ISO 16603 and YY/T 0700: Synthetic Blood Penetration of Protective Clothing

ISO 16603 and YY/T 0700: Synthetic Blood Penetration of Protective Clothing

Synthetic blood penetration of protective clothing and surgical gown materials is specified as YY/T 0700-2008 in Chinese documents and as ISO 16603:2004 by overseas buyers. Both describe one test: the material is clamped in a test cell with synthetic blood on one side, pressurized in a set sequence, and checked for visible penetration on the other side. This guide explains how the 0 to 20 kPa steps are applied, how results become a "highest pressure passed" or a "class", and how the method relates to ASTM F1670, then sets out what the SUNGO Shanghai laboratory is accredited to report.

Key points

  • YY/T 0700-2008 is an identical adoption of ISO 16603:2004. Its foreword lists editorial changes only. The test principle and procedures are the same, although a few YY/T 0700 clauses carry extra wording (see the list of differences below).
  • The method gives pass/fail results, not classes. Procedures A and B hold 14 kPa for 1 min. C and D start at 0 kPa, then hold 1.75, 3.5, 7, 14 and 20 kPa for 5 min each. The report gives each specimen's result at each step.
  • Classes come from product standards. GB 19082-2023 Table 4 (classes of resistance to synthetic blood and bacteriophage penetration) maps the six pressures from 0 to 20 kPa to classes 1 to 6. Both synthetic blood (clause 4.4.3, at least class 2) and bacteriophage (clause 4.5.1, at least class 1) are graded against it. EN 14126:2003 classifies only on the bacteriophage result and uses the synthetic blood test as a screen.
  • Same origin as ASTM F1670, not the same test. ISO 16603 is based on ASTM F1670-98. Its Procedures A and B use 14.0 kPa against 13.8 kPa (2.0 psig) in ASTM F1670/F1670M, and it adds stepped Procedures C and D up to 20 kPa.
  • SUNGO Shanghai accreditation. Both standards are in the Shanghai CNAS schedule (test object "protective clothing") and CMA schedule (test object "surgical gowns"), item "synthetic blood penetration". They are not in the IAS scope, so no IAS report is possible. We issue test-method reports only: no conformity conclusions against GB 19082, EN 14126 or other product standards, and no complete test programmes for whole garments under those product standards.

The standards and what they cover

ISO 16603:2004

Full title: Clothing for protection against contact with blood and body fluids — Determination of the resistance of protective clothing materials to penetration by blood and body fluids — Test method using synthetic blood. It is the first edition, published in April 2004 by ISO/TC 94/SC 13 (Protective clothing). As of 2026-10-22, the ISO website lists it as confirmed and states that it was last reviewed and confirmed in 2025 and remains current.

Clause 1 describes a laboratory method in which synthetic blood stays in continuous contact with the material under specified conditions in the ISO 13994 apparatus. It warns that the method is not always effective for materials with thick inner liners that readily absorb synthetic blood.

The introduction sets the limits. The method assesses materials and certain constructions such as seams, not garment design or interfaces, and does not necessarily simulate real use, so the data suit broad comparison between materials. It suggests considering shelf life, laundering, sterilization, flexing, abrasion and pre-wetting where relevant.

How YY/T 0700-2008 maps to ISO 16603

The cover of YY/T 0700-2008 reads "ISO 16603:2004, IDT". It was issued by the then State Food and Drug Administration on 2008-10-17 and took effect on 2010-01-01. As of 2026-10-22, China's national standards information platform lists it as a current, recommended pharmaceutical industry standard. Its foreword lists three editorial changes: "this International Standard" becomes "this standard", a decimal point replaces the decimal comma, and the ISO foreword is deleted.

Comparing the full YY/T 0700 text with the public preview of ISO 16603 shows a few more differences:

  • References. Thickness and mass per unit area are measured to GB/T 3820 and GB/T 4669, and clause 1 calls for the apparatus of YY/T 0699 (corresponding to ISO 13994). Clause 2 of YY/T 0700 also adds GB/T 5549, used in Annex A to measure the surface tension of the synthetic blood; clause 2 of ISO 16603 has no corresponding entry.
  • Two additions. Clause 6.4 specifies a timer accurate to at least 1 s, where ISO 16603 clause 6.4 says only "stopwatch, or electronic timer". Clause 7.1.1 adds a sentence after the three-specimen rule: if a false positive result is found, another specimen is taken and tested. ISO 16603 has no such sentence at that point.
  • Foreword and annexes. With the foreword removed, the ISO statement that the standard is based on ASTM F1670-98 does not appear in YY/T 0700, although its bibliography still lists ASTM F1670-98. ISO 16603 has informative Annexes A (synthetic blood formula) and B (sources of apparatus); YY/T 0700 contains only Annex A.

So the choice between a YY/T 0700 and an ISO 16603 report depends on where the report will be used and which number the receiving document cites. The procedures are essentially the same; for details, the text of the cited standard governs.

Product standards that call it up

The method usually appears as one item inside a product standard. The two examples below are background only. Neither product standard is within SUNGO's accreditation.

  • GB 19082-2023, Disposable protective clothing for medical use. A mandatory national standard, published on 2023-11-27 and in force from 2025-12-01, replacing GB 19082-2009. It cites YY/T 0700-2008 normatively. Clause 4.4.3 requires at least class 2 in Table 4. Clause 5.5.3 specifies sampling locations (including seams), testing by YY/T 0700-2008 with synthetic blood prepared as in YY 0469, and requires all specimens to meet 4.4.3. Table 4 also grades bacteriophage penetration (clause 4.5.1), tested by Procedure C or D of YY/T 0689-2008 (clause 5.6.1).
  • EN 14126:2003, Protective clothing against infective agents. It covers reusable and limited-use protective clothing; surgical team clothing and drapes are outside its scope. Clause 4.1.4.1 classifies materials tested to ISO 16603 and ISO 16604 by the bacteriophage (ISO 16604) result, and its note says the synthetic blood test is for screening, to predict where strikethrough can be expected in the bacteriophage test. It cites ISO 16603 at FDIS stage.

Both use the same six pressures from 0 to 20 kPa. GB 19082-2023 applies Table 4 to both synthetic blood and bacteriophage, so the synthetic blood result is itself graded; EN 14126:2003 classifies only on the bacteriophage result. Whether other tests are needed, and how a class is assigned, depends on the target product standard.

How the test is run

Principle and synthetic blood

Clause 4: synthetic blood, as a body fluid simulant, is applied to the material for a specified time and pressure sequence, with the material separating it from the viewing side of the cell. Any evidence of penetration means failure, and results are reported as pass/fail.

Clause 5 specifies the synthetic blood: surface tension (0.042 ± 0.002) N/m, pH 7.3 ± 0.1, viscosity (2.7 ± 0.3) mPa·s and conductivity (12.0 ± 1.2) mS/cm. Annex A gives one way to prepare it. The introduction explains that blood and body fluids (excluding saliva) have surface tensions of about 0.042 to 0.060 N/m. The synthetic blood is set near the lower end of that range to simulate their wetting behaviour.

Specimen selection and conditioning

  • Representative construction. A specimen may be a single layer or a composite of several layers in the same order as the actual garment (7.1.1).
  • Locations and seams. Different materials or thicknesses by location are sampled separately; if seams are claimed to protect as well as the base material, seam specimens are added (7.1.1).
  • Size and number. Squares at least 70 mm on a side, 75 mm preferred. Three specimens are taken at random from each material, composite, area (for heterogeneous designs) or other condition; YY/T 0700 adds that a false positive calls for a retest on a new specimen. For quality control or broad claims, a statistical design over more data is recommended, for example ISO 2859-1 (GB/T 2828.1).
  • Edge sealing for sandwich constructions. Materials with an impervious layer between two fabric layers can give false failures through wicking at the edges. The specimen edges are sealed, leaving a central 57 mm square for testing (7.1.2).
  • Conditioning. Each specimen is conditioned for at least 24 h at (21 ± 5) °C and (60 ± 10) % relative humidity. Where allowed, other preconditioning such as sterilization may be used to assess possible damage to protective performance (7.2).

Preliminary measurements and mounting

Thickness is measured to the nearest 0.02 mm and mass per unit area to the nearest 10 g/m². A 10 µL drop of synthetic blood on the inside surface confirms that penetration would be visible; talc or indirect lighting may help (8.1). The outside surface faces the synthetic blood. Extensible materials are supported by a retaining screen (8.2.1) with an open area above 50 % and deflection of 5 mm or less (6.2).

The four procedures and the pressure steps

Clause 8.3.1 asks the user to select an appropriate procedure from Table 1. The table below follows Table 1 of YY/T 0700-2008, an identical adoption of ISO 16603:

ProcedurePressure and time sequenceRetaining screenMain points of the standard's remarks
A0 kPa for 5 min, then 14 kPa for 1 min, then 0 kPa for 4 minNoSelecting materials for critical areas with large amounts of blood or body fluid, direct contact, pressing or leaning
BAs Procedure AYesWhen failure in Procedure A is suspected to be caused by deformation of the material
C0 kPa for 5 min, then 1.75, 3.5, 7, 14 and 20 kPa for 5 min eachNoWhere blood or body fluids may be present at different contact pressures; the protection level is chosen after analysing the work and likely contact
DAs Procedure CYesWhen failure in Procedure C is suspected to be caused by deformation of the material

Clause 8.3 in brief: slowly fill the cell with 60 mL of synthetic blood, stopping if liquid penetrates during filling; pressurize at no more than 3.5 kPa/s and hold each level for the set time; at the end of each interval, check the viewing side for synthetic blood or wetting. On penetration, stop and record the time and pressure. A specimen with no liquid or characteristic discoloration throughout passes.

The C and D pressure steps line up one-to-one with the pressures in GB 19082-2023 Table 4. The procedure actually used should follow the product standard, customer specification or risk assessment, and should be stated on the test request.

Reporting the "highest pressure passed"

Clause 9 requires, among other items, material and sampling details (including any sterilization), thickness and mass per unit area of each specimen, the procedure and retaining screen, pass or fail for each specimen at each time and pressure interval, and, where required, the time to failure.

With Procedure C or D, the highest pressure step each specimen passed without visible penetration can therefore be read directly from the report. For example, if penetration is seen at the end of the 7 kPa interval, the highest pressure passed is 3.5 kPa. The note to Table 1 also says that the highest pressure with no visible penetration from Procedure C or D can be used in YY/T 0689. YY/T 0689-2008 corresponds to ISO 16604:2004, the bacteriophage penetration method.

Two cautions. The method reports specimen by specimen; combining specimens into one product conclusion is a product-standard rule (GB 19082-2023 clause 5.5.3 requires every specimen to meet 4.4.3). And a "class" is likewise decided under the product standard, not by the method report.

How it compares with ASTM F1670

ASTM F1670/F1670M is the US counterpart. The SUNGO Shanghai CNAS schedule lists the F1670/F1670M-24 edition; as of 2026-10-22, the ASTM website shows -24 as "Historical" and F1670/F1670M-25 as the active edition. ASTM's public summary, worded the same in both editions, says the method is based on ASTM F903, gives a visual pass/fail result, and is used to select materials for further viral barrier testing. Its Procedures A and B pressurize to 13.8 kPa (2.0 psig), against 14.0 kPa in ISO 16603. For synthetic blood surface tension, F1670/F1670M clause 5.2 gives about (0.040 ± 0.005) N/m, while ISO 16603 clause 5 requires (0.042 ± 0.002) N/m. ISO 16603 also has stepped Procedures C and D. Whether one report can stand in for the other depends on which standard the target document cites.

SUNGO accreditation scope

The rows below are copied from the accreditation schedules. The schedules give no clause and no limitation for these rows.

CertificateTest objectItemStandard and clauseNotes / limitations
CNAS L22083Protective clothingSynthetic blood penetrationYY/T 0700-2008 (no clause given)No limitation stated
CNAS L22083Protective clothingSynthetic blood penetrationISO 16603:2004 (no clause given)No limitation stated
CMA 240015340087Surgical gownsSynthetic blood penetrationYY/T 0700-2008 (no clause given)No limitation stated; CMA is valid only within China
CMA 240015340087Surgical gownsSynthetic blood penetrationISO 16603:2004 (no clause given)No limitation stated; CMA is valid only within China
CNAS L22083 (for comparison)Surgical gownsSynthetic blood penetration testASTM F1670/F1670M-24 (no clause given)No limitation stated; the schedule lists the -24 edition; not in the CMA schedule
IAS TL-1019 (for comparison)Not listedWhole standardASTM F1670/F1670MThe IAS scope gives no edition year; confirm the edition with the laboratory before ordering an IAS report to a specific edition

The Shanghai IAS scope (TL-1019) does not list ISO 16603 or YY/T 0700, and none of the three Hefei certificates covers this test. GB 19082, EN 14126 and ISO 16604 are not on any SUNGO certificate.

Where SUNGO performs it and under which accreditation

The test is performed by the laboratory of Shanghai SUNGO Medical Technology Co., Ltd., 958 Kangqiao East Road, Pudong New Area, Shanghai. Available report marks:

  • CNAS (L22083): both standards, test object "protective clothing". The CNAS certificate states that CNAS is a signatory to the ILAC and APAC mutual recognition arrangements.
  • CMA (240015340087): both standards, test object "surgical gowns". CMA covers only scheduled items and is valid within the People's Republic of China.
  • IAS (TL-1019): neither standard, so no IAS report. ASTM F1670/F1670M is in the IAS scope but is a different method.

Because the two schedules name the test object differently, please confirm before ordering whether your sample can carry both marks. Hefei SUNGO Medical Technology Co., Ltd. is a separate legal entity whose scope does not include this test. Whether a regulator or customer accepts a report is their decision.

Preparing a submission

  1. Standard and report mark. YY/T 0700-2008 or ISO 16603:2004; CNAS or CMA; report language.
  2. Procedure. A, B, C or D, with or without retaining screen. If unsure, tell us the intended use and target document.
  3. Construction and orientation. Single layer or composite, layer order as in the product, and the normal outside and inside surfaces marked.
  4. Sample source and locations. Roll goods or garments; flag location differences and seam claims. The laboratory will tell you how much material to send.
  5. Material information. Supplier, lot, and whether and how the material was sterilized. These go into the report.
  6. Preconditioning. If you want to assess the effect of sterilization, ageing, laundering or similar treatment, complete the treatment before shipping and state the conditions on the test request. Preconditioning is not part of the accredited synthetic blood penetration item; if you need the laboratory to arrange it, confirm separately.
  7. Special constructions. Tell us about sandwich constructions (edge sealing), thick absorbent liners (the method may not be effective) and dark inner layers that could hide synthetic blood.

FAQ

Will a YY/T 0700 test give a different result from an ISO 16603 test?

YY/T 0700-2008 adopts ISO 16603:2004 identically, so the principle, procedures and pass/fail rule are the same; a few YY/T 0700 clauses add wording, such as retesting after a false positive. Choose the number your target document cites; both are in the Shanghai CNAS and CMA schedules.

Will the report state "class 5" or "class 6"?

The method report gives pass or fail for each specimen at each pressure step. Assigning a class is a product-standard decision, and SUNGO does not issue conformity conclusions against product standards such as GB 19082 or EN 14126.

Should I choose Procedure A or Procedure C?

A and B check a single pressure of 14 kPa; C and D step from 0 to 20 kPa and show where penetration occurs, so only they match a 0 to 20 kPa class table. The product standard or customer specification has the final say.

Can you issue an IAS report?

No. The Shanghai IAS scope (TL-1019) does not include ISO 16603 or YY/T 0700. ASTM F1670/F1670M is in the IAS scope, but its pressure and procedures differ, so it is not a substitute.

Can you test a whole protective garment to GB 19082 or EN 14126? What about isolation gowns?

No. GB 19082 and EN 14126 (including ISO 16604) are outside SUNGO's accreditation; the only rows with "protective clothing" as the test object are the two synthetic blood rows above, and no row names isolation gowns. Liquid barrier classification is a separate standard, ANSI/AAMI PB70: the Shanghai IAS scope lists it (whole standard, no edition year; confirm the edition with the laboratory), and the Shanghai CNAS schedule lists three AAMI PB70:2022 items for surgical gowns (synthetic blood penetration test, hydrostatic pressure, water penetration resistance). Ask us separately about a given sample.

Requesting a quote

For synthetic blood penetration testing to YY/T 0700-2008 or ISO 16603:2004, start with our surgical gown and drape testing service. To request a quote, use the contact page to send the material construction, sampling locations, intended procedure, any preconditioning already done, and the report mark you need. We will reply item by item against the standard clauses and our accreditation scope.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.

Need testing? Tell us your product and standards

Contact us directly
Phone
+86 132 4819 8029 (Mr. Gan)+86 177 5560 4650 (Mr. Luo)
E-mail
haixiao.gan@sungoglobal.comjun.luo@sungoglobal.com
Mon–Fri 9:00-18:00
WeChat QR CodeWeChatScan to add us on WeChat
Send us a message
About your personal information
  • Who collects it: Shanghai SUNGO Medical Technology Co., Ltd., the operator of this website. If the work you ask about is carried out by Hefei SUNGO Medical Technology Co., Ltd., we pass your enquiry to that company’s contact person for the same purpose.
  • What we collect: only what you type into the form above. The website’s access logs routinely record IP address and browser type, used only for security, troubleshooting and traffic statistics. Please do not enter ID numbers, medical records or other sensitive information.
  • Why: to reply to this testing enquiry. It is not used for bulk marketing; except for passing it to Hefei SUNGO Medical Technology Co., Ltd. as described above, it is not given to anyone else.
  • How long: 12 months from the day you submit it, then it is deleted. It is stored on a server in mainland China.
  • Your rights: contact us by the phone numbers or e-mail addresses above at any time to see, correct or delete what you submitted.