion. Some port tests apply only to lines longer than 3 m or to long-distance lines (defined in subclause 3.1.9 of IEC 61326-1:2020). Take the test levels from the standard itself; this article does not quote them. When testing to Edition 4, take Tables 101 and 102 from the Edition 4 text.

Edition 4 makes Table 103 a specification for enclosure-port immunity to RF wireless communications equipment and adds Table 104 for proximity magnetic fields, with informative Annex G on applying it. If testing to Edition 4, plan for proximity magnetic fields early.

Pass/fail criteria: start from risk

Subclause 6.4 of Part 1 defines performance criteria A, B and C, and Tables 101 and 102 of Part 2-6 Edition 3 assign one of them to each test; under Edition 4, the pass/fail criteria follow its subclause 6.401. The replaced subclause 6.2 in Edition 3 is titled "Risk assessment and consideration of EMC immunity requirements". Its main points:

  • IVD equipment does not keep patients alive or resuscitate them, so a malfunction would not directly cause death or serious injury; but an incorrect reading can lead to a wrong therapeutic decision (misdiagnosis), and for some analytes an incorrect result could seriously harm the patient.
  • Larger IVD equipment can also malfunction in ways that directly threaten the operator, such as unexpected mechanical movements.
  • The manufacturer shall perform risk management according to ISO 14971.

Edition 3 defines basic safety (entry 3.105, freedom from unacceptable risk to the operator directly caused by physical hazards) and essential performance (entry 3.106, performance of a clinical function whose loss or degradation beyond the limits in the user documentation results in unacceptable risk). Edition 4 adds subclauses 4.101 (essential performance), 4.102 (basic safety) and 6.401 (pass/fail criteria), plus informative Annex F on risk management and immunity pass/fail criteria, with examples for an IVD immunoassay analyzer.

In the test plan, write criteria that can be observed and decided: how much deviation in a result is acceptable; whether aborting a test, alarming and withholding the result is an acceptable response; and what in sample handling or mechanical motion counts as failure.

Emissions: CISPR 11 group and class

Clause 1 of IEC 61326-1:2020 requires equipment to be classified as Class A or Class B under the requirements and procedure of CISPR 11, with emission requirements in Clause 7. Under entries 3.1.2 and 3.1.3 (derived from CISPR 11:2015, 5.2), Class B equipment suits residential environments and establishments directly connected to a low-voltage network supplying domestic buildings; Class A suits all other locations. For what Clause 7 of Part 2-6 itself specifies, refer to the standard text.

The Chinese guidance gives two examples: Group I Class A under GB/T 18268.1 and GB/T 18268.26 for specific protein immunoassay analyzers, and Group 1 Class B under GB 4824 for home-use IVD instruments. For how the classification relates to GB 4824, see EMC Emission Reports: GB 4824 Sits under a Higher Layer. Edition 4 also adds requirements on instructions for use (subclauses 9.101 to 9.103); the EMC information in the instructions should match the test report.

SUNGO accreditation scope: Hefei IAS (TL-1119)

The table reproduces the directly relevant rows of the IAS scope of Hefei SUNGO Medical Technology Co., Ltd. (TL-1119, effective 2026-06-04). Item titles follow the certificate ("Electrical equipment for measurement, control and laboratory use" omitted); the scope lists complete standards with no edition year, clauses or limitations.

ItemStandard and clauseNotes / limitations
EMC requirements – Part 2-6: Particular requirements – In vitro diagnostic (IVD) medical electrical equipmentIEC 61326-2-6, complete standardNo limitation stated; no edition year (confirm edition before ordering)
EMC requirements – Part 2-6: Particular requirements – In vitro diagnostic (IVD) medical electrical equipment (European version)EN IEC 61326-2-6, complete standardNo limitation stated; no edition year (confirm edition before ordering)
EMC requirements – Part 26: Particular requirements – In vitro diagnostic (IVD) medical equipmentGB/T 18268.26, complete standardNo limitation stated; no edition year (confirm edition before ordering)

The basic test methods used in Tables 101 and 102, IEC 61000-4-2, -4-3, -4-4, -4-5, -4-6, -4-8 and -4-11, together with GB/T 17626.2, .3, .4, .5, .6, .8 and .11 and, for emissions, CISPR 11 and GB 4824, are also listed separately in the Hefei IAS scope as complete standards.

Not within SUNGO's accreditation: a stand-alone report to IEC 61326-1 or GB/T 18268.1, and electrical safety standards for IVD instruments such as the IEC 61010 series and the GB 4793.1 and YY 0648 standards cited in the Chinese guidance.

Where SUNGO tests and under which accreditation

EMC testing of IVD instruments is carried out by the Hefei laboratory, within its IAS accreditation (TL-1119). Reports are issued by Hefei SUNGO Medical Technology Co., Ltd. (Zipeng Town, Feixi County, Hefei, Anhui), a legal entity separate from Shanghai SUNGO Medical Technology Co., Ltd. The Shanghai laboratory's scope (Kangqiao East Road 958, Pudong) does not include EMC.

  • Reports available: reports within the IAS scope to IEC 61326-2-6, EN IEC 61326-2-6 or GB/T 18268.26 (the particular requirements are applied together with Part 1). Whether such a report also satisfies a citation of a given edition of GB/T 18268.1 in the product technical requirements is for the reviewers to decide; confirm before ordering.
  • Not within accreditation: a stand-alone report to IEC 61326-1 or GB/T 18268.1.
  • Edition: the IAS scope carries no edition year; confirm before ordering whether the report will follow the 2020 or 2025 edition of IEC 61326-2-6, or which edition of GB/T 18268.26.
  • Acceptance: whether a report is accepted for a US, EU or Chinese submission is decided by the relevant regulator or notified body; please confirm in advance.

Service overview: EMC testing.

Preparing for testing

  1. Intended use environment: professional facility, home (self-testing) or point of care, with reasoning; this decides the immunity tables and the emission class.
  2. Editions: Edition 3 or 4 of IEC 61326-2-6, which edition of GB/T 18268.26, and the matching Part 1 edition.
  3. Draft test plan: per subclauses 5.2 to 5.5 of IEC 61326-1, the EUT composition, auxiliary equipment, each port's cable and maximum length, earthing, operating modes and software version, plus the most severe power option for each mode and its risk-analysis basis.
  4. Criteria and risk documents: identified essential performance and basic safety, EMC-related parts of the risk management file, and pass/fail criteria per test.
  5. Consumables: controls, calibrators, reagents and consumables used to judge results during testing, and arrangements for any biological material, agreed in advance.
  6. Other information: radio module bands, rating plate data and draft EMC information for the instructions for use.

FAQ

Can an IVD instrument simply be tested to IEC 60601-1-2 or YY 9706.102?

Check whether it is ME equipment. If not, it falls outside the IEC 60601 series and goes to IEC 61326-1 with IEC 61326-2-6; the Chinese guidance cites GB/T 18268.1 and GB/T 18268.26.

Is IEC 61326-1 alone enough, and can SUNGO issue it on its own?

IEC 61326-1 is the general part; the IVD-specific requirements are in Part 2-6, and the two are used together. SUNGO issues IAS reports to Part 2-6 or GB/T 18268.26; a stand-alone Part 1 report is not within its accreditation.

Should we test to the 2020 or the 2025 edition?

The current IEC edition is Edition 4 (2025), and FDA accepts declarations of conformity to Edition 3 until 2028-07-02. The choice depends on your target markets and submission timing; put it on the test order and confirm it with the laboratory.

What changes for home-use, self-testing IVD instruments?

Under subclause 6.2 of Edition 3, the home healthcare environment uses Tables 102 and 103; Edition 4 also has Table 104 (proximity magnetic fields), so check the standard text for whether it applies. The Chinese home-use IVD guidance asks for Group 1 Class B under GB 4824 in the instructions, and subclause 9.102 of Edition 4 adds instructions-for-use requirements for home healthcare equipment.

What happens when the new GB/T 18268.26 is published?

The revision plan foresees implementation 24 months after publication. The edition to test to will then follow the registration requirements; confirm with the laboratory before ordering a report to the new edition.

Requesting a quote

Please send the product name and model, intended use environment, target markets and planned standard editions, product photos, a list of ports and cables, and any risk management and essential performance documents through our contact page. The Hefei laboratory will confirm the editions it can test to and the test arrangements in its reply.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.